You are on page 1of 7

METROLOGIA-2003 – Metrologia para a Vida

Sociedade Brasileira de Metrologia (SBM)


Setembro 01−05, 2003, Recife, Pernambuco - BRASIL

FLASH METHOD STANDARDIZATION IN BRAZIL FOR THE


MEASURMENT OF SOLID THERMOPHYSICAL PROPERTIES
Paulo Couto1, Helcio R. B. Orlande1, Renato M. Cotta1, Sérgio P. Oliveira2,
Jorge A. Cruz2, Paulo R. F. Santos2, Cláudio S. Pinto Costa1 and Henrique Massard1.
1
Federal University of Rio de Janeiro - EE/COPPE, Rio de Janeiro, RJ, 21945-970, Brazil
2
National Institute of Metrology, Standardization and Industrial Quality – INMETRO, Xerém, Brazil

Abstract: The flash method, proposed by Parker, Butler, Additionally, the large size of the samples imposed intoler-
Jenkins, and Abbott from the U.S. Navy Radiological De- able limitations, usually tied to heat losses and contact resis-
fense Laboratory in 1960, is the most popular method of tance between the specimen, and its associated heat sources,
measuring the thermal diffusivity of solids. In this method heat sinks and measurement devices. The Flash Method,
the front surface of a small sample is subjected to a very first proposed by Parker et al. [2], eliminated the problem of
short burst of radiant thermal energy. The resulting tempera- contact resistance of the classical methods, and minimized
ture rise on the opposite surface of the sample is measured, the heat losses by making the measurement time short
and the thermal diffusivity is computed from the tempera- enough so that very little cooling can take place. This
ture rise vs. time data. Also, the heat capacity can be com- method consists of heating the front surface of a thermally
puted from the temperature vs. time data, for the calculation insulated specimen with a high-intensity short-duration
of the thermal conductivity. Several theoretical models are radiative heat pulse and measuring the temperature evolu-
available for the flash method, which include adiabatic tion on the back surface by the use of an infrared (IR) detec-
boundary conditions, heat losses, surface coating effects, tor. The non-intrusive backside measurement method elimi-
among many other aspects. In this paper, these models are nates the concern and issues with sensor attachment to the
briefly discussed. This paper also describes the implementa- sample, and removes all uncertainties associated with con-
tion of the flash method in Brazil as a standard method for tact resistance and sensor measurement accuracy.
the identification of thermal diffusivity. This work is under-
Although different industries and research institutes are
taken in the context of a joint project between the Graduate
already using the Flash Method for thermal properties
School in Engineering of the Federal University of Rio de
measurements in Brazil, the lack of a national standardized
Janeiro (COPPE-UFRJ) and the National Institute of Me-
procedure does not allow the inter-laboratory comparison of
trology, Standardization and Industrial Quality
thermal properties, which reduces the measurement trace-
(INMETRO).
ability. Therefore, the National Institute of Metrology, Stan-
Keywords: flash method standardization, thermal diffusiv- dardization and Industrial Quality (INMETRO) in a joint
ity measurement, and thermal conductivity. project with the Federal University of Rio de Janeiro
(COPPE/UFRJ), funded by the National Council for Scien-
tific and Technological Development (CNPq), is working
1. INTRODUCTION towards the implementation of a primary method for the
measurement of the thermal diffusivity of solids materials
The second half of the twentieth century has seen the subject
based on the Flash Method, according to the ASTM stan-
of thermal properties measurement change from an exercise
dard E1461-0 [3]. This standardized primary method will
mostly of academic interest to a complex issue employing
stimulate the development of new techniques for thermo-
the attention of a large number of people in different sectors
physical properties identifications, which is of main interest
of industry, government and universities. The explosion in
of several Brazilian industries, decreasing its external de-
materials development for many emerging technologies has
pendence of qualified materials testing and thermal charac-
produced very substantial quantities of improved materials
terization.
and composites in different forms. This has led to a corre-
sponding requirement for evaluation of their thermophysical This paper describes the actions performed by INMETRO
properties more rapidly and often, under conditions that and COPPE/UFRJ within the scope of this joint project, as
cannot be satisfied using the appropriate classical methods well as a description of the method itself. A brief literature
[1]. These classical methods, which involved the fitting of review about the Flash Method is summarized and a discus-
steady state and non-steady-state experimental temperature sion of the traditional theoretical models considered for this
data to theoretical models, were usually time expendable. method is performed.
2. FLASH METHOD STANDARDIZATION IN BRA- ties. Another field of application of reference materials is
ZIL testing, measurement or analysis as a basis for important
decisions, e.g. for authoritative evidence. The definition of a
The objective of the work presented here is performing and
“reference procedure” presumes the existence of several
assessing thermophysical tests of materials and systems
procedures for a specified task or of different realizations of
including the provision of reference procedures and refer-
the same methodology. Therefore, a reference procedure is
ence materials.
qualified by the uncertainty of results, proven to fit for a
Reference materials and standards are essential tools to given purpose. Moreover it has to be accepted as such by the
