Professional Documents
Culture Documents
i-2 2014/07
Viva-ProE ™ System Operator’s Guide
2014/07 i-3
Viva-ProE ™ System Operator’s Guide
i-4 2014/07
Viva-ProE ™ System Operator’s Guide
Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-2
Absorbance Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-3
Measuring Absorbance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-3
Measurement Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4
Calculating Concentrations from Absorbances . . . . . . . . . . . . . . . . . . . 16-5
Measurement with Reagent Blank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-5
Measurement with Reagent Blank and Sample Blank . . . . . . . . . . . . . . . . . . . . . . . 16-5
Bichromatic Endpoint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-6
Interpolation on Calibration Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-6
Calibration Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-7
Standard Cubic Spline Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-7
Modified Cubic Spline Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-8
NLLS Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-8
4 Parameter Logit-log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-8
5 Parameter Logit-log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-8
5 Parameter Exponential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-9
Linear Regression and Point-to-Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-9
Explanation of System Features . . . . . . . . . . . . . . . . . . . . . . . . 16-10
Integrity Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-10
Absorbance Check for Calibrated Tests . . . . . . . . . . . . . . . . . . . . . . . . . 16-10
Westgard Rules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-10
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
2014/07 1-1
System Features Viva-ProE ™ System Operator’s Guide
Intended Use
ELITech Clinical system Viva-ProE ™ is an automated clinical chemistry system
intended for use in clinical laboratories. It is intended to be used for a variety of
assay methods that have been applied to spectrophotometric techniques. The
system has a spectrophotometric system for measurement of analytes using
spectrophotometric techniques, such as end point, rate and turbidimetric
assays.
1-2 2014/07
Viva-ProE ™ System Operator’s Guide System Features
System Users
This Operator’s Guide provides information for the following clinical laboratory
professionals who use the system:
User Description
Operator Performs daily operating tasks such as start of day tasks, calibration and
control measurements, loading and unloading reagents and samples,
examining test results, and basic maintenance.
Supervisor In addition to the Operator functions, the Supervisor can also accept or
reject flagged results, accept or reject pending calibration results, and add
comments to a QC history controlled result.
Administrator In addition to the Supervisor functions, the Administrator can create new or
edit existing controls, create new calibrators, lots, and tests, enable or
disable tests, edit test and instrument parameters, extend expired control
results, override Westgard rules, and exclude controls from the QC history.
2014/07 1-3
System Features Viva-ProE ™ System Operator’s Guide
Hardware Overview
Figure 1-1: Viva-ProE ™ System
1 Reagent Rotor
2 Probes
3 Cuvette Rotor
4 Sample Rotor
1-4 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Cuvette Rotor
The cuvette rotor contains 48 cuvettes. The system requires a measuring
volume between 200 µL - 400 µL. The cuvette rotor uses a heater installed
below the rotor and a heated cover to keep the cuvettes at
37 °C (98.6 F).
The stirrer stirs the reaction mixture. It operates by lowering the probe into the
eccentric washer; the eccentric washer vibrates the probe in the reaction
mixture. The cuvette rotor then rotates the sample until it reaches the
measurement position. Once the system completes all measurements, the
wash arm washes and dries the cuvette, then deposits the waste into the
waste container.
Figure 1-2: Cuvette Rotor
1 Cuvette Rotor
2 Wash Arm
3 Stirrer
2014/07 1-5
System Features Viva-ProE ™ System Operator’s Guide
1 Reagent Arm
2 Reagent Probe
3 Reagent Rotor
5 Rinsing Stations
6 Sample Arm
7 Sample Probe
8 Sample Rotor
1-6 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Sample Rotor
The sample rotor holds the sample tubes. All positions are numbered and
displayed on the rotor screens. The samples are divided over three rings.
The sample rotor is covered to protect the samples from light and other
ambient influences. The cover has openings to let the sample probe pass
through.
Figure 1-4: Sample Rotor
2014/07 1-7
System Features Viva-ProE ™ System Operator’s Guide
Reagent Rotor
The reagent rotor is designed to hold both 14 mL and 28 mL reagent bottles.
An optional adapter is also available that allows 14 mL bottles to be used in a
28 mL position. Reagent positions are defined in the system software.
The reagent rotor is covered to protect the reagent from light, heat, and other
ambient influences. The cover has openings to let the reagent probe pass
through. The base of the reagent rotor is cooled.
Figure 1-5: Reagent Rotor
1 28 mL Position
2 14 mL Position
1-8 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Pipettor
The pipettor unit uses a 1 mL syringe for pipetting reagent and a 100 µL
syringe for pipetting sample materials.
