What we do –
develop, manufacture and market medical devices that cater to
hospital needs around the world
How we do it –
vertically integrated business model; R&D oriented operation; A
value player
Who we are today –
a Chinese market leader with growing international presence
Negative pressure
Air Pump
Positive
Pressure
Three segments :
PS2 PS4 PS3 PS1
250KPa
160KPa
40KPa
70KPa
Air Filter Drier
T376
Drier/Filter GF1
T431
T325 40KPa
T376
RGV2 FCM Pressure,
T427
T326 T428 constant
160KPa Regulator
70KPa
T379
GF1
RGV3 Other DP
DIL Chamber Pressure
T380 WC1 Chamber Pressure
70KPa Regulator
HGB Bubble pressure
T368
RGV6 250KPa
Air filter
Drier
Outlet
Inlet:
Air Filter
Reagent Chamber
Diaphragm pump
Cylinder
Pinch Valve
300
250
200
150 250Kpa
100
50
0
0 10 20 30 40 50 60
160
159
158
157
156
155 160Kpa
154
153
152
151
150
0 10 20 30 40 50 60
80
70
60
50
70Kpa
40
40Kpa
30
20
10
0
0 10 20 30 40 50 60
0
0 10 20 30 40 50 60
-10
-20
-30
-40Kpa
-40
VAC
-50
-60
-70
-80
OPEN
-40KPa
VAC
Main parts : Vacuum Relief
valve -40KPa
WC1 Vacuum:
Vacuum Relief Valve 1. Drain waste from baths
2. Collect SRV waste
3. Impedance bath sample
Throttle pipe tubeT314 Rear panel T321 preparation and flush
4. Collect Wipe-block waste
T314 T320
Anti-overflow bath T277
VAC
Operation parts
TC2
Features :
To get -40KPa by vacuum Relief valve
Anti-reverse flow
TC2
Solution
1. For No. 1 , Switch PS2 & PS3 Tube
3. For No. 3 , connect the tube from regulator directly to pressure sensor
Possible reason :
Regulator fail or leak
Tube leak after regulator
Switch tube on PS5 & PS6 上, if the same value, the sensor is ok, otherwise it’s
sensor faulty
Connect pneumatic output directly on PS6 , If value is above -80KPa ( sea level,
at 3000M place , should be -60KPa ), it means the pneumatic is no problem
Block the tube after overflow valve , if PS5 and PS6 reach -80KPa above ( sea
level, at 3000M place , should be -60KPa ), it means the pneumatic is no
problem, but overflow valve leak
25 © 2011 Mindray Confidential
Typical error from Pneumatic system
Abnormal (red)
0
-50 t 6.1 7.5 8.9 0.3 1.7 3.1 4.5 5.9 7.3 8.7 0.1 1.5 2.9 4.3 5.7 7.1 8.5 9.9 1.3 2.7 4.1 5.5 6.9 8.3 9.7 1.1 2.5 3.9 5.3 6.7 8.1 9.5 0.9 2.3
1 1 1 1 1 1 1 2 2 2 2 2 2 2 2 3 3 3 3 3 3 3 4 4 4 4 4 4 4 5 5
-100
-150
-200 正常
-250 喷液1
-300
-350
-400
-450