You are on page 1of 9

AJCP / ORIGINAL ARTICLE

Magnesium Sulfate as an Alternative In Vitro


Anticoagulant for the Measurement of Platelet
Parameters?
Steffen Mannuß, Peter Schuff-Werner, Katrin Dreißiger, and Peter Kohlschein

From the Rostock University Medical Centre, Institute of Clinical Chemistry and Laboratory Medicine, Rostock, Germany.

Key Words: Platelet count; Mean platelet volume; Preanalytics; Cell swelling; Anticoagulation; Magnesium sulfate; EDTA; Citrate

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


Am J Clin Pathol June 2016;145:806-814

DOI: 10.1093/AJCP/AQW066

ABSTRACT In the routine hematology laboratory, EDTA and citrate


are the most commonly used anticoagulants for measuring
Objectives: There are conflicting reports on the reliable
platelet parameters; for special analyses such as impedance
measurement of platelet count and mean platelet volume aggregometry, hirudin is the optimal anticoagulant. The ad-
(MPV) using EDTA or citrate. The anticoagulant properties vantages and disadvantages of these anticoagulants have
of magnesium sulfate (MgSO4) are known from the litera- been reviewed in detail by Banfi and colleagues.1 Exact
ture. The aim of this study was to evaluate MgSO4 as an platelet count is of importance in thrombocytopenia, espe-
in vitro anticoagulant for platelet count, MPV, platelet dis- cially if the necessity for platelet transfusion is discussed.2
tribution width, and platelet activation. The exact and reproducible determination of platelet vol-
Methods: Whole blood from volunteers was anticoagulated ume is of interest as a predictive biomarker, for example, in
by EDTA, citrate, or MgSO4. Platelets were counted by the cardiovascular disease.3,4 However, there are conflicting re-
XE 5000 (Sysmex, Norderstedt, Germany) impedance and ports on the correct anticoagulant and the best method for
fluorescence optical technique. platelet counts and volume measurements; therefore, further
standardization is necessary.
Results: The mean impedance platelet counts were 227.7, EDTA was introduced as an in vitro anticoagulant in
197.0, and 201.1  109/L in EDTA-, citrate-, or MgSO4- hematology because of its calcium-complexing capacity,
anticoagulated blood, respectively. The counts were 4.7% which prevents clotting and therefore stabilizes the fluidity of
higher (EDTA) after 3 hours of storage but 4% lower in cit- whole blood. Blood smears from EDTA-anticoagulated blood
rate-anticoagulated blood. The counts in magnesium sam- were shown to be optimal for morphologic blood cell differen-
ples remained stable. The MPV was 10.4 fL (EDTA), 9.5 fL tiation using May-Grünwald-Giemsa (Pappenheim) staining.
(citrate), and 9.3 fL (MgSO4). EDTA samples showed cell The only but relevant disadvantage of EDTA-anticoagulated
swelling within the first 3 hours. This was lower in citrate blood is the time-dependent swelling of platelets.5
and only marginal in magnesium samples. High activation Sodium citrate is mainly used for the study of the plas-
of platelets was observed only in EDTA samples. matic coagulation system but is also suitable for performing
a CBC.6
Conclusions: Magnesium anticoagulation might be advan-
Magnesium sulfate (MgSO4) has been known as an
tageous for more reliable MPV measurements. Although
in vitro anticoagulant since Sahli7 introduced it for conven-
platelet count is underestimated when the impedance
tional platelet count by referring to Bizzozero’s experiences
method is used, the platelet count reveals similar results
with the use of MgSO4 for avoiding platelet clumps in glass
when measured by the fluorescent optical method.
tubes. In the late 1980s, this anticoagulant was successfully
used in individuals with EDTA-introduced pseudothrombo-
cytopenia (PTCP)8 and also for hematology tests,9 but it
was never introduced for routine purposes. Our own positive

806 Am J Clin Pathol 2016;145:806-814 © American Society for Clinical Pathology, 2016. All rights reserved.
DOI: 10.1093/ajcp/aqw066 For permissions, please e-mail: journals.permissions@oup.com
AJCP / ORIGINAL ARTICLE

experiences with MgSO4 in patients with PTCP led to the 9NC; fill volume of blood 2.9 mL), and MgSO4 (S-Monovette
commercial launch of a blood sampling tube prepared with ThromboExact; fill volume of blood 2.7 mL), all manufac-
MgSO4 as an anticoagulant. In a multicenter study10 of anti- tured by Sarstedt.
coagulant-induced PTCP, magnesium anticoagulation was The order of blood sampling was the following: (1)
shown to be effective, although platelet counts and platelet EDTA, (2) citrate, and (3) MgSO4. Blood from the healthy
volume seemed to be underestimated compared to when volunteers was used for the above-described time kinetics
using EDTA blood.11 and measured within 5 to 10 minutes after sampling.
The real platelet volume is still a matter of controversy, Patients’ blood sampling was performed in the casualty
especially if one takes into account that EDTA as well as units of the hospital, located either in the same or in the
citrate anticoagulation leads to early platelet swelling.5 This nearby buildings. Due to this fact, the time period between
is a critical issue of preanalytical standardization for meas- sampling and measurement was between 30 and 45 minutes.
uring the mean platelet volume (MPV) as a possible but
controversially discussed predictive biomarker.12-14

