INSTRUMENT MAINTENANCE PROCEDURE


Calibration Procedure Calibration procedure of nuclear level switch Type GM - 16

Procedure Number: IMP02/001/4/1999/01 Approval Name Signature

Revision: 1 Date

Prepared by

Checked by

Approved by

PT. PETROKIMIA NUSANTARA INTERINDO

Follow the steps below based on the use of the detector , either as a LOW LEVEL SWITCH or a HIGH LEVEL SWITCH . ALARM / TIME INTERVAL LEDS ( RED )

INTENSITY

LED ( YELLOW

OPERATE / STAND-BY Button

CALIBRATION Button

MODE LED ( GREEN ) 1 . Ensure that the interconnection cables have been connected correctly , then switch the source power 110 VAC to ON . 2 . Press the OPERATE / STAND-BY button ( button on the left ) , so that the gage is in the stand-by mode ( indicated by the GREEN LED being OFF ) 3 . Set the radiation field in the desired ALARM condition : a . As a LOW LEVEL ALARM : Open the source radiation . b . As a HIGH LEVEL ALARM : close the source radiation . 4 . Press and release the CALIBRATION button ( button on the right ) , the GREEN led will begin Flashing and the RED led will ON , wait for about 2 ( two ) minutes ( auto calibrate ) , and then both the GREEN led and the RED led will turn OFF . 5 . Set the radiation field in the desired NORMAL condition : a . As a LOW LEVEL ALARM : Close the source radiation . b . As a HIGH LEVEL ALARM : Open the source radiation . 6 . Press and release the OPERATE / STAND-BY button ( button on the left ) the green LED will begin flashing and red LED turn OFF, the gauge will be calibrate for two minutes . 7 . When calibration is completed the green LED will shift from flashing to ON , and the gage automatically begin operation.

164772699.doc

2 of 3

PT. PETROKIMIA NUSANTARA INTERINDO

Interpretation of GREEN and RED LEDS : RED GREEN OFF GREEN FLASHING GREEN ON RED OFF STAND-BY ALARM CALIBRATION # 2 NON-ALARM CONDITION OPERATE MODE NO ALARM RED FLASHES AT END OF TIME INTERVAL RED FLASHING FAILSAFE CALIBRATION # 1 ALARM CONDITION OPERATE MODE PROCESS ALARM ON

Wiring Connection Diagram : 01 02 03 04 05 06 07 08 09 10 11 Process Alarm 1 COMMON Process Alarm 1 OUTPUT ( N.O. ) Process Alarm 1 OUTPUT ( N.C. ) Process Alarm 2 COMMON Process Alarm 2 OUTPUT ( N.O. ) Process Alarm 2 OUTPUT ( N.C. ) FAILSAFE Alarm COMMON FAILSAFE Alarm OUTPUT ( N.O. / N.C. selectable ) INPUT POWER GROUND L 2 ( COMMON ) L 1 ( 115 / 230 VAC )

Connect the existing cable for INPUT POWER to terminal 10 / 11, and PROCESS ALARM OUTPUT to 01 / 02 ( N.O. ) action . Switch the Main POWER OFF from the Equipment Room first before disconnect or connect the cable to the terminal and ensure that the Tag No is correct .