You are on page 1of 4

VOL. 33 NO.

1 (2021)

ARTICLE

Traceable measurements by
atomic spectrometry
Mike Sargent
Metrology Consultant, Surbiton, UK. metrology@misar.uk; https://orcid.org/0000-0001-6325-7944

The concept of traceable measure- case traceable calibration values will the mole which is the base unit for the
ments is simple and well established.1 probably not be adequate on their own amount of a chemical substance. Until
Traceability solves the problem of ensur- to achieve consistent results. Additional that point the Metre Convention was
ing that measurements of the same precautions such as formal method vali- concerned solely with physical measure-
quantity are comparable even when dation are then necessary. This applies to ments.
made by different people, in different most measurements made by analytical
places or at different moments in time. chemists and is discussed later. Traceability in chemistry
Almost all measurements require a cali- Nevertheless, analytical chemists have
bration of some kind. If measurements of Global traceability become accustomed to using the phys-
the same quantity always use calibration For a long time each community requir- ical measurement system because it
standards derived ultimately from the ing traceable measurement results underpins traceable calibrations for
same unique reference point, or primary devised its own units and primary stand- balances and volumetric glassware. This
standard, all will be based on the same ards as well as the means to dissemi- is important, but not an end in itself,
measurement unit and hence compa- nate them. However, the 19th Century because analytical chemistry is about
rable with each other. It is rarely possi- industrial revolution created the need more than determination of mass and
ble to calibrate all measurements directly for a single, global system. A conven- volume per se; the analyst must meas-
against the unique primary standard. The tion held in Paris in 1875 resulted in an ure the amount of the required element
answer is a traceability “chain” compris- international treaty known as the Metre or compound. To achieve this, most
ing a series of calibrations using interme- Convention.2 The structure envisaged in instrumental methods require calibration
diate reference standards, the values and the Metre Convention to support trace- of the instrument using solutions or solid
uncertainties of which are passed to the able measurements has continued into artefacts containing known amounts of
next link in the chain. the 21st Century. Originally there were the target analyte. With direct analysis of
Provided that the measurement proce- just three base units—for mass, length solid samples the instrument response
dure is relatively straightforward, such as and time. Further base units were subse- varies substantially with the overall
with everyday measurements of length quently added and there are presently composition. In this case reliable calibra-
and mass, comparability is sufficient seven as shown in Table 1. These form tion requires the use of known values
to ensure consistent (or equivalent) the basis of the International System of from “previously analysed materials” of
measurement results. However, meas- Units, the SI.3 More than seven units are very similar composition to the sample,
urements such as amount of a chemi- in common use, but the others can be i.e. matrix-matched reference materi-
cal substance increase the complexity directly derived from those in the table. als. Both these and calibration solu-
of the measurement procedure. In this The most recent addition, in 1971, was tions are chemical analysis standards

Table 1. The base units of the International System of Units (SI).

Physical quantity Name of unit Abbreviation


Mass kilogram kg
DOI: 10.1255/sew.2021.a1
Length metre m
© 2021 The Author Time second s
Published under a Creative Commons Temperature kelvin K
BY-NC-ND licence Amount of substance mole mol
Electric current ampere A
Luminous intensity candela cd

www.spectroscopyeurope.com SPECTROSCOPYEUROPE 11
VOL. 33 NO. 1 (2021)

