Professional Documents
Culture Documents
Thermoelectrics
www.advelectronicmat.de
Adv. Electron. Mater. 2019, 1900167 1900167 (1 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Adv. Electron. Mater. 2019, 1900167 1900167 (2 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Table 1. Summary of the steady-states methods presented in this review for measuring the thermal transport properties of the OHT materials.
∆T
qs = κ A (5)
L
Adv. Electron. Mater. 2019, 1900167 1900167 (3 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Th − Tc
Rs = (6)
qs
d
Rs = + Rcontact (7)
Aκ
Figure 2. Two configurations for thermal conductivity measurement according to the ASTM In recent years, the methods based on the
D5470 standard. ASTM D5470 standard have been used to
measure the thermal conductivity of several
both studies, the thermal conductivity at room temperature is thermoelectric carbon nanotube (CNT)/polymer nanocompos-
around 0.3 W m−1 K−1. ites.[24,26–29] The thickness of the specimens ranged typically
Efforts have been made to carry out the thermal conductivity from several tens of micrometers to hundreds of micrometers.
measurement above room temperature with the same experi- It has been found that the polymer composites with randomly
mental configuration. Zaitsev et al. have demonstrated an appa- distributed CNTs demonstrate fairly low thermal conductivities.
ratus which employed a radiation shield thermally anchored For example, Yu et al. prepared a series of samples by mixing
to both the heater and the heat sink to simulate a temperature single-wall carbon nanotube (SWCNT) with poly(3,4-ethylen
gradient that existed in the sample.[22] To further reduce the edioxythiophene):poly(styrenesulfonate) (PEDOT:PSS) and/
radiation heat loss, the space between sample and the heat or polyvinyl acetate polymers. The thermal conductivities for
shield was filled with thermally insulating powder. With this the composites with 60 wt% CNT were found to be 0.2–0.4 W
setup, the authors were able to simultaneously determine the m−1 K−1 at room temperature.[28] The values are close to those
thermal conductivity, electrical conductivity, and Seebeck coef- reported for multiwalled carbon nanotube (MWCNT)/polyani-
ficient of Mg2Si1−xSnx solid solution in a temperature range of line (PANI) composites, which lie between 0.4–0.5 W m−1 K−1
300–870 K.[22] when the MWCNT content was varied from 20 to 80 wt%.[27]
Another absolute steady-state method relies on test stacks In general, the absolute method is relatively easy to imple-
constructed from metal bars. The related test apparatus ment. By properly setting up the experimental apparatus, it is
and test procedures have been specialized in ASTM D5470 possible to measure the thermal conductivity along with the
standard.[23] Figure 2 shows two test configurations based on electrical conductivity and the seebeck coefficient for the bulk
this method. In the first configuration, the sample is pressed samples.[22] On the other hand, the heat losses and the thermal
between two metal bars (also known as meter bars) which are contact resistance are two major concerns when carrying out
made from highly thermally conductive materials with well- the absolute method. For the measurement of the OHT mate-
documented thermal conductivity (Figure 2a). A linear array rials, how to effectively reduce the thermal contact resistance
of temperature sensors is placed in each metal bar. When the between the sample surface and testing components (e.g.,
steady-state heat flow is reached, the heat flow through the meter bars, heater, heat sink, temperature sensors, etc.) is
sample can be measured by the metal bars. The interfacial especially critical, since the polymer or other organic mate-
temperatures between the sample and the metal bar (Th and Tc) rials usually possess relatively large thermal expansion coef-
are extrapolated from the linear temperature distributions on ficient which may be as high as 1.2 × 10−4 K−1.[9] To elucidate
both hot side and cold side.[24] In the second configuration, the problems caused by the samples with the large thermal
a guard heater is installed around the main heater in order expansion coefficient, it can be assumed that the sample is
to significantly reduce the heat leakage from the latter one[25] measured by the ASTM D5470 apparatus and the temperature
(Figure 2b). In this case, qs can be deduced from the power difference across the sample is large (e.g., >30 K). Under such
dissipated from the heater with necessary corrections for the a scenario, the thermal expansion at the hot surface of the
heat losses.[25] Th and Tc are obtained from the temperature sample may be significantly larger than that at the cold sur-
sensors that are installed in extreme proximity to the inter- face. This could cause the sample to bow and to generate gaps
faces between the sample and the metal bars. In both con- at the interfaces between the sample and the metal bars.[10]
figurations, compressive load and thermal interface materials The formation of the gaps leads to an increase of the thermal
such as thermal grease could be applied to minimize the contact resistance which may eventually deteriorate the quality
thermal contact resistances between the sample and the metal of the measurement . Therefore, to measure the OHT mate-
bars.[23] rials with relatively large thermal expansion coefficients, the
Adv. Electron. Mater. 2019, 1900167 1900167 (4 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
temperature difference should be carefully kept at a level (e.g., where L and A are the length and the cross-sectional area of the
several Kelvins[24]) that is small enough to avoid the formation sample, respectively.
of the gaps but is still large enough for carrying out reliable The technique is capable of obtaining the thermal conduc-
measurements. tivity of the small bulk samples with a relatively simple experi-
mental setup. It does not require special sample treatments
2.2. Parallel Thermal Conductance Method and the sample mounting is simple and rapid. Recently, Wan
et al. have applied this method to obtain the in-plane thermal
The parallel conductance method was developed by Tritt and conductivities of TiS2 single crystal and an inorganic/organic
co-workers for measuring small bulk samples at the millimeter superlattice material TiS2[(HA)0.08(H2O)0.22(DMSO)0.03]
scale which often bring about difficulties to the temperature at room temperature, which were found to be 4.45 and
and heat flow measurements.[30] As shown in Figure 3a, the 0.69 W m−1 K−1, respectively.[31] The value of the TiS2 single
sample holder consists of a heater (hot end) and a heat sink crystal is close in magnitude to the result previously reported
(cold end), which are connected by a supporting post with low by Imai et al.[32] Varghese et al. have measured the in-plane
thermal conductance. A differential thermocouple is used to thermal conductivity of an ≈100 µm thick free-standing ther-
measure the temperature difference between the hot end and moelectric film made by screen printing colloidal inks com-
the cold end. The measurement is carried out in two steps. In posed of Bi2Te2.8Se0.2 nanocrystals. The value (≈0.56 W m−1 K−1
the first step, a baseline thermal conductance Gbaseline is meas- at room temperature) is within 5% of the cross-plane thermal
ured for the empty sample holder, which accounts for the heat conductivity determined by the laser flash method, which indi-
losses associated with the sample holder itself. To do this, mul- cates the isotropic nature of the sample.[33]
tiple DC heating currents (I’s) are applied to the heater suc- Two issues should be paid attention when implementing the
cessively to generate corresponding steady-state temperature technique. First of all, to maintain a reasonably high measure-
differences (ΔT’s) across the supporting post. The heating ment sensitivity, the baseline thermal conductance associated
power P varies linearly with ΔT as denoted by the following with the sample holder Gbaseline needs to be made as low as
equation possible.[30] Second, the thermal contact resistance between the
specimen and the sample holder should be kept much smaller
P = I 2Re,H = Gbaseline ∆T (8) than the thermal resistance of the specimen itself to ensure the
accuracy of the measurement.
where Re,H is the resistance of the heater. Gbaseline is extracted
from the slope of the P ~ ΔT plot. The sample is then attached
to the sample holder, and the total thermal conductance Gtotal is 2.3. Comparative Method
measured in a similar way
P = I 2Re,H = Gtotal ∆T (9) The comparative method determines the thermal conductivity
of bulk materials without measuring the heat flow through
Both Gbaseline and Gtotal are measured in a high vacuum envi- the sample qs. In this method, a reference sample with known
ronment to reduce the convective heat losses.[31] The thermal thermal conductivity is thermally connected in series with the
conductivity κ of the sample is determined from the following specimen to be investigated. The stack structure is sandwiched
equation between the heater and the heat sink. Two or more temperature
κ = (Gtotal − Gbaseline ) L/A (10) sensors are arranged along the test sample and the reference
sample, respectively (Figure 4). Under the steady-state condi-
tion, the heat flow through the sample should be equal to that
Figure 3. Schematic diagram of the parallel thermal conductance method. Figure 4. Schematic of the experimental setup for the comparative
a) Empty sample holder. b) Sample holder attached with a sample. method.
Adv. Electron. Mater. 2019, 1900167 1900167 (5 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
through the reference sample. The thermal conductivity of the considered to be a merit for this modification. However, sig-
sample is then given by nificant error in measuring κ has been found in their studies
presumably due to high residual gas pressure remained in
A ∆T L their measurement chamber.[40] This implies that the issues
κ s = κ ref ref ref s (11)
As ∆TsLref such as heat losses and thermal contact resistance require
extremely careful handling when characterizing the small
where subscripts “s” and “ref” denote the specimen to be tested samples.[38]
and the reference sample, respectively. The method achieves
the best accuracy if the thermal conductivity of the sample is
comparable with that of the reference sample.[34] 2.4. Völklein Method
Test standards like ASTM E1225[35] and ASTM E1530[36]
describe the relevant apparatuses and test procedures for the Völklein et al. have developed a chip-based method to measure
comparative method. When applying this method to the OHT the in-plane thermal conductivity of the sub-micrometer thick
materials, the influence of the uneven thermal expansions thin film.[42–46] The chip (known as λ-chip) includes a rectan-
on the thermal contact resistance should be minimized[10] as gular suspended dielectric (e.g., SiNx) membrane with a thick-
for the absolute method discussed in Section 2.1. Marconnet ness of 50 to 100 nm, on which the film specimen can be
et al. have characterized the thermal conductivity of aligned deposited (Figure 5a). A thin metal stripe is created along the
CNT-thermoset epoxy composites using a setup conforming long axis of the membrane which serves as both a heater and
to the ASTM E1225 standard.[37] In this study, the tempera- thermometer. The silicon rim surrounding the membrane acts
ture sensing was carried out with an IR microscope instead as a heat sink. All measurements should be conducted in the
of thermocouples. The axial thermal conductivity of the com- vacuum environment. The thermal conductivity of the bare
posites was found to increase from 0.46 to 4.87 W m−1 K−1 as membrane needs to be determined first. To do this, a dc cur-
the volume fraction of CNT varied from 1 to 16.7 vol%, which rent is applied to the metal stripe and the resulted heat will
is comparable with that of unfilled and aligned CNT arrays flow through the membrane to the heat sink. At the same time,
investigated in the same study. On the other hand, as men- radiation heat loss will occur from the surfaces of the mem-
tioned in Section 2.1, Pisoni et al. have measured the tem- brane and the metal stripe. From the temperature rise and
perature-dependent thermal conductivity of the CH3NH3PbI3 the input electrical power, an apparent thermal conductance
sample[21] by adopting the experimental configuration shown in associated with the chip (G) can be calculated. With certain
Figure 4 and the obtained results were confirmed later by other approximations, it can be shown that the thermal conductance
authors.[20] of the membrane due to heat conduction and radiation (GM)
Verebelyi has extended this method to measure the is roughly equal to G.[44] Since GM is a function of unknown
thermal conductivity of Pb-doped Bi2Sr2Ca1Cu2Ox whisker thermal conductivity κM,∥ and emissivity ε of the membrane,
with d imensions less than 1000 × 100 × 100 µm3 within a it is necessary to carry out two measurements with two dif-
temperature range of 30–200 K.[38] The measurement error ferent membrane geometries so that one may obtain κM,∥ and ε
for such a small sample was estimated to be less than 25% simultaneously.[44]
with the application of the temperature matching shielding The thermoelectric thin film to be tested can be deposited
and micro-thermocouples. Later, Yoshino et al. have applied on the top side (top configuration)[46] or bottom side (bottom
this method to measure the thermal conductivity of mil- configuration)[45] of the membrane through various vapor dep-
limeter-sized organic crystals. By properly designing the osition procedures such as thermal evaporation, magnetron
sample holder, they were able to simultaneously obtain the sputtering, atomic layer deposition, etc. If the film is depos-
values of the thermal conductivity, the Seebeck coefficient, ited on the top side of the membrane, the metal stripe should
and the electrical conductivity from the same sample.[39–41] be covered with a thin insulating layer in order to avoid elec-
The capability of measuring κ together with S and σ is trical shortcuts (Figure 5b). The top configuration allows one to
Figure 5. a) Schematic illustration of the measurement chip used in the Völklein method. b) Deposition of the thermoelectric film on top of the
suspended membrane.