validate techniques and apparatus based on primary methods relevant target groups. In the case of quantitative results the
of measurement and for use in the calibration of techniques uncertainty comprises trueness and precision combined in
based on secondary or indirect methods. In order to cover the sense of measurement uncertainty. In the case of qualita-
the wide ranges of material types and their overall conduc- tive results the uncertainty is an estimate of the probability
tivity and temperature ranges of use it is seen that a broad of erroneous results.
variety of materials and specimens are required to satisfy the
Accordingly, the developed procedures and techniques
total needs.
within the scope of this project are implementations of pub-
During the last years there has been a huge increase in the lic standard methods. It will not be include test methods for
development and use of new and modified materials for a which special legal regulations apply, e.g. in the field of
very broad range of engineering, physical, chemical, bio- dangerous goods and substances, nor healthy and environ-
logical and medical applications. In a majority of cases mental rules. Provision of reference procedures at metro-
involving heat transfer, the thermal properties (e.g., thermal logical level will be oriented towards current needs and
conductivity, thermal diffusivity and specific heat) are nec- tasks; the reference procedures will be reviewed on a regular
essary to model, design and safely operate the system. The basis and an updated list will be published at an Internet site
levels of accuracy and precision required can vary depend- and in every new edition of a booklet catalogue.
ing upon the application. Reliability and confidence in the
In addition to the development of standard procedures in
measured value is the primary intention and use of reference
Brazil for the identification of thermophysical properties,
materials is an essential ingredient towards attaining this
the present project will also permit the validation and ad-
goal. It has been found that values, particularly of thermal
vancement of new techniques for the identification of ther-
conductivity for a material, vary significantly many times by
mophysical properties, developed within the framework of
amounts well outside the required precision levels. In addi-
research projects conducted by the group of the Laboratory
tion, some concern has been expressed by the occurrence of
of Heat Transmission and Technology at COPPE/UFRJ.
cases where significant discrepancies appear to exist be-
Such novel techniques have been under development by
tween directly measured thermal conductivities and values
using inverse analysis and design of optimum experiments
derived from measurements of thermal diffusivity and spe-
approaches [4]. Among the techniques already available, we
cific heat for certain well-known isotropic materials. Clearly
can find those for the identification of the thermal conduc-
such differences and uncertainties create serious problems
tivity components of orthotropic materials [5-9], of thermo-
for the scientists and engineers requiring reliable data for
physical properties of ablating materials [10], of the heat
whatever material and/or application of concern to them.
capacity and thermal conductivity of solids [11,12] and of
The final goal is the development, certification and provi- the heat and mass transfer properties of drying bodies
sion of reference materials, both in Brazil and abroad, de- [13,14].
signed to meet the increasing scientific and commercial
requirements. Reference materials (RM) are materials or
3. LITERATURE REVIEW
substances with sufficiently well established properties,
which can be used for calibrating instruments, assessing The thermal diffusivity, α, of a medium is the thermophysi-
measurement methods or assigning values to materials. cal property that is associated with the propagation of heat
Reference Materials are the key to guarantee reliability and into a medium during changes of temperature in time. The
accuracy for technical measurement data. higher the thermal diffusivity, the faster is the propagation
In addition to reference materials we can refer to reference of heat. The thermal diffusivity is related to the thermal
procedures (RP). We define reference procedures as those conductivity, k, specific heat, Cp, and density, ρ, as follows:
that have been designed to deliver results, which can be used α = k /( ρC p ) (1)
as a reference. The term “reference procedure” applies to
testing, measurement and analysis, i.e. all procedures for The thermal diffusivity has the dimension length2/time and it
determining characteristics of materials, products and proc- is usually expressed in the unit m2/s. The most popular
esses. These characteristics can be of quantitative or qualita- method used for measuring thermal diffusivity is the flash
tive form, and they can be defined independently or by the method. It has the advantage of being fast while providing
procedure itself. values with excellent accuracy and reproducibility. After the
Reference procedures are used to validate other procedures, sample has been stabilized at a desired uniform temperature,
to characterize reference materials or reference objects, as T0, a nearly instantaneous pulse of energy (usually laser,
well as for determining reference values of materials proper- Xenon lamp, or other discharge source) is imposed on its
front surface, and the temperature increase on the rear sur-
face of the sample is recorded as a function of time (Fig. 1). IR detector
The thermal diffusivity is then determined by comparing Signal
this thermogram with theoretical models that describe this amplifier
transient heat conduction phenomenon. These models are Data
discussed next. aquisition