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


Automated Determination of Platelet Parameters With
To test MgSO4 as an alternative to EDTA and citrate, the XE 5000
we compared the platelet parameters measured up to
All specimens were routinely analyzed with the auto-
180 minutes from healthy individuals. Finally, we con-
mated hematology analyzer (XE-5000; Sysmex, Norderstedt,
firmed our findings by a comparative study of platelet
Germany). This analyzer provides two options for platelet
counts and MPV in 834 patients’ blood samples.
counting: impedance and fluorescence optical technology.
The platelet count correlates with the number of im-
pulses triggered from cells sized between 2 and 30 fL. The
Materials and Methods height of the impulse size is calculated and leads to a histo-
Platelet counts and MPV were measured 20, 40, 60, and gram of platelet volumes. Platelet number and MPV are
180 minutes after sampling from blood of 21 healthy volun- estimated on the basis of the platelet histogram surface,
teers. The samples were anticoagulated by EDTA, citrate, which is limited by the lower and upper volume discrimin-
and MgSO4. Only apparently healthy individuals without ator.15 For comparison reasons, we used the fluorescence
medication up to 7 days before blood donation and not older method of platelet counting in parallel.
than 65 years or younger than 18 years were included in this Because of the dilution effect, the platelet counts in cit-
study. The group of volunteers involved 11 men (mean age rate-anticoagulated samples were multiplied by a factor of 1.1.
33.8 years, ranging from 22-65 years) and 10 women (mean The analyzer was calibrated and maintained according
age 44.3 years, ranging from 24-62 years). to the manufacturer’s instructions. Internal quality control
Platelet count and volume of the first consecutive 834 (QC) was performed and samples were assayed only if all
patients of an ongoing PTCP prevalence study (371 women QC criteria were fulfilled.
with a mean age of 59.3 years, ranging from 19-93 years, The imprecision of platelet count, displayed by the co-
and 463 men with a mean age of 60.4 years, ranging from efficient of variation, of the impedance technology/fluores-
14-93 years) were analyzed after the exclusion of EDTA- or cence optical technology using differently anticoagulated
citrate-induced PTCP by comparing the platelet counts with sampling devices was 2.07%/2.19% (EDTA), 1.37%/3.71%
those from MgSO4-anticoagulated blood. (citrate), and 1.23%/2.28% (MgSO4), respectively. The im-
This study was approved by the Ethics Committee of precision of MPV determination was 2.37% (EDTA),
the University of Rostock (registration no. A 2013-0104). 2.46% (citrate), and 1.56% (MgSO4).
Written consent is generally given by the patient when
admitted to the hospital but can be withdrawn by the patient Statistical Analysis
for personal reasons. The statistical analysis (Wilcoxon signed-rank test and
Spearman rank-order correlation) was performed with the
Blood Sampling software SigmaPlot 11. To show the distribution of all quan-
After releasing the tourniquet, whole blood was collected titative values within a group of samples, they are displayed
from lying or sitting patients and volunteers by venipuncture by box plots. Each box describes the 25th and 75th percent-
of the cubital vein using a 21-gauge butterfly needle iles, including 50% of the measured values. The median
(Sarstedt, Nümbrecht, Germany) with a respective adapter. value is depicted by a line inside the box. The whiskers
Blood was collected in commercial tubes prefilled with mark the 10th and 90th percentiles. The black dots mark the
the respective anticoagulant: EDTA (S-Monovette K3-EDTA; 5th and 95th percentiles. Box plots allow the graphical com-
fill volume of blood 2.6 mL), natrium-citrate (S-Monovette parison of data and their variation.16 Standard error of the

© American Society for Clinical Pathology Am J Clin Pathol 2016;145:806-814 807


807 DOI: 10.1093/ajcp/aqw066
Mannuß et al / MAGNESIUM SULFATE AND PLATELET ANALYSIS

mean was used for plotting the results of comparing the im- 400
pedance and the fluorescent optical method for platelet
350
count in differently anticoagulated blood samples.

Platelet Count (× 109/L)


300

250
Results
200
Time Dependency of the Measurement of Platelet
150
Parameters in EDTA-, Citrate-, and MgSO4-
Anticoagulated Blood Samples Within 180 Minutes of 100
Venipuncture
Platelet count and MPV of 21 volunteers were meas- 50

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


<10 20 40 60 180
ured in differently anticoagulated blood samples using the Time After Sampling (min)
impedance method of the XE 5000. In nine of the 21 sam-
ples, the optical method was applied in parallel. The first Figure 1 Platelet count measured by the XE 5000 (imped-
measurement was done within 10 minutes of venipuncture, ance method) in samples of 21 volunteers anticoagulated
followed by further determinations after 20, 40, 60, and with EDTA (white), citrate (light gray), and magnesium sul-
180 minutes. fate (dark gray).