Adv. Electron. Mater. 2019, 1900167 1900167 (6 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
prepare the OHT samples by different kinds of solution-based 2.5. Microfabricated Suspended Device
methods such as spin-coating, dip-coating, drop casting, ink-jet
printing, etc.[47,48] Aiming for eliminating the parasitic heat transfer from the
To obtain the in-plane thermal conductivity of the depos- sample into the substrate, another steady-state method based
ited film (κf,∥), the thermal conductivity of the composite on microfabricated suspended devices (MFSD) has been devel-
layer (including the test film and the supporting mem- oped to measure the thermal conductivity of nanowires or
brane) κC,∥ is measured by employing the same procedure nanotubes along their axial direction or narrow thin film stripes
of obtaining κM,∥. κf,∥ is subsequently extracted from the fol- along their in-plane direction.[50–54] The temperature sensing of
lowing equation[44,45] this technique is performed at both ends of the sample and all
measurements are carried out in the vacuum environment to
reduce the convective heat loss. Two typical configurations will
κ C, dC − κ M, dM
κ f , = (12) be reviewed in this paper.
df One configuration proposed by Shi et al.[50] is schematically
shown in Figure 6a in which the test platform consists of two
where df, dM, dC are the thicknesses of the thin film, the mem- adjacent suspended SiNx membranes each supported by five
brane, and the composite layer, respectively. By applying this identical long and thin SiNx beams. A platinum resistance ther-
method, Linceis et al. have recently measured κf,∥ for a 110 nm mometer (PRT) is fabricated on each membrane to serve as the
thick Bi87Sb13 film over a temperature range of 120–450 K,[45] heater or the thermometer. The sample is transferred onto the
which shows good agreement with previously published device to bridge the two membranes. One membrane is used as
results.[49] heating membrane on which the PRT is heated by a dc current,
The use of the ultrathin suspended membrane is a major while another one is used as sensing membrane. It is assumed
advantage of this method, since the parasitic heat transfer from that the temperature rise on both membranes is uniform and
the sample into the substrate is eliminated,[43] which is critical to the heat losses due to gas flow, convection, and radiation can
achieving reliable measurements on the sub-micrometer thick be neglected. The Joule heat generated by the heater and the
thermoelectric films.[44,45] Additionally, the insulation layer two Pt current leads on the heating membrane are PH and 2PL,
and metal stripes are prefabricated on the test chip, which not respectively. A certain amount of the generated heat is trans-
only significantly simplifies the sample preparations but also ferred from the heating membrane directly to the environment
prevents the samples from undergoing harsh environments through the five beams. The rest heat flows into the sensing
of depositing the insulation layer or performing lithography membrane through the sample and is further conducted to
process. This is very important for the characterization of the the environment through the five beams connected with the
vulnerable OHT materials.[13] Moreover, the current experi- sensing membrane. It can be shown that the measured thermal
mental configuration has been integrated with measurement conductance of the sample Gs is expressed as[50]
techniques of the other TE properties on a single chip, there- ∆Tc (PH + PL )
fore all quantities needed for calculating the in-plane ZT-value Gs = (13)
∆Th2 − ∆Tc2
can be obtained from the same sample in one measurement
run.[45] where ΔTh and ΔTc are the temperature rise at the heating
On the other hand, it should be noted that the method is membrane and the sensing membrane, respectively. Both
incapable of measuring the cross-plane thermal conductivity of ΔTh and ΔTc can be measured from the resistance change of
the thin films. The measurement must be carried out in the the PRT. On the other hand, the measured thermal resistance
vacuum environment. The measurement accuracy depends on (i.e., Gs−1) consists of the intrinsic thermal resistance of the sample
the relative magnitudes of the thermal conductance associated Rn, and the thermal contact resistance between the sample and
with the deposited film and the bare membrane. If possible, the the two membranes, Rcontact
thermal conductance of the bare membrane should be made
smaller than that of the deposited film (κM,∥dM < κf,∥df). Gs−1 = R n + Rcontact (14)
Figure 6. a) Schematic illustration of the experimental configuration proposed by Shi et al.[50] b) Another configuration proposed by Sultan et al.[53]
Adv. Electron. Mater. 2019, 1900167 1900167 (7 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Only if Rcontact is small enough, the thermal conductivity of the including oxidized polyacetylene nanofibers (74–126 nm
sample κ can be calculated directly from Gs−1 by in diameter),[59] amorphous polythiophene nanofibers
L L (71–245 nm in diameter),[60] PEDOT:PSS and PEDOT:Tosylate
κ = s ≈ −1s (15) (Tos) films (130–230 nm thick),[61] and polymer treated CNT
Rn As Gs As
films or polymer:CNT hybrid films (≈90 nm thick in ref. [62]
where Ls and As are the length and cross-sectional area of the and 80–110 nm thick in ref. [64]).[62–64] Among these studies,
sample, respectively. Therefore, how to reduce Rcontact is one of Weathers et al. have systematically investigated the in-
major concerns when applying this technique. In practice, Rcontact plane thermal conductivities of a series of PEDOT:PSS and
can be reduced by metal or amorphous carbon deposition at PEDOT:Tos films, which demonstrated an obvious correlation
the locations where the sample is contacted with the mem- with the in-plane electrical conductivity σ∥. As σ∥ changed from
branes.[50] Alternatively, new techniques have been invented 20 to 500 S cm−1, the thermal conductivity could vary from
later to explicitly obtain the thermal contact resistance and about 0.5 W m−1 K−1 to as high as 1.8 W m−1 K−1 at 300 K.[61]
the intrinsic thermal resistance of the sample by, for example, The value of 0.5 W m−1 K−1 is close to the in-plane thermal con-
making use of the thermoelectric voltage of the sample to deter- ductivity (≈0.58 W m−1 K−1) for a low-σ∥ PEDOT:Tos film meas-
mine the temperature drops at the contacts.[51,55] ured by Ushirokita and Tada using a chip-based 3ω method that
Figure 6b illustrates another experimental configuration pro- will be discussed in Section 4.3.[65]
posed by Sultan et al. in which two identical suspended SiNx It is also worth mentioning that the test platforms can be
membranes each equipped with the heater and thermometer designed to couple with the characterizations of the Seebeck
and supported by four legs are linked together by a narrow SiNx coefficient and the electrical conductivity of the same sample.[61]
bridge.[53,54,56] One membrane is used as the heating mem- On the other hand, the SiNx bridge in Figure 6b allows direct
brane excited by a dc current, while another one as the sensing depositions of the films on it by various physical vapor deposi-
membrane. It can be demonstrated that the steady-state tem- tion (PVD) techniques[53,54] as well as solution-based ultrasonic
perature rises on the two membranes are related to the heating spray (UP) process.[62,63] The combination of the MFSD with
power P by[53] the UP process will facilitate the measurements of those solu-
tion- processable OHT films.[62,63]
GL + GB The major drawbacks associated with the method are that
∆Th = P (16)
(2GB + GL )GL it requires extensive microfabrication works and the transfer
of as-prepared samples to the microdevices could be chal-
GB lenging. The thermal contact resistance and the measure-
∆Tc = P (17)
( 2GB + GL )GL ment sensitivity are also critical issues that needed to be
carefully considered. Since the measurement needs to be car-
where GB and GL are the thermal conductances of the bridge ried out under the high vacuum condition, the technique is
and the supporting legs, respectively. GB and GL can be solved not applicable to the humidity-controlled tests for the OHT
from the slopes of the ΔTh ∼ P and ΔTc ∼ P straight lines. Before materials.[66]
transferring the test sample onto the bridge, the thermal con-
ductance of the bare bridge GB,Si-N needs to be measured first.
Then the thermal conductance GB,C of the composite structure 2.6. Steady-State Infrared Thermography
consisting of the sample and the SiNx bridge is measured in the
same manner. The difference between GB,C and GB,Si-N is used The steady-state infrared thermography has been employed
to compute the thermal conductivity of the sample. It should to determine the in-pane thermal conductivity of the free-
be noticed that thermal contact resistance between the sample standing PEDOT:PSS film[67] and PEDOT:PSS/Si nanoparticle
and the two membranes should also be kept as low as possible. composite films.[68] As shown in Figure 7, the thin film sample
Besides the thermal contact resistance, another impor- is suspended over a cavity in an opaque substrate with high
tant issue associated with the MFSD method is the measure- thermal conductivity. The sample is homogenously illuminated
ment sensitivity of the device, which sets the lower limit of the from below by visible light with an absorbed power density QA
thermal conductance that can be measured for a sample.[8] This
value is around 1 nW K−1 for a typical device with the configu-
ration shown in Figure 6a.[50] The measurement sensitivity is
limited by several factors, including the heat transfer through
residual gas molecules, the radiation from the membranes,
and temperature fluctuation of the sample environment.[57]
Recently, significant progresses have been made to increase
the measurement sensitivity, it is now possible to measure the
sample with a low thermal conductance at the order of 10−10
or 10−11 W K−1.[57,58] For the configuration shown in Figure 6b,
the major factor that affects the measurement sensitivity is the
background thermal conductance of the bridge.
Both configurations have been used to measure the thermal
conductivity of polymer or OHT materials in recent years, Figure 7. Schematic of steady-state infrared thermography.
Adv. Electron. Mater. 2019, 1900167 1900167 (8 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
in a vacuum chamber. The wavelength of the light is carefully thermal diffusivity and the thermal conductivity of the bulk
selected to cause strong optical absorptance for generating sub- samples, respectively. The time-domain thermoreflectance
stantial heating. The substrate serves both as a shadow mask method, on the other hand, is suitable for characterizing the
for the illumination and as a heat sink for the heat flow. The samples ranging from the bulks to sub-micrometer thick films.
temperature distribution is recorded by an infrared camera. Table 2 summarizes the major merits and limitations of these
Theoretically, the temperature at a distance r from the cavity three methods.
center is given by[67]
Table 2. Summary of the time-domain methods presented in this review for measuring the thermal transport properties of the OHT materials.
Adv. Electron. Mater. 2019, 1900167 1900167 (9 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Figure 9. a) Schematic of the temporal evolution of the thermoreflectance signal. b) Sketch of the TDTR experimental setup based on an
ultrafast-pulsed-laser system.
Adv. Electron. Mater. 2019, 1900167 1900167 (10 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
conductivity of the sub-micrometer thick thin films, since the could reach 1.8 W m−1 K−1 when σ∥ is around 400–500 S
influences from the substrate can be eliminated.[97] Depending cm−1. The difference might be attributed to the fact that the
on the implementation strategies of the method, the extraction PEDOT samples in both studies have been prepared in dif-
of the thermal conductivity from the measurement data may or ferent manners.
may not require that the volumetric capacity CV of the sample The major limitations of the TDTR method include its
be predetermined.[98–102] In fact, it is possible to measure κ complicated experimental setup and sophisticated data-
and CV simultaneously by using a frequency-dependent TDTR reduction procedure as well as the surface coating. More-
method.[101,102] Though the TDTR method was originally devel- over, to perform the TDTR measurement, the surface of
oped to measure the cross-plane thermal conductivity κ⊥, κ∥ the sample needs to be smooth in order to minimize diffuse
may be measured by orientating the sample in such a way reflections.[14] The requirement may bring challenge to the
that the incident laser beam is parallel to the in-plane direc- sample preparations for certain OHT materials whose sur-
tion of the sample.[69,103] Alternatively, the TDTR method can face roughness might be difficult to control. For example,
be extended to characterize κ∥ by adopting variable spot size undulations or striations can naturally occur on the surface
configuration or beam-offset configuration.[14,96,104] of spin-coated or casted polymer films.[13] The OHT materials
The TDTR method provides a rapid and noncontact way can also possess strong anisotropy in thermal conductivity.
