Professional Documents
Culture Documents
Glasmästar Et Al 2003 Silicone Transfer During Microcontact Printing
Glasmästar Et Al 2003 Silicone Transfer During Microcontact Printing
Microcontact printing (µCP) is a widely used method to make miniaturized patterns on surfaces. In this
work, the issue of the possible transfer of stamp material from the stamp to the substrate during stamping
was addressed. Poly(dimethylsiloxane) was used to stamp Milli-Q water or buffer on substrates of SiOx,
TiO2, and Au. X-ray photoelectron spectroscopy (XPS) and time-of-flight secondary ion mass spectrometry
(ToF-SIMS) were used to measure and characterize the substrate before and after stamping to detect the
possible transfer of stamp-related material to the substrates. Both the XPS and the ToF-SIMS analyses
show that silicone-related material is transferred from flat stamps and that even more material is transferred
See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.
from patterned stamps. Interestingly, a UV/ozone treatment (essentially oxidation of the surface) of the
stamps before inking and stamping significantly reduces the silicone transfer. Two application examples
are used to illustrate the importance of silicone transfer to the substrates during µCP: water condensation
patterns and supported lipid bilayer formation.
Downloaded via SEJONG UNIV on March 22, 2024 at 01:34:05 (UTC).
deviation from two data points per sample type. PDMS was Sylgard
184 elastomer, untreated or UV/ozone-treated (10 min).
nitrobenz-2-oxa-1,3-diazol-4-yl)amino]dodecanoyl-1-hexadecanoyl-
sn-glycero-3-phosphocholine (NBD-HPC; Molecular Probes,
U.S.A.). Vesicles produced by this method have an average
diameter of 40-60 nm (by dynamic light scattering). Glass
coverslips (Merck) were cleaned by sequential ultrasonic agitation
in acetone, 2-propanol, and Milli-Q water, dried, and UV/ozone-
treated for 10 min. The cleaned coverslips were then stamped
and subsequently incubated in a solution of vesicles (30 µL in 1
mL of buffer) for 10 min, rinsed with buffer, and imaged using
an Axioplan 2 Imaging epifluorescence microscope (Zeiss).
Figure 2. High-resolution XPS Si(2p) spectra from PDMS Results
stamps (Sylgard 184 elastomer), untreated or UV/ozone-treated.
The curves are displaced along the y axis for clarity and are XPS Analyses of Stamps. To be able to trace the
normalized with respect to the total signal. The spectra were contaminations coming from PDMS used in this study,
recorded using an electron gun to compensate for the charging XPS analyses of untreated and UV/ozone-treated flat
and were later shifted to C(1s) ) 284.4 eV for a charging PDMS stamps were performed (Sylgard 184 elastomer).
reference. The arrow points at a shoulder, indicating the
presence of more oxidized species. Figure 2 shows the high-resolution XPS spectra of the
Si(2p) region for untreated and UV/ozone-treated PDMS.
The XPS analysis of PDMS (Sylgard 184 elastomer) stamps For both untreated and UV/ozone-treated PDMS, the Si-
were performed using equipment and settings the same as those (2p) peak is asymmetric. This is much more pronounced
described previously, with the addition of an electron flood gun for the untreated stamp. The most dominant peak at 101.9
to compensate for the charging. The electron gun was adjusted eV of the untreated PDMS corresponds well to the
so that the O(1s) signal of PDMS was at about 532 eV. A result reference spectra for PDMS (101.8 eV).25 The shoulder at
of the charging neutralization process was that the total intensity a higher BE, at the arrow in Figure 2, indicates the
varied between the samples. Therefore, we chose to normalize presence of more oxidized species. After UV/ozone treat-
the curves of Figure 2 with respect to the total signal.
