Multi-Angle Feature
Feature Overview and
Activation
General Overview
Available in Phased Array mode only
User designates any three angles (focal laws) within the S-
Scan range
The A-scans from each these three angles are overlaid on
top of each other in the A-scan
User can view all three A-scans at the same time
Each individual angle is color coded
Perfect when using conventional sizing methods requiring
evaluation at 45, 60 and 70 degrees!
Required Items
Transducer and Wedge
– 5L16-9.6X10-A10P-P-2.5-OM
probe
– SA10P-N55S wedge
Test Block
– EP1000-PABLOCK-1
Couplant
– D-12
Parameter Setup
Once unit has been properly calibrated…
– Set Range to 3in
– Set Gain to 18dB
– Set Angle to 45 degrees
Feature Activation
Under the Display Setup menu:
– Press P2 A-Scan Setup
– Change “AScan” to “Multi-Angle”
– Choose the 3 angles for the Multi-Scan AScans
– Press ESC to return to the home screen
Adjusting the focal law in fine adjustment will adjust the angle
over the S-Scan range in 1 degree increments
and so on…
A-Scan Colors
The angles in the A-Scan are color coded
– Red = Lowest number angle (e.g. 45)
– Yellow = Middle number angle (e.g. 60)
– Green = Highest number angle (e.g. 70)
White=45 Red=45
Yellow=60 White=51
Green=70 Yellow=60
Green=70
Multi-Angle-Single Toggle
It is possible to toggle between single A-scan (standard) mode and
Multi-Angle mode after activating the Multi-Angle feature
Even though 45
degrees is the active
angle, you can measure
signals at the other
selected angles!
Notch Detection (cont.)
Switch angle to 60 degrees
– The peaked up waveform that was
yellow will now become the white angle!