to obtain the thermal conductivity for the materials ranging To properly extract the cross-plane thermal conductivity of
from sub-micrometer thick films to bulks with a high lateral the anisotropic materials, it has been suggested that a large
spatial resolution.[103] It is not involved with the complicated spot size of the laser and high modulation frequency should
sample preparation process as required by the conventional be used to eliminate radial heat transport.[96] The measure-
3ω method that will be discussed Section 4.1. The method ment of the in-plane thermal conductivity based on variable
has attracted much attention in recent years as an important spot size configuration or beam-offset configuration[14,96,104]
technique for characterizing the thermal conductivity of var- seems even more challenging, since more complicated experi-
ious OHT materials including PEDOT:PSS films (100–400 nm mental setup and data reduction models are involved. Jiang
thick in ref. [99], 2.5 µm thick in ref. [100], and 80 nm thick et al. have suggested that it is difficult to measure the in-
in ref. [105]),[69,99,100,105] poly(3-hexylthiophene-2,5-diyl) (P3HT) plane thermal conductivity that is less than 5 W m−1 K−1 by
thin films (77–200 nm thick),[98] graphene based organic nano- using the configurations mentioned above due to limits in
composite (65 nm thick),[106] and tetracyanoquinodimethane the lowest modulation frequency and the smallest spot size
incorporated Cu3(BTC)2 (BTC = benzene tricarboxylate) metal– of the laser used in the current TDTR setups.[14] Although
organic framework films (200 nm thick).[101] With this method, it is technically possible to further reduce the low limit of
Liu et al. have observed that the cross-plane thermal conduc- the modulation frequency and the laser spot size in the
tivity κ⊥ of the drop-cast DMSO-mixed PEDOT:PSS films future,[14] the application of these configurations in meas-
does not change with the in-plane electrical conductivity σ∥ at uring the in-plane thermal conductivity of the OHT materials
≈325 K.[69] The average value of κ⊥ is about 0.3 W m−1 K−1, may be limited at this moment, since many OHT materials
which is consistent with the data obtained by the conven- of interest have an in-plane thermal conductivity less than
tional 3ω technique.[107] In the same study, the authors were 5 W m−1 K−1.[47,61,63,64,67]
able to obtain κ∥ for the PEDOT:PSS films by properly ori-
entating the sample embedded in an epoxy matrix with
respect to the direction of the laser beam.[69] The κ∥ value 3.3. Hot Disk Method
demonstrates an increasing trend with σ∥, which has also
been observed by Weathers et al. via the MFSD technique The hot disk method, also known as transient plane source
for the suspended PEDOT:Tos films prepared by vacuum (TPS) method,[108–111] uses a thin metal strip or a metal disk
vapor phase polymerization and chemical polymerization.[61] (usually with a bifilar spiral pattern as shown in Figure 10a)
The κ∥ value estimated from the published data in ref. [69] as a plane heat source which is sealed between two thin insu-
is ≈1.05 W m−1 K−1 for σ∥ ≈ 520 S cm−1, while κ∥ in ref. [61] lating polymer films. To perform the measurement, the encap-
sulated hot disk is sandwiched between two
identical samples and a constant electrical
current is applied to the heater (Figure 10b).
The temperature rise is monitored by meas-
uring the resistance change of the heater
as a function of time. To ensure that the
sample can be treated as thermally infinite-
size solid, a criterion has been put forward
as a guideline for preparing the sample with
proper dimensions, which states that the
distance from the sensor edge to the nearest
free boundary of the sample should be
larger than 2 α tmax , where α is the thermal
Figure 10. a) The bifilar spiral structure of a hot-disk sensor. b) Cross-sectional view of the diffusivity of the sample and tmax is the max-
experimental setup for the hot-disk method. imum transient recording time.[109,110,112]
Adv. Electron. Mater. 2019, 1900167 1900167 (11 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Theoretically, the temperature increase detected by the bifilar It should be mentioned that Coquard et al. recently
spiral sensor is given by[109] suggested that the accuracy of the TPS method becomes
questionable when it is applied to low density materials
P
∆T (τ ) = D (τ ) (20) (such as polymer foams) due to the fact that their thermal
π 3/2rsenκ inertia is usually two orders of magnitude smaller than
that of the probe.[119] Therefore, if the method is applied to
αt aerogel-based or polymer foam-based OHT materials,[123,124] it
τ= (21)
rsen might be a good idea to crosscheck the results with another
method.[125,126]
where P is the output power of the heater; rsen is the radius
of the sensor; t is the time and τ is a dimensionless quantity
known as characteristic time ratio; D(τ) is a complicated dimen- 4. Frequency-Domain Methods
sionless time-function that can be evaluated numerically.[112]
An advantage of the hot disk method is that both thermal In the frequency-domain methods, the periodic thermal per-
conductivity κ and the thermal diffusivity α can be obtained turbations are often created by the resistive Joule heating or
from a single transient recording.[109,110] Hence, the volumetric laser heating. The thermal responses captured by various
specific heat can be obtained at the same time according to types of sensors (e.g., the resistive thermometer, thermo-
Equation (2). One way of extracting the thermal transport prop- couple, IR sensor, photodiode, acoustic sensor, etc.) are usu-
erties from the raw data is given as follows. α is varied as a trial ally fed into lock-in amplifiers to improve the signal-to-noise
parameter to yield a series of ΔT(τ) versus D(τ) plots. The cor- ratio. In this section, we will review eleven frequency-domain
rect value of α corresponds to the plot that ΔT(τ) varies linearly methods, including eight contact methods and three noncon-
with D(τ). From the slope of the ΔT(τ) versus D(τ) curve (i.e., tact methods. The contact methods include the traditional 3ω
P/π3/2rsenκ), one may immediately obtain the thermal conduc- method, the self-heating 3ω method, the 3ω method based
tivity κ.[112] To improve the accuracy of the result, Equation (20) on the suspended ultrathin dielectric membranes, the scan-
can be further modified by including a number of factors such ning thermal microscopy (SThM)-3ω method, the Ångström’s
as the heat capacity of the probe, the thermal resistance due method, the temperature wave analysis method, the light
to the insulation layer of the sensor, resistance change of the irradiated ac-calorimetric method, and the photo-thermoelec-
sensor with temperature, and so on.[109,112,113] An inconven- tric method, while the frequency-domain thermoreflectance
ience associated with the usual hot disk method is that the method, the modulated photothermal radiometry, and the
measurement requires two pieces of samples each with an photoacoustic method belong to the noncontact methods. The
identical planar surface. Recently, a new technique known as merits and disadvantages of these methods are summarized in
modified transient plane source method has been introduced Table 3.
to address this issue. The new method is based on one-sided
sensor, which allows one to carry out the test with one single
piece of sample.[114–118] 4.1. Traditional 3ω Method
The technique provides a nondestructive and rapid tool to
investigate the thermal conductivity and thermal diffusivity of The traditional 3ω method developed by Cahill and Pohl in
various materials over a wide temperature range.[119] Compared 1980s was initially used to measure the thermal conductivity
with the steady state methods, the size of the sample could be of bulk materials and thick films.[127,128] It was later extended
significantly small and the sample with irregular shape can be to measure the cross-plane thermal conductivity of the thin
accepted. During the test, the temperature rise caused by films on substrate.[129,130] The technique utilizes a thin metal
the transient heating is typically less 5 K,[111] which represents stripe (with a width ranging from several micrometers to
a small temperature perturbation imposed on the sample. This tens of micrometers) deposited on top of the specimen to
will benefit the measurement of the OHT materials which may serve as both a heater and thermometer. For electrically
suffer mechanical instability caused by the uneven thermal conductive specimens, a thin insulation layer is needed to
expansion under a large temperature difference as discussed separate the metal stripe from the sample (Figure 11). An
in Section 2.1. In recent several years, the technique has been ac current at frequency ω in the form of I(t) = I0cos(ωt) is
extensively used for characterizing the thermal conductivity of injected into the heater/sensor to create the Joule heating at
the OHT materials with a thickness ranging from several hun- frequency 2ω
dred micrometers to 1–2 millimeters.[115–118,120–123] Unfortunately,
the method is not applicable to measuring the thermal conduc- I 02Re,H
tivity of sub-micrometer thick films. With this technique, Wu
P(t ) =
2
(1 + cos(2ω t)) (22)
et al. measured the thermal conductivity of pellet samples (σ is
around 0.2–0.8 S cm−1) mechanically pressed from p-toluenesul- where I0 is the amplitude of the ac current and Re,H is the
fonic acid-doped PANI nanowires or nanorods.[117] The thermal resistance of the metal stripe. A temperature oscillation at
conductivity is about 0.3–0.4 W m−1 K−1 at 300 K, which is frequency 2ω (ΔT2ω) arises in response to the ac heating, which
in good agreement with that measured via the absolute in turn brings about a resistance fluctuation at the same fre-
steady-state method for a PANI sample (κ ≈ 0.38 W m−1 K−1 at quency. Multiplication of the 2ω resistance change by the ac
σ = 0.45 S cm−1) prepared in a similar fashion.[27] driving current gives rise to a small 3ω voltage component that
Adv. Electron. Mater. 2019, 1900167 1900167 (12 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Table 3. Summary of the frequency-domain methods presented in this review for measuring the thermal transport properties of the OHT materials.
can be detected by the lock-in amplifier. The 3ω voltage signal tance that heat diffuses within the sample during one cycle of
V3ω is given by[131] the ac power heating period[127]) is much larger than the half-
width of the metal stripe b, the metal strip can be treated as a
1
V3ω = I 0 Re,H ∆T2ω β H (23) line source. The complex temperature rise of the metal stripe
2 can be approximated by[127,129,131]
1 dR
where βH βH = e,H
is the temperature coefficient of Pl 1 1 αs π
Re,H dT ∆Ts = − ln(2ω ) + ln 2 + η − i (24)
resistance (TCR) of the metal stripe. The thermophysical prop- πκ s 2 2 b 4
erties such as the thermal conductivity of the sample can be
retrieved by fitting proper theoretical transport models to the where κs is the thermal conductivity of the solid and b is the
ΔT2ω data.[127,132] half-width of the metal stripe. Pl is the amplitude of the ac
For a semi-infinite isotropic solid with a thermal diffusivity heating power per unit length dissipated by the metal stripe.
αs, if its thermal penetration depth LP = α s /2ω (i.e., the dis- η is a constant. Equation (24) demonstrates that the thermal
Adv. Electron. Mater. 2019, 1900167 1900167 (13 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Pl 1
1/2
∞ sin 2 λ df κ f ,
1/2
∆Tf =
π κ f ,⊥κ f , ∫ λ3
tanh λ
b
κ f ,⊥
dλ (27)
0
Adv. Electron. Mater. 2019, 1900167 1900167 (14 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
quality for isolating the electrically conducting materials to be simultaneously obtain the thermal conductivity and the thermal
measured from the metal stripe.[131,145] To apply the traditional diffusivity (or volumetric capacity) of the sample by performing
3ω method on the OHT materials, more precautions should proper data fittings.
be taken for the sample preparations due to the fact that some On the other hand, when the measurement is carried out at
OHT materials are vulnerable to the relatively high processing the low frequency limit ωγt → 0, V3ω,RMS is almost independent
temperature (e.g., 573 K in a typical plasma enhanced chemical on frequency
vapor deposition (PECVD) process for depositing the insula-
tion layer[13]), and the harsh chemicals used in the lithography
3
4I RMS 2
LRe,s βs
V3ω ,RMS = (29)
procedure. In order to address the issues mentioned above, π κ s As
4
Adv. Electron. Mater. 2019, 1900167 1900167 (15 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
heat sink (i.e., silicon chip) through the suspended membrane The in-plane thermal conductivity is measured for both
(bare dielectric membrane or composite layer composed of the the bare dielectric membrane and the composite structure
dielectric membrane and deposited film). By measuring the 3ω deposited with the film of interest. The thermal conductivity
voltage V3ω across the metal stripe, the 2ω temperature oscil- of the deposited film can be subsequently deduced from
lation ΔT2ω at the location where the stripe is located can be Equation (12).
given by[44] Linseis et al. have applied this method to measure the
temperature-dependent in-plane thermal conductivity of
2V3ω the PEDOT:PSS films with a thickness ranging from sub-
∆T2ω = (30)
I 0 Re,H β H micrometers (100–383 nm)[47] to 15 µm.[48] In ref. [47], the
error bar range associated with the measurement increases
where βH and Re,H are the TCR and electrical resistance of the with decreasing the film thickness and the results for the
metal stripe, respectively. 100 nm thick film demonstrate a high degree of uncertainty.
