You are on page 1of 45

{Instal ationInstructs}{Production}{Carestream Health}{Restricted} Publication No.

II5258-1
21JUL09
Supersedes II5258-1
01JUN09
Restricted

Installation Instructions
for the
KODAK DIRECTVIEW Classic/Elite CR Systems
Service Code: 5258, 5259
Important
• Qualified service personnel must install this equipment.
• See the Service Portal for possible updates before using this document.
• When doing the procedures in this document, you must use safe work practices and wear
the correct personal protective equipment (for example, safety eyewear) according to your
company’s standard operating procedures.

H219_0001GC

© CARESTREAM HEALTH, INC.


INSTALLATION INSTRUCTIONS
21JUL09
II5258-1 PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the subject
Page matter gained by Carestream Health, Inc. prior to publication.
2 of 45
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes no
warranty, express or implied, with respect to this information. Carestream Health shall not be liable
for any loss or damage, including consequential or special damages, resulting from any use of this
information, even if loss or damage is caused by Carestream Health’s negligence or other fault.

This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description Page

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Packing List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Necessary Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Preparing for the Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installing the Wall Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Installing the Floor Stand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Unpacking the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installing the Equipment on a Wall Stand . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Installing the Equipment on a Floor Stand . . . . . . . . . . . . . . . . . . . . . . . . . 32
Installing the Equipment on Customer-Supplied Equipment . . . . . . . . . . 36
Connecting the Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Energizing the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Doing the Configuration for the Local User Interface (LUI). . . . . . . . . . . . 40
Doing the Configuration for the Software . . . . . . . . . . . . . . . . . . . . . . . . . . 44
INSTALLATION INSTRUCTIONS Safety
21JUL09
II5258-1
Page
Section 1: Safety
3 of 45
This information defines the safety and information icons used in this publication.

Radiation
This icon is used for conditions that could cause personal injury from radiation.

Laser Warning
This icon is used for conditions when a laser beam could cause injury to a person.

Caution
This icon is used for conditions that could cause injury to a person, or damage to the
equipment or software data.

ESD
Possible damage from electrostatic discharge.

This icon is used for conditions that could cause damage to the equipment.
Electrostatic discharge (ESD) is a primary source of:
• equipment failure
• equipment repairs
A person cannot detect an electrical charge of less than 3,500 V, but 30 V can cause
damage to components in the equipment.
INSTALLATION INSTRUCTIONS Safety
21JUL09
II5258-1 Preventive Measures
Page
4 of 45 • Check for an ESD warning label before doing any procedure with ESD-sensitive
components. All sensitive components have graphic labels that frequently include
instructions. Use all label instructions.
• Wear a grounding strap when you touch ESD-sensitive components. Check that the CLIP
remains fastened to a ground that has a clean surface with no paint.
• Repair components in an ESD-protection area or use a portable grounding mat.
• When moving ESD-sensitive components from area to area, insert and transport the
components in the special material made for the transport of these components.

Important
This icon is used for important information.

Note
This icon is used for additional information.
INSTALLATION INSTRUCTIONS Packing List
21JUL09
II5258-1
Page
Section 2: Packing List
5 of 45

Part No. Description Quantity


--- KODAK DIRECTVIEW Classic/Elite CR System 1
7E9471 ERASE LAMPS 10
5F0625 NETWORK CABLES 2
--- Documentation kit 1
6F1753 Shipping document - Japan 1
379212 WIRE TIES 4
--- DISKETTE - user preferences 1
--- DISKETTE - DICOM print 1
INSTALLATION INSTRUCTIONS Necessary Materials
21JUL09
II5258-1
Page
Section 3: Necessary Materials
6 of 45
Tools
Part No. Description
6H6825 Site Information Sheet for the Software Version 5.x for the KODAK
DIRECTVIEW CR 825/850/950/975 Systems and the KODAK DIRECTVIEW
Classic/Elite CR Systems
--- Standard FE tool kit
--- Modification Instructions and Software Installation Instructions for Software
Version 5.x for the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the
KODAK DIRECTVIEW CLASSIC ELITE CR Systems
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
Section 4: Installation
7 of 45
Preparing for the Installation
1 Are you installing the Classic/Elite CR System on:
Wall Stand Floor Stand Customer-Supplied Equipment
Continue with Installing Advance to Installing the a. Prepare a stable surface for
the Wall Stand. Floor Stand. the equipment.
b. Advance to Unpacking the
Equipment.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Installing the Wall Stand
Page
8 of 45 1 Unpack and check the components of the WALL STAND.

