Model 3. Aquilion Start Datasheet
Model 3. Aquilion Start Datasheet
Product Data
No. MPDCT0765EAF
APPLICATION
Aquilion Start is a multislice helical CT system that supports
whole-body imaging. Aquilion Start is packed with premium
technology originally developed for our high-end CT
systems, carefully crafted and adapted keeping in mind the
changing requirements of healthcare organizations,
including the need for faster workflows, universal
accessibility, and sound economics.
FEATURES
Quality care for all
Aquilion Start delivers premium performance in a compact
system. It incorporates advanced technologies to optimize
patient care and accelerate clinical decision making.
*1: Option
*2: For the short patient couch version
COMPOSITION PERFORMANCE SPECIFICATIONS
Standard composition (Model: TSX-037A/1, /2) Scan parameters
• Gantry......................................................................................................................1 • Gantry aperture: 780 mm in diameter
• Patient couch.....................................................................................................1 • Rotation: 360° continuous
• Console*1......................................................................................................1 set • Rotation times Unit: s
• Power distributor.............................................................................................1 Half scan 0.48
• Accessories...................................................................................................1 set
Axial scan 0.75, 1.0, 1.5, 2.0, 3.0
– Inter-unit cables
– Manuals Dynamic scan,
0.75, 1.0, 1.5
– Set of phantoms, phantom holder Helical scan,
SURE (This covers all 3 modes)
– Scan support accessories Start
• Time between scans
Optional items – S & S: Min. 1.8 s
• Cerebral blood-flow analysis system (CBP-study) – S & V: Min. 2.5 s
(CSCP-002A) • Continuous scan: Max. 50 s (2.0 MHU)
• Colon view (CSCV-001A) Max. 100 s (3.5 MHU)
• Display system for dental application (CDP-07A) • Acquisition modes
• Fat index view (CSFM-001A) – Conventional scan (S & S, S & V)
• FlyThrough software (CFT-03A)
16 row scan 1 mm (S & S only)
• Lung volume analysis (CSLV-001A)
4 row scan 1 and 4 mm
• SURESubtraction angio (CSSA-001A)
• SURESubtraction iodine mapping (CSSI-001A) – Dynamic volume scan
• SURESubtraction lung (CSSL-001A) 4 row scan 4 mm
• SURESubtraction ortho (CSSO-001A)
– Helical scan
• Vessel view (CVV-001A)
• Double slice kit (CSDS-003A) 16 row scan 1 mm
• Extended field of view (CSTC-005A) 4 row scan 1 mm
• Injector synchronization system, CAN protocol Class 1 • Scan field
(CKIS-004A) – CT scan Unit: mm
• Injector synchronization system, CAN protocol Class 4
M L
(CKIS-005A)
• Orbital synchronized scan system (CKOS-001A) φ350 φ430 (φ500*3)
• SURESubtraction scan system (CHSS-001A) – Scanoscopy Unit: mm
• Color printer interface (CCP-03A)
Axial direction Longitudinal direction
• DICOM fast transfer system (COT-45A)
Adjustable from
• DICOM MPPS (COT-33D)
Up to 384 200 to 1780*4
• DICOM MWM (COT-32D) 200 to 1480*2
• DICOM PGP profile (COT-44A)
• DICOM Q/R SCP (COT-34D) • Tube position for
• DICOM Q/R SCU (COT-35D) scanoscopy: 0°, 90°, 180° and 270°
• DICOM storage commitment SCU (COT-41D) Any desired angle can be
• DICOM storage SCP (COT-30D) specified (in 5° increments).
• Couch footswitches (CAFS-008A)
• Table-top stroke shortening kit (CBZH-010A)*2
Note: Some options may not be available in your country or region.
Please check with your sales representative.
