FULLY AUTO
BIOCHEMISTRY ANALYZER
Version 2023.07.01
Basic parameter
a. 24+5 positions for sample, 1 position for calibration, 1 position for quality control, 1
position for emergency. The total volume of the sample cups is 850 μL
b. 27 double reagent positions, outer track is NO.1-NO.27, bottle volume is 8ml. Inner
track is NO.28-NO.54, bottle volume is 24ml. NO.27 is system position, please add
pure water to the bottle, cannot set as reagent position
c. 81 cuvettes, 9 cuvettes/group, 9 groups totally
d. Wavelength:340 405 450 500 546 578 630 670nm
e. Methodology : Two-point、End-point、Rate
f. Power supply: 50Hz AC100-240V
g. Power: 150W
h. Dimension: 605mm(L)*350mm(W)*330mm(H)
i. Net weight: 22kg
Testing
★ One sample can apply for up to 20 test items
★ Don't add reagent more than 2/3 of reagent bottle
★ Put pure water at No.27 Reagent position → for probe liquid level detection
1
★ Check that the reagents and samples are properly placed before starting the test
★ Please ensure that the sample is sufficient to meet the test requirements. The test
results are not accurate when the total volume of serum in the sample cup is less than
300 μL, sample tube test volume can‘t less than 1.5ml.
★ If the machine detects a lack sample/reagent, result is * . *. User need reapply.
★ here will be a liquid level induction error warning when the machine is starting up.
Restart under the condition that the outlet water of the cleaning pool is normal. Set
to enter the none liquid level detection mode at this time: Not detect the volume of
samples and reagents, so please add enough reagents and samples to ensure the
accuracy of test results
1. Normal Test
a. Select sample number, ID number do not need to be modified
b. Select item combination、test item singly
c. Select the testing times(The default is 1 time), when testing repeatability we can
choose only one item
d. Press “Enter”, application is successful
e. Press “Start” , starting the test
f. ★ When you apply for the test after running out of 24 sample positions, you will
automatically jump to no. 1 to reorder, and the results of the previous test will be
automatically covered. Therefore, only 29(24 sample cup+5 tube) sample tests can
be started at a time.
2. QC test
a. Press QC, the system defaults quality control position is No. 26 (QC position)
b. Select the quality control test item
c. Press “Enter”, application is successful
d. Press “Start” , starting the test
3. Calibration test
a. Press “CAL”, the single point calibration is defaulted by No.25 position(CAL), and
the multi point calibration is defaulted by No.1 to No.5 positions
b. Set item parameters, select calibration item
c. Press “Enter" , “CAL” turn to blue, application is successful
2
d. Press “Start" , starting the test
e. The system automatically updates the F value of the item parameters after
calibration test is completed, the machine automatically prints the F value of the
calibration item
f. ★ In order to ensure the reliability of the calibration results, no operation is
allowed during the calibration test
g. ★ Single-point and multi-point calibration can not be applied for and started at
the same time, multi-point calibration can only apply for and start one item at a
time, single-point calibration can apply for and start more than one item at a time
4. Emergency test
a. Press “ST”, system defaults ST position is No.27(ST position)
b. Select ST item
c. Press “Enter”, application is successful
d. Press “Start" , starting the test
5. Eliminate
a. The tests that have been applied(blue) cannot be deleted
b. The sample that has been finished(green) can be deleted, press the sample
position to delete
3
7. Stop
a. In an emergency, press “Stop” key to stop the machine
b. ★ When press “Stop” key, the machine stops all actions immediately, and the
unfinished tests are all invalid and cannot be resumed
c. ★ When you need to retest after stopping, you need to turn off the machine then
restart
8. View results
4
consistency of the water blank test data
c. Water blank should be tested regularly during the continuous use of the same
cuvettes in order to eliminate the interference of the cuvettes aging to the water
blank. The best test period of the water blank is 1 time/day (after turn on the
machine, a water blank test before the test), Or the user should adjust the water
blank test cycle automatically according to the using frequency of the cuvettes
d. Ensure that the cuvettes are new cuvettes or washed cuvettes before the water
blank test
e. Press the water blank key for five seconds. The machine automatically tests the
water blank. After the test is finished, stop the machine and update the water
blank automatically. The water blank test process is: suction - cleaning once -
adding pure water - test the water blank - two times suction – stop
5
Setting
1. Item parameters setting
6
: Test type(Colorimetric/Blood/Turbidimetry)
2. Adding rule
The first period add R1, the Sixth period add S, the 13th add R2, maximum 25 period, 30
seconds/period
Incubation time → Incubation point: Incubation time(s)/30
For example: Incubate 300 seconds after added S, incubation point:300/30=10
Note:(Double reagent: Calculate incubation time after added R2 only, “R1+S” incubation time is
fixed time and does not need to be calculated and cannot be changed)
A: End-point
1. Single reagent End-point:
a. S.point: 3~5 (set any point)
b. E.point: 6+Incubation point, For example: Incubate 420s, E.point:6+14=20
7
Set: S.point=5, E.point=20
8
d. Immune turbidimetry (if requires S.point is after added R2)
• S.point:14
• E.point: 14+incubation point For example:Incubate 300s,
E.point:14+10=24
•
Set: S.point=14, E.point=24
9
2. Double reagent Kinetic&Two-point Kinetic
a. S.point:13+ Incubation time(s)/30 For example: Incubate 30s, S.point:
13+1=14
b. E.point: 13+Incubation point+Reading point For example:
Incubate 60s, then reading 2 minutes abs. E.point: 13+2+4=19
10
3. Item combination setting
a. Press “set” key——Input password to setting interface
b. Press “Item setting”
c. Select item combination button
d. Select or deselect the item
e. Press “Enter” key to save
f. ★ A item combination contains twenty items at the most
g. Press “modify” and then press “item combonation” to modify name
4. Debug command
11
5. Date and time
Press the corresponding column of date and time to modify
12
h. When the test results are completed, the machine should not be shut down while
it is cleaning, wait for the cleaning till complete
i. Refrigerate the reagent immediately after the test to ensure the stability of the
reagent
j. Clean the probe tip and cleaning pool with alcohol ever y day to ensure
cleanliness
k. All substances in contact with the machine are considered infectious,pay
attention to personal safety during operation
1. Test period:
a. The 1st period add R1
b. The 6th period add S
c. The 13th period add R2
13
e. 25 periods totally
f. The machine will clean the used cuvettes without adding sample in the period
g. The machine will clean all used cuvettes after the reaction is over until all the cups
have been cleaned
h. If turn off the machine in an abnormal condition, the machine will automatically look
for clean cuvettes for testing in the next test. If there is no usable cuvettes, it will clean
the cuvettes first till there are enough usable cuvettes to test
2. Calculation methods
a. End-point method:(end-point N)
b. Reaction range = end-point absorbance((N+N-1)/2) – star t point
absorbance
c. 2-point method:Reaction range = end-point absorbance - starting point
absorbance /60
d. Rate method:Reaction range=The rate of change in absorbance per minute
between the starting point and the end point
e. If set the dual wavelength,the main wavelength minus the secondary wavelength
and then calculate
3. Calibration formula
C(concentration)=Reaction amplitude * F value + b value
F value and B value are got by calibration, or inputting the values in the item
parameters
14
Data Data IP
Header Command Data Bytes Data description
length checksum checksum
0x0001 2 Sample ID
Syringe initial position 1. Sensor fault. 2. Syringe Restart,If they still appear, please contact
6
sensor fault mechanical fault. with maintenance engineer
Reaction tray Code 1. Sensor fault. 2. Reaction tray Restart,If they still appear, please contact
13
disc fault mechanical fault. with maintenance engineer
Reaction tray initial 1. Sensor fault. 2. Reaction tray Restart,If they still appear, please contact
14
position sensor fault mechanical fault. with maintenance engineer
15
Code Fault description Fault causes Solution
Lack of Water blank Machine cannot read tests of Retest water blank test, if they still appear,
29
tests water black please contact with maintenance engineer
Washing head sensor 1.Sensor fault. 2. Washing Restart,if they still appear, please contact
83
fault head mechanical fault with maintenance engineer
Probe up-down 1.Sensor fault. 2. Probe up- Restart,if they still appear, please contact
84
sensor fault down mechanical fault with maintenance engineer
Probe rotate sensor 1.Sensor fault. 2. Probe rotate Restart,if they still appear, please contact
85
fault mechanical fault with maintenance engineer
16