Important
The SUPPORT RAIL must be installed:
2 SUPPORT BRACKETS
• by the customer
• before you start to assemble the
SUPPORT RAIL
remaining components of the WALL
STAND
• to match local codes
2 SET • to support a total weight of 45.5 kg
SCREWS (100.0 lbs)

2 Loosen the 2 SET SCREWS on each of


the 2 SUPPORT BRACKETS.
3 Slide the 2 SUPPORT BRACKETS into
the top of the SUPPORT RAIL.
4 Place the 2 SUPPORT BRACKETS in
the approximate position, and tighten
the 2 SET SCREWS for each.

H219_3035GCA
H219_3035GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 5 Rotate to install the COMPUTER
Page
9 of 45 COMPUTER BRACKET BRACKET at the top of the SUPPORT
RAIL.
6 Tighten the 2 SCREWS.
7 Install:
• FRONT COMPUTER BRACKET
COMPUTER
BRACKET
• 2 THUMB SCREWS - leave loose
2 SCREWS

FRONT
COMPUTER
BRACKET
SUPPORT
RAIL

2 THUMB
H219_3036GCA SCREWS
H219_3036GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 8 Install:
Page
10 of 45 • MONITOR ARM
• WASHER
• SCREW
MONITOR ARM • KEYBOARD ARM
• WASHER
• SCREW
9 Are you installing a KEYBOARD TRAY?
Yes No
Continue with Advance to
WASHER
Step 10. Step 11.
SCREW

KEYBOARD ARM
WASHER
H219_3037GCA
H219_3037GC SCREW
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 10 Install:
Page KEYBOARD
11 of 45 TRAY • KEYBOARD TRAY
• 4 SCREWS
• BAR CODE READER HOLDER
• 4 SCREWS
• BOOT
BAR CODE
READER
• SCREW
HOLDER

4 SCREWS

H219_3038ACA
SCREW H219_3038AC

11 Install directly to the KEYBOARD ARM:


BAR CODE • BAR CODE READER HOLDER
READER
KEYBOARD
HOLDER
ARM • SCREW
12 Advance to Unpacking the Equipment.

SCREW

H219_3039ACA
H219_3039AC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Installing the Floor Stand
Page
12 of 45 1 Unpack:
KEYBOARD SHELF • FLOOR STAND
• KEYBOARD SHELF - optional
• SCAN GUN BOOT - optional
2 Check the packing lists.

SCAN GUN BOOT

FLOOR STAND

H219_3017GCA
H219_3017GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
13 of 45

STAND

8 SCREWS

8 TABS

H219_3019HCA BASE
H219_3019HC

3 Place the BASE on the floor.


4 Install the 8 TABS of the STAND into the BASE.
5 Carefully place the assembly on its side and install the 8 SCREWS.
6 Place the assembly on its BASE.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 7 Install to the BASE:
Page
14 of 45 • LOCATOR
BASE
• 2 SCREWS - leave loose

2 SCREWS
LOCATOR H219_3020ACA
H219_3020GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page MONITOR MONITOR
15 of 45
SHELF SHELF