Figure 1-6: Pipettor
1 Probe Connection
2 Valve
3 Sample Syringe
4 Plunger Handle
5 Reagent Syringe
6 Water Connection
2014/07 1-9
System Features Viva-ProE ™ System Operator’s Guide
2 Float Switch
3 Waste Container
4 Water Filter
1 - 10 2014/07
Viva-ProE ™ System Operator’s Guide System Features
1 Cap
2 Float Switch
2014/07 1 - 11
System Features Viva-ProE ™ System Operator’s Guide
1 Filter
2 Quartz-iodine Lamp
1 - 12 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Cooling Unit
The Viva-ProE ™ System uses a cooling unit to extend the on-board stability of
reagents on the system. The cooling unit keeps the reagent’s temperature
between 8 - 12 °C (46 - 53 °F), up to an ambient temperature of 25° C (77° F).
Above 25° C (77° F) the reagent temperature increases relative to the ambient
temperature. The cooling unit is located behind the front panel of the system.
The cooling unit uses a pump to circulate the cooling liquid and has a sight
window with markers that indicate minimum and maximum cooling liquid
levels.
Figure 1-10: Cooling Unit
1 Cap
2014/07 1 - 13
System Features Viva-ProE ™ System Operator’s Guide
2 Sample Rotor
1 - 14 2014/07
Viva-ProE ™ System Operator’s Guide System Features
2014/07 1 - 15
System Features Viva-ProE ™ System Operator’s Guide
Computer
The Viva-ProE ™ System includes a Personal Computer (PC) on an adjustable
mounting arm. The panel PC runs the Microsoft® Windows® operating system
and the user interface software for the system. The mechanical components of
the system are controlled by electronic boards located in the back of the
system.
The touchscreen offers an easy-to-use direct user interface to the system
functions.
Note: Ensure the operator’s fingers or gloves are clean when
touching the screen. A clean, plastic stylus may also be used.
1 - 16 2014/07
Viva-ProE ™ System Operator’s Guide System Features
2014/07 1 - 17
System Features Viva-ProE ™ System Operator’s Guide
Software Overview
System Access
The Viva-ProE ™ System software requires authentication by the operator. The
operator must have a user name and password to access the system. The
operator must login to the system using the following access screens:
• Windows Login
• System Software Login
Windows Login
In the Windows system, the following user accounts are available:
Table 1-2: Windows Login Account Types
Analyzer Operator
User Description
Operator Performs daily operations of the system, such as start of day tasks, calibration
and control measurements, loading reagents and samples, requesting and
running tests, examining test results, and repeat or re-running tests for some
flagged results.
Supervisor In addition to the Operator functions, the Supervisor accepts or rejects flagged
test results.
Superuser Login
Switching to another user is possible without logging out the previous user.
This login allows a supervisor to temporarily login to perform tasks requiring
supervisor level access.
1 - 18 2014/07
Viva-ProE ™ System Operator’s Guide System Features
2014/07 1 - 19
System Features Viva-ProE ™ System Operator’s Guide
1 11
Start of Day Screen Icon Print Screen Icon
2 12
Reagents Screen Icon Help Screen Icon
3 13
Load Calibration Screen Icon Events Screen Icon
4 14
Load Controls Screen Icon Quality Control Screen Icon
6 16
Load Patients Screen Icon Tasks Screen Icon
7 17
Results Screen Icon Reset Icon
8 18
Unload Screen Icon System Stopped Icon
9 19
End of Day Screen Icon Stop System Icon
10 20
Login / Logout Icon Open or Close Information Panel window icon
1 - 20 2014/07
Viva-ProE ™ System Operator’s Guide System Features
1 Start of Day
2 Description
3 Status
2014/07 1 - 21
System Features Viva-ProE ™ System Operator’s Guide
Reagent Screen
The Reagent screen displays the status of reagents and is used for loading,
unloading, replacing, and extending reagents.
Figure 1-17: Reagent Screen Components
1 Clear
2 Reagent Rotor
3 Print
4 Reagent Info
5 Switch Rotor
6 Extend
7 Replace
8 Unload
9 Load
Color Status
1 - 22 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Indicator Description
2014/07 1 - 23
System Features Viva-ProE ™ System Operator’s Guide
Item Displays
Reagent The role of the reagent, such as first (R1), second (R2), or third
(R3).
Note: For special fluids, the name of the fluid will be displayed.
Status The current state of the reagent. See Table 1-8 “Reagent Status
Icons” for more information.
Onboard Stability The remaining time during which the reagent can still be used
when it is kept on the system.