Furnace
Furnace
3.1. The Parker Analysis Test sample power
Parker et al [2] developed the earliest method of analysis in supply
1961. The authors modeled the heat pulse diffusion by con-
duction in a perfectly insulated solid of thickness L, using a Heat pulse
one-dimensional transient formulation. Parker et al [2] power
source Computer, and
assumed that all the light energy Q was absorbed instanta- supply
(Flash) software
neously in a very thin layer g of the sample material facing
the heat pulse source (zero pulse width), and that the mate- Figure 1. Schematic of Flash Method
rial properties were constant over the small temperature thermal diffusivity measurement
range of the measurement. Under these conditions, the tem-
perature at the rear face (x = L) of the solid at any time can
be written as [15]: We note the interesting fact that the time corresponding to
half of the maximum temperature, which is the selected time
Q  ∞
 αt  to compute the thermal diffusivity with equation (6), results
T ( L, t ) = 
ρC p L 
1 + 2 ∑ ( −1) n exp − n 2π 2 2 
 L 
(2) in a large sensitivity coefficient [16]. Therefore, the use of
n =1 such time results in an accurate estimated value for thermal
conductivity, if the experimental implementation of the
where ρ, Cp and α are the density, the specific heat, and the
method is in accordance with the proposed theoretical
thermal diffusivity of the solid, respectively. Equation (2)
model.
can be written in a dimensionless format, by using the fol-
lowing non-dimensional parameters: Parker et al. [2] compared data obtained from Eq. (6) with
several other sources, obtaining good agreement and maxi-
Θ = T ( L, t ) / Tm (3) mum deviations on the order of ± 5%. However, this model
is somewhat limited in applicability because it neglected the
τ = αt / L2 (4)
heat loss effect from the material surfaces (adiabatic condi-
where Tm is the maximum temperature rise at the rear sur- tion), and it assumed instantaneous pulse heating in the front
face [Tm = T(L, t → ∞)]. By substituting Eq. (3) and (4) into surface of the solid. These effects become more evident at
Eq. (2) we obtain: high temperatures.

