Power factor of the choke coil Cos = (I1- I2- I3)
2 I2 I3
Voltage across the choke coil V = I2 R
Impedance of the coil (Z) = V/ I3
Resistance of the coil (R) = Z Cos
Reactance of the coil (X) = Z Sin
Induction of the coil (L) = X/2
Observation table for 3 Voltmeter method:
S.No V1 V2 V3
P=
(V1- V2- V3) 2R
Cos=
(V1- V2- V3) 2 V2 V3
Sin
I=V2
R
Z=V3
I
R=Zcos XL=Zsin L=XL/2 f
Electrical Measurements Lab
EEE Department Page 14
Average Inductance = Average Resistance =
Observation table for 3 Ammeter method:
S.N
o
I1 I2 I3
P=
(I1- I2- I3)
2R
Cos=
(I1- I2- I3)
2 I2 I3
Sin V= I2R Z=V/I3 R=Zcos XL=Zsin L=XL/2 f
Average Inductance =
Average Resistance=
Result:
Parameters of a given choke coil by using 3 Voltmeter &
3 Ammeter methods is measured.
VIVA-VOICE
1. what is inductance ?
2. what is formula for inductive reactance ?
3. what is formula for capacitive reactance ?
4. what is capacitance?
5. what is rating of dimmer stat?
6. What is meant by choke coil ?
7. What is the difference between MC &MI instruments?
8. what is resistance?
9. what is meant by power factor?
10. what is power?
Electrical Measurements Lab
EEE Department Page 15
5.MEASUREMENT OF MUTUAL INDUCTANCE
Aim:
To measure the mutual inductance of a given transformer.
Name Plate details of transformer:
Power 3 KVA
Voltage 230V
Frequency 50 Hz
Phase 1-
Apparatus:
S.No Name Type Range Qty
1 Transformer --- 230V, 3KVA 1
2 Voltmeter M.I
0-150V 1
0-300V 1
3 Ammeter M.I 0-2A 1
4 Wattmeter LPF 0-2A, 0-150V 1
Circuit Diagram:
Fig: Measurement of mutual inductance
Theory:
Mutual inductance has been used in great number of bridges
for a variety of purposes known mutual inductances are used in some
circuits for the measurements of known mutual inductances. Variable
Electrical Measurements Lab
EEE Department Page 16
mutual standard inductances have been used as components in bridges
for the measurement of self inductance, capacitance and frequency.
Many such bridges are found in literature, but some of the circuits
represents minor modifications of other bridges that in changes to
achieve greater accuracy or special range of unknown quantities of
terminals of two coils whose mutual inductance is to be measured are
available.
Self inductance= flux linkages per unit current
L= (N/I)
Mutual inductance is phenomenon but two coils coupled together
M= (N21/I1)
Procedure:
1. Make the connections as per the circuit diagram.
2. L.V side is taken as primary for magnification of no load
current.
3. Keep the autotransformer at minimum voltage output
position and switch on the supply.
4. Now initially apply some voltage to the primary winding
and the corresponding no load current, wattmeter reading
and e.m.f induced in the secondary side (V2) should be
note down.
5. Calculate M using the formula M= V2 / (Im)
6. Repeat the experiment for deferent values of voltage.
7. Take one reading just above the reated voltage.
Calculations:
No load power factor, Cos0 = W/ (I0V1)
Ic = I0 Cos0 ; Im = I0 Sin0
V2 = Induced e.m.f in the secondary side = MIm
Mutual induction (M) = V2/( Im) Where =2f
Observation Table:
S.No
Primary
voltage
(V2)
Primary
current
(I0)
Wattmeter
reading
(W0)
Im = I0 Sin0
Secondary
voltage
(V2)
M= V2/( Im)
Graph:
1. No load losses verses primary voltage.
2. Im verses no load voltage of primary.
3. Mutual induction verses voltage of primary.
Electrical Measurements Lab
EEE Department Page 17
Model graph:
M,W,Im
V1
Result:
The Mutual inductance of a given transformer is calculated.
Precautions:
1. Loose connections should be avoided.
2. Check the connections before giving supply.
3. Note the readings carefully.
VIVA-VOICE
1. What is inductance?
2. What is mutual inductance?
3. What is the formula for voltage across inductance?
4. What is the formula for voltage across capacitance?
5. What is the difference between inductance and mutual
6. Why we are using LPF wattmeter in this experiment?
7. What is LPF wattmeter?
8. What is meant by exciting current?
9. What is meant by magnetizing current?
10. What is the formula for magnetizing current?
6.CROMPTON D.C POTENTIOMETER.
