4.1 Set Time
Enter the me by the keypad then confirm with Enter it will ask for date
Enter the year, month & date by the keypad and confirm by pressing Enter, it will go back to
service menu
2. Func on
1. History
1 . Profile history
2 . Batch history
2. Quality Control
1. QC Set Up
2. QC Edit
3. QC Graph
3. Set Up
1. Print
2. Auto Sense
3. Auto transmission
4. Pa ent ID
4. Service
5. Factory
SERVICE MANUAL
5. Factory
1. Blank Check
2. Set Temp
3. Ch Gain
4. Motor Speed
5. User Code
1. Blank Check
This menu is to set the blank value of the channels.
CH 1 : 32564
CH 2 : 367
Block the measuring channel then it will show the blank values for each channel press enter
to save the blank values. A er press enter it return to factory menu
2. Set Temp
This is to set the temperature of the measuring block. For this check with thermometer in
the cuve e holder for 10 min.
Off Set : 20
Heater : Off
Temperature: 37.2 C
To change the temperature adjust offset value.
eg. If the actual temperature is 39 .5 and display temp is 37.0 then increase the offset to
adjust the display temp by 25 . Then display temp also will become 39.5 C. To change offset
press 2 to select and enter by key pad. For confirming press Enter a er se ng to save press 3 ,
then it will return to factory menu.
SERVICE MANUAL
3. Ch Gain
This is to set the gain of the Channel 1 .
This will be adjus ng the ADC out put value by calcula ng the so ware gain.
The Maximum AD value is :
Adc –Ch1 : 27896
From this menu press enter to calculate and save the Gain. This is required for only
Channel1. A er press enter it will return to factory mode.
4. Motor Speed
This is to adjust the mixing motor speed.
Motor Speed : 6
To increase or decrease use the up and down key. Press Enter to confirm. Then it will return
to Factory mode
5. User Code
This is not used.
SERVICE MANUAL
7.3 Power Supply Module
SERVICE MANUAL 16
7.4 Printer
17
SERVICE MANUAL
7.5 Measurement Unit
SERVICE MANUAL 18
SERVICE MANUAL
8. Top Cover Assembled
SERVICE MANUAL 20
SERVICE MANUAL
10. Wiring Diagram
SERVICE MANUAL
Knonauerstrasse 54-6330 Cham - Switzerland
Tel: +41 41 780 60 10 Fax: +41 41 780 60 11 info@agappeswiss.com