3.2. Analyses Accounting for Heat Losses
Θ(τ ) = 1 + 2 ∑ (−1) n exp(−n 2π 2τ ) (5)
Cowan [17] derived in 1963 equations for the temperature
n =1
distribution within a thin slab of solid material, which has
Equation (5) is a function of the non-dimensional time τ received a short pulse of energy on one surface, for the case
only. Figure 2 shows the variation of the non-dimensional in which energy losses at the surfaces by radiation or con-
temperature Θ as a function of τ. According to Parker et al. vection to surroundings at T0 are not negligible. When heat
[2], the thermal diffusivity α can be determined from Eq. (5) is lost from the sample surface, the temperature vs. time
and Fig. 2, at the point where the non-dimensional tempera- curve reaches its maximum at a lower temperature when
ture Θ is equal to 0.5. At this point, the non-dimensional compared to the adiabatic case. After the peak, the solid
time τ is equal to 0.1388, and from Eq. (5) it follows that: temperature drops off due to the cooling effect provided by
the heat loss. Additionally, it was also assumed that the
0.1388L2 energy pulse was rectangular in shape and of relatively short
α= (6)
t0.5 duration (pulse time tpulse << t0.5). The expression for the rear
surface temperature derived by Cowan [17] is:
where t0.5 is the time required for the back surface tempera-

ture to reach half of the maximum temperature rise. There- Q  2 2 
fore, the thermal diffusivity α can be directly determined Θ(τ ) =
ρC p L  ∑
exp( − aτ ) + 2 ( −1) exp( − n π τ )
n
(7)
from Eq. (6) by measuring the time where the non- n =1 
dimensional temperature Θ equals 0.5 (t0.5). It is important where τ is defined by Eq. (4) and Θ(τ) = T(L, τ) – T0. The
to observe that it is not necessary to know the amount of non-dimensional parameter a in the first term on the right
energy absorbed in the front surface in order to determine hand side of Eq. (7) accounts for the heat losses at the solid,
the thermal diffusivity with this method, which depends and it is given by:
only on the shape of the temperature vs. time curve (Fig. 2)
and thickness of the solid.  1  dW  1  dW  
a = L   +    (8)
 k  dT  x = 0 k  dT  x = L 
1,0 1,0
0,9 0,9

0,8 0,8

0,7 0,7

CpΘ( )L/Q
0,6
0,6
Θ(τ)

0,5 0,5
0,4 a=0
0,4
0,3 a = 0.01
0,3 a = 0.1
0,2 a = 0.5
0,2
0,1 a=1
0,1
0,0
0,0
0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9 1,0 0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9 1,0
τ = t/L2
Figure 2. Thermal diffusivity measurement Figure 3. Thermal diffusivity measurement
according to Parker et al. [2]. according to Cowan [5].

where W is the heat flux lost to the ambient at the surfaces at different dimensionless temperatures θ as measured on the
x = 0 and x = L. It is clear that if there are no heat losses temperature rise curve. These times were t0.2, t0.3, t0.4, t0.7 and
from the solid surfaces, a = 0 and Eq. (7) turns into Eq. (5), t0.8. They developed a theoretical relationship between the
which represents the case modeled by Parker et al. [2]. dimensionless time τ in Eq. (4) and the following three sets
of time ratios:
This method requires the knowledge of the value of a at
least approximately, for maximum accuracy. In the case of Θ(t0.7)/Θ(t0.3), Θ(t0.8)/Θ(t0.4), and Θ(t0.8)/Θ(t0.2) (11)
radiation losses only, a rough knowledge of the sample
The three values resulting from these ratios are very close
emissivity ε may give a with sufficient accuracy from [17]:
and are averaged to compute the thermal diffusivity value.
a ≈ 2.3 × 105{[1 + (TL / T0 ) 3 ](T / 1000) 3 (εL / k )} (9) Koski [18] presented a two-dimensional transient analysis of
the heat transfer problem, including a dimensionless heat
Since Equation (8) correlate Θ and τ as a function of a, an loss factor to characterize the radiative losses form the sam-
alternative approach for the thermal diffusivity estimation is ple.
to experimentally measure the ratios:

