Professional Documents
Culture Documents
Amendments
Revision 1 January 2006 First Release
Revision 2 June 2006 Change to reflect current format
Revision 3 November 2007 Deletion in specifications
Revision 4 March 2012 WEEE information added. FTV information added
Revision 5 July 2013 TPK-R information added, photographs updated
Revision 6 August 2015 Minor amendments
Revision 20 January 2020 Detail Changes
Revision 21 November 2020 TPK image change
2
INSTRUCTION MANUAL
CONTENTS PAGE
1 Introduction ....................................................................................................... 4
1.1 The TPK and TPK-R .................................................................................... 4
1.2 General description and overview of operation ............................................ 5
2 Setting Up .......................................................................................................... 7
2.1 Location........................................................................................................ 7
2.2 Assembly ...................................................................................................... 7
3 Operation ........................................................................................................... 8
3.1 Determining the Test Flow Rate ................................................................... 8
3.2 Measuring Delivered Dose Uniformity of DPIs ............................................. 9
3.3 Measuring the Aerodynamic Particle Size of DPIs ..................................... 10
4 Maintenance and Cleaning ............................................................................. 12
4.1 Maintenance ............................................................................................... 12
4.2 Cleaning ..................................................................................................... 12
5 Specification .................................................................................................... 13
5.1 TPK and TPK-R General ............................................................................ 13
5.2 TPK and TPK-R Physical Dimensions........................................................ 13
5.3 TPK and TPK-R Operating Environment .................................................... 13
5.4 TPK and TPK-R Solenoid Valve ................................................................. 13
5.5 TPK and TPK-R Absolute Pressure Transducer ........................................ 13
5.6 TPK and TPK-R Differential Pressure Transducer ..................................... 13
5.7 TPK and TPK-R Timer ............................................................................... 13
6 Decommissioning, dismantling, disabling and scrapping. ......................... 14
3
INSTRUCTION MANUAL
1 Introduction
Flow control
Timer Button valve
Outlet connector
(to pump)
P1 port
Inlet connector
(to test apparatus)
4
INSTRUCTION MANUAL
The Critical Flow Controller Model TPK incorporates the following components (see
also Figure 1.3 – items highlighted in red):
An in-built differential pressure meter, for measuring the pressure drop over
the inhaler (P1).
A timer controlled two-way solenoid valve to control flow duration.
A flow control valve to adjust flow rate.
Absolute pressure meter for measuring absolute pressures at points P2 and
P3, immediately upstream and downstream of the flow control valve.
An ON/OFF switch on the rear panel.
An outlet hose connector on for connecting to the vacuum pump.
An inlet hose connector for connecting to the test apparatus.
A small pressure port for measuring the differential pressure P1.
A timer which can be set up to 999.9 seconds for controlling the two-way
solenoid valve.
P1, P2 and P3 buttons with corresponding LEDs.
LED display for displaying pressure (in kPa).
Timer button for starting the timer.
5
INSTRUCTION MANUAL
The Critical Flow Controller Model TPK and TPK-R comprises the parts described in
Figure 1.4 and Figure 1.5 below:
LED Display
Timer
P1, P2 and P3
buttons and LEDs
Outlet
Connector
(to pump)
P1 port
Inlet connector
(to test apparatus)
6
INSTRUCTION MANUAL
2 Setting Up
2.1 Location
Site the TPK on a level surface capable of withstanding the weight of the unit (see
specification).
Ensure that the TPK is readily accessible to the user and placed at a convenient
height for access to the on/off switch, the inlet and outlet connectors, and the front
panel.
Ensure that the TPK is in a clean and dust-free environment.
Minimise as far as possible the distance between the TPK, vacuum pump and
test apparatus in order to reduce the tubing length between each item.
2.2 Assembly
Once in position, plug the mains cable into the socket on the rear of the TPK and
switch on.
Allow 15 minutes after switching on for the electronics of the TPK to stabilise
before use.
The flow inlet (on the right panel) will have a small (12 mm O.D.) hose connector
in position. This is for connection directly to the test apparatus (cascade impactor
or dosage unit sampling apparatus) using suitable tubing.
If the TPK is being used with a Next Generation Pharmaceutical Impactor (NGI)
then it may be necessary to replace the small hose connector with the large (17
mm O.D.) hose connector supplied with the TPK.
NOTE: A 24mm spanner/wrench will be required for changing hose connectors – not
supplied.
7
INSTRUCTION MANUAL
Before using the TPK, ensure that the mains power supply is correct
for the TPK being used. Details of the TPK voltage and frequency can
be found on the specification label on the rear panel of the TPK.
Incorrect mains power supply could cause irreparable damage to the
TPK.
3 Operation
NOTE: It is highly recommended that test monographs contained within European
Pharmacopoeia Section 2.9.18 and the United States Pharmacopoeia Section 601
are used in conjunction with this instruction manual.
Connect the TPK to the Sampling Apparatus for DPIs as shown in Figure 3.1.