Aim: Calibration of PMMC voltmeter and ammeter by using Crompton
D.C Potentiometer.
Apparatus:
1. DC Potentiometer.
2. Standard cell
3. Volt ratio box
4. Sensitive Galvanometer
M vs V
1
W vs V
1
I
m
vs V
1
Electrical Measurements Lab
EEE Department Page 18
5. DC power supply : (0-30V,1A)
6. Volt meter: (0-30V), MC
7. Ammeter: (0-1A),MC
8. Rheostat: 50,5A
Circuit diagram:
Fig: DC Crompton Potentiometer
Circuit diagram:
Electrical Measurements Lab
EEE Department Page 19
Calibration of Voltmeter:
Fig: calibration of voltmeter
Theory:
Cromptons DC Potentiometer
Cromptons DC potentiometer is a slide wire type of potentiometer. The
long slide wire is awkward, and ever for the length shown cannot be read
to a very great degree of precision. Modern laboratory type potentiometer
use calibrated dial resistors and a small circular wire of one (or) more
turns, there by reducing the size of the instrument. The circuit of a
simple laboratory type potentiometer consists of one dial switch with
filter steps ,each having a precession resistors. There is also a single turn
wire. This potentiometer consists of unknown cell. A key and a protective
resistance is used in the galvanometer circuits.
Procedure:
1. Make the connections as per the circuit diagram.
2. Calibrate the DC potentiometer by adjust coarse and fine
rheostats. When Galvanometer has null deflection and
function must be in STD mode.
Electrical Measurements Lab
EEE Department Page 20
3. The voltage across voltmeter is stepped down to a value
suitable for application to a potentiometer with the help of
volt ratio box.
4. For accurse of measurement it is necessary to measure
voltage near the maximum range of the potentiometer.
Calibration of Ammeter:
Fig: Calibration of Ammeter
Procedure:
1. Made the connections as per the circuit diagram.
2. A resistance of suitable value and sufficient current carrying
capacity is placed in series with the ammeter under calibration.
3. The voltage across the standard resistor is measured with the help
of potentiometer and the current trough the standard resistance
can be computed.
4. Current I = VS / R
Where VS = Voltage across the standard resistor as indicated by
the potentiometer.
R = Resistance of standard resistor.
Electrical Measurements Lab
EEE Department Page 21
Tabular column:
Calibration of Voltmeter:
S.No Voltage from
DC potentiometer
(Vact)
Voltmeter
(Vin)
% Error =
((I out - I in)/ I out) x 100
Calibration of Ammeter:
S.No Iin
(Amps)
I out = Vs /
S
(Amps)
% Error= ((I out - I in)/ I out) x 100
RESULT: Thus PMMC Voltmeter and Ammeter using DC Crompton
Potentiometer are calibrated.
VIVA-VOICE
1. How is dc potentiometer made direct reading?
2. How is the dc potentiometer is standardized?
3. What is dc potentiometer ?
4. What is ac potentiometer?
5. What is the difference between dc and ac potentiometer?
6. What is polar type potentiometer ?
7. What is coordinate type potentiometer?
8. What is difference between polar and coordinate type potentiometer?
9. Why are ac potentiometer not very widely used?
10. On what basis ac potentiometer are classified?
Electrical Measurements Lab
EEE Department Page 22
7.C.T TESTING BY SILSBEES METHOD
AIM:
To test the given current transformer by Silsbees method and to
determine the Phase Angle and Percentage Ratio error.
APPARATUS:
THEORY:
Silsbees method is a comparison method. There are two types of
Silsbees methods: deflection and null. Only deflection method is
described here. Here the ratio and phase angle of the test transformer X
are determined, in terms of that of a standard transformer S having the
same nominal ratio. The two transformers are connected with their
primaries in series. An adjustable burden is put in the secondary circuit
of the transformer under test.
An ammeter is included in the secondary circuit of the standard
transformer so that the current may be sent to the desired value .
w1 is a wattmeter whose current coil is connected to carry the
secondary current of the standard transformer. The current coil of
wattmeter w2 carries current I which is the difference between the
secondary currents of the standard and test transformer s. The voltage
Circuits of the wattmeters (I .e; their pressure coils) are supplied in
parallel from a phase shifting transformer at a constant voltage v.