The spectra shown in Figures 2 and 3 were corrected for
ment, the peak at 101.9 eV is shifted toward a higher BE
charging effects by shifting the peaks as follows: PDMS spectra (102.5 eV), which is in agreement with the literature.9,26,27
were shifted so that the binding energy (BE) for C(1s) was 284.4 However, the peak is not shifted completely to the SiO2
eV; spectra from the Au samples were shifted to Au(4f7/2) ) 84.0 value (103.3 eV),28 indicating that the surfaces of the UV/
eV; spectra from the SiOx samples were shifted to Si(2p3/2) ) 99.3 ozone-treated stamps have a surface chemistry somewhere
eV; and spectra from the TiO2 samples were shifted to Ti(2p3/2) between PDMS and SiO2. The atomic composition of the
) 458.8 eV. stamp surface (Table 1) shows a decrease in carbon and
ToF-SIMS Analysis. The ToF-SIMS analyses were performed an increase in oxygen content upon UV/ozone treatment.
using a ToF-SIMS IV (ION-TOF GmbH) system in static mode,23 XPS Analyses Using Flat Stamps. The first set of
that is, the ion dose was below 1013 ions/cm2. Gallium ions (25
stamping experiments was made with untreated and UV/
kV; 1 pA) were used as the primary beam. Both high mass
resolution spectra (m/∆m ) 7000) and high lateral resolution ozone-pretreated stamps, using Milli-Q water and buffer
images (<1 µm) were recorded. Flat stamps and stamps with solutions, to stamp SiOx, TiO2, and gold surfaces. Parts
line patterns, with and without UV/ozone pretreatment, were A-F of Figure 3 present, for each surface type, the XPS
used to stamp Milli-Q water on Au substrates. The analyses spectra taken before stamping, after stamping with
started at about 80 min after preparation. untreated stamps, and after stamping with UV/ozone
Light Microscopy. Dark-field light microscopy was used to pretreated stamps, using Milli-Q water as the ink. The
inspect certain aspects of the quality of the stamped patterns. main observations are that (i) the transfer of material
The microscope used was a Zeiss Axiotech microscope equipped occurs from untreated stamps on all the surface types
with a digital camera. Water condensation patterns were made
examined and (ii) a pretreatment of the stamp with UV/
by placing the sample on an ice-cooled piece of copper, under the
microscope objective, in an environment of humid air and by ozone lowers the amount of PDMS (Sylgard 184, Dow
letting water condense on the sample surface. Corning) transferred from flat stamps to substrates of
Another useful control system, sensitive to the chemical state Au, SiOx, and TiO2.
of the surface, that has been extensively studied in our group, The XPS analysis of PDMS transfer to the SiOx substrate
is the formation of supported phospholipid bilayers by vesicle is complicated by the fact that the substrate itself has a
adsorption and rupture (see, e.g., ref 24). A lipid bilayer will strong Si signal. The transfer from untreated PDMS
form on a clean glass surface if it is exposed to lipid vesicles. stamps is, however, still detectable by the Si(2p) signal
Fluorescently labeled unilamellar lipid vesicles were made by in the XPS spectra (Figure 3A), although a quantification
extrusion (11 × 100 nm followed by 11 × 30 nm polycarbonate is not possible without extensive (and often unreliable)
membranes, Whatman, U.S.A.) of a 5 mg/mL lipid mixture in a
Tris buffer of 1-palmitoyl-2-oleoyl-sn-glycero-3-phosphocholine
(25) Beamson, G.; Briggs, D. High resolution XPS of organic polymers;
(POPC; Avanti Polar Lipids, AL) and 0.5% (w/w) of 2-[12-(7- John Wiley & Sons: Chichester, U.K., 1992.
(26) Bar, G.; Delineau, L.; Hafele, A.; Whangbo, M. H. Polymer 2001,
(23) Vickerman, J. C., Briggs, D., Eds.; ToF-SIMS: Surface analysis 42, 3627.
by mass spectrometry; IM Publications and SurfaceSpectra Limited: (27) Murakami, T.; Kuroda, S.; Osawa, Z. J. Colloid Interface Sci.
Huddersfield, U.K., 2001. 1998, 202, 37.