*1: Option
3
Patient couch X-ray detection
Load • Detection system: Solid-state detectors
Maximum 220*1 (485) • Main detector: 720 channels × 16 elements
(kg [lb])
Table type Long Short • Number of elements: 11520
Width • Data acquisition: 720 channels × 16 rows
470 • Reference detector: 1 set
(mm)
Step feed Range 0.5 - 600 • View rate: Max. 1200 views/s
(mm) Increments 0.5
Height Maximum 900 Data processing
(mm) Minimum 312 • Reconstruction matrix: 512 × 512
Stroke Vertical 588 • Picture element (pixel) size
(mm) 2190 1890 – CT image Unit: mm
1760*2
Scan field M L
Horizontal 1560*2
*to *to
1360*2 Pixel size
0.68 0.84 (0.98*2)
1160*2
Scan range (with headrest) 1830 1530 *: Depending on the Vari-Area or Zoom factor
(mm) 1400*2 – Scanogram Unit: mm
1200*2
Enlargement ratio (area) Standard
1000*2
800*2 LL 1.00
Horizontal reproducibility L 1.00
± 0.25 Pixel size
(mm) M 1.00
Speed 16 - 24 (50 Hz) S 0.50
Up
(mm/s) 19 - 28 (60 Hz)
• Dose reduction functions
Down 20 - 30
Horizontal 10 / 130 – Quantum Denoising Software (QDS)
System driver Vertical Hydraulic – Adaptive Iterative Dose Reduction 3D (AIDR 3D)
Horizontal Motor / Manual – AIDR 3D Enhanced
• Metal artifact reduction function
X-ray generation – SEMAR
• X-ray exposure: Continuous • Reconstruction filter functions
• X-ray tube voltage: 80, 100, 120 and 135 kV – Abdomen with BHC
• X-ray tube current: 10 mA to 300 mA*3 – Abdomen without BHC
600 mA max. equivalent – Brain with BHC
with AIDR 3D* (3.5 MHU) – Brain without BHC
72 kW equivalent – Inner ear and bone
with AIDR 3D* (3.5 MHU) – Lung
• X-ray tube heat capacity – High-resolution mode for evaluation of resolution
– TSX-037A/1 2 MHU parameters
5 MHU equivalent with AIDR 3D – Auditory ossicles and spine with high-resolution
– TSX-037A/2 3.5 MHU processing
7.5 MHU equivalent with AIDR 3D – Maintenance
• X-ray tube cooling rate: Max. 337.8 kHU/min (2 MHU) • Reconstruction time: Min. 0.066 s/image*
Max. 735 kHU/min (3.5 MHU) (Up to 15 images/s)
• Focal spot size *: Depending on the scan conditions and reconstruction conditions.
– IEC 60336: 2005 • Real-time scanoscopy
nominal : 2 MHU • Data processing unit
1.1 mm × 1.3 mm (small) – CPU: 64 bit
1.7 mm × 1.7 mm (large) – Memory size: 32 GB
3.5 MHU – Magnetic disk unit: Raw data, 180 GB
1.1 mm × 1.1 mm (small) Image data, 300 GB
1.7 mm × 1.7 mm (large) *1: Patient weight Max. 205 kg (452 lb) + Accessories 15 kg (33 lb)
*: Applying AIDR 3D, the same standard deviation on water phantom *2: Option
images can be obtained at lower mA with less tube output. *3: kV dependent
4
MPDCT0765EAF
5
• Slice counter IMAGE QUALITY
• Sleep mode
• Access control (NEMA XR-26) Noise
Standard deviation Less than 0.7%
Dose management Scan parameters
• CTDIvol (or CTDIw)/DLP/Geometric efficiency in z-direction Tube voltage 120 kV
Tube current 300 mA
• Dose check (NEMA XR-25)
Scan time 1s
• DICOM SC Exposure summary
Slice thickness 8 mm
• DICOM SR compliant Dose report
(4 mm × 4 rows : 2 stack)
Reconstruction function FC70
3D color image processing Scan field M
High-quality 3D images can be obtained rapidly and with Phantom φ18 cm water
easy operation.
• 3D surface rendering
Spatial resolution
– Clipping, texture or non-texture
Resolution 18.0 lp/cm at MTF 0%* (3.5 MHU)
• 3D volume rendering
14.5 lp/cm at MTF 2%
– Maximum intensity projection (Max-IP)
8.0 lp/cm at MTF 50%
– Minimum intensity projection (Min-IP) *For reference
– X-ray volume rendering Scan parameters
– Intensity volume rendering Tube voltage 120 kV
– Shaded volume rendering Focal spot Small
(a desired opacity curve can be set) Slice thickness 2 mm
• Display/processing function (1 mm × 4 rows : 2 stack)
Zooming, panning, measurement (distance, angle), Reconstruction function FC90
annotation, cutting, drilling Scan field M
• Cine display Phantom IRIS QA phantom
• MPR
3 orthogonal planes/oblique image
High-contrast detectability
Curved MPR
X-ray tube heat capacity 2 MHU 3.5 MHU
• Easy accurate bone elimination function
X-Y plane 0.35 ± 0.05 mm 0.31 mm
• High resolution mode
Scan parameters
Tube voltage 120 kV 120 kV
Image transfer Tube current 130 mA 150 mA
• 1000BASE-T, 100BASE-TX, 10BASE-T Scan time 1.5 s 1.5 s
• TSB protocol Slice thickness 1 mm
• DICOM storage SCU (1 mm × 4 rows : No stack)
• Enhanced DICOM*1 Reconstruction method MUSCOT
• TIFF conversion Reconstruction function FC90
Scan field M
Filming Phantom Catphan® 500 phantom
• Ethernet: TSB protocol (CTP528 module)
DICOM PRINT SCU Z-direction 0.85 mm 0.85 mm
• Sheet editing function using virtual film Scan parameters
• T-mode: Related information items such Tube voltage 120 kV 120 kV
as the patient name are displayed Tube current 130 mA 150 mA
in the footer area using a larger Scan time 1.5 s 1.5 s
font. Slice thickness 1 mm
(1 mm × 16 rows : No stack)
Note: To use T-mode, the laser imager must support
2048 × 2404 pixels for a 1 × 1 frame layout.
