Multislice HELICAL CT SCANNER

Product Data
No. MPDCT0765EAF

APPLICATION
Aquilion Start is a multislice helical CT system that supports
whole-body imaging. Aquilion Start is packed with premium
technology originally developed for our high-end CT
systems, carefully crafted and adapted keeping in mind the
changing requirements of healthcare organizations,
including the need for faster workflows, universal
accessibility, and sound economics.

FEATURES
Quality care for all
Aquilion Start delivers premium performance in a compact
system. It incorporates advanced technologies to optimize
patient care and accelerate clinical decision making.

• Designed to Maximize Clinical Flexibility • Designed to Realize Business Potential


– PUREViSION Detector technology, which was introduced in – From starting a new venture with an optimal investment,
our premium systems, has been adapted to achieve the to expanding business to new horizons, Aquilion Start is
perfect balance of image quality and patient dose the right choice for realizing all your business goals.
optimized for outstanding speed in routine clinical – Aquilion Start has a minimal footprint of only 9.8 m2*1, *2,
practice. compact enough to meet even the most restrictive siting
– Canon Medical Systems advanced clinical solutions requirements.
ensures excellent visualization of vasculature in CTA – Innovative Adaptive Power Management technologies
studies, while Iodine Mapping provides an extra level of dramatically decrease energy requirements, reducing
detail as part of your chest CTA exams. running costs and easing the environmental impact.
– SEMAR (Single Energy Metal Artifact Reduction) helps
eliminate metal artifacts, improving visualization of
implants and supporting bone and adjacent soft tissue
for a clearer and more confident diagnosis.

• Designed to Optimize Productivity Safely


– The Aquilion Start gantry features design innovations to
improve the scanning experience for patients while
providing excellent operability and ensuring safety.
– The spacious 780 mm wide bore and 470 mm wide
couch allow comfortable scanning for even the largest
patients. The couch-top can be lowered to a minimum
height of 312 mm, which makes it easier to transfer
patients from a wheelchair.
– Intuitive automation and a host of intelligent functions
enable fast and efficient workflows that not only save
time, but also provide the perfect clinical solution.

*1: Option
*2: For the short patient couch version
COMPOSITION PERFORMANCE SPECIFICATIONS
Standard composition (Model: TSX-037A/1, /2) Scan parameters
• Gantry......................................................................................................................1 • Gantry aperture: 780 mm in diameter
• Patient couch.....................................................................................................1 • Rotation: 360° continuous
• Console*1......................................................................................................1 set • Rotation times Unit: s
• Power distributor.............................................................................................1 Half scan 0.48
• Accessories...................................................................................................1 set
Axial scan 0.75, 1.0, 1.5, 2.0, 3.0
– Inter-unit cables
– Manuals Dynamic scan,
0.75, 1.0, 1.5
– Set of phantoms, phantom holder Helical scan,
SURE (This covers all 3 modes)
– Scan support accessories Start
• Time between scans
Optional items – S & S: Min. 1.8 s
• Cerebral blood-flow analysis system (CBP-study) – S & V: Min. 2.5 s
(CSCP-002A) • Continuous scan: Max. 50 s (2.0 MHU)
• Colon view (CSCV-001A) Max. 100 s (3.5 MHU)
• Display system for dental application (CDP-07A) • Acquisition modes
• Fat index view (CSFM-001A) – Conventional scan (S & S, S & V)
• FlyThrough software (CFT-03A)
16 row scan 1 mm (S & S only)
• Lung volume analysis (CSLV-001A)
4 row scan 1 and 4 mm
• SURESubtraction angio (CSSA-001A)
• SURESubtraction iodine mapping (CSSI-001A) – Dynamic volume scan
• SURESubtraction lung (CSSL-001A) 4 row scan 4 mm
• SURESubtraction ortho (CSSO-001A)
– Helical scan
• Vessel view (CVV-001A)
• Double slice kit (CSDS-003A) 16 row scan 1 mm
• Extended field of view (CSTC-005A) 4 row scan 1 mm
• Injector synchronization system, CAN protocol Class 1 • Scan field
(CKIS-004A) – CT scan Unit: mm
• Injector synchronization system, CAN protocol Class 4
M L
(CKIS-005A)
• Orbital synchronized scan system (CKOS-001A) φ350 φ430 (φ500*3)
• SURESubtraction scan system (CHSS-001A) – Scanoscopy Unit: mm
• Color printer interface (CCP-03A)
Axial direction Longitudinal direction
• DICOM fast transfer system (COT-45A)
Adjustable from
• DICOM MPPS (COT-33D)
Up to 384 200 to 1780*4
• DICOM MWM (COT-32D) 200 to 1480*2
• DICOM PGP profile (COT-44A)
• DICOM Q/R SCP (COT-34D) • Tube position for
• DICOM Q/R SCU (COT-35D) scanoscopy: 0°, 90°, 180° and 270°
• DICOM storage commitment SCU (COT-41D) Any desired angle can be
• DICOM storage SCP (COT-30D) specified (in 5° increments).
• Couch footswitches (CAFS-008A)
• Table-top stroke shortening kit (CBZH-010A)*2
Note: Some options may not be available in your country or region.
Please check with your sales representative.

