The calibration is then finished, and the correct weight value is now displayed in SIWATOOL.
Read back DR3
During the calibration, the SIWAREX internally changes its calibration digits and the
calibration weight.
This means that obsolete parameter values are now present in SIWATOOL.
These are displayed in red, e.g.:
The DR3 must be read back in order to again achieve calibration digits and the calibration
weight in the SIWATOOL which are consistent with the SIWAREX:
SIWAREX WP521/WP522
62 Manual, 04/2016, A5E36695225A