ARTICLE
available from a wide variety of suppli- properly validate measurement proce- accredited in accordance with ISO 17034.
ers. However, for much of the 20 th dures, estimate uncertainty and ensure However, it is still prudent to carefully
Century no attempt was made to apply full documentation. Calibration stand- assess the materials. For example, ISO
the concept of traceability to ensure the ards or matrix reference materials used in 17034 allows non-certified values to be
consistency of their values. This was not the procedure should have a traceability included on a certificate. This is useful,
feasible at the time due to the lack of a chain which is complete and fully docu- but such values are unsuitable for cali-
global chemical measurement infrastruc- mented. Commercial availability of suita- bration purposes and are unlikely to be
ture comparable to the physical one. ble materials is generally good for many covered by the certificate’s traceability
In the 1980s a move began to address applications of atomic spectrometry. This statement. There is no standard format
numerous quality issues identified with is particularly true for most elemental for traceability statements and many are
chemical measurements. In particu- analysis applications using techniques limited to unhelpful phrases such as
lar, measurement laboratories were such as atomic absorption spectrome- “traceable to the SI”.5 Users should look
encouraged to seek accreditation to try, inductively coupled plasma-optical for more detail, either in a more compre-
the ISO/IEC 17025 or ISO 15189 stand- emission spectroscopy and induc- hensive statement or as additional infor-
ards. Subsequently, this was extended to tively coupled plasma-mass spectrom- mation supplied with it. This usually
include accreditation of reference mate- etry (ICP-MS) which are calibrated with includes the source of the traceable cali-
rial producers to ISO 17034. Meeting single or multi-element solutions. Solid bration values used in assigning values
the requirements of these standards for sampling techniques such as X-ray fluo- to the material.
traceable chemical measurement results rescence or arc/spark optical emission Assignment of values to matrix refer-
necessitated the creation of an appro- spectroscopy are more difficult, due to ence materials often requires compli-
priate international infrastructure. This the need for matrix-matched calibrations. cated methods and the producer should
was done by extending the one which However, older CRM stocks for applica- provide information about these, includ-
had been developed for physical meas- tions such as steels, alloys and ores are ing how they were validated. Matrix
urements over more than 100 years.4 steadily being replaced with modern reference materials values are frequently
Calibrants with traceable values are now materials having traceable values. assigned by inter-laboratory compari-
widely available, including many needed Wherever possible purchase commer- sons. Claims such as “traceable to the
for applications of atomic spectrometry. cial reference materials from a supplier methods used by the participants” are
Making use of these materials requires
laboratories to become the last link in
the traceability chain. In addition to the
chemical calibration standard or solu-
tion there will be another chain for cali- kilogram
bration of the balance and perhaps also
for volumetric glassware. Whether or not
calibration values must be traceable
depends on whether they contribute
significantly to the overall uncertainty of
candela metre
the measurement result. Hence, matrix
materials used for method validation
should have traceable values. This is
obvious if a matrix correction or recov-
ery factor is subsequently applied to the
results. However, even when a certified SI Units
reference material (CRM) is used only
to confirm 100 % recovery of analyte,
ampere second
the uncertainty of the certified value(s)
still contributes to the overall uncertainty
estimation for the measurement results.

Traceability in the
laboratory
The tasks needed by the laboratory to mole kelvin
take its place at the end of the tracea-
bility chain are straightforward, because
they comprise best practice specified
for laboratory accreditation. In summary,