*1: Desk not included.


*2: For the short patient couch version
*3: Option
*4: For the long patient couch version
2
MPDCT0765EAF

Helical scan Dynamic scan


• Continuous scan time: Max. 50 s (2 MHU) • Programmable time: Max. 1 hour/eXam Plan
Max. 100 s (3.5 MHU) • Number of
• Scan start time delay: Min. 5 s programmable scans: Max. 20
Setting is possible in increments Max. 50 s/scan (2 MHU)
of 0.1 s. Max. 100 s/scan (3.5 MHU)
• Scan length (with headrest) Unit: mm • Scan plan
Long patient couch 1780 – Scan interval: Min. 1 s in increments of 0.1 s.
version Note: When a scanning mode with patient couch movement is
Short patient couch 1480, used, the minimum scan interval is limited by the time
required for movement.
version 1350*1, 1150*1, 950*1, 750*1
• Scan start delay time: Min. 0.5 s
• Multi-helical: Up to 10 scan plans are Setting is possible in increments
programmable in one eXam Plan. of 0.1 s.
(Multiple and/or multidirectional • Scan rate: Max. 66 scans/100 s (2 MHU)
helical) (0.75 s scan, 66 rotations)
• Couch-top speed: The couch-top speed can be Max. 133 scans/100 s (3.5 MHU)
specified in a range between (0.75 s scan, 133 rotations)
1.6 mm/s and 43 mm/s. • Image reconstruction
• CT pitch factor – Image interval: 0.1 s
For 16 row scanning: Scan field M
0.625 to 1.0, 1.125 to 1.5 SURE
Start
Scan field L • Scan start mode: Automatic
0.625 to 1.0 Manual
For 4 row scanning: 0.625 to 0.875, 1.125 to 2.0 • Continuous scan time: Max. 100 s
• SUREExposure 3D: Function for continuously varying • Acquisition mode: Intermittent, continuous,
the X-ray tube current to ensure combination
the minimal X-ray dose during • Region of interest (ROI): Max. 3
helical scanning. • CT number
measurement interval: 0.083 s
• Image reconstruction
• Scan start delay time: Min. 3 s
time: Up to 15 images/s* • Display function: Mean CT number within the ROI,
(0.066 s/image) elapsed time
*: Depending on the scan conditions and reconstruction conditions.
• Real-time helical Voice-recorded instruction and scan system
reconstruction time: 12 images/s (0.083 s/image) (VoiceLink)
(1 slice, 512 × 512 matrix) As part of the eXam Plan, voice instructions to the patient
• Specification of can be recorded by the operator and automatically played
reconstruction position: By entering the couch-top back during scan sequences.
position or using the scanogram • Number of messages: Max. 200
• Reconstruction method: MUSCOT* • Recording time: Max. 29 s per message
TCOT**+ • Delay time setting: The delay time between the end
• Reconstruction mode: Full image of the message and the start of
Half image scanning can be set for up to
Detail image 10 s, in increments of 1 s.
*: Multi-slice Cone-beam Tomography
**: True Cone-beam Tomography

