Professional Documents
Culture Documents
NOTES 2O18
Instrumentation, Ionization methods, Mass analyzers,
qualitative and quantitative analysis, Fourier
transform Mass spectrometry, tandem mass spectroscopy,
secondary ion mass spectroscopy.
Coupled
techniques such as GC-MS, CE-MS.
Basic theory of mass spectrometry
128
80
60
40
20 102
64 77 c
c c
51 b
b b b
0
50 100 150
Mass to charge ratio m/z
Mass spectrum of CO2. Note that the molecular The mass spectrum of naphthalene with
ion appears at m/z = 44 (C = 12, O = 16). Frag- electron impact ionization by 70 eV electrons.
ment ions appear at m/z values of 28, 16, and a, molecular ion and base peak(C10H+8, 100%);
12. These correspond to CO+, O+, and C+, resp- , b, 13C isotope peak;
ectively. c, fragment ion peaks.
Introduction to mass spectrometry
Mass spectrometers use the difference in mass-to-charge ratio (m/e) of ionized
atoms or molecules to separate them from each other. Mass spectrometry is
therefore useful for quantitation of atoms or molecules and also for determining
chemical and structural information about molecules. Molecules have distinctive
fragmentation patterns that provide structural information to identify structural
components.
All commonly used mass analyzers use electric and magnetic fields to apply a
force on charged particles (ions). The relationship between force, mass, and the
applied fields can be summarized in Newton's second law and the Lorentz force
law:
F = ma (Newton's second law)
F= e(E+ v × B) (Lorentz force law)
where F is the force applied to the ion, m is the mass of the ion, a is the
acceleration, e is the ionic charge, E is the electric field , v B is the vector cross
product of the ion velocity and the applied magnetic field.
From Newton's second law, it is apparent that the force causes an acceleration that
is mass-dependent, and the Lorentz force law tells us that the applied force is also
dependent on the ionic charge. Therefore, it should be understood that mass
spectrometers separate ions according to their mass-to-charge ratio (m/z) rather
than by their mass alone.
The general operation of a mass spectrometer is:
• create gas-phase ions
• separate the ions in space or time based on their mass-to-charge ratio
• measure the quantity of ions of each mass-to-charge ratio
The ion separation power of a mass spectrometer is described by the resolution R,
which is usually defined as:
where m is the ion mass and delta m is the difference in mass between two
resolvable peaks in a mass spectrum.
Mass Spectrometry
Theory
In mass spectrometry, a small sample of a chemical compound is vaporized,
bombarded with high energy electrons to ionize the sample, and the ions
produced are detected based on the charge to mass ratio of the ions.
Sample Data
Interfacing Ion Source Analyzer Detector
introduction System
EI Quadrupole
GC Library
CI Sector
LC FAB TOF
SFC MALDI FT-ICR
IC ESI Ion Trap
CE Vacuum System:
APCI
DLI DP or TMP
Thermospray
ICP Particle Beam 10–5 ~ 10–8 torr
Solid probe
MS/MS – Q/TOF, TOF/TOF
HRMS
LRMS
Schematic diagram of mass spectrometer.
Vacuum system
Vacuum System:
- Pumps must be maintained ; Ultra-clean ; Vacuum regulation and
maintenance
MS Configuration for Ionization
EI
High vacuum
- EI (electron impact)
- CI (chemical ionization)
- FAB (fast atom bombardment)
- ESI (electrospray ionization)
- MALDI (matrix assisted laser desorption ionization)
- APCI (atmospheric pressure chemical ionization)
Sample introduction / ionization method:
M + e– M+· + e – + e –
70 eV Molecular ion ~ 55 eV 0.1 eV
7 × 12C = 84 7 × 1H = 7
1 × 14N = 14
1 × 16O = 16
Total 121
This type of calculation is called the unit mass molecular ion. Please refer to the
spectrum for benzamide.
N,N-diethylmethylamine has one nitrogen and its molecular mass (m/z = 87) is an odd number.
A majority of the fragment ions have even-numbered masses (ions at m/z = 30, 42, 56 & 58 are
not labeled), and are even-electron nitrogen cations.
Electron ionization (70 eV) mass spectra of molecular ion region of benzene
(C6H6) and biphenyl (C12H10).
