SLE6000 Specifications

(V2.0.50 software)

The SLE6000 is a high-specification, compact infant ventilator that offers a


range of conventional modes with additional options for non-invasive ventilation
(NIV), high frequency oscillation ventilation (HFOV) and high flow oxygen therapy.
SpO2 and etCO2 monitoring options are supported with the addition of plug-in
modules. OxyGenie (Auto O2) can also be integrated.

Core Invasive Ventilation Core Non-Invasive Ventilation

u CPAP (Dual limb, ET) u CMV (Dual limb, ET) u nCPAP D (Dual limb)
for passive nCPAP interfaces e.g. SLE Miniflow
Inspiratory Time (Ti) 0.1 to 3.0 s Respiratory Rate (RR) 1 to 150 BPM
CPAP 0 to 35 mbar Inspiratory Time (Ti) 0.1 to 3.0 s Inspiratory Time (Ti) 0.1 to 3.0 s

PIP 0 to 65 mbar PEEP 0 to 35 mbar CPAP 0 to 35 mbar

O2 Concentration 21 to 100% PIP 0 to 65 mbar PIP 0 to 65 mbar


Volume Targeted (Added with VTV module) O2 Concentration 21 to 100%
w Additional Parameters
Ventilation (VTV) 2 to 300 ml † w Additional Parameters
RR Backup 1 to 150 BPM
O2 Concentration 21 to 100% RR Backup 1 to 150 BPM
Rise Time 0 to 3.0 s
w Additional Parameters Rise Time 0 to 3.0 s
Trigger Sensitivity
with flow sensor: 0.2 to 20 l/min Rise Time 0 to 3.0 s Trigger Sensitivity 1 to 100%
without flow sensor: 1 to 100%
u SIMV (Dual limb, ET)
Respiratory Rate (RR) 1 to 150 BPM u NIPPV D (Dual limb)
u PTV & PSV (Dual limb, ET) Inspiratory Time (Ti) 0.1 to 3.0 s Respiratory Rate (RR) 1 to 150 BPM
Respiratory Rate (RR) 1 to 150 BPM PEEP 0 to 35 mbar Inspiratory Time (Ti) 0.1 to 3.0 s
Inspiratory Time (Ti) 0.1 to 3.0 s PIP 0 to 65 mbar PEEP 0 to 35 mbar
PEEP Pressure 0 to 35 mbar Volume Targeted (Added with VTV module) PIP 0 to 65 mbar
Ventilation (VTV) 2 to 300 ml † O2 Concentration 21 to 100%
PIP Pressure 0 to 65 mbar
O2 Concentration 21 to 100%
Volume Targeted (Added with VTV module) w Additional Parameters
Ventilation (VTV) 2 to 300 ml † w Additional Parameters Rise Time 0 to 3.0 s
O2 Concentration 21 to 100% Rise Time 0 to 3.0 s
w Additional Parameters P Support 0 to 65 mbar
Rise Time 0 to 3.0 s Trigger Sensitivity
Trigger Sensitivity with flow sensor: 0.2 to 20 l/min
with flow sensor: 0.2 to 20 l/min without flow sensor: 1 to 100%
without flow sensor: 1 to 100% Termination Sensitivity
5 to 50%
(% of peak insp flow)
Termination Sensitivity
(% of peak insp flow) 5 to 50% Termination Sensitivity parameter is not shown when
(PSV only) pressure support (P Support) is off.

Optional Ventilation Modes

u HFOV (Dual limb, ET) u nHFOV D (Dual limb) u NIPPV Triggered (Dual limb)
Frequency 3 to 20 Hz for passive nCPAP interfaces e.g. SLE Miniflow for passive nCPAP interfaces e.g. SLE Miniflow