*1: Option
3
Patient couch X-ray detection
Load • Detection system: Solid-state detectors
Maximum 220*1 (485) • Main detector: 720 channels × 16 elements
(kg [lb])
Table type Long Short • Number of elements: 11520
Width • Data acquisition: 720 channels × 16 rows
470 • Reference detector: 1 set
(mm)
Step feed Range 0.5 - 600 • View rate: Max. 1200 views/s
(mm) Increments 0.5
Height Maximum 900 Data processing
(mm) Minimum 312 • Reconstruction matrix: 512 × 512
Stroke Vertical 588 • Picture element (pixel) size
(mm) 2190 1890 – CT image Unit: mm
1760*2
Scan field M L
Horizontal 1560*2
*to *to
1360*2 Pixel size
0.68 0.84 (0.98*2)
1160*2
Scan range (with headrest) 1830 1530 *: Depending on the Vari-Area or Zoom factor
(mm) 1400*2 – Scanogram Unit: mm
1200*2
Enlargement ratio (area) Standard
1000*2
800*2 LL 1.00
Horizontal reproducibility L 1.00
± 0.25 Pixel size
(mm) M 1.00
Speed 16 - 24 (50 Hz) S 0.50
Up
(mm/s) 19 - 28 (60 Hz)
• Dose reduction functions
Down 20 - 30
Horizontal 10 / 130 – Quantum Denoising Software (QDS)
System driver Vertical Hydraulic – Adaptive Iterative Dose Reduction 3D (AIDR 3D)
Horizontal Motor / Manual – AIDR 3D Enhanced
• Metal artifact reduction function
X-ray generation – SEMAR
• X-ray exposure: Continuous • Reconstruction filter functions
• X-ray tube voltage: 80, 100, 120 and 135 kV – Abdomen with BHC
• X-ray tube current: 10 mA to 300 mA*3 – Abdomen without BHC
600 mA max. equivalent – Brain with BHC
with AIDR 3D* (3.5 MHU) – Brain without BHC
72 kW equivalent – Inner ear and bone
with AIDR 3D* (3.5 MHU) – Lung
• X-ray tube heat capacity – High-resolution mode for evaluation of resolution
– TSX-037A/1 2 MHU parameters
5 MHU equivalent with AIDR 3D – Auditory ossicles and spine with high-resolution
– TSX-037A/2 3.5 MHU processing
7.5 MHU equivalent with AIDR 3D – Maintenance
• X-ray tube cooling rate: Max. 337.8 kHU/min (2 MHU) • Reconstruction time: Min. 0.066 s/image*
Max. 735 kHU/min (3.5 MHU) (Up to 15 images/s)
• Focal spot size *: Depending on the scan conditions and reconstruction conditions.
– IEC 60336: 2005 • Real-time scanoscopy
nominal : 2 MHU • Data processing unit
1.1 mm × 1.3 mm (small) – CPU: 64 bit
1.7 mm × 1.7 mm (large) – Memory size: 32 GB
3.5 MHU – Magnetic disk unit: Raw data, 180 GB
1.1 mm × 1.1 mm (small) Image data, 300 GB
1.7 mm × 1.7 mm (large) *1: Patient weight Max. 205 kg (452 lb) + Accessories 15 kg (33 lb)
*: Applying AIDR 3D, the same standard deviation on water phantom *2: Option
images can be obtained at lower mA with less tube output. *3: kV dependent
4
MPDCT0765EAF

