You are on page 1of 2

Instructions 8250/8252/8254/8256/8258

NHT/NAT/NAH/NAE/NAR
Pressure Transmitter
Warning: Before installation, commissioning, and operation, ensure that the pressure transmitter is suitable for the appli-

! cation in terms of measuring range, design and environmental conditions. Non-observance can result in serious injury and/
or damage to equipment.
Intended use: This instrument converts pressure into an electrical signal.
The instrument has been designed and built solely for the intended use described here and may only be used accordingly. If the equip-
ment is used in a different manner, the protection provided by the equipment may be impaired and Trafag shall not be liable for any
claims at all.

Datasheets
NHT 8250 NAT 8252 NAH 8254 NAE 8256

09/2023
H73303v Trafag AG
www.trafag.com/H72338 www.trafag.com/H72303 www.trafag.com/H72304 www.trafag.com/H72305
NAR 8258

We reserve the right to make alternations as technical progress may warrant.


www.trafag.com/H72307

Pressure switch parameterization tool


Hardware: Sensor Master Interface SMI Software: Sensor Master Communicator SMC

Download the Android app


in the Google Play Store

www.trafag.com/H72618 Download the Windows app in the Microsoft store


https://www.microsoft.com/store/productId/9NK63QX8XXBK

Connection of the measuring equipment


Switching output Connection of loads Current output Voltage output
to switch contacts 2-wires 3-wires
shield shield shield

Us (pos. Supply)
P
Us (pos. Supply)
SP1
P US +
US –
Us (pos. Supply)
Out (Output)
Out SP2
I Us (neg. Supply)
Us (neg. Supply) SP 1, SP 2 Us (neg. Supply)
RL
earth/housing earth/housing
earth US –

www.trafag.com
Type label description (example)
Measuring range Serial number Type code Electrical connection Manufacturing date (W/Y)

Type 825X.XX.XXXX.XX
Max. overpressure
S/N YYYYYYYY-ZZZ │ XX/XX

Range -1...9 bar-G │ max.: XXXX bar


Out (1) 4...20mA │ b (4)
US+ (2) 9...32 VDC
!

Output signal Supply positive Pin assignment Supply voltage Pin assignment

Electrical connections
Industrial M12x1 M12x1
Designation MIL-C 26482 Cable
standard 4–pole 5–pole

Type code 825X.XX.XXXX.01 825X.XX.XXXX.32 825X.XX.XXXX.35 825X.XX.XXXX.02 825X.XX.XXXX.22/24/08/88

Ambient 22: -5°C ... +60°C 08: -40°C ... +125°C


-40°C ... +125°C -40°C ... +125°C -40°C ... +125°C -40°C ... +125°C
temperature range 24: -40°C ... +70°C 88: -40°C ... +100°C
RD: red
A
3 3 2 BK: black
2 5 F B WH:white
Pin 2
3
1 GN: green
configuration 4
4 4 C BN: brown
1 E YE: yellow
1 D
YE/GN: yellow/green

Output Load resistance I U


DT04 DT04 SUPPLY SUPPLY
Designation 4 ... 20 mA see graphic 9 ... 32 VDC
3-pole 4-pole 0 ... 6 VDC ranges ≥ 5.0 kΩ to Us – ≤ 20 mA 9 ... 32 VDC
0 ... 10.1 VDC ranges ≥ 5.0 kΩ to Us – ≤ 15 mA 15 ... 32 VDC
Type code 825X.XX.XXXX.D3 825X.XX.XXXX.D4
0.5 ... 4.5 VDC ratiom. ≥ 5.0 kΩ to Us – ≤ 10 mA 4.75 ... 5.25 VDC
2 PNP Transistors ≤ 10 mA 9 ... 32 VDC
Ambient
-40°C ... +125°C -40°C ... +125°C 1 PNP Transistor ≤ 10 mA 9 ... 32 VDC
temperature range
4...20mA: min./max resistor vs. supply voltage @ Pmax = 100%
B A 4 1 1200
Customer Load Resistor (Ohm)

Pin B A 4 1 1000 ....... Rmax


Rmin 100°C
configuration 3 2
Rmin 125°C
800
C
C
3 2
600

400

200
Functions switching output
0
0 5 10 15 20 25 30
Hysteresis Window Delay Customer Supply Voltage (V)
P P P
SP FH SP

rP
rP FL

t t t
ON ON ON
Hno Fno Hno
OFF OFF OFF

ON ON ON
Hnc Fnc Hnc
OFF OFF OFF

delay delay
www.trafag.com