Half amplitude / 6 dB drop technique :

Half amplitude technique for defining the ends of a discontinuity is


used when it is longer than the crystal size of the probe. This is the
most popular technique for discontinuity length measurement.
After maximizing the discontinuity indication, set this signal to a
convenient height [ ~ 80 % ]. The probe is then moved towards right,
parallel to the discontinuity or the weld seam as appropriate, up to
the position where the signal drops to half the set amplitude. A point
is marked with a marker above the discontinuity location
corresponding to the centerline of the probe crystal. A similar point is
found by moving the probe towards the left. The distance between In this method the edge of the ultrasound
these two points is the length of the discontinuity. beam is used to find the starting edge of the
The discontinuity length measured by this technique always results discontinuity.
in an over estimation of the actual size. This is because the half Once the presence of the discontinuity is
intensity line is located at an angle to the main axis of the beam confirmed, the scanning starts from a position
[ centerline of the crystal ]. where the discontinuity signal is not visible on
The error will depend on the size and frequency of the probe and the CRT screen.
the beam path distance to the discontinuity. The shape at the ends The probe is now slowly moved towards the
of the discontinuity will also influence the measurement. Small crystal discontinuity up to the position where the
probes [ 4 MHz, 8 X 9, 10 mm ] should be preferred for the discontinuity indication just begins to appear on
measurements. the CRT screen. A point above the discontinuity
This measurement error can be minimized by plotting the half location is marked which coincides with the line
amplitude beam boundary in the horizontal plane using the IOW passing through the sidewall of the probe,
beam profile block and using the graph thus obtained for correcting facing the discontinuity.
the length at the applicable flaw depth. A similar point is to be found by scanning the
Edge approach technique : discontinuity by probe movement from the other
When the discontinuity size is much smaller than the size of the end. The distance between these two points is
crystal, half amplitude technique of measurement does not produce the length of the discontinuity.
satisfactory result. A different method called edge approach technique
is used.

Compiled by KALYAN PRASAD CHATTERJEE/ K. SHIVAKUMAR – INTERNATIONAL TRAINERS