{PreventiveMaint}{Production}{Carestream Health Inc.

}{Confidential}

Publication No. PM4825-1, Rev. B


2018-10-10
Supersedes PM4825-1
2015-11-17

Preventive Maintenance
for the
KODAK DIRECTVIEW CR 825/850 Systems
Service Codes: 4825, 5634
Important
• Qualified service personnel must do these procedures.
• See the Carestream Health Service Portal for possible updates before doing these procedures.
• When doing procedures associated with this equipment, you must use safe work practices and wear the correct
personal protective equipment (for example, safety eyewear) according to your company’s standard operating
procedures.

H177_0500AC

Confidential

© Carestream Health, Inc. 2018


PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes
no warranty, express or implied, with respect to this information. Carestream Health shall not be
liable for any loss or damage, including consequential or special damages, resulting from any
use of this information, even if loss or damage is caused by Carestream Health’s negligence or
other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description Page

Necessary Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Maintenance Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Checking the Error Logs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Removing Images from the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Checking the Scan Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cleaning the Computer Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Checking the CMOS Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Cleaning the PMT/DAS Board, Parts and Collector Assembly . . . . . . . . . . . . . 8
Cleaning the PMTs and Collector Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cleaning and Applying Lubricant to the Cassette Handling Assembly. . . . . . . . 10
Cleaning the Bar Code Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Cleaning the Extraction Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Cleaning and Applying Lubricant to the Slow Scan Slides. . . . . . . . . . . . . . . . . 12
Cleaning the Erase Lamp Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Cleaning the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installing the Light Gaskets on the Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installing the Clamp Bar Plush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installing the Right and Left Light Baffels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installing the Rear Clip Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installing the New Light Seal Plush Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Installing the Cassette Drive Rollers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking the CAM and Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking the Batteries on the UPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Checking the Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Checking the Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Publication History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

2 Revision B – PM4825-1
Necessary Materials

Part No. Description Quantity Notes


DG4825-1 Diagnostics for the KODAK DIRECTVIEW CR 825/850 1 Available on the Carestream
Systems Health Service Portal
AR4825-1 Adjustments and Replacements for the KODAK 1 Available on the Carestream
DIRECTVIEW CR 825/850 Systems Health Service Portal
II4825-1 Installation Instructions for the KODAK DIRECTVIEW 1 Available on the Carestream
CR 825/850 Systems Health Service Portal
--- Soap - solution --- can acquire locally
UN3E0899 Lint-free cloth (300 package) --- can acquire locally
IC8271 Lint Free Cloth (150 package) --- can acquire locally
UN1C8026 Lubricant - Grease SUPERLUBE TEF ---
TLUN-4071 Vacuum cleaner ---
UNSP1C80 Lens cleaner --- can acquire locally
00
UN952807 Lens paper --- can acquire locally
UN5E6777 Spindle oil (3-in-1 Oil, SAE 10W oil) --- can acquire locally
SP1C8067 Canned air (2 package) --- can acquire locally
SP6J5334 CMOS Battery Replacement Kit 1 Includes:
As required: See Service Bulletin 8J6266 • 5 CMOS batteries
• 5 labels
7H9035 PM kit 1
6E5160 Gasket - light seal 2 Included in PM KIT
1F2215 Plush, 3/8 clamp bar 2
1F5142 Light baffle - right 1
1F6140 Light baffle - left 1
1F7859 Front light seal 1
1F2941 Rear clip assembly 1
1F5817 Plush - light seal 750 HGH 2
1F5661 Roller assembly- cassette drive 2
5H9743 Right light seal assembly 1

PM4825-1 – Revision B 3
Preventive Maintenance

Maintenance Procedures

Biohazard
• Field personnel must complete the safety training “Bloodborne Pathogen Program” before doing these
procedures.
• You must use NITRILE or NEOPRENE GLOVES when you clean the equipment.

Important
• Every 6 months: Do all procedures in the sequence indicated in the Service publications. Refer to Periodic
Maintenance Checklist, AG3345, for frequency of Preventive Maintenance.
• Ask the customer about problems with the equipment.

Important
Ensure the latest cyber security updates have been applied to the product. See Service Bulletin (SBAH2718 for
details).

