3DQuorum TM

Software

Service Manual Supplement


For Selenia Dimensions Systems and 3Dimensions™ Systems
® ®

Part Number MAN-06228


Revision 001
August 2019

Technical Support
USA: +1.877.371.4372 Asia: +852 37487700
Europe: +32 2 711 4690 Australia: +1 800 264 073
All Other: +1 781 999 7750 Email: BreastHealth.Support@hologic.com

© 2019 Hologic, Inc. Printed in the USA. This manual was originally written in English.

Hologic, 3DQuorum, 3Dimensions, Selenia, Dimensions, and associated logos are trademarks and/or registered trademarks of
Hologic, Inc., and/or its subsidiaries in the United States and/or other countries. All other trademarks, registered trademarks,
and product names are the property of their respective owners.

This product may be protected by one or more U.S. or foreign patents as identified at www.Hologic.com/patents.
3DQuorum Software Service Manual Supplement
Table of Contents

Table of Contents
1: Introduction __________________________________________________________________1
1.1 About 3DQuorum Software .................................................................................................................................. 1
1.2 3DQuorum Software Theory of Operation ......................................................................................................... 1
1.3 Summary of User Interface Changes.................................................................................................................... 2
1.3.1 Image Display Screen ............................................................................................................................... 3
1.3.2 Display Modes .......................................................................................................................................... 4
1.3.3 Slice Indicator ............................................................................................................................................ 4
1.3.4 Close a Patient ........................................................................................................................................... 5
1.3.5 Select an Output Group ........................................................................................................................... 5
1.3.6 Output Groups .......................................................................................................................................... 5

2: Installation and Configuration__________________________________________________7


2.1 3DQuorum Installation .......................................................................................................................................... 7
2.1.1 Install the 3DQuorum Software License ............................................................................................... 7
2.2 3DQuorum Configuration of Default Settings.................................................................................................... 9
2.2.1 Create or Select a Primary Output Setting ............................................................................................ 9
2.3 Install the Default Output Groups ..................................................................................................................... 24
2.3.1 Add New Output Group ....................................................................................................................... 25
2.3.2 Verify Configuration of Existing Output Groups .............................................................................. 27
2.4 Observe SmartSlices on Connected Devices ..................................................................................................... 28

MAN-06228 Revision 001 iii


3DQuorum Software Service Manual Supplement
1: Introduction

1 Introduction
1:

1.1 About 3DQuorum Software


Note
3DQuorum™ imaging technology may not be available in all regions. Contact your
sales representative for information.

3DQuorum™ software uses image data available from a high-resolution breast


tomosynthesis acquisition to generate SmartSlices, which are "slabbed" 3D images. The
SmartSlices are created without the need for an additional digital tomosynthesis
exposure. A SmartSlice can be perceived as a "thick" version of a tomosynthesis slice and
is equivalent to six slices (6 mm) from the original tomosynthesis reconstruction. The
SmartSlice volume consists of sequentially synthesized slices. Each SmartSlice shares half
of its contents with the SmartSlice before and the other half with the SmartSlice after.
Therefore, when reviewing SmartSlices, the images are advanced three 1-mm
tomosynthesis slices at a time. The resulting number of images to be reviewed by a
radiologist in a study is one-third the number of a tomosynthesis 1-mm reconstruction
image set.
SmartSlices are designed to reduce the total tomosynthesis screening study interpretation
time by decreasing the total number of images a radiologist is required to review.
SmartSlices images are not intended to be used without the synthesized 2D images to
make a clinical decision or diagnosis.

1.2 3DQuorum Software Theory of Operation


Overview
3DQuorum software is an image processing application for post-processing a limited
number of slices of tomosynthesis data, captured on a Hologic tomosynthesis imaging
system, into a combined 3D image. The SmartSlices image set reduces the number of
images to one-third compared to tomosynthesis reconstructed slices data set while
maintaining clinical information. As a result, read times for the radiologist can be
reduced by about 13%. And for archival and transmission purposes, file sizes can be
reduced by over 50%. SmartSlices can be used in place of the tomosynthesis
reconstructed slices when reviewing a screening study employing tomosynthesis.
SmartSlices achieve the same sensitivity as reading 1-mm 3D™ slices.

