Advanced Training Material

Basic Operation of Splicing

© INVAS Technologies Pvt. Ltd.


Stripping the fiber

Please click on the URL for the video:

https://
www.youtube.com/wa
tch?v=t12ctguyvA8

CONFIDENTIAL INVAS Technologies Private Limited


Cleaning the fiber

Please click on the URL for the video:

https://
www.youtube.com/wa
tch?v=EJOUJ_QNgfo

Note:
Clean the bare fibre from the end of the Fibre coating
with a lint free gauze pad moistened with pure alcohol.

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaving the fiber

1. Open the Top clamp lever Please click on the URL for the video:
and lid of the Single fiber
Adapter
https://
www.youtube.com/wa
2. Move the Blade carriage in the tch?v=0V9J1MhVfF8
forward position

3. Place the fiber in to the


correct groove of the Single
fiber Adapter at the desired
mark on the ruler.

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaving the fiber

4. Close the lid of the Single fiber Please click on the URL for the video:

Adapter and Toper Clamp lever


https://
and slide the Blade carriage to
the rear side as shown in the www.youtube.com/wa
Below image. tch?v=0V9J1MhVfF8

CONFIDENTIAL INVAS Technologies Private Limited | 10


Inserting the fibres into the Splicing machine

CONFIDENTIAL INVAS Technologies Private Limited | 10


Inserting Protection Sleeve, Splicing and splice protection

Please click on the URL for the video:

https://
www.youtube.com/wa
tch?v=IeA6PD_rYk0

CONFIDENTIAL INVAS Technologies Private Limited | 10


To keep excellent splice quality, regular cleaning and
inspection are required. Especially cleaning should be
performed before and after the each use. We recommend
your Splicer to be checked through our service center
once in a year.
Daily Cleaning &
Check-up Cleaning
- Turn of the Z2C before maintenance work.
- Clean each part with cotton swab(ear buds)
moistened with Isopropyl Alcohol.
- Clean components before and after the use.

CONFIDENTIA
INVAS Technologies Private Limited | 8
L
Maintenance Tool kit

CONFIDENTIAL INVAS Technologies Private Limited | 7


Component Names

CONFIDENTIAL INVAS Technologies Private Limited | 9


Component Names

CONFIDENTIAL INVAS Technologies Private Limited | 10


Splicing machine Do’s & Don’ts(Z2C/82C/T-82M12/T-400S)
Do’s Don’ts
1. Replace the Electrodes after 6000 Splices and other consumable 1. Do not use the fake Electrodes and Cleaver Blades. Using fake
spare parts like Blade Cleaver pads etc. for optimal splice quality. Electrodes may affects the Arc circuit board.
2. Please reset arc count only when electrodes are replaced. 2. Do not sharpen the Electrodes because the Arc power and
Otherwise, proper electrodes change timing will be known Electrode position get unstable.
correctly. 3. Do not use the splicer in a higher or lower temperature than
3. Perform the Arc Test when there are changes in operational the one described in the specification sheet.
Environment for better splice quality. 4. Do not charge the Battery frequently. Too many charging and
4. When the Splicing machine is not used for many days we discharging cycles may shorten the Battery cycle and battery
recommend you to Put silica gel inside a storage box (to prevent life.
humidity) 5. If you do not use fusion splicer for long period(more than one
5. Cleaning is the most important for daily maintenance. We month), we recommend to remove a battery pack from the
recommend to clean V-Grooves, LEDs, Bare fiber pads, Microscope splicer and store within "-20 ℃~+20 ℃" environment without
lens protection glass etc. please use pure Isopropyl alcohol and dew condensation.
cotton buds. 6. Do not use the machine if it gets wet. Dry the machine
6. Only use the recommended Battery pack, AC Adapter, and Power completely before the operation or contact service center.
cord. Failure to do so can cause these items or Splicing machine 7. Do not use the gas spray for cleaning the machine. Doing so
may heat resulting in fire or electric shock. may contaminate the inside parts of the machine.
7. Only use the specified Battery charger. Failure to do so can cause 8. Do not touch the Electrodes during or after the splice. Doing
the Battery to be overcharged, bulged, over heated, bursting, fire. so can cause personal injury or electric shock.
8. This product is a precision instrument. When transporting the 9. Do not use the Heat shrink oven with Isopropyl alcohol.
machine, use its specified Hard carrying case to protect the
machine from excessive shock impact.
CONFIDENTIAL INVAS Technologies Private Limited | 10
Cleaning the V-Grooves

Please click on the URL for the video:

https://www.youtube.c
om/watch?v=wSa-RGI1
HEg

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaning the LEDs

Please click on the URL for the video:

https://www.youtube.c
om/watch?v=q9FwiYA
MzG8

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaning the Bare fiber pads

Please click on the URL for the video:

https://www.youtube.c
om/watch?v=4uLML0v
N0kM

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaning the Microscope lens protection covers

Please click on the URL for the video:

https://www.youtube.c
om/watch?v=tBZe4xBR
6nc

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaning the Heat shrink oven
(1) Ensure the ovens are cool before cleaning. (1) Ensure the ovens are cool before cleaning.
Please click on the URL for the video:

https://
www.youtube.com/wa
tch?v=SwbnJ8Mzdw8

(2) Open the heat shrink oven clamps and (4) Clean the clamps with a cotton
transparent lids. swab moistened with pure alcohol.