I:E Ratio 1:1 / 1:2 / 1:3 Frequency 3 to 20 Hz Respiratory Rate (RR) 1 to 150 BPM
MAP 0 to 45 mbar I:E Ratio 1:1 / 1:2 / 1:3 Inspiratory Time (Ti) 0.1 to 3.0 s
Delta P 4 to 180 mbar Mean Airway Pressure 0 to 45 mbar PEEP 0 to 35 mbar
VTV 2 to 50 ml † Delta P 4 to 180 mbar PIP 0 to 65 mbar
O2 Concentration 21 to 100% O2 Concentration 21 to 100% O2 Concentration 21 to 100%
w Additional Parameters w Additional Parameters
w Additional Parameters
Sigh RR 1 to 150 BPM Rise Time 0 to 3.0 s
Sigh RR 1 to 150 BPM
Sigh Ti 0.1 to 3.0 s Trigger Sensitivity 1 to 100%
Sigh Ti 0.1 to 3.0 s
Sigh P 0 to 45 mbar
Sigh P 0 to 45 mbar u DuoPAP (Single limb)
u HFOV+CMV (Dual limb, ET) Respiratory Rate (RR) 1 to 60 BPM
u nCPAP S (Single limb)
for active (fluidic-flip) nCPAP interfaces (e.g. Inspiratory Time (Ti) 0.1 to 3.0 s
Respiratory Rate (RR) 1 to 150 BPM
SLE1000 generator or Infant Flow) PEEP 2 to 15 mbar
Inspiratory Time (Ti) 0.1 to 3.0 s
Inspiratory Time (Ti) 0.1 to 3.0 s PIP 2 to 25 mbar
Frequency 3 to 20 Hz
CPAP 2 to 15 mbar O2 Concentration 21 to 100%
PEEP 0 to 35 mbar
PIP 2 to 25 mbar
PIP 0 to 65 mbar
O2 Concentration 21 to 100%
Delta P 4 to 180 mbar O2 Therapy
w Additional Parameters
O2 Concentration 21 to 100%
RR Backup 1 to 10 BPM u High Flow Oxygen Therapy
w Additional Parameters
Trigger Sensitivity 1 to 100% (Single limb)
Oscillation on both high
HFO Waveform and low cycles or oscillation Flow Rate 2 to 30 l/min
on low cycle only. O2 Concentration 21 to 100%
Oscillation pause 60 s


VTV control, when enabled, becomes Vte Target control. Page 1 of 2
Optional Module features

u SpO2 (Masimo™ SET®) u EtCO2 (Oridion Microstream™) u OxyGenie®


Saturation (fraction of User selectable Adds additional (start/stop) option to
CO2 units
oxyhaemoglobin to (mmHg or kPa or Vol%) FiO2 parameter controller.
Displayed parameters functional haemoglobin), EtCO2 range 0-99.9 mmHg Range selector in SpO2 utilities menu.
pulse rate, Signal IQ and Controls Ranges are: 90 - 94%, 91 - 95%
plethysmogram EtCO2 resolution 1 mmHg
(default), 92 - 96%, 94 - 98%
SpO2 and Pulse rate for 0-38 mmHg: ± 2 mmHg Manual override (timed, for 30
Trends 39-150 mmHg: ± (5% of
previous 14 days CO2 accuracy seconds)
reading + 0.08 x [reading -
Absorption 39 mmHg]) Additional SpO2 screen can show
Measuring method
spectrophotometry any one ventilation parameter plus
50 ml/min (+15 ml/ Waveforms
ODU-type plug (red). plethysmogram and trends of SpO2
Ventilator connector CO2 sampling flow rate min, -7.5 ml/min) flow and FiO2.
Powered from ventilator. measured by volume
Dimensions (mm) 24 (h) x 33 (w) x 92 (l) Alarms automatically set on SpO2
Waveform sampling 20 samples/s software, corresponding with target
Weight (excluding 40 s (typical, includes range (1% above high and 1% below
0.122 kg
sensor) Initialisation time power-up and low). Can be manually set as well.
Alarms
Fractional Pulse Rate initialisation time) Alarm indications shown in Alarm bar.
SpO2 (%) (BPM)
ODU-type plug (yellow). Alarm level indicators on SpO2 and
Display Range 0% - 100% 25 - 240 BPM Ventilator connector
Powered from ventilator. FiO2 graphs.
Calibration Dimensions (mm) 70 (w) x 93.3 (l) x 50.3 (h) Status panel shows OxyGenie
70% - 100% 25 - 240 BPM
range status such as ‘Auto’, ‘Manual
Weight 240 g Indicator
No motion Override’ (with countdown) or
± 2.0% ± 3.0 BPM
accuracy (rms) For further specifications and operating temperature, ‘Waiting for Signal’.
Motion pressure and humidity ranges for SpO2 and etCO2 Trending information for SpO2 and
± 3.0% ± 5.0 BPM please see User Manuals.
accuracy (rms) FiO2 can be shown simultaneously.
Trends
Resolution ≤ 0.1% ≤ 1 BPM The Microstream technology is designed for use Up to 14 days of data are stored for
during invasive ventilation in conventional modes. each parameter.
Averaging time 2-4, 4-6, 8, 10,
- It is currently not recommended for use in NIV or
(seconds) 12, 14, 16
during HFOV.