Time-Dependent Changes in Platelet Counts


measured in EDTA blood by impedance (PLTimp, 217 6 49
As depicted in Figure 1 , the mean 6 SD platelet count  109/L) and by fluorescence (PLTopt, 228 6 54  109/L).
in EDTA-anticoagulated blood from 21 healthy volunteers When measured after 180 minutes, the impedance
increased within 180 minutes from 228 6 49  109/L (me- platelet count increased slightly to a PLTimp of 198 6
dian, 220) to 240 6 59  109/L (median, 237). 44109/L, whereas the mean platelet counts measured by
In citrate-anticoagulated blood, the initial platelet count the fluorescence optical method remained stable over time
of 197 6 42  109/L (median, 190) was markedly lower within the range of the counts from EDTA-anticoagulated
compared with EDTA-anticoagulated blood. Within the blood drawn in parallel from the same donors and measured
storage time of 180 minutes, the platelet count decreased to using both methods.
189  109/L (median, 190). In contrast to these findings, in citrate-anticoagulated
The platelet count of blood anticoagulated with MgSO4 blood (Figure 2B), platelet counts were lower regardless of
was determined after blood sampling with 201 6 41  109/ the method used. However, more platelets were counted by
L (median, 193). It remained relatively stable over time and the fluorescence optical method compared with the imped-
reached 206 6 42  109/L (median, 194) after 180 minutes. ance measurement (PLTopt, 204 6 48  109/L; PLTimp,
In nine of the volunteer donors, platelet counts gener- 184 6 38  109/L). During storage up to 180 minutes, the
ated by impedance were measured in parallel by the fluores- platelet count showed a tendency to decrease slightly to
cence optical channel of the XE 5000. The results are shown 181 6 39  109/L and to 195 6 44  109/L when measured
in Figure 2 . The time-dependent comparison clearly shows by impedance or fluorescence, respectively.
that both methods measure comparable platelet counts in
EDTA-anticoagulated blood (Figure 2A), although the first
counts were higher when measured by fluorescence
(PLTopt, 228 6 54; PLTimp, 217 6 49  109/L). From Time-Dependent Changes in the MPV
40 minutes onward, both counts were similar, ranging be- The influence of different anticoagulants and time after
tween 226 6 52 and 229 6 53  109/L for impedance and blood sampling is shown in Figure 3 . When the platelet
between 223 6 54 and 229 6 58  109/L for optical count. volume was measured in EDTA-anticoagulated blood, the
In MgSO4-anticoagulated blood, the mean platelet initial MPV from 21 healthy donors was 10.40 6 1.18 fL,
count (Figure 2C) measured by impedance was markedly which increased continuously to 11.12 6 1.32 fL after
lower (PLTimp, 190 6 42 109/L) compared with the values 180 minutes of storage at room temperature.
measured in the EDTA-anticoagulated samples. However, The platelet volume of platelets from MgSO4-
when measured by the fluorescent optical channel, the plate- anticoagulated blood was markedly lower than the MPV
let count (PLTopt, 223 6 53  109/L) was similar to that measured in EDTA blood. Immediately after blood

808 Am J Clin Pathol 2016;145:806-814 © American Society for Clinical Pathology


808 DOI: 10.1093/ajcp/aqw066
AJCP / ORIGINAL ARTICLE

14
A
13
250
Platelet Count (× 109/L)

240
12
230
220

MPV (fL)
11
210
200
190 10
180
170 9
160
0 20 40 60 180 8
Time After Sampling (min)
7

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


<10 20 40 60 180
B
Time After Sampling (min)

250 Figure 3 Mean platelet volume (MPV) measured by the XE


Platelet Count (× 109/L)

240 5000 (impedance method) in samples of 21 volunteers anti-


230
coagulated with EDTA (white), citrate (light gray), and mag-
220
210 nesium sulfate (dark gray).
200
Different Anticoagulants and Their Influence on Platelet
190
180 Distribution Width and Platelet Large Cell Ratio
170 The influence of EDTA, citrate, and MgSO4 on the plate-
160 let distribution width (PDW) is depicted in Figure 4 by box
0 20 40 60 180
Time After Sampling (min) plots. As already shown, platelets swell with time if stored in
EDTA-anticoagulated blood. This can be confirmed by the ob-
C vious increase of the initial PDW (12.44 6 1.98 fL; median,
12.2 fL) to 14.10 6 2.69 fL (median, 13.8 fL) after
250 180 minutes. This is also expressed by the platelet large cell
Platelet Count (× 109/L)

240 ratio (P-LCR), which increased from 28.2% 6 8.8% (median,


230
28.8%) to 34.4% 6 10.0% (median, 33.0%) after 180 minutes.
220
210 Corresponding to the MPV changes, the PDW in mag-
200 nesium-anticoagulated blood is clearly lower than that in
190 EDTA: at the beginning of the measurement series, the
180 PDW was 10.35 6 1.50 fL (median, 10.4 fL), while after
170
160
180 minutes, it was 11.15 6 2.02 fL (median, 11.0 fL).
0 20 40 60 180 The initial PDW in citrate-anticoagulated blood
Time After Sampling (min) (10.87 6 2.06 fL; median, 10.6 fL) and that after 180 minutes
(11.26 6 1.96 fL; median, 11.20 fL) were compatible with
Figure 2 Comparison of the measured platelet count by im-
the respective MPV values measured at the same time.
pedance (black) and optical fluorescence method (gray) in dif-
The initial P-LCR, determined in magnesium-
ferently anticoagulated samples of nine volunteers. The
anticoagulated blood (19.73% 6 6.93%; median, 21.1%), is
whiskers represent the standard error of the mean. A,
moderately different from that of citrate blood (21.79% 6
EDTA. B, Citrate. C, Magnesium sulfate.
8.29%; median, 20.9%). After 180 minutes, the P-LCR of
platelets anticoagulated by magnesium was 22.53% 6
7.83% (median, 22.3%), whereas the P-LCR in citrate blood
sampling, it was 9.31 6 0.97 fL and increased slightly to
was 23.69% 6 7.24% (median, 25.4%).
9.66 6 1.06 fL after 180 minutes.
When measured in citrate-anticoagulated blood, the ini-
tial MPV was 9.54 6 1.12 fL (median, 9.45 fL); within Platelet Counts in Blood Samples Anticoagulated by
60 minutes, the MPV increased to 9.99 6 1.13 fL (median, EDTA, Citrate, and MgSO4
10.0 fL) and then dropped to 9.80 6 0.97 fL, although the Platelets were counted in differently anticoagulated
median remained at 10.0 fL. blood samples from 834 patients using the impedance

