Professional Documents
Culture Documents
SLE6000 Datasheet - V25 Specifications
SLE6000 Datasheet - V25 Specifications
(V2.0.50 software)
u CPAP (Dual limb, ET) u CMV (Dual limb, ET) u nCPAP D (Dual limb)
for passive nCPAP interfaces e.g. SLE Miniflow
Inspiratory Time (Ti) 0.1 to 3.0 s Respiratory Rate (RR) 1 to 150 BPM
CPAP 0 to 35 mbar Inspiratory Time (Ti) 0.1 to 3.0 s Inspiratory Time (Ti) 0.1 to 3.0 s
u HFOV (Dual limb, ET) u nHFOV D (Dual limb) u NIPPV Triggered (Dual limb)
Frequency 3 to 20 Hz for passive nCPAP interfaces e.g. SLE Miniflow for passive nCPAP interfaces e.g. SLE Miniflow
I:E Ratio 1:1 / 1:2 / 1:3 Frequency 3 to 20 Hz Respiratory Rate (RR) 1 to 150 BPM
MAP 0 to 45 mbar I:E Ratio 1:1 / 1:2 / 1:3 Inspiratory Time (Ti) 0.1 to 3.0 s
Delta P 4 to 180 mbar Mean Airway Pressure 0 to 45 mbar PEEP 0 to 35 mbar
VTV 2 to 50 ml † Delta P 4 to 180 mbar PIP 0 to 65 mbar
O2 Concentration 21 to 100% O2 Concentration 21 to 100% O2 Concentration 21 to 100%
w Additional Parameters w Additional Parameters
w Additional Parameters
Sigh RR 1 to 150 BPM Rise Time 0 to 3.0 s
Sigh RR 1 to 150 BPM
Sigh Ti 0.1 to 3.0 s Trigger Sensitivity 1 to 100%
Sigh Ti 0.1 to 3.0 s
Sigh P 0 to 45 mbar
Sigh P 0 to 45 mbar u DuoPAP (Single limb)
u HFOV+CMV (Dual limb, ET) Respiratory Rate (RR) 1 to 60 BPM
u nCPAP S (Single limb)
for active (fluidic-flip) nCPAP interfaces (e.g. Inspiratory Time (Ti) 0.1 to 3.0 s
Respiratory Rate (RR) 1 to 150 BPM
SLE1000 generator or Infant Flow) PEEP 2 to 15 mbar
Inspiratory Time (Ti) 0.1 to 3.0 s
Inspiratory Time (Ti) 0.1 to 3.0 s PIP 2 to 25 mbar
Frequency 3 to 20 Hz
CPAP 2 to 15 mbar O2 Concentration 21 to 100%
PEEP 0 to 35 mbar
PIP 2 to 25 mbar
PIP 0 to 65 mbar
O2 Concentration 21 to 100%
Delta P 4 to 180 mbar O2 Therapy
w Additional Parameters
O2 Concentration 21 to 100%
RR Backup 1 to 10 BPM u High Flow Oxygen Therapy
w Additional Parameters
Trigger Sensitivity 1 to 100% (Single limb)
Oscillation on both high
HFO Waveform and low cycles or oscillation Flow Rate 2 to 30 l/min
on low cycle only. O2 Concentration 21 to 100%
Oscillation pause 60 s
†
VTV control, when enabled, becomes Vte Target control. Page 1 of 2
Optional Module features
Misc. Specifications
w Flow Sensor w Power AC w Connectors (Rear mounted)
10 mm dual-hot-wire Mains voltage 100-240V / 50-60Hz RS232 & USB data ports
Flow sensor type:
anemometer. (Single-use Power 115 VA VGA (Video out) port
(Electrically isolated)
or autoclavable versions).
T2.5AH 250V USB Power port for nebuliser
Applied part Type BF Fuses (x2)
(5x20 mm) Nurse Call
Flow rate 0.2 to 30 l/min Typical 3+ hour battery 24V DC input
Accuracy ±8% maximum Battery back-up life (in all modes) in
normal use SpO2 & etCO2
Dead space 1 ml
Weight 10 g Full charge: 18 hours RJ45 Ethernet networking port
Battery charging
80% charge: 8 hours
w Flow w IP Rating
w Power DC
Flow rate 0 to 99 l/min Type of protection
IP21
Voltage 24V 4A against ingress of water
w Volume
w Operating Environment w Environmental Storage
Expiratory tidal Conditions
0 to 999 ml Temperature +10°C to +40°C
volume
10 to 90% Ambient Temperature -20°C to +50°C
Expiratory minute Relative Humidity
0 to 18 L (non-condensing) 10% to 90%
volume Relative Humidity
non-condensing
w Measured Parameters w Dimensions
w 330 mm x w Sound levels
Leak 0 to 99%
Size, ventilator only h 369 mm x Sound pressure level 49 dBA
Respiratory rate 0 to 999 BPM d 548 mm
Compliance 0 to 99.9 ml/mbar Height on pole 1310 mm
C20/C 0 to 9999 Weight (Ventilator only) ≤ 22 kg
Resistance 0 to 999 mbar/(l/s)
w Pneumatic Connectors
Inspiratory time 0 to 9.99 s
15 mm F / 22 mm M
Expiratory time 0 to 9.99 s Exhalation port
conical (ISO5356-1)
Vmin 0 to 99.99 l Proximal airway 5 mm non-conical
Trigger resolution 1 15 mm M conical
Fresh gas port
Vte 0 to 99.9 ml (ISO5356-1)
Nebulizer port (on rear) 5 mm non-conical
DC02 0 to 9999
w Classification (Electrical)
I:E Ratio 1:9.9 to 9.9:1
Type of protection Class 1
Oxygen concentration 0 to 999% against electric shock: Unit must be earthed.
Peak pressure 0 to 999 mbar Degree of protection
Type BF, applied part
PEEP pressure 0 to 999 mbar against electric shock:
Mean pressure -999 to 999 mbar
Delta P 9 to 999 mbar
Trending Data logged @ 1 Hz
Above values are obtained under ATPD (ambient
temperature and pressure, dry) conditions.
SLE Ltd, Twin Bridges Business Park, 232 Selsdon Road, South Croydon, Surrey, CR2 6PL, UK Page 2 of 2
t +44 (0)20 8681 1414 f +44 (0)20 8649 8570 e sales@sle.co.uk w www.sle.co.uk © 2019 SLE Ltd | Subject to Change | G0171/0UK/001 07/19 (v2.0.50)