In order to establish the relationship between V3ω and the The thermal conductivities of the 276 and 383 nm thick
thermophysical properties of the membrane, a 1D transient samples (σ∥ < 20 S cm−1) vary approximately from 0.39 to
heat transfer equation has been solved by imposing proper 0.54 W m−1 K−1 at room temperature, which are comparable
boundary conditions.[159] The amplitude of V3ω in the low fre- with those of the low-σ∥ samples measured by the MFSD[61]
quency limit can be written as[48] and TDTR[69] techniques. On the other hand, the in-plane
thermal conductivity of the 15 µm thick film (σ∥ ≈ 200 S cm−1)
β HRe,H
2
I 03 prepared by drop casting was determined to be ≈0.35 W m−1
V3ω = (31)
4K P K−1 at room temperature,[48] which is smaller than that of the
sub-micrometer thick samples with low σ∥.[47] However, the
2κ dLM value is very close to the κ⊥ value of the PEDOT:PSS films
KP = (32)
w which has been found to be independent on σ∥.[69] The low κ∥
value of the thick film probably reflects the isotropic nature of
where w and LM are the width and length of the mem- the sample.
brane as defined in Figure 5a. κ∥ and d are the in-plane Ushirokita and Tada have taken another approach to com-
thermal conductivity and the thickness of the membrane, bine the 3ω technique with the suspended ultrathin mem-
respectively. brane.[65] Their test platform is based on a double stripe
It has been found that the values of κ∥ calculated from configuration[161–163] as shown in Figure 13. Two metal stripes
Equations (31) and (32) are subject to errors caused by the (denoted as S1 and S2, respectively) are fabricated on top of a
radiation heat loss from the membrane surfaces.[48] Therefore, SiNx membrane window. A 50 nm thick SiO2 layer is depos-
a correction procedure based on two membranes with different ited on top of the stripes as an insulating layer for subsequent
dimensions has been proposed as follows, from which the deposition of the conducting sample. An ac current in the form
thermal conductivity κ∥ and emissivity ε of the membrane are of I(ω) = I0sin(ωt) is passed through the strip S1 to induce the
obtained together.[48] 2ω temperature oscillation. The heating frequency is selected
It can be shown that the thermal conductance associated to be high enough to make the thermal penetration length
with the two membranes due to heat conduction and heat radi- Lp inside the membrane much smaller than the width of the
ation is written as[44,48] membrane window. Under the 1D heat transfer approximation,
the amplitude of the temperature oscillation at S1 (|ΔTS1|) and
PAC,i wi
GM,i = = 2κ dLM,i µ coth µ (33)
∆T2ω ,i 2
8εγ T03
µ2 = (34)
κ d
where PAC,i and |ΔT2ω,i| are the amplitude of the ac heating
power and amplitude of the temperature rise associated with
the ith (i = 1,2) membrane, respectively. γ is the Stefan–Boltz-
mann constant and T0 is the temperature of the heat sink. The
ratio of the thermal conductance of the two membranes is con-
sequently given by[48]
LM,1 coth µ 1
w
GM,1 2
= (35)
LM,2 coth µ 2
GM,2 w
2
Adv. Electron. Mater. 2019, 1900167 1900167 (16 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
the phase shift of the temperature oscillation at S2 (ΔϕS2) are ink-jet printing).[48] On the other hand, to ensure a sufficient
given by[65,161] measurement sensitivity, the thermal conductance of the test
film should be as large as possible when compared to that of
∆TS1 α 1 the bare membrane.[65] In contrary to the conventional 3ω tech-
= (36)
Pl 2 2dκ ω nique, this method should be conducted under the vacuum
environment, which precludes its use in exploring the thermal
and conductivity of the OHT materials in ambient or humidity-con-
trolled conditions.[66]
ω
∆ϕS2 = − L + ϕ0 (37)
α
4.4. SThM-3ω Method
respectively. Here, Pl is the amplitude of the ac heating power
per length. κ∥ and α∥ are the in-plane thermal conductivity and SThM brings a microprobe which acts as thermometer and
thermal diffusivity of the suspended membrane, respectively. heater to be in contact with or in close proximity to the sample
d is the thickness of the membrane. L is the interval between surface.[165–167] By delivering heat locally to the sample and
the two stripes and ϕ0 is a constant. |ΔTS1| is derived from the measuring corresponding thermal responses, it is possible to
measurement of the 3ω voltage. To obtain ΔϕS2, a dc “probe” probe the thermal conductivity of the sample with high lateral
current is passed through S2 to detect its resistance vari- resolution. More specifically, the probe could be a thermo-
ation. Since the temperature rise of S2 has a 2ω component, couple junction[168,169] or a resistive metallic probe.[170–174] In
the resistance of S2 gains a small ac component of the same the latter case, the temperature variation is detected by meas-
frequency. The product of the ac resistance with the “probe” uring the resistance change of the probe. If an ac current with
current gives rise to a small voltage component oscillated at angular frequency ω is used to heat the resistive metallic probe,
2ω.[162,164] The phase shift of this small voltage signal is ΔϕS2 the resulted 2ω temperature oscillation of the probe should be
which can be measured by the lock-in amplifier. In the experi- proportional to its voltage rise in the third harmonic (i.e.,V3ω)
ment, both ΔϕS2 and |ΔTS1| are recorded as a function of ω. α∥ that can be detected by a lock-in amplifier. An experimental
can be obtained by linearly fitting ΔϕS2 as a function of ω . setup adopted by Lefèvre and Volz for performing the SThM-3ω
With known α∥, κ∥ is derived from the slope of the |ΔTS1| versus measurement is schematically shown in Figure 14.[172]
ω−1/2 curve. Based on this configuration and Equation (12), the In order to link the observable quantity V3ω with the thermal
thermal conductivity of poly(methyl methacrylate) (PMMA) and conductivity, a number of analytical and numerical models
several thin conjugated polymer films (200 nm–1 µm thick) have been proposed, which take into accounts of various factors
have been characterized at room temperature by Ushirokita and such as the geometrical/dimensional parameters and mate-
Tada.[65] The room temperature value of the PMMA film was rials properties of the probe, the effective thermal resistance
found to be 0.17 W m−1 K−1, which is in good agreement with between the sample and the probe, the effective coefficient of
the literature result.[88] They also reported a correlation between heat losses by the whole probe surface to its environment, and
κ∥ and the in-plane electrical conductivity σ∥ for the PEDOT:Tos so on.[171,172,175–178]
films, which confirms the previous observations made on the The technique allows relatively fast characterization of a
PEDOT:PSS and PEDOT:Tos samples by applying the MFSD[61] wide variety of sample types (including bulks, films, or nano-
and TDTR[69] techniques. All these studies actually yield close structures)[166,172,176,178–181] and requires only minimal sample
κ∥ values (around 0.5–0.7 W m−1 K−1) for the low-σ∥ samples preparations. It provides a high lateral spatial resolution[166]
(σ∥ < 50 S cm−1).[61,65,69] The κ∥ value for the sample with that is not attainable by other techniques. For the bulk or ther-
σ∥ = 240 S cm−1 was found to be 0.86 W m−1 K−1 in ref. [65], mally thick films, the thermal conductivity can be derived by
which is coincident with the result (≈0.82 W m−1 K−1 at treating the sample as a semi-infinite medium. If the film on
σ∥ ≈ 250 S cm−1) obtained via the TDTR technique[69] and is a substrate is thermally thin and heat spreading in the in-plane
about 28% smaller than the value interpolated from the MFSD directions is non-negligible, a 2D heat transfer model for a
data (≈1.2 W m−1 K−1).[61] multilayered system should be employed to obtain the thermal
To summarize, the 3ω method based on the suspended
ultrathin membrane can be used to measure the in-plane
thermal conductivity of the films with various thicknesses of
sub-micrometers to more than ten micrometers by eliminating
the substrate effect. The chip-based measurement platform
has been designed to measure the Seebeck coefficient and
the electrical conductivity of the sample at the same time.[47]
Unlike the conventional 3ω method, this method utilizes the
prefabricated insulation layers and metal stripes, which is an
advantage for characterizing the OHT materials that may not
be able to sustain hostile processing conditions. The test-chip
is designed to allow facile sample deposition by PVD methods
(e.g., thermal evaporation and magnetron sputtering) and Figure 14. Schematic of the experimental setup for the SThM-3ω
solution-based techniques (e.g., spin coating, drop casting, or measurement adopted by Lefèvre and Volz.[172]
Adv. Electron. Mater. 2019, 1900167 1900167 (17 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
c onductivity.[176] In recent years, the method has been used to characterizing the thermal transport properties of the bulk
measure the thermal conductivity of several inorganic nano- thermoelectric materials, the Ångström’s method still remains
structures such as the nanowires of Si,[178,179] SiGe,[179] and useful due to its simplicity and versatility. Recently, Shin
Bi2Te3,[180] and some OHT materials including P3HT nanow- et al. have used this method to characterize the Bi0.5Sb1.5Te3
ires (100 µm long and 120–350 nm in diameter),[181,182] doped and Bi2Te2.7Se0.3 based samples prepared by a combination of
and undoped poly[N-9′-heptadecanyl-2,7-carbazole-alt-5,5-(4′,7′- screen-printing and hot pressing procedures.[191] Tan et al. have
di-2-thienyl-2′,1′,3′-benzothiadizole)] (PCDTBT) films (several used this method to measure the thermal diffusivity of a kind
micrometers thick),[176,183] and so on. In order to evaluate the of carbon nanoparticle hybrid aerogel with extremely low den-
reliability of the SThM-3ω technique, Wilson et al. carried out sity and high porosity.[124] The derived thermal conductivity for
characterizations on calibrated samples, including PANI com- the aerogel sample was about 0.02–0.03 W m−1 K−1, which is
posites mixed with 5% and 7% graphene nanoplatelets,[176] comparable with that of a 3D CNT network measured via the
which have been investigated previously by the comparative laser flash and hot disk methods at low air pressure (20 Pa).[126]
method.[184] The measured thermal conductivities were found The main limitation of the Ångström’s method is that it does
to be 0.47 and 0.68 W m−1 K−1, respectively, which are in good not directly yield the value of κ, although certain attempt has
agreement with the calibrated data. Since the SThM measure- been made to modify the method in order to achieve a simulta-
ments are usually performed in air condition, the method is neous measurement of α and κ.[186]
in principle applicable to the investigations of the humidity
dependent thermal conductivity[66] of the OHT materials. On
the other hand, in order to extract the value of the thermal con- 4.6. Temperature Wave Analysis Method
ductivity, the sophisticated models and extensive calibrations
are generally needed. The accuracy of the results relies on how The temperature wave analysis (TWA) method is a contact
those effective parameters associated with the theoretical model method for measuring the cross-plane thermal diffusivity of the
have been obtained.[171,172,175–178,185] sample.
In this method, the sample of thickness d is sandwiched
between one heater and one resistive temperature sensor[192–194]
4.5. Ångström’s Method (Figure 16). The heater creates a thermal wave at the front sur-
face of the sample by the ac Joule heating at frequency f mod
The Ångström’s method is a widely accepted method for deter- ( f mod = ω mod /2π ) , which diffuses toward the rear surface
mining the thermal diffusivity of the solid materials.[7,186,187] In along the thickness direction. The temperature oscillation at
this method, the sample is heated periodically at one end, while the rear surface is measured by detecting the resistance varia-
the temperatures are measured at two locations along its length tion of the sensor with the lock-in technique. If the sample is
(Figure 15). When the propagation of the thermal wave reaches thermally thick ( i.e., d π f mod /α 1), the phase lag of the tem-
a steady state, temperature variations are recorded to calculate perature variation between the front surface and the rear sur-
the thermal diffusivity by[188] face, Δϕ, is approximated by[193,194]
π f mod L2 ω mod π
α= (38) ∆ϕ = − d− (39)
∆ϕ ln ( ∆Th / ∆Tc ) 2α 4
where f mod is the heating frequency. L is the distance between The thermal diffusivity α is determined from the slope of Δϕ
xh and xc where two temperature sensors are located. |ΔTh| and versus ω mod plot.
|ΔTc| are the measured amplitudes of the temperature wave at The TWA method only needs simple instrumentation and
xh and xc, respectively, while Δϕ is the phase lag between the minimum sample handlings. The major limitation of the
two locations.
The Ångström’s method is a traditional method for obtaining
the thermal diffusivity of the rod-like or stripe-like sample.