© American Society for Clinical Pathology Am J Clin Pathol 2016;145:806-814 809


809 DOI: 10.1093/ajcp/aqw066
Mannuß et al / MAGNESIUM SULFATE AND PLATELET ANALYSIS

20 500

18
400

Platelet Count (× 109/L)


16

300
PDW (fL)

14

12 200

10
100
8

6 0
EDTA Citrate MgSO4

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


<10 min 180 min

Figure 4 Platelet distribution width (PDW) measured by the Figure 5 Comparison of platelet counts in 834 blood sam-
XE 5000 (impedance method) in samples of 21 volunteers ples drawn at the same time but using different anticoagu-
anticoagulated with EDTA (black), citrate (light gray), and lants and measured by the impedance and the optical
magnesium sulfate (dark gray). methods using the XE 5000 as a routine analyzer. The results
of the platelet counts are depicted by box plots as described
under statistical methods (PLTimp, blank; PLTopt, gray).
MgSO4, magnesium sulfate.
(PLTimp) and fluorescence optical (PLTopt) method of the
XE 5000 Figure 5 and Table 1 in parallel. The mean platelet counts measured in citrate and EDTA showed a
platelet count in EDTA-anticoagulated blood was 243.6 slightly worse correlation (r ¼ 0.932; y ¼ 0.83x þ 0.15).
109/L (32-769  109/L) using the impedance method. The
mean platelet count measured by the optical method of the
XE 5000 was 236.5  109/L (39-799  109/L).
The mean PLTimp count for citrate-anticoagulated MPV Measured in Blood Samples Anticoagulated by
blood was 171.2  109/L (24.2-585.2  109/L), although EDTA, Citrate, and MgSO4
the dilution effect had already been corrected by a factor of Platelets anticoagulated with EDTA, citrate, or MgSO4
1.1. The mean value for PLTopt was 195.7  109/L (30.8- differ markedly by volume when measured with the Sysmex
756.8  109/L). XE 5000. The values are depicted as box plots in Figure 6 .
In blood samples anticoagulated with magnesium salt, The respective medians and 5th, 25th, 75th, and 95th per-
the mean PLTimp was 205  109/L (32-642  109/L), and centiles of the platelet counts are summarized in Table 1.
the respective mean value for PLTopt was 226.5  109/L The MPV in EDTA-anticoagulated samples from 834
(34-787  109/L). The differences between PLTimp and patients was 10.4 fL with a minimum of 7.9 fL and a max-
PLTopt were statistically significant independent from the imum of 14.0 fL. In samples from the same patients
anticoagulant used (EDTA, P < .001; citrate, P < .001; anticoagulated with citrate, the mean volume amounted to
MgSO4, P < .001). These findings are depicted as box plots 9.9 fL, ranging from 7.8 to 13.0 fL, and the mean volume in
in Figure 5. magnesium-anticoagulated samples was 9.4 fL with a range
The platelet counts from EDTA-anticoagulated samples from 7.3 to 12.4 fL. These differences in MPV measured in
and those measured from parallel samples anticoagulated by differently anticoagulated whole blood are all highly signifi-
citrate or MgSO4 were correlated by Spearman rank-order cant (P < .001).
correlation. The MPVs measured in MgSO4- and citrate-
Platelet counts measured by impedance in EDTA blood anticoagulated blood were correlated with the respective
(PLTimp) correlated better with platelet counts measured in counts in EDTA as a reference. The correlation coefficient r
MgSO4-anticoagulated blood than platelet counts from cit- for MPVs measured in MgSO4- or EDTA-anticoagulated
rate blood, r ¼ 0.967 (y ¼ 0.80x þ 9.48) and r ¼ 0.834 blood was better than the correlation of the MPVs from
(y ¼ 0.64x þ 15.28), respectively. citrated blood and from EDTA blood (r ¼ 0.902 and
Using the fluorescence optical method, the platelet r ¼ 0.859, respectively). The corresponding regression lines
counts (PLTopt) in MgSO4- and EDTA-anticoagulated blood are defined by the equations y ¼ 0.78x þ 1.24 and
were correlated with r ¼ 0.977 (y ¼ 0.95x þ 2.46), whereas y ¼ 0.78x þ 1.74.