(24) Keller, C. A.; Glasmästar, K.; Zhdanov, V. P.; Kasemo, B. Phys. (28) Chastain, J., Ed.; Handbook of X-ray Photoelectron Spectroscopy;
Rev. Lett. 2000, 84, 5443. Perkin-Elmer Corporation: Eden Prairie, MN, 1992.
5478 Langmuir, Vol. 19, No. 13, 2003 Glasmästar et al.
Figure 3. High-resolution XPS spectra from samples not stamped or stamped with untreated or UV/ozone-treated flat stamps
(Sylgard 184 elastomer) using Milli-Q water as the ink. All the spectra from the SiOx samples were shifted to a BE for Si(2p3/2)
of 99.3 eV, the TiO2 samples to a BE for Ti(2p3/2) of 458.8 eV, and the Au samples to a BE for Au(4f7/2) of 84.0 eV. For clarity, the
curves are displaced on the y axis. (A) Si(2p) spectra of SiOx samples. (B) C(1s) spectra of SiOx samples. (C) Si(2p) spectra of TiO2
samples. (D) O(1s) spectra of TiO2 samples. (E) Si(2s) spectra of Au samples. (F) O(1s) spectra of Au samples.
curve fitting. Inorganic silica (SiO2) has a peak at 103.3 BE (∼102 eV), which is not observed on the samples where
eV,28 which is clearly visible on all the SiOx samples. UV/ozone-treated stamps were used. The XPS spectra from
However, for the samples that were stamped with an the untreated PDMS stamps themselves (Figure 2) show
untreated stamp, this peak shows a shoulder at a lower a nice correlation with this shoulder observed at 102 eV
Silicone Transfer Langmuir, Vol. 19, No. 13, 2003 5479
Table 2. Atomic Composition by XPS Survey Spectra of the Samples Stamped with Flat PDMS Stampsa
substrateb stamp C(1s) O(1s) Si(2p/2s)c Au(4f) Ti(2p)
Au untreated stamp 29.3 ( 5.7 22.3 ( 2.4 6.2 ( 1.6 41.5 ( 3.4
treated stamp 34.5 (5.5 14.0 ( 2.0 0.2 ( 0.4 50.3 ( 4.9
not stamped 32.3 ( 8.8 16.2 ( 3.8 0 48.8 ( 4.8
TiO2 untreated stamp 24.3 ( 4.8 50.7 ( 3.7 7.4 ( 3.4 15.6 ( 3.4
treated stamp 15.0 ( 0.9 58.1 ( 0.6 3.6 (0.5 20.7 ( 1.1
not stamped 21.7 ( 7.4 55.5 ( 4.3 1.5 ( 0.1 19.1 ( 2.9
SiOx untreated stamp 16.9 ( 2.1 38.8 ( 0.7 44.3 ( 1.4
treated stamp 13.9 ( 2.0 40.3 ( 0.5 45.6 ( 1.4
not stamped 12.9 ( 0.6 40.9 ( 1.1 46.1 (0.7
a Mean values are noted in percent plus or minus the standard deviation of at least four data points per sample type. b For some of the
Au samples, a contamination of Cu was found (max 4%), probably a result of a contamination of the evaporation source. All the Ti samples
had fluorine contamination (max 2%) from the oxygen plasma equipment. c The Si(2p) peak was used for all but the Au substrates, where
the Si(2s) peak was used. The ink used was Milli-Q water; the contact time was 3 min. The stamps were made of Sylgard 184 elastomer,
untreated or UV/ozone-treated (10 min).