The implementation of the method is relatively easy and the
sample preparations are simple. It has been used to measure
the thermal diffusivity of the thermoelectric materials since
1960s.[188–190] Though many choices are available today for
Figure 16. The experimental setup for the temperature wave analysis
Figure 15. The experimental configuration for Ångström’s method. method.
Adv. Electron. Mater. 2019, 1900167 1900167 (18 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Adv. Electron. Mater. 2019, 1900167 1900167 (19 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Adv. Electron. Mater. 2019, 1900167 1900167 (20 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
Adv. Electron. Mater. 2019, 1900167 1900167 (21 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
graphene oxide composite,[227] and PANI/silver composite,[228] Experimentally, the PA signal can be recorded with the front-
as well as the transverse thermal diffusivity of millimeter-sized side detection configuration (Figure 20a) and the rear-side
organic semiconductors (up to 600 µm thick in the transverse detection configuration (Figure 20b). In order to eliminate the
direction).[229] The thermal conductivity of the PANI sample frequency dependence of the instrumental responses and non-
could be used as a benchmark for verifying the reliability essential signal-related proportionality factors, normalization
of the measurements. With the PTR technique. Islam of the PA signals is typically needed. For example, in order to
et al.[227] reported a room temperature thermal conductivity of measure the thermal diffusivity of the solids, several normaliza-
≈0.44 W m−1 K−1 for a PANI sample with an electrical conduc- tion procedures have been proposed. Pessoa et al. performed
tivity of less than 0.1 S cm−1, which is consistent with that of the normalization by taking the signal’s ratio in the rear-side
the low-σ sample (0.46 W m−1 K−1 at σ = 0.48 S cm−1) meas- and front-side detection configurations for one-single layered
ured by the comparative method by Abad et al.[184] The result sample at a single modulation frequency.[234] Balderas-López
is also comparable with the cross-plane thermal conductivity took the signal ratio from the investigated sample and a refer-
of 0.59 W m−1 K−1 measured by the conventional 3ω tech- ence sample which were both measured in the rear-side detec-
nique for a 5 µm thick PANI film with an electrical conduc- tion configuration.[235] Alternatively, the test sample can be
tivity of 93 S cm−1.[13] measured in the rear-side detection configuration, while the
reference sample in the front-side configuration.[236]
In principle, the thermal conductivity of the bulk sample can
4.10. Photoacoustic (PA) Method be determined by, for instance, putting the sample in thermally
contacted with a backing materials whose thermophysical prop-
In the PA method,[230–232] an intensity modulated radiation erties have been known.[237] On the other hand, by performing
heats a sample located in a closed gas cell (PA cell). A gas layer the measurements at a frequency up to 20 kHz and applying
in contact with the sample surface is heated periodically and the advanced multilayer model[233] for data reduction, the tech-
is driven to expand and contract like a thermal piston. The nique has been extended to measure the thermal conductivity
periodic change of pressure produces an acoustic wave that is of the thin film with sub-micrometer thickness.[238,239]
detected by a microphone, which can be used to derive the ther- The PA technique is a noncontact method featured with
mophysical properties of the sample. a unique and inexpensive detecting system. It can actually
The PA effect was first theoretically explained by Rosencwaig measure the cross-plane thermal conductivity of the films and
and Gersho who revealed that the PA signal (or the variations the bulk samples with rough surfaces.[238] The toleration of the
of the gas pressure) is given by[230] method for the surface conditions may allow one to measure
the thermal conductivity of those OHT materials whose sur-
γ g µgPg,0θ i(ω mod t − π4 ) faces are so rough that the use of the techniques such as the
∆Pg (t ) = e (50)
2LgT0 conventional 3ω method and the TDTR (or FDTR) method
should be precluded. A disadvantage of this technique is that
where Pg,0 is the static pressure and ω mod is the angular modu- it needs extensive calibration and normalization procedures
lation frequency. μg(µg = 2α g /ω mod , where αg is the thermal to eliminate the PA signal distortions caused by the electronic
diffusivity of the gas) and γg are the thermal diffusion depth circuitry.[234–236,238,240] Additionally, the surface coating for the
and the ratio of specific heats of the gas, respectively. T0 is sample is needed to enhance the laser power absorption, which
the average temperature. Lg is the gas column length. θ is a will limit the use the sample for other purposes. To date, the
complex expression for the temperature oscillation at the gas/ method has been utilized to measure the thermal conductivity
sample interface. of several OHT materials including undoped PCDTBT film,[183]
The theory was soon extended by McDonald and Wetsel Bi2Te3 nanowire array/SU-8 epoxy resin nanocomposite,[241]
who suggested that mechanical vibration of the sample sur- poly(3-octylthiophene-2,5 diyl) (P3OT) film and MWCNT/P3OT
face due to thermal expansion exerts work and subsequently composites,[242] polythiophene film and nanofibre arrays,[60]
causes additional temperature change on the gas medium.[231] as well as P3HT nanotube array,[243] etc. The thickness of the
Recently, Hu et al. proposed a generalized expression for the OHT samples usually ranges from several micrometers[183] to
PA effect in a multilayer system.[233] several tens of micrometers.[241–243] Among these materials,
Figure 20. Schematic sketch of the photoacoustic measurement setup. a) Front-side detection configuration. b) Rear-side detection configuration.
Adv. Electron. Mater. 2019, 1900167 1900167 (22 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
the P3OT film was found to have a thermal conductivity coefficient in order to achieve a high measurement accuracy,
of 0.21 W m−1 K−1,[242] which is in agreement with the values although the samples with a Seebeck coefficient as low as
reported for structurally similar polymers such as P3HT.[98,143] 5.8 µV K−1 were still measurable.[244]
The value of the undoped PCDTBT film was measured to be
0.20 W m−1 K−1, which is in reasonable agreement with the result
obtained from the SThM-3ω measurement (0.25 W m−1 K−1).[183] 5. Summary
This paper presents a brief review on the commonly used and
4.11. Photo-Thermoelectric Method newly developed techniques that can be used for characterizing
the thermal conductivity of the OHT materials. Each individual
The photo-thermoelectric method was proposed by Kuria- technique mentioned above has its own merits and limitations.
kose et al.[244,245] who used the transient photo-thermoelectric A suitable method is selected primarily based on the factors such
response of the sample as a probe to explore its own thermo- as the characteristics of the sample (e.g., dimensionality, size,
physical properties. The experimental setup is schematically morphology, mechanical properties, optical properties, thermal
illustrated in Figure 21. The top and bottom surfaces of the ther- properties, etc.), the temperature range of the measurement,
moelectric sample are covered with thin metal electrodes for the required measurement accuracy, and so on. When a specific
extracting the thermoelectric voltage. The bottom of the sample method has been selected, it is crucial to understand its theo-
is thermally contact with a substrate with known thermal dif- retical backgrounds and possible error sources. Careful sample
fusivity and thermal effusivity. The top surface of the sample is preparations or instrumentation calibrations are also important
heated by a modulated laser beam. At a fixed modulation fre- for carrying out reliable measurements. To further ensure the
quency, the thermoelectric signal ΔVth obtained with presence accuracy and creditability of the results, some researchers prefer
of the substrate is normalized by the signal ΔVth,0 obtained with to crosscheck their results by different techniques.[126,183,244]
air as the backing material[244] With the continuing development of the novel OHT materials,
some measurement techniques may receive more attention in
∆Vth the future. For example, the characterization of the anisotropic
Vn = (51)
∆Vth,0 thermal conductivity has become an important topic in the OHT
community. The combined use of the relevant tools (such as the
where Vn is the normalized signal. By fitting Vn to a theoretical various 3ω techniques, the thermoreflectance techniques, the
model,[244] both thermal diffusivity α and thermal effusivity e light irradiated ac-calorimetric method, the modulated photo-
of the sample can be calculated. By applying Equation (4), the thermal radiometry, etc.) might be helpful for tackling the issue.
thermal conductivity κ of the sample is determined. On the other hand, it has been shown in many studies that the
The method provides a simple and rapid way to measure thermal conductivity can be simultaneously measured with the
the thermal diffusivity and thermal conductivity for the bulk Seebeck coefficient and the electrical conductivity on the same
thermoelectric sample. The OHT materials such as 6.6 wt% sample.[22,61,157] This feature is highly desirable, since calculating
oxidized SWNTs/doped PANI composite have been tested the ZT value from the TE properties measured on different sam-
by this method. Its thermal conductivity was measured to be ples may cause errors. This is due to the fact that the samples
0.83 W m−1 K−1, which agrees well with that obtained from the are usually more or less inhomogeneous even when they are
conventional photothermal radiometry.[244] Another interesting chosen from the same batch.[34]
feature of this method is that it can be adapted to measure the
thermal conductivity of thermoelectric liquids, such as tetrado-
decylammonium nitrate salt/1-octanol mixture.[245] It should be Acknowledgements
noted that the sample should possess a relatively large Seebeck
The authors would like to acknowledge the financial supports from
the National Natural Science Foundation of China under grant nos.
51573190 and 61176083, and the State Key Development Program for
Basic Research of China under grant no. 2011CB932801.
Conflict of Interest
The authors declare no conflict of interest.
Keywords
hybrid thermoelectric materials, organic thermoelectric materials,
thermal conductivity measurement
Adv. Electron. Mater. 2019, 1900167 1900167 (23 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
[1] Q. Zhang, Y. Sun, W. Xu, D. Zhu, Adv. Mater. 2014, 26, 6829. [37] A. M. Marconnet, N. Yamamoto, M. A. Panzer, B. L. Wardle,
[2] Y. Chen, Y. Zhao, Z. Liang, Energy Environ. Sci. 2015, 8, 401. K. E. Goodson, ACS Nano 2011, 5, 4818.
[3] C. Gao, G. Chen, Compos. Sci. Technol. 2016, 124, 52. [38] D. T. Verebelyi, Rev. Sci. Instrum. 1997, 68, 2494.
[4] B. Russ, A. Glaudell, J. J. Urban, M. L. Chabinyc, R. A. Segalman, [39] H. Yoshino, G. C. Papavassiliou, K. Murata, J. Therm. Anal. Cal-
Nat. Rev. Mater. 2016, 1, 16050. orim. 2008, 92, 457.
[5] G. Chen, W. Xu, D. Zhu, J. Mater. Chem. C 2017, 5, 4350. [40] H. Yoshino, G. C. Papavassiliou, K. Murata, Synth. Met. 2009, 159,
[6] A. Zevalkink, D. M. Smiadak, J. L. Blackburn, A. J. Ferguson, 2387.
M. L. Chabinyc, O. Delaire, J. Wang, K. Kovnir, J. Martin, [41] H. Yoshino, H. Aizawa, K. Kuroki, G. C. Anyfantis,
L. T. Schelhas, T. D. Sparks, S. D. Kang, M. T. Dylla, G. J. Snyder, G. C. Papavassiliou, K. Murata, Phys. B 2010, 405, S79.
B. R. Ortiz, E. S. Toberer, Appl. Phys. Rev. 2018, 5, 021303. [42] F. Völklein, Thin Solid Films 1990, 188, 27.
[7] R. Taylor, in CRC Handbook of Thermoelectrics (Ed: D. M. Rowe), [43] F. Völklein, E. Kessler, Measurement 1987, 5, 38.
CRC Press, Boca Raton 1996, Ch. 16. [44] F. Völklein, H. Reith, A. Meier, Phys. Status Solidi A 2013, 210,
[8] Y. Liu, M. Zhang, A. Ji, F. Yang, X. Wang, RSC Adv. 2016, 6, 106.
48933. [45] V. Linseis, F. Völklein, H. Reith, P. Woias, K. Nielsch, J. Mater. Res.
[9] S. Imai, K. Adachi, IEEE Trans. Electr. Electron. Eng. 2015, 10, 101. 2016, 31, 3196.
[10] C. Stacey, A. J. Simpkin, R. N. Jarrett, Int. J. Thermophys. 2016, 37, [46] V. Linseis, F. Völklein, H. Reith, P. Woias, K. Nielsch, J. Electron.