CONFIDENTIAL INVAS Technologies Private Limited | 10


Arc Test

CONFIDENTIAL INVAS Technologies Private Limited | 10


Arc Test
1. Insert the fiber into the machine 3. Tap on Arc Test icon on setting panel

4. After “Arc Test ready” appear, tap on Set icon to


2. Tap the setting panel icon circled in Yellow start the Arc test.

CONFIDENTIAL INVAS Technologies Private Limited | 10


Arc Test
5. The fiber will melt back distance and display the Arc position values.

6. After “Good arc state” displayed, begin the fiber preparation for splicing.

Note:
If “Arc Power Too Weak”, “Arc Power Too Strong”, “Arc
Centre” is displayed, prepare the fiber again and repeat
the test until the “Good arc state” displayed.

CONFIDENTIAL INVAS Technologies Private Limited | 10


Arc Test?

Please click on the URL for the video:

https://
www.youtube.com/wa
tch?v=SwbnJ8Mzdw8

CONFIDENTIAL INVAS Technologies Private Limited | 10


Importance of Electrode replacement

Question from the customer


It's a pain to change out the electrodes, so sometimes we end up using
them even past the recommended number of connections. What kinds
of problems might this cause?

Answer from the engineer


When electrodes are used beyond their recommended number of
splices, silica deposits accumulate on the electrode tips. The silica is an
insulator and can affect the arc direction. Together these changes
create arc instability which can increase splice loss. Some workers try
to remove the silica deposits with sandpaper or other abrasive
materials. But this risks changes the shape of the electrode tip, which
again can lead to further instability of the arc.

CONFIDENTIAL INVAS Technologies Private Limited | 10


How to replace the Electrodes
(1) Turn off the splicer. (3) Remove the battery from the splicer.
Please click on the URL for the video:

https://www.youtube.c
om/watch?v=il_MAjw_
Ehc

(2) Disconnect the power cord. (4) Open the hood.

CONFIDENTIAL INVAS Technologies Private Limited | 10


How to replace the Electrodes
(5) Loosen the thumbscrew and lift the electrode cover plate.
(Take care not to touch the electrode tip against anything.)
Please click on the URL for the video:
Repeat the previous step for the other electrode
https://www.youtube.c
om/watch?v=il_MAjw_
Ehc

(6) Take care not to touch the metallic part of the electrode.

CONFIDENTIAL INVAS Technologies Private Limited | 10


How to replace the Electrodes
(7) Install new electrodes.(Take care not to touch the electrode tip against anything.)
Please click on the URL for the video:

https://www.youtube.c
om/watch?v=il_MAjw_
Ehc

If the electrode is not installed in the splicer properly, it will be lifted up and higher
than the electrode which is installed properly

CONFIDENTIAL INVAS Technologies Private Limited | 10


Cleaver Maintenance – Blade replacement

IMPORTANT
Do not rotate the blade with
metallic tools

INVAS Technologies Private Limited | 2


Cleaver Maintenance – Blade rotation

INVAS Technologies Private Limited | 2


Basic Operation of OTDR

© INVAS Technologies Pvt. Ltd.


EXFO OTDR General Maintenance

1. Always inspect fiber-optic connectors before using them and clean


them if necessary.
2. Keep the unit free of dust.
3. Clean the unit casing and front panel with a cloth slightly dampened
with water.

4. Store unit at room temperature in a clean and dry area. Keep the unit
out of direct sunlight.
5. Avoid high humidity or significant temperature fluctuations.
6. Avoid unnecessary shocks and vibrations.
7. If any liquids are spilled on or into the unit, turn off the power
immediately, disconnect from any external power source, remove the
batteries and let the unit dry completely.

INVAS Technologies Private Limited | 2


EXFO OTDR General Maintenance

Cleaning EUI Connectors


Regular cleaning of EUI connectors will help maintain optimum
performance. There is no need to disassemble the unit.
IMPORTANT: If any damage occurs to internal connectors, the module
casing will have to be opened and a new calibration will be required.

WARNING: Looking into the optical connector while the light source is
active WILL result in permanent eye damage. EXFO strongly
recommends to TURN OFF the unit before proceeding with the cleaning
procedure.

INVAS Technologies Private Limited | 2


EXFO OTDR General Maintenance
To clean EUI connectors:
 Remove the EUI from the instrument to expose the connector
baseplate and ferrule.

 Moisten a 2.5 mm cleaning tip with one drop of isopropyl alcohol


(alcohol may leave traces if used abundantly).
 Slowly insert the cleaning tip into the EUI adapter until it comes out on
the other side.

INVAS Technologies Private Limited | 2