810 Am J Clin Pathol 2016;145:806-814 © American Society for Clinical Pathology


810 DOI: 10.1093/ajcp/aqw066
AJCP / ORIGINAL ARTICLE

Table 1
Summarized Results of Platelet Counts Measured by Impedance (PLTimp) or Fluorescent Optical Technique (PLTopt) and Mean
Platelet Volume (MPV) From Whole-Blood Samples Anticoagulated by EDTA, Citrate, or Magnesium Sulfate (MgSO4) From 834
Individuals
EDTA Citrate MgSO4
PLTimp, PLTopt, PLTimp, PLTopt, PLTimp, PLTopt,
Characteristic  109/L  109/L MPV, fL  109/L  109/L MPV, fL  109/L  109/L MPV, fL
Minimum 32.0 39.0 7.9 24.2 30.8 7.8 32.0 34.0 7.3
5% 139.7 134.7 9.1 86.5 102.3 8.7 115.7 129.3 8.3
25% 193.0 186.0 9.8 132.3 150.7 9.3 164.0 178.0 8.8
Median 229.0 223.5 10.4 165.0 187.0 9.9 194.0 214.0 9.3
75% 284.0 276.0 11.0 201.3 224.4 10.4 238.8 263.0 9.8
95% 380.7 367.4 12.2 267.7 310.2 11.3 319.4 353.1 10.9
Maximum 769.0 799.0 14.0 585.2 756.8 13.0 642.0 787.0 12.4

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


Mean 243.6 236.5 10.4 171.2 195.7 9.9 205.0 226.5 9.4

14 100

80
12
CD62+ Platelets (%)

60
MPV (fL)

10
40

8
20

6 6
EDTA Citrate MgSO4 7-9.9 10-10.9 ≥11
MPV (fL)
Figure 6 Mean platelet volume (MPV) measured by the im-
Figure 7 Expression of platelet activation markers deter-
pedance method (Sysmex XE 5000) in samples of 834 pa-
mined in whole blood from 21 volunteers and anticoagulated
tients anticoagulated with EDTA, citrate, or magnesium
with different anticoagulants (EDTA, triangle; magnesium
sulfate (MgSO4). The averaged results are depicted by box
sulfate, square; citrate, dot). MPV, mean platelet volume.
plots as described under statistical methods.

Discussion
MPV and Spontaneous Platelet Activation Today, routine platelet counts and volume measure-
It has been reported that the MPV correlates with plate- ments are carried out from peripheral venous blood,
let activation. We therefore measured the spontaneous ex- anticoagulated with EDTA or citrate, using an automated
pression of activation markers on platelets anticoagulated hematology analyzer. An important requirement for the cor-
with all three anticoagulants. In EDTA-anticoagulated rect determination of platelet count and volume is the stand-
blood, the platelets express CD62 to a variable degree, ardization of both, the preanalytical and the analytical
whereas in MgSO4- and citrate-anticoagulated blood, only procedure.17 There is a large body of evidence that the
slight activation is measurable. In Figure 7 , the measured blood-collecting conditions, including the choice of anti-
activity is plotted for three different MPV tertiles (7-9.9 fL/ coagulant, are critical for the reliability of platelet count,
10-10.9 fL/11 fL). It is obvious that the spontaneous activ- MPV,18 and platelet activation.19
ity of EDTA-anticoagulated platelets shows a weak but not On the basis of our experiences with MgSO4 in individ-
significant correlation with the MPV, whereas the otherwise uals with anticoagulant-induced in vitro platelet aggrega-
anticoagulated platelets do not. tion,10 so-called pseudothrombocytopenia, we studied the

© American Society for Clinical Pathology Am J Clin Pathol 2016;145:806-814 811


811 DOI: 10.1093/ajcp/aqw066
Mannuß et al / MAGNESIUM SULFATE AND PLATELET ANALYSIS

suitability of a recently introduced blood-sampling device, This might be explained by a loss of platelets when they
prefilled with MgSO4 (ThromboExact-Monovette; Sarstedt) are in contact with citrate, although there were no signs of
for the measurement of platelet count and volume in com- cell destruction. Another explanation might be the fact that
parison to the established anticoagulants EDTA and citrate. the MPV is markedly lower in citrate-anticoagulated blood,
The rationale and aim of the study was to evaluate the pos- so very small platelets may bypass the lower platelet dis-
sible advantages of MgSO4 as an in vitro anticoagulant, es- criminator. However, this does not explain the low platelet
pecially under the aspect of some shortcomings of the counts measured by the fluorescence method.
established anticoagulants. EDTA is an irreversible calcium chelator, whereas cit-
The cell-swelling and shape-changing effect of EDTA rate binds calcium in a pH-dependent manner. The release
within the first hours after venipuncture is known from the of calcium might lead to the formation of platelet aggregates
literature20,21 and represents one of the major problems in during the storage and repeated agitation of citrate-
standardizing platelet volume measurement. anticoagulated blood. Aggregated platelets are then counted