Data storage – OEM (Organ Effective


• Magnetic disk Modulation): Function for reducing exposure
– Raw data: Max. 3600 rotations dose from specific directions to
– Image data: Max. 260000 tissues such as the eyes and the
• DVD-RAM: 4.7 GB breasts.
– DICOM images: 8000 • CT image processing
• DVD-R: 4.7 GB – ROI
– DICOM images: 7500 · Shape: Point, rectangular, polygonal,
elliptical, irregular
Image display · Processing: Mean value, standard deviation,
• Display monitor: 48.1 cm (19 inch color) LCD unit area, number of pixels, maximum
• Monitor matrix: 1280 × 1024 value, minimum value
• Image matrix: 1024 × 1024 (max.) · Display: Max. 10/image
• CT number · Control: Size, position, rotation
– Display range: From -1536 to +8191 – Measurement of distance and angle between two points
Note: The CT number measurement range is from -32768 to – Profile (oblique profile also available)
+32767. – Histogram
• Window width/level: Continuously variable – CT number display
(adjustable at variable speed) – Mark display (grid display, scale display)
• Preset windows: 3/image – Volume calculation
• Window types: Linear, non-linear – Enlargement, reduction, panning
(including user-programmable – Addition/subtraction between images
and double windows) – Band display (non-linear windowing)
• Image retrieval – Comment and arrow insertion
– Method: On-screen menus and keyboard – Top/bottom, right/left, black/white reversal of image
– Mode: Image, series and patient – Image filtering
• Autoview function: Software control, function key – Image rotation (arbitrary rotation)
• Multi-frame display: Reduction/cutoff display, – Screen save
ROI processing – High-speed axial interpolation
• Inset scanogram display – MultiView (Auto MPR)
• Information display: User selectable – Quantum denoising software (QDS)
• Cine display: Variable speed – Boost3D
• Scanogram/CT image – Z-sharpening
switching: Show/hide scano line, zoom • Raw data processing
• Slice-feed playback – Zooming reconstruction
(CineView): High-speed image feed using the – Stack reconstruction
mouse or keyboard – Protect/Unprotect
– Half-view reconstruction of helical scan raw data
– High-resolution reconstruction for raw data acquired by
Image processing
• Scanogram processing helical scanning
– Slice position display – Play/Reverse reconstruction (helical and dynamic scan)
(display of planned slice, preset slice – Priority reassignment in reconstruction queue
and last scanned slice)
– Anatomical scale System management
(display of position, relative to selected zero position) • Warm-up function
– Slice position setting • Calibration data acquisition
– Enlargement • Patient data input
– Flex e-Tilt: Functional tilting from the • Patient appointment function
comfort of the console to enable • Examination summary
faster scan planning and • eXam Plan editing
optimized angled scans to • Modification of related information
generate images at the desired • Operation environment settings
anatomical reading plane.