γ n2 (γ n2 + Bi 2 ) cos(γ n )
∑ exp( −γ n2τ )
Θ(5t0.5)/Θ(t0.5) or Θ(10t0.5)/Θ(t0.5) (10) Q
Θ( L,τ ) = 2 (12)
ρC p n =1 γ n2 + Bi 2 + 2 Bi
and then find the value of τ for a given heat loss estimation.
When there is no heat loss from the sample (a = 0), the rear where γn are the roots of:
surface temperature remains constant after reaching its
maximum and the ratio Θ(5t0.5)/Θ(t0.5) [or Θ(10t0.5)/Θ(t0.5)] 2γ n Bi
tan γ n = (13)
becomes 2. The non-dimensional time τ in equation (7) then γ n2 − Bi 2
becomes 0.1388 as in Parker’s [2] analysis. Although this
method accounts for the heat losses, two aspects of this The radiative losses parameter is written in a Biot number
model may be considered as source of errors: 1) the derived format, and it is given by:
equations require estimation of parameters, such as the
4σεT 3 L
amount of heat loss per square meter; 2) the use of Eq. (9) is Rloss ≡ Bi = (14)
tied to the solid thermal conductivity k, which must be k
known a priori. where σ is the Stefan-Boltzmann constant, ε is the surface
Clark and Taylor [18] proposed in 1975 another method for emissivity, T is the average absolute temperature of sample
the thermal diffusivity measurement based on Cowan’s [17] front and back faces, L is the sample thickness, and k is the
procedure. However, instead of analyzing the cooling part of sample thermal conductivity. The radiative Biot number is
the temperature rise curve, they proposed a method for de- determined in the Koski [17] analysis and provides a meas-
termining thermal diffusivity based on analysis of the heat- ure of the heat loss during the measurement. When
ing part of the curve (i.e., before the back surface tempera- Bi = 0, there is no heat loss from the sample. Equation (14)
ture reaches its maximum). Clark and Taylor [18] deter- is useful in that it shows how the various variables influence
mined five points on the time scale corresponding to five the loss factor. Koski's [17] analysis also allows the laser
pulse width, tpulse to be relatively long, as compared to the 3.3. Heat Capacity Estimations
characteristic heating time.
The specific heat of a material is defined as the amount of
Koski [17] uses five time and temperature ratios including energy required to raise a unit mass of material by one unit
those four indicated in Eqs. (10) and (11), and an additional of temperature at constant pressure:
ratio:
Q
Θ(tpulse)/Θ(t0.5) (15) Cp = (19)
m∆T
Koski [17] generated a series of coefficients that can be used
in conjunction with various combinations of the five tem- where Cp is the specific heat, Q is the energy, m is the mate-
perature ratios and temperature ratios to calculate three rials mass, and ∆T is the change in temperature.
values of the dimensionless time τ parameter in Eq. (4) Specific heat can be measured with the flash method by
based on Clark and Taylor's time ratios, and one value of the comparing the temperature rise of the sample to the
τ parameter based on Cowan's temperature ratio. The three temperature rise of a reference sample of known specific
Clark and Taylor results are then averaged. Loss factors (Eq. heat tested under the same conditions [2]. This temperature
14) are also computed for each case. The results of the rise is recorded during the diffusivity measurement, so the
Koski [17] analysis are limited to cases where the loss factor specific heat can be calculated from the same data with a
Bi is less than 1.2 and the pulse time is less than t0.5. Flash suitable calibration. Assuming that the light pulse energy
method measurements based on a Xenon Lamp usually have and its coupling to the sample remain essentially unchanged
a minimum pulse width of 100µs. This means that t0.5 must between samples, Eq. (19) can be rewritten as follows:
be greater than 100µs in all cases. Even when testing thin
samples of high diffusivity material, t0.5 is likely to be Q = Absorbed energy = (∆T m Cp)sample = (∆T m Cp)ref (20)
greater than 100µs. The Koski [17] method thus produces Or, simply:
two values of thermal diffusivity and loss factor for each
test. The results are typically very close. Koski [17] recom- ( m∆TC p ) ref
mends that the results based on the Clark and Taylor [16] (C p ) sample = (21)
( m∆T ) sample
ratios be selected for most conventional tests.
Degiovanni and Laurent [18] developed a different method A reference sample is measured at each temperature of in-
of analyzing the experimentally obtained temperature rise terest to calibrate the change in output voltage of the IR
curve in 1986. Degiovanni and Laurent [18] used the time detector (∆V) resulting from the absorbed laser energy. The
integral of the temperature rise curve, Θ vs. t, which repre- measured ∆V divided by the detector amplifier gain (G) will
sent the area under the curve, to describe its shape. In addi- be proportional to the temperature rise (∆T), as long as the
tion, a second time integral is calculated representing the temperature rise is small. This calibration gives the absorbed
area under the curve Θ/t vs. t, which can easily be derived energy or calibration factor in terms of the mass, specific
from the temperature history curve. The integration is car- heat, ∆V and G for the reference sample measurement.
ried out between the limits t0.1 and t0.8: When a sample of unknown specific heat is measured, the
absorbed energy divided by the product of the mass and
t0.8
∆V/G of the test sample, Equation (21), gives the test sample
m0 = ∫ Θdt (16)
specific heat.
t0.1