MAN-06228 Revision 001 Page 1


3DQuorum Software Service Manual Supplement
1: Introduction

Configuration
3DQuorum software has no user-configurable settings that influence the appearance of
the resulting SmartSlices. SmartSlices are stored and transmitted as a series of frames
using the DICOM Breast Tomosynthesis Image object format. The DICOM header for
SmartSlices contains the information needed to differentiate it from any accompanying
tomosynthesis reconstructed slices in the same view. An annotation ("3DQuorum") is
also burned into the SmartSlices pixel data.
During the initial installation of 3DQuorum software, the system is configured to allow
sending of tomosynthesis reconstructed slices and SmartSlices to PACS and diagnostic
workstations. The default output groups can be configured to send only SmartSlices, only
tomosynthesis reconstructed slices, or both to the preferred devices automatically. The
site PACS administrator, in consultation with Hologic connectivity engineers, can select
the output formats most appropriate for the site’s IT infrastructure and workstations.
As with the introduction of any new imaging technology, Hologic advises testing
SmartSlices with your PACS, workstations, and other Imaging Enterprise devices to
ensure compatibility.
Workflow
As with any imaging study, the technologist selects the patient and identifies the type of
imaging procedure that will be done. For tomosynthesis screening exams with
3DQuorum software, the process of imaging the subject and completing the study is all
that is necessary. The 3DQuorum software itself operates with no direct human
intervention.

1.3 Summary of User Interface Changes


3DQuorum software affects the user interface on the control monitor in two major ways:
• Adds the SmartSlices button to the Display Modes.
• Changes preconfigured output groups and storage devices using standardized
names.

Page 2 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
1: Introduction

1.3.1 Image Display Screen


After you make an exposure, the acquired image opens on the Image Display monitor.
Patient and procedure information can be displayed on the screen. To turn the
information on or off, go to the Tools tab in the Procedure screen and select the Patient
Information button.

Figure 1: Image Display Screen (SmartSlices option shown)

Figure Legend

1. Patient information and exam date


2. 3DQuorum software logo
3. Slice Indicator
4. Exposure information

MAN-06228 Revision 001 Page 3


3DQuorum Software Service Manual Supplement
1: Introduction

1.3.2 Display Modes


Use the buttons in the Display Modes area of the screen to select the type of view to show
on the Image Display monitor. You can change between conventional, generated 2D,
projections, reconstructions, and SmartSlices to show the combination images.

Figure 2: Display Modes

Figure Legend
1. Conventional button shows conventional 2D images.
2. Generated 2D button shows a 2D image generated from the acquired tomographic
image.
3. Projections button shows the projection images from the 15° sweep.
4. Reconstructions button shows 1-mm reconstructed slices.
5. SmartSlices button shows 6-mm slices from the 3DQuorum software.

1.3.3 Slice Indicator


The Slice Indicator on the Image Display monitor shows only on tomographic
reconstructions, including SmartSlices.

Figure Legend
1. Up and Down arrows let you change between slices
that contain a lesion target and slices that are tagged
for printing.
2. Up and Down arrows let you change between slices
that contain notices.
3. “H” (anatomical reference to head direction)
4. Height within the breast volume (for SmartSlices)
5. Number of the tomosynthesis reconstructed slice or
SmartSlice image
6. Scroll bar moves through the slices of the
reconstruction.
7. Slices that contain targets or are tagged for printing.
8. Slices that contain notices.
9. “F” (anatomical reference to foot direction)

Figure 3: Slice Indicator


(SmartSlice option shown)

Page 4 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
1: Introduction

1.3.4 Close a Patient


Follow the standard procedure to close a patient. If there are SmartSlices that have not
completed processing, a dialog box opens with the following message:

Figure 4: Processing Message

Wait for processing to complete before you open the next patient.