5
• Slice counter IMAGE QUALITY
• Sleep mode
• Access control (NEMA XR-26) Noise
Standard deviation Less than 0.7%
Dose management Scan parameters
• CTDIvol (or CTDIw)/DLP/Geometric efficiency in z-direction Tube voltage 120 kV
Tube current 300 mA
• Dose check (NEMA XR-25)
Scan time 1s
• DICOM SC Exposure summary
Slice thickness 8 mm
• DICOM SR compliant Dose report
(4 mm × 4 rows : 2 stack)
Reconstruction function FC70
3D color image processing Scan field M
High-quality 3D images can be obtained rapidly and with Phantom φ18 cm water
easy operation.
• 3D surface rendering
Spatial resolution
– Clipping, texture or non-texture
Resolution 18.0 lp/cm at MTF 0%* (3.5 MHU)
• 3D volume rendering
14.5 lp/cm at MTF 2%
– Maximum intensity projection (Max-IP)
8.0 lp/cm at MTF 50%
– Minimum intensity projection (Min-IP) *For reference
– X-ray volume rendering Scan parameters
– Intensity volume rendering Tube voltage 120 kV
– Shaded volume rendering Focal spot Small
(a desired opacity curve can be set) Slice thickness 2 mm
• Display/processing function (1 mm × 4 rows : 2 stack)
Zooming, panning, measurement (distance, angle), Reconstruction function FC90
annotation, cutting, drilling Scan field M
• Cine display Phantom IRIS QA phantom
• MPR
3 orthogonal planes/oblique image
High-contrast detectability
Curved MPR
X-ray tube heat capacity 2 MHU 3.5 MHU
• Easy accurate bone elimination function
X-Y plane 0.35 ± 0.05 mm 0.31 mm
• High resolution mode
Scan parameters
Tube voltage 120 kV 120 kV
Image transfer Tube current 130 mA 150 mA
• 1000BASE-T, 100BASE-TX, 10BASE-T Scan time 1.5 s 1.5 s
• TSB protocol Slice thickness 1 mm
• DICOM storage SCU (1 mm × 4 rows : No stack)
• Enhanced DICOM*1 Reconstruction method MUSCOT
• TIFF conversion Reconstruction function FC90
Scan field M
Filming Phantom Catphan® 500 phantom
• Ethernet: TSB protocol (CTP528 module)
DICOM PRINT SCU Z-direction 0.85 mm 0.85 mm
• Sheet editing function using virtual film Scan parameters
• T-mode: Related information items such Tube voltage 120 kV 120 kV
as the patient name are displayed Tube current 130 mA 150 mA
in the footer area using a larger Scan time 1.5 s 1.5 s
font. Slice thickness 1 mm
(1 mm × 16 rows : No stack)
Note: To use T-mode, the laser imager must support
2048 × 2404 pixels for a 1 × 1 frame layout.
Reconstruction method TCOT+ with 0.5 mm SR
Reconstruction function FC70
• Auto filming in eXam Plans Scan field M
Phantom Catphan 500 phantom
(CTP528 module)

*1: Option
6
MPDCT0765EAF

Low contrast detectability (3.5 MHU) X-ray tube


Object size (A) 2 mm at 0.3% This is a large-capacity, high-cooling-rate X-ray tube that is
CTDIvol 17.7 mGy able to withstand continuous operation as in helical
Object size (B) 3 mm at 0.3% scanning.
CTDIvol 9.5 mGy • Heat capacity
Scan parameters 10 mm (with AIDR 3D) – TSX-037A/1 2 MHU
Phantom φ20 cm Catphan 5 MHU equivalent with AIDR 3D
– TSX-037A/2 3.5 MHU
7.5 MHU equivalent with AIDR 3D
Low contrast detectability (2.0 MHU)
• Cooling rate: Max. 337.8 kHU/min (2 MHU)
Object size (A) 2 mm at 0.3%
Max. 735 kHU/min (3.5 MHU)
CTDIvol 18.3 mGy
Object size (B) 3 mm at 0.3%
CTDIvol 9.9 mGy
Patient couch
The patient couch is positioned in front of the gantry and
Scan parameters 10 mm (with AIDR 3D)
supports the patient. The entire unit moves vertically, and
Phantom φ20 cm Catphan
the couch-top moves longitudinally. In an emergency, the
couch-top can be pulled out manually with very little effort.
SYSTEM COMPONENTS AND THEIR The couch-top can also be lowered to a minimum height of
FUNCTIONS approx. 312 mm from the floor, facilitating transfer of the
Gantry patient from a low bed or stretcher.
The scanner is composed of the gantry and the patient
couch. The scanner uses a fan-shaped continuous X-ray Console
beam to scan the region to be examined. Transmitted X-rays The console is provided with a hybrid keyboard, a monitor
are detected and converted into electrical signals by the and a mouse.
SSMD. • Functions for scanning
The gantry includes the main body and its support – Selection of scan parameters
mechanism. The X-ray tube and the SSMD are mounted – Scanoscope control
facing each other on either side of the gantry aperture, and – Scan control
the X-ray tube and detectors rotate continuously around the – Couch-top movement control
gantry aperture. A slipring is employed to transmit power – Flex e-Tilt
between the gantry and the rotating X-ray high-voltage • Functions for image processing
generator assembly. – Window level and window width adjustment
Three-dimensional alignment lights are provided for setting – Other mouse-operated image processing functions
slice positions. Gantry and patient couch operating controls
are provided on both sides of the front of the gantry OPERATING FEATURES
housing. The patient guide display indicates the scan status
Patient handling and positioning
to the operator and the patient. The X-ray high-voltage
• The couch-top can be lowered to approx. 312 mm from
generator is built into the gantry, and the system employs a
the floor, facilitating transfer of the patient to and from a
high-frequency inverter for generating and stabilizing the
bed or stretcher.
high voltage supplied to the X-ray tube. The generator
• Alignment lights are provided in the gantry aperture for
includes electronic circuits for controlling the speed of the
fast and accurate patient positioning.
rotating anode in the X-ray tube. Use of a high-frequency
• High-precision couch-top positioning is possible from the
inverter system results in high power output combined with
integrated console or by manual operation from the control
excellent stability. In addition, the system is compact and
panel and clear digital readouts are provided on the gantry.
lightweight.
• The couch-top can be pulled out manually in an
emergency.
X-ray generator
This unit supplies stable high voltage to the X-ray tube unit.
The high-frequency inverter method is employed, resulting
in a light and compact design. This unit is incorporated in
the gantry.
• Max. power: 50.4 kW