Procedure
Checking the Error Logs
Checking the Scan Operation
Cleaning the Computer Filter
Checking the CMOS Battery
Cleaning the PMT/DAS Board, Parts and Collector Assembly
Cleaning the PMTs and Collector Assembly
Cleaning and Applying Lubricant to the Cassette Handling Assembly
Cleaning the Bar Code Reader
Cleaning the Extraction Bar
Cleaning and Applying Lubricant to the Slow Scan Slides
Cleaning the Erase Lamp Assembly
Cleaning the System
Installing the Light Gaskets on the Top Cover
Installing the Clamp Bar Plush
Installing the Right and Left Light Baffels
Installing the Rear Clip Assembly
Installing the New Light Seal Plush Assembly
Installing the Cassette Drive Rollers
Checking the CAM and Motor
Checking the Adjustments
Checking the Operation
Checking the Batteries on the UPS

4 Revision B – PM4825-1
Checking the Error Logs

Important
You must use the Checking the Error Logs procedure in Diagnostic Tests for the KODAK DIRECTVIEW CR 825/850
Systems, DG4825-1.
[1] Do the Checking the Error Logs procedure.

Important
If you see more than one error code of the same component or assembly, you might have to install new parts or do
adjustments during this procedure.
[2] Do an evaluation of the error logs.
[3] Continue with Checking the Scan Operation.

Removing Images from the System


Note
If there are undelivered images, request the images from the Key Operator.
[1] Verify that all images have been delivered.
[2] Notify the Key Operator that you must delete all images and patient records.
[3] From the Main Menu, select:
• Key Operator Functions
• Manage Patient Exam Records
• Delete All Patients
• Yes at the message prompt
• Yes at the next message prompt
• OK at the next message prompt
[4] Select Find Local to verify that the deletion is complete. The message “No Matches Found” indicates that tall
images and patient records were deleted.
[5] Continue with Checking the Scan Operation.

Checking the Scan Operation

Important
You must check the scan operation before de-energizing the system.
[1] Observe the scan operation or all cassette sizes and make repairs before doing preventive maintenance
procedures.
[2] Continue with Cleaning the Computer Filter.

PM4825-1 – Revision B 5
Preventive Maintenance

Cleaning the Computer Filter

2.4 GHz
Caution
computer Dangerous Voltage.

Important
You must use the appropriate procedures in the
Installation Instructions for the KODAK DIRECTVIEW
CR 825/850 Systems, II4825-1.
[1] De-energize the system using the Energizing the
System procedure in the Installation Instructions
for the KODAK DIRECTVIEW CR 825/850
Systems, II4825-1.
[2] Remove the rear panel. See Rear Panel in the
Adjustments and Replacements for the KODAK
FILTER DIRECTVIEW CR 825/850 Systems, AR4825-1.
[3] Use the vacuum cleaner to remove dust from the
filter in the computer.
[4] Install the rear panel.
[5] Continue with Checking the CMOS Battery.

H194_0029GCA
H194_0029GC

6 Revision B – PM4825-1
Checking the CMOS Battery
[1] Check the age of the CMOS battery by reading the replacement date on the label located on the PC. Replace
the battery if it is 5 or more years old.
[2] If there is no date label indicating the last time the CMOS battery was replaced, determine the approximate age
of the CMOS battery by referring to the manufacture date on the data plate of the CR Scanner and to the
following table of computer implementation dates:
Computer Console Model Implementation Date
8187-Q9U 2004-11 (November 2004)
8187-SEU 2005-03 (March 2005)
8212 2006-04 (April 2006)
8810 2007-10 (October 2007)
HP 5700 2008-12 (December 2008)
HP 5800 2012-09 (September 2012)

[3] If the battery is less than 5 years old, attach a label to the computer console, and write the manufacture date
found on the data plate of the CR Scanner or the implementation date listed in the previous table, whichever is
more recent.

Note
Attach the label in a discreet place, but where the FE is likely to see it during the next PM visit.
[4] Replace the CMOS battery if it is 5 or more years old. Add a label to the PC with the replacement date. See:
Publication Section Procedure
Adjustments and Replacements for the KODAK Replacements CMOS Battery
DIRECTVIEW CR 825/850 Systems, AR4825-1 or
Replacement Instructions for CMOS Battery in the
Computer Console on CR Systems, PN 6J5133.