12 SPECTROSCOPYEUROPE www.spectroscopyeurope.com
VOL. 33 NO. 1 (2021)

ARTICLE
unhelpful, so look for additional informa- taken. A routine spectroscopic scan will purity but there is a problem. Organic
tion. Materials with a very long shelf life, confirm the absence of other elements compounds close to 100 % pure are
such as those used in metallurgy and even at trace levels. If impurities are rarely available, so an accurate purity
geochemistry, also need caution. Many identified at significant concentrations, assessment is essential to avoid signifi-
still on sale today were certified before it may be easier to find an alternative cant errors. This is not a trivial task7 and
current best practice using informal inter- source rather than attempt to measure presents major difficulties with large
laboratory exercises, often without trace- their amount accurately in order to apply molecules.
able calibrants. Moreover, these older a correction. In many cases, metals sold There is, however, an alternative cali-
materials may sometimes be used as as “4 or 5 nines pure” are suitable but bration approach for speciation applica-
calibration standards for certification of there are two potential problems. These tions using hyphenated ICP-MS. As the
modern reference materials. metals may have pockets of occluded instrument response corresponds to the
gases which could introduce an error in total amount of the metal or heteroatom,
Traceability in practice the calculated concentration of the solu- it can be calibrated using an inorganic
Sometimes traceable elemental calibra- tion. Some metals are prone to oxidation, solution of the target atom. Calibration
tion solutions are not available commer- producing a surface oxide layer and the solutions with traceable values are read-
cially or laboratories are not in a position surface should be cleaned before weigh- ily purchased or prepared as described
to purchase them. Commercial solutions ing them. Both errors can be reduced for elemental analysis. This measure-
of analytes such as organo-­metallic or by using rod or wire instead of turnings ment does, of course, determine the
isotopic materials are also less common. or fine powders. It is also essential to total amount of the element and not
The laboratory then needs to prepare ensure that metal is taken into solution the target organic molecule. However,
them in-house and assign traceable quantitatively, which may not always be the element-specific chromatogram
values. In principle this is not difficult a trivial task. will reveal the presence of any impu-
and was once the norm for analyti- Solutions of metals such as rhodium, rity peak(s) containing the target atom
cal laboratories. However, the devil lies which are available in high purity form so a traceable value for the target mole-
in the detail. A common approach is to but are difficult to dissolve, or of non- cule can be calculated. This approach
take a known mass of the analyte from a metallic elements, may be prepared from has been used for compounds such as
sample having known purity and dissolve a suitable salt. However, dissolution of a selenoproteins,8 ferritin-bound iron9 and
it in a known mass or volume of solvent. known weight of the salt is rarely suffi- carboplatin adducts.10
Provided the balance and volumet- ciently accurate due to the lack of reli-
ric glassware have traceable calibration able data on stoichiometry, water content Isotopic analysis
values the laboratory can demonstrate or impurities. The preferred option is to It is sometimes forgotten that ICP-MS
the traceability of the assigned value. assay a solution of approximate concen- always makes isotopic measurements.
The devil is the purity of the solid analyte tration using a technique such as EDTA A specific isotope of the target element
sample. It does not necessarily need to titrimetry. EDTA solutions with certified is compared between the sample and
be ultra pure, but the value assigned to concentration are readily available or can calibrant solutions. This works provided
the purity must be reliable and tracea- be assayed using a certified solution of a the isotope’s abundance is the same
ble. Ideally, the pure material should be metal such as silver. To ensure traceable for both. Unless the element is mono
a CRM but these are often unavailable. results, each such calibration step must isotopic there is no guarantee that this
It is tempting to use commercial reagent be accurate and its measurement uncer- is true. For example, the natural isotopic
grade materials instead, but these usually tainty documented. abundance of lead varies quite widely
come with only an approximate estimate and isotopic fractionation of some
of purity or a batch specification. Such Speciation analysis elements can be caused by processing
values are inappropriate for the prepara- The hyphenation of ICP-MS with liquid techniques. Hence, to obtain traceable
tion of calibration solutions. Some possi- or gas chromatography, or field-flow- results it is necessary to determine the
ble alternatives for three application areas fractionation allows analysis of organic isotopic abundance of both the calibrant
are discussed below. compounds containing a metal atom and the sample, applying a correction if
or heteroatoms such as P and S. Used necessary.
Elemental analysis with appropriate analytes, this offers Calibration of mass spectrometry using
Unfortunately, few pure elements for greater sensitivity than most conven- isotope dilution analysis (IDMS) requires
preparation of calibration solutions tional organic mass spectrometers and addition to the sample of an isotopi-
are available as CRMs and undertak- selectivity for molecules containing the cally labelled material (the “spike”). This
ing a full assay procedure is not a triv- target atom. It also allows an easier route comprises, or is labelled with, an isotope
ial task. 6 However, many metals are to traceable measurement results than having low abundance in the natural
widely available for other purposes in organic mass spectrometry. The latter sample. Its ratio to a naturally abundant
very high purity forms. These may be also uses calibration solutions prepared isotope is then determined. The popu-
used provided suitable precautions are by weighing analyte samples of known larity of IDMS calibration has increased

www.spectroscopyeurope.com SPECTROSCOPYEUROPE 13
VOL. 33 NO. 1 (2021)