7
Scanning Image display and processing
• Scanoscope function provides a projection image of the • Reconstructed images are automatically displayed
patient for high-precision advance planning of scan areas. according to the window settings preset in the eXam Plan.
• On the scanogram, the length of the scan area can be • The window save function allows the user to store an
adjusted over a range of up to 1780 mm*1 or 1480 mm*2. image with different window settings from those set in the
Because the images are reconstructed in real time, the eXam Plan.
scan can be aborted at any time. This allows the patient • Filter parameters can be customized through simple
exposure dose to be minimized. on-screen menu selections. These parameters include the
• The auto index function allows automatic incremental number of filtering passes, matrix size and filter
couch-top movement based on the slice positions coefficients.
determined through the scanogram. • Images can be rotated and reversed
• The eXam Plan function allows simple selection of pre- (right/left, top/bottom, black/white).
programmed scanning parameters for routine • The Multi-frame feature allows up to 15 images to be
examinations, maximizing patient throughput. retrieved and displayed simultaneously on the screen.
• Protocol comments can be saved in each eXam Plan, • The three-dimensional image display function allows color
providing interactive onscreen instructions for all studies, three-dimensional and MPR images in real-time to be
reducing the need to refer to a separate protocol book. generated from the volumetric scan data acquired by
• The Vari-area function allows the user to preselect a region helical scanning. This results in higher definition and
of interest for zooming using raw data, permitting image quality than images reconstructed from
immediate post-scan analysis. Zooming using raw data conventional single-slice scanning. This is because helical
yields higher resolution than enlarging an image that has scanning provides superior data continuity along the
already been reconstructed. patient axis compared with conventional scanning.
• Dynamic and rapid-sequence scan modes are provided.
• Multislice helical scan acquires raw data by rotating the Image filming
X-ray tube continuously while moving the patient • Filming of images can be performed manually or
continuously through the scanner. The acquired volume automatically from the console.
data can be used to reconstruct axial slices at any desired • Automatic filming sends an entire study to the laser printer.
position. This scan mode is best used for rapid patient Filming is performed in background mode so that other
scanning during a single breath-hold and for high- scanner and image processing functions can be performed
definition three-dimensional and MPR imaging. without interruption or delay.
• Real-time helical reconstruction mode makes it possible to • When T-mode is used, related information items displayed
observe the images being scanned in real time at a together with an image (surrounding the image, in a small
maximum of 12 frames per second. This mode shows any font) are displayed in the footer area using a larger font,
shift in the slice position in real time and allows the permitting not only easier reading but also simpler film
operator to check the scan field on the image, contrast management.
study timing, patient body motion, etc. The patient can Note: To use T-mode, the laser imager must support
therefore be released immediately after scanning. 2048 pixels × 2404 pixels for a 1 × 1 frame layout.
• The SUREStart function allows the operator to start helical
scanning at the timing of maximum enhancement in Patient throughput
contrast studies. SUREStart monitors the CT number in real Patient throughput and cost effectiveness were major
time to detect the arrival of contrast medium in the image. objectives in the design and production of the system.
When the CT number reaches the predefined threshold, • The system incorporates a 2 MHU or 3.5 MHU X-ray tube
dynamic volume scan or helical scan automatically starts. with a fast cooling rate of 337.8 kHU/min or 735 kHU/min
This technique ensures optimal contrast enhancement, in actual use.
independent of individual differences in blood flow • High-speed scans can be performed in as little as
velocity, and at the same time minimizes the dose of 0.75 second per scan.
contrast medium. • Real-time scanoscopy.
• Ease of operation is ensured through the incorporation of
Data processing a hybrid keyboard, mouse-driven menus and a large color
• A variety of reconstruction algorithms are available and LCD screen.
can be selected according to the anatomical region • The couch-top can be lowered to a position close to the
examined and the clinical objective of the study. These floor, facilitating patient transfer.
include algorithms for the abdomen, head, bones, lung, *1: For the long patient couch version
small structures, soft tissues, etc. *2: For the short patient couch version
8
MPDCT0765EAF