Remaining Tests The number of tests that can be performed with the amount of
remaining reagent.
Completed Tests The number of tests that were performed using the current
reagent bottle.
Icon Description
Reagent Empty
Reagent Expired
1 - 24 2014/07
Viva-ProE ™ System Operator’s Guide System Features
1 Load
2 Unload
3 Sample Rotor
4 Review
Item Displays
Tests The name of the test for the requested Calibration and or
Reagent Blank.
Calibrator Lot The lot number and lot expiration of the Calibrator.
Note: The lot number is listed on the package label of the
Calibrator. Do not use beyond the expiration date.
2014/07 1 - 25
System Features Viva-ProE ™ System Operator’s Guide
1 Summary
2 Details
3 Calibration
4 Reagent Blank
5 Cancel
6 Accept
7 Close
1 - 26 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Item Displays
Tests The name of the test for the requested Calibration and or
Reagent Blank.
Blank Expiration The time remaining before the reagent blank expires.
Calibration Expiration The time remaining before the calibration measurement expires
Note: If N/A is displayed, no calibration was performed and was
not required for the test requested.
2014/07 1 - 27
System Features Viva-ProE ™ System Operator’s Guide
Button Description
Use Previous Discards the calibration and revert to the previous calibration
results.
Accept Confirms the request and opens the Load Calibration screen.
Icon Description
Calibrator loaded; request to change a test cannot not be made at this time.
1 - 28 2014/07
Viva-ProE ™ System Operator’s Guide System Features
1 Load
2 Unload
3 Sample Rotor
4 Review
Item Displays
Tests The name of the test for which the Control is requested.
2014/07 1 - 29
System Features Viva-ProE ™ System Operator’s Guide
1 Summary
2 Details
3 Cancel
4 Accept
5 Close
6 Extend
Item Displays
Tests The name of the test for which the Control is requested.
1 - 30 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Item Displays
Tests The name of the test for which the Control is requested.
Control Expiration The time remaining before the Control Measurement expires.
Note: The Control Expiration displays when the Control
Measurement expires, not when the Control lot expires. The date
is displayed as Expiration in the control information section to
the right of the tests list.
Item Displays
2014/07 1 - 31
System Features Viva-ProE ™ System Operator’s Guide
1 Delete
2 Edit
3 Create
4 Load
5 Unload
6 Sample Rotor
7 Print List
Item Displays
Search ID A text box used to enter a new patient or search for an existing
patient identification number.
1 - 32 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Icon Description
The sample has been loaded onto the rotor and is ready for processing.
2014/07 1 - 33
System Features Viva-ProE ™ System Operator’s Guide
1 Routine 10 Female
4 Replicates 13 Category
7 Accept 16 Physician
Note: * If selected, remembers the last test selection for use with
the next sample.
1 - 34 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Item Displays
2014/07 1 - 35
System Features Viva-ProE ™ System Operator’s Guide
1 Reagent Blanks
2 Loaded
3 Recent
5 Patients
6 Search
7 Cancel
8 Reject
9 Accept
10 Dilute
11 Repeat
12 Export
13 Resend to LIS
14 Print
15 Summary Report
1 - 36 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Button Description
View Calibration Views the last calibration results for a desired test.
View Blank Views the last reagent blank results for a desired test.
Item Displays
ID Sample identification.
Icon Description
2014/07 1 - 37
System Features Viva-ProE ™ System Operator’s Guide
1 Reagent Blanks
3 Patients
4 Search
5 Print
6 Export
1 - 38 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Button Description
2014/07 1 - 39
System Features Viva-ProE ™ System Operator’s Guide
The Results Screen - Calibrations and Controls details screen displays the
following information:
Table 1-23: Results Screen - Calibrations and Controls List Box Screen
Item Displays
Icon Description
1 - 40 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Unload Screen
The Unload screen displays a list of currently loaded samples and an image of
the sample rotor. The screen is used to unload completed samples.
Note: Only validated samples can be unloaded. If one or more tests
results for the sample are flagged, operator intervention is
required before the sample can be released for unloading.
Figure 1-26: Unload Screen
1 Type
2 Name
3 Sample ID/Level
4 Pos
5 Unload All
6 Unload
7 Rotor
2014/07 1 - 41
System Features Viva-ProE ™ System Operator’s Guide
1 End of Day
2 Description
3 Status
1 - 42 2014/07
Viva-ProE ™ System Operator’s Guide System Features
Events Screen
The Events screen lists all events that were recorded and is used to
acknowledge events and or silence the audio alarm. The list displays the time
and date of each event with a description and information about a possible
action that was performed.