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


Therefore, citrate has already been used as an anticoagu- as single large platelets or as leukocytes, thus explaining the
lant for the measurement of platelet activation, and platelet continuous drop of platelets during our time course experi-
function analysis22,23 was also recommended for a more ments but not the initial low counts.
reliable determination of MPV. On the other hand, whether Another possible explanation for the lower platelet
morphologic differentiation of WBCs is comparable to that counts in citrate-anticoagulated blood was discussed by
of EDTA-anticoagulated blood is controversial.24 Bournazos et al,29 who observed more platelet-leukocyte
The anticoagulant potential of MgSO4 has been known conjugates in citrate than in EDTA blood.
since the beginning of the past century, when Bizzozero Our study clearly shows that in MgSO4-anticoagulated
introduced it for platelet count, as cited by Sahli.7 Much blood, platelet counts were significantly lower if measured
later, magnesium became known as an electrolyte with anti- by impedance and compared with those in EDTA blood.
coagulant properties and was therefore investigated in clin- Using the fluorescent optical method, the platelet count was
ical medicine as an in vivo platelet aggregation inhibitor, largely similar to platelet counts in EDTA blood. This
mainly to prevent cardiovascular complications.25 means that there was obviously no substantial loss of plate-
The antiplatelet activity of magnesium was studied and lets due to magnesium salt anticoagulation11; they seem to
reviewed in more detail by Sheu et al.26 Magnesium inter- be only partially captured by impedance in contrast to the
feres with the binding of fibrinogen to collagen-stimulated fluorescence optical approach.
platelets and obviously has the potential to inhibit intracel- The fact that higher platelet counts are measured with
lular Ca2þ mobilization and thromboxane A2 formation in the XE 5000 fluorescence optical method refutes the argu-
stimulated platelets. Furthermore, it induces the formation ment that platelets are wrecked by the anticoagulant. A par-
of cyclic adenosine monophosphate in platelets and averts allel estimation of platelet counts by microscopy using the
phosphorylation of the protein P47 by inhibiting protein kin- Fonio30 approach led to comparable counts in EDTA and
ase C activity.27 magnesium, thus confirming the fluorescence optical find-
Prior to our study, MgSO4 was used as an in vitro ings (data not shown).
anticoagulant for blood cell count by a group of researchers We suppose that the lower platelet count must be dis-
from Japan,8,9 but it was never introduced into the routine cussed together with the finding that the MPV in citrate-
hematology laboratory. and MgSO4-anticoagulated blood samples was markedly
As shown by others28 and now confirmed by our study, lower than the MPV in EDTA blood. We therefore tend to
the comparability and reproducibility of platelet count de- believe that the impedance platelet count algorithm of the
pend on the anticoagulant and on the method used for plate- XE 5000 excludes platelets with a volume below a distinct
let counting. Platelet counts in EDTA-anticoagulated blood volume threshold.
were largely identical regardless of whether they were The data analysis of platelet counts in a series of 834 in-
counted by impedance or by the fluorescent optical method dividuals clearly confirms that the platelet counts in
of the Sysmex XE 5000. MgSO4- and citrate-anticoagulated samples are significantly
In citrate-anticoagulated blood, platelet counts were lower compared with EDTA anticoagulation.
markedly lower compared with the respective platelet However, the impedance method and the respective
counts in EDTA blood when measured by the impedance software of the analyzer are supposed to underestimate the
method. Platelet counts from the same blood sample meas- real platelet count in MgSO4-anticoagulated samples, pos-
ured in parallel by fluorescence were higher but still below sibly because the analyzers are adjusted for platelet count in
the counts measured in EDTA by both methods. EDTA-anticoagulated blood only.

812 Am J Clin Pathol 2016;145:806-814 © American Society for Clinical Pathology


812 DOI: 10.1093/ajcp/aqw066
AJCP / ORIGINAL ARTICLE

Lower platelet counts in magnesium-anticoagulated There are some indications from other investigators
blood using the routine impedance method underestimate that the MPV is related to the degree of platelet activation.35
the EDTA platelet count by 5% to 10%. At first glance, this This is in accordance with our findings that the spontaneous
might be critical for patients with thrombocytopenia at a expression of CD62p and CD63 in EDTA-anticoagulated
defined critical transfusion threshold of 109/L, for ex- blood tends to be correlated with the MPV. In contrast, cit-
ample.31 This low threshold was set with the intention to re- rate and especially MgSO4 anticoagulation is accompanied
duce the need for and frequency of platelet transfusion only by marginal spontaneous activation; nevertheless, the
because of economic reasons. In such a situation, platelets platelets can still be activated by suitable agonists (data not
from magnesium-anticoagulated blood have to be measured shown).
by the fluorescent optical method. At least, underestimation We think that the reliability of the MPV is such an im-
of platelet counts is not disadvantageous for the individual portant prerequisite for its use as a predictive and prognostic
patient. biomarker that MgSO4 might be advantageous for this issue;