[5] Continue with Cleaning the PMT/DAS Board, Parts and Collector Assembly.

PM4825-1 – Revision B 7
Preventive Maintenance

Cleaning the PMT/DAS Board, Parts and Collector Assembly

Important
You must use the appropriate procedures in the Adjustments and Replacements for the KODAK DIRECTVIEW CR
825/850 System, AR4825-1.

ESD
PMT/DAS BOARD
Possible damage from electrostatic discharge.

Wear an ESD strap when handling or cleaning the PMT


board and collector assembly. The system is
susceptible to static discharge.
[1] Remove the PMT/DAS board using
Replacements/PMT/DAS Board in Adjustments
and Replacements for KODAK DIRECTVIEW CR
825/850 System, AR4825-1.

Caution
Use the vacuum cleaner to remove dust before you use
the canned air.
[2] Use the vacuum cleaner to remove dust from the
PMT/DAS board.
[3] If necessary, use the canned air to remove dust
from the PMT/DAS board.
Continue with Cleaning the PMTs and Collector
Assembly.

H194_1076GCA
H194_1076GC

8 Revision B – PM4825-1
Cleaning the PMTs and Collector Assembly
[1] Remove the PMTs and collector assembly using
2 PMTs
the procedures in Replacements/PMTs and
Collector Assembly in Adjustments and
Replacements for the KODAK DIRECTVIEW CR
COLLECTOR AY
825/850 Systems, AR4825-1.
[2] Remove
• 2 screws
• cover
• foam
• 2 PMTs
• collector assembly

Caution
• Use the vacuum cleaner to remove dust before you
use the canned air.
• Use the canned air to move debris from the optical
cavity.
• Use care when cleaning mirror surfaces. Surfaces
are made of glass and reflective film mirror tape
and scratch easily. Use only canned air to remove
debris. See SB8J5790 for cleaning procedures.
OPTICAL
H194_1077GCA
H194_1077GC CAVITY
Important
Pay attention to the laser path. It must be free from any
fiber or images will be affected by vertical lines.
[3] Use the vacuum cleaner to remove dust from:
• 2 PMTs
• collector assembly
• optical cavity
[4] If necessary, use the canned air to remove dust
from:
• 2 PMTs
• collector assembly
• optical cavity
[5] Use the procedures in the Adjustments and
Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1, to install:
• PMTs and collector assembly
• PMT/DAS board
[6] Perform collector alignment using procedures to
avoid damage to the collector glass using
procedures in the Adjustments/Collector
Assembly in Adjustments and Replacements for
the KODAK DIRECTVIEW CR 825/850 System,
AR4825-1.

PM4825-1 – Revision B 9
Preventive Maintenance

Cleaning and Applying Lubricant to the Cassette Handling Assembly


[1] Remove the cassette handling assembly. See the
Adjustments and Replacements for the KODAK
DIRECTVIEW CR 825/850 Systems, AR4825-1.

Caution
Use the vacuum cleaner to remove dust before using
the canned air.
[2] Use the vacuum cleaner to remove dust from the
cassette handling assembly.
[3] If necessary, use the canned air to remove dust
from the cassette handling assembly.

CASSETTE
HANDLING AY
H194_1078ACA
H194_1078AC

DRIVE
ROLLER AY

CAM BLOCK

STOP PIN

RIGHT LIGHT
SEAL AY
LEFT PLUSH CAM BLOCK
ASSEMBLY
STOP PIN FRONT LIGHT SEAL
H194_1079_C1

[4] Apply lubricant to:


• stop pin for the right light seal assembly
• stop pin for the left light seal assembly
• CAM block for the front light seal
• CAM block for the drive roller assembly
[5] Continue with Cleaning the Bar Code Reader.

10 Revision B – PM4825-1
Cleaning the Bar Code Reader
[1] If necessary, remove the cassette handling
assembly. See the Adjustments and
Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1.

