Professional Documents
Culture Documents
Definition of automation
Immunoassay
Fluorometric or Fluorescent polarization
Chemiluminescent, Nephelometric
Sample Pipettor and Pump
Probe or pipettor
Wash solution
Detector
Manual Spectrophotometer
Automated Chemistry
Analyzer
Advantage & Disadvantages of Manual Methods
Disadvantages
Slower turnaround time (TAT)
Requires more hands-on Technologist time
Less precision and accuracy in Quality Control
Tests may be less sensitive
Often require larger sample volumes
Increased safety risk to Techs (biological and reagent)
Advantages
May be less expensive and easier to operate
Less skilled manpower 20
Advantage and Disadvantages of Automated
Methods
Advantages
Quicker TAT
Less handling of biological specimens by Techs
Greater precision and accuracy in Quality Control
Greater sensitivity
Require smaller sample size
Can run and report many more test results in shorter
period of time
Disadvantages
May be more expensive
More spare parts need to be available
Requires higher skilled technologists
Semi-Automation versus Automation
Chemistry Analyzers
In Semi-automation there are steps performed by the
technologist:
prepares sample for testing
loads sample cup onto analyzer
manually inputs Patient ID and Test(s) Requested into
instrument via keyboard
In full automation:
specimens are placed directly on the analyzer
Barcode reader inputs Patient ID and test request
Most semi-automated and automated instruments are:
*Random-Access Analyzers
Can run 50 or more tests per hour
Technologist has more time for other lab responsibilities
(walk- away)
Reports from the instrument are printed in “ready to send
format” to the patient’s hospital chart or hand deliver to
patient or clinician
24
How to Unpack an Instrument (The instrument
is here, now what should I do?)
A new automated instrument has
just been received in the
laboratory.
The technologist must now set up,
calibrate and validate the
instrument, comparing it to the
manual method.
Automation Analyzer Vendors or Engineers
Your instrument vendor (engineer)
may help with the installation,
calibration and validation of your
new instrument
Engineers (vendors) are an
important resource:
Keep their contact information
taped to your instruments
Initial Instrument Setup
Pre-installation Steps
Out-of-the box steps
Establish a working procedure
Pre-installation Requirements
Check for the following items:
Proper electrical wiring and voltage, and uninterrupted
power
UPS surge protectors and /or stabilizers
Generators should be available
Temperature controlled
Adequate distilled water supply
Check for the following items:
Proper waste disposal capability
Adequate air circulation / flow around instrument
Dust free environment (cover when not in use)
Stable bench top
Before Beginning the Instrument Validation
Turn on the power
Permit instrument to stabilise/equilibrate
Let all components reach proper temperature
Set in any parameters that may be required
Ranges, Temperatures
Initial Instrument Calibration
Follow manufacturers recommendation for reagent and control
preparation
Place calibrator(s) on instrument in proper order sequence
Run prepared calibrator(s) on both automated and manual
methods
3-5 standards are analyzed for analyte
Absorbance is measured
Absorbance versus concentration graph is plotted
Linearity is determined
Concentration curve is stored in analyzer
QC samples verify concentration curve
Calibration Curve
1.00
absorbance
0.80
0.60
0.40
0.20
0.00
0 20 40 60 80 100 120
concentration
Ongoing Automation Analyzer Calibration
Determine QC results by standard curve or factor
If controls within stated value for analyzer and
methodology, then move to instrument validation
Test calibration will be repeated
When there is a new lot number of reagent or
In accordance with manufacturer’s recommendations
On more frequent basis as indicated by quality control
results
Evaluation of a New Automation Method
1. Precision
2. Comparison
3. Linearity
4. Reference range
5. Interferences
Establishing Reference Ranges
Reference ranges should be established for manual and
automated methods
Materials used
20 healthy patient samples previously obtained are run
over a period of at least 3 consecutive days
Calculation of Reference Ranges
51