An IntelliBridge module is also available

Misc. Specifications
w Flow Sensor w Power AC w Connectors (Rear mounted)
10 mm dual-hot-wire Mains voltage 100-240V / 50-60Hz RS232 & USB data ports
Flow sensor type:
anemometer. (Single-use Power 115 VA VGA (Video out) port
(Electrically isolated)
or autoclavable versions).
T2.5AH 250V USB Power port for nebuliser
Applied part Type BF Fuses (x2)
(5x20 mm) Nurse Call
Flow rate 0.2 to 30 l/min Typical 3+ hour battery 24V DC input
Accuracy ±8% maximum Battery back-up life (in all modes) in
normal use SpO2 & etCO2
Dead space 1 ml
Weight 10 g Full charge: 18 hours RJ45 Ethernet networking port
Battery charging
80% charge: 8 hours
w Flow w IP Rating
w Power DC
Flow rate 0 to 99 l/min Type of protection
IP21
Voltage 24V 4A against ingress of water
w Volume
w Operating Environment w Environmental Storage
Expiratory tidal Conditions
0 to 999 ml Temperature +10°C to +40°C
volume
10 to 90% Ambient Temperature -20°C to +50°C
Expiratory minute Relative Humidity
0 to 18 L (non-condensing) 10% to 90%
volume Relative Humidity
non-condensing
w Measured Parameters w Dimensions
w 330 mm x w Sound levels
Leak 0 to 99%
Size, ventilator only h 369 mm x Sound pressure level 49 dBA
Respiratory rate 0 to 999 BPM d 548 mm
Compliance 0 to 99.9 ml/mbar Height on pole 1310 mm
C20/C 0 to 9999 Weight (Ventilator only) ≤ 22 kg
Resistance 0 to 999 mbar/(l/s)
w Pneumatic Connectors
Inspiratory time 0 to 9.99 s
15 mm F / 22 mm M
Expiratory time 0 to 9.99 s Exhalation port
conical (ISO5356-1)
Vmin 0 to 99.99 l Proximal airway 5 mm non-conical
Trigger resolution 1 15 mm M conical
Fresh gas port
Vte 0 to 99.9 ml (ISO5356-1)
Nebulizer port (on rear) 5 mm non-conical
DC02 0 to 9999
w Classification (Electrical)
I:E Ratio 1:9.9 to 9.9:1
Type of protection Class 1
Oxygen concentration 0 to 999% against electric shock: Unit must be earthed.
Peak pressure 0 to 999 mbar Degree of protection
Type BF, applied part
PEEP pressure 0 to 999 mbar against electric shock:
Mean pressure -999 to 999 mbar
Delta P 9 to 999 mbar
Trending Data logged @ 1 Hz
Above values are obtained under ATPD (ambient
temperature and pressure, dry) conditions.

SLE Ltd, Twin Bridges Business Park, 232 Selsdon Road, South Croydon, Surrey, CR2 6PL, UK Page 2 of 2
t +44 (0)20 8681 1414 f +44 (0)20 8649 8570 e sales@sle.co.uk w www.sle.co.uk © 2019 SLE Ltd | Subject to Change | G0171/0UK/001 07/19 (v2.0.50)