Caution
Use the vacuum cleaner to remove dust before using
the canned air.
[2] Use the vacuum cleaner to remove dust from the
bar code reader.
[3] If necessary, use the canned air to remove the
dust from the bar code reader.

BAR CODE
READER H194_1080ACA
H194_1080AC

[4] If the canned air does not remove the contamination from the bar code reader, clean the bar code reader with:
• lens cleaner solution, UNSP1C8000
• lens paper, UN952807
• Lint free cloth or a microfiber cloth if needed

Caution
Obtain the lens cleaner SDS or AIS and familiarize yourself with it before starting this procedure.
[5] Continue with Cleaning the Extraction Bar.

Cleaning the Extraction Bar


[1] If necessary, remove the cassette handling
assembly. See the Adjustments and
Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1.

Caution
Use the vacuum cleaner to remove dust before using
the canned air.
[2] Clean the extraction bar with a lint-free cloth.
[3] Install the cassette handling assembly. See the
Adjustments and Replacements for the KODAK
DIRECTVIEW CR 825/850 Systems, AR4825-1.
EXTRACTION BAR
[4] Continue with Cleaning and Applying Lubricant to
the Slow Scan Slides.
H194_1081ACA
H194_1081AC

PM4825-1 – Revision B 11
Preventive Maintenance

Cleaning and Applying Lubricant to the Slow Scan Slides

Carriage
holes
Extraction Bar

Slow Scan
Slides
Lead Screw

Slow Scan Ay

H196_3003BCA
H196_3003BC PCB Slide

[1] Use a lint-free cloth to clean:


• Slow scan slides
• Top of the extraction bar

Caution
Do not apply spindle oil on the lead screw or nut. It will
destroy the casting can cause the slide to fail. Do not
apply ANY lubricant of any kind to the lead screw!

Caution
Obtain the spindle oil SDS or AIS and familiarize
yourself with it before starting this procedure.
holes
[2] Apply 2 drops of spindle oil on each of the 2 slow
scan slides and wipe the oil along the slides.
[3] Rotate the lead screw up and down to distribute
Lead Screw the spindle oil across the surface of the 2 slow
scan slides.
Slow Scan [4] For the PCB configuration, apply at least 2 drops
Slide
of spindle oil in the holes on the carriage.
[5] For the Kerk Slide configuration, remove the 3
aluminum disks on the carriage and insert at least
2 drops of spindle oil in each of the 3 holes on the
carriage.
[6] Continue with Cleaning the Erase Lamp
Assembly.

H206_0841_C1 Kerk Slide

12 Revision B – PM4825-1
Cleaning the Erase Lamp Assembly
[1] Disengage the latch.
LATCH [2] Pull the erase lamp assembly.
[3] Use a lint-free cloth to clean erase lamps.
[4] Use the vacuum cleaner to remove dust from the
erase lamp assembly.
[5] If necessary, use the canned air to remove the
dust from the erase lamp assembly.
4
5
[6] Continue with Cleaning the System.
1

ERASE
2
3
LAMPS 4
5

ERASE
LAMP AY
H177_2104GCB
H177_2104GC

Cleaning the System

Caution
Dangerous Voltage.

Important
You must use the appropriate procedures in the Installation Instructions for the KODAK DIRECTVIEW CR 825/850
Systems, II4825-1.
[1] De-energize the system using the Energizing the System procedure in the Installation Instructions for the
KODAK DIRECTVIEW CR 825/850 Systems, II4825-1.
[2] Use the vacuum cleaner to remove dust from the inside of the system.
[3] Continue with Installing the Light Gaskets on the Top Cover.

PM4825-1 – Revision B 13
Preventive Maintenance

Installing the Light Gaskets on the Top Cover

TOP COVER narrow ends of


LIGHT SEALS Top View back edge of SLOT

LEFT
LIGHT
SEAL

RIGHT
LIGHT
SEAL H194_1083BCA
front edge of SLOT H194_1083BC

[1] Open the top cover. See the Adjustments and Replacements for the KODAK DIRECTVIEW CR 825/850
Systems, AR4825-1.

