You are on page 1of 2

URIT

URIT-910
Electrolyte Analyzer

Electrolyte Analyzer Series


URIT -91 OA K/Na/CI
URIT -91 OB K/Na/CI/TC02
.URIT -91 OC K/Na/Cl/iCa/nCa/TCa/pH
URIT -910 K/Na/Cl/iCa/nCa/TCa/pH/TCO/AG

..... .... • •
la URrT-91 0 .... •

I
---

Global Diagnostics Supplier

c ./
URIT -910
Electrolyte Analyzer
11 Specifications I

Analysis Method: ion selective electrode (ISE)

Parameters: K, Na, Cl, iCa, nCa, tCa, pH, TC02, AG

Test Speed: 45 senconds/sample

Sample: serum, plasma, cerebrospinal fluid and

diluted urine
Calibration: Self-calibrating

Storage:1000test results (can be extended to more than 10000)


----
Printer: Built-in printer

Ambient: 5-35 DC

Power: 220V± 10V, 50Hz-60Hz,can transfer to 11OV,50-60Hz

Interface:Serial port(RS232)

11 Range and Precision I


Parameter Range of Measurement Resolution Precision(CV% )

K+ 0.5-15.0mmoI/L 0.01 mmol/L ~1.0%


+ ~1.0%
Na 30-200mmo1/L 0.1mmol/L
C1- 30-200mmo1/L 0.1 mmol/L ~1.0%
Ca++ 0.5-5.0mmoI/L 0.01 mmol/L ~2.0%
PH 6-9pH 0.01 pH ~3.0%
TC02 2.0-70.0mmoI/L 0.1mmol/L ~4.0%

11 Features I

,.. High-accurate and long life electrode and TC02 sensor

,.. Simple YES/NO prompted operation

,.. Auto-prompting and eliminating malfunction

,.. Power failure protection to avoid data losing

,.. 4 types selection for requirements of different customers


,.. Individual reagent to avoid waste

URIT MEDICAL ELECTRONIC (GROUP) CO., LTD.


Add:No.4 EAST ALLEY, JIUHUA ROAD, GUILlN, GUANGXI 541001, P.R.CHINA
Tel:+86 773 2288582/2288583
E-mall:export@uritesLcom
Fax:+86 773 2288560/2288559
Http://www.uritesLcom
(€ 1~1
15013485
Rev: 201004-5(20100412)