and The measured ∆V will be affected by heat loss during the


measurement, so in order to use ratios of the measured ∆V in
t0.8Θ Equation (21), the heat loss factors (see section the Koski
m−1 = ∫ dt (17) [19] analysis) of the reference and test sample should be
t0.1 t
similar. Tests have shown that the light pulse energy is sta-
Between certain limits of the value of m–1, a unique relation- ble from pulse to pulse within about ±2%, so the average of
ship exists between m0 and m–1 that can be expressed as: several laser pulses at each temperature are used in the cal-
culations. The absorptive efficiency of the front surface of
m0 = ( L2 / α )[ F ( m−1 )] (18) the samples to the light pulse and the radiative efficiency of
the back surface to the IR detector are controlled by coating
where F(m–1) is a polynomial function of m–1, whose coeffi- the reference and test samples with a thin layer of graphite.
cients were derived by Degiovanni and Laurent [18]. The
thermal diffusivity can be computed from Eq. (18) once the 3.4. Thermal Conductivity Estimations
two integrals have been calculated from the experimental
curve. Degiovanni and Laurent [20] assumed that the laser The thermal conductivity of a sample can be calculated from
pulse width is small compared to the half-rise time t0.5 (i.e., Eq. (1) using the diffusivity and specific heat data obtained
there is no finite pulse width correction). The analysis al- by the flash method, that is,
lows for heat loss from all sides of the sample, and it should
k = α(ρCp) (22)
yield accurate results on thick samples of low thermal diffu-
sivity. The primary data measured in the Flash Method is the ther-
mal diffusivity (α), while the specific heat (Cp) is a secon-
dary data. The thermal conductivity is not measured; it is [2] Parker W. J., Jenkins W. J., Butler C. P., and Abbott G. L.
derived from a primary and a secondary measurement data. “Flash Method of Determining Thermal Diffusivity, Heat
Capacity and Thermal Conductivity”. In: Journal of Applied
It is important now to observe that the density (ρ) of the Physics, Vol. 32, No. 9, pp. 1679-1684, Sept. 1961.
sample must be known for the thermal conductivity calcula-
[3] ASTM Test Method E1461-01. “Standard Test Method for
tion, while for the specific heat, the mass (m) of a sample Thermal Diffusivity of Solids by the Flash Method”.
must be known. Therefore, special care must be taken when American Society for Testing and Materials, Annual Book of
measuring properties of porous materials. More specifically: ASTM Standards, Vol. 14.02, 2001.
(i) If the measurements are performed in a high vacuum
[4] Ozisik, M.N. and Orlande, H.R.B., Inverse Heat Transfer:
environment, the heat transmission will occur only in the
Fundamentals and Applications, Taylor & Francis, New
solid phase of the porous material. On the contrary, convec- York, 2000.
tion with the pores will affect the temperature vs. time curve
shape on the back surface of the sample; (ii) If the mass and [5] Mejias, M. M.; Orlande, H. R. B.; Ozisik, M. N. Effects Of