S.No Name Type Range Qty
1 Phase shifting transformer --- ---- ----
2 Precision C.T (0-5A)
1
3 commercial C.T (0-5A)
1
4 Ammeters MI (0-5A)
4
5 Wattmeter LPF
UPF
(0-2.5A, 0-300V)
(0-2.5A, 0-300V)
1
1
6 Rheostat (50, 5A)
1
Electrical Measurements Lab
EEE Department Page 23
CIRCUIT DIAGRAM:
Procedure:
1. The connections are made as per the ckt diagram.
2. With the (10x4) rheostat cutout the supply switch to the auto
transformer is closed.
Electrical Measurements Lab
EEE Department Page 24
3. Using the auto transformer the current through the ct primaries
is gradually increased.
4. The current ID1, bring the difference of C.Ts,s secondary
current (Iss-Isx),, should be zero If it is twice that of Iss or Isx,
the supply is opened & the connections to any one of the C.T
secondaries are reversed.
5. This ensures proper polarity of C.T secondaries the current (Ip)
through the C.T primaries is made equal to a fixed value using
auto transformer .
6. The burden ( 10x ) rheostat is introduced in the secondary
ckt . of the C.T to be Compared. This resistance is gradually cut
in fill the current Id Is equal to a fixed value.
7. BY operating the phase shifter, the wattmeter Ws is made to
indicate zero .The readings Of the ammeter & wattmeter WD2 are
noted.
8. Now Ws is adjusted to indicate maximum reading by means of
phase shifter and the Reading of the ammeters and wattmeters
WD1 are noted.
9. The voltage V applied to the wattmeters potential coils is also
noted. The above Procedure is repeated for different primary
currents like 15 amp & 20amp .
Calculations :
Nominal ratio of standard C.T = 2
Ratio error of standard C.T = 0.5%
Phase angle error of standard C.T = 8
Actual ratio of standard C.T = 2/(1+0.005) =Na
Actual ratio of C.T under test is
given by Na = Na[1+WD/VIss]
%Error = NR-AR/AR
Phase angle error for the C.T
under test = x = [s + (WD2/VIss)(180/)]
Tabular column:
S.NO Ip(A) Iss(A) ID(A) Ws(A) WD1(W) WD2(W) Isx(A) RATIO
ERROR
Phase
angle
error
Electrical Measurements Lab
EEE Department Page 25
Phasor diagram:
V b
V a
ISX ISS
s
r
IP
RESULT: Hence the phase angle and percentage ratio error are
calculated by using Silsbees method.
VIVA-VOICE
1. What is meant by the term instrument transformer?
2. What is instrument transformer?
3. How do instrument transformer is differ from power transformer?
4. What do you understand by ammeter shunt?
5. What do you understand by voltmeter multiplier?
6. Why instrument transformers are used?
7. What is meant by turn burden of an instrument transformer?
8. How do current transformer is differ from potential transformer?
9. What is formula for ratio error?
10. What is formula for phase angle error?
8.a SCHERING BRIDGE
AIM: To determine the value of given capacitor and obtain its
dissipation factor.
APPARATUS:
1. Schering bridge
2. Function generator
3. CRO
4. Probes and connecting wires
5. Digital voltmeter
THEORY: Schering bridge is widely used for capacitance and
the dissipation factor measurement in fact schering bridge is one
of the most important of the A.C bridges, it is extensively used in
measurement of capacitance in general and in particular in the
measurement of proper he of insulators capacitor bushing ,
insulating oil and other insulating materials . The bridge is
particularly suitable for small capacitance and is then usually
Electrical Measurements Lab
EEE Department Page 26
supplied from a high frequency or a high voltage sources . The
measurement done on small capacitance suffer from many
disadvantages it carried out at low voltage high voltage schering
bridge is preferable for such measurement.
Fig: Circuit diagram of schering bridge
TABULAR FORM:
R1() R2() C3(f) CX=(R1/R2)C3 Dissipation
factor
Average capacitance =
Dissipation factor=
Average Dissipation factor=
Electrical Measurements Lab
EEE Department Page 27
PROCEDURE:
1) The bridge is connected as shown in circuit diagram.
2) The oscilloscope is connected across the point b and d.
3) The bridge is balanced by adjusting Cy and Ry.
4) The reading are tabulated and C4 and R1 is calculated.
5) A number of reading are taken and average values are
calculated.
CALCULATIONS:
C1=Unknown capacitance
R1= Resistance representing losses in the unknown capacitance
C3=Standard capacitor
R3=Standard non-inductive resistance
R4=Variable non-inductive resistance
C4= Variable capacitor
CX=(R1/R2)C3
Dissipation factor=2fCxR1
RESULT:
Thus the value of given capacitance and Dissipation Factor
are measured.