107. Mater. 2018, 47, 3203.
[11] S.-Y. Fu, in Polymers at Cryogenic Temperatures (Eds: S. Kalia, [47] V. Linseis, Z. M. Hassan, H. Reith, J. Garcia, K. Nielsch,
S.-Y. Fu), Springer, Berlin, Heidelberg 2013, Ch. 2. H. Baumgart, E. Redel, P. Woias, Phys. Status Solidi A 2018, 215,
[12] C. Tseng, M. Yamaguchi, T. Ohmori, Cryogenics 1997, 37, 305. 1700930.
[13] P. B. Kaul, K. A. Day, A. R. Abramson, J. Appl. Phys. 2007, 101, [48] V. Linseis, F. Völklein, H. Reith, K. Nielsch, P. Woias, Rev. Sci.
083507. Instrum. 2018, 89, 015110.
[14] P. Q. Jiang, X. Qian, R. G. Yang, J. Appl. Phys. 2018, 124, 161103. [49] F. Völklein, E. Kessler, Phys. Status Solidi (b) 1987, 143, 121.
[15] B. Abad, D. A. Borca-Tasciuc, M. S. Martin-Gonzalez, Renewable [50] L. Shi, D. Y. Li, C. H. Yu, W. Y. Jang, D. Kim, Z. Yao, P. Kim,
Sustainable Energy Rev. 2017, 76, 1348. A. Majumdar, J. Heat Transfer 2003, 125, 881.
[16] T. M. Tritt, D. Weston, in Thermal Conductivity, Physics of Solids and [51] J. Kim, E. Fleming, Y. Zhou, L. Shi, J. Phys. D: Appl. Phys. 2018, 51,
Liquids (Ed: T. M. Tritt), Springer, Boston, MA 2004, Ch. 2.1, p. 187. 103002.
[17] A. Rudajevova, M. Svantner, D. Vasylyev, O. Musil, V. Lang, Int. J. [52] B. L. Zink, F. Hellman, Solid State Commun. 2004, 129, 199.
Thermophys. 2006, 27, 1241. [53] R. Sultan, A. D. Avery, G. Stiehl, B. L. Zink, J. Appl. Phys. 2009, 105,
[18] J. Sebek, E. Santava, J. Phys.: Conf. Ser. 2009, 150, 012044. 043501.
[19] K. A. Borup, J. de Boor, H. Wang, F. Drymiotis, F. Gascoin, X. Shi, [54] R. Sultan, A. D. Avery, J. M. Underwood, S. J. Mason, D. Bassett,
L. Chen, M. I. Fedorov, E. Müller, B. B. Iversen, G. J. Snyder, Energy B. L. Zink, Phys. Rev. B 2013, 87, 214305.
Environ. Sci. 2015, 8, 423. [55] A. Mavrokefalos, M. T. Pettes, F. Zhou, L. Shi, Rev. Sci. Instrum.
[20] A. Kovalsky, L. Wang, G. T. Marek, C. Burda, J. S. Dyck, J. Phys. 2007, 78, 034901.
Chem. C 2017, 121, 3228. [56] B. L. Zink, A. D. Avery, R. Sultan, D. Bassett, M. R. Pufall, Solid
[21] A. Pisoni, J. Jacimovic, O. S. Barisic, M. Spina, R. Gaal, L. Forro, State Commun. 2010, 150, 514.
E. Horvath, J. Phys. Chem. Lett. 2014, 5, 2488. [57] A. Weathers, K. Bi, M. T. Pettes, L. Shi, Rev. Sci. Instrum. 2013, 84,
[22] V. K. Zaitsev, M. I. Fedorov, E. A. Gurieva, I. S. Eremin, 084903.
P. P. Konstantinov, A. Y. Samunin, M. V. Vedernikov, Phys. Rev. B [58] M. C. Wingert, Z. C. Chen, S. Kwon, J. Xiang, R. Chen, Rev. Sci.
2006, 74, 045207. Instrum. 2012, 83, 024901.
[23] ASTM, ASTM D5470–06, ASTM International, West Consho- [59] A. Weathers, S. Matsushita, M. T. Pettes, M. Goh, K. Akagi, L. Shi,
hocken, PA, USA 2006. J. Appl. Phys. 2013, 114, 194302.
[24] C. Yu, Y. S. Kim, D. Kim, J. C. Grunlan, Nano Lett. 2008, 8, 4428. [60] V. Singh, T. L. Bougher, A. Weathers, Y. Cai, K. Bi, M. T. Pettes,
[25] H. Huang, C. H. Liu, Y. Wu, S. Fan, Adv. Mater. 2005, 17, 1652. S. A. McMenamin, W. Lv, D. P. Resler, T. R. Gattuso, D. H. Altman,
[26] D. Kim, Y. Kim, K. Choi, J. C. Grunlan, C. H. Yu, ACS Nano 2010, K. H. Sandhage, L. Shi, A. Henry, B. A. Cola, Nat. Nanotechnol.
4, 513. 2014, 9, 384.
[27] C. Meng, C. Liu, S. Fan, Adv. Mater. 2010, 22, 535. [61] A. Weathers, Z. U. Khan, R. Brooke, D. Evans, M. T. Pettes,
[28] C. Yu, K. Choi, L. Yin, J. C. Grunlan, ACS Nano 2011, 5, 7885. J. W. Andreasen, X. Crispin, L. Shi, Adv. Mater. 2015, 27, 2101.
[29] G. P. Moriarty, K. Briggs, B. Stevens, C. Yu, J. C. Grunlan, Energy [62] A. D. Avery, B. H. Zhou, J. Lee, E.-S. Lee, E. M. Miller, R. Ihly,
Technol. 2013, 1, 265. D. Wesenberg, K. S. Mistry, S. L. Guillot, B. L. Zink, Y.-H. Kim,
[30] B. M. Zawilski, R. T. Littleton, T. M. Tritt, Rev. Sci. Instrum. 2001, J. L. Blackburn, A. J. Ferguson, Nat. Energy 2016, 1, 16033.
72, 1770. [63] B. A. MacLeod, N. J. Stanton, I. E. Gould, D. Wesenberg, R. Ihly,
[31] C. Wan, X. Gu, F. Dang, T. Itoh, Y. Wang, H. Sasaki, M. Kondo, Z. R. Owczarczyk, K. E. Hurst, C. S. Fewox, C. N. Folmar,
K. Koga, K. Yabuki, G. J. Snyder, R. Yang, K. Koumoto, Nat. Mater. K. Holman Hughes, B. L. Zink, J. L. Blackburn, A. J. Ferguson,
2015, 14, 622. Energy Environ. Sci. 2017, 10, 2168.
[32] H. Imai, Y. Shimakawa, Y. Kubo, Phys. Rev. B 2001, 64. [64] K. Choi, S. L. Kim, S. I. Yi, J. H. Hsu, C. Yu, ACS Appl. Mater. Inter-
[33] T. Varghese, C. Hollar, J. Richardson, N. Kempf, C. Han, faces 2018, 10, 23891.
P. Gamarachchi, D. Estrada, R. J. Mehta, Y. Zhang, Sci. Rep. 2016, [65] H. Ushirokita, H. Tada, Chem. Lett. 2016, 45, 735.
6, 33135. [66] G.-H. Kim, J. Kim, K. P. Pipe, Appl. Phys. Lett. 2016, 108, 093301.
[34] T. M. Tritt, in Thermoelectrics Handbook: Macro to Nano (Ed: [67] A. Greppmair, B. Stoib, N. Saxena, C. Gerstberger, P. Muller-
D. M. Rowe), CRC Press, Boca Raton 2006, Ch. 23, p. 23. Buschbaum, M. Stutzmann, M. S. Brandt, Rev. Sci. Instrum. 2017,
[35] ASTM, ASTM E1225–09, ASTM International, West Consho- 88, 044903.
hocken, PA, USA 2009. [68] N. Saxena, M. Čorić, A. Greppmair, J. Wernecke, M. Pflüger,
[36] ASTM, ASTM E1530–11, ASTM International, West Consho- M. Krumrey, M. S. Brandt, E. M. Herzig, P. Müller-Buschbaum,
hocken, PA, USA 2011. Adv. Electron. Mater. 2017, 3, 1700181.
Adv. Electron. Mater. 2019, 1900167 1900167 (24 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
[69] J. Liu, X. Wang, D. Li, N. E. Coates, R. A. Segalman, D. G. Cahill, [102] J. Liu, J. Zhu, M. Tian, X. Gu, A. Schmidt, R. Yang, Rev. Sci.
Macromolecules 2015, 48, 585. Instrum. 2013, 84, 034902.
[70] Q. Wei, M. Mukaida, K. Kirihara, T. Ishida, ACS Macro Lett. 2014, [103] X. Zheng, D. G. Cahill, J. C. Zhao, Adv. Eng. Mater. 2005, 7, 622.
3, 948. [104] J. P. Feser, J. Liu, D. G. Cahill, Rev. Sci. Instrum. 2014, 85, 104903.
[71] W. J. Parker, R. J. Jenkins, C. P. Butler, G. L. Abbott, J. Appl. Phys. [105] M. Scholdt, H. Do, J. Lang, A. Gall, A. Colsmann, U. Lemmer,
1961, 32, 1679. J. D. Koenig, M. Winkler, H. Boettner, J. Electron. Mater. 2010, 39,
[72] P. Schoderböck, H. Klocker, L. S. Sigl, G. Seeber, Int. J. Thermo- 1589.
phys. 2009, 30, 599. [106] T. Juntunen, H. Jussila, M. Ruoho, S. Liu, G. Hu, T. Albrow-Owen,
[73] V. Khuu, M. Osterman, A. Bar-Cohen, M. Pecht, IEEE Trans. L. W. T. Ng, R. C. T. Howe, T. Hasan, Z. Sun, I. Tittonen, Adv.
Compon., Packag., Manuf. Technol. 2011, 1, 1015. Funct. Mater. 2018, 28, 1800480.
[74] D. Zhao, X. Qian, X. Gu, S. A. Jajja, R. Yang, J. Electron. Packag. [107] G. H. Kim, L. Shao, K. Zhang, K. P. Pipe, Nat. Mater. 2013, 12,
2016, 138, 040802. 719.
[75] H. Yan, N. Sada, N. Toshima, J. Therm. Anal. Calorim. 2002, 69, [108] S. E. Gustafsson, E. Karawacki, M. A. Chohan, J. Phys. D: Appl.
881. Phys. 1986, 19, 727.
[76] W. Q. Ao, L. Wang, J. Q. Li, F. Pan, C. N. Wu, J. Electron. Mater. [109] S. E. Gustafsson, Rev. Sci. Instrum. 1991, 62, 797.
2011, 40, 2027. [110] M. Gustavsson, E. Karawacki, S. E. Gustafsson, Rev. Sci. Instrum.
[77] K. Zhang, Y. Zhang, S. Wang, Sci. Rep. 2013, 3, 3448. 1994, 65, 3856.
[78] W. Zheng, P. Bi, H. Kang, W. Wei, F. Liu, J. Shi, L. Peng, Z. Wang, [111] V. Bohac, M. K. Gustavsson, L. Kubicar, S. E. Gustafsson, Rev. Sci.
R. Xiong, Appl. Phys. Lett. 2014, 105, 023901. Instrum. 2000, 71, 2452.
[79] Y. Liu, Z. Song, Q. Zhang, Z. Zhou, Y. Tang, L. Wang, J. Zhu, [112] Y. He, Thermochim. Acta 2005, 436, 122.
W. Luo, W. Jiang, RSC Adv. 2015, 5, 45106. [113] Y. Li, C. Shi, J. Liu, E. Liu, J. Shao, Z. Chen, D. J. Dorantes-Gonzalez,
[80] Q. Yao, Q. Wang, L. Wang, L. Chen, Energy Environ. Sci. 2014, 7, X. Hu, Meas. Sci. Technol. 2014, 25, 015006.
3801. [114] J. Cha, J. Seo, S. Kim, J. Therm. Anal. Calorim. 2012, 109, 295.
[81] T. Baba, N. Taketoshi, T. Yagi, Jpn. J. Appl. Phys. 2011, 50, 11RA01. [115] Y. Sun, P. Sheng, C. Di, F. Jiao, W. Xu, D. Qiu, D. Zhu, Adv. Mater.
[82] C. A. Paddock, G. L. Eesley, J. Appl. Phys. 1986, 60, 285. 2012, 24, 932.
[83] P. M. Norris, A. P. Caffrey, R. J. Stevens, J. M. Klopf, J. T. McLeskey, [116] J. Wu, Y. Sun, W.-B. Pei, L. Huang, W. Xu, Q. Zhang, Synth. Met.