1.3.5 Select an Output Group


Select an output device group like PACS, Diagnostic Workstations, CAD devices, and
printers from the Output Groups drop-down list in the Procedure screen.

Figure 5: Standardized default Output Group used with 3DQuorum software

Note
Images are not sent if an Output Group is not selected.

1.3.6 Output Groups


The accepted images are sent automatically to the output devices in the selected Output
Group. The system configuration controls if the images are sent after a patient is closed
or immediately after each image is accepted.

Note
Tomosynthesis images are not sent to a print device in the selected Output Group. You
can print selected tomosynthesis slices and SmartSlices from the Print screen.

MAN-06228 Revision 001 Page 5


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

2 Installation and Configuration


2:

2.1 3DQuorum Installation


After performing the setup, configuration, and connectivity tasks per the instructions in
the Selenia Dimensions or 3Dimensions Service Manual, follow the steps in this section to
install the 3DQuorum software license.

2.1.1 Install the 3DQuorum Software License


When installing a new licensed feature on a system, follow these steps.
1. Obtain the Sales Order number that includes the 3DQuorum software feature.
2. Launch System Tools (refer to the System (Service) Tools topic in the system Service
Manual).
3. Navigate to the AWS > Licensing. The Licensing screen opens (see the following
figure).

Figure 6: System Tools > AWS > Licensing screen

MAN-06228 Revision 001 Page 7


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

4. Generate a feature request file (xxxxxx.holx). This file includes all the features
currently activated on the system and the MAC address of the computer. Refer to the
previous figure for item references.
a. Select Export Current (item 1).
b. Select Download Request (item 2).
c. In the browser dialog, select Save as.
d. Navigate to the folder where you want to save the xxxxxx.holx file. Make sure
that HOLX File is selected as file type.
e. In the browser dialog, select Save.
5. Create an email that includes the following:
a. To Line: 3D@Hologic.com
b. Subject Line: 3DQuorum Upgrade License request.
c. SN# of the system.
d. 3DQuorum Sales order # (if installing on an existing system).
e. Attach the generated xxxxx.holx request file saved earlier to C:\Licenses.
f. Your Name / Contact info.
6. Send the email to 3D@Hologic.com.
7. You should receive a return email from 3D@Hologic.com with a new activation file
(xxxxxx.holx). This file includes all the features currently activated on the system
plus the new requested level of software.
8. In System Tools, go to the AWS > Licensing page and import the new activation file.
Refer to the previous figure for item references.
a. Select Browse (item 3).
b. Navigate to the folder containing the new activation file, select the file, and select
Open.
c. Select Import.
d. Verify that both the “ClarityHD” and “Quorum3D” features are listed with a
“Licensed” state in the feature table (as well as the existing licensed features).
e. Select Validate (item 5) to confirm that the proper software has been loaded for
your system. (If successful, a "Validate Complete, no errors detected" message
appears).
9. Exit from System Tools and reboot the computer.
10. In the system Capture application, go to Admin > About screen > Licensing tab and
verify that the new software licenses are installed.

Page 8 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

2.2 3DQuorum Configuration of Default Settings


This section describes how to prepare the default output settings. These settings are used
during clinical application training and may be further customized for the particular
workflow requirements of a customer.
Upon initial installation of 3DQuorum software, users must have control over whether
SmartSlices, 3D 1 mm Slices, or both are sent to storage devices, such as PACS,
workstations, and/or workflow routers (such as SecurXchange). Use the instructions in
this section to create three installed device configurations for each physical device to
allow the required control over data image types.

Note
The output devices configured for SmartSlices must support receiving the DICOM
Breast Tomosynthesis Image SOP Class (BTO).

