Calibration Certificate for Fluke Multimeter

The calibration certificate dated 04/04/2023 for a True RMS Multimeter (Model: 117, Serial No: 22941596) indicates that it was calibrated for KBR with a calibration expiry date of 05/04/2024. The document details various measurements, including DC voltage, AC voltage, DC current, AC current, frequency, and resistance, with corresponding reference values, measured values, deviations, and uncertainties. The instrument is noted to be in good condition and no adjustments were made during calibration.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
81 views3 pages

Calibration Certificate for Fluke Multimeter

The calibration certificate dated 04/04/2023 for a True RMS Multimeter (Model: 117, Serial No: 22941596) indicates that it was calibrated for KBR with a calibration expiry date of 05/04/2024. The document details various measurements, including DC voltage, AC voltage, DC current, AC current, frequency, and resistance, with corresponding reference values, measured values, deviations, and uncertainties. The instrument is noted to be in good condition and no adjustments were made during calibration.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd

CALIBRATION CERTIFICATE

Date 04/04/2023 Certificate No: LIF230406-32


:
Client Name : KBR
Unit under Calibration : True RMS Multimeter Job No. : LIF/JN/23/4310
Manufacturer : Fluke Calibration Date : 06/Apr/2023
Calibration Expiry
Model : 117 : 05/Apr/2024
Date
Indication Type : Digital Full Range/Resolution : Multi-Range
Reference Procedure Laboratory
: LIF/CP-02 : 22.4 °C
No. Temperature
Serial No. : 22941596 Laboratory Humidity : 40.9 %RH
Calibration Reference Standard
The UUC has been calibrated against:
Serial No of Calibration
Equipment Name Certificate No Validity Up to
Equipment Date
Fluke 5520 A CZ1018883 230618-10 18/06/2022 17/06/2023
DC Voltage

Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)


mV mV mV ±mV
-500.0 -499.998 -0.002 0.1
-100.0 -99.998 -0.002 0.1
-1.0 -1.000 0.000 0.1
0.0 0.000 0.000 0.1
1.0 1.000 0.000 0.1
2.0 2.000 0.000 0.1
10.0 10.000 0.000 0.1
50.0 50.000 0.000 0.1
100.0 100.000 0.000 0.1
200.0 199.998 0.002 0.1
400.0 399.998 0.002 0.1
500.0 499.998 0.002 0.1
V V V ±(V)
-900.0 -899.9980 -0.0020 0.6
-500.0 -499.9980 -0.0020 0.1
-100.0 -100.0000 0.0000 0.1
0.000 0.0000 0.0000 0.001
1.000 1.0000 0.0000 0.001
2.000 2.0000 0.0000 0.001
10.00 10.0000 0.0000 0.01
50.00 50.0001 -0.0001 0.01
100.0 100.0001 -0.0001 0.1
200.0 199.9998 0.0002 0.1
400.0 399.9998 0.0002 0.1
600.0 600.0000 0.0000 0.1
700.0 700.0000 0.0000 0.6
900.0 900.0000 0.0000 0.6
AC Voltage Certificate No: LIF230406-32

Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)


V V V ±V
180 V@50Hz 179.9 V 0.1 V 0.2
360 V@50Hz 359.9 V 0.1 V 0.1
540 V@50Hz 539.9 V 0.1 V 0.1
540 V@500Hz 539.9 V 0.1 V 0.1
900 V@50Hz 901.0 V 1.0 V 0.3
DC Current

Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)


A A A ±A
20 A 20.0 A 0.0 A 1.0
-20 A -20.1 A 0.1 A 1.0
540 A 545.2 A 5.2 A n.c
900 A 909.0 A 9.0 A n.c
AC Current

Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)


A A A ±A
20 A@50Hz 60.020 0.020 1.0
20 A@400Hz 599.905 0.095 1.0
540 A@50Hz 6.004 0.004 n.c
900 A@50Hz 59.997 0.003 n.c

Frequency
Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)
Hz Hz Hz ±Hz
40 Hz@10A 50.0 Hz 0.0 Hz 0.3
440 Hz@10A 440.0 Hz 0.0 Hz 0.1
Resistance

Reference Value UUC Measured Value Deviation (M-A) Uncertainty (U)


Ω Ω Ω ±Ω
4k Ω 2 wire 4.0k Ω 0.0k Ω 0.1
40k Ω 2 wire 40.0k Ω 0.0k Ω 0.1
400k Ω 2 wire 400.0k Ω 0.0k Ω 0.1

Note:
1. Ye
The instrument is "used" and in good condition on receiving.
s
2. No The instrument was new on receiving.
3. No The instrument is adjusted and post-adjustment values are reported.
4. Ye
No adjustment is carried out and measurements in this certificate are as received figures.
s

Comments:
The quoted Uncertainty refers only to the measured value and does not carry any implication regarding the
long-term stability of the instrument.
Calibrated By: Approved By
SHAHID ABBAS ABDUL
HASEEB