Mispa COUNT

CONTENT
TECHNICAL CHARACTERISTICS
HEMATOLOGY PARAMETERS
TYPES OF MEASUREMENT
MEASUREMENT PRINCIPLE
HEMAGLOBIN MEASURING PRINCIPLE
INSTRUMENT VIEWS
A CROSS SECTIONAL VIEW
FRONT COVER
SAMPLING MODULE
NEEDLE RINSING ASSAY
SAMPLING MODULE X AXIS
SAMPLING MODULE Y AXIS
SAMPLING MODULE X & Y AXIS SECURING
COUNTING MODULE
COUNTING MODULE EXPLODED VIEW
COUNTING MODULE MANIFOLD
COUNTING CHAMBERS
COUNTING VALVE MODULE
SYRINGE VALVE MODULE
WASTE SYRINGE MODULE
WASTE SYRINGE BODY
WASTE SYRINGE MOTOR SUPPORT
REAGENT SYRINGE MODULE
REAGENT SYRINGE BODY
REAGENT SYRINGE MOTOR SUPPORT
FLUIDIC DIAGRAM
TUBINGS DIAGRAM
DILUENT LINE
LYSE LINE
CLEANER LINE
FLUIDIC PARAMETERS
CONNECTION BOARD
CPU BOARD
CPU MAIN COMPONENTS
CPU LED INDICATORS
CPU BOARD CONNECTIONS
LCD SCREEN
POWER SUPPLY CABLE
WASTE SYRING CABLE
ELECTROVALVES CABLE
FLUIDIC DOOR SWITCH CABLE
MOTOR CABLE
ON/OFF CABLE
USB CABLE
HB CABLE
WBC COAXIAL CABLE
RBC COAXIAL CABLE
START KEY SWITCH CABLE
TROUBLESHOOTING
Messages Description Action/Troubleshooting

Auto clean cycle initiated. Press on and wait the end of the cycle.

Note: Auto Clean frequency is adjustable in


SETTINGS/OTHERS screen.

Auto rinse cycle initiated Press on and wait the end of the cycle.

Note : Auto Rinse cycle is triggered after 30


minutes of standby.

Appears when trying to run a cycle after an Press on and perform a control cycle.
emergency stop.

Control cycle is mandatory after emergency


stop.
Vacuum stability check failed during syringe Press on to cancel the message.
vacuum
Perform VACUUM TEST then verify the value (410 mb +/-5%) and the
stability of the vacuum.

Vacuum failure during the needle rinsing. Press on to cancel the message.

1. Perform VACUUM TEST then verify the value (410 mb +/-5%) and
the stability of the vacuum.
2. Check tub #10 (pinched or clogged).
3. Check if diluent comes properly to rinsing head.
4. Check if needle O-ring damaged.
5. Check if needle guide released.
6. Check valve #3 (clogged or damaged).

Vacuum failure at the beginning of bath Press on to cancel the message.


draining
1. Verify diluent comes to the baths.
2. Perform VACUUM TEST then verify the value (410 mb +/-5%) and
the stability of the vacuum.
3. Verify valve #1; #2 (do not close properly).
Pressure failure during waste syringe Press on to cancel the message
draining.
1. Check tub #5 (pinched or clogged).
2. Check tub #23 (pinched or clogged).
3. Check valve #7 (clogged or damaged).

System detects waste container full. Press on , empty the waste container and reset waste level in
reagent menu.

Note: Waste container capacity is adjusted in


reagent menu.

The system detected the fluidic door opened. If the fluidic door is closed, verify the good functionality of the fluidic
door switch and its physical adjustment.

The adjustment of the HGB LED is failed. 1. Verify that WBC bath is filled with diluent during HGB LED
adjustment.
2. Change HGB board.
The system detected HGB command error Press on to cancel the window.
just after switching ON the instrument.

1. Check the HGB cable connection on HGB board.


Note: When this error occurs the HGB LED is 2. Check the HGB cable connection on CPU board.
not lighted ON. 3. Replace the HGB board.
4. Replace the CPU board.

Motors initialization is not done. Press on to cancel the window.

Perform motors initialization.


Ex: Instrument was switched ON with fluidic
door opened.

Steps lost on the needle motor. Press on to cancel the window.

1. Check if mechanical hard point on needle axis.


2. Check if needle bent.
3. Check the needle belt tension.
4. Check O-ring in rinsing head.