Professional Documents
Culture Documents
The accurate term for the affiliated thermo analytical method is then Differen-
tial Thermal Analysis (DTA). For a short description of the principle of DTA
measuring systems in comparison with DSC, see Appendix 1.
1 Introduction 3
Owing to the new materials and sensors used and the application of modern
mechanical manufacturing (micro and nano technologies), advanced data pro-
cessing by modern electronic systems, all calorimeters available today - includ-
ing the so-called "classic" calorimeters - are instruments which allow precise
measurements to be carried out with high sensitivity and repeatability (see Sect.
7.2), their operation being relatively simple or perhaps even automated. Caloric
methods are used in many fields for quality assurance purposes (cf. series of
standards ISO 9000). The standards to be applied in the field of quality assur-
ance (e.g., the standard ISO/IEC 17025, 1999: General requirements for the
competence of testing and calibration laboratories) demand a large variety of
measures to ensure reliable measurement results, for example, the application of
well-proven methods for the preparation of samples, reliable measurement
methods, calibration methods and calibration materials traceable to standards
(see also Chapter 4).
In the following, reference is made to some of the fields in which calorimeters
are widely used today (see Hemminger, Hohne, 1984; Wunderlich, 1990; Oscar-
son, Izatt, 1992). A suggestion how to classify calorimeters is given in Appendix 2.
- With bomb calorimeters, combustion heats (gross calorific values) are mea-
sured on a large scale in industry (costs of fossil fuels) (Sunner, Mansson,
1979).
- Gas calorimeters (flow calorimeters) are used for the continuous or discon-
tinuous measurement of the calorific value of fuel gases (e.g., natural gas),
both in frontier-crossing commercial transactions (between supplier and
buyer) and for the calculation of the costs for the individual consumer (Hyde,
Jones, 1960; Hemminger, 1988; Sarge, 1997; Ulbig, Hoburg, 2002; Jaeschke,
2002; Dale et al., 2002; Alexandrov, 2002).
- Drop calorimeters (usually self-made) allow mean heat capacities or enthalpy
differences to be quickly measured. Drop sample temperatures of up to
2000°C are realized (Chekhovskoi, 1984).
- Different types of isoperibol mixing calorimeters are used to investigate reac-
tions between two fluids or between a fluid and a solid (reaction heats, heats
of solution, heats of mixing, adsorption heats) (see, e.g., Parrish, 1986).
- Reaction calorimeters and safety calorimeters allow model tests of procedures
applied in industrial chemistry to be carried out. These instruments provide
valuable support in development and optimization tasks, as all test parame-
ters (temperature, time, addition of substances etc.) are completely docu-
mented and can be varied automatically (Landau, 1996).
Questions of process technology which are also important for the dimen-
sioning of production facilities, may be well in the fore here (Regenass,
1985). Reaction heat released is connected with the degree of conversion
and exploitation, for example, in biotechnology, where calorimetric studies
are carried out on industrial production plants (large-scale calorimetry,
mega-calorimetry, see, e.g., von Stockar, Marison, 1991 in Lamprecht et aI.,
1991).
4 1 Introduction
- There are not yet any practicable and experimentally tested recommenda-
tions for temperature and heat calibration and for measurement procedures
which are internationally accepted (see Chapter 4; and some examples in
Della Gatta et al., 2000).
- There is no international agreement on a single set of substances for the tem-
perature and heat calibration of DSC which have been measured with suffi-
cient accuracy, including a metrologically sound traceability to national/in-
ternational standards; instead there is a confusing variety of "certified refer-
ence materials" whose characteristic data are in part contradictory. Sub-
stances which are certified on a metrologically basis (i. e., traceable to the SI
units) are offered by the German Metrology Institute (PTB) (cf. Sect. 4.6).
Due to the rapidly increasing use of DSCs in various fields of application, some
negligence has gained ground - favored by the ease of operation and evaluation
- which would be inconceivable in "classic" calorimetry. Remedial measures
should be taken; for example, the national societies of the International Con-
federation for Thermal Analysis and Calorimetry (ICTAC) should offer special
training courses, and precise specifications for instruments and programs
should be drawn up.
The following can frequently be observed:
- A realistic estimate of the uncertainty of measurement is rarely made (cf.
Sect. 7.3). The calibration capability (cf. Chapter 4), different influencing
quantities (cf. Sect. 5.2) and known theoretical considerations (cf. Chapter 3)
should be taken into account. In many cases, the repeatability of a DSC is, for
example, simply, but wrongly, indicated as accuracy of the measured data (cf.
Sect. 7.3).
- Interpretation of the DSC measurement results is often insufficient or faulty,
when
- uncertainties of measurement are not taken into consideration;
- systematic error sources are disregarded;
- the measured curve is not "desmeared" (cf. Sect. 5.4);
- the laws of thermodynamics, kinetics, are not taken into account;
- uncritical confidence is placed in the evaluation programs provided by the
manufacturer;
- the results are not confirmed by other measuring methods.
A great number of calorimetric methods has not been mentioned here as, in
comparison with DSC, they are used only in specific fields. To name a few ex-
amples:
- low-temperature calorimetry for measuring heat capacities (see, e. g., Gmelin,
1987),
- more recent techniques of cp measurement (see, e.g., Lakshmikumar, Gopal,
1981; Maglic et aI., 1984),
- measurement of the energy of particle radiation (see, e. g., Domen, 1987),
- measurements on biological systems (see, e. g., Lamprecht et al., 1991; Wadso,
1993 and Kemp, 1993; Kemp et aI., 1998; Hansen, 2000),
1 Introduction 7
Summary