ARTICLE
with the wide use of ICP-MS,11 but it is values for their own products. Many of chemical measurements”, Accred.
then essential to compensate for a vari- the national institutes also have active Qual. Assur. 24, 407–415 (2019).
ation in instrument response between research programmes addressing trace- https://doi.org/10.1007/s00769-
the two isotopes, the so-called mass ability issues, including novel applications 019-01407-6
bias correction. Traceable values can be such as organo-metallics, protein analy- 8. M.E. del Castillo Busto, C. Oster,
achieved by obtaining or preparing solu- sis and isotopic measurements. Thus, we S. Cuello-Nuñez, C.L. Deitrich, A.
tions of the isotopic spike with known can be assured that traceable measure- Raab, A. Konopka, W.D. Lehmann,
amounts of the chosen isotope. An alter- ments in atomic spectrometry will have H. Goenaga-Infante and P. Fisicaro,
native approach is known as the “double” a long and successful future. “Accurate quantification of seleno-
or “reverse” IDMS technique 12 which proteins in human plasma/serum
uses a calibration solution of the natural References by isotope dilution ICP-MS: focus
analyte. This requires a relatively compli- 1. M. Sargent, “Quality assurance: on selenoprotein P”, J. Anal. At.
cated measurement procedure involving Traceabilit y ”, Encyclopedia of Spectrom. 31, 1904 (2016). https://
a sequence of “blends” in which both Analytical Science, 3rd Edn. Elsevier, doi.org/10.1039/C6JA00122J
the sample and calibration solutions are Vol. 8, pp. 527–535 (2019). https:// 9. A. Tchaikovsky, A. Schoeber, H.
spiked with the isotopic material. doi.org/10.1016/B978-0-12-409547- Schueff, A. Raab, S. Emin, A. Slany,
2.14398-7 P. Heffeter, G. Koellensperger and
Where next? 2. BIPM, The Metre Convention. C. Swart, “Quantification of ferritin-
As we have seen, traceable results have https://www.bipm.org/en/world- bound iron in murine samples for
a long and successful history with phys- wide-metrology/metre-convention/ Alzheimer’s disease studies using
ical measurements. Fortunately, the 3. NPL, The International System of species-specific isotope dilution
global measurement infrastructure has Units (SI). https://www.npl.co.uk/ mass spectrometry”, Metrologia
now been extended to support tracea- si-units 57, 042101 (2020). https://doi.
ble values in analytical chemistry. Hence, 4. BIPM, Consultative Committee for org/10.1088/1681-7575/ab8c9f
any spectrometry laboratory should Amount of Substance: Metrology 10. R. Larios, M.E. Del Castillo Busto, D.
be able to provide traceable measure- in Chemistry and Biology (CCQM). Garcia-Sar, C. Ward-Deitrich and H.
ment results by adopting current best https://www.bipm.org/en/commit- Goenaga-Infante, “Accurate quan-
practice. However, this depends on the tees/cc/ccqm/ tification of carboplatin adducts
availability of calibration standards and 5. M. Sargent, “The provision and use with serum proteins by monolithic
matrix reference materials with traceable of traceability statements for refer- chromatography coupled to ICPMS
values. Supply is already good for many ence materials”, Accred. Qual. Assur. with isotope dilution analysis”, J.
elemental calibration solutions, but there 25, 367–372 (2020). https://doi. Anal. At. Spectrom. 34, 729–740
is much more to do with matrix-matched org/10.1007/s00769-020-01450-8 (2019). https://doi.org/10.1039/
standards. These are essential, for exam- 6. J. Vogl, The IAWG Roadmap for the C8JA00409A
ple, with the solid sampling techniques Purity Determination of Pure Metallic 11. M. Sargent, H. Goenaga-Infante, K.
used in metallurgy and geochemistry. On Elements. BIPM, CCQM-IAWG/17-28 Inagaki, L. Ma, J. Meija, A. Pramann,
the positive side, developing the chem- (2017). https://www.bipm.org/wg/ O. Rienitz, R. Sturgeon, J. Vogl, J.
ical measurement infrastructure has CCQM/IAWG/Allowed/April_2017/ Wang and L. Yang, “The role of
encouraged many more national insti- CCQM-IAWG17-28.pdf ICP-MS in inorganic chemical metrol-
tutes to produce chemical standards and 7. K.A. Lippa, D.L. Duewer, M.A. Nelson, ogy”, Metrologia 56, 034005 (2019).
CRMs. These are available to laboratories S.R. Davies and L.G. Mackay, “The https://doi.org/10.1088/1681-7575/
but also provide the means for commer- role of the CCQM OAWG in provid- ab0eac
cial producers to establish traceable ing SI traceable calibrators for organic 12. M. Sargent, C. Harrington and R.
Harte, Guidelines for Achieving
High Accuracy in Isotope Dilution
Mass Spectrometry. Royal Society of
Chemistry, Cambridge, UK (2002).
Dr Mike Sargent is an analytical chemist with over ISBN 0-85404-418-3, https://doi.
50 years experience of atomic and molecular org/10.1039/9781847559302
spectrometry, reference materials production,
analytical quality assurance, laboratory manage-
ment and chemical metrology. He played a leading
role in developing UK and international chemical
metrology and chairs or participates in a number of
committees for analytical chemistry and reference
material production.

14 SPECTROSCOPYEUROPE www.spectroscopyeurope.com

You might also like