KEYBOARD
SHELF SCAN GUN
BOOT

4 NUTS SCREW
SCAN GUN BOOT
3 NUTS

FLOOR STAND
SCREW

2 BIN LINERS

H219_3021HCA
H219_3021HC

Important
• Some sites may not install:
– KEYBOARD SHELF
– SCAN GUN BOOT
• When you are not installing the optional KEYBOARD SHELF, the SCAN GUN BOOT can
be installed to the MONITOR SHELF, as shown in the above graphic.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 8 Install:
Page
16 of 45 • MONITOR SHELF
• 3 NUTS
• KEYBOARD SHELF - optional
• 4 NUTS
• SCAN GUN BOOT - optional
• SCREW
• 2 BIN LINERS
9 Place the FLOOR STAND in the correct position at the site.
10 Level the FLOOR STAND.
11 Move the LOCATOR fully down to the floor.
12 Tighten the 2 SCREWS.
13 Continue with Unpacking the Equipment.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Unpacking the Equipment
Page
17 of 45 1 Pull and lift to remove the FRONT
COVER.

FRONT COVER

H219_4033ACA
H219_4033AC

2 SCREWS CONNECTOR
2 Loosen the 2 SCREWS to allow access
COVER to the 2 CONNECTORS.
3 Rotate the CONNECTOR COVER.
4 Disconnect the 2 CONNECTORS.

H219_3001ACA
2 CONNECTORS H219_3001AC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 5 Remove:
Page
4 SCREWS
18 of 45 TOP COVER • 4 SCREWS
• TOP COVER

H219_5032GCA
H219_5032GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 6 Remove:
Page LEFT COVER
19 of 45 • 4 SCREWS
4 SCREWS
• LEFT COVER
• 4 SCREWS
• RIGHT COVER

4 SCREWS
H219_5031GCA
H219_5031GC RIGHT COVER
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page 5 SCREWS 9 SCREWS
20 of 45

BACK EMI
PANEL

LEFT EMI PANEL

H219_3061HCA
H219_3061HC

7 Remove:
• 5 SCREWS
• LEFT EMI PANEL
• 9 SCREWS
• BACK EMI PANEL - rotate and lift
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 8 Check the TILT WATCH on the back of
Page
21 of 45 the frame.
9 Is the TILT WATCH red?
Yes No
Stop the installation Continue with
procedure and call the Step 10.
Technical Support
Center
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
22 of 45

2 SCREWS

2 SCREWS 2 SCREWS
REAR SHIPPING REAR SHIPPING
BRACKET BRACKET

FRONT
FRONT SHIPPING
SHIPPING
H219_3062HCA 4 SCREWS BRACKET
H219_3062HC BRACKET 2 SCREWS

10 For the FRONT SHIPPING BRACKET:


a. Remove the 4 SCREWS.
b. Slide the FRONT SHIPPING BRACKET toward the back of the system.
c. Install the 2 SCREWS.
11 For the REAR SHIPPING BRACKET:
a. Loosen the 2 SCREWS on the side.
b. Remove the 2 SCREWS in the back.
c. Slide the SHIPPING BRACKET toward the front of the system.
d. Tighten the 2 SCREWS on the side.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page remove 2 SCREWS
23 of 45 loosen 2 SCREWS
SHIPPING
BRACKET POWER
SUPPLY

tighten
2 SCREWS

SLOW SCAN
TOWER

H219_3071BCA
H219_3071BC SHIPPING BRACKET
12 Remove and discard the 2 SCREWS from the SLOW SCAN TOWER portion of the
SHIPPING BRACKET.
13 Loosen the 2 SCREWS on the POWER SUPPLY portion of the SHIPPING BRACKET.
14 Pull the SHIPPING BRACKET away from the SLOW SCAN TOWER and towards the
outside of the system.
15 Tighten the 2 SCREWS on the POWER SUPPLY portion of the SHIPPING BRACKET.