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


Another reason for controversy in platelet analysis is reliable conclusions from disease-related MPV can only be
the absence of standardized MPV measurements. The reli- made if the study protocol strictly standardizes the preana-
ability of this parameter also depends on the kind of anti- lytical and the analytical conditions, including the kind of
coagulant and the time until measurement because of anticoagulation and the method of automated platelet count,
platelet swelling, as confirmed in this study.5 Despite these since there is a marked variability of MPV determination by
reservations, MPV is discussed as a useful prognostic bio- impedance or light scatter analysis.35-37
marker, especially in patients with cardiovascular Our results presented here suggest that MgSO4 could be
diseases.14,32 advantageous for studying MPV as a predictive biomarker,
In most of these studies, EDTA was used as an anti- because the values remain more stable after venipuncture
coagulant, but relatively often the time between blood sam- compared with those measured in the presence of citrate or
pling and MPV measurement was different or not stated. EDTA. Investigators should be aware of the fact that the
This lack of standardization of an important preanalytical lack of standardizations for measurement techniques and
criterion is the most critical weak point of MPV biomarker the use of unsuitable anticoagulants might lead to the misin-
studies. This is justifiably criticiszed by Lancé et al. 33 terpretation of MPV as a predictive or prognostic marker or
One reason to evaluate MgSO4 as an alternative anti- only as a surrogate marker. The use of MgSO4 as an anti-
coagulant to EDTA and citrate was the minimal cell swel- coagulant is also suitable to effectively exclude anticoagu-
ling during sample storage. lant-induced pseudo-thrombocytopenia.
The initial MPV measured by impedance in 834 un-
selected patients was approximately 1 fL lower in MgSO4-
Corresponding author: Peter Schuff-Werner, Rostock University
anticoagulated blood compared with EDTA anticoagulation. Medical Centre Institute of Clinical Chemistry and Laboratory
This finding cannot be explained by time differences be- Medicine, Ernst-Heydemann-Straße, 8 D-18057 Rostock
tween venipuncture and volume analysis, because the sam- Germany; pschuffw@med.uni-rostock.de.
ples were taken and processed at the same time. The mean
MPV of cells anticoagulated with citrate was between the
MPV of EDTA- and MgSO4-anticoagulated samples.
Although the MPV in magnesium-anticoagulated blood in-
References
creased by around 0.35 fL, it is significantly different from
the MPV in EDTA-anticoagulated blood, which increased 1. Banfi G, Salvagno GL, Lippi G. The role of ethylenediamine
tetraacetic acid (EDTA) as in vitro anticoagulant for diag-
by approximately 0.7 fL within 3 hours. nostic purposes. Clin Chem Lab Med. 2007;45:565-576.
Both PDW and P-LCR confirm the MPV changes 2. Slichter SJ. Evidence-based platelet transfusion guidelines:
described for the first 180 minutes of storage at room tem- Haematology/the education program of the American
perature. There is an increase in PDW because cells Society of Hematology. Am Soc Hematol Educ Program.
2007;172-178.
are swelling and the proportion of larger platelets (P-LCR)
3. Ki YJ, Park S, Ha SI, et al. Usefulness of mean platelet vol-
increases when platelets are anticoagulated by EDTA. ume as a biomarker for long-term clinical outcomes after per-
These parameters increase less if citrate or MgSO4 is used cutaneous coronary intervention in Korean cohort: a
as an anticoagulant. Although the real clinical relevance of comparable and additive predictive value to high-sensitivity
these new platelet parameters seems to be restricted to the cardiac troponin T and N-terminal pro-B type natriuretic
peptide. Platelets. 2014;25:427-432.
differentiation of thrombocytopenia,34 it is useful to know
4. Lippi G, Filippozzi L, Salvagno GL, et al. Increased mean
the influence of different anticoagulants on PDW and platelet volume in patients with acute coronary syndromes.
P-LCR. Arch Pathol Lab Med. 2009;133:1441-1443.

© American Society for Clinical Pathology Am J Clin Pathol 2016;145:806-814 813


813 DOI: 10.1093/ajcp/aqw066
Mannuß et al / MAGNESIUM SULFATE AND PLATELET ANALYSIS

5. Lancé MD, van Oerle R, Henskens YMC, et al. Do we need 22. Ahnadi CE, Chapman ES, Lépine M, et al. Assessment of
time adjusted mean platelet volume measurements? Lab platelet activation in several different anticoagulants by the
Hematol. 2010;16:28-31. Advia 120 haematology system, fluorescence flow cytometry,
6. Perrotta G, Roberts L, Glazier J, Schumacher HR. Use of and electron microscopy. Thromb Haemost. 2003;90:940-948.
sodium citrate anticoagulant for routine haematology ana- 23. Heilmann EJ, Kundu SK, Sio R, et al. Brief communication
lysis on the CELL-DYNV R 4000: an opportunity to enhance comparison of four commercial citrate blood collection sys-
efficiency in the clinical laboratory. Lab Hematol. tems for platelet function analysis by the PFA-100TM system.
1998;156-162. Thromb Res. 1997;87:159-164.
7. Sahli H. Lehrbuch der klinischen Untersuchungs-Methoden für 24. Narasimha A, Kumar H, Prasad CS. Anticoagulant induced

Studierende und praktische Arzte mit 291 teilweise farbigen artefacts in peripheral blood smears. Indian J Hematol Blood
Holzschnitten im Texte und 5 lithographierten Tafeln/von H. Transfus. 2008;24:43-48.
Sahli. 4th ed. Wien, Leipzig: Deutike; 1905. 25. Shechter M, Merz CN, Paul-Labrador M, et al. Oral magne-
8. Nakamoto K, Sugibayashi S, Takahashi A, et al. Platelet sium supplementation inhibits platelet-dependent throm-
count in EDTA-dependent pseudothrombocytopenia—ap- bosis in patients with coronary artery disease. Am J Cardiol.
plication of MgSO4 as an anticoagulant. Rinsho Byori. 1999;84:152-156.

