Professional Documents
Culture Documents
Gas Chromatography
A.) Introduction:
Gas Chromatography (GC) - chromatographic technique where the mobile phase is a gas.
GC is currently one of the most popular methods for separating and analyzing compounds. This is due
to its high resolution, low limits of detection, speed, accuracy and reproducibility.
GC can be applied to the separation of any compound that is either naturally volatile (i.e., readily goes
into the gas phase) or can be converted to a volatile derivative. This makes GC useful in the
separation of a large number of small organic and inorganic compounds.
B.) Equipment:
An inert gas such as helium is passed through the column as a carrier gas and is the
moving phase. A sample is injected into a port which is much hotter than the column and
is vaporized. The gaseous sample mixes with the helium gas and begins to travel with
the carrier gas through the column. As the different compounds in the sample have
varying solubility in the column liquid and as these compounds cool a bit, they are
deposited on the column support. However, the column is still hot enough to vaporize the
compounds and they will do so but at different rates since they have different boiling
points. The process is repeated many, many times along the column. Eventually the
components of the injected sample are separated and come off of the column at different
times (called "retention times").
5
There is a detector at the end of the column which signals the change in the nature of
the gas flowing out of the column. Recall that helium is the carrier gas and will have a
specific thermal conductivity, for example. Other compounds have their own thermal
conductivities. The elution of a compound other than helium will cause a change in
conductivity and that change is converted to an electrical signal. The detector, in turn,
sends a signal to a strip chart recorder or to a computer. Detectors come in several
varieties, for example, thermal detectors, flame-ionization and electron capture
detectors.
6 Theory
Water + I2 Water + I2
CCl4 CCl4 + I2
CCl4
+ I2
As the gas moves the solute (analyte) through and over the stationary phase, the
solute will be in equilibrium with the gas and the solid phase. Since there is a
mobile phase, the separation will appear as a chromatogram showing the
separation of the analytes.
GC Theory
12
As the gas moves the solute (analyte) through and over the stationary phase, the
solute will be in equilibrium with the gas and the solid phase. Since there is a
mobile phase, the separation will appear as a chromatogram showing the
separation of the analytes.
Analyte
To detector
Column + packing
Time
Separations
13
To detector
tR
to t’ R
k’ = (tr - t0) / t0
k’ = (t’r ) / t0
.
Inject
Unretained
Solute
15 GC Process
Column Efficiency (N)
16
Solutes are placed on an GC column in a narrow band
• Each solute band spreads as it moves through the column due
to diffusion and mass transfer affects
• The later eluting bands will spread more
to
t1
t2
Band spreading eventually causes peaks to merge into the baseline. We want to
minimize band spreading as much as possible.
17 Chromatogram nomenclature
18 GC Peak parameters
19 Typical separation
20 GC General
The chromatogram shows the order of elution (order of
components coming off the column), the time of elution
(retention time), and the relative amounts of the components
in the mixture.
pentane 36
hexane 69
cyclohexane 80
iso-octane 99
toluene 110
4-methyl-2-pentanone 117
octane 126
22 GC Chromatogram
• The observed elution pattern appears below. Notice the
reversed elution of toluene and 4-methyl-2-pentanone.
23 Column temperature
For precise work, column temperature must be controlled to within tenths of a
degree. The optimum column temperature is dependant upon the boiling point of
the sample. As a rule of thumb, a temperature slightly above the average boiling
point of the sample results in an elution time of 2 - 30 minutes. Minimal
temperatures give good resolution, but increase elution times. If a sample has a
wide boiling range, then temperature programming can be useful. The column
temperature is increased (either continuously or in steps) as separation proceeds.
24 Mobile Phase (gas)
25 GC Under the hood
26 GC Column and Oven
27 Typical GC (dual column)
28 Sample Injections
Next, the sample injection system. Here it is important that
the sample be injected onto the column as a plug and of a
suitable size. Also, the injector should provide consistent and
reproducible injections.
The micro-syringe is used to load the sample onto the
column. The syringe should be clean and accurate and gas
tight. The syringe is injected through a rubber septum. The
septum should be replaced after many injections to insure gas
tightness onto the column. An auto sampler can be used to
inject the samples. Typical volumes range from 0.2 to 20 Ls.
With capillary columns it is necessary to use a splitter. (See
Figure.) A suitable solvent is also necessary for the proper
separations and injections.
29 GC Injector
30 Carrier Gas
This is the mobile phase and should be pure gas so as not to react with the column or
analyte. Gas is usually He, Ar, N2, or H2. Choice will depend on the type of detector used.
He and H2 give better resolution (smaller plate height) than N2. Pressure is also important
and as expected the system comes with regulators. Can you find where in GC equations
that are dependent on pressure?
31 Columns
There are two general types of column, packed and capillary (also known as open tubular).
Packed columns contain a finely divided, inert, solid support material (commonly based on
diatomaceous earth) coated with liquid stationary phase. Most packed columns are 1.5 -
10m in length and have an internal diameter of 2 - 4mm.
Capillary columns have an internal diameter of a few tenths of a millimeter. They can be
one of two types; wall-coated open tubular (WCOT) or support-coated open tubular
(SCOT). Wall-coated columns consist of a capillary tube whose walls are coated with
liquid stationary phase. In support-coated columns, the inner wall of the capillary is lined
with a thin layer of support material such as diatomaceous earth, onto which the stationary
phase has been adsorbed. SCOT columns are generally less efficient than WCOT columns.
Both types of capillary column are more efficient than packed columns.
In 1979, a new type of WCOT column was devised - the Fused Silica Open Tubular
(FSOT) column;
34 FSOT column
35 Capillary column
C.) Mobile Phase:
36 GC separates solutes based on their different interactions with the mobile and stationary
phases.
- solute’s retention is determined mostly by its vapor pressure and volatility
- solute’s retention is controlled by its interaction with the stationary phase
- gas mobile phase has much lower density
‚decreased chance for interacting with solute
‚increased chance that solid or liquid stationary phase interacts with solute
C.) Mobile Phase:
37 Carrier gas – main purpose of the gas in GC is to move the solutes along the column, mobile
phase is often referred to as carrier gas.
Stationary phase in GC is the main factor determining the selectivity and retention
39 of solutes.
- same material is used as both the stationary phase and support material
- common adsorbents include:
‚ alumina
‚ molecular sieves (crystalline aluminosilicates [zeolites] and clay)
‚ silica
‚ active carbon
advantages:
- long column lifetimes
- ability to retain and separate some compounds not easily resolved by other
GC methods
‚ geometrical isomers
‚ permanent gases
disadvantage:
- very strong retention of low volatility or polar solutes
- catalytic changes that can occur on GSC supports
- GSC supports have a range of chemical and physical environments
‚ different strength retention sites
‚ non-symmetrical peaks
‚ variable retention times
2.) Gas-liquid chromatography (GLC)
- stationary phase is some liquid coated on a solid support
41 - over 400 liquid stationary phases available for GLC
‚ many stationary phases are very similar in terms of their retention properties
- material range from polymers (polysiloxanes, polyesters, polyethylene glycols) to
fluorocarbons, molten salts and liquid crystals
Based on polarity, of the 400 phases available only 6-12 are needed for most separations.
The routinely recommended phases are listed below:
OV-17 50% Phenyl methyl 375 119 158 162 243 202
OV-210 50% Trifluoropropyl 270 146 238 358 468 310 Higher the number the
OV-225 25% Cyanopropyl- 250 238 369 338 492 386 higher the absorption.
25% phenyl
Silar-SCP 50% Cyanopropyl- 275 319 495 446 637 531
50% phenyl
SP-2340 75% Cyanopropyl 275 520 757 659 942 804
disadvantage:
- liquid may slowly bleed off with time
‚ especially if high temperatures are used
‚ contribute to background
‚ change characteristics of the column with time
3.) Bonded-Phase Gas chromatography
- covalently attach stationary phase to the solid support material
43 - avoids column bleeding in GLC
- bonded phases are prepared by reacting the desired phase with the surface of a silica-
based support
reactions form an Si-O-Si bond between the stationary phase and support
or
reactions form an Si-C-C-Si bond between the stationary phase and support
- many bonded phases exist, but most separations can be formed with the following
commonly recommended bonded-phases:
‚ Dimethylpolysiloxane
‚ Methyl(phenyl)polysiloxane
‚ Polyethylene glycol (Carbowax 20M)
‚ Trifluoropropylpolysiloxane
‚ Cyanopropylpolysiloxane
O Si O Si O Si
HO C C O H
advantages:
- more stable than coated liquid phases
- can be placed on support with thinner and more uniform thickness than
liquid phases
E.) Support Material:
45 A common problem to all chromatographic techniques is that in any one sample there may
be many solutes present, each retained by the column to a different degree:
Gradient elution - change column condition with time which changes retention of solutes to
overcome general elution problem
Temperature Programming – changing the temperature on the column with time to simulate
gradient elution in GC since a solute’s retention in GC is related to its volatility.
46
Comparison of a GC separation using isothermal conditions and temperature programming:
ISOTHERMAL
Column temp. 120oC
Programmed temp.
(30oC to 180oC) (5o/min)
The choice of detector will depend on the analyte and how the GC method is being
used (i.e., analytical or preparative scale)
Process
- measures a bulk property of the mobile phase leaving the column.
- measures ability to conduct heat away from a hot-wire (i.e., thermal conductivity)
- thermal conductivity changes with presence of other components in the mobile phase
Design
- based on electronic circuit known as a Wheatstone bridge.
49 - circuit consists of an arrangement of four resistors with a fixed current applied to them.
- thermal conductivity changes with presence of other components in the mobile phase.
- the voltage between points (+) and (-) will be zero as long as the resistances in the
different arms of the circuit are properly balanced
advantages:
- truly universal detector
‚ applicable to the detection of any compound in GC
- non-destructive
‚ useful for detecting compounds from preparative-scale columns
‚ useful in combination with other types of GC detectors
disadvantage:
- detect mobile phase impurities
- sensitive to changes in flow-rates
- limit of detection
‚ ~ 10-7 M
‚ much higher then other GC detectors
2.) Flame Ionization Detector (FID)
- most common type of GC detector
51
- “universal” detector capable of measuring the presence of almost any organic and many
inorganic compound
Process
- measures the production of ions when
a solute is burned in a flame.
52
advantages:
disadvantage:
- destructive detector
3.) Nitrogen-Phosphorus Detector (NPD)
53
- used for detecting nitrogen- or phosphorus containing compounds
- also known as alkali flame ionization detector or thermionic detector
Process
- same basic principal as FID
Alkali Bead
- Like FID, does not detect common mobile phase impurities or carrier gases
- limit of detection: NPD is 500x better than FID in detecting nitrogen- and
phosphorus- containing compounds
disadvantage:
- destructive detector
55 - radiation-based detector
- selective for compounds containing electronegative atoms, such as halogens
Process
- based on the capture of electrons by
electronegative atoms in a molecule
- selective for halogen- (I, Br, Cl, F), nitro-, and sulfur-containing compounds
GC quality control analysis of food products can confirm the presence and quantities of the
analytes For example, fruits, fruit derived foodstuffs, vegetables and soft drinks, tea and
coffee, were analyzed for their polybasic and hydroxy acid contents (citric, maleic acids) as
TMS derivatives.
All food and beverage products on sale today must be carefully assayed for contamination
with pesticides, herbicides and many other materials that are considered a health risk. The
analysis of food involves separating and identifying very complex mixtures, the components
of which are present at very low concentrations. GC is the ideal technique for use in food
and beverage assays and tests. Furthermore, the origin of many herbs and spices can often
be identified from the peak pattern of the chromatograms from their head space analysis.
61 Food and Cancer
There are still numerous GC applications involving both quantitative and qualitative
identification of the active components and possible contaminants, adulterants or
characteristic features which may indicate the source of the particular sample. Forensic
analysis frequently users GC to characterize drugs of abuse, in some cases the characteristic
chromatographic fingerprint gives an indication of the source of manufacture of the sample
or worldwide source of a vegetable material such as cannabis.
Analytical procedures, chromatographic methods and retention data are published for over
600 drugs, poisons and metabolites. These data are extremely useful for forensic work and
in hospital pathology laboratories to assist the identification of drugs.
63 Pyrolysis gas chromatography
Combustion of fossil fuel, disposal of waste materials and products, treatment of crops with
pesticides and herbicides have all contributed to the problems. Technological developments
have enabled man to study these problems and realize that even trace quantities of pollutants
can gave detrimental effects on health and on the stability of the environment.
There is a vast amount of literature on the use of GC for studying a wide variety of these
problems.