(285.1 eV) for the samples stamped with untreated stamps deviation. To be compared to when flat stamps were used, see Table
as compared to that of the samples that were not stamped 2. Patterned stamps were 5-µm-wide lines separated by 15-µm
or stamped with UV/ozone-treated stamps. This is prob- spaces. The ink used was Milli-Q water; the contact time was 3
ably a mixed peak of atmospheric hydrocarbons and carbon min. The stamps were made of Sylgard 184 elastomer, untreated
or UV/ozone-treated (10 min).
from PDMS, the latter having a peak at 284.4 eV.25
For the stamped TiO2 samples, the results were similar
to the ones described for the SiOx samples previously. An deposited on the surface as a result of the ozone treatment.
intensity change of the Si(2p) peak (Figure 3C) was (Our BE values are slightly higher than those reported
observed on the samples stamped with untreated stamps. by Krozer and Rodahl.) The O(1s) signal from the samples
The Si(2p) peak (Figure 3C) shows that some Si-containing stamped with untreated PDMS results in a dominant peak
species are also present on the sample that was not at a higher BE (531.8 eV), which can be associated with
stamped. The latter probably originates from wafer dust the presence of PDMS (532 eV).25
(remember that Ti was deposited on a Si wafer) produced When the observations in Figure 3A-F are summarized,
when snapping the chips or maybe from the reactive ion the most important result is that the UV/ozone-treated
etcher used to oxidize the TiO2 samples. The atomic stamps transfer less PDMS-derived material than the
compositions, as were calculated from the survey spectra untreated ones. This result is further supported by the
(Table 2), show a significantly higher Si content for the following description of the ToF-SIMS results. If we
stamped samples than that for the not stamped ones. compare the different surface types, we find that the UV/
Furthermore, the UV/ozone treatment of the stamp seems ozone pretreatment is more efficient in lowering the
to reduce the transfer of Si-containing species to the TiO2 transfer to gold surfaces than that to TiO2 surfaces (Table
surface. 2). Unfortunately, it is hard to perform a quantification
of the transfer to SiOx surfaces because the silicon peaks
The “normal” shape of the O(1s) peak for TiO2 surfaces
from PDMS and the substrate overlap.
is observed for the sample that was not stamped (Figure
That PDMS sticks less to Au than to TiO2 substrates
3D). The tail at a higher BE is assigned to the organic
is actually supported by observations from the casting
compounds bound to the surface.29 An additional peak at
process: PDMS cured on masters of plain SiOx or TiO2-
532.3 eV is found for the samples stamped with the
coated wafers was extremely hard to release. In contrast,
untreated stamps, which corresponds to the O(1s) peak
when Au-coated wafers were used as masters, PDMS was
for PDMS (532.4 eV, not shown). Also, the fact that the
quite easily released after the curing. Hence, it appears
intensity of the TiO2 oxygen peak is reduced when
that the direct chemical interaction between PDMS and
stamping with an untreated stamp supports the conclusion
the substrate is stronger for the two oxide surfaces
of a transferred adlayer.
compared to that of gold.
For the Au samples, the Si(2p) signal cannot be used
All the data presented here are from the µCP of Milli-Q
to identify the PDMS transfer because it is hidden in the
water using the Sylgard 184 elastomer. Trends the same
large background of the Au(4f) peak. However, the Si(2s)
as those described previously were found when stamping
signal (Figure 3E and Table 2) clearly shows that
PBS and also when using another brand of PDMS that is
Si-containing species are transferred to the gold surface
less commonly used for µCP: Rhodorsil RTV 1556 (data
from untreated stamps whereas the UV/ozone-treated
not shown).
stamps transfer only trace amounts (if any). A shape
XPS Analyses Using Patterned Stamps. Types of
change and shift of the O(1s) signal were also observed
experiments the same as those described previously were
(Figure 3F). For not-stamped, UV/ozone-treated Au, a
performed using the gold samples and patterned stamps
double peak is observed in the BE range of 529-531 eV.
having 5-µm-wide lines separated by 15-µm-wide spaces
According to the literature,30 these peaks correspond to
(depth 10 µm). The elemental concentration from the XPS
chemisorbed oxygen (529.1 eV) and gold oxide (530.1 eV)
survey spectra of gold surfaces stamped with patterned
(29) Lausmaa, J.; Kasemo, B.; Mattsson, H. Appl. Surf. Sci. 1990,
stamps are shown in Table 3. These stamps seem to
44, 133. transfer more Si-containing species to gold substrates than
(30) Krozer, A.; Rodahl, M. J. Vac. Sci. Technol., A 1997, 15, 1704. the flat stamps do. This conclusion is drawn from the
5480 Langmuir, Vol. 19, No. 13, 2003 Glasmästar et al.
Figure 4. ToF-SIMS images for selected fragments characteristic of PDMS. The chemical formula of each fragment is marked
under the respective image. (A) A gold surface stamped with an untreated PDMS stamp (Sylgard 184; Milli-Q water as the ink).