A. N. Smith, Rev. Sci. Instrum. 2003, 74, 400. 2014, 196, 173.
[84] P. E. Hopkins, J. R. Serrano, L. Phinney, S. P. Kearney, [117] J. Wu, Y. Sun, W. Xu, Q. Zhang, Synth. Met. 2014, 189, 177.
T. W. Grasser, C. T. Harris, J. Heat Transfer 2010, 132, 10. [118] M. Aghelinejad, S. N. Leung, Composites, Part B 2018, 145, 100.
[85] D. G. Cahill, Rev. Sci. Instrum. 2004, 75, 5119. [119] R. Coquard, E. Coment, G. Flasquin, D. Baillis, Int. J. Therm. Sci.
[86] J. Zhu, D. Tang, W. Wang, J. Liu, K. W. Holub, R. Yang, J. Appl. 2013, 65, 242.
Phys. 2010, 108, 094315. [120] S. P. Ashby, J. García-Cañadas, G. Min, Y. Chao, J. Electron. Mater.
[87] O. W. Käding, H. Skurk, K. E. Goodson, Appl. Phys. Lett. 1994, 65, 2013, 42, 1495.
1629. [121] B. Dörling, J. D. Ryan, J. D. Craddock, A. Sorrentino,
[88] D. Chu, M. Touzelbaev, K. E. Goodson, S. Babin, R. F. Pease, A. E. Basaty, A. Gomez, M. Garriga, E. Pereiro, J. E. Anthony,
J. Vac. Sci. Technol., B: Microelectron. Nanometer Struct.–Process., M. C. Weisenberger, A. R. Goñi, C. Müller, M. Campoy-Quiles, Adv.
Meas., Phenom. 2001, 19, 2874. Mater. 2016, 28, 2782.
[89] D. J. Yang, Q. Zhang, G. Chen, S. F. Yoon, J. Ahn, S. G. Wang, [122] D. Kiefer, L. Yu, E. Fransson, A. Gómez, D. Primetzhofer,
Q. Zhou, Q. Wang, J. Q. Li, Phys. Rev. B 2002, 66, 165440. A. Amassian, M. Campoy-Quiles, C. Müller, Adv. Sci. 2017, 4,
[90] M. A. Panzer, G. Zhang, D. Mann, X. Hu, E. Pop, H. Dai, 1600203.
K. E. Goodson, J. Heat Transfer 2008, 130, 052401. [123] R. Kroon, J. D. Ryan, D. Kiefer, L. Yu, J. Hynynen, E. Olsson,
[91] A. Neubrand, J. Dadda, E. Mueller, S. Perlt, T. Höche, J. Electron. C. Müller, Adv. Funct. Mater. 2017, 27, 1704183.
Mater. 2013, 42, 2165. [124] D. Tan, J. Zhao, C. Gao, H. Wang, G. Chen, D. Shi, ACS Appl.
[92] R. J. Stoner, H. J. Maris, Phys. Rev. B 1993, 48, 16373. Mater. Interfaces 2017, 9, 21820.
[93] H. E. Elsayed-Ali, T. B. Norris, M. A. Pessot, G. A. Mourou, Phys. [125] O. Almanza, M. A. Rodriguez-Perez, J. A. De Saja, J. Polym. Sci.,
Rev. Lett. 1987, 58, 1212. Part B: Polym. Phys. 2004, 42, 1226.
[94] R. W. Schoenlein, W. Z. Lin, J. G. Fujimoto, G. L. Eesley, Phys. Rev. [126] J. K. Chen, X. C. Gui, Z. W. Wang, Z. Li, R. Xiang, K. L. Wang,
Lett. 1987, 58, 1680. D. H. Wu, X. G. Xia, Y. F. Zhou, Q. Wang, Z. K. Tang, L. D. Chen,
[95] R. B. Wilson, B. A. Apgar, L. W. Martin, D. G. Cahill, Opt. Express ACS Appl. Mater. Interfaces 2012, 4, 81.
2012, 20, 28829. [127] D. G. Cahill, R. O. Pohl, Phys. Rev. B 1987, 35, 4067.
[96] A. J. Schmidt, X. Chen, G. Chen, Rev. Sci. Instrum. 2008, 79, [128] D. G. Cahill, Rev. Sci. Instrum. 1990, 61, 802.
114902. [129] S. M. Lee, D. G. Cahill, J. Appl. Phys. 1997, 81, 2590.
[97] W. S. Capinski, H. J. Maris, T. Ruf, M. Cardona, K. Ploog, [130] S. M. Lee, D. G. Cahill, R. Venkatasubramanian, Appl. Phys. Lett.
D. S. Katzer, Phys. Rev. B 1999, 59, 8105. 1997, 70, 2957.
[98] J. C. Duda, P. E. Hopkins, Y. Shen, M. C. Gupta, Appl. Phys. Lett. [131] S. T. Huxtable, Ph.D. Thesis, University of California, Berkeley,
2013, 102, 251912. 2002.
[99] D. Beretta, A. J. Barker, I. Maqueira-Albo, A. Calloni, G. Bussetti, [132] T. Borca-Tasciuc, A. R. Kumar, G. Chen, Rev. Sci. Instrum. 2001, 72,
G. Dell’Erba, A. Luzio, L. Duo, A. Petrozza, G. Lanzani, M. Caironi, 2139.
ACS Appl. Mater. Interfaces 2017, 9, 18151. [133] Y. S. Ju, K. Kurabayashi, K. E. Goodson, Thin Solid Films 1999, 339,
[100] A. K. K. Kyaw, T. A. Yemata, X. Wang, S. L. Lim, W. S. Chin, 160.
K. Hippalgaonkar, J. Xu, Macromol. Mater. Eng. 2018, 303, [134] K. C. See, J. P. Feser, C. E. Chen, A. Majumdar, J. J. Urban,
1700429. R. A. Segalman, Nano Lett. 2010, 10, 4664.
[101] K. J. Erickson, F. Leonard, V. Stavila, M. E. Foster, C. D. Spataru, [135] D. Rodin, S. K. Yee, Rev. Sci. Instrum. 2017, 88, 014902.
R. E. Jones, B. M. Foley, P. E. Hopkins, M. D. Allendorf, A. A. Talin, [136] J. Sun, M. L. Yeh, B. J. Jung, B. Zhang, J. Feser, A. Majumdar,
Adv. Mater. 2015, 27, 3453. H. E. Katz, Macromolecules 2010, 43, 2897.
Adv. Electron. Mater. 2019, 1900167 1900167 (25 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
[137] O. Bubnova, Z. U. Khan, A. Malti, S. Braun, M. Fahlman, [168] H. H. Roh, J. S. Lee, D. L. Kim, J. Park, K. Kim, O. Kwon,
M. Berggren, X. Crispin, Nat. Mater. 2011, 10, 429. S. H. Park, Y. K. Choi, A. Majumdar, J. Vac. Sci. Technol., B: Micro-
[138] Y. Yao, M. Shahi, M. M. Payne, J. E. Anthony, J. W. Brill, J. Mater. electron. Nanometer Struct.–Process., Meas., Phenom. 2006, 24,
Chem. C 2016, 4, 8817. 2398.
[139] D. Yuan, D. Huang, C. Zhang, Y. Zou, C. A. Di, X. Zhu, D. Zhu, [169] H. H. Roh, J. S. Lee, D. L. Kim, J. Park, K. Kim, O. Kwon,
ACS Appl. Mater. Interfaces 2017, 9, 28795. S. H. Park, Y. K. Choi, A. Majumdar, J. Vac. Sci. Technol., B: Micro-
[140] D. Huang, H. Yao, Y. Cui, Y. Zou, F. Zhang, C. Wang, H. Shen, electron. Nanometer Struct.–Process., Meas., Phenom. 2006, 24,
W. Jin, J. Zhu, Y. Diao, W. Xu, C. A. Di, D. Zhu, J. Am. Chem. Soc. 2405.
2017, 139, 13013. [170] S. Lefèvre, S. Volz, J.-B. Saulnier, C. Fuentes, N. Trannoy, Rev. Sci.
[141] A. R. Abramson, W. C. Kim, S. T. Huxtable, H. Yan, Y. Wu, Instrum. 2003, 74, 2418.
A. Majumdar, C. L. Tien, P. Yang, J. Microelectromech. Syst. 2004, [171] S. Lefèvre, J. B. Saulnier, C. Fuentes, S. Volz, Superlattices Micro-
13, 505. struct. 2004, 35, 283.
[142] N. E. Coates, S. K. Yee, B. McCulloch, K. C. See, A. Majumdar, [172] S. Lefèvre, S. Volz, Rev. Sci. Instrum. 2005, 76, 033701.
R. A. Segalman, J. J. Urban, Adv. Mater. 2013, 25, 1629. [173] L. David, S. Gomès, M. Raynaud, J. Phys. D: Appl. Phys. 2007, 40,
[143] S. Rausch, D. Rauh, C. Deibel, S. Vidi, H. P. Ebert, Int. J. Thermo- 4337.
phys. 2013, 34, 820. [174] E. Puyoo, S. Grauby, J.-M. Rampnoux, E. Rouvière, S. Dilhaire, Rev.
[144] C. Dames, Annu. Rev. Heat Transfer 2013, 16, 7. Sci. Instrum. 2010, 81, 073701.
[145] Y. Jin, S. Nola, K. P. Pipe, M. Shtein, J. Appl. Phys. 2013, 114, [175] A. Altes, R. Heiderhoff, L. J. Balk, J. Phys. D: Appl. Phys. 2004, 37,
194303. 952.
[146] L. Lu, W. Yi, D. L. Zhang, Rev. Sci. Instrum. 2001, 72, 2996. [176] A. A. Wilson, M. Muñoz Rojo, B. Abad, J. A. Perez, J. Maiz,
[147] W. Yi, L. Lu, D. L. Zhang, Z. W. Pan, S. S. Xie, Phys. Rev. B 1999, J. Schomacker, M. Martín-Gonzalez, D.-A. Borca-Tasciuc,
59, R9015. T. Borca-Tasciuc, Nanoscale 2015, 7, 15404.
[148] T. Y. Choi, D. Poulikakos, J. Tharian, U. Sennhauser, Appl. Phys. [177] M. Chirtoc, J. Gibkes, R. Wernhardt, J. Pelzl, A. Wieck, Rev. Sci.
Lett. 2005, 87, 3. Instrum. 2008, 79, 093703.
[149] J. S. Heron, T. Fournier, N. Mingo, O. Bourgeois, Nano Lett. 2009, [178] E. Puyoo, S. Grauby, J.-M. Rampnoux, E. Rouvière, S. Dilhaire,
9, 1861. J. Appl. Phys. 2011, 109, 024302.
[150] K. M. Lee, T. Y. Choi, S. K. Lee, D. Poulikakos, Nanotechnology [179] S. Grauby, E. Puyoo, J.-M. Rampnoux, E. Rouvière, S. Dilhaire,
2010, 21, 125301. J. Phys. Chem. C 2013, 117, 9025.
[151] G. Li, D. Liang, R. L. J. Qiu, X. P. A. Gao, Appl. Phys. Lett. 2013, [180] M. M. Rojo, S. Grauby, J. M. Rampnoux, O. Caballero-Calero,
102, 043104. M. Martin-Gonzalez, S. Dilhaire, J. Appl. Phys. 2013, 113, 054308.