Figure 1-28: Events Screen
1 Silence Alarm
2 Stop Instrument
3 Acknowledge
4 Reconnect LIS
5 Stay Disconnected
Item Displays
2014/07 1 - 43
System Features Viva-ProE ™ System Operator’s Guide
1 - 44 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
2014/07 2-1
System Operation Viva-ProE ™ System Operator’s Guide
2-2 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
Preparing Samples
The quantity of patient sample needed to perform the desired tests is a
combination of sample volume per test, buffer volume per test, and dead
volume per test.
The following is a definition of the Buffer and Dead Volumes:
• The buffer volume is a small amount of the sample aspirated prior to
testing to remove water from the interior wall lining of the sample probe.
• The dead volume is the patient sample material that remains in the sample
tube or cup after testing is completed.
CAUTION: Do not use the remaining sample volume material; what remains
is too low to be accurately aspirated by the system.
Use the table below to determine volume requirements for the sample tubes
or cups.
Table 2-1: Sample Volume Requirements
Sample Volume Per Test* Buffer Volume Per Test Dead Volume Per Test**
1 - 10 µL 5 µL 250 µL
10 - 20 µL 10 µL 250 µL
20 - 30 µL 15 µL 250 µL
* Note: This is the sample volume required for each test, as specified in the reagent package insert.
** Note: When using pediatric cups with the pediatric adaptors, the dead volume is ~ 100 µL.
CAUTION: Ensure that the sample/reagent mixture does not contain any
bubbles, blood clots, dust, or other insoluble contaminants.
Insoluble contaminants contained within a sample may
negatively affect patient results.
2014/07 2-3
System Operation Viva-ProE ™ System Operator’s Guide
2-4 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
1 Quiet Zone
• Ensure that the label does not touch the tube closure.
• Ensure that the label is not transparent to prevent reading the barcode on
an underlying label.
• Ensure that the label has no creases.
• Do not place clear tape over the label.
• Ensure that the label is free of moisture.
• Ensure that the adhesive is strong enough to prevent the label from
peeling loose or becoming partially detached from the tube.
2014/07 2-5
System Operation Viva-ProE ™ System Operator’s Guide
2-6 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
Daily Setup
Turning on the System
1. On the back of the system, turn main disconnect switch On.
2. Turn the system PC On.
3. From the Windows screen, select Analyzer.
Login
1. From the Home screen, select the Login icon.
2. Enter Username and Password then select Login.
2014/07 2-7
System Operation Viva-ProE ™ System Operator’s Guide
1 Cap
2 Float Switch
3 Waste Container
2-8 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
1 Cap
3 Water Filter
1. Open the Treated Water Container door using the handle at the bottom
to pull the door outward.
2. Remove the Treated Water Container cap.
3. Slide the Treated Water Container out of the system.
4. Fill the Treated Water Container with 12.5 mL of system solution and 5 L
of distilled or purified water (ASTM Type II Grade or above).
5. Push the Treated Water Container inside the system and ensure that it is
in the proper position.
6. Replace and tighten the cap; close the cabinet door.
2014/07 2-9
System Operation Viva-ProE ™ System Operator’s Guide
1 Cap
2 Float Switch
2 - 10 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
Checking Syringes
Figure 2-6: Reagent and Sample Syringes
1 Probe Connection
2 Valve
3 Sample Syringe
4 Probe Connection
5 Reagent Syringe
6 Water Connection
1. Open the Pipettor door using the handle at the bottom to pull the door
outward.
2. Inspect the Reagent and Sample syringes and verify that no air bubbles or
leaks are present.
2014/07 2 - 11
System Operation Viva-ProE ™ System Operator’s Guide
2 - 12 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
2014/07 2 - 13
System Operation Viva-ProE ™ System Operator’s Guide
3. From the Home screen, select the End of Day icon. The screen
displays a list of tasks that must be performed at the end of each day.
4. Perform all tasks that show an orange flag or exclamation mark.
5. Select the checkbox next to the task to mark as complete.
2 - 14 2014/07
Viva-ProE ™ System Operator’s Guide System Operation
Emergency Shutdown
When the Viva-ProE ™ System must be turned off immediately to ensure the
safety of personnel and equipment and a controlled shutdown is not possible,
use the following procedures to perform an emergency shutdown.
CAUTION: Selecting Stop will immediately halt all sample processing and
testing, which could result in the loss of patient data.
CAUTION: The power to the system is not switched off when an emergency
stop is performed. If the electrical power to the system must be
interrupted completely, go to the back of the system and turn the
main disconnect switch Off and or disconnect the Power Cable.
WARNING: Opening the system cover while the system is working does
Not stop pipetting and may cause injuries or damages by
touching a moving probe or rotor. Do not open the analyzer
cover unless the system is stopped or paused.
2014/07 2 - 15
System Operation Viva-ProE ™ System Operator’s Guide
4. From the Windows desktop, double-select the Pro Series Software icon.
5. Select Start Application Software.
6. From the Home screen, select the Login icon.
2 - 16 2014/07
Viva-ProE ™ System Operator’s Guide Reagents
Chapter 3: Reagents
2014/07 3-1
Reagents Viva-ProE ™ System Operator’s Guide
Reagents
Checking Reagents
3-2 2014/07
Viva-ProE ™ System Operator’s Guide Reagents
c. Status
Table 3-1: Reagent Status Icons
Icon Description
Reagent Empty
Reagent Expired
Replacing a Reagent
2014/07 3-3
Reagents Viva-ProE ™ System Operator’s Guide
Loading a Reagent
Figure 3-1: Load Reagent Screen
5 Cancel 11 Test
6 Accept
3-4 2014/07
Viva-ProE ™ System Operator’s Guide Reagents
10. Remove the Reagent bottle cap and insert the bottle into the desired
rotor position.
11. Select Accept.
Note: When loading probe rinse or acid solution fluid, select
Special Fluid radio button, than follow steps 7 through 11.
Unloading a Reagent
2014/07 3-5
Reagents Viva-ProE ™ System Operator’s Guide
4. Set the Flange aside and carefully lift the Reagent Rotor out of the
system.
5. Store the reagents and rotor following the standard laboratory operating
procedures.
3-6 2014/07
Viva-ProE ™ System Operator’s Guide Reagents
2. Slightly rotate the Reagent Rotor left and right until it falls over the
positioning pin and locks into place.
3. Reinstall the Flange.
2014/07 3-7
Reagents Viva-ProE ™ System Operator’s Guide
3-8 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
2014/07 4-1
Sample Management Viva-ProE ™ System Operator’s Guide
Controls
4-2 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
Icon Description
Control Expired
2014/07 4-3
Sample Management Viva-ProE ™ System Operator’s Guide
Unloading Controls
4-4 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
Patients
Requesting Patient Tests
2014/07 4-5
Sample Management Viva-ProE ™ System Operator’s Guide
1. Using only the outer two load positions of the Sample Rotor, place the
Sample Tubes on the rotor with the barcodes labels facing out.
Figure 4-1: Sample Rotor with Internal Barcode Shown
2. Repeat the above procedure until all of the Sample Tubes are loaded.
3. Install the Sample Rotor cover.
4. Close the system cover.
4-6 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
2014/07 4-7
Sample Management Viva-ProE ™ System Operator’s Guide
4-8 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
WARNING: Opening the system cover while the system is working does
NOT stop pipetting and may cause injuries or damage by
touching a moving probe or rotor. Do not open the system
cover unless the system has stopped or paused.
2014/07 4-9
Sample Management Viva-ProE ™ System Operator’s Guide
WARNING: Opening the system cover while the system is operating does
NOT stop pipetting and may cause injuries or damage if a
moving probe or rotor is touched. Do not open the system
cover unless the system has stopped or is paused.
4 - 10 2014/07
Viva-ProE ™ System Operator’s Guide Sample Management
9. Select Repeat.
Note: Only users with Supervisor access or above can perform
step 10 of this procedure.
10. Select Accept to accept or Reject to reject flagged results.
Unloading a Sample
2014/07 4 - 11
Sample Management Viva-ProE ™ System Operator’s Guide
4 - 12 2014/07
Viva-ProE ™ System Operator’s Guide Calibration
Chapter 5: Calibration
2014/07 5-1
Calibration Viva-ProE ™ System Operator’s Guide
Calibration
Loading Calibrators
Performing Calibrations
Icon Description
5-2 2014/07
Viva-ProE ™ System Operator’s Guide Calibration
Unloading Calibrators
2014/07 5-3
Calibration Viva-ProE ™ System Operator’s Guide
5-4 2014/07
Viva-ProE ™ System Operator’s Guide Quality Control
2014/07 6-1
Quality Control Viva-ProE ™ System Operator’s Guide
3 Reset Westgard Rules Resets the Westgard Rules for a control-test combination.
Note: Available only if the Control uses Westgard Rules.
4 Override Rules Overrides the rules (Westgard or Basic) for the control-test
combination.