The stamp was in contact with the gold surface at 5-µm-wide lines separated by 15-µm-wide spaces. The scan size was 101 × 101
µm2. (B) Same as part A except that the stamp in this case was UV/ozone-treated for 10 min before stamping. The scan size was
101 × 101 µm2.
Table 4. ToF-SIMS Signals from the Au Samples gold samples. The Au (m/z ) 197+) signal is used as an
Stamped with Flat or Patterned Stamps intensity reference, and we focus on the signals charac-
flat stamp patterned stamp teristic for PDMS in the low-mass-range spectra,31 that
UV/O3 UV/O3
is, m/z ) 73+, 147+, 207+, 221+, and 281+. The contact
signal untreated treated untreated treated area between the stamp and the gold surface consists of
5-µm-wide, elevated lines, separated by 15-µm-wide
gold/total 249 814 89 487 spaces. A Au substrate stamped with an untreated stamp
ion count
and analyzed by ToF-SIMS in image mode shows the
a The numbers given are averages over the whole image (101 ×
expected pattern (Figure 4A). Most of the silicone transfer
101 µm). The lower the gold signal, the more adsorbates on the has occurred in the regions where the stamp was in contact
surface. For comparison, the gold/total ion count signal is 917 for with the surface during stamping (i.e., the 5-µm-wide lines
a sample that was not stamped. Patterned stamps were 5-µm-wide
lines separated by 15-µm spaces. The ink used was Milli-Q water; are the brightest). However, when the stamp was pre-
the contact time was 3 min. The stamps were made of Sylgard 184 treated with UV/ozone, the ToF-SIMS image shows more
elastomer, untreated or UV/ozone-treated (10 min). intensity from the PDMS-related fragments (relative to
the Au signal) in the areas of noncontact (see Figure 4B),
comparison of the data in Table 3 with the corresponding which is surprising.
values for flat stamps on gold in Table 2. Apart from the The surface topography influences the probability of
peak intensities, no qualitative differences (such as ionization of the surface molecules when imaging with
differences in the relative peak heights or peak shapes) ToF-SIMS.23 To rule out the possibility of such effects as
were observed between the spectra from surfaces stamped the cause of the unexpected ToF-SIMS images obtained
with flat or patterned stamps. with UV/ozone-treated stamps, tapping mode atomic force
There is a possibility that silicone monomers/fragments microscopy was used to image a gold sample stamped with
were transferred through the vapor phase in the regions patterned, untreated PDMS stamps (lines). No lines were
where the patterned stamp was not in contact with the detected in the height image (0-10 nm), but lines of
surface during stamping. To test this hypothesis, the XPS dimensions the same as those of the contact area were
analyses were performed for the gold samples exposed to clearly seen in the phase contrast image (not shown). The
vapors from the untreated or UV/ozone-treated flat stamps latter observation suggests that there is a chemical pattern
(see Materials and Methods for details). No silicon was on the surface.32
detected in either case [Si(2s); pass energy 5.85 eV; The most important conclusions from the ToF-SIMS
acquisition time 17 min over 10 eV]. analysis are: (i) It supports the conclusion from the XPS
ToF-SIMS Analyses Using Flat Stamps. The ToF- results that UV/ozone pretreatment of the stamp is
SIMS analyses of the gold samples, stamped with flat efficient in lowering the PDMS transfer from the stamp
stamps, showed a trend the same as that in the XPS to the substrate. (ii) When PDMS transfer occurs from
analyses: less PDMS is transferred if the stamp is UV/ozone-treated stamps, transferred PDMS is not located
pretreated with UV/ozone than if it is untreated. The only in the areas where the stamp and substrate were in
intensity of the gold peak (m/z ) 197+ normalized to the direct contact. (iii) There is more transfer from patterned
total ion intensity) is lower for the samples stamped with than from flat stamps.