Note
You must keep the 2 SCREWS in the POWER SUPPLY portion of the SHIPPING BRACKET,
or a light leak could occur. The PLATE SUPPORT ASSEMLY could hit the SHIPPING
BRACKET if you position the SHIPPING BRACKET incorrectly.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
24 of 45
SWITCH
CB1
4 VOLTAGE
JUMPERS

WIRE TIE

JUMPER

H219_3025HCA VOLTAGE CONNECTOR


H219_3025HC

Important
You must determine the correct JUMPER to connect to the VOLTAGE CONNECTOR. The
correct JUMPER might not be the same JUMPER removed in the graphic.

16 Locate the 4 VOLTAGE JUMPERS.


17 Determine the correct JUMPER for the voltage at the site.
18 Cut the WIRE TIE and remove the JUMPER.
19 Connect the correct JUMPER to the VOLTAGE CONNECTOR.
20 Check that SWITCH CB1 is set to the right.
21 If necessary, move SWITCH CB1 to the right.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 22 Loosen the 2 SCREWS.
Page
25 of 45 23 Slide the BRACKET to the left.
24 Tighten the 2 SCREWS.

Important
Do not remove the ERASE LAMP
ASSEMBLY.
2 SCREWS

25 Pull to fully extend the ERASE LAMP


ASSEMBLY.

BRACKET
H219_3063GCA
ERASE LAMP AY H219_3063GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 26 Remove and discard the REMOVE AT
Page
26 of 45 INSTALL TAB.

A E
T M
IN O
R
S VE
TA
LL

REMOVE AT
H219_3064GCA INSTALL TAB
H219_3064GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 27 Install:
Page
27 of 45 • BACK EMI PANEL
• 9 SCREWS
• LEFT EMI PANEL
• 5 SCREWS
• LEFT COVER
• 4 SCREWS
• RIGHT COVER
• 4 SCREWS
28 Install:
• TOP COVER
• 4 SCREWS
29 Connect the 2 CONNECTORS.
30 Rotate the CONNECTOR COVER.
31 Tighten the 2 SCREWS.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 32 Install the 10 ERASE LAMPS into the
Page ERASE LAMP AY
28 of 45 ERASE LAMP ASSEMBLY.
33 Push the ERASE LAMP ASSEMBLY
into the system.
34 Install the FRONT COVER.

10 ERASE LAMPS
H219_3065GCA
H219_3065GC

35 Unpack any other accessories for the Classic/Elite CR System.


36 Are you installing the Classic/Elite CR System on:
Wall Stand Floor Stand Customer-Supplied Equipment
Continue with Installing Advance to Installing the Advance to Installing the
the Equipment on a Wall Equipment on a Floor Equipment on Customer-
Stand. Stand. Supplied Equipment.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Installing the Equipment on a Wall Stand
Page
29 of 45 1 Install:
4 SCREWS
• MONITOR
• 4 SCREWS

MONITOR

H219_3040ACA
H219_3040AC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
30 of 45
FRONT COMPUTER
BRACKET

2 THUMB SCREWS
COMPUTER

MONITOR ARM

MONITOR

SUPPORT RAIL

KEYBOARD
2 SUPPORT BRACKETS

KEYBOARD ARM
KEYBOARD TRAY

BAR CODE READER MOUSE


H219_3023HCA
H219_3023HC
BAR CODE READER HOLDER
2 Install the remaining equipment on the WALL STAND:
• COMPUTER
• KEYBOARD - optional
• MOUSE - optional
• BAR CODE READER - optional
3 Tighten the 2 THUMB SCREWS for the FRONT COMPUTER BRACKET.
4 Place the Classic/Elite CR System adjacent to the WALL STAND.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Figure 1 Model 8810 Computer Figure 2 Model HP5700 Computer
Page
31 of 45