Note
SmartSlices and 3D 1 mm Slice images are not sent to a print device in the selected
Output Group by functional design. Users can print selected SmartSlice and 3D 1 mm
Slice images manually from the Print screen.

Note
Unless directed to configure the system differently, upon installation of 3DQuorum
software, use the standardized labels in this section to aid in the initial training and
adoption of SmartSlices. The installed device configurations and output group names
may be further customized once long-term 3D imaging data norms are established.

2.2.1 Create or Select a Primary Output Setting


1. Launch System Tools.
Note
Depending on the system, there may be a shortcut to System Tools on the Windows
desktop that allows you to launch System Tools without starting the Capture
application.

If no shortcut is available on the desktop:


a. Launch the Capture application and log in as a Service user.
b. At the Select Patient screen, select the Admin button.
c. At the Admin screen, select System Tools.

MAN-06228 Revision 001 Page 9


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

2. The System Tools Login screen appears. Login as a Service user and enter the
appropriate password. Select Log In.

Figure 7: System Tools Login screen

3. The System Tools Welcome main screen appears.


4. Navigate to AWS > Connectivity > Installed Devices. The Installed Devices screen
appears.
The following figure shows an example archive named “PACS” and an example
workstation named “WS” that are installed.

Figure 8: Installed Devices screen

5. If the store device of interest of has not yet been installed, consult the
Dimensions/3Dimensions Service Manual for instructions (refer to the Install an
Archive Interface and/or Install a Workstation Interface sections). Complete the output
devices installation and return here.
In the procedures that follow, the existing installed devices configuration settings are
referred to as the "primary output setting” for the physical device.

Page 10 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Configure Primary Output Settings - All Desired Data except SmartSlices


and 1 mm Slices
1. In the Installed Devices screen in System Tools, select (highlight) the primary output
setting device.
2. Mouse over Advanced and select Advanced Store.

Figure 9: Installed Devices - Advanced Store button

3. The Advanced Store screen opens. Scroll down and locate the Support DICOM Types
section under the Config group section (see the following figure).

MAN-06228 Revision 001 Page 11


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Figure 10: System Tools > AWS > Connectivity > Installed Devices > Advanced Store

4. Use the table below to record which of the following DICOM types appear in the
Selected column for the installed device being edited. This is referenced later in the
section Verify Configuration of Existing Output Groups on page 27.
Device Name

3D Slabs (BTO)

3D Slices (BTO)

3D Slices (SCO)

5. If present, move the DICOM Types “3D Slabs (BTO)”, “3D Slices (BTO)” and “3D
Slices (SCO)” from the “Selected” column on the right to the “Unselected” column on
the left. To move a DICOM type, select and drag the DICOM type from the Selected
box to the Unselected box. See the following sequence of screen figures.

Page 12 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Figure 11: Moving Support Dicom Types from Selected to Unselected states
Keep in mind the following:
• DICOM types located in the Selected box are sent to the device.
• DICOM types located in the Unselected box are not sent to the device.
6. Select Update.
7. Select << Back.
8. On the Installed Devices screen, mouse over Ping, then select Ping Store. The system
indicates if the ping was successful or failed. If the ping failed, check the IP address
entered for that device.
9. Mouse over Dicom Ping, then select Dicom Ping Store. The system indicates if the
DICOM ping was successful or failed. If the DICOM ping failed, check the Called AE
Title, Calling AE Title and/or Port fields.
10. Repeat this section (starting at step 1) for all existing output devices appropriate for
choosing to send SmartSlices and/or 3D 1 mm Slices.

Configure SmartSlices Output Settings

MAN-06228 Revision 001 Page 13


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Note
The output devices configured for SmartSlices must support receiving the DICOM
Breast Tomosynthesis Image SOP Class (BTO).

1. In the Installed Devices screen in System Tools, select (highlight) the primary output
setting device configured in the previous section.
2. Select Clone this Device.

