Rotor Fitment and Details

CenturionScientific Limited
Welcome
It is of p
a
import ramount
read thisance you
doc
before u ument
se

to our Rotor Fitment, balancing,


care and safety manual
Centrifuges can be dangerous if rotors are not correctly fitted, samples
are not correctly loaded or poor samples tubes are used.
Rotor fitment is from Page 8-41. Please look up the rotor to be used to
go to the relevant pages. Read carefully and follow the instructions.

We recommend: Tubes can break or rupture causing


damage or sample failure. We know
● Quality tubes are used how much time & work it may take
● Y
 ou balance within to get to this point to be ruined by a
1 gram cheap tube. Although our imbalance
system is sophisticated and will shut
● Y
 ou balance the rotor down if exceeded. Centrifugation is
correctly. about seperation and good balancing
● Y
 ou always fit the lid gives better separations.
supplied Always place balanced tubes opposite
● W
 e recommend that to each other, as shown, in the pictures.
you check the rotor for Lids are used to prevent aerosol effects
cracks reguarly. and reduce air flow and therefore noise.

! If a crack is found. DO NOT USE.


Rotors are under a constant strain and need to be kept clean and checked for
strain.
DO NOT exceed the life of the rotor 7 years maximum.
For samples of an unknown nature we recommend you check before centrifugation
by using the Chemical Resistance Table, page 42 onward.
We do not recommmend the use of strong acids, strong Alkalis or strong
radioactive samples, these can damage the rotors and the Centrifuge.

2 www.centurionscientificglobal.com
Index
Page 4: Balance. Haematocrit and Micro
Page 5: Balance fixed angle
Page 6: Balance swing outs
Page 7: Balance Cytology & Cell wash
Page 8 - 9: BRK5401
Page 10 - 11: BRK5424. BRK5436. BRK5448. BRK5494
Page 12 - 13: BRK5407. 5408
Page 14 - 15: BRK5412. BRK5406
Page 16 - 17: BRK5308. BRK5324. BRK5100
Page 18 - 19: BRK5431. BRK5434
Page 20 - 21: BRK5508.
Page 22 - 23: BRK5505.
Page 24 - 25: BRK5510. New BRK1000, with 100ml bucket
Page 26 - 27: BRK5510. New BRK1000, with 250ml bucket
Page 28 - 29: BRK5553. New BRK3000 & BRK4000 (Same fitment)
Page 30 - 31: BRK5540
Page 32 - 33: BRK5206. BRK5212
Page 34 - 35: BRK5210. BRK5208. BRK5224
Page 36 - 37: BRK5256
Page 38 - 39: 4420. 4430. 4460
Page 40 - 41: 5450. C washer rotor
Page 42 Cleaning
Page 43: Nomograph
Page 44 - 54: Chemical Resistance table
Page 55: Contact us
Page 56 - 59: Certificates

Rotor Fitment and Details 3


Rotor Balance
Haematocrit Rotor

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
CHECK the Rim seal gasket and replace regularly (Part number 8011)
CHECK the rotor screw fixing tightness regularly

Micro Rotors

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
CHECK the two top seals and replace regularly (Part number 8044)
CHECK the rotor screw fixing tightness regularly

4 www.centurionscientificglobal.com
Small fixed angle Rotors

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
CHECK The rotor screw fixing tightness regularly

Large fixed angle Rotors

✔ ✘ ✘

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
CHECK The rotor screw fixing tightness regularly

Rotor Fitment and Details 5


Small swing out Rotor

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL the Tube Holders and any supports
CHECK The rotor screw fixing tightness regularly

Medium – Large Swing out Rotors

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL 4 Buckets and 4 Adaptors supplied
CHECK the rotor screw fixing tightness regularly

Microtitor Rotor

✔ ✘
! ALWAYS load your plates opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit ALL 2 Buckets supplied
CHECK the rotor screw fixing tightness regularly

6 www.centurionscientificglobal.com
Cytology Rotors

✔ ✘ ✘
! ALWAYS load your holders opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit the lid supplied
Check the two top Gaskets and replace regularly (Part Number 8055)
CHECK the rotor screw fixing tightness regularly

Manual Cell Washing Rotor

✔ ✘ ✘
! ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
Regularly smear a small amount of Silicon grease over the O rings
Check the inner and outer O rings and replace regularly (Part Number 8066)
CHECK the rotor screw fixing tightness regularly

Rotor Fitment and Details 7


Rotor fitment
BRK5401
5
1. Rotor
1 2. Threaded Fastener
2 34 6
3. M6 Spring Washer
4. Rotor Screw (M6 x 20mm)
5. Rotor Lid
7
6. Rotor Lid Knob
7. Allen Key

1 Place the rotor on


the shaft of the
motor. 6 Place samples in
as required.
2 Locate threaded
fastener (Item
2) into the three
interlocking pin
holes on the rotor.

3 Place M6 Spring 7 Place rotor lid


washer (Item3) (Item 5) on
onto rotor screw threaded fastener.
(Item 4).
4 Insert screw
and washer into
central hole of
threaded fastener.

5 Hand tighten with 8 Hand tighten lid


Allen Key (Item 7) using rotor knob
(Item 6).

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED ON A


REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid.

8 www.centurionscientificglobal.com
Rotor fitment

Rotor BRK5401
Rotor type 24 x capillary/12 x 2ml
Tube size max 2 x 75mm/11 x 40mm
Minimum speed Rpm 500
Maximum speed Rpm 12,000
Maximum Rcf (G) 13,500
Radius max cms 8.4
Sample tube angle ° 0/60
Acceleration time (secs) 25
Deceleration time (secs) 25
Autoclavable (frequecncy) 121°C (10)

Accessories
Part Number Description
8012 Plain Capillary tubes pack 1000
8013 Heparanised Capillary tubes pack 1000

8003 PCV reader 8014 Sealing Clay

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 9


Rotor fitment
BRK5424. BRK5436.
BRK5448. BRK5494
1

4 2
1. Rotor Screw (M6 x 20mm)
3 2. M6 Spring Washer
3. Threaded Fastener
5 4. Torque Key
5. Rotor Lid

1 Locate threaded Hand tighten with


fastener (Item 4 Torque Key
3) into the three (Item 4)
interlocking pin
holes on the rotor.