VOLTAGE
unused
SWITCH POWER
POWER PS/2 PORTS

KEYBOARD COM3/unused
MOUSE (ignore the
COM1/unused
painted PC label
that says COM2)
MONITOR
TOUCHSCREEN
BAR CODE
READER USB
KEYBOARD
USB MOUSE
MONITOR COM 4

unused
COM 3 BAR CODE
extra NIC
READER
COM 1

TOUCHSCREEN CUSTOMER
CUSTOMER NETWORK
H219_0011GCA NETWORK H226_1027GCA
H219_0011GC H226_1027GC

Important
• You will first connect the COMPUTER to the Classic/Elite CR System.
• After you do the software installation procedures, you will then connect both to the
hospital network at the site:
– COMPUTER
– Classic/Elite CR System

5 See the graphic above to connect the equipment.


INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 6 Route the wires.
Page
32 of 45 7 On the COMPUTER, set the VOLTAGE SWITCH:
Setting Site Voltage
115 100 - 120 Vac
230 200 - 240 Vac
8 Advance to Connecting the Equipment.

Installing the Equipment on a Floor Stand


FLOOR STAND
1 Place the FLOOR STAND in the correct
position at the site.
LOCATOR 2 Move the Classic/Elite CR System onto
the FLOOR STAND.
3 Align the 2 WHEELS into the 2
NOTCHES in the LOCATOR.

2 NOTCHES

H219_3022ACA
H219_3022AC 2 WHEELS
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
MONITOR
4 Place the COMPUTER in front of the
Page
33 of 45 FLOOR STAND.
KEYBOARD 5 Install the equipment on the FLOOR
STAND:
• MONITOR
• KEYBOARD - optional
• MOUSE - optional
MOUSE
• BAR CODE READER - optional

BAR CODE
READER

FLOOR STAND

H219_3016GCA COMPUTER
H219_3016GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 6 See the graphic to route the wires.
Page
34 of 45

H219_3024GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Figure 3 Model 8810 Computer Figure 4 Model HP5700 Computer
Page
35 of 45

VOLTAGE
unused
SWITCH POWER
POWER PS/2 PORTS

KEYBOARD COM3/unused
MOUSE (ignore the
COM1/unused
painted PC label
that says COM2)
MONITOR
TOUCHSCREEN
BAR CODE
READER USB
KEYBOARD
USB MOUSE
MONITOR COM 4

unused
COM 3 BAR CODE
extra NIC
READER
COM 1

TOUCHSCREEN CUSTOMER
CUSTOMER NETWORK
H219_0011GCA NETWORK H226_1027GCA
H219_0011GC H226_1027GC

Important
• You will first connect the COMPUTER to the Classic/Elite CR System.
• After you do the software installation procedures, you will then connect both to the
hospital network at the site:
– COMPUTER
– Classic/Elite CR System

7 See the graphic above to connect the equipment.


INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 8 On the COMPUTER, set the VOLTAGE SWITCH:
Page
36 of 45 Setting Site Voltage
115 100 - 120 Vac
230 200 - 240 Vac
9 Place the COMPUTER inside the FLOOR STAND.

Installing the Equipment on Customer-Supplied Equipment


1 Place the equipment on a stable surface:
• KEYBOARD
• MOUSE
• COMPUTER
• MONITOR
• BAR CODE READER
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Figure 5 Model 8810 Computer Figure 6 Model HP5700 Computer
Page
37 of 45

VOLTAGE
unused
SWITCH POWER
POWER PS/2 PORTS

KEYBOARD COM3/unused
MOUSE (ignore the
COM1/unused
painted PC label
that says COM2)
MONITOR
TOUCHSCREEN
BAR CODE
READER USB
KEYBOARD
USB MOUSE
MONITOR COM 4

unused
COM 3 BAR CODE
extra NIC
READER
COM 1

TOUCHSCREEN CUSTOMER
CUSTOMER NETWORK
H219_0011GCA NETWORK H226_1027GCA
H219_0011GC H226_1027GC

Important
• You will first connect the COMPUTER to the Classic/Elite CR System.
• After you do the software installation procedures, you will then connect both to the
hospital network at the site:
– COMPUTER
– Classic/Elite CR System

2 See the graphic above to connect the equipment.


INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 3 On the COMPUTER, set the VOLTAGE SWITCH:
Page
38 of 45 Setting Site Voltage
115 100 - 120 Vac
230 200 - 240 Vac
4 Continue with Connecting the Equipment.