Figure 12: Installed Devices - Clone this Device button

3. Select (highlight) the clone of the primary output setting.

Page 14 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

4. Mouse over Basic, then select Basic Store.

Figure 13: Installed Devices - Basic Store button

5. The ImageStorage Basic Store screens opens. Edit the Name field by adding “ -
SmartSlices” to the existing device name.

Figure 14: Basic Store - Edit Name

6. Edit the Notes field as appropriate.


7. Select Update.
8. Select << Back.
9. On Installed Devices screen, move your mouse pointer over Advanced, then select
Advanced Store.

MAN-06228 Revision 001 Page 15


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

10. The Advanced Store screen opens.

Figure 15: Installed Devices > Advanced Store

11. Scroll down and locate the Support DICOM Types section under the Config group
section (see the following figure).

Page 16 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

12. Move all of the DICOM from the “Selected” column on the right to the “Unselected”
column on the left. To move a DICOM type, select and drag the DICOM type from
the Selected box to the Unselected box. See the following sequence of screen figures.

Figure 16: Moving Selected DICOM types to Unselected section

MAN-06228 Revision 001 Page 17


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

13. Use the Scroll bar in the Unselected section to locate the entry “3D Slabs (BTO)”.
Select 3D Slabs (BTO) and drag the selection to the Selected section. See the
following set of figures.

Figure 17: Moving 3D Slabs (BTO) to Selected

14. Select Update.


15. Select << Back.
16. On the Installed Devices screen, mouse over Ping, then select Ping Store. The system
indicates if the ping was successful or failed. If the ping failed, check the IP address
entered for that device.
17. Mouse over Dicom Ping, then select Dicom Ping Store. The system indicates if the
DICOM ping was successful or failed. If the DICOM ping failed, check the Called AE
Title, Calling AE Title and/or Port fields.
18. Repeat this section starting at Step 1 for all existing output devices to which the
technologist will want control over sending SmartSlices, 3D 1 mm Slices, or both
types of Digital Breast Tomosynthesis reconstructed slices.

Page 18 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Configure 3D 1 mm Slices Output Settings


Note
The output devices configured for 3D 1 mm Slices must support receiving the DICOM
Breast Tomosynthesis Image SOP Class (BTO).

1. In the Installed Devices screen in System Tools, select (highlight) the primary output
setting device configured in the previous section.
2. Select Clone this Device.

Figure 18: Installed Devices - Clone this Device button

3. Select (highlight) the clone of the primary output setting.

MAN-06228 Revision 001 Page 19


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

4. Mouse over Basic, then select Basic Store.

Figure 19: Installed Devices - Basic Store button

5. The ImageStorage Basic Store screens opens. Edit the Name field by adding “ – 3D 1
mm Slices” to the existing device name.

Figure 20: Basic Store - Edit Name

6. Edit the Notes field as appropriate.


7. Select Update.
8. Select << Back.

Page 20 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

9. On Installed Devices screen, move your mouse pointer over Advanced, then select
Advanced Store. The Advanced Store screen opens.

Figure 21: Installed Devices > Advanced Store

MAN-06228 Revision 001 Page 21


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

10. Scroll down and locate the Support DICOM Types section under the Config group
section (see the following figure).
11. Move all the DICOM types from the “Selected” column on the right to the
“Unselected” column on the left. To move a DICOM type, select and drag the
DICOM type from the Selected box to the Unselected box. See the following sequence
of screen figures.

Figure 22: Moving Selected DICOM types to Unselected section

Page 22 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

12. Use the Scroll bar in the Unselected section to locate the entry “3D Slices (BTO)”.
Select 3D Slices (BTO) and drag the selection to the Selected section as shown in the
following sequence of screen figures.

Figure 23: Move 3D Slices (BTO) to Selected

13. Select Update.


14. Select << Back.
15. On the Installed Devices screen, mouse over Ping, then select Ping Store. The system
indicates if the ping was successful or failed. If the ping failed, check the IP address
entered for that device.
16. Mouse over Dicom Ping, then select Dicom Ping Store. The system indicates if the
DICOM ping was successful or failed. If the DICOM ping failed, check the Called AE
Title, Calling AE Title and/or Port fields.

