Professional Documents
Culture Documents
Central Medical Gas System
Central Medical Gas System
VENKAT THEJA.B
IHMR-B/PGDHM2011-2013/51
DIFFERENCE BETWEEN TRADITIONAL
MEDICAL GAS DELIVERY SYTEM AND
CENTRAL MEDICAL SUPPLY
Introduction
Gases administered to the patients are called as
Medical Gases
Centralized medical gas system provides as efficient,
economical and dependable medical life support
network that supplies :
a. Medical gases (oxygen and nitrous oxide)
b. Vacuum (suction)
c. Compressed air for operating rooms and ICUs
Introduction
Centralized gas and vacuum supply service is a
modern system of piped supply of medical gases from
a central storage area called manifold room to all
delivery points in hospital
It provides a very efficient, economical and highly
dependable life support service
Makes better patient care in all the areas of hospital
CENTRAL MEDICAL GAS
SYSTEM
Functions
Supply of right medical gases at right pressure
Supply of compressed air at right pressure
Supply of clinical vacuum at right pressure
Proper planned maintenance of all equipments,
including distribution network
Optimum level of cleanliness and pollution free
environment
Prevention of hazards such as fire, explosion or
contamination of gases supplied
Components
Source of supply: Central supply room with control
equipments and panels
Distribution system: piping
Point of use delivery connections: Suitable station
outlet valves and pendants
Monitoring and control equipment and alarms
Manifold Room
Consists of a cylinder manifold and a control panel
Manifold can be of 2 banks of 2 cylinders each or 2
banks of 20 cylinders each
Control panel: primary and secondary pressure
regulators; warning lamp
Pressure gauges
Manifold Room
Vacuum Unit
Vacuum pump with an electric motor
Cylindrical reservoir tank: stabilizes the pressure of the
pipeline system at all outlet points
Motor has switch for automatic start and stop
Vacuum pumps are duplexed
Manifold Room
Compressed air unit
Compressor with electric motor, after cooler, air receiver
and air dryer
Instantly provides compressed air
Compressor are duplexed
Manifold Room
Primary supply
Secondary supply when primary supply is exhausted
Primary and secondary provides normal operating
supply
When operating supply fails, a reserve supply takes
over
Manifold Room
In case of breakdown of control panel of oxygen and
nitrous oxide, an emergency kit ensures supply of gas
through the pipes
Emergency kit comprises a regulator and high pressure
tubing to a bulk cylinder and the gas is fed directly to
the pipeline through a service outlet
In case the vacuum and air supply system fails, a
standby motor-cum-pump is used to provide
uninterrupted supply
Manifold Room
Half of the total daily consumption of oxygen and
nitrous oxide is kept in the manifold room as reserve
Oxygen and nitrous oxide should be stored seperately
from flammable gases and liquids
Storage location should be free of combustible
materials
If exceeding 2,000 cubic feet storage should be outside
the building
Must have generator backup, adequate ventilation,
lighting and telephone communication
Advantages of Centralized Medical
Gas System
For the patients
No distressing sight of oxygen cylinders at the bedside
Elimination of irritating noise from movement of
cylinders
Protection of contamination due to movement of
cylinders
Uninterrupted and clean supply at desired location
Advantages of Centralized Medical
Gas System
For Hospital Staff
Instant availability of gas on taps
Clean, safe and reliable delivery of gases
Continuous flow of gases when and where required
Minimal accident hazards due to mishandling of
cylinders
Advantages of Centralized Medical
Gas System
For Hospital Administrator
Easy purchase of gases in bulk quantities at favourable
terms
Economy on purchase of cylinders
Fewer breakages
Rationalization in ordering, storing and transporting a
wide variety and size of gas cylinders
Minimum damage to building due to handling of
cylinders
Oxygen
Increased demand of oxygen
May be supplied from a bulk tank or from cylinders
Can be stored as gas or liquid
Oxygen containers may be stationary or movable
Oxygen and Nitrous Oxide
Bulk oxygen system: assemblage of equipment such as
oxygen containers, pressure regulators, safety devices,
vaporizer, manifolds and interconnecting piping with a
storage capacity of more than 20,000 cu ft of oxygen
including reserves
The system terminates at the point where oxygen first
enters the supply line at service pressure
Vacuum
Used for patient draining, aspiration and suction
Also used in hospital laboratory
Distributed through network of pipes
Central vacuum supply system with control
equipments, an alarm and piping
The piping terminates with outlet valves at user points
Vacuum
Central vacuum consists of two (or more) vacuum
pumps that operate either simultaneously or alternately
Each pump should be capable of maintaining 75% of
calculated demand during peak time
The pumps should alternate automatically
Pumps should be equipped with a motor starting
device and overload protection
Failure of one vacuum should not affect other
OUTLETS COLOR CODING
Compressed Air
Compressed air is used for both medical and non-
medical puposes
Medical use – Laboratory work, inhalation therapy
equipment and powering of surgical tools
Non-medical use – maintain tools and equipments and
in engineering department
Compressed Air
Should be free of dust and moisture
Done by means of oil filters, dust filters, moisture
seperator and the dryer at equipment level
Separate dryer for air compressor system
If compressed air not used properly, it can damage
equipment and contaminate chemicals, food and drugs
Pipes
Should be seamless type, non-ferrous, non-arsenic, of
high quality copper tubing
Should be protected against physical damage and
corrosion
Exposed oxygen pipelines should not be installed in
areas like kitchen, laundry and rooms where
combustible materials are stored
Colour coded as per the gas content
Pipes
Before pipes are erected, all pipes, tubes and fittings
should be cleaned thoroughly and washed with
tetrachloride
Pipes should be blown clear using oil-free air or
nitrogen
When whole system is in place, it should be subjected
to a test pressure of 150 psi or 1 ½ times the working
pressure for 24 hours to check, if it can withstand the
pressure as well as leakages
This test, called pressure lock is done section by section
Terminal units
Gas outlets / vacuum inlet units
Connected to source equipment through the
distribution system
Located at actual usage point
Outlets should be gas specific and non-interchangeable
Fitted with locking system and non-return valves
Alarm system
Audio-visual
Monitoring of pressure
Meant to warn maintenance dept
Location and space
Ground floor in rear part
Near maintenance area
Seperate enclosure for full cylinders, empty cylinders,
compressed air, vacuum supply
No fire hazard
Compresor unit in sound proof area
Liquid oxygen area shall be cordoned off
Centralized Medical Gas System
Staffing
No specific norms
Total 6 technicians (including supervisors and 25%
leave relief)
Under chief engineer
Quality Indicators
Number of complaints of malfunctioning, from the
users
Correctness of pressure level in the oxygen, nitrous
oxide and compressed air pipeline
Correctness of negative pressure in the vacuum line
Bacteriological culture of the gases and the air
delivered
Leakage at the outlet points
Functioning of audio-visual alarm system
Thank You