density of the sample includes the effects of the fluid mass The Heating Process And Body Dimensions On The
Estimation Of The Thermal Conductivity Components Of
within the porous structure, the measured thermophysical
Orthotropic Solids, Inverse Problems In Engineering, Vol.
properties will be the effective specific heat and thermal 11, no1, pp.423-440, 2003.
conductivity (solid + fluid).
[6] Mejias, M. M.; Orlande, H. R. B.; Ozisik, M. N. Design Of
Optimum Experiments For The Estimation Of The Thermal
4. CONCLUSIONS Conductivity Components Of Orthotropic Solids, Hybrid
Methods In Engineering, New York, V. 1, N. 1, P. 37-53,
Generally, from the work conducted so far within the scope 1999
of this joint project between COPPE and INMETRO, we
can mention the following, regarding the major advantages [7] Orlande, H. R. B.; Mejias, M. M.; Ozisik, M. N. On The
of the flash method: Choice Of Boundary Conditions For The Estimation Of The
Thermal Conductivity Components Of Orthotropic Solids In.
• It requires a very small sample, usually 12 mm diameter 2000 ASME National Heat Transfer Conference, 2000,
and few millimeters thick; Pittsburgh, 2000
[8] Mejias, M. M.; Orlande, H. R. B.; Ozisik, M. N. A
• It is very fast; the actual measurement takes few seconds;
Comparison Of Different Parameter Estimation Techniques
• The same instrument can be made to test a very broad For The Identification Of Thermal Conductivity Components
range of thermal diffusivity materials (10–7 to 10–3 m2/s), Of Orthotropic Solids, 3rd International Conference On
while usual thermal conductivity instruments barely cover Inverse Problems In Engineering, Port Ludlow, WA, USA,
pp. 325-332, 1999.
a 10 to 1 ratio range;
[9] F. A. Rodrigues, H. R. B. Orlande And M. M. Mejias, Use
• It can also be used to measure other thermophysical Of A Single Heated Surface For The Estimation Of Thermal
properties (specific heat and thermal conductivity); Conductivity Components Of Orthotropic 3d Solids, 4th
International Conference On Inverse Problems In
• It can be used to test samples up to much higher tempera- Engineering, Rio De Janeiro, Brazil, 2002.
tures than steady state methods.
[10] Orlande, H. R. B.; Silva, D. Estimation Of Thermal
On the other hand, major disadvantages of the flash method Properties Of Ablating Materials In. International
are: Symposium On Inverse Problems In Engineering Mechanics
- 2001, Nagano, pp. 49-54, 2001
• The equipment is usually quite expensive, due to the heat
pulse generation, optical detection and high-speed data [11] Oliveira, A. P.; Silva, D.; Souza, R.; Orlande, H. R. B.
acquisition systems; Estimation Of Thermal Conductivity And Volumetric Heat
Capacity In Transient Experiments, In. Xv Congresso
• Porous and non-homogeneous materials require special Brasileiro De Engenharia Mecânica - Cobem, 1999, Águas
care to properly process the temperature data. de Lindóia, 1999
[12] F. Salas, A. Oliveira, D. Silva, H. Orlande, Experimental
Identification Of Thermal Conductivity And Volumetric
Heat Capacity, Paper CIT02-0117, Thermal Sciences
ACKNOWLEDGEMENTS Meeting - Encit, Caxambu, 2002.