MAN-06228 Revision 001 Page 23


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

17. Repeat this section starting at Step 1 for all existing output devices to which the
technologist will want control over sending SmartSlices, 3D 1 mm Slices, or both
types of Digital Breast Tomosynthesis reconstructed slices. See the following figure
for an example installation for an archive (PACS) and a SecurView workstation (WS).

Figure 24: Installed Devices screen

Note
Log out of System Tools and reboot the system for all changes to take effect.

2.3 Install the Default Output Groups


When installing 3DQuorum software, the following Output Groups are added to any
previously existing list of Output Groups:
Table 1: Output Groups for 3DQuorum Software
Output Group Name Explanation of the Configuration
SmrtSlces,DBT,2D Sends all configured data types

SmartSlices,2D Sends SmartSlices and 2D data types (No 1-mm reconstructed


slices)

DBT,2D Sends 1-mm reconstructed slices and 2D data types (No


SmartSlices)

Custom Provides on-demand selection of outputs and data types

In addition to the standardized Output Groups, edits to previously existing Output


Groups may be required to preserve current data operations. The following procedures
are provided for the necessary edits to previously existing Output Groups to preserve
current operations.

Page 24 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

2.3.1 Add New Output Group


1. In the Capture application, at the Select Patient screen, select the Admin button.
2. At the Admin screen, select Manage Output Groups.
3. Select the New button. The Add Output Set screen appears.

Figure 25: Admin > Manage Output Groups > Add Output Set

4. Enter the Name and Description of the grouping, along with the choice of outputs
per the following table:
Table 2: Detailed Description of Output Groups
Output Group Description Included Devices
SmrtSlces,DBT,2D All image types sent All Desired Primary Output Settings
All SmartSlices Output Settings
All 3D 1 mm Slice Output Settings
All other devices included in normal
workflow, such as CAD.

SmartSlices,2D No 3D 1 mm slices All Desired Primary Output Settings


sent
All SmartSlices Output Settings
All other devices included in normal
workflow, such as CAD.

DBT,2D No SmartSlices sent All Desired Primary Output Settings


All 3D 1 mm Slice Output Settings
All other devices included in normal
workflow, such as CAD.

MAN-06228 Revision 001 Page 25


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

The following figures show an example of each output group setting type when the
installed devices include an archive (PACS) and workstation (WS).
• SmrtSlces,DBT,2D

• SmartSlices,2D

• DBT,2D

5. To save the Output Group, select Add, then select OK in the Update Successful
message.
6. Repeat steps 3 through 5 until all three Output Groups have been added.

Note
Log out of System Tools and reboot the system for all changes to take effect.

Page 26 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

7. To set Output Group defaults, you can select any group to set as the default for
different study types. See the following figure.

Figure 26: Manage Output Groups screen

2.3.2 Verify Configuration of Existing Output Groups

IMPORTANT
When 3DQuorum software is being added to an existing system, be mindful of existing
output group settings that remain and could be continued to be used. This section
explains how to add back any data types removed from the “primary output setting”
for each modified device.

To verify previously existing output groups:


1. Refer to the table in Configure Primary Output Settings - All Desired Data except
SmartSlices and 1 mm Slices on page 11, noting any image types removed from the
primary output setting for each device.
2. In the Capture application, at the Select Patient screen, select the Admin button.
3. At the Admin screen, select Manage Output Groups.
4. Select and review each existing output group and determine if any primary output
settings edited to support 3DQuorum software is active for the group.