Connecting the Equipment


1 Check the voltage at the site.
2 Select the correct POWER CORD for
the site.

Caution
Dangerous Voltage

3 Connect the new POWER CORD to:


NETWORK • POWER CONNECTION
CONNECTION
• main power at the site
4 Connect the new NETWORK CABLE to:
POWER
CONNECTION • NETWORK CONNECTION
H219_3010ACA
H219_3010AC
• network at the site
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 5 Place the system in the correct position
Page
39 of 45 at the site.
6 Remove the FRONT PANEL.
7 Loosen the 2 LOCKING NUTS.
8 Rotate the 2 HOLD DOWN FEET until
the HOLD DOWN FEET make contact
with the floor.
9 Tighten the 2 LOCKING NUTS.

2 LOCKING
NUTS
FRONT
PANEL

2 HOLD
DOWN FEET
H219_3066GCA floor
H219_3066GC
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Energizing the System
Page
40 of 45
Laser Warning
• This equipment uses a visible red LASER. Prevent direct exposure to the beam from the
LASER when the system is energized.
• Dangerous Voltage

1 Energize:
• COMPUTER
• system
• MONITOR

Doing the Configuration for the Local User Interface (LUI)

Service Menu BUTTON

1 At the LUI, touch the Service Menu BUTTON.


INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
41 of 45

green CHECK MARK

2 Type the service password: 4213


3 Touch the green CHECK MARK.

4 Touch:
• [Network Configuration].
• [Ethernet Settings]
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
42 of 45

green ARROW

5 Select the correct “Link Speed and Duplex Value” for the site.
6 Touch:
• green arrow
• [TCP/IP Settings]
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1
Page
43 of 45

VALUE BOX

7 Do Step 8 - Step 10 to enter each of the values:


• “IP Address”
• “Gateway”
• “Netmask”
8 Touch the VALUE BOX.

9 Type the correct value.


10 Touch the green CHECK MARK.
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 11 Touch:
Page
44 of 45 • green arrow
• green arrow
12 Press the POWER BUTTON to de-energize the Classic/Elite CR System.
13 Energize the Classic/Elite CR System.

Doing the Configuration for the Software

Important
• You must do the necessary procedures for a new system.
• You must see the Service Knowledge Portal for the latest software instructions.

1 Do the modification instructions and software installation instructions for Software Version
5.x for the KODAK DIRECTVIEW CR 825/850/950/975 Systems and the KODAK
DIRECTVIEW Classic/Elite CR Systems.
2 Disconnect the COMPUTER from the Classic/Elite CR System.
3 Connect to the network at the site:
• COMPUTER
• Classic/Elite CR System
INSTALLATION INSTRUCTIONS Installation
21JUL09
II5258-1 Publication History
Page
45 of 45
Publication Publication Changed
ECO No. File Name Notes
Date No. Pages
13NOV07 II5258-1 - - - --- ii5258_1.fm New Publication
28NOV07 II5258-1 - - - 1 ii5258_1.fm Revision
07APR08 II5258-1 - - - All ii5258_1.fm Revision
01JUN09 II5258-1 - - - All ii5258_1.fm Revision
21JUL09 II5258-1 - - - 1 ii5258_1.fm Revision

DIRECTVIEW is a trademark of Carestream Health, Inc.


The Kodak trademark and trade dress are used under license from Kodak.

Printed in U.S.A. • ii5258_1_21Jul09.fm

Carestream Health, Inc.


150 Verona Street
Rochester, NY 14608

You might also like