1. POWER ON
2. PLACE DIST TILLED WATER AND PRESS ASPIRATE BUTTON
3. WAIT
*** APPERANCE OF THE TEST MENU***
(CALIBRATION OF REAGENT)
1. SELECT TEST TO BE CALIBRATED (CHECK PARAMETERS IF IT IS
THE SAME AS IN THE MANUAL OF WORKING PROCEDURES)
2. SELECT OK
3. TO THE QUESTION ASKED BY THE MACHINE: “DO YOU WANT TO
TEST REAGENT BLANK?” SELECT YES
• REAGENT BLANK IS DONE ONLY DURING CALIBRATION,
EXCEPT FOR BUN AND CREATININE-JAFFE WHICH IS ALSO
DONE EVERY TEST RUN
• AST/SGOT AND ALT/SGPT DO NOT REQUIRE REAGENT
BLANK
4. PLACE DISTILLED WATER IN POSITION ( MAKE SURE SIPPER IS
SUBMERGE IN LIQUID)
5. PRESS ASPIRATE BUTTON
6. WAIT
7. PLACE REAGENT BLANK IN POSITION
8. PRESS ASPIRATE BUTTON
9. WAIT
10. SELECT CAL
11. PLACE CALIBRATOR IN POSITION
12. PRESS ASPIRATE BUTTON
13. WAIT (THERE MAY BE INSTANCES WHERE IN MACHINE WILL
REQUEST TO REPEAT CALIBRATOR TEST, USING THE SAME
SAMPLE, ASPIRATE CALIBRATOR)
14. RESULT WILL APPEAR
15. RETURN (BACK TO TEST MENU)
16. WASH WITH DISTILLED WATER, BY PLACING DISTILLED WATER ON
POSITION AND PRESS RINSE BUTTON BEFORE AND AFTER
PROCEEDING TO THE NEXT TEST
TO ADD A NEW TEST
1. GO TO TEST MENU
2. SELECT BLANK BAR
3. SELECT EDIT
4. SELECT CHANGE MODE: TO CHOOSE BETWEEN TWO POINT, END
POINT or KINETICS
5. TO CHANGE THE UNIT, TEMPERATURE AND WAVELENGTH USE
THE ARROW POITING DOWNWARDS BESIDE
6. TO CHANGE THE SAMPLE VOLUME, REAGENT VOLUME, ASPIRATE
VOLUME, DELAY TIME, TEST TIME, NORMS, FACTOR, NAME AND
LINEAR SELECT THE KEYBOARD AT THE LOWER RIGHT CORNER
OF THE SCREEN.