6060 Spine Road, Boulder, CO U.S.A. 80301 (303) 444-2009 FAX (303) 444-9543
BULLETIN TOPIC:
B. IMPORTANT FACTS:
Negative TOC values are usually the result of measuring low levels of TOC with relatively high levels of IC,
and can be statistically relevant results on a validated and properly functioning TOC analyzer, however, many
customers find this undesirable. In an effort to mitigate this situation, the following procedure is designed to
fine tune both the IC and TC channels of the analyzer.
In the case of off-line water samples, the TOC may be low, and the IC relatively high (off-line samples absorb
carbon dioxide from the atmosphere). Off-line samples may equilibrate over a 250-400 ppb IC range.
Depending on the extent to which samples have equilibrated with the atmosphere, there could be different
levels of IC for the same water samples if they are analyzed at different times.
800 19025 C
Page 1 of 2
C. TC ZERO OFFSET:
If it has been determined by an Ionics Instruments Service Representative to make an adjustment to the TC zero
offset, perform the following:
Note: Be sure to record all TC zero offset adjustments and communicate the history to the Ionics Instruments
Service Representative assisting you. If you have made more than two adjustments to the TC zero offset in the last
six months, there may be underlying issues with your water system or analyzer. Please inform our representatives
accordingly.
Modification of operation and maintenance manual procedure in chapter 8:
1. Obtain a 1 L or larger flask (preferably a beaker to offer a large area of exposure to the air) and fill the
container approximately half-full with low TOC water. Leave the sample open to the atmosphere overnight so
that it may equilibrate with room CO2 levels (200-400 ppb typically).
2. If necessary, loosely cover the mouth of the flask with parafilm (being careful to leave openings) to minimize
organic contamination that may be present in the lab.
3. Turn the UV lamp off.
4. Set up the instrument for sample mode: On Line sampling.
5. Set reagent flows: Acid = 0.20 L/min and Oxidizer = .01 L/min.
6. Analyze the water sample prepared in Step 1 (usually done overnight).
7. Calculate the average value of the last 5 TOC measurements.
8. Calculate the new TC Zero Offset using the following equation:
New TC Zero Offset = (Initial TC Zero Offset) (Avg. TOC value from step7)*
*Be sure to note the sign of the Average TOC value and input the value accordingly; it may be negative.
9. Enter in the new TC Zero offset.**
10. Rerun with new offset for 1-2 hours. TOC should be close to zero PPB, if not repeat Steps 7-10.
**If the absolute values of the new TC Zero Offset is greater than 50 ppb, contact an Ionics Instruments Service Representative
before inputting the new TC Zero Offset value.
D. CALIBRATION ADJUSTMENT:,
After speaking with an Ionics Instruments Service Representative and eliminating the possibility of other potential
reasons for negative TOC results, it may be determined that adjustment of either the IC or TC calibration constants
is necessary. You should perform a calibration adjustment only if you have been directed by a service
representative to do so (See operation and maintenance manual: IC calibration constant procedure calibration
chapter).
ATTACHMENTS:
YES
CAN BE RELEASED TO
NO
IDENTIFICATION:
CUSTOMERS:
YES
800 19025 C
Page 2 of 2