Distance and Distance semiannual Distance checked Physicist Assuring distance and
indicators should with tape centering are accurate,
centering
be +- 10 %, and measure, visual avoid the need for
indicators centering inspection of repeated exposure
accuracy indicators should collimator light
be +- 2% beam.
Angulator or Angles should be semiannual Large protractor Physicist Accurate readings for
+- 1%, with and level positioning baselines
protractor
adjustments
accuracy made when
necessary
Kilovoltage +- 5 kVp semiannual Computerized Physicist Maintaining image
dosimeters quality
accuracy
Timer accuracy +- 5% of label semiannual Computerized Physicist Exposure cannot be
dosimeters or a predicted unless time
spinning top for 1 is accurate
phase and
motorized
synchronous top
for 3 phase
mR/mAs and +- 10% semiannual Comparing Physicist Exposure cannot be
mR/mAs predicted unless mA is
milliamperage
measurements accurate, maintain
linearity after both time iamge quality
and kVp accuracy
have been
verified
Computerized
Quality Control Test Table
Create a table of commonly performed QC tests. Include the categories listed.
Sammy Puccini
Notebook 11
dosimetry
AEC
Exposure OD +-0.1 Annually Same procedure Physicist Same as diagnostic
but with subject radiography systems
reproducibility
phantom
Radiographs must
be produced and
measured with
densitometer
Quality Control Test Table
Create a table of commonly performed QC tests. Include the categories listed.
Sammy Puccini
Notebook 11
Technologist Responsibilities
Daily QC: General system inspection: sensitometry (laser-generated seneitometry strip, film densities
measurement, inspect images (dust, scratches, and marks), network queue check (send images to PACS)
Artifact identification
Problem reporting