Important
• The narrow end of the light seal must be toward the front edge of the slot.
• The light seal must make a seal against the back edge of the slot.
[2] Remove:
• left light seal
• right light seal
[3] Remove the existing gasket from:
• left light seal
• right light seal
[4] Install the new gasket on:
• left light seal
• right light seal
[5] Install:
• left light seal
• right light seal
[6] Continue with Installing the Clamp Bar Plush.

14 Revision B – PM4825-1
Installing the Clamp Bar Plush
[1] Remove:
LEFT CASSETTE GUIDE
• 2 covers
RIGHT CASSETTE GUIDE • 4 screws
• left cassette guide
• right cassette guide
• 2 screws
2 COVERS
4 SCREWS • cassette load cradle

2 SCREWS

H194_1062GCA
CASSETTE
H194_1062GC LOAD CRADLE

[2] Remove:
CLAMP BAR PLUSH • 2 screws
• clamp bar plush
CLAMP BAR [3] Install:
• new clamp bar plush
• 2 screws
2 SCREWS
[4] Continue with Installing the Right and Left Light
Baffels.

H194_1084GCA
H194_1084GC

PM4825-1 – Revision B 15
Preventive Maintenance

Installing the Right and Left Light Baffels


[1] Remove:
• 2 screws
• cassette load cradle
CASSETTE
LOAD CRADLE

H194_1074ACA
2 SCREWS H194_1074AC

[2] Remove:
2 SCREWS
• 2 screws
LEFT LIGHT 2 SCREWS
BAFFLE • left light baffle - discard
• 2 screws
• right light baffle - discard
[3] Install:
• left light baffle
• 2 screws
• new right light baffle
RIGHT LIGHT • 2 screws
BAFFLE
[4] Reverse Step 1.
[5] Continue with Installing the Rear Clip Assembly.
H194_1085ACA CASSETTE
H194_1085AC LOAD CRADLE

16 Revision B – PM4825-1
Installing the Rear Clip Assembly
[1] Open the top cover. See the Adjustments and
Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1.
[2] Remove:
REAR BRACKET AY
• 4 screws
• rear bracket assembly

4 SCREWS

H194_1086GCA
H194_1086GC

[3] Remove:
• 2 screws
• rear clip assembly - discard
REAR CLIP AY [4] Install:
• rear clip assembly
• 2 screws
[5] Reverse Step 2.
[6] Continue with Installing the New Light Seal Plush
Assembly.

H194_1087ACA
2 SCREWS H194_1087AC

PM4825-1 – Revision B 17
Preventive Maintenance

Installing the New Light Seal Plush Assembly


[1] Remove:
LEFT CASSETTE GUIDE
• 2 covers
RIGHT CASSETTE GUIDE • 4 screws
• left cassette guide
• right cassette guide
• 2 screws
2 COVERS
4 SCREWS • cassette load cradle
• 2 screws
• Left plush assembly
• 2 screws
• Right plush assembly
[2] Install
• Light seal plush - right
• Light seal plush - left
2 SCREWS

H194_1062GCA
CASSETTE
H194_1062GC LOAD CRADLE

[3] Remove:
• 2 screws
• Front light seal
• 2 screws
• Right light seal assembly
[4] Install:
• front light seal assembly
• 2 screws
• Right light seal assembly
• 2 screws
front light seal
[5] Reverse Step 1.
[6] Continue with Installing the Installing the Cassette
Drive Rollers
2 screws light seal plush

right light seal


assembly

right plush
assembly 2 screws h194_1088_c1

18 Revision B – PM4825-1
Installing the Cassette Drive Rollers
[1] Remove the cassette drive assembly. See the
Adjustments and Replacements for the KODAK
DIRECTVIEW CR 825/850 Systems, AR4825-1.
[2] Remove and discard the 2 cassette driver rollers.
See the Adjustments and Replacements for the
KODAK DIRECTVIEW CR 825/850 Systems,
AR4825-1.
[3] Use the procedures in the Adjustments and
Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1, to install:
CASSETTE • 2 new cassette drive rollers
DRIVE AY • cassette drive assembly
[4] Continue with Checking the CAM and Motor.

