NABL Accredited Calibration Services as ISO/IEC 77025:20L7

CALI BRATION CERTI F ICATE

SS/FF-20lDMN-03
Page 1 of 2
! Name and Address of Customer : M/s. Sarvashree
L-95, sth cross, 1st Main Kirloskar Colony,
1Sth Main Road, 3rd Stage, Basaveshwar Nagar,
Bengaluru, Karnataka - 560 079.

! Customer Reference
2.1 SRF No. : 1536 Dated . 22 June 2022
2.2 Cerlificate No. : 55/2211536-01 ULR No.: CC229122000002250F
2.3 Date of Calibration : 22 June 2022
2.4 Next Calibration Due : 21 June 2023

! Details of Device Under Calibration(DUCl.


3.1 Nomenclature : Digital Micrometer
3.2 Make : INSIZE Model :
3109-25A
3.3 Sl No : 19040590 lD No
: SS-MECH-DMN-012
3.4 No. of Pages :2 Range : 0 to 25 mm
3.5 Calibration Procedure No : SOP-DMN-O3 LC : 0.001 mm
3.6 DUC Condition : Satisfactory Location : Dimension lab
3.7 Calibration Done at : Dimension Lab, Sarvashree

4 Environmental Condition
Temperature 20t2 "C Humidity 40 to 60 % Rh

! Master lnstruments used for Calibration:

Traceable
Nomenclature Make / Model Sl. No. Validity
Certificate No.
Gauge Block Set (Grade 0) KCP / M4711 1 0675 2021t101010 08-10-2023

Optical Parallels Prism lndia / --- QSr/0535/22l03 15-03-2024

0 Other Details :
6.1. Kindly refers to Note(s) section mentioned over leaf.

Authorised

:::ru
(Galibration Engineer)
ffi
"Quality You Can Quantify"
NABL Accredited calibration services as per lso/lE
c 1,7025:2017

Certificate No. sst22t1536-01 U LR No. : CC229122000002250F Page 2 of 2

Name of the lnstrument : Digital Micrometer

Z Results : (All readings in mm )


Sl.No. Slip Gauge Readings UUC Readings Observed Deviation
1 0 0 000 0 000
2 5 5.000 0.000
3 10 10.000 0 000
4 15 14 999 -0 001
A
20 19 999 -0 001
6 25 24.999 -0.001

Flatness of Measuring Faces

Parallelism of Measuring Faces

I Uncertaintv Of Measurement Result :

Expanded Uncertainty of Micrometer (For k=2, 95.45 % C.L.) t 1.22 pm

! Reference :
9.1. Standard and Device Under Calibration readings are average of five repeated observations.
9.2. All the reported values have been reported in Sl units.

Au[horised By

l)^"*
t'tagarll m R
(Cali bration Engineer) (Lab ln-Charge)
*** End of Certificate ---