untreated stamps than those stamped with UV/ozone- These samples were also investigated by dark-field
treated stamps; see Table 4. This suggests that more microscopy. We found that on both the TiO2 and the SiOx
material is transferred to the sample from an untreated surfaces, PDMS transferred from patterned, untreated
stamp than from a UV/ozone-treated stamp, that is, the stamps could be seen as brighter lines with dark-field
lower the gold intensity, the thicker the layer of adsorbed microscopy. Nothing could be seen when the stamp was
molecules. There is also a lower gold signal after stamping UV/ozone-treated before stamping or on the Au surfaces
with a patterned stamp compared to that of a flat one, that were stamped with untreated stamps.
indicating that more material is transferred from a
patterned stamp than from a flat. (31) Dong, X.; Proctor, A.; Hercules, D. M. Macromolecules 1997, 30,
63.
ToF-SIMS Analyses Using Patterned Stamps. Pat- (32) Pang, G. K. H.; Baba-Kishi, K. Z.; Patel, A. Ultramicroscopy
terned stamps were used to stamp Milli-Q water on the 2000, 81, 35.
Silicone Transfer Langmuir, Vol. 19, No. 13, 2003 5481
Discussion
µCP Using Flat Stamps. There is clear evidence from
the present study for a transferred overlayer of molecules
derived from PDMS, on substrates subjected to µCP with
untreated PDMS stamps (both flat and patterned), when
buffer or Milli-Q water is used as the ink. This conclusion
is based on the XPS and ToF-SIMS data of the present
study.
Surprisingly few studies in the literature have aimed
at elucidating the problem of transferral of stamp material.
Graham et al.21 recently published the only (to our
knowledge) more detailed study on this issue. They
compared the self-assembled monolayers on gold made
by either adsorption from ethanolic solution or µCP with
flat stamps. They found that PDMS is transferred to the
surface by the µCP process (using XPS, ToF-SIMS, and
mulitvariate analyses). The authors present an extensive
cleaning protocol for the PDMS stamps, including extrac-
tion in hexane and sonication in an ethanol/water mixture
to avoid the transfer. The cleaning protocol takes about
Figure 5. (A) Dark-field light microscopy (20× objective) of a 1 week to perform according to Graham et al.21 Addition-
gold sample stamped with a UV/ozone-treated stamp (Sylgard ally, they blotted the stamp on a cleaned piece of silicon
184; Milli-Q water as the ink) having a 10 by 10 µm grid pattern
(the squares were not in contact). Water was subsequently
before stamping and claim that this was an effective way
allowed to condense on the surface, producing this pattern. (B) to remove residual silicone that had migrated to the surface
Fluorescence microscopy image of NBD-labeled phospholipid during cleaning. By using this cleaning protocol, they were
vesicles backfilled onto a glass surface “pre-stamped” with an able to lower the PDMS transfer from flat stamps during
untreated PDMS stamp (Sylgard 184; Milli-Q water as the ink) printing in ethanol to levels below that of XPS detection,
of a design the same as that in part A. The upper left corner although PDMS was still found by ToF-SIMS analysis.