2 CASSETTE
DRIVE ROLLERS
H194_1007GCB
H194_1007GC

Checking the CAM and Motor


[1] Remove the CAM motor assembly. See the
MOTOR CAM Adjustments and Replacements for the KODAK
DIRECTVIEW CR 825/850 Systems, AR4825-1.
[2] Pull the CAM toward the motor. Check that the
CAM does not move on the shaft of the motor by
pulling to CAM toward the motor.
[3] Does the CAM move on the shaft?
Yes No
Install new CAM motor Continue with Step 4.
assembly.

[4] Use the procedures in the Adjustments and


Replacements for the KODAK DIRECTVIEW CR
825/850 Systems, AR4825-1, to install the CAM
SHAFT motor assembly.
H177_3031ACA [5] Continue with Checking the Adjustments.
H177_3031AC

PM4825-1 – Revision B 19
Preventive Maintenance

Checking the Batteries on the UPS

UPS Operational Panel

A B C D

/TEST Bypass
Line On Line

On Battery

Replace Battery LED


H226_1028ACA
H226_1028AC

[1] Turn off the computer and leave the UPS on.
[2] Press and hold the I/Test button for a few seconds.
The LEDs A, B, C, and D will turn on for several seconds depending on the battery charge level.

Note
If the batteries are in good condition, all of the LEDs turn on. If they are not all on, consider replacing the batteries
even if the Replace Battery LED doesn’t blink. See publication 8H9517.

Checking the Adjustments

Important
You must use the procedures in the Adjustments and Replacements for the KODAK DIRECTVIEW CR 825/850
Systems, AR4825-1, to do the adjustments.
[1] Do:
• fold mirror
• collector assembly
• CAM sensor board A8
• CAM motor assembly
• sled CAM follower
• clamp bar
• cassette end stop
• rear cassette sensor
• extraction bar home position
• plate positioning assembly
• right and left plush assembly - alignment
• left plush assembly - vertical position
• internal bar code reader
• cassette load cradle
• intermediate plate (only if required after the cassette handling adjustment)

20 Revision B – PM4825-1
Important
Check to see if the cassette is opened/scanned/closed/ejected correctly. Perform any necessary adjustments/
replacements. The items in this list interact and need to work properly.
[2] Continue with Checking the Operation.

PM4825-1 – Revision B 21
Preventive Maintenance

Checking the Operation


Before doing the calibration, do the following:
[1] Make a flat field exposure of a good plate for 35 x 43 and 24 x 30 mammography cassette (if available), the
same procedure as a calibration.
[2] Scan the plate.
[3] Adjust the image to a window width (WW) of approximately 500 and a window level (W/L) approximately the EI
value.
The resulting flat image should be visible. The EI should be approximately 3300 (GP/PQ) or 3000 (EHR/SNP).

Note
If you experience:
• vertical lines: check the laser opening on the collector for fibers present.
• blurred vertical bands: check fold mirror for oxidation (only for silver mirrors)
Do not calibrate if you observe vertical lines. First, resolve the problem by cleaning the collector aperture. If
horizontal artifacts appear, these are due to plate transport problems and cannot be resolved with a calibration.
[4] Do the following procedures from the Adjustments and Replacements for the KODAK DIRECTVIEW CR 825/
850 Systems, AR4825-1:
• Calibration for the CR 825/850 System
• Checking the Operation
[5] Continue with Checking the Batteries on the UPS.

22 Revision B – PM4825-1
Publication History

Publication Changed
Date Revision Pages Updates
2004-01-09 n/a --- New Publication
2005-12-10 n/a 3,6 Revised
2006-12-16 n/a 3 Revised
2009-04-30 n/a 3-6, 15,16 Revised
2011-06-27 n/a Various Added a procedure for checking the CMOS battery; new cover
format; EN1041:2008 dates
2015-07-10 n/a 3-9, 11, 12, Revised
19, 20, 23
2015-11-17 n/a Various Revised
2018-10-10 A All Revision A added to match PLI nomenclature.
Updated high voltage symbols.
4 Updated reference to Periodic Maintenance Checklist, AG3345.
Added cyber security warning.

PM4825-1 – Revision B 23
CARESTREAM and DIRECTVIEW are trademarks of Carestream Health, Inc.
The Kodak trademark and trade dress are used under license from Kodak.

Printed in U.S.A.