This work is supported by CNPq through the grant [13] L. B. Dantas, H. R.B. Orlande and R. M. Cotta, Estimation
460.203/01-0. of Dimensionless Parameter of Luikov’s System for Heat
and Mass Transfer in Capillary Porous Media, Int. J. Therm.
Sci., vol. 41, pp. 217-227, 2002.
REFERENCES [14] L. B. Dantas, H. R.B. Orlande and R. M. Cotta, An Inverse
[1] Campbell, R. C. and Smith, S. E. “Flash Diffusivity Method: Problem of Parameter Estimation for Heat and Mass
A Survey of Capabilities”. In: Electronics Cooling, Vol. 8, Transfer in Capillary Porous Media, Int. J. Heat Mass
no. 2, May 2002. Also available at http://www.electronics- Transfer, vol. 46, pp. 1587-1598, 2003.
cooling.com/html/articles.html
[15] Carslaw, H. S. and Jaeger, J. C. “Conduction of Heat in Helcio R. B. Orlande, PhD.: Associate Professor, Department of
Solids”. 2nd Edition, Oxford University Press, New York, Mechanical Engineering – PEM/EE/COPPE, Federal University of
1959. Rio de Janeiro – UFRJ, Cidade Universitária, Cx. Postal 68503,
[16] H.R.B. Orlande and M.N. Özisik. Simultaneous Estimation Rio de Janeiro, RJ, 21945-970, Brazil. Phone/Fax: (21) 2562 8405
of Thermal Diffusivity and Relaxation Time with Hyperbolic Renato M. Cotta, Ph.D.: Professor, Department of Mechanical
Heat Conduction Model, Proceedings of the 10th Engineering – PEM/EE/COPPE, Federal University of Rio de
International Heat Transfer Conference, Paper No15-CI-20, Janeiro –UFRJ, Cidade Universitária, Cx. Postal 68503, Rio de
Vol. 6, pp.403-408, Brighton, UK, 1994. Janeiro, RJ, 21945-970, Brazil. Phone: (21) 2562 8567, Phone/Fax:
(21) 2562 8566
[17] Cowan, R. D. Pulse Method of Measuring Thermal
Diffusivity at High Temperatures, Journal of Applied Sérgio P. Oliveira: National Institute of Metrology, Standardiza-
Physics, Vol. 34, No. 4 (part 1), pp. 926 – 927, April 1963. tion and Industrial Quality – INMETRO, Av. Nossa Senhora das
Graças 50, Xerém, RJ, 25250-020, Brazil.
[18] Clark, L. M. and Taylor, R. E., Radiation Loss in the Flash
Method for Thermal Diffusivity, Journal of Applied Physics, Jorge A. Cruz: National Institute of Metrology, Standardization
Vol. 46, No. 2, pp.714 – 719, February 1975. and Industrial Quality – INMETRO, Av. Nossa Senhora das Gra-
ças 50, Xerém, RJ, 25250-020, Brazil.
[19] Koski, J. A. Improved Data Reduction Methods for Laser
Pulse Diffusivity Determination with the Use of Paulo R. F. Santos: National Institute of Metrology, Standardiza-
Minicomputers, Proc. of the 8th Symposium on tion and Industrial Quality – INMETRO, Av. Nossa Senhora das
Thermophysical Properties, Volume II, pp. 94, 1981. Graças 50, Xerém, RJ, 25250-020, Brazil.
[20] Degiovanni, A. and Laurent, M., Une Nouvelle Technique Cláudio S. P. Costa, M.Sc.: Researcher, Department of Mechani-
d'Identification de la Diffusivite Thermique pour la Methode cal Engineering – PEM/EE/COPPE, Federal University of Rio de
Flash, Revue Phys. Appl., Vol. 21, pp. 229-237, (1986). Janeiro –UFRJ, Cidade Universitária, Cx. Postal 68503, Rio de
Janeiro, RJ, 21945-970, Brazil. Phone: (21) 2562 8566
Paulo Couto, Dr.Eng.: Visiting Professor, Department of Me- Henrique Massard: Undergraduate Student, Department of Me-
chanical Engineering – PEM/EE/COPPE, Federal University of chanical Engineering – PEM/EE/COPPE, Federal University of
Rio de Janeiro - UFRJ, Cidade Universitária, Cx. Postal 68503, Rio Rio de Janeiro –UFRJ, Cidade Universitária, Cx. Postal 68503, Rio
de Janeiro, RJ, 21945-970, Brazil. Phone/Fax: (21) 2562 8566 de Janeiro, RJ, 21945-970, Brazil. Phone: (21) 2562 8566

You might also like