Reconstruction method TCOT+ with 0.5 mm SR
Reconstruction function FC70
• Auto filming in eXam Plans Scan field M
Phantom Catphan 500 phantom
(CTP528 module)
*1: Option
6
MPDCT0765EAF
7
Scanning Image display and processing
• Scanoscope function provides a projection image of the • Reconstructed images are automatically displayed
patient for high-precision advance planning of scan areas. according to the window settings preset in the eXam Plan.
• On the scanogram, the length of the scan area can be • The window save function allows the user to store an
adjusted over a range of up to 1780 mm*1 or 1480 mm*2. image with different window settings from those set in the
Because the images are reconstructed in real time, the eXam Plan.
scan can be aborted at any time. This allows the patient • Filter parameters can be customized through simple
exposure dose to be minimized. on-screen menu selections. These parameters include the
• The auto index function allows automatic incremental number of filtering passes, matrix size and filter
couch-top movement based on the slice positions coefficients.
determined through the scanogram. • Images can be rotated and reversed
• The eXam Plan function allows simple selection of pre- (right/left, top/bottom, black/white).
programmed scanning parameters for routine • The Multi-frame feature allows up to 15 images to be
examinations, maximizing patient throughput. retrieved and displayed simultaneously on the screen.
• Protocol comments can be saved in each eXam Plan, • The three-dimensional image display function allows color
providing interactive onscreen instructions for all studies, three-dimensional and MPR images in real-time to be
reducing the need to refer to a separate protocol book. generated from the volumetric scan data acquired by
• The Vari-area function allows the user to preselect a region helical scanning. This results in higher definition and
of interest for zooming using raw data, permitting image quality than images reconstructed from
immediate post-scan analysis. Zooming using raw data conventional single-slice scanning. This is because helical
yields higher resolution than enlarging an image that has scanning provides superior data continuity along the
already been reconstructed. patient axis compared with conventional scanning.
• Dynamic and rapid-sequence scan modes are provided.
• Multislice helical scan acquires raw data by rotating the Image filming
X-ray tube continuously while moving the patient • Filming of images can be performed manually or
continuously through the scanner. The acquired volume automatically from the console.
data can be used to reconstruct axial slices at any desired • Automatic filming sends an entire study to the laser printer.
position. This scan mode is best used for rapid patient Filming is performed in background mode so that other
scanning during a single breath-hold and for high- scanner and image processing functions can be performed
definition three-dimensional and MPR imaging. without interruption or delay.
• Real-time helical reconstruction mode makes it possible to • When T-mode is used, related information items displayed
observe the images being scanned in real time at a together with an image (surrounding the image, in a small
maximum of 12 frames per second. This mode shows any font) are displayed in the footer area using a larger font,
shift in the slice position in real time and allows the permitting not only easier reading but also simpler film
operator to check the scan field on the image, contrast management.
study timing, patient body motion, etc. The patient can Note: To use T-mode, the laser imager must support
therefore be released immediately after scanning. 2048 pixels × 2404 pixels for a 1 × 1 frame layout.
• The SUREStart function allows the operator to start helical
scanning at the timing of maximum enhancement in Patient throughput
contrast studies. SUREStart monitors the CT number in real Patient throughput and cost effectiveness were major
time to detect the arrival of contrast medium in the image. objectives in the design and production of the system.
When the CT number reaches the predefined threshold, • The system incorporates a 2 MHU or 3.5 MHU X-ray tube
dynamic volume scan or helical scan automatically starts. with a fast cooling rate of 337.8 kHU/min or 735 kHU/min
This technique ensures optimal contrast enhancement, in actual use.
independent of individual differences in blood flow • High-speed scans can be performed in as little as
velocity, and at the same time minimizes the dose of 0.75 second per scan.
contrast medium. • Real-time scanoscopy.
• Ease of operation is ensured through the incorporation of
Data processing a hybrid keyboard, mouse-driven menus and a large color
• A variety of reconstruction algorithms are available and LCD screen.
can be selected according to the anatomical region • The couch-top can be lowered to a position close to the
examined and the clinical objective of the study. These floor, facilitating patient transfer.
include algorithms for the abdomen, head, bones, lung, *1: For the long patient couch version
small structures, soft tissues, etc. *2: For the short patient couch version
8
MPDCT0765EAF
COMPLIANCE Grounding
Grounding must be provided in accordance with local
Council Directive 93/42/EEC and subsequent amendments
regulations for medically used electrical equipment.