COMPLIANCE Grounding
Grounding must be provided in accordance with local
Council Directive 93/42/EEC and subsequent amendments
regulations for medically used electrical equipment.
IEC 60601-1: 2005+Amd.1: 2012
IEC 60601-1-2: 2007 Power distribution board
IEC 60601-1-2: 2014 For 3 phase, 400 V
IEC 60601-1-3: 2008+Amd.1: 2013 Ground resistance:
As per applicable
IEC 60601-1-6: 2010+Amd.1: 2013 legal requirements.
IEC 60601-1-9: 2007+Amd.1: 2013 NFB
100 A to 150 A
IEC 60601-2-28: 2010
IEC 60601-2-28: 2017
IEC 60601-2-44: 2009+Amd.1: 2012+Amd.2: 2016 Ground bar
Power distributor
IEC 60825-1: 2014
IEC 62304: 2006+Amd.1: 2015
IEC 62366: 2007+Amd.1: 2014 Ambient conditions
Temperature Humidity Heat generation
DIMENSIONS AND MASS Scan room
Gantry 20°C to 26°C 40% to 80% Approx.
Dimensions W × L × H Mass Tolerance: ±2°C No 9720 kJ/h*
Unit
mm (in) kg (lb) condensation 32070 kJ/h**
2050 × 960 × 1910 1300 Patient couch 20°C to 26°C 40% to 80% Approx.
Gantry Tolerance: ±2°C No 1080 kJ/h*
(80.7 × 37.8 × 75.2) (2866)
condensation 1800 kJ/h**
Long patient 630 × 2690 × 450 485 Control room
Patient couch version (24.8 × 105.9 × 17.7) (1069) Console 16°C to 28°C 40% to 80% Approx.
couch Short patient 630 × 2390 × 450 455 No 10800 kJ/h
condensation
couch version (24.8 × 94.1 × 17.7) (1003) Power distributor 16°C to 28°C 40% to 80% Approx.
430 × 690 × 645 75 No 2883 kJ/h*
Console (NAVI Box) condensation 6400 kJ/h**
(16.9 × 27.2 × 25.4) (165)
700 × 695 × 973 470 *: When scanning is not performed.
Power distributor **: When scanning is performed continuously at the maximum rated
(27.6 × 27.4 × 38.3) (1036) output of the system.

SITING REQUIREMENTS
Power requirements
• Phase: Three-phase
• Voltage: 380 V, 400 V, 415 V, 440 V, 460 V,
480 V*
• Frequency: 50 Hz or 60 Hz ±1 Hz
• Line capacity: 50 kVA
• Voltage fluctuation
due to load variation: Less than 5%
• Power voltage
fluctuation: Less than 10%**
*: For other voltages or in the event of excessive power fluctuation,
consult with your sales representative.
**: Represents the total voltage fluctuation due to load and power
variation.

