TSX-D02029

Improper setting of DICOM Transfer Image for DigitexPRO

In DigitexPRO (DAR-3000 1M CCD), use “Set_net.exe” to set Destination Host for DICOM
Image Transfer (option), and “Set_id.exe” to register the serial number of DAR. When
entering the setting data and pressing the button “to DAR”, these setting tools transfer the
file including the set data from PC-Gateway to DAR main body through SCSI Bus.
However, the users reported that some systems were not able to transfer the file properly
by using “to DAR” button.
If proper transfer of the file fails, store the set files described below in a floppy without
pressing the button “to DAR”. Then, conduct serial transfer of set data to DAR main body
by using laptop PC.

Name of Set File Location of File before Program to be Item to be set


Transfer created
HOSTLST.DCM C:\Option\bin\ConfigDAR Set_net.exe Destination Host List
DAR_AET.DCM C:\Option\bin\ConfigDAR Set_net.exe DAR AE Title
DAR_ID.DCM C:\Option\bin\ConfigDAR Set_ID.exe Model Number
* Transfer only files in which some changes are added. You do not need to transfer all files.
Carry out serial transfer of all files only when you confuse files including some changes
with others.
* Use capital letters to name the file transferred to destination host as descried in the list.

1
Fig.1 File for Destination Host List / DAR AE Title Setting Created by Set_set.exe

Fig.2 File for registering serial number created by Set_ID.exe