IEC 60601-1: 2005+Amd.1: 2012
IEC 60601-1-2: 2007 Power distribution board
IEC 60601-1-2: 2014 For 3 phase, 400 V
IEC 60601-1-3: 2008+Amd.1: 2013 Ground resistance:
As per applicable
IEC 60601-1-6: 2010+Amd.1: 2013 legal requirements.
IEC 60601-1-9: 2007+Amd.1: 2013 NFB
100 A to 150 A
IEC 60601-2-28: 2010
IEC 60601-2-28: 2017
IEC 60601-2-44: 2009+Amd.1: 2012+Amd.2: 2016 Ground bar
Power distributor
IEC 60825-1: 2014
IEC 62304: 2006+Amd.1: 2015
IEC 62366: 2007+Amd.1: 2014 Ambient conditions
Temperature Humidity Heat generation
DIMENSIONS AND MASS Scan room
Gantry 20°C to 26°C 40% to 80% Approx.
Dimensions W × L × H Mass Tolerance: ±2°C No 9720 kJ/h*
Unit
mm (in) kg (lb) condensation 32070 kJ/h**
2050 × 960 × 1910 1300 Patient couch 20°C to 26°C 40% to 80% Approx.
Gantry Tolerance: ±2°C No 1080 kJ/h*
(80.7 × 37.8 × 75.2) (2866)
condensation 1800 kJ/h**
Long patient 630 × 2690 × 450 485 Control room
Patient couch version (24.8 × 105.9 × 17.7) (1069) Console 16°C to 28°C 40% to 80% Approx.
couch Short patient 630 × 2390 × 450 455 No 10800 kJ/h
condensation
couch version (24.8 × 94.1 × 17.7) (1003) Power distributor 16°C to 28°C 40% to 80% Approx.
430 × 690 × 645 75 No 2883 kJ/h*
Console (NAVI Box) condensation 6400 kJ/h**
(16.9 × 27.2 × 25.4) (165)
700 × 695 × 973 470 *: When scanning is not performed.
Power distributor **: When scanning is performed continuously at the maximum rated
(27.6 × 27.4 × 38.3) (1036) output of the system.
SITING REQUIREMENTS
Power requirements
• Phase: Three-phase
• Voltage: 380 V, 400 V, 415 V, 440 V, 460 V,
480 V*
• Frequency: 50 Hz or 60 Hz ±1 Hz
• Line capacity: 50 kVA
• Voltage fluctuation
due to load variation: Less than 5%
• Power voltage
fluctuation: Less than 10%**
*: For other voltages or in the event of excessive power fluctuation,
consult with your sales representative.
**: Represents the total voltage fluctuation due to load and power
variation.
9
Room layout example Installation requirements
1324 (1024*)
• The scanner emits radiation. X-ray shielding must be
provided around the scan room and the entrance in
accordance with all local requirements and regulations.
• The ceiling height should be at least 2500 mm when a
ceiling mounted contrast injector is installed.
5230 (4630*)
1 • Wiring pits and ducts are required for routing cables that
connect the various units.
2 Control room
7630 (7030*)
*1: Option
10
MPDCT0765EAF
Breaker box
Patient couch
Power 13
Gantry
distributor
17
17 Speaker
17 (Scan room)
9 Speaker
(Control room)
Console
11
OUTLINE DRAWINGS
12
MPDCT0765EAF
OUTLINE DRAWINGS
870
(34.3)
2050 340
(80.7) (13.4)
(3 80
(75.2)
1910
7)
7
0.
φ
(37.8)
960
960
(37.8)
Gantry
Unit: mm (in)
13
OUTLINE DRAWINGS
(18.5)
(24.8)
470
630
2690
(105.9)
588 (23.1)
STROKE
(40.9)
1038
(17.7)
450
Patient couch (for the 220 kg (485 lb) long patient couch version)
2390
(94.1)
588 (23.1)
STROKE
(40.9)
1038
(17.7)
450
Patient couch (for the 220 kg (485 lb) short patient couch version)
Unit: mm (in)
14
MPDCT0765EAF
OUTLINE DRAWINGS
(25.4)
645
430 690
(16.9) (27.2)
Console
Unit: mm (in)
15
MPDCT0765EAF
OUTLINE DRAWINGS
700
(27.6)
(38.3)
973
Anchor plate
866 695
(34.1) (27.4)
Power distributor
Unit: mm (in)
https://global.medical.canon
©Canon Medical Systems Corporation 2018-2021. All rights reserved.
Design and specifications are subject to change without notice.
Model number: TSX-037A MPDCT0765EAF 2021-02 CMSC/Produced in Japan
Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.
Aquilion, SEMAR, SURESubtraction, SUREStart, SUREExposure, Boost3D and Made for Life are trademarks of Canon Medical Systems Corporation.
Catphan is a registered trademark of The Phantom Laboratory.