6-2 2014/07
Viva-ProE ™ System Operator’s Guide Quality Control
Button Description
Item Displays
2014/07 6-3
Quality Control Viva-ProE ™ System Operator’s Guide
6-4 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Chapter 7: Maintenance
2014/07 7-1
Maintenance Viva-ProE ™ System Operator’s Guide
Introduction
As an important part of a complete quality assurance program, proper service
and maintenance:
• Minimizes down time
• Provides optimal performance
• Preserves records for inspection and accreditation
Note: Some maintenance work may be performed by service
engineers of the system supplier.
7-2 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7-3
Maintenance Viva-ProE ™ System Operator’s Guide
Maintenance Schedule
Table 7-1 indicates the recommended frequency of preventative maintenance
procedures. It can be used as a Maintenance Quick Reference Guide.
Frequency Procedure
7-4 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Maintenance - Daily
Note: It may be necessary to perform these procedures more than
once daily, depending on the laboratory workload.
• Emptying Waste
• Filling Treated Water Container
• Emptying Concentrated Waste Container
• Checking Syringes
• Checking Cuvette Blank
• Checking / Refilling 10% Cleaning Solution
• Checking Daily Report
• Checking 0.1N HCl Cleaning Solution
• Priming the System
• Checking Reagent Inventory
• Archiving Results Database
2014/07 7-5
Maintenance Viva-ProE ™ System Operator’s Guide
1 Cap
2 Float Switch
3 Waste Container
7-6 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
1 Cap
3 Water Filter
1. Open the Treated Water Container door using the handle at the bottom
to pull the door outward.
2. Remove the Treated Water Container cap.
3. Slide the Treated Water Container out of the system.
4. Fill the Treated Water Container with 12.5 milliliters of system solution
and 5 liters of distilled or purified water.
5. Push the Treated Water Container inside the system and ensure that it is
in the proper position.
6. Replace and tighten the cap then close the cabinet door.
2014/07 7-7
Maintenance Viva-ProE ™ System Operator’s Guide
1 Cap
2 Float Switch
7-8 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Checking Syringes
Figure 7-4: Reagent and Sample Syringes
1 Probe Connection
2 Valve
3 Sample Syringe
4 Probe Connection
5 Reagent Syringe
6 Water Connection
1. Open the Pipettor door using the handle at the bottom to pull the door
outward.
2. Inspect the Reagent and Sample syringes and verify that no air bubbles or
leaks are present.
2014/07 7-9
Maintenance Viva-ProE ™ System Operator’s Guide
b. Expiration date.
Note: If the cuvette blank has expired or about to expire, proceed to
step 3.
3. Select Run to perform a cuvette blank.
4. Select Yes at The cuvette blank operation will take some time to
complete, during which the instrument will be unavailable for tests.
Do you wish to continue? prompt.
Note: When the cuvette blank results are within the defined
acceptance ranges, the measurement is automatically
accepted and the information on the screen updated. No
further action is required.
7 - 10 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 11
Maintenance Viva-ProE ™ System Operator’s Guide
Maintenance - Weekly
Item Description
7 - 12 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Maintenance - Monthly
• Cleaning the Treated Water and Waste Containers
• Cleaning the Concentrated Waste Container
Item Description
0.1N NaOH
2014/07 7 - 13
Maintenance Viva-ProE ™ System Operator’s Guide
2 Float Switch
3 Waste Container
4 Water Filter
1. Open the Treated Water and Waste Container door using the handle at
the bottom to pull the door outward.
2. Remove both the Treated Water and Waste Container caps.
3. Slide the Treated Water and Waste Containers out of the system.
4. Empty the Treated Water and Waste Containers following standard
laboratory operating procedures.
5. Pour a small amount of 0.1N NaOH solution (enough to cover the
bottom) into the Treated Water and Waste Containers.
6. Swirl the 0.1N NaOH solution around the container then empty the
Treated Water and Waste Containers following standard laboratory
operating procedures.
7. Rinse both containers three times using distilled or purified water.
8. Fill the Treated Water Container with 12.5 mL of system solution and 5
liters of distilled or purified water.
9. Push the Treated Water and Waste Containers inside the system and
ensure they are in the proper position.
10. Replace and tighten the container caps then close the cabinet door.
7 - 14 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Item Description
0.1N NaOH
1 Cap
2 Float Switch
2014/07 7 - 15
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 16 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 17
Maintenance Viva-ProE ™ System Operator’s Guide
3. Select Replace Drying Block to open the Replace Drying Block screen.
4. Select Replace Drying Block.
5. Disconnect the tubing [2] from the front end of the Drying Block and
move the tubing aside.
6. Remove the Positioning Block Screw [1], using caution not drop the
screw into the system.