2 Place M6 Spring 5 Always fit the lid


washer (Item 2) provided
onto rotor screw
(Item 1) 6 To fasten the
rotor lid, Place
3 rotor lid (Item
Insert screw
and washer into 5) on threaded
central hole of fastener hand
threaded fastener. tighten using the
rotor knob.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

10 www.centurionscientificglobal.com
Rotor Details
Microtube rotors

Rotor BRK5424 BRK5436 BRK5448 BRK5494


Rotor type 24 x 2.2ml 36 x 0.5ml 48 x 0.2 to 0.4ml 4 x Pcr Strips
Tube size max 11 x 50mm 8 x 30mm 6 x 40mm 8 per strip
Minimum speed Rpm 500 500 500 500
Maximum speed Rpm 15,000 15,000 15,000 15,000
Maximum rcf (G) 22,000 22,000 22,000 22,000
Radius max cms 8.75cms 8.75cms 8.75cms 8.75cms
Sample tube angle ° 45 45 45 45
Acceleration time (secs) 20 20 20 20
Deceleration time (secs) 20 20 20 20
Autoclavable (frequecncy) 121°C (10) 121°C (10) 121°C (10) 121°C (10)

Reducers

Rotor BRK5424
Part number RS04 (0.2- 0.4ml)
Tube size max 6 x 30mm
Part number RS05 (0.5ml)
Tube size max 8 x 30mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 11


Rotor fitment
BRK5407. BRK5408

2
3
1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key
1

1 Place the rotor on Place the rotor


the shaft of the 3 fixing screw
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (Item 3).
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

12 www.centurionscientificglobal.com
Rotor Details

Rotor BRK5407 BRK5408


Rotor type 4 x 15ml & 4 x 10ml 8 x 15ml tube
Tube size max 17 x 120mm & 13 x 110mm 17 x 120mm
Minimum speed Rpm 500 500
Maximum speed Rpm 8,000 6,000
Maximum Rcf (G) 6,100 3,500
Radius max cms 8.6 8.5
Sample tube angle ° 30 30
Acceleration time (secs) 30 30
Deceleration time (secs) 30 30
Autoclavable (frequecncy) 121°C (10) 121°C (10)

Reducers

Rotor BRK5407 / BRK5408


Part number RM05 (5ml)
Tube size max 13 x 80mm
Part number RM10 (10ml)
Tube size max 16 x 100mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 13


Rotor fitment
BRK5412. BRK5406

1
1. M6 Rotor Screw
2. M6 Washer

1 Place the rotor on 3 Place the M6


rotor fixing (item
the shaft of the
rotor. 1) into the central
hole of the rotor.

2 Place the M6 4 Hand tighten the


Washer (item 2) M6 rotor fixing.
over the centre
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

14 www.centurionscientificglobal.com
Rotor Details

Rotor BRK5412 BRK5406


Rotor type 12 x 15ml 6 x 50ml
Tube size max 17 x 120mm 30 x 120mm
Minimum speed Rpm 500 500
Maximum speed Rpm 6,000 6,000
Maximum rcf (G) 3,800 3,800
Radius max cms 9.5 9.5
Sample tube angle ° 30 30
Acceleration time (secs) 25 25
Deceleration time (secs) 25 25
Autoclavable (frequecncy) 121°C (10) 121°C (10)

Reducers

Rotor BRK5412 BRK5406


Part number RM05 (5ml) RM15 (15ml)
Tube size max 13 x 80mm 17 x 120mm
Part number RM10 (10ml) RM25 (25ml)
Tube size max 16 x 100mm 25 x 100mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 15


Rotor fitment
BRK5308. BRK5324.
BRK5100

1 1. M6 Washer
2
2. M6 Handwheel

1 Place the rotor on Place the rotor


the shaft of the 3 fixing screw
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Allen Key (Item 3).
over the centre
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

16 www.centurionscientificglobal.com
Rotor Details
Rotors 6000 RPM

Rotor BRK5324 BRK5308 BRK5100


Rotor type 24 x 15ml 8 x 50ml 6 x 100ml
Size max 17 x 120mm 30 x 120mm 45 x 125mm
Minimum speed Rpm 500 500 500
Maximum speed Rpm 6,000 6,000 6,000
Maximum Rcf (G) 4,800 4,800 4,800
Radius max cms 12 12 12
Sample tube angle ° 30 30 30
Acceleration time (secs) 35 35 35
Deceleration time (secs) 35 35 35
Autoclavable (frequency) 121°C (10) 121°C (10) 121°C (10)

Refrigerated Centrifuges Only


Minimum Temperature 4°C 4°C 4°C
At maximum speed (relative to room temperature at 23°C)

Pack of 4 Pack of 4
Reducers
Rotor BRK5324 BRK5308 BRK5100
Part number RM05 (5ml) RM15 (15ml) RL10 (10ml)
Tube size max 13 x 80mm 17 x 120mm 16 x 100mm
Part number RM10 (10ml) RM25 (25ml) RL25 (25ml)
Tube size max 16 x 100mm 25 x 100mm 25 x 100mm
RL50 (50ml)
35 x 110mm
RL85 (85ml)
39 x 110mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 17


Rotor fitment
BRK5431. BRK5434

1 1. M6 Handwheel
2
2. M6 Washer

1 Place the rotor on Place the M6


the shaft of the 3 rotor fixing knob
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten M6


washer (Item 1) rotor fixing knob.
over the centre
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

18 www.centurionscientificglobal.com
Rotor Details
Rotors 4000 RPM

Rotor BRK5431 BRK5434


Rotor type 24 x 15ml 8 x 50ml
Tube size max 17 x 120mm 30 x 120mm
Minimum speed Rpm 500 500
Maximum speed Rpm 4,000 4,000
Maximum rcf (G) 2,200 2,200
Radius max cms 12.6 12.6
Sample tube angle ° 30 30
Acceleration time (secs) 35 35
Deceleration time (secs) 35 35
Autoclavable (frequecncy) 121°C (10) 121°C (10)

Reducers

Rotor BRK5431 BRK5434


Part number RM05 (5ml) RM15 (15ml)
Tube size max 13 x 80mm 17 x 120mm
Part number RM10 (10ml) RM25 (25ml)
Tube size max 16 x 100mm 25 x 100mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 19


Rotor fitment
BRK5508, S, M, L
3

2
1. M6 Washer
1 2. Rotor Screw M6 x 20 mm
3. Torque Key

1 Place the rotor on Place the M6


the shaft of the 3 rotor fixing screw
motor. (Item 2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (item 3). Ensure
hole on the rotor. all buckets are
fully inserted
before use.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

20 www.centurionscientificglobal.com
Swing out Rotor - 8 x 5ml, 8 x 10ml & 8 x 15ml

Rotor BRK5508S BRK5508M BRK5508L


Rotor type 8 x 5ml Swing Out 8 x 10ml 8 x 15ml
Tube size max 13 x 75mm max 15 x 115mm 17 x 125mm
Minimum Speed Rpm 500 Rpm 500 Rpm 500 Rpm
Maximum Speed Rpm 4,000 Rpm 4.000 Rpm 4.000 Rpm
Maximum Rcf (G) 1500 2200 2,600
Radius max cms 8.5 12.4 14.6
Sample tube angle (°) 0 0 0
Acceleration time (secs) 20 20 20
Deceleration time (secs) 20 20 20
Autoclavable (frequency) 121°C (20) 121°C (20) 121°C (20)

Rotor Fitment and Details 21


Rotor fitment
BRK5505
3 1. M6 Washer
2. Rotor Screw
M6 x 20 mm 1 Place the rotor on
2 the shaft of the
3. Torque Key motor.

2 Place the M6 Hand tighten with


washer (Item 1) 4 Torque Key
over the centre (item 3).
hole on the rotor.