9
Room layout example Installation requirements

3300 Scan room


• Before installing the gantry, check the maximum
permissible floor load.

1324 (1024*)
• The scanner emits radiation. X-ray shielding must be
provided around the scan room and the entrance in
accordance with all local requirements and regulations.
• The ceiling height should be at least 2500 mm when a
ceiling mounted contrast injector is installed.
5230 (4630*)

1 • Wiring pits and ducts are required for routing cables that
connect the various units.

2 Control room
7630 (7030*)

• An observation window is required for monitoring the


scan room. X-ray shielding of the window glass must be
provided in accordance with all local requirements and
regulations, and the bottom of the window frame should
be 90 cm from the floor.
• Wiring pits and ducts are required for routing cables that
(100)

connect the various units.


• The control room should have entrances for access to the
corridor and the scan room.
2300

Checks before bringing-in the unit


700

• Check in advance the width of the corridor, the dimensions


of the entrance, and the dimensions and maximum
Unit: mm
allowable load of the stairs and elevators to ensure that it
1 Gantry 3 Console
is possible to bring in the unit safely and without difficulty.
2 Patient couch 4 Power distributor
• The minimum dimensions of the entrance used for
*: For the short patient couch version.
bringing in the unit are as follows
– Width: 1060 mm (41.7 in)
Minimum area for installation – Height: 2130 mm (83.9 in)
Long patient couch version 17.2 m2 • The corners of corridors should be as illustrated below.
Scan room area 12.6 m2 • Elevator load rating: At least 2000 kg (4400 lb)
Control room area 4.6 m2
Short patient couch version
Scannable range 800 mm*1 14.4 m2 W>
= 1800 mm (70.9 in)
W W
Scan room area 9.8 m2
Control room area 4.6 m2

*1: Option
10
MPDCT0765EAF

Cable connections between units in meters

Breaker box

Patient couch
Power 13
Gantry
distributor
17

17 Speaker
17 (Scan room)

9 Speaker
(Control room)

Console

11
OUTLINE DRAWINGS

Gantry and patient couch

12
MPDCT0765EAF

OUTLINE DRAWINGS

870
(34.3)
2050 340
(80.7) (13.4)
(3 80

(75.2)
1910
7)
7
0.
φ

(37.8)
960

960
(37.8)

Gantry

Unit: mm (in)

13
OUTLINE DRAWINGS

2362 2190 (86.2)


(93.0) COUCH-TOP STROKE

(18.5)
(24.8)
470
630

2690
(105.9)
588 (23.1)
STROKE
(40.9)
1038

(17.7)
450

Patient couch (for the 220 kg (485 lb) long patient couch version)

2062 1890 (74.4)


(81.2) COUCH-TOP STROKE
(18.5)
(24.8)
470
630

2390
(94.1)
588 (23.1)
STROKE
(40.9)
1038

(17.7)
450

Patient couch (for the 220 kg (485 lb) short patient couch version)

Note: The footswitch is not included in the standard configuration.

Unit: mm (in)

14
MPDCT0765EAF

OUTLINE DRAWINGS

(25.4)
645

430 690
(16.9) (27.2)

Console

Note: The console desk is not included in the standard configuration.


Some of the units shown in the photograph on the front page differ from those shown in the drawings above.

Unit: mm (in)

15
MPDCT0765EAF

OUTLINE DRAWINGS

700
(27.6)

(38.3)
973

Anchor plate

866 695
(34.1) (27.4)

Power distributor

Unit: mm (in)

1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan

https://global.medical.canon
©Canon Medical Systems Corporation 2018-2021. All rights reserved.
Design and specifications are subject to change without notice.
Model number: TSX-037A MPDCT0765EAF 2021-02 CMSC/Produced in Japan

Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.

Aquilion, SEMAR, SURESubtraction, SUREStart, SUREExposure, Boost3D and Made for Life are trademarks of Canon Medical Systems Corporation.
Catphan is a registered trademark of The Phantom Laboratory.
This document may include trademarks or registered trademarks of their respective owners.