7. Remove the Positioning Block [6] by pulling it up and off the
Needle Plate [5].
8. Remove the Drying Block [3] by pulling it up and to the right.
9. Place the new Drying Block [3] into the slit and press down into the bush.
10. Reinstall the Positioning Block [6].
11. Replace the Positioning Block Screw [1] but do not tighten.
12. Select Continue.
13. Tighten the Positioning Block Screw [6].
14. Reattach the tubing [2] to the Drying Block [3].
15. Select Continue.
16. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
17. Install the Cuvette Rotor Cover.
7 - 18 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
1 Pulley
2 Stirrer Belt
3 Eccentric Pulley
1. Starting from the Pulley [1], use a pair of tweezers to pull the
Stirrer Belt [2] off the Pulley [1].
2. Starting from the Eccentric Pulley [3], use the same tweezers to pull the
new Stirrer Belt [2] around the Pulley [1].
2014/07 7 - 19
Maintenance Viva-ProE ™ System Operator’s Guide
7. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
1 Cap
3 Tubing
4 Filter Holder
5 Water Filter
1. Open the Treated Water door using the handle at the bottom to pull the
door outward.
2. Remove the Treated Water Container cap [1].
7 - 20 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
3. Remove the tubing [3] with the filter out of the container.
4. Unscrew the filter [5] from the filter holder [4].
5. Screw the new filter [5] onto the filter holder [4].
6. Place the tubing with the filter into the Treated Water Container.
7. Push the Treated Water Container inside the system and ensure it is in
the proper position.
8. Replace and tighten the container cap then close the cabinet door.
13. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
2014/07 7 - 21
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 22 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Maintenance - Yearly
• Replacing Syringes
• Replacing the Photometer Lamp
Note: Customers with a Siemens maintenance contract will have
these services performed by a Siemens service technician.
Replacing Syringes
Figure 7-10: Replacing Syringes
1 Valve Block
2 End Cap
3 Glass Barrel
4 Plunger Handle
5 Screw
6 Drive Pin
2014/07 7 - 23
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 24 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
20. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
2014/07 7 - 25
Maintenance Viva-ProE ™ System Operator’s Guide
WARNING: The lamp gets very hot when switched on. Ensure the
lamp has cooled down before removing it from the system.
CAUTION: Wear latex gloves when replacing the lamp. Quartz -iodine lamps
are very sensitive. Touching the lamp with your hands will
dramatically reduce its lifetime.
Figure 7-12: Replacing the Photometer Lamp
1 Quartz-iodine Lamp
2 Lamp Holder
3 Mounting Screw
7 - 26 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2 Screw
2014/07 7 - 27
Maintenance Viva-ProE ™ System Operator’s Guide
1 Lamp
2 Fastening Screw
3 Tapped Hole
7 - 28 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
WARNING: Looking into the lamp can hurt the Operator’s eyes.
Avoid looking directly into the lamp while it is switched on.
Figure 7-15: Adjusting the Lamp Unit
1 Fastening Screw
2 Vertical Adjustment
3 Horizontal Adjustment
1. Remove the Front Panel by lifting up the panel from the bottom and
pulling it forward.
2. Remove the Lamp Unit Cover:
2014/07 7 - 29
Maintenance Viva-ProE ™ System Operator’s Guide
2 Screw
7 - 30 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
14. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
15. Perform a Cuvette Blank Measurement.
2014/07 7 - 31
Maintenance Viva-ProE ™ System Operator’s Guide
1 Stirrer Knob
2 Center Knob
CAUTION: Wear latex gloves when replacing the Cuvette Rotor. Hold the
Cuvette Rotor by the Center Knob [2] only. Do not touch the
Cuvettes.
6. Use the Knob [1] to pull the Stirrer Unit up.
7. Use the Center Knob [2] to remove the Cuvette Rotor.
7 - 32 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
8. Install the new Cuvette Rotor by carefully aligning the four notches of the
new Cuvette Rotor above the slits on the Rotor Holder.
9. Push down the Stirrer Knob [1].
10. Select Continue.
11. Select Yes at the “Reset Cuvette Rotor Counter” prompt.
12. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
13. Replace the Cuvette Rotor Cover.
14. Perform a Cuvette Blank.
2014/07 7 - 33
Maintenance Viva-ProE ™ System Operator’s Guide
Maintenance - As Needed
Perform the following procedures as needed:
• Cleaning the System
• Cleaning the Reagent Rotor Compartment
• Filling the Cooling Fluid
7 - 34 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
c. Set the Flange aside and carefully lift the Reagent Rotor out of the
system.