3 Place the M6 5 Ensure all four


rotor fixing screw buckets are fully
(Item 2) into the inserted before
central hole of the use.
rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

22 www.centurionscientificglobal.com
Adaptors for BRK5505
Capacity Size

Rotor Details
Part No. Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m 8x20 AM505 36
1.5ml 11x38 AM520 24
2.0ml 11x38 AM520 24
0.2ml 6x20 AM502 36
0.4ml 6x30 AM504 36
Tube type: Plain no cap Shajje: round
1ml 6x45 AR501 36
3ml 10x60 AR503 28
5ml 12x75 AR505* 28
6ml 12x82 AR505* 28
7ml 12x100 AR507* 28
9/10ml 14x100 AR509* 24
15ml 17x100 AR515* 8
25ml 24x100 AR525* 8
Rotors BRK5505 50ml 34x100 AR550* 4
100ml 45x100 AR5100* 4
Rotor type 4 x 100ml 150ml 52x100 N/A
250ml 62x100 N/A
Tube size max 45 x 100mm Tube type: Falcon with cap Shape: point
Minimum speed Rpm 500 15ml 17x120 AF515* 8
Maximum speed Rpm 4,000 50ml 29x115 AF550* 4
175ml 61x118 N/A
Maximum rcf (G) 2,200 15ml 17x120 N/A
Radius max cms 12.5 Tube type: Corning with cap* Shape: point
250ml 60x172 See K242/R
Sample tube angle ° 0 500ml 98x148 See K242/R
Acceleration time (secs) 25 Tube type: Falcon with cap Shape: square
Deceleration time (secs) 25 12ml 17x100 AFS512 8
25ml 25x90 AFS525 4
Autoclavable (frequecncy) 121°C (20) 30ml 25x110 AFS530 4
50ml 29x115 AFS550* 4
15ml 17x120 N/A N/A
Tube type: Nalgene/Oakridge Shape: round
10ml 16x80 ANO510* 8
30ml 26x95 ANO530* 4
50ml 29x107 ANO550 4
100ml 38x106 ANO585 4
Tube type: Nalgene/Oakridge* Shape: flat
250ml 62x130 See K241/R
750ml 98x153 See K243/R
Tube type: Monovette Shape: square
1.1-1.4ml 8x82 AM5014 24
Sealed lids available B5419 2.7-3ml 11x82 AM503 24
2.6-2.9ml 13x81 AM529* 24
4.5-5ml 11x108 AM503* 24
7.5-8.2ml 13x106 AM579* 20
4.5-5ml 15x92 AM550* 20
9-10ml 16x108 AM590* 20
PLEASE ENSURE THE ROTOR IS Tube type: Vacutainer Shape: round
CHECKED REGULARLY FOR CRACKS 1.6-5ml 13x75 AV516* 24
4-7ml 13x100 AV550* 16
OR DEFECTS 8.5-10ml 16x100 AV580* 12
Note: - We can not be held responsible if the tube manufacturers change the design and/or size of the tube.*
Limited capacity with sealed bucket lids (B5419 and B5319)

Rotor Fitment and Details 23


Rotor fitment
BRK5510. New BRK1000
3 1. M6 Washer
2. Rotor Screw
M6 x 20 mm 1 Place the rotor on
2 the shaft of the
3. Torque Key motor.

2 Place the M6 Hand tighten with


washer (Item 1) 4 Torque Key
over the centre (item 3).
hole on the rotor.

3 Place the M6 5 Ensure all four


rotor fixing screw buckets are fully
(Item 2) into the inserted before
central hole of the use.
rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

24 www.centurionscientificglobal.com
Adaptors for Swing out rotors
Swing Out Rotor Capacity Size
To fit buckets B5010
- 400ml max Part No.

Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m 8x20 AM405 36
1.5ml 11x38 AM420 24
2.0ml 11x38 AM420 24
0.2ml 6x20 AM402 36
0.4ml 6x30 AM404 36
Tube type: Plain no cap Shape: round
1ml 6x45 AR401 36
3ml 10x60 AR403 28
5ml 12x75 AR405* 28
6ml 12x82 AR405* 28
7ml 12x100 AR407* 28
9/10ml 14x100 AR409* 24
15ml 17x100 AR415* 8
25ml 24x100 AR425* 8
50ml 34x100 AR450* 4
BRK1000 Swing out Rotor 100ml 45x100 AR4100* 4
150ml 52x100 N/A
Shown with B5010 Buckets and 250ml 62x100 N/A
sealed lids B5419 Tube type: Falcon with cap Shape: point
15ml 17x120 AF415* 8
Rotor / buckets BRK1010 50ml 29x115 AF450* 4
175ml 61x118 N/A
Tube size max 45 x 100mm 15ml 17x120 N/A
Minimum speed 500Rpm Tube type: Corning with cap* Shape: point
Maximum speed 4000Rpm 250ml 60x172 See K242/R
500ml 98x148 See K242/R
Maximum Rcf (G) 2200
Tube type: Falcon with cap Shape: square
Radius max 12.5cm 12ml 17x100 AFS412 8
Tube angle 0 degree 25ml 25x90 AFS425 4
30ml 25x110 AFS430 4
Acceleration time 25 seconds 50ml 29x115 AFS450* 4
15ml 17x120 N/A N/A
Deceleration time 25 seconds
Tube type: Nalgene/Oakridge Shape: round
Autoclavable (times) 121°C (20) 10ml 16x80 ANO410* 8
Minimum Temperature 4°C (at 23°C 30ml 26x95 ANO430* 4
ambient) 50ml 29x107 ANO450 4
100ml 38x106 ANO485 4
Tube type: Nalgene/Oakridge* Shape: flat
B5010 bucket (set 4) REQUIRED 250ml 62x130 See K241/R
750ml 98x153 See K243/R
100ml max per bucket
Tube type: Monovette Shape: square
1.1-1.4ml 8x82 AM4014 24
2.7-3ml 11x82 AM403 24
2.6-2.9ml 13x81 AM429 24
4.5-5ml 11x108 AM403* 24
7.5-8.2ml 13x106 AM479* 24
4.5-5ml 15x92 AM450* 20
9-10ml 16x108 AM490* 20
Tube type: Vacutainer Shape: round
1.6-5ml 13x75 AV416* 24
4-7ml 13x100 AV450* 16
8.5-10ml 16x100 AV480* 12

Note: - We can not be held responsible if the tube manufacturers change the design and/or 25
size of the tube.* Limited capacity with sealed bucket lids (B5419 and B5319)
Rotor fitment
BRK5510. New BRK1000
3 1. M6 Washer
2. Rotor Screw
M6 x 20 mm 1 Place the rotor on
2 the shaft of the
3. Torque Key motor.

2 Place the M6 Hand tighten with


washer (Item 1) 4 Torque Key
over the centre (item 3).
hole on the rotor.

3 Place the M6 5 Ensure all four


rotor fixing screw buckets are fully
(Item 2) into the inserted before
central hole of the use.
rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

26 www.centurionscientificglobal.com
Adaptors for Swing out rotors
Swing Out Rotor Capacity Size
To fit buckets B5250
1 Litre Max Part No.

Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m 8x20 AM605 40
1.5ml 11x38 AM620 28
2.0ml 11x38 AM620 28
0.2ml 6x20 AM602 48
0.4ml 6x30 AM604 48
Tube type: Plain no cap Shape: round
1ml 6x45 AR601 36
3ml 10x60 AR603 28
5ml 12x75 AR605 28
6ml 12x82 AR605 28
7ml 12x100 AR607 28
9/10ml 14x100 AR609* 24
15ml 17x100 AR615* 28
25ml 24x100 AR625 8
50ml 34x100 AR650 4
BRK1000 Swing out Rotor 100ml 45x100 AR6100 4
150ml 52x100 AR601
Shown with B5250 Buckets and 250ml 62x100 BUCKET 4
sealed lids B5419 Tube type: Falcon with cap Shape: point
15ml 17x120 AF615* 16
Rotor / buckets BRK1025 50ml 29x115 AF650* 4
175ml 61x118 AF6175* 4
Tube size max 62 x 100mm 15ml 17x120 N/A 4
Minimum speed 500Rpm Tube type: Corning with cap* Shape: point
Maximum speed 4000Rpm 250ml 60x172 See K242/R
500ml 98x148 See K242/R
Maximum Rcf(G) 2600
Tube type: Falcon with cap Shape: square
Radius max 14.6cm 12ml 17x100 AFS612* 16
Tube angle 0 degree 25ml 25x90 AFS625* 8
30ml 25x110 AFS630 8
Acceleration time 25 seconds 50ml 29x115 AFS650 4
15ml 17x120 AFS614 4
Deceleration time 25 seconds
Tube type: Nalgene/Oakridge Shape: round
Autoclavable (times) 121°C (20) 10ml 16x80 ANO610 16
Minimum Temperature 4°C (at 23C 30ml 26x95 ANO630 4
ambient) 50ml 29x107 ANO650 4
100ml 38x106 ANO685 4
Tube type: Nalgene/Oakridge* Shape: flat
B5250 bucket (set 4) REQUIRED 250ml 62x130 Buckets 5250 4
750ml 98x153 See K243/R
250ml max per Bucket
Tube type: Monovette Shape: square
1.1-1.4ml 8x82 AM6014 40
2.7-3ml 11x82 AM603 40
2.6-2.9ml 13x81 AM629 40
4.5-5ml 11x108 AM603* 40
7.5-8.2ml 13x106 AM679* 40
4.5-5ml 15x92 AM650 28
9-10ml 16x108 AM690* 28
Tube type: Vacutainer Shape: round
1.6-5ml 13x75 AV616 40
Set of 4 4-7ml 13x100 AV650 40
8.5-10ml 16x100 AV680 28

Note: - We can not be held responsible if the tube manufacturers change the design and/or 27
size of the tube.* Limited capacity with sealed bucket lids (B5419 and B5319)
Rotor fitment
BRK5553. New BRK3000 &
3
BRK4000 (Same fitment)

2
1. M6 Washer
1 2. Rotor Screw M6 x 20 mm
3. Torque Key

1 Place the rotor on Place the M6


the shaft of the 3 rotor fixing screw
motor. (Item 2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (item 3). Ensure
hole on the rotor. all four buckets
are fully inserted
before use.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

28 www.centurionscientificglobal.com
Adaptors for 5520/5553
Capacity Size

Rotor Details
Part No. Tubes per
rotor
Tube type: Micro with cap Shape: point
0.5m 8x20 AM805 120
1.5ml 11x38 AM820 84
2.0ml 11x38 AM820 84
0.2ml 6x20 AM802 168
0.4ml 6x30 AM804 168
Tube type: Plain no cap Shape: round
Illustrative Purposes
1ml 6x45 AR801 168
3ml 10x60 AR803 96
Rotors BRK5553 5ml 12x75 AR805 96
6ml 12x82 AR805 96
Buckets B3000 (set 4 required) 7ml 12x100 AR807 96
9/10ml 14x100 AR809 84
Rotor type 4 x 750ml 15ml 17x100 AR815 48
Tube size max 98 x 170mm 25ml 24x100 AR825 28
50ml 34x100 AR850 16
Minimum Speed Rpm 500 100ml 45x100 AR8100 8
Maximum Speed Rpm 4,000 150ml 52x100 AR8150 4
250ml 62x100 AR8250 4
Maximum Rcf (G) 3,600
Tube type: Falcon with cap Shape: point
Radius max cms 20.5
15ml 17x120 AF815 48
Sample tube angle (°) 0 50ml 29x115 AF850 16
Acceleration time (secs) 45 175ml 61x118 AF8175 4
15ml 17x120
Deceleration time (secs) 45
Tube type: Corning with cap* Shape: point
Autoclaveable (frequency) 121°C (20)
250ml 60x172 AF8250 4
500ml 98x148 AF8500 4
Rotors BRK5520
Tube type: Falcon with cap Shape: square
Buckets B2000 (set 4 required) 12ml 17x100 AFS812 48
25ml 25x90 AFS825 28
Rotor type 4 x 500ml 30ml 25x110 AFS830 28
Tube size max 98 x 155mm 50ml 29x115 AFS850 20
Minimum Speed Rpm 500 15ml 17x120 AFS850 48
Maximum Speed Rpm 4,000 Tube type: Nalgene/Oakridge Shape: round
10ml 16x80 ANO810 48
Maximum Rcf (G) 3,400 30ml 26x95 ANO830 28
Radius max cms 19.5 50ml 29x107 ANO850 24
Sample tube angle (°) 0 100ml 38x106 ANO885 12
Tube type: Nalgene/Oakridge* Shape: flat
Acceleration time (secs) 45
250ml 62x130 ANO8250 4
Deceleration time (secs) 45 750ml 98x153 See K243/R
Autoclaveable (frequency) 121°C (20) Tube type: Monovette Shape: square
1.1-1.4ml 8x82 AM8014 96
Refrigerated Centrifuges Only 2.7-3ml 11x82 AM803 76
2.6-2.9ml 13x81 AM829 76
Minimum Temperature 4°C 4.5-5ml 11x108 AM803 76
7.5-8.2ml 13x106 AM879 76
At maximum speed (relative to room temperature 4.5-5ml 15x92 AM850 64
at 23°C) 9-10ml 16x108 AM890 64
Tube type: Vacutainer Shape: round
1.6-5ml 13x75 AV816 76
Sealed lids available B5319 4-7ml 13x100 AV850 76
8.5-10ml 16x100 AV880 64

Note: - We can not be held responsible if the tube manufacturers change the design and/or size of the tube.*
Limited capacity with sealed bucket lids (B5419 and B5319)

Rotor Fitment and Details 29


Rotor fitment
BRK5540
3

2
1. M6 Washer
1 2. Rotor Screw M6 x 20 mm
3. Torque Key

1 Place the rotor on Place the M6


the shaft of the 3 rotor fixing screw
motor. (Item 2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (item 3). Ensure
hole on the rotor. all two buckets
are fully inserted
before use.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

30 www.centurionscientificglobal.com
Microtitor Plate Rotor 4 x Standard or 2 x High Plates

Rotor BRK5540
Buckets Complete with buckets
Sealed Lids Available with
Rotor type 4 x STD Plates
Tube size max 85mm x 128mm
Minimum Speed Rpm 500 Rpm
Maximum Speed Rpm 3500 Rpm
Maximum Rcf (G) 2000
Radius max cms 14.6
Sample tube angle (°) 0°C (10)
Acceleration time (secs) 30
Deceleration time (secs) 30
Autoclavable (frequency) 121°C (20)

