CALIBRATION CERTIFICATE

Equipment Name : TRANSFORMER OIL BDV TEST KIT


Capacity : 0 - 100 KV, AC
Brand / Make : S.B.ELECTROTECH
Model No. : SBET - OBTS 100MO
Sl. No. : SBET/1056/23-24
Date Of Calibration : 01.04.2024
Calibration Valid upto : 31.03.2025
TESTS:-
1. Control Sequential operations such as ‘Mains ON-OFF’, ‘HT ON-OFF’, ‘DOOR INTERLOCK’,
‘ZERO START INTERLOCK’ and ‘Memory’ Circuit, etc. checked and found working satisfactorily.
2. Overload tripping circuit has been checked and found working properly.
3. Open Circuit voltage has been raised up-to 100 KV and kept for 1 Minutes, And the set withstood
the test successfully.
(i) Set Output KV meter Verification:-

Input : 0 - 230 V, AC Display : 0 - 100.0 KV Make : CABS ELECTRA


MODEL NO. : CE - 0102 V SL. NO. : C 93017
Reading On Corresponding Actual
Deviation % Deviation
Standard Reading on UUT Reading on UUT
Volt Kilo Volt Kilo Volt Kilo Volt %
46 20.0 20.1 0.10 0.50
92 40.0 40.2 0.20 0.50
138 60.0 60.3 0.30 0.50
184 80.0 80.5 0.50 0.63
230 100.0 100.6 0.60 0.60
(ii) Set Output KV Verification:-

Between Left HT Between Right HT


Terminal and Earth Terminal and Earth Total Total
Reading Reading on Reading Reading on Reading Reading on
On Unit On Unit on Unit %
Standard Under Standard Under Standard Under Deviation Deviation
Test Test Test
A B C D (A+C) (B+D)
KV KV KV KV KV KV KV %
10.09 10 10.09 10 20.18 20 -0.18 0.90
20.16 20 20.16 20 40.32 40 -0.32 0.80
30.24 30 30.24 30 60.48 60 -0.48 0.80
40.32 40 40.32 40 80.64 80 -0.64 0.80
50.42 50 * 50.41 50 * 100.84 100 -0.84 0.84
The value marked as ' * ' has been got from the calculations and not from observed value.

1
 Standards Used :-
Instruments Used to Measure the Parameters with Calibration Details :
Description Make/Model SL. No. Cal. Cert. No. Valid Up-to Traceability
Digital Multi Meter RISHABH / 2112005261 CAL/23-24/CC/087-5 01.05.2024 NEC,
RISH multi 16 S NASHIK
POTENTIAL S.B. SBET / 01 / CAL/23-24/CC/087-1 01.05.2024 NEC,
TRANSFORMER ELECTROTECH 13-14 NASHIK

 REMARKS:
1. Our Standard Instruments’ Calibration are traceable to ‘Nashik Engineering
Cluster’, Nashik, Which in turn traceable to National / International Laboratories.

For S.B.ELECTROTECH