MAN-06228 Revision 001 Page 27


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

5. For any groups affected, select the group and then select Edit.
• If the DICOM Type 3D Slices (BTO) was removed from the primary output
setting, select the installed device labeled “Device Name – 3D 1 mm Slices” to the
output group.
• If the DICOM Type 3D Slices (SCO), 3D Slices (CTO), or 3D Slabs (BTO) were
removed from the primary output setting, consult with your Connectivity
Specialist or Technical Support and the customer to determine if additional
installed devices need to be added to the existing output group. Given 3D Slices
(BTO) was added to one of the new devices configured above, be careful not to
send 3D Slices (SCO) [or 3D Slices (CTO)] to one device and 3D Slices (BTO) to
another. Also be careful not to send both 3D Slices (SCO) [or 3D Slices (CTO)]
and 3D Slices (BTO) to the same destination.
6. When finished, select Save, then select OK in the Update Successful message.
7. Repeat steps 4 -6 for all previously existing output groups.

Note
Log out of System Tools and reboot the system for all changes to take effect.

2.4 Observe SmartSlices on Connected Devices


Acquisition Workstations
1. Following instructions in the system User Guide, create a test patient Tomo or Tomo
HD screening procedure with naming and demographics consistent with customer
testing policies.
2. Select the output group that will send SmartSlices, 3D 1 mm Slices, and 2D image
data to the connected devices to be tested. The following figure shows the output
group to select when the standardized default labeling is used.

Figure 27: Standardized default Output Group used with 3DQuorum software

3. Following instructions in the system User Guide, acquire images for the Tomo or
Tomo HD procedure and close the procedure as Completed.
4. Observe the Manage Queue utility to ensure the image store jobs are successfully
sent to the connected devices active in the Output Group.

Page 28 MAN-06228 Revision 001


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Diagnostic Workstations
1. Observe the presentation of both the SmartSlices and the 3D 1 mm Slices on the
diagnostic (DX) workstation using the following chart as a guide. Contact your
Connectivity Specialists or Technical Support and Clinical Application Training
Specialists for any questions or concerns regarding your observations.
Table 3: Presentation of SmartSlices and 3D 1 mm Slices on the DX Workstation
Observe that SmartSlices can be discerned from the 3D 1 mm Slices. Use the differing DICOM field content
and features to identify each image type.

3D 1 mm Slices SmartSlices

Series Description (0008,103E) Laterality+View Breast Laterality+View Breast


Tomosynthesis Image Tomosynthesis Image Slabs
RCC Breast Tomosynthesis Image RCC Breast Tomosynthesis Image
Slabs

Series Number (0020,0011) 73200000 73500000

Burned in marker None 3DQuorum

Observe the default Window Center and Window Width values. Indicate if they are not the values listed
below 512/512 for 3D 1 mm Slices, 916/570 for SmartSlices. Determine if the alternative SmartSlices WW/WC
values are accessible via the workstation tools.

Window Width (WW) 512 916\760\1120


(5200,9229.0028,9132.0028,1051)

Window Center (WC) 512 570\540\650


(5200,9229.0028,9132.0028,1050)

Window Center & Width N/A Normal\High contrast\Low contrast


Explanation
(5200,9229.0028,9132.0028,1055)

When using the slice scrolling feature, observe that the total number of slices and slice thickness values are
displayed. Scroll through the slices and SmartSlices noting any unexpected behavior on the part of the
workstation.

Number of Frames (0028,0008) X: For example: 60 slices X / 3: For example: 20 slices

Slices Thickness 1 6
(5200,9230.0028,9110.0018,0050)

MAN-06228 Revision 001 Page 29


3DQuorum Software Service Manual Supplement
2: Installation and Configuration

Archives (PACS)
1. If query/retrieve from PACS is available to the Acquisition Workstation, delete the
test patient from the system. Make sure there are no outstanding store or print jobs
for this patient, else the patient will not be deleted.
2. Retrieve the test patient from the archive.
3. Open the patient and view both the SmartSlices and 3D 1 mm Slices on the
Acquisition Workstation. Use the previous table to observe the retrieved images,
noting anything unexpected.

Page 30 MAN-06228 Revision 001