VIVA VOICE QUESTIONS:
1. Why a spark is connected across resistance arms in a
Schering bridge?
2. The most useful ac bridge for comparing capacitances of
two air capacitor is______________________________
3. Dissipation factor of a capacitor can be determined by using a
_______________________
4. The capacitance and dielectric loss of a capacitor is generally
measured by ___________
5. A bridge used for measurement of dielectric loss and power
factor is _________________
6. The bridge used for measuring inter electrode capacitance
is____________________
7. The bridge used for measuring dissipation factor of a
capacitor is___________________
8. Most commonly used AC bridge circuit for the measurement
of capacitance is________
9. The bridge suitable for measurement of capacitance of a
capacitor at high voltage is ___________
10. Why is Schering bridge particularly suitable for
measurement at high voltage?
Electrical Measurements Lab
EEE Department Page 28
8.b ANDERSON BRIDGE
AIM: To calculate the value of unknown inductance by using
Anderson Bridge.
Apparatus:
1. Anderson Bridge trainer circuit.
2. Connecting wires.
3. Head phones.
4. Unknown inductance.
5. Auto frequency Oscilloscope.
CIRCUIT DIAGRAM:
FIG: ANDERSON BRIDGE
Procedure:
1. Connect the audio oscillator and Head phones to proper
terminals.
2. 230 V supply is given to the Bridge oscillator.
3. In the Bridge the value of P, Q, R is equal to 1000 .
Minimum sound can be obtained by varying S and M
alternatively.
4. The best way to get balance is to vary S first to dicers the
sound in the headphones. The final minimum sound can be
obtained by varying M.
Electrical Measurements Lab
EEE Department Page 29
5. It is to be note that the perfect silence cannot be obtained in
the headphones. But only minimum of sound can be
achieved.
6. The value of M, S, and C can be noted. The value of L can
be calculated from formula.
L= C [RQ + (R+S) M]
Where
L is in henries
C is in farads
Other resistances are in Ohms.
S includes the resistance of self-inductance also for all
calculations.
7. The experiment for the same inductance can be reputed
By selecting of deferent value of C and gating balance by
varying S and M. The mean vale of L can be calculated.
Bridge arms:
P = Non inductive resistance of 1000
Q = Non inductive resistance of 1000
R = Non inductive resistance of 1000
S = A variable non inductive resistance in the form of 3 decades of
10x1, 10x10, 10x100 .
M = A variable non inductive resistance in the form of 3 decades of
10x1, 10x10, 10x100 .
C = A standard capacitance in the form 4 values of 0.005, 0.01,
0.02,
0.5 fd selected by a selector switch
Precautions:
1. The value of C should be small so as to allow sufficient
variations of M.
2. The A.C balance should be obtained by varying S and M
alternately.
Result:
Hence the unknown value of inductance is measured by using Anderson
Bridge .
Electrical Measurements Lab
EEE Department Page 30
VIVA VOICE
1. Maxwell bridge is used for measurement of
_________________________
2. Maxwells bridge is very convenient and useful bridge for
determination of
Inductance of a coil having ________________________
3. why there are two conditions of balance in ac bridges ?
4.why is high grade insulation employed in high impedance bridges?
5. why are highly sensitive detectors undesirable for the operation
of ac bridges ?
6. In an Anderson bridge , the unknown inductance is measured in
terms of________________
7.Anderson bridge is used for the measurement of ______________________
8Anderson bridge is a modification of _________________________________
9.Anderson bridge is used to measure______________________
9.CALIBRATION OF LPF WATTMETER BY PHANTOM LOAD TEST
Aim: To calibrate the given LPF wattmeter using phantom loading.
Apparatus:
Name Type Range Qty
Voltmeter M.I 0-150V 1
ammeter M.I 0-5/10A 1
wattmeter LPF 0-150V,5A 1
Autotransformer .. 0-270V,10A,1-Ph 2
CIRCUIT DIAGRAM :
Electrical Measurements Lab
EEE Department Page 31
Fig: Calibration of LPF wattmeter
Theory:
Calibration of LPF watt meter by phantom loading :
When the current rating of a meter under test is high as test with
actual loading arrangements would involve considerable wastage of
power. In order to avoid thisphantom or frictions loading is done.
Phantom loading consist of supply the pressure circuit from a
circuit of a required normal voltage and a current circuit from a low
voltage supply. It is possible to circulate the rated current throw the
current circuit with a low voltage supply as the impedance of the circuit
is very low with this arrangement the total power supplied for the test is
that due to the small pressure coil current at normal voltage, pulse that
due to the circuit current supplied at low voltage. The total power ,
required for testing the meter is comparatively very small.