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 31


Rotor fitment
BRK5206. BRK5212

2 3
1. M6 Washer
2. Rotor Screw M6 x 20 mm
1 3. Torque Key

1 Place the rotor on Place the rotor


the shaft of the 3 fixing screw
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (Item 3).
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

32 www.centurionscientificglobal.com
Rotor Details
Rotors 10,000 Rpm

Rotor BRK5212 BRK5206


Rotor type 12 x 15ml 6 x 50ml
Tube size max 17 x 120mm 30 x 120mm
Minimum speed Rpm 500 500
Maximum speed Rpm 10,000 10,000
Maximum Rcf (G) 10,600 10,600
Radius max cms 9.5 9.5
Sample tube angle ° 30 30
Acceleration time (secs) 45 45
Deceleration time (secs) 45 45
Autoclavable (frequency) 121°C (20) 121°C (20)

Refrigerated Centrifuges Only


Minimum Temperature 4°C 4°C
At maximum speed (relative to room temperature at 23°C)

Reducers
(Pack of 4)

Rotor BRK5212 BRK5206


Part number RM05 (5ml) RM15(15ml)
Tube size max 13 x 80mm 17 x 120mm
Part number RM10 (10ml) RM25 (25ml)
Tube size max 16 x 100mm 25 x 100mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 33


Rotor fitment
BRK5210. BRK5208.
BRK5224

2 3
1. M6 Washer
2. Rotor Screw M6 x 20 mm
1 3. Torque Key

1 Place the rotor on 3 Place the rotor


fixing screw
the shaft of the
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (Item 3).
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

34 www.centurionscientificglobal.com
Rotor Details
Rotors 6,000 Rpm

Rotor BRK5224 BRK5208 BRK5210


Rotor type 24 x 15ml 8 x 50ml 6 x 100ml
Size max 17 x 120mm 30 x 120mm 45 x 125mm
Minimum speed Rpm 500 500 500
Maximum speed Rpm 10,000 10,000 10,000
Maximum Rcf (G) 13,400 13,400 13,400
Radius max cms 12 12 12
Sample tube angle ° 30 30 30
Acceleration time (secs) 50 50 50
Deceleration time (secs) 50 50 50
Autoclavable (frequency) 121°C (20) 121°C (20) 121°C (20)

Refrigerated Centrifuges Only


Minimum Temperature 4°C 4°C 4°C
At maximum speed (relative to room temperature at 23°C)

Pack of 4 Pack of 4 Pack of 6

Reducers
Rotor BRK5224 BRK5208 BRK5210
Part number RM05 (5ml) RM15 (15ml) RL10 (10ml)
Tube size max 13 x 80mm 17 x 120mm 16 x 100mm
Part number RM10 (10ml) RM25 (25ml) RL25 (25ml)
Tube size max 13 x 100mm 25 x 100mm 25 x 100mm
RL50 (50ml)
35 x 110mm
RL85 (85ml)
39 x 110mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 35


Rotor fitment
BRK5256 (Now in silver)

1. M6 Washer
2. Rotor Screw M6 x 20 mm
3. Torque Key

1 Place the rotor on Place the rotor


the shaft of the 3 fixing screw
motor. (Item2) into the
central hole of the
rotor.

2 Place the M6 4 Hand tighten with


washer (Item 1) Torque Key
over the centre (Item 3).
hole on the rotor.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

36 www.centurionscientificglobal.com
Rotor Details
Rotors 10,000 RPM

Rotor BRK5256
Rotor type 6 x 250ml
Size max 62 x 130mm
Minimum speed Rpm 500
Maximum speed Rpm 10,000
Maximum Rcf (G) 16,100
Radius max cms 14.4
Sample tybe angle ° 30
Acceleration time (secs) 60
Deceleration time (secs) 85
Autoclavable (frequecncy) 121°C (20)

Refrigerated Centrifuges Only


Minimum Temperature 4°C
At maximum speed (relative to room temperature at 23°C)

Reducers
Rotor BRK5256
Part number RX10 (10ml) RX25 (25ml) RX50 (50ml)
Tube size max 16 x 100mm 25 x 100mm 35 x 110mm

Part number RX85 (85ml) RX100 (100ml) RX175 (175ml)


Tube size max 39 x 110mm 48 x 110mm 62 x 121mm

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 37


Rotor fitment
4420. 4430. 4460
1. M6 Washer
2. Rotor Screw
M6 x 20 mm 1 Place the rotor on
the shaft of the
3. Torque Key motor
1
2 3

2 Place the M6 Hand tighten with


washer (Item 1) 4 Torque Key (item
over the centre 3)
hole on the rotor.

3 Place the M6 5 Ensure lid is fully


rotor fixing screw tightened before
(Item 2) into the
central hole of the use
rotor

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

38 www.centurionscientificglobal.com
Rotor Details

BRK4420 Illustrated

Rotor 4420 (4 Place) 4430 (8 Place) 4430 (12 Place)


Rotor type 4 x 0.2 to 6ml 8 x 0.2 to 6ml 12 x 0.2 to 6ml
Tube size max Single or double Single or double Single or double
Minimum Speed Rpm 200 200 200
Maximum Speed Rpm 2,000 2,000 2,000
Maximum Rcf (G) 500 500 500
Radius max cms 12 12 12
Sample tube angle (°) 0 0 0
Acceleration time (secs) 25 25 25
Deceleration time (secs) 25 25 25
Autoclaveable (frequency) 121°C (10) 121°C (10) 121°C (10)

Cytology rotor includes Rotor, Sealed Lid, Clips, 24 holders with filter card & 100 slides

Accessories

4446 4444 4600 4462


Double sample holder Single sample holder Double sample holder Stainless steel clips
with card (up to 1ml) with card (up to 1ml) with card (up to 6ml)

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 39


Rotor fitment
5450. Cell washer rotor

Please ensure that the 1


rotor Shaft is Clear of
any debris.

2 Place the rotor on the shaft.

3 Turn the knob on the rotor until


hand tight.

Make sure all the


4
adaptors are fully
inserted before use.

PLEASE ENSURE THAT THE FIXING IS CHECKED AND HAND TIGHTENED


ON A REGULAR BASIS
Failure to follow these instructions could cause serious damage to the centrifuge and
possible injury to persons nearby and would make the warranty invalid

40 www.centurionscientificglobal.com
Rotor Details
5450. Cell washer rotor
Speed max 500-3500 Rpm
RCF max 1300 G

• High quality Polypropylene


• Autoclaveable
The rotor in this device can be
detached And efficiently function
as a holder in Which test tubes can
be fitted, Centrifuges, transported,
incubated and Emptied without the
necessity for Individual handling.

2 sets of adaptors supplied,


12 of each size

Our adaptors enable tohe rotor to


handle both 10 x 75mm and
12 x 75mm tubes.
This included rotor carries 12 tubes
but can operate with any balanced
load of fewer tubes and any
combination of sizes as long as these
are properly balanced.
This offers comlete flexibility for the
user.

PLEASE ENSURE THE ROTOR IS CHECKED REGULARLY FOR CRACKS OR DEFECTS

Rotor Fitment and Details 41


Technical Support
Caring for your Centrifuge & Buckets
Always keep buckets & adapters clean and free from
! moisture at all times, this will help prevent corrosion.

Use only official Centurion Scientific accessories with


your product Always ensure samples are distributed in a
! diametric pattern to ensure the machine runs balanced.

Rotor pins should be kept clean and lubricated on a regular


! basis.

It is recommended that your centrifuge is switched off and


the lid left open overnight to ensure that any moisture build
! up can evaporate.

Always ensure that lids and accessories are correctly


! tightened before running the unit.

Ensure that all buckets are present in the rotor when


running your centrifuge, as using the unit with an
! unbalanced rotor can cause distortion and excessive wear.

Do not use any kind of scourers or abrasive material to


clean the buckets and ensure that a pH neutral cleaner is
! always used.

Aluminium buckets should be kept clean and moisture free


! to keep them in a good, safe operating condition.

It is recommended that regular maintenance and cleaning


! of the unit and accessories is carried out.

42 www.centurionscientificglobal.com
Technical Support
Nomograph for RCF Determination

Example
To find the relative centrifugal
force at a radial distance of 10
cm from the center of rotation
when operating the centrifuge
at a speed of 3000 r.p.m., place
a straightedge on the chart
connecting the 10cm point on the
Rotating Radius Scale (A) with the
3000 r.p.m. point on the Speed
Scale (B).
Read the point at which the
straightedge intersects the
Relative Centrifugal Force Scale
(C) – in this case, 1000 x gravity.
Similarly, if the desired “r.c.f.” is
known, the necessary speed for
a given rotating radius may be
determined by connecting the
two known points and reading the
intersection of the straightedge
with the Speed Scale.

Equation for Calculating R.C.F.


R.C.F. = .00001118 x r x N2
R.C.F. = relative centrifugal force (gravities)
r = radius from center of rotation to bottom of tube (cm)
N = rotating speed (rev. per minute)

Rotor Fitment and Details 43


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

CHEMICAL
2-Butanol S U S S
2-Butyl Alcohol S U S S
2-mercaptoethanol S S S S S S S S S
2-Propanol / U S S
3-Pentanone U M S
Acetaldehyde S / / / U M S U /
Acetic Acid (10%) M S S
Acetic Acid (5%) S M S M S S S
Acetic Acid (5%) S S M S S S S S M
Acetic Acid (60%) S U S U S S M
Acetic Acid (60%) S S U S U M S M U
Acetic Acid (Glacial) S S U S U U S U U
Acetic Anhydride / / U M S
Acetone M S M S U S S M M
Acetonitrile S S S S U M S S S
Acetophenone / U M S
Acetylene / U M S
Alconox™ U U S S M S S S S
Allyl Alcohol / / S / S S S / /
Aluminium Chloride U U S S S S U
Aluminium Fluoride C S S S
Aluminium Potassium Sulphate C S S S
Aluminium Sulphate / M S S
Aluminum Chloride U U U S S S S M U
Ammonia U S S
Ammonium Acetate S S S S S S S S S
Ammonium Carbonate M S S S U S S S M
Ammonium Carbonate, saturated U S S
Ammonium Chloride / M S S
Ammonium Chloride, saturated U M S S
Ammonium Hydroxide (10%) U U M S U S S S S
Ammonium Hydroxide (28%) U U M S U S S S S
Ammonium Hydroxide (conc.) U U M / U S S S S
Ammonium Nitrate / M S S

44 www.centurionscientificglobal.com
Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Ammonium Persulphate / M S S
Ammonium Phosphate U / S S S S S S M
Ammonium Sulphate U M U S S S S S U
Ammonium Sulphate (77%) M S S
Ammonium Sulphate (sat.) U S S
Ammonium Sulphide U S S
Ammonium Sulphide, sat. U S S
Amyl Acetate / U U S
Amyl Alcohol S / S / S M / U /
Aniline S S S S U M S S S
Aqua Regia U / U / U U / / /
Barium Salts M U S S S S S S M
Benzaldehyde / / U M S
Benzenamine / U M S
Benzene S S M S U U S U U
Benzoic Acid / / S S S
Benzyl Alcohol / M / U U S /
Beta-Mercaptoethanol S S S
Boric Acid U S U S S S S S S
Butane / U S S
Butyl Acetate / U S U
Calcium Bisulphate / U S S
Calcium Chloride M U S S M S S S M
Calcium Hypochlorite M / M / M S S M U
Calcium Salts U S S
Carbon Dioxide / S S S
Carbon Disulphide / U M S
Carbon Tetrachloride U M S U S S M
Cesium Acetate M / S S S S S S M
Cesium Bromide M S S S S S S S M
Cesium Chloride M S S S S S S S M
Cesium Formate M S S S S S S S M
Cesium Iodide M S S S S S S S M
Cesium Sulphate M S S S S S S S M
Chlorine, dry gas / M U S
Chlorine, water solution / / M M S

Rotor Fitment and Details 45


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Chlorine, wet gas / M M S


Chlorobenzene / / U U S
Chloroethane U U S
Chloroform U U M S U M S U U
Chlorosulphonic Acid U U S
Chromic Acid (10%) U S S M U S
Chromic Acid (30%) / M S S
Chromic Acid (50%) / U / U S S /
Chromic Acid (80%) U U S
Chromic Acid -10% U / U S M S S M U
Chromic Acid -50% U / U / M S S / U
Citric Acid / M / M S S /
Citric Acid (10%) M S M S S S S S S
Copper Salts S S S
Copper Sulphate / / S S S
Cresol Mixture S S S S U U S S S
Culture Media S S S
Cyclohexane S S S S U U S U M
Cyclohexanone / / U M S
Decalin / U U S
Deoxycholate S S S S S S S S S
Dextran M S S S S S S S M
Dextran (Sulphate) S S S
Diacetone / U S S
Dibutyl Phthalate S / U M S
Dichloroethylene / U U S
Diethyl Ether S S S S U U S S S
Diethyl Ketone S / M S U M S / /
Diethylpyro-carbonate S S S S U S S S S
Dimethylsulfoxide S S S S U S S S S
Dioxane M S M S U M S S S
Distilled Water S S S S S S S S S
Ethanol (10%) M S S
Ethanol (100%) U S S
Ethanol (5%) M S S
Ethanol (50%) S U S S

46 www.centurionscientificglobal.com
Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Ethanol (70%) U S S
Ethanol (95%) S U S S
Ethanol (96%) U S S
Ethanolamine / U S S
Ethyl Acetate M M M S U S S M M
Ethyl Alcohol (10%) M S S
Ethyl Alcohol (100%) U S S
Ethyl Alcohol (5%) M S S
Ethyl Alcohol (50%) S S M S U S S S M
Ethyl Alcohol (50%) U S S
Ethyl Alcohol (70%) U S S
Ethyl Alcohol (95%) S S M S U S S S U
Ethyl Alcohol (95%) / M / U S S /
Ethyl Alcohol (96%) / M / U S S /
Ethyl Chloride / U U S
Ethylene Dichloride S / S / U U S U /
Ethylene Glycol S S S S U S S S M
Ethylene Oxide (100%) / U U S
Ethylene Oxide Gas / M S S
Ethylene Oxide Vapour S / / S M S S U S
Fatty Acids, saturated S U S S
Fatty Acids, unsaturated U U S S
Ferric Chloride U U M / / S / M U
Ferric Sulphate / S S S
Ferrous Chloride / M S S
Ferrous Sulphate / S S S
Ficoll-Hypaque™ M S S S S S S S M
Fluoboric Acid / / S S
Formaldehyde (40%) / S / M S S /
Formaldehyde (5%) M S S
Formaldehyde -40% M M S S S S S S M
Formalin M M S
Formalin (40%) / / M M S
Formalin (5%) M S S
Formic Acid / / U S S
Formic Acid (10%) / M S S

Rotor Fitment and Details 47


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Formic Acid (100%) / U / U S S /


Formic Acid (3%) / M S S
Formic Acid -100% / S U / M S S / U
Fuel Oil / / M M S
Furfural / U U S
Gallic Acid U S S
Gasoline / U U S
Gluanidine Hydrochloric / S / S S S /
Glutaraldehyde S S S S S S / S S
Glycerine / / S S S
Glycerol M S S S S S S S S
Guanidine Hydrochloride U U S S S S S S U
Haemo-Sol™ S S S S S S S S S
Hexane S S S S U S S U S
Hydrochloric Acid (10%) U M S S
Hydrochloric Acid (30%) U M S
Hydrochloric Acid (37%) C U M S
Hydrochloric Acid (5%) / S S S
Hydrochloric Acid (50%) / U / U S S /
Hydrochloric Acid (conc.) U U U / U S S U U
Hydrochloric Acid -10% U U U S U S S S U
Hydrochloric Acid -50% U U U S U S S M U
Hydrocyanic Acid / S S S
Hydrofluoric Acid (10%) / U / M S S /
Hydrofluoric Acid (100%) U M S
Hydrofluoric Acid (38%) U S S
Hydrofluoric Acid (50%) / U / U S S /
Hydrofluoric Acid (60%) U M S
Hydrofluoric Acid (70%) U M S
Hydrofluoric Acid -10% U U U / M S S U U
Hydrofluoric Acid -50% U U U / U S S U U
Hydrofluoric Acid conc. U M S
Hydrogen Peroxide (10%) / U / S S S /
Hydrogen Peroxide (100%) / U U S
Hydrogen Peroxide (3%) / S / S S S /
Hydrogen Peroxide (30%) / / S S S

48 www.centurionscientificglobal.com
Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Hydrogen Peroxide (conc.) U U S


Hydrogen Peroxide -3% S M S S S S S S S
Hydrogen Peroxide-10% U U U S S S S S M
Hydrogen Sulphide, dry gas M S S
Hydrogen Sulphide, wet gas C M S S
Hydroquinone / / S S
Idoacetic Acid / S / S S S /
Iodine Solution M S S
Iodoacetic Acid S S S S S S S M S
Iso-Butanol U S S
Isobutyl Alcohol / / S / S S S S /
Iso-Butyl Alcohol (10%) U S S
Iso-Octane U U S
Iso-Propanol (100%) S / U S S
Isopropyl Alcohol M M S S U S S S M
Isopropyl Alcohol (100) U S S
Kerosene S S S S M M S U S
Lacquer Thinner U / U U S
Lactic Acid (100%) / / / / S S S M S
Lactic Acid (20%) / / / / S S S M S
Lactic Acid (50%) M S S
Lauryl Alcohol / S S
Lead Acetate / / M S S
Magnesium Chloride M S S S S S S S M
Magnesium Hydroxide / U S S
Magnesium Hydroxide, saturated U S S
Magnesium Sulphate / S S S
Maleic Acid / M S S
Manganese Salts / / S
Mercaptoacetic Acid U S S S U U S U S
Mercuric Chloride / S S S
Mercury / U S S
Mercury Salts S S S
Methanol S U S S
Methanol (10%) M S S
Methanol (100%) U S S

Rotor Fitment and Details 49


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Methyl Alcohol S S M S U S S S M
Methyl Alcohol (10%) / M S S
Methyl Alcohol (100%) / U S S
Methyl Ethyl Ketone S S M S U S S S S
Methyl Isobutyl Ketone / U S S
Methylene Chloride U U S S U U S S M
Metrizamide™ M S S S S S S S M
Milk / S S S
N, N-Dimethyl- formamide S S S S U S S M S
N/Butyl Alcohol S / S / M S S M /
N/Butyl Phthalate S S S S U U S M M
Naphthalene (100%) / U M S
N-Butyl Alcohol / S / U S S /
N-Butyl Phthalate S S S U M
Nickel Chloride S S S
Nickel Salts U S / S S S S S M
Nickel Sulphate S S S
Nitric Acid (10%) U S U S S S S M S
Nitric Acid (30%) M S S
Nitric Acid (50%) U S U S M M S U S
Nitric Acid (60%) U U S
Nitric Acid (70%) / U U S
Nitric Acid (95%) U / U / U M S U S
Nitric Acid conc. / U U S
Nitric Acid, fuming U U S
Nitrobenzene C / U M S
N-N-Dimethylformamide / S / U S S /
n-Propanol U S S
Oil (Cottons/d) M S S
Oil (Lins/d) / U M S
Oil (Mineral) / / U U S
Oil (Paraffin) U U S
Oil (Silicone) / M M S
Oil (Vegetable) M S S
Oils (Other) S / S S S S S / S
Oils (Petroleum) S S S S M U S U S

50 www.centurionscientificglobal.com
Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Oleic Acid S / U S S S S M U
Oxalic Acid U U U S U S S S U
Oxalic Acid (50%) S S S
Oxygen S S S
Palmitic Acid / S S S
Paraffin / M M S
Paraffin, White M M S
Perchloric Acid (10%) / U / U S S /
Perchloric Acid (70%) / U / U M S /
Perchloric Acid -10% U / U S U M S U /
Perchloric Acid -70% U U U S U M S U U
Percholric Acid (conc.) U M S
Petrol / / U U S
Petroleum Ether U U S
Phenol U M S
Phenol (5%) U S M S U S S U M
Phenol (50%) U S M S U M S U U
Phenol, liquefied (92%) U U S
Phosphate Trichloride / U U S
Phosphoric Acid (10%) / U / S S S /
Phosphoric Acid (25%) M S S
Phosphoric Acid (30%) U S S
Phosphoric Acid (50%) U / U S S
Phosphoric Acid (85%) / U S S
Phosphoric Acid (95%) U S S
Phosphoric Acid (conc.) U U U / M M S U M
Phosphoric Acid -10% U U U S S S S U M
Physiologic Media (Serum, Urine) M S S S S S S S S
Picric Acid S S S S S S S U M
Potassium Acetate S S S
Potassium Bromide U S S S S S S S M
Potassium Carbonate M U S S U S S S S
Potassium Carbonate, saturated U S S
Potassium Chloride U S S S S S S S U
Potassium Hydroxide (10%) U S S
Potassium Hydroxide (45%) U U S S

Rotor Fitment and Details 51


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Potassium Hydroxide (5%) U U M S U S S M U


Potassium Hydroxide (conc.) U U M M U M U / U
Potassium Iodide / S S
Potassium Permanganate S S S S S M S S M
Potassium Sulphate / S S S
Propane U U S
Pyridine / / U M S
Pyridine (5%) U M S
Pyridine (50%) U S U U U S S S U
Rubidium Bromide M S S S S S S S M
Rubidium Chloride M S S S S S S S M
Sera S S S
Silver Cyanide S S S
Silver Nitrate S / S S S
Sodium Bicarbonate / S S S
Sodium Bisulphate / S S S
Sodium Borate M S S S S S S S M
Sodium Bromide U S S S S S S S M
Sodium Carbonate S / M S S
Sodium Carbonate (2%) / S / M S S /
Sodium Carbonate -2% M U S S U S S S S
Sodium Chlorate / S S S
Sodium Chloride (10%) / S / S S S /
Sodium Chloride (sat’d) U / S / S S / S S
Sodium Chloride -10% S / S / S S / S S
Sodium Cyanide / / S S
Sodium Dichromate / M S S
Sodium Dodecyl Sulphate S S S S S S S S S
Sodium Hydroxide (<1%) / / SS M S S /
Sodium Hydroxide (1%) U U S S
Sodium Hydroxide (10%) / U S M S S /
Sodium Hydroxide (conc.) U S S
Sodium Hydroxide -10% U / U M U S S M S
Sodium Hypochlorite M M S
Sodium Hypochlorite (5%) U U U S S M S M U

52 www.centurionscientificglobal.com
Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Sodium Iodide M S S S S S S S M
Sodium Metaborate U S S
Sodium Nitrate S S S S S S S U S
Sodium Nitrate (10%) S S S
Sodium Perborate / / M S
Sodium Peroxide / / S M S
Sodium Phosphate M S S
Sodium Silicate C U S S
Sodium Sulfate U S S S S S S S M
Sodium Sulfide S / / / U S / S S
Sodium Sulfite S S S M S S S S S
Sodium Sulphate / S / S S S /
Sodium Sulphite / S / S S S /
Sodium Thiosulphate C M S S
Solution 555 (20%) S S S S S S S S S
Stearic Acid S / S S S S S M M
Sucrose M S S S S S S S S
Sucrose / S / S /
Sucrose, Alkaline M S S S U S S S M
Sulfosalicylic Acid U U S S S S S S U
Sulfuric Acid (10%) M U U S S S S U U
Sulfuric Acid (50%) M U U S U S S U U
Sulfuric Acid (conc.) M U U / U S S U U
Sulphosalicyclic Acid / S / S S S /
Sulphur Chloride / U U S
Sulphur Dioxide / M S S
Sulphuric Acid (10%) / U / M S S /
Sulphuric Acid (50%) / U / U S S /
Sulphuric Acid (60%) / / U U S
Sulphuric Acid (75%) / U M S
Sulphuric Acid (conc.) / U / U S S /
Sulphuric Acid, fuming U U S
Tannic Acid / S S S
Tartaric Acid / / S S S
Tetrachloride U U M S U M M M M
Tetrahydrofuran S S U S U U S U S

Rotor Fitment and Details 53


Chemical resistance table

ANODIC COATING

SILICONE RUBBER
POLYCARBONATE

POLYPROPYLENE
for ALUMINIUM

STAINLESS ST/L
ALUMINUM
Key

TEFLON™
DELRIN™
S - Satisfactory

GLASS
M - Moderate attack
U - Unsatisfactory

Tetralin U U S
Toluene S S M S U U S U S
Trichloroacetic Acid U U U S M S S U U
Trichloroacetic Acid (10%) U S S S
Trichloroethane S / M / U U S U /
Trichloroethylene / / / / U U S U /
Triethanolamine U M S
Tris Buffer (neutralpH) U S S S S S S S S
Trisodium Phosphate / / M / / S S / /
Triton X/100™ S S S S S S S S S
Turpentine / / U U S
Urea S / S / M S S S M
Urine / / S S S
White Spirits U U S
Xylene S S M S U U S U M
Zephiran Chloride (1%) S S S
Zephiran Chloride (7%) M S S
Zinc Chloride U U U S S S S S U
Zinc Sulphate U S S S S S S S S

54 www.centurionscientificglobal.com
Head Office Factory 1
The Old Stables (Ambient Manufacturing)
Church Farm The Old Grain Barn, Church Farm,
Stoughton Stoughton, Chichester, PO18 9JL.
Chichester
Factory 2
West Sussex
PO18 9JL (Refrigeration Manufacturing)
United Kingdom. The Old Drier Barn, Church Farm,
Stoughton, Chichester, PO18 9JL.
Office 2 (Research & Service) Factory 3 (Logistics)
The Old Equine Barn, Church Farm, The Old Storage Barn, Church Farm,
Stoughton, Chichester. PO18 9JL. Stoughton, Chichester, PO18 9JL.

Call us on: Tel +44(0)2392 631225


or email
General: info@centurionscientific.co.uk
Sales: sales@centurionscientific.co.uk

www.centurionscientificglobal.com
Follow Centurion Scientific on social media

We reserve the right to make changes with ongoing research and development

55
EC Declaration of Conformity
Declares that product: PrO-ASTM, PrO-Analytical, PrO-Research, Pro-Hospital, PrO-Ed, PrO-Vet,
PrO-Xtract, PrO-Road, PrO-PRP Centrifuge Series
Product options: Rotor plus Accessories

This instrument is fully guaranteed against manufacturing defect for a period of three years from
date of purchase.
Service and technical manual can be obtained via our website and address below.

Conform to the following Product Specifications:


IEC 1010-1, 1010-2 020, IEC 1010-2-0202 020, 7.2.101
EN 61010, EN 61010-1, 7.5 expelled parts, BS 5724, BS 7687:2.20
BS EN 61010-2-020 1195, 73/23/EEC Low voltage directive
EMC EU 89 336 (INCLUSIVE 93/68/EEC)
Supplementary Information CE Marking
The product herewith complies with the requirements of the following Directives and carries the
CE mark accordingly.
The Low voltage Directive 73.23/EEC
The EMC Directive 89/336/EEC (inclusive 93/68/EEC)
This product was tested in a typical configuration with Centurion Scientific Ltd and other test
House Facilities.
Quality Control. ISO 90001:2008
internal systems for testing sub assemblies (5) and final product.
IVD 98/79CE - European Directive.
Self certified group C.
IVD 9879EC
I hereby declare that the equipment named above has been designed to comply with the relevant
sections of the above referenced specifications. The unit complies with all the applicable essential
requirements of directives.

Certified by: K. Cooper QA Manager

Centurion Scientific Limited

The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
Certificate of Origin
Centurion Scientific Ltd of The Old Stables, Church Farm,
Stoughton, Chichester, West Sussex, PO18 9JL,
UNITED KINGDOM, hereby confirm that all centrifuges and
accessories supplied by us are of UNITED KINGDOM origin.
Signed for and on behalf of Centurion Scientific Ltd
Granted 2016

Company Representative

SC I ENTI FI
ON C
RI
LI
U

MI
T
• CEN

TED

CertifiCate
• CEN

Of Origin
TED
MI

TU
LI

IO
IC NS
C I ENT I F

Centurion Scientific Limited

The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
Centurion Scientific Limited

58 The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
www.centurionscientificglobal.com
Rotor Fitment and Details 59
To review all of our ranges
please visit:
www.centurionscientificglobal.com