Intensity of M+1 relative to molecular ion for C nHm :
Intensity = n × 1.08% + m × 0.012%
Contribution from 13C Contribution from 2H
Methyl Bromide: An example of how isotopes can aid in
peak identification.
The ratio of peaks containing 79Br and its isotope 81Br (100/98) confirms
the presence of bromine in the compound.
Relative Isotope Abundance of Common Elements:
Carbon 12
C 100 13
C 1.11
Hydrogen H
1
100 2
H .016
Nitrogen 14
N 100 15
N .38
Oxygen 16
O 100 17
O .04 18
O .20
Sulfur 32
S 100 33
S .78 34
S 4.40
Chlorine 35
Cl 100 37
Cl 32.5
Bromine 79
Br 100 81
Br 98.0
Electron ionization mass spectrum Mass spectrum showing natural isotopes
(70 eV) of 1-bromobutane. of Pb observed as an impurity of brass.
Molecular Mass and Nominal Mass
The unit of atomic mass is the dalton, Da, defined as 1/12 of the mass of 12C.
Mass specrometrists perfer the symbol “u” for “unified atomic mass unit”.
“Da” and “u” are synonymous.
Atomic mass is the weighted average of the masses of the isotopes of an
element.
Ex. Bromine 79
Br 78.91834 Da 50.69%
81
Br 80.91629 Da 49.31%
Br = 78.91834 × 0.5069 + 80.91629 × 0.4931
= 79.904 Da
The molecular Mass of a molecule or ion is the sum of atomic masses listed in
a periodic table.
Ex. C2H5Br = 12.0107 × 2 + 1.00794 × 5 + 79.904 × 1 = 108.965
The nominal mass of a molecule or ion is the the integer mass of the species
with the most abundant isotope of each of the constituent atoms.
Ex. C2H5Br = 12 × 2 + 1 × 5 + 79 × 1 = 108
Chemical ionization (CI)
CI uses a reagent ion to react with the analyte molecules to form ions by either
a proton or hydride transfer:
MH + C2H5+ M+ + C2H6
mass analysis:
A process by which a mixture of ionic or neutral species is identified according
to the mass- to- charge (m/ z) ratios (ions) or their aggregate atomic masses
(neutrals). The analysis may be qualitative and/ or quantitative. [IUPAC
Compendium]
Common mass analyzer designs:
Quadrupole Mass Spectrometer
Introduction
A quadrupole mass filter consists of four parallel metal rods arranged as in the
figure below. Two opposite rods have an applied potential of (U+Vcos(wt))
and the other two rods have a potential of -(U+Vcos(wt)), where U is a dc
voltage and Vcos(wt) is an ac voltage. The applied voltages affect the
trajectory of ions traveling down the flight path centered between the four
rods. For given dc and ac voltages, only ions of a certain mass-to-charge ratio
pass through the quadrupole filter and all other ions are thrown out of their
original path. A mass spectrum is obtained by monitoring the ions passing
through the quadrupole filter as the voltages on the rods are varied. There are
two methods: varying w and holding U and V constant, or varying U and V
(U/V) fixed for a constant w.
Instrumentation
Quadrupole mass spectrometers consist of an ion source, ion optics to
accelerate and focus the ions through an aperture into the quadrupole filter,
the quadrupole filter itself with control voltage supplies, an exit aperture, an
ion detector and electronics, and a high-vacuum system.
where m is the mass of the ion, v is it's velocity, e is the charge of the ion and V is the
applied voltage of the ion optics.
The ions enter the flight tube between the poles of a magnet and are deflected by the
magnetic field, H. Only ions of mass-to-charge ratio that have equal centrifugal and
centripetal forces pass through the flight tube:
mv2 / r = Hev
This equation shows that the m/e of the ions that reach the detector can be varied by
changing either H or V.
Magnetic sector mass spectrometer.
Instrumentation
Introduction
The ion-trap mass spectrometer uses three electrodes to trap ions in a small
volume. The mass analyzer consists of a ring electrode separating two
hemispherical electrodes. A mass spectrum is obtained by changing the
electrode voltages to eject the ions from the trap. The advantages of the ion-
trap mass spectrometer include compact size, and the ability to trap and
accumulate ions to increase the signal-to-noise ratio of a measurement.
Ion Trap Analysis
Theory
The quadrupole ion trap mass analyser (see figure) consists of three hyperbolic
electrodes: the ring electrode, the entrance endcap electrode and the exit endcap
electrode. These electrodes form a cavity in which it is possible to trap and analyse
ions. Both endcap electrodes have a small hole in their centres through which the
ions can travel. The ring electrode is located halfway between the two endcap
electrodes.
Ions produced from the source enter the trap through the inlet focussing system
and the entrance endcap electrode. Various voltages are applied to the electrodes to
trap and eject ions according to their mass-to-charge ratios. The ring electrode RF
potential, an a.c. potential of constant frequency and variable amplitude, is applied
to the ring electrode to produce a 3D quadrupolar potential field within the
trapping cavity. This will trap ions in a stable oscillating trajectory confined within
the trapping cell. The nature of the trajectory is dependent on the trapping potential
and the mass-to-charge ratio of the ions. During detection, the electrode system
potentials are altered to produce instabilities in the ion trajectories and thus eject
the ions in the axial direction. The ions are ejected in order of increasing mass-to-
charge ratio, focussed by the exit lens and detected by the ion detector system.
Ion-trap mass spectrometer.
(a) Mass analyzer consists of two end caps (left and right) and central ring
electrode. (b) Schematic diagram.
Time-of-Flight Mass Spectrometry (TOF-MS)
Introduction
Theory
K.E. = qV
1/2 mv2 = qV
v = (2qV/m)1/2
The transit time (t) through the drift tube is L/V where L is the length of the
drift tube.
t = L / (2V/m/q)1/2
Instrumentation
This schematic shows ablation of ions from a solid sample with a pulsed
laser. The reflectron is a series of rings or grids that act as an ion mirror.
This mirror compensates for the spread in kinetic energies of the ions as
they enter the drift region and improves the resolution of the instrument.
The output of an ion detector is displayed on an oscilloscope as a function
of time to produce the mass spectrum.
Equation 2:
Equation 3:
Where:
• F = Lorentz force (observed by the incident ion)
• B0 = ICR magnetic field
http://www-methods.ch.cam.ac.uk/meth/ms/theory/fticr.html
Ion Detection
Once the ion passes through the mass analyzer it is then detected by the ion
detector, the final element of the mass spectrometer. The detector allows a
mass spectrometer to generate a signal current from incident ions by
generating secondary electrons, which are further amplified. Alternatively,
some detectors operate by inducing a current generated by a moving charge.
Among the detectors described, the electron multiplier and scintillation
counter are the most commonly used and convert the kinetic energy of
incident ions into a cascade of secondary electrons
- Faraday cup
- Electron Multiplier
- High-energy dynodes with electron multiplier
- Array
- FT-MS
Tandem Mass Spectrometry
In contrast to electron ionization (EI) which produces many fragment ions, the
new ionization techniques are relatively gentle and do not produce a significant
amount of fragment ions. To obtain more information on the molecular ions
generated in the electrospray ionization and MALDI ionization sources, it has
been necessary to apply techniques such as tandem mass spectrometry (MS/MS)
to induce fragmentation. Tandem mass spectrometry (abbreviated MSn - where n
refers to the number of generations of fragment ions being analyzed) allows one
to induce fragmentation and mass analyze the fragment ions. This is
accomplished by collisionally generating fragments from a selected ion and then
mass analyzing the fragment ions. Fragmentation can be achieved by inducing
ion/molecule collisions by a process known as collision-induced dissociation
(CID) (also known as collision-activated dissociation (CAD)). Collision-induced
dissociation is accomplished by selecting an ion of interest with a mass analyzer
and introducing that ion into a collision cell. The selected ion then collides with a
collision gas (typically argon or helium) resulting in fragmentation. The
fragments are then analyzed to obtain a fragment ion spectrum. The abbreviation
MSn is applied to processes which analyze beyond the initial ions (MS) to the
fragment ions (MS2) and subsequent generations of fragment ions (MS3, MS4 and
…). Tandem mass analysis is primarily used to obtain structural information.
Multiple Mass Spectrometry ; MS/MS/MS:
Provides even greater certainty of identification and additional characterization
information than electrospray ionization/ tandem mass spectrometry. Fully
automated. [CHI Proteomics report]
When more than two stages are involved, the technique is called multi-
dimensional MS (MS n where n indicates the number of stages).