Vacuum
Pump TPK P1 Port
Insert the inhaler into the inlet of the sample tube of the Sampling Apparatus
for DPIs, using a suitable mouthpiece adapter.
Connect the tube marked P1 from the TPK to the pressure tap marked P1 in
the sample tube of the Sampling Apparatus for DPIs.
Switch on the vacuum pump.
Set the timer to a high value (e.g. 500 seconds) and then open the two-way
solenoid valve by pressing the “Timer” button.
Press the "P1" button and adjust the flow control valve until the differential
pressure on the LED display indicates 4.0 kPa.
Press the "P2" button and then the "P3" and note the respective readings on
the LED display.
P2 must not be less than 2 x P3 for sonic flow to occur. If sonic flow is not
achieved then do not proceed with the test.
Replace the inhaler with a suitable flow meter and flow meter adapter and
then measure the flow rate (Q) as shown in Figure 3.2.
8
INSTRUCTION MANUAL
P1 Port
TPK
NOTE: If 4 kPa cannot be achieved at P1 (in the case of extremely low resistance
inhalers) before a flow rate of 100 L/min is exceeded, then adjust the flow control
valve to achieve a flow rate of 100 L/min.
NOTE: The purpose of the absolute pressure measurements (P2 and P3) is to
ensure that critical or sonic flow conditions exist in the flow control valve so that
the flow rate through the particular test apparatus remains constant and
unaffected by fluctuations in the pump.
NOTE: If the vacuum pump is fitted with an unregulated vacuum inlet, then use
this to bypass the flow control valve that may be fitted to the pump. If sonic flow is
still not achieved, then a more powerful vacuum pump, or a secondary pump,
may be required. Contact Copley Scientific for further details.
NOTE: Calculate the test time (T) such that 4 L of air is drawn through the inhaler
(i.e. T = 240/Q).
Replace the inhaler and discharge the powder into the Sampling Apparatus for
DPIs, by activating the timer controlled solenoid valve, by pressing the "Timer"
button.
Repeat as necessary to achieve the minimum recommended dose.
Assay the contents of the Sampling Apparatus for DPIs for drug after rinsing
the filter and interior with a suitable solvent.
9
INSTRUCTION MANUAL
Connect the TPK to the cascade impactor and attach a suitable flow meter to
the inlet the induction port as shown in Figure 3.3.
Flow Meter
Adapter
Flow Meter
TPK
Cascade
Impactor
Figure 3.3 Measuring the Aerodynamic Particle Size of DPIs – Flow Rate Determination
NOTE: The purpose of the absolute pressure measurements (P2 and P3) is to
ensure that critical or sonic flow conditions exist in the flow control valve so that
the flow rate through the particular test apparatus remains constant and
unaffected by fluctuations in the pump.
NOTE: If the vacuum pump is fitted with an unregulated vacuum inlet, then use
this to bypass the flow control valve that may be fitted to the pump. If sonic flow is
still not achieved, then a more powerful vacuum pump, or a secondary pump may
be required. Contact Copley Scientific for further details.
10
INSTRUCTION MANUAL
NOTE: Calculate the test time (T) such that 4 L of air is drawn through the inhaler
(i.e. T = 240/Q).
Insert the mouthpiece of the primed/loaded inhaler into the inlet of the
induction port using a suitable mouthpiece adapter as shown in Figure 3.4.
Vacuum
Pump TPK
Discharge the powder into the Sampling Apparatus for DPIs, by activating the
timer controlled solenoid valve, by pressing the "Timer" button.
Repeat as necessary to achieve the required number of doses.
Assay the contents of impactor in accordance with the impactor
manufacturer's instructions, in conjunction with the USP and EP monographs.
11
INSTRUCTION MANUAL
Operational verification during calibration can be aided by the ‘Flow Time Verification
Kit’ which measures the actual duration that the vacuum is present on the inlet to the
TPK (or TPK-R). Please contact Copley Scientific Ltd. for details.
4.2 Cleaning
The metal case and front panel can be cleaned with mild soapy water, taking care
not to allow water to run down the sides of the machine or through any of the
apertures/sockets/switches in the case.
Dry immediately after cleaning.
Internal cleaning is not required.
12
INSTRUCTION MANUAL
5 Specification
5.1 TPK and TPK-R General
230 V, 50 Hz 110V, 60 Hz
Mains supply See specification label See specification label
Class I - MUST BE Class I - MUST BE
Isolation classification
EARTHED EARTHED
Fuse 1 x 2 A (Type T) 1 x 2 A (Type T)
13
INSTRUCTION MANUAL
14
INSTRUCTION MANUAL
7 Technical Support
Should you have any queries, or require technical information regarding the Critical
Flow Controller Model TPK, or any other equipment we supply, please do not
hesitate to contact us.
www.copleyscientific.com
e-mail: support@copleyscientific.co.uk
web site: www.copleyscientific.com
Copley Scientific AG
Erlenstrasse 27
Postfach 152
CH-4106 Therwil
Switzerland
e-mail: service@copleyscientific.ch
web site: www.copleyscientific.com
15