shows a black area where the sample dried during the
experiment and, thus, is free of lipid. Comparing our data for the samples stamped with
untreated, flat stamps with the corresponding data for
Condensation Figures and Bilayer Formation. To samples stamped with UV/ozone-treated stamps (Figure
test for differences in chemistry between the underlying 3A-F and Table 2), we find that 10 min of UV/ozone
substrate and any transferred material, water from humid treatment reduces the transfer significantly. We, thus,
air was allowed to condense on the gold samples stamped show that the simple and fast UV/ozone treatment of the
with patterned, UV/ozone-treated or untreated stamps stamp, for the present conditions, significantly reduces
(with Milli-Q water as the “ink”). The pattern of the stamp the amount of PDMS transferred during µCP to levels at
was a square grid of 10-µm-wide, elevated lines. The or below that of XPS detection. The same trends were
squares between the lines were 10 by 10 µm and were not found for a second brand of PDMS elastomer: Rhodorsil
in contact with the surface during stamping. Upon RTV 1556 (only XPS data).
condensation, clear patterns were formed. The details of µCP Using Patterned Stamps. The combined results
the patterns varied from run to run because the conden- from XPS and ToF-SIMS indicate that PDMS is trans-
sation is very sensitive to minor variations in the ferred both in the noncontact and in the contact regions
condensation conditions, primarily the humidity level and between the untreated stamp and the gold substrate
temperature. One condensation pattern is shown in Figure (Tables 2-4). This is despite only one-fourth of the stamped
5A for a UV/ozone-treated stamp. A clear pattern of the surface being in contact with the patterned stamp, which
expected dimensions was observed, demonstrating a intuitively would suggest less transfer. The XPS values
contrast in hydrophobicity/hydrophilicity between the are mean values over the analyzed area and, thus, a mean
contact/noncontact areas. The patterns were also seen with over the areas that were in contact with the stamp and
untreated stamps (not shown). the areas of noncontact. (These experiments were only
In another “application example”, fluorescently labeled, performed on the gold samples.) It is likely that the
extruded unilamellar vesicles of POPC were adsorbed on mechanism of transfer of PDMS is different in the areas
glass surfaces “pre-stamped”, as in the condensation of contact versus the areas of noncontact.
experiment previously described, using the same types of As for flat stamps, the transfer of PDMS from patterned
stamps, that is, untreated or UV/ozone-treated. It is known stamps is reduced by a UV/ozone pretreatment of the
that lipid bilayers are formed by the adsorption and stamp. However, the overall reduction of PDMS transfer
subsequent rupture of vesicles on clean glass surfaces24 is smaller for patterned stamps compared to flat stamps
and that they can be studied using fluorescence imaging (Tables 2-3). The distribution of the transferred PDMS
and fluorescence recovery after photobleaching.33 The first on the stamped surface, which for UV/ozone-treated
test results show that the fluorescent patterns formed on stamps is predominantly detected in the regions of
the surface when the stamp was both untreated (Figure noncontact, may indicate that the majority of the material
5B) and UV/ozone-treated (not shown). Homogeneous is not transferred across the contact interface. Some
fluorescence indicates that the lipid film on the surface possible mechanisms for the transfer of PDMS onto the
is continuous and of the same thickness everywhere. On portions of the surface not in contact with the stamp are
the basis of the intensity variations, one could attribute shown schematically in Figure 6. One mechanism that
the bright areas to bilayers and the less bright areas to may involve volatile PDMS fragments from the “ceiling”
monolayers. above the areas of noncontact being transported across
the vapor phase during the stamping process (option 1 in
(33) Boxer, S. G. Curr. Opin. Chem. Biol. 2000, 4, 704. Figure 6) seems to be ruled out by the XPS experiment
5482 Langmuir, Vol. 19, No. 13, 2003 Glasmästar et al.
pattern for both treated and untreated, patterned stamps. in Borås, Sweden, for running the ToF-SIMS and for
The transfer of PDMS to the surface during µCP is relevant fruitful discussions about the results. We also thank Sofia
to and calls for attention by those who are using this Svedhem for making the fluorescently labeled vesicles
method in applications where control of the surface and Sarunas Petronis for producing the masters. Financial
chemistry is of importance for the application. support for this study was provided by the Swedish
Foundation for Strategic Research through the Screening
of Tissue Integrated Materials project.
Acknowledgment. The authors thank Peter Sjövall
at the SP Swedish National Testing and Research Institute LA026558X