Procedure:
1. Connect the circuit as per the circuit diagram.
2. Initially keep the two autotransformers at minimum output
voltage.
3. By varying the autotransformer in pressure circuit the voltmeter
reading is adjusted to a rated value.
4. By slowly varying the autotransformer reading is adjusted at
different values in steps.
5. The same procedure is repeated by connecting ammeter and
wattmeter 10A range.
Calculations:
True power=VI COS
%Error=(W-Wt/Wt) X 100
Observation table:
For Current 5A:
Voltmeter
(V)
Ammeter
(A)
Wattmeter
(W)
True
power
VICOS
%Error
(W-Wt/Wt) X
100
Electrical Measurements Lab
EEE Department Page 32
Observation table:
For Current 2.5A:
Voltmeter
(V)
Ammeter
(A)
Wattmeter
(W)
True
power
VICOS
%Error
(W-Wt/Wt) X
100
Model Graph:
+ %Error (0-5A) (0-25A)
W
-%Error
Result: Hence calibrations of the given LPF Wattmeter using Phantom
loading are done.
VIVA-VOICE
1.What is meant by phantom loading?
2.What is meant by low power factor?
3.How is electrostatic wattmeter superior to other types of wattmeter?
4.How does LPF wattmeter are differ from ordinary wattmeter?
5.What is formula for low power factor?
6.What are the errors in wattmeter?
7.What is meant by fictitious load?
8.What are the special features of a wattmeter suitable for working on
LPF circuits?
9.What is the rating of dimmer stat?
Electrical Measurements Lab
EEE Department Page 33
10.MEASUREMENT OF 3- REACTIVE POWER WITH 1-
WATTMETER
AIM: To measure the reactive power and to draw a characteristics curve
between KVAR and P.F with load current of 3 - system with 1-
wattmeter method.
NAME PLATE DETAILS:
power --- 3 KVA
voltage --- 415V
current --- 4.5A
frequency -----50 HZ
speed -----1440 rpm
APPARATUS:
NAME TYPE RANGE QTY
Voltmeter M.I 0-5A 1
Ammeter
M.I 0-600V 1
Wattmeter UPF 600V,5A
2
Tachometer --------- ---------- 1
Theory:This method can be used only when the load is balanced. The
connections are shown in the figure. The current coil is connected in one
of the lines and one end of the pressure coil to the same line, other end
being connections alternatively to the other two lines. Power
measurements are made in high voltage circuit connecting the wattmeter
to circuit through current and potential transformers. Voltmeters and
Ammeters are affected by only ratio errors while watt meters are
influenced in additional by phase angle errors.
CIRCUITDIAGRAM:
FIG:Measurement of 3 Reactive Power with 1- Wattmeter
Electrical Measurements Lab
EEE Department Page 34
Procedure:
1. Connect the ckt as per the ckt diagram.
2.The motor is started with the help of starter.
3.Varying the load on the motor by tighten the spring balance over the
pully of the motor.
4. Take the corresponding reading of the voltmeter, ammeter and
wattmeter.
5.Repeat the process for different values up to rated current
6.Calculate the reactive power.
Calculations:
Power factor, cos = W/VI
Reactive power, Q = VI sin
KVAR = Q/1000
Model graph:
IL(A)
Precautions:
1. loose connections should be avoided.
2. operate the starter care fully.
3. loads should not exceeding beyond their rating.
4. proper rating meters should be used.
5. check the connections before giving supply.
Result:
Hence the 3- phase reactive power is measured by 1-phase wattmeter
method and characteristics curve between KVAR and P.F with load
currents are plotted.
VIVA-VOICE
1. What is electrodynamometer type wattmeter ?
2. What is meant by KVAR ?
3. What is formula for reactive power?
4. What is meant by UPF wattmeter?
5. What is rating of three phase induction motor?
6. What is meant by TPST switch?
7. What is meant by DOL starter?
8. What is meant by balanced load?
9. What is meant by unbalanced load?
10. What is model graph of reactive power measurement?
KVAR
P.F
KVAR
P.F
F
Electrical Measurements Lab
EEE Department Page 35
Electrical Measurements Lab
EEE Department Page 36
Electrical Measurements Lab
EEE Department Page 37
Electrical Measurements Lab
EEE Department Page 38
Electrical Measurements Lab
EEE Department Page 39
Electrical Measurements Lab
EEE Department Page 40