Professional Documents
Culture Documents
Turn the adjuster feet by hand until the sampler rises an even 5 mm
off the countertop.
• Check that the instrument has a clearance of about 2 cm from the
countertop
NOTE:
There are two hooks on each valve. Always check that the
tube is pinched in the REAR hook with an arrow mark.
III. Setting Up Eluents and Hemolysis Washing Solution :
1. Connect the fluid level detection sensor cord of hemolysis washing solution bottle.
Plug the fluid level detection sensor cord that extends from the
bottle cap H, into the WASH terminal on the rear panel.
Fit the tubes from the bottle caps A , B and CV into the grooves
in the tube guide
4. Attach the bottle caps with nozzle to the packs and bottle.
Insert the nozzle of the bottle cap A into the eluent A pack.
Tighten the cap securely.
5. Connecting the Drain Tubes
Connect the instrument to your lab drainage system or bottle for liquid waste. This instrument has two
drains: one for the optical unit and the other for liquid waste.
6. Connect the drain tubes to the instrument.
Fit one end of the optical unit drain tube into joint “D1”.
Fit one end of the liquid waste drain tube into joint “D2”.
Printer paper: See “4.2.3. Replacing the Printer Paper” on page 4-10.
Display contrast: See “5.5.1. If Nothing Appears on the Display” on page 5-22.
Date and time: See “3.5.1. Setting the Date and Time” on page 3-22.
Measurement mode: See “2.4. Measurement Mode Selection (Variant or Fast)” on page 2-14.
Reagent information: See “3.8. Reagent Information Settings” on page 3-38.
With liquid overflowing from the IN side, securely tighten the IN side push screw by hand and immediately wrap
the OUT side of the column with tissue paper.
• Liquid will overflow from the OUT side of the column.
Press[ Suspend ] or STOP when the appropriate column pressure appears on the display.
• Fluid pumping will stop
8 Close the covers.
Close the temperature control box cover, then the front cover.
9 Complete maintenance.
Press [ Finish ].
If “Reset counter?” appears, press .OK
• The instrument will record the maintenance date.
• The column measurement counter is reset and the [Reagent replacement menu] screen will appear again.
Perform HbA1C calibration (Calib) :
Perform HbA1C calibration before the instrument is used for the first time after installation.
Calibration provides operation coefficients (calibration coefficients) for correcting measurement results.
IMPORTANT:
Carefully read the package insert that comes with the calibrator before use.
Use the calibration rack to perform HbA1C calibration.
1. Setting the HbA1C Measurement Conditions :
On the standby screen, select [Menu]
<3 Measurement condition menu> and <4 HbA1c
measurement setup> (3400)
Set the HbA1C calibration conditions (when
1
needed).
IMPORTANT:
Do not set a sample cup in the port next to a sample tube with control. The sample tube will be recognized for
barcode reading and the control in the tube will not be measured. Also, the barcode of the sample tube will be
assigned to the sample in the sample cup.
NOTE:
Set sample tubes with barcode labels in the hemolysis control rack with the barcode labels facing the rear of
the rack.
4 Load the hemolysis control rack onto the sampler.
Load the hemolysis control rack onto the sampler.
IMPORTANT:
+ Measurement accuracy is controlled separately for the Variant mode and Fast mode.
+ Accuracy control reports can be printed for each mode or both modes (see “3.7.3. Printing Accuracy Control
Reports” on page 3-31).
FINISH .