Part Number: SP00209
Rev: -
Date: 14 January 2002
© 2002 Draeger Medical, Inc.
Julian
Pfgf/Psys Pressure PCB
8603361 Installation
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installation:
NOTE: If pressure sensor Pfgf or Psys is 6. To install the new pressure sensor PCB
faulty, the existing Pressure Sensor (8603361), you must first remove the
PCB (8201761) must be replaced piggyback PCB to access the mounting
with new Pressure PCB (8603361). holes. Transfer the O-rings from the old
pressure PCB (8201761) to the new
1. Switch off Julian and remove AC power. pressure PCB. Secure the pressure PCB
with screws, then reinstall the
2. Remove Julian’s rear panels and piggyback PCB.
disconnect the battery.
CAUTION: The upper and lower boards of
3. Remove the breathing system and pull the Pressure PCB are matched. Plug in only
out the pneumatic plug-in unit. the corresponding piggyback board.
4. Remove the EDOS block (see Anesthesia 7. Mount the connector of the enclosed
Equip & Monitoring CD-ROM/Julian/ cable harness onto the Actuator PCB
Repair Instructions/EDOS/Removing (see Figure 1). Plug the 5-pin white
EDOS). connector into X18 (A) (left-aligned).
Remove jumper X25 (B) and mount the
5. Remove the Pressure Sensor PCB 3-pin black connector including the
(8201761) from the face of the EDOS connector protection. The two cables of
block. the same color point downwards (jumper
simulation).
CAUTION: Be careful when pulling out the
sensors. Do not lose the O-rings; you will 8. Reinstall the EDOS block and join both
need them for later installation. connectors to the Pressure PCB (see
Figure 2).
1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
X18 X8 X2
1
1 A X19
B1
X20 X40
Silicone hose B2
C91
G1
X14
1
X16
X9 L1
C95 +
X41
K1
X43 X5
X12
X15
N16
X25 B X42
X21 X7
X6
X22
1
X13
X1 X3
1
Reset Button
2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Checking:
3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS