Professional Documents
Culture Documents
INTERPRETAION
1
Radiography is an inspection methods that are
based on the differential absorption of
penetrating radiation.
Uses of Radiography
The ability to detect internal flaws
The ability to detect significant variations in
composition
Permanent recording of raw inspection data
Limitation
Cracks cannot be detected unless they are
essentially parallel to the radiation beam.
Laminations are nearly impossible to detect with
2
Radiographic Interpretation is the
art of extracting the maximum
pertinent information from a
radiographic image required
judgment by the interpreter .It is
influenced, not only by the
viewing condition & vision acuity,
but also by the interpreter
knowledge and experience.
3
Ensure that Interpreter has
adequate vision acuity under
proper viewing condition.
Establish proper viewing room.
Assess the quality of the
radiographic image.
4
A qualified interpreter must:
Define the quality of the radiographic image,
which includes a critical analysis of the
radiographic procedure and the image-
developing procedure
Analyze the image to determine the nature and
extent of any abnormal condition in the test
piece.
Evaluate the test piece by comparing interpreted
information with standards or specifications.
Report inspection results accurately, clearly, and
within proper administration channels.
5
Duties of a Radiographic Interpreter
7
The quality of the radiograph depends mainly on
film density, gradient, graininess, and fog, which
are functions of film type and development
procedure. The faster the film, the larger the
graininess and the lower the gradient--and vice
versa.
Film gradient, also called film contrast, is a
measure of the slope of the characteristic curve.
If the difference is great, the gradient (contrast)
is said to be high. If the difference is slight, the
gradient is said to be low.
Graininess is an inherent property of the
emulsion, but can be influenced somewhat by the
conditions of exposure and development.
8
Radiographic Film
Base
cellulose acetate / polyester
Base must be :-
• Transparent - To allow white light to go through
• Chemically inert
• Must not be susceptible to expansion and contraction
• High tensile strength
• Flexibility
9
Radiographic Film
Subbing
Base
Subbing
The film base is usually tinted blue and is about 0.18 mm
thick. A very thin but tough coating of gelatin called a
protective overcoat covers the emulsion to protect it
against minor abrasion. The total thickness of the x-ray
film is approximately 0.23 mm
Subbing layer -the adhesive between the emulsion and base
10
Radiographic Film
Supercoat
Subbing
Base
Subbing
Supercoat
11
AgNo3(Aq) + KBr(Aq) AgBr(s)+KNo3(Aq)
A crystal of silver bromide is a regular cubic
12
The film can be exposed by the direct action of X-rays, but more
commonly the X-ray energy is converted into light by intensifying
screens and this light is used to expose the film.
13
Radiographic Film
• Improve contrast
• Reduce the exposure time
14
Image formation
When radiation passes through an object it is differentially
absorbed depending upon the materials thickness and any
differing densities
The portions of radiographic film that receive sufficient
amounts of radiation undergo minute changes to produce the
latent image (hidden image)
1. The silver halide crystals are partially converted into
metallic silver to produce the latent image
2. The affected crystals are the amplified by the developer,
the developer completely converts the affected crystals
into black metallic silver
3. The radiograph attains its final appearance by fixation
15
Film Types
Grain Size Speed Quality Film factor
Coarse Fast Poor 10
Medium Medium Medium 35
Fine Slow Good 90
Ultra Fine V.Slow V.Good 200
16
Film emulsion produced by mixing solutions of
nitrate and salt such as potassium bromide.
The rate and temperature determine the grain
structures
Rapid mixing at low temperature - Finest grain
structure
Slow mixing at high temperature - Large grain
structure
17
Processing Film
18
Processing Systems
Developer
Running water
Fixer
Stop
bath
Manual System
19
Processing Systems
Development
Metallic Silver converted into Black metallic silver
3-5 min at 20OC
Main Constituents
Developing agent metol-hydroquinone
Accelerator keeps solution alkaline
Restrainer ensures only exposed silver halides converted
Preservative prevents oxidation by air
Replenishment
Purpose – to ensure that the activity of the developer and the
developing time required remains constant
20
1. Reducing Agent: reduce exposed silver halide to
black metallic silver
-Hydroquinone (Blacks)
-Phenidone (Grays)
2. Activator: softens gelatin, maintains alkaline pH
(increase pH)
-Sodium Carbonate
3. Hardener: prevents damage to the film from over
swelling of gelatin in automatic processors.
-Gluteraldehyde
21
4. Preservative: Antioxidant that prevents
oxidation of developer
-Sodium Sulfite
5. Restrainer: prevents chemical fog protects
crystals from chemicals
-Potassium Bromide
6. Solvent: dissolves & ionizes the developer
chemicals
-Water
22
Processing Systems
Stop Bath
23
Fixer
25
Processing Systems
26
Control of Temperature and Time : At
temperatures exceeding 24 °c, the radiograph
may become fogged, and the emulsion may be
loosened from the base, causing permanent
damage to the radiograph.
Agitation During Development: Agitating the
film during the course of development is the
main factor that eliminates streaking on the
radiograph.
Check the activity of Developer Solution: The
developing power of the solution decreases when
film after film is developed, Even when the
developer is not used, its activity will slowly
decrease because of oxidation of the developing
agent.
27
28
Clean, dry location
40 – 60 % Humidity 70 º Fahrenheit
Away from chemical fumes
Safe from radiation exposure
Standing on edge
Expiration date clearly visible
29
X rays are part of the electromagnetic spectrum
and are really a form of light. Like visible light, X-
ray photons travel at the speed of light, and they
can produce a latent image on film. Unlike visible
light, X rays can penetrate most opaque matter,
make some materials fluorescent, and produce
ionization of some materials. Energy is always
exchanged in discreet particles of light known as
photons. The loss of energy by the electron as it is
deflected by the heavy nuclei in the anode target
produces a very high energy photon of light called
an x-ray. The dental x-ray tube produces
Bremsstrahlung radiation.
30
X-RAY TUBE
31 31
SENSITOMETRY
The relationship between the logarithm of the
radiation exposure and the optical density is
plotted, we obtain a curve known as the
Characteristic Curve or Sensitometric curve.
32
Characteristic Curves
• Increasing exposures applied to successive
areas of a film.
• After development the densities are measured
• The density is then plotted against the log of
the exposure is called:
• Characteristic curve
• Sensitometric Curve
• Hunter & Driffield curve
33
34
Characteristic Curves
Density
Shoulder
(Log)
Straight line
section
Toe
36
Characteristic Curves
A B C D E
39
Changing Density
Density achieved 1.5
Density
Density required 2.5
2.5
Determine interval between logs
1.8 - 1.3 = 0.5
1.5
Antilog of 0.5 = 3.16
41
is a device that measures the degree of
darkness (the optical density) of a
photographic .
Transmission densitometers that measure
transparent materials.
Reflection densitometers that measure light
42
Changing Film
Density
Obtain Logs for Films A and A B
B at required density 2.5
44
Radiographic definition is the abruptness of change from one
density to another or sharpness of a radiograph , specified in
line pairs per millimeter or by the minimum
distance by which two features can be physically
distinguished.
Radiographic sensitivity is generally refers to the
size of the smallest detail that can be seen on a
radiograph.
Subject contrast is the ratio of radiation intensities
transmitted by various portions of a test piece.
Film latitude, which is the range of test piece
thickness that can be recorded with a single
exposure.
45
where Is is the intensity of the scattered radiation and Id is the intensity of the
direct (or primary) radiation passing through the tes tpiece.
46
DEFINITION
Radiographic Definition
47
Film un-sharpness specifies the spatial
resolution of the film and must not be
confused with film graininess.
Film un-sharpness depends not only on the
50
Penumbra Ug)
Focal spot
size, F
Ug= F x ofd
fod
fod
ffd (Ug = 0.25mm)
ofd
51
Penumbra (Ug)
Penumbra = F x OFD
FFD – OFD
F = 4mm
OFD = 25mm
FFD = 275
= 4 x 25
275 - 25
Penumbra = 0.4mm
53
Penumbra Calculations
- -
- -
- -
- -
55
56
Inherent Unsharpness
•Lead screens
•Fluorescent screens
•Fluoro-metallic screens
58
Lead Intensifying Screens
Film placed between 2 intensifying screens
Intensification action achieved by emitting
particulate radiation (electrons)
Generally lead of 0.05mm to 0.15mm
Frontscreen shortens exposure time and
improves quality by filtering out scatter
Back screen acts as a filter only
59
Salt Intensifying Screens
Film placed between 2 intensifying screens
Intensification action achieved by emitting
Light radiation (Visible or UV-A)
Intensification action twice that of lead
screens
No filtration action achieved
Salt screen used calcium tungastate
60
Film Latitude
61
Fluorometallic Intensifying Screens
62
Scatter
63
Scatter
64
Scatter
65
Scatter
• Side scatter : walls and nearby objects in the
path of the primary beam.
66
Scatter
67
Control of Scatter
• Protection from back scatter
• Beam filtration
• Blocking
• Grids
• Increased beam energy
• Use of masks, diaphragms, collimators
& lead screen.
68
Use of a lead diaphragm to limit the included angle of the x-ray beam, and use
of a filter to reduce subject contrast and to eliminate much of the secondary
radiation that causes undercutting.
69
Radiation Diffraction is observed most often in
the radiography of thin test pieces having a grain
size large enough. Castings made of austenitic
corrosion-resistant and heat-resistant stainless
steel or of In-conel and other nickel-base alloys
are the products most likely to exhibit diffraction
in radiographs. Diffraction patterns can be
distinguished from these conditions in the test
piece by making successive radiographs with the
test piece rotated between exposures 1 to 5°
about an axis perpendicular to the beam.
70
A mottled pattern due to porosity or
segregation will be only slightly changed,
but a pattern due to diffraction effects
will show a marked change. The Mottling
due to diffraction can be reduced, and
sometimes eliminated, by raising x-ray
tube voltage and by using lead screens.
Filters will usually aid in the control of
diffraction.
71
Sensitivity
72
IQI sensitivity
The image on a radiograph which is used to
determine the quality level.
Defect sensitivity
Ability to assist the sensitivity and locate a
defect on a radiograph.
(Depend on the defect orientation)
73
Image Quality Indicator
74
Image Quality Indicators
IQI’s / Penetrameters are used to measure
radiographic sensitivity and the quality of the
radiographic technique used.
They are not used to measure the size of defects
detected.
Standards for IQI’s include:
BS 3971
BS EN 462-1
DIN 54 109
ASTM E747
75
BS EN 462-1,Europian standard wire type IQIs each
consist of 7 wires taken from a list of 19 wires.
76
EN 462-1 wire type IQIs
77
Designation Diameter(mm)
W1 3.2
BS EN 462-1 wire diameters
W2 2.5
W3 2.0
W4 1.6
W5 1.25
W6 1.0
W7 0.8
W8 0.63
W9 0.5
W10 0.4 Easy to remember the wire diameters:
W11 0.32
W12 0.25
Remember the diameters of the first
W13 0.2 wire 3.2 & then multiply by 0.8 to get
W14 0.16 subsequent wire dia. And multiply
W15
W16
0.125
0.1
from 0.10 to 16th wire to get upper
subsequent wire dia.
78
ASTM E 747
The series consists of 21 wires ranging from 0.08 mm to
8.1 mm in diameter; there are 4 overlapping groups of 6
wires, each designated by a letter (A, B, C or D)
79
80
Image Quality Indicators
7FE12
1T
82
Image Quality Indicators
83
ASME Image Quality Indicators
4T dia
T dia
Penetrmeter Design
2T dia
Minimum Penetrmeter Thickness 0.5mm
(2% of the weld thickness) 12mm
17
Minimum Diameter for 1T Hole 0.5mm
Minimum Diameter for 2T Hole 1.0mm
Minimum Diameter for 4T Hole 2.00mm 38mm
T
IQI Sensitivity
1 Hole visible = 4T
2 Holes visible = T
3 Holes visible = 2T
84
Image Quality Indicators
86
IQI Sensitivity
87
IQI Sensitivity
88
IQI Sensitivity
Using the same IQI pack 6-12 Din 62, How many IQI wires
must be visible to give an IQI sensitivity of 2 %, thickness of
material 16mm
Sensitivity % = Thickness of thinnest step/wire visible x 100
Total object thickness
89
Exposure Control
90
Exposure control
E1 D1 2
=
E2 D2 2
E1 = New exposure time
E2 = Original exposure time
D1 = New FFD
D2 = Original FFD
91
Exposure control
For FFD/SFD change
Example: Calculate new exposure time for FFD = 600mm
Original exposure at FFD= 500mm was 10 min
E1 D1 2
=
E2 D2 2
92
If a good radiograph was produced using
an exposure of 100 curie minutes at a
source to film distance of 850 mm what
exposure will produce a good radiograph
if the source to film distance is changed
to 550 mm (assuming that all other
factors remain equal)?
E1 D1 2
=
E2 D2 2 93
SFD = 800 mm. If all other factors remain equal
what exposure is needed at an exposure chart for
iridium 192 that has been constructed for SFD =
500 mm gives an exposure of 100 Ci-min for 25
mm of steel. The specification calls for a
minimum specified SFD?
E1 D1 2
=
E2 D2 2
94
A radiographic technique produces a good
radiograph, the settings are:
kV =175, mA = 5, FFD = 440 mm and
Exposure time = 2 mins 12 secs
What exposure time will be required if the
settings are changed as follows?
kV = 175,mA = 3.5, FFD = 500
E1 D1 2
=
E2 D2 2
95
Exposure calculation
E = exposure (mA.min)
I = Tube current (mA)
T = Exposure time (min)
96
Exposure calculation
97
A satisfactory radiograph is produced in 3
minutes at 8 mA. Assuming that all other
factors remain the same, what exposure time
is required if the mA is reduced to halved?
98
Radiographic Techniques
99
Radiographic Techniques
Single Wall Single Image (SWSI)
- film inside, source outside
Single Wall Single Image (SWSI) panoramic
- film outside, source inside (internal exposure)
Double Wall Single Image (DWSI)
- film outside, source outside (external exposure)
Double Wall Double Image (DWDI)
- film outside, source outside (elliptical exposure)
10
0
Single wall single image SWSI
Film
Film
Film
Film
• IQI’s are placed on the film side
• Source outside film outside (multiple exposure)
• This technique is intended for pipe diameters over 100mm
10
3
Double wall single image DWSI
Identification
• IQI placing
A B
• Pitch marks indicating
readable film length ID MR11
Radiograph
10
4
Double wall double image DWDI elliptical
exposure
Film
• IQI’s are placed on the source side
• Source outside film outside (multiple exposure)
• A minimum of two exposures
• This technique is intended for pipe diameters less
than 100mm 10
5
Double wall double image DWDI
Identification 4 3
• Unique identification EN W10
• IQI placing
• Pitch marks indicating 1 2
readable film length ID MR12
Shot A Radiograph
10
6
Double wall double image (DWDI)
perpendicular exposure
Film
• IQI’s are placed on the source side
• Source outside film outside (multiple exposure)
• A minimum of three exposures
• Source side weld is superimposed on film side weld
• This technique is intended for small pipe diameters
10
7
The double-wall, double-image technique is
mainly applicable to sections with an outside
diameter of no more than 90 mm.
Two radiographs 90° apart are required to
provide complete coverage when the ratio of
outside diameter to inside diameter is 1.4 or less.
When the ratio of outside diameter to inside
diameter is greater than 1.4 that is, when radio
graphing a thick-wall tube the number of views
required to provide complete coverage can be
determined by multiplying that ratio by 1.7 and
rounding off to the next higher integer.
108
Sandwich Technique
FILM A FILM A
FILM B FILM B
LEAD
SCREENS
111
It must have the following:
Heat controller
Intensity control switch
Diffused light : minimize variation in the
intensity of light across the area being
viewed. Diffusing glass must be positioned
between the light source and the viewing
area.
Masks to evaluate a small portion of a large
radiograph.
11
2
Viewing conditions
• Darkened room
• Clean viewer
• Minimum adequate illumination from the viewer is
3000cd/m2
• Eyesight must be adjusted to the darkened conditions
• Comfortable viewing position and environment
• Avoid fatigue
11
3
Radiographic Quality
Density - relates to the degree of darkness
11
4
Factors Influencing Sensitivity
Sensitivity
Contrast Definition
11
5
Contrast
Contrast Definition
Contrast Definition
12
0
Radiographic Contrast
Poor contrast
Poor contrast
High contrast
12
1
The amount of blackness on the film is called
the optical density, D
12
2
Radiographic Density
Incident light
Density = Log10
Transmitted light
12
3
Radiographic Density
12
4
Measuring Radiographic Density
12
5
What is a good radiograph?
A good radiograph satisfies the
inspection requirement.
12
6
Determination of Exposure
Film density to be achieved
Wavelength - Gamma fixed, X-ray variable
Intensity - Gamma curries fixed, X-ray mA variable
Film speed
Source to film distance
Material type
Material thickness
12
7
Determination of Gamma Exposures
Gamma exposures are calculated by the use of a
gamma calculators/slide rule
Gamma calculators take into consideration
12
8
Determination of X-ray Exposures
12
9
Consideration for Exposures
- Density
- Contrast
- Definition
13
0
Consideration for Exposures
13
1
Consideration for Exposures
X–ray Exposure calculation
13
2
Consideration for Exposures
X-ray Equipment
13
3
Consideration for Exposures
Intensifying Screen
• Reduce the exposure to attain the required density
• Fluorescent & flourometallic screen have an adverse
affect on the definition
• Using lead screen (x-ray) obey the Law of Reciprocity
(E = mA x T)
13
4
Consideration for Exposures
Film Type
13
5
A steel section 50 mm thick requires an
exposure of 700 curie-minutes using Iridium
192 with a source to film distance of 1 metre
using Kodak CX film and lead intensifying
screens.
13
6
An exposure which produces an acceptable radiograph may
have been determined to be, 36 mAmins at 200 kV.
13
7
Interpretation of radiographs takes place in three
basic steps:
(1) detection,
(3) evaluation.
13
8
i) Rounded Indications – indications with a
maximum length of three times the width or less.
These indications may be circular,
elliptical, conical, or irregular in shape.
ii) Aligned Indications – a sequence of four or more
139
Discontinuities are interruptions in the
typical structure of a material
Defects discontinuities, which do not meet
140
Types of discontinuities
Inherent discontinuities: results of its initial
141
Welding Discontinuities
Cold lap is a condition where the weld filler
metal does not properly fuse with the base
metal or the previous weld pass material
(inter pass cold lap). The arc does not melt
the base metal sufficiently and causes the
slightly molten puddle to flow into the base
material without bonding.
142
143
Porosity is the result of gas entrapment in
the solidifying metal. Porosity can take many
shapes on a radiograph but often appears as
dark round or irregular spots . This is the
result of gas attempting to escape while the
metal is still in a liquid state and is called
wormhole porosity.
Hollow Bead: is an elongated gas void that is
144
145
Clusterporosity is caused
when flux coated electrodes
are contaminated with
moisture. The moisture turns
into a gas when heated and
becomes trapped in the weld
during the welding process.
The indications will be
grouped close together.
146
147
Slaginclusions are nonmetallic
solid material entrapped in
weld metal or between weld
and base metal. In a
radiograph, dark, jagged
asymmetrical shapes within
the weld or along the weld
joint areas are indicative of
slag inclusions.
148
149
LOP occurs when the weld
metal fails to penetrate the
joint. The appearance on a
radiograph is a dark area with
well-defined, straight edges
that follows the land or root
face down the center of the
weldment.
150
151
Incomplete fusion is a condition
where the weld filler metal does not
properly fuse with the base
metal. appears as a dark line or
lines oriented in the direction
of the weld seam along the weld
preparation or joining area.
15
2
153
Internal concavity or suck
back is a condition where the weld
metal has contracted as it cools
and has been drawn up into the
root of the weld. On a radiograph it
looks similar to a lack of
penetration but the line has
irregular edges and it is often quite
wide in the center of the weld
image.
154
155
Internal or root undercut
is an erosion of the base metal
next to the root of the weld. In
the radiographic image it
appears as a dark irregular line
offset from the centerline of the
weldment. Undercutting is not as
straight edged as LOP because it
does not follow a ground edge.
156
157
External or crown undercut
is an erosion of the base
metal next to the crown
of the weld. In the
radiograph, it appears as
a dark irregular line
along the outside edge of
the weld area.
158
159
Offset or mismatch
are terms associated with a condition
where two pieces being welded together
are not properly aligned. The
radiographic image shows a noticeable
difference in
density between the two pieces. The
difference in density is caused by the
difference in material thickness. The
dark, straight line is caused by the failure
of the weld metal to fuse with the land
area.
160
161
Inadequate weld reinforcement
is an area of a weld where the
thickness of weld metal
deposited is less than the
thickness of the base material.
The image density in the area
of suspected inadequacy will
be higher (darker) than the
image density of the
surrounding base material
162
163
Excess weld reinforcement
is an area of a weld that has
weld metal added in excess of
that specified by engineering
drawings and codes. The
appearance on a radiograph is
a localized, lighter area in the
weld.
164
Cracks are fracture or ruptures of the weld metal
occuring when the stress in a localized area exceed
the weld metal’s ultimate tensile strength.It
will appear as jagged and often very faint irregular lines.
Cracks can sometimes appear as "tails" on inclusions or
porosity.
There are basically two types of crack one is Hot & other cold.
Hot crack: occur while the weld metal is in plastic condition
Cold crack: occurs after the weld metal has cooled.
It is also classified as Longitudinal Crack, Transverse crack,
Under bead crack(occur in the HAZ), Crater crack(star shaped
occurs in a depression at the end of a weld bead.
165
166
Tungsten inclusions
Tungsten is a brittle and inherently
dense material used in the electrode
in tungsten inert gas welding. If
improper welding procedures are
used, tungsten may be entrapped in
the weld. Radiographically, tungsten
is more dense than aluminum or
steel therefore it shows up as a
lighter area with a distinct outline on
the radiograph.
167
168
Whiskers are short lengths
of weld electrode wire,
visible on the top or bottom
surface of the weld or
contained within the weld.
On a radiograph they
appear as light, "wire like"
indications.
169
Burn-Through results when too
much heat causes excessive weld
metal to penetrate the weld zone.
Often lumps of metal sag through
the weld, creating a thick globular
condition on the back of the weld.
These globs of metal are referred
to as icicles. On a radiograph,
burn-through appears as dark
spots, which are often surrounded
by light globular areas (icicles).
170
171
CASTING DEFECTS:
172
GAS POROCITY GAS VOID
17
3
are nonmetallic oxides, which appear on the
radiograph as irregular, dark blotches.
17
4
is a form of discontinuity that appears as
dark spots on the radiograph. Shrinkage
assumes various forms, but in all cases it
occurs because molten metal shrinks as it
solidifies,
17
5
176
Cracks are thin (straight or jagged) linearly disposed
discontinuities that occur after the melt has solidified. They
generally appear singly and originate at casting surfaces.
Cold shuts generally appear on or near a surface of cast
metal as a result of two streams of liquid meeting and
failing to unite. They may appear on a radiograph as cracks
or seams with smooth or rounded edges.
Hot tears are linearly disposed indications that represent
fractures formed in a metal during solidification because of
hindered contraction.
Mottling is a radiographic indication that appears as an
indistinct area of more or less dense images. The condition
is a diffraction effect that occurs on relatively vague,thin-
section radiographs, most often with austenitic stainless
steel.
177
crack in a steel weldment.
Crack in a steel
HOT TEAR
weldment
178
Dark outline of a shape similar to the shape of
the chaplet
itself .Chaplets are used to maintain cores in
their correct position during the casting
process. When a chaplet is not fused
completely by the molten metal, a discontinuity
is formed.
17
9
ARTIFACTS
developer temperature.
Density too low reason
Underexposure Check exposure (time and
radiation intensity
Underdevelopment Increase development time or
180
FOG
Light leaks in darkroom With darkroom unlighted, turn on all
lights in adjoining rooms; seal any light leaks.
Exposure to safelight Reduce safelight wattage. Use proper
safelight filters.
Stored film inadequately protected from radiation
Attach strip of lead to loaded film holder and place in film-
storage area. Develop test film after 2 to 3 weeks; if image of
strip is evident, improve radiation shielding in storage area.
Film exposed to heat, humidity, or gases Store film in a cool,
dry place not subject to gases or vapors.
Overdevelopment Reduce development time or developer
temperature.
Developer contaminated Replace developer.
Exposure during processing Do not inspect film during
processing until fixing is completed.
181
Dark circular marks: Film splashed with developer prior to
immersion Immerse film in developer with care.
Dark spots :Insufficient fixing Use fresh fixer solution and
proper fixing time.
Static discharge: Unwrap film carefully. Do not rub films
together. Avoid clothing productive of static electricity.
Dark fingerprints :Touching undeveloped film with
chemically contaminated fingers Wash hands thoroughly and
dry, or use clean, dry rubber gloves.
Light fingerprints Touching undeveloped film with oily or
greasy fingers Wash hands thoroughly and dry, or use clean,
dry rubber gloves.
Light circular patches: Air bubbles on film during
development
Agitate immediately upon immersion of film in developer.
182
Reticulation (leather grain appearance):
183
lower density compared to adjacent areas.
Causes : localized application of pressure
18
4
crescent-shaped defects that are darker
(higher density)than adjacent film areas
sharp bending of the film after exposure and
18
5
Crescent-shaped defects that are lighter (lower density)
than adjacent film areas.
Causes: Sharp, abrupt bending of the film before
exposure causes crimp marks of this type.
18
6
Appear as either jagged, branching dark lines
or irregular
Causes: The dissipation of static electricity
Occurs during removing film quickly from its
18
7
small spots of a lower density than the
surrounding film area.
Before development, a splash of fixer, even
DEVELOPER SPOTS
HAIR LINES
18
8
A screen mark will appear as a dark line on
a negative. Screen marks will also appear
as white spots on a negative.
A dark line is caused by a deep scratch on
18
9
Defined as a line, smear, or band. In this case
it can be black or white.
Caused by any of several types of chemical
contamination.
The contamination often occurs in poor
darkroom setups or because of improper
processing technique.
Streaks of this type happen when chemicals
19
0
Relevant Indications – only those rounded
indications that exceed the following
dimensions shall be considered relevant and
compared to the acceptance charts for
disposition.
t /10 for t less than 3 mm (1/8 in)
0.4 mm (1/64 in) for t greater than or equal
to 3 mm (1/8 in) and less than or equal to 6
mm (1/4 in)
0.8 mm (1/32 in) for t greater than 6 mm
(1/4 in) and less than or equal to 50 mm (2 in)
19
1
1.5 mm (1/16 in) for t greater than 50 mm (2
in)
Maximum Size of Rounded Indication – the
192
6. Aligned Rounded Indications – aligned rounded
indications are acceptable when the
summation of the diameters of the indications is
less than t in a length of 12t .
Minimum group spacing 3L where L is the longest
imperfection in the group.
7. Clustered Indications –The length of an
193
Any crack or zone of incomplete fusion or
lack of penetration
ii) Any other linear indication that has a
or equal to 57 mm
19 mm for t greater than 57 mm
RC is acceptable if density is less than the
19
4
Required commercial quality : storage for
3-10 years.
Required archival quality: storage for 20 years
or more
If radiography films are not properly washed,
fixer chemicals thiosulfate salts remain in the
emulsion, affect the storage life. If it exceeds
a certain max allowable limit, the radiograph
becomes fade or change in color during long
term storage.
19
5
Methylene blue test : performed within two
weeks of processing the film.
Silver densitometric test: can be performed at
any time. Both are laboratory procedure.
Rough estimation can be performed using 710
ml of water, 120 ml of acetic acid, & 7ml of
silver nitrate and diluted with 950 mlof water
prior to use. When a drop of this solution is
placed on a clear area of a radiograph ,it turns
brown. Now matching the brown spot with
patches on a standard test strip.
19
6
THE END
THANKS
197