[152] M. N. Ou, S. R. Harutyunyan, S. J. Lai, C. D. Chen, T. J. Yang, [181] M. M. Rojo, J. Martín, S. Grauby, T. Borca-Tasciuc, S. Dilhaire,
Y. Y. Chen, Phys. Status Solidi (b) 2007, 244, 4512. M. Martin-Gonzalez, Nanoscale 2014, 6, 7858.
[153] S. W. Finefrock, Y. Wang, J. B. Ferguson, J. V. Ward, H. Fang, [182] M. M. Rojo, J. Martín, S. Grauby, T. Borca-Tasciuc, S. Dilhaire,
J. E. Pfluger, D. S. Dudis, X. Ruan, Y. Wu, Nano Lett. 2013, 13, 5006. M. Martin-Gonzalez, Nanoscale 2015, 7, 4256.
[154] K.-Y. Chun, S. H. Kim, M. K. Shin, Y. T. Kim, G. M. Spinks, [183] J. Maiz, M. Muñoz Rojo, B. Abad, A. A. Wilson, A. Nogales,
A. E. Aliev, R. H. Baughman, S. J. Kim, Nanotechnology 2013, 24, D.-A. Borca-Tasciuc, T. Borca-Tasciuc, M. Martín-González, RSC
165401. Adv. 2015, 5, 66687.
[155] Y. Sun, L. Qiu, L. Tang, H. Geng, H. Wang, F. Zhang, D. Huang, [184] B. Abad, I. Alda, P. Diaz-Chao, H. Kawakami, A. Almarza,
W. Xu, P. Yue, Y.-s. Guan, F. Jiao, Y. Sun, D. Tang, C.-a. Di, Y. Yi, D. Amantia, D. Gutierrez, L. Aubouy, M. Martín-González,
D. Zhu, Adv. Mater. 2016, 28, 3351. J. Mater. Chem. A 2013, 1, 10450.
[156] J. Zhang, G. Song, L. Qiu, Y. Feng, J. Chen, J. Yan, L. Liu, X. Huang, [185] J. Bodzenta, J. Juszczyk, M. Chirtoc, Rev. Sci. Instrum. 2013, 84,
Y. Cui, Y. Sun, W. Xu, D. Zhu, Macromol. Rapid Commun. 2018, 39, 093702.
1800283. [186] Y. Zhu, Int. J. Heat Mass Transfer 2016, 92, 784.
[157] F. Huewe, A. Steeger, K. Kostova, L. Burroughs, I. Bauer, [187] A. Prasad, A. Ambirajan, Int. J. Therm. Sci. 2018, 134, 216.
P. Strohriegl, V. Dimitrov, S. Woodward, J. Pflaum, Adv. Mater. [188] A. Green, L. E. J. Cowles, J. Sci. Instrum. 1960, 37, 349.
2017, 29, 1605682. [189] D. J. McNeill, J. Appl. Phys. 1962, 33, 597.
[158] H. Zhang, Y. Yao, M. M. Payne, J. E. Anthony, J. W. Brill, Appl. Phys. [190] R. Homma, Y. Hasegawa, H. Terakado, H. Morita, T. Komine, Jpn.
Lett. 2014, 105, 073302. J. Appl. Phys. 2015, 54, 026602.
[159] A. Sikora, H. Ftouni, J. Richard, C. Hébert, D. Eon, F. Omnès, [191] S. Shin, R. Kumar, J. W. Roh, D. S. Ko, H. S. Kim, S. I. Kim, L. Yin,
O. Bourgeois, Rev. Sci. Instrum. 2012, 83, 054902. S. M. Schlossberg, S. Cui, J. M. You, S. Kwon, J. Zheng, J. Wang,
[160] A. Sikora, H. Ftouni, J. Richard, C. Hébert, D. Eon, F. Omnès, R. Chen, Sci. Rep. 2017, 7, 7317.
O. Bourgeois, Rev. Sci. Instrum. 2013, 84, 029901. [192] T. Kurihara, J. Morikawa, T. Hashimoto, Int. J. Thermophys. 1997,
[161] X. Zhang, C. P. Grigoropoulos, Rev. Sci. Instrum. 1995, 66, 1115. 18, 505.
[162] A. T. Ramu, J. E. Bowers, Rev. Sci. Instrum. 2012, 83, 124903. [193] H. J. Chang, J. Morikawa, T. Hashimoto, J. Appl. Polym. Sci. 2006,
[163] N. Bodenschatz, A. Liemert, S. Schnurr, U. Wiedwald, P. Ziemann, 99, 1104.
Rev. Sci. Instrum. 2013, 84, 084904. [194] J. Morikawa, T. Hashimoto, J. Appl. Phys. 2009, 105, 113506.
[164] H. Wang, F. Yang, Y. Guo, K. Peng, D. Wang, W. Chu, S. Zheng, [195] C. L. Choy, W. P. Leung, Y. K. Ng, J. Polym. Sci., Part B: Polym. Phys.
Measurement 2019, 131, 204. 1987, 25, 1779.
[165] D. G. Cahill, W. K. Ford, K. E. Goodson, G. D. Mahan, [196] Y. Hiroshige, M. Ookawa, N. Toshima, Synth. Met. 2007, 157, 467.
A. Majumdar, H. J. Maris, R. Merlin, S. R. Phillpot, J. Appl. Phys. [197] L. Zhang, T. Goto, I. Imae, Y. Sakurai, Y. Harima, J. Polym. Sci.,
2003, 93, 793. Part B: Polym. Phys. 2017, 55, 524.
[166] S. Gomès, A. Assy, P.-O. Chapuis, Phys. Status Solidi A 2015, 212, 477. [198] K. Suemori, T. Kamata, Synth. Met. 2017, 227, 177.
[167] G. Wielgoszewski, T. Gotszalk, in Advances in Imaging and Electron [199] A. J. Schmidt, R. Cheaito, M. Chiesa, Rev. Sci. Instrum. 2009, 80,
Physics, Vol. 190 (Ed: P. W. Hawkes), Academic Press, Elsevier Ltd., 094901.
Waltham 2015, Ch. 4, p. 177. [200] A. J. Schmidt, R. Cheaito, M. Chiesa, J. Appl. Phys. 2010, 107, 5.
Adv. Electron. Mater. 2019, 1900167 1900167 (26 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim
www.advancedsciencenews.com
www.advelectronicmat.de
[201] W.-L. Ong, E. S. O’Brien, P. S. Dougherty, D. W. Paley, C. Fred [222] K. Horne, H. Ban, A. Mandelis, A. Matvienko, Mater. Sci. Eng., B
Higgs III, A. J. McGaughey, J. A. Malen, X. Roy, Nat. Mater. 2017, 2012, 177, 164.
16, 83. [223] A. Wolf, P. Pohl, R. Brendel, J. Appl. Phys. 2004, 96, 6306.
[202] W.-L. Ong, S. M. Rupich, D. V. Talapin, A. J. H. McGaughey, [224] A. Mendioroz, R. Fuente-Dacal, E. Apinaniz, A. Salazar, Rev. Sci.
J. A. Malen, Nat. Mater. 2013, 12, 410. Instrum. 2009, 80, 074904.
[203] J. A. Malen, K. Baheti, T. Tong, Y. Zhao, J. A. Hudgings, [225] L. I. Giri, S. Tuli, M. Sharma, P. Bugnon, H. Berger, A. Magrez,
A. Majumdar, J. Heat Transfer 2011, 133, 081601. Mater. Lett. 2014, 115, 106.
[204] I. Hatta, Y. Sasuga, R. Kato, A. Maesono, Rev. Sci. Instrum. 1985, [226] M. Streza, S. Longuemart, E. Guilmeau, K. Strzalkowski, K. Touati,
56, 1643. M. Depriester, A. Maignan, A. Hadj Sahraoui, J. Phys. D: Appl.
[205] R. Kato, A. Maesono, R. P. Tye, Int. J. Thermophys. 2001, 22, Phys. 2016, 49, 285601.
617. [227] R. Islam, R. Chan-Yu-King, J. F. Brun, C. Gors, A. Addad,
[206] X. Y. Yu, L. Zhang, G. Chen, Rev. Sci. Instrum. 1996, 67, 2312. M. Depriester, A. Hadj-Sahraoui, F. Roussel, Nanotechnology 2014,
[207] R. Kato, A. Maesono, I. Hatta, Jpn. J. Appl. Phys. 1993, 32, 3656. 25, 475705.
[208] H. Kato, K. Nara, M. Okaji, Cryogenics 2001, 41, 373. [228] F. Roussel, R. Chen Yu King, M. Kuriakose, M. Depriester,
[209] D. A. Borca-Tasciuc, G. Chen, J. Appl. Phys. 2005, 97, 084303. A. Hadj-Sahraoui, C. Gors, A. Addad, J.-F. Brun, Synth. Met. 2015,
[210] T. Borca-Tasciuc, D. A. Borca-Tasciuc, G. Chen, IEEE Trans. 199, 196.
Compon. Packag. Technol. 2007, 30, 609. [229] J. W. Brill, M. Shahi, M. M. Payne, J. Edberg, Y. Yao, X. Crispin,
[211] H. Zhang, J. W. Brill, J. Appl. Phys. 2013, 114, 043508. J. E. Anthony, J. Appl. Phys. 2015, 118, 235501.
[212] D. A. Borca-Tasciuc, G. Chen, A. Prieto, M. S. Martín-González, [230] A. Rosencwaig, A. Gersho, J. Appl. Phys. 1976, 47, 64.
A. Stacy, T. Sands, M. A. Ryan, J. P. Fleurial, Appl. Phys. Lett. 2004, [231] F. A. McDonald, G. C. Wetsel, J. Appl. Phys. 1978, 49, 2313.
85, 6001. [232] H. Vargas, L. C. M. Miranda, Phys. Rep. 1988, 161, 43.
[213] B. A. Lunn, J. Unsworth, N. G. Booth, P. C. Innis, J. Mater. Sci. [233] H. Hu, X. Wang, X. Xu, J. Appl. Phys. 1999, 86, 3953.
1993, 28, 5092. [234] O. Pessoa, C. L. Cesar, N. A. Patel, H. Vargas, C. C. Ghizoni,
[214] A. Malti, J. Edberg, H. Granberg, Z. U. Khan, J. W. Andreasen, L. C. M. Miranda, J. Appl. Phys. 1986, 59, 1316.
X. Liu, D. Zhao, H. Zhang, Y. Yao, J. W. Brill, I. Engquist, [235] J. A. Balderas-López, Rev. Sci. Instrum. 2003, 74, 1338.
M. Fahlman, L. Wagberg, X. Crispin, M. Berggren, Adv. Sci. 2016, [236] J. A. Balderas-López, Rev. Sci. Instrum. 2006, 77, 064902.
3, 1500305. [237] A. Lachaine, P. Poulet, Appl. Phys. Lett. 1984, 45, 953.
[215] F. Jiao, A. Naderi, D. Zhao, J. Schlueter, M. Shahi, J. Sundström, [238] X. Wang, H. Hu, X. Xu, J. Heat Transfer 2001, 123, 138.
H. Granberg, J. Edberg, U. Ail, J. Brill, T. Lindström, M. Berggren, [239] B. A. Cola, R. Karru, C. Cheng, X. Xu, T. S. Fisher, IEEE Trans.
X. Crispin, J. Mater. Chem. A 2017, 5, 16883. Compon. Packag. Technol. 2008, 31, 46.
[216] R. Santos, L. C. M. Miranda, J. Appl. Phys. 1981, 52, 4194. [240] J. A. Balderas-López, A. Mandelis, Rev. Sci. Instrum. 2003, 74, 5219.
[217] L. Fabbri, P. Fenici, Rev. Sci. Instrum. 1995, 66, 3593. [241] K. G. Biswas, T. D. Sands, B. A. Cola, X. Xu, Appl. Phys. Lett. 2009,
[218] M. Depriester, P. Hus, S. Delenclos, A. H. Sahraoui, Rev. Sci. 94, 223116.
Instrum. 2005, 76, 074902. [242] T. A. El-Brolossy, S. S. Ibrahim, Phys. Scr. 2014, 89, 105701.
[219] R. Fuente, E. Apinaniz, A. Mendioroz, A. Salazar, J. Appl. Phys. [243] M. K. Smith, V. Singh, K. Kalaitzidou, B. A. Cola, ACS Nano 2015,
2011, 110, 033515. 9, 1080.
[220] P. Martinez-Torres, A. Mandelis, J. J. Alvarado-Gil, J. Appl. Phys. [244] M. Kuriakose, M. Depriester, R. Chan Yu King, F. Roussel, A. Hadj
2009, 106, 114906. Sahraoui, J. Appl. Phys. 2013, 113, 044502.
[221] J. A. Garcia, A. Mandelis, B. Farahbakhsh, C. Lebowitz, I. Harris, [245] K. Touati, M. Depriester, M. Kuriakose, A. H. Sahraoui, Rev. Sci.
Int. J. Thermophys. 1999, 20, 1587. Instrum. 2015, 86, 094901.
Adv. Electron. Mater. 2019, 1900167 1900167 (27 of 27) © 2019 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim