In DigitexPRO (DAR-3000 1M CCD), use “Set_net.exe” to set Destination Host for DICOM
Image Transfer (option), and “Set_id.exe” to register the serial number of DAR. When
entering the setting data and pressing the button “to DAR”, these setting tools transfer the
file including the set data from PC-Gateway to DAR main body through SCSI Bus.
However, the users reported that some systems were not able to transfer the file properly
by using “to DAR” button.
If proper transfer of the file fails, store the set files described below in a floppy without
pressing the button “to DAR”. Then, conduct serial transfer of set data to DAR main body
by using laptop PC.