Downloaded from https://academic.oup.com/ajcp/article/145/6/806/2836709 by guest on 23 March 2024


1986;34:167-173. 26. Sheu J, Hsiao G, Shen M, et al. Mechanisms involved in the
9. Kondo H, Kobayashi E, Itani T, et al. Haematology tests of antiplatelet activity of magnesium in human platelets. Br J
blood anti-coagulated with magnesium sulphate. Southeast Haematol. 2002;119:1033-1041.
Asian J Trop Med Public Health. 2002;33:6-9. 27. Hwang DL, Yen CF, Nadler JL. Effect of extracellular magne-
10. Schuff-Werner P, Steiner M, Fenger S, et al. Effective esti- sium on platelet activation and intracellular calcium mobil-
mation of correct platelet counts in pseudothrombocytopenia isation. Am J Hypertens. 1992;5:700-706.
using an alternative anticoagulant based on magnesium salt. 28. Buttarello M. Quality specification in haematology: the
Br J Haematol. 2013;162:684-692. automated blood cell count. Clin Chim Acta. 2004;346:45-54.
11. François D, Masure A, Atallah N, et al. Underestimation of 29. Bournazos S, Rennie J, Hart SP, et al. Choice of anticoagu-
platelet count on magnesium salt-anti-coagulated samples. lant critically affects measurement of circulating platelet-
Clin Chem Lab Med. 2014;52:e95-e97. leukocyte complexes. Arterioscler Thromb Vasc Biol.
12. Balta S, Demirkol S, Celik T, et al. Mean platelet volume as 2008;28:E2-E3.
a surrogate marker of long-term mortality in patients €
30. Fonio A. Uber ein neues Verfahren der
undergoing percutaneous coronary intervention. Am J Blutpl€attchenz€ahlung. Deutsch Zeitschr Chirurg.
Cardiol. 2013;112:142. 1912;117:176-194.
13. Shah B, Oberweis B, Tummala L, et al. Mean platelet vol- 31. Segal HC, Briggs C, Kunka S, et al. Accuracy of platelet
ume and long-term mortality in patients undergoing percu- counting haematology analysers in severe thrombocytopenia
taneous coronary intervention. Am J Cardiol. and potential impact on platelet transfusion. Br J Haematol.
2013;111:185-189. 2005;128:520-525.
14. Chu SG, Becker RC, Berger PB, et al. Mean platelet volume 32. Bath P, Butterworth RJ. Platelet size measurement, physi-
as a predictor of cardiovascular risk: a systematic review and ology and vascular disease. Blood Coagul Fibrinolysis.
meta-analysis. J Thromb Haemost. 2010;8:148-156. 1996;7:157-161.
15. Briggs C, Harrison P, Machin SJ. Continuing developments 33. Lancé MD, Sloep M, Henskens Yvonne MC, et al. Mean
with the automated platelet count. Int J Lab Hematol. platelet volume as a diagnostic marker for cardiovascular dis-
2007;29:77-91. ease: drawbacks of pre-analytical conditions and measuring
16. Liu Y. Box plots: use and interpretation. Transfusion. techniques. Clin Appl ThrombHaemost. 2012;18:561-568.
2008;48:2279-2280. 34. Kaito K, Otsubo H, Usui N, et al. Platelet size deviation
17. Diaz-Ricart M, Brunso L, Pino M, et al. Preanalytical width, platelet large cell ratio, and mean platelet volume
treatment of EDTA–anti-coagulated blood to ensure stabil- have sufficient sensitivity and specificity in the diagnosis of
ization of the mean platelet volume and component meas- immune thrombocytopenia. Br J Haematol.
ured with the ADVIA counters. Thromb Res. 2005;128:698-702.
2010;126:e30-e35. 35. Shah B, Valdes V, Nardi MA, et al. Mean platelet volume re-
18. Thompson CB, Diaz DD, Quinn PG, et al. The role of anti- producibility and association with platelet activity and anti-
coagulation in the measurement of platelet volumes. Am J platelet therapy. Platelets. 2014;25:188-192.
Clin Pathol. 1983;80:327-332. 36. Lippi G, Pavesi F, Pipitone S. Evaluation of mean platelet
19. Mody M, Lazarus AH, Semple JW, et al. Pre-analytical volume with four haematological analysers: harmonisation is
requirements for flow cytometric evaluation of platelet acti- still an unresolved issue. Blood Coagul Fibrinolysis.
vation: choice of anticoagulant. Transfus Med. 2015;26:235-237.
1999;9:147-154. 37. Macey M, Azam U, McCarthy D, et al. Evaluation of the
20. Vogelaar SA, Posthuma D, Boomsma D, et al. Blood sam- anticoagulants EDTA and citrate, theophylline, adenosine,
ple stability at room temperature for counting red and and dipyridamole (CTAD) for assessing platelet activation
white blood cells and platelets. Vascul Pharmacol. on the ADVIA 120 haematology system. Clin Chem.
2002;39:123-125. 2002;48:891-899.
21. White JG. Effects of ethylenediamine tetraacetic acid
(EDTA) on platelet structure. Scand J Haematol.
1968;5:241-254.

814 Am J Clin Pathol 2016;145:806-814 © American Society for Clinical Pathology


814 DOI: 10.1093/ajcp/aqw066

You might also like