5. Clean the Reagent Rotor:
a. Pour all condensation fluid out of the rotor.
b. Dry the rotor on the inside and outside with paper tissue or a clean,
dry cloth.
6. Dry the Stator (base on which the rotor rests) with a paper tissue or a dry,
clean cloth.
7. Reinstall the Reagent Rotor:
a. Carefully lower the Reagent Rotor ensuring to align the white dot
on the Reagent Rotor[2] with the white dot on the driving axle [1].
(See Figure 4-9.)
2014/07 7 - 35
Maintenance Viva-ProE ™ System Operator’s Guide
b. Slightly rotate the Reagent Rotor left and right until it falls over the
positioning pin and locks into place.
c. Reinstall the Flange.
8. Reinstall the Reagent bottles.
9. Install the Reagent Rotor cover.
7 - 36 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
1. Remove the Front Panel by lifting up the panel from the bottom and
pulling it forward.
2. Check the Cooling Fluid Level. The level should be between the
Minimum [3] and Maximum [2] markers.
Note: If the cooling fluid level is below the Minimum [3] marker,
proceed to step 3.
Note: Cooling liquid must be diluted with distilled water. Use 1 part
cooling liquid on 4 parts distilled water.
3. Fill the Cooling Fluid:
a. Remove the Cooling Fluid Fill Cap [1].
b. Add Cooling Fluid until the level is just below the Maximum [2]
marker.
c. Wait a few minutes and then recheck the Cooling Fluid level. If the
Cooling Fluid level is between Minimum [3] and Maximum [2],
proceed to step D.
2014/07 7 - 37
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 38 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
Maintenance Logs
Use the Maintenance Logs on the following pages to record Daily, Weekly, and
Monthly, Every 3 Months, Every 6 Months, Yearly, Every 10,000 Tests, and As
Needed maintenance procedures. Recording these tasks helps track
instrument performance. Complete each column for each task completed.
2014/07 7 - 39
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 40 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 41
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 42 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 43
Maintenance Viva-ProE ™ System Operator’s Guide
Archive Results
Date Initials
Database
7 - 44 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 45
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 46 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 47
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 48 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 49
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 50 2014/07
Viva-ProE ™ System Operator’s Guide Maintenance
2014/07 7 - 51
Maintenance Viva-ProE ™ System Operator’s Guide
7 - 52 2014/07
Viva-ProE ™ System Operator’s Guide Troubleshooting
Chapter 8: Troubleshooting
2014/07 8-1
Troubleshooting Viva-ProE ™ System Operator’s Guide
Troubleshooting Overview
This chapter describes Viva-ProE ™ System errors and provides instructions on
possible causes and solutions for resolving them.
Note: Only trained operators can perform these procedures.
8-2 2014/07
Viva-ProE ™ System Operator’s Guide Troubleshooting
Error Signaling
A variety of error conditions may occur. Each error code is signaled differently
depending on the type of error condition. The following is a list of error and
system states:
• Audible Signal
• Dialog Signal
• Icon Signal
Repetitive Audio Alert A Critical Error has Resolve the critical event then perform
occurred. the following procedure:
1. Select Events.
2. Select Silence Alarm.
Single Audio Alert A Non-Critical Error has No action required. Open the Events
occurred. screen to view the event.
2014/07 8-3
Troubleshooting Viva-ProE ™ System Operator’s Guide
Reagent cooling temperature out of range. Check the Cooling Liquid level.
See “Filling the Cooling Fluid” on page 37 of
chapter 7.
Treated water running low. Fill the Treated Water Container as soon as
possible.
See “Filling Treated Water Container” on
page 7 of chapter 7.
8-4 2014/07
Viva-ProE ™ System Operator’s Guide Troubleshooting
flagged results are listed on the Results screen and included on printed
reports.
Table 8-2: Test Related Error Codes, Code Description, Cause, and Resolution
a Reagent Absorbance Limit • The reagent absorbance is out of • Check to ensure reagents are stored in
Violation range. the proper rotor positions.
• Poor reagent quality. • Refresh reagent.
• Reagents were placed in the
wrong rotor positions.
A Calibrator Limit Violation • A calibrator measurement is out • Check calibration parameters with data
of range. on the test package insert.
• Incorrect calibration parameters. • Ensure calibrators are stored in the
• Poor calibrator quality. proper rotor positions.
• Calibrators were placed in the • Refresh calibrators.
wrong rotor positions. • Refresh reagent.
• Poor reagent quality.
B Barcode Not Matching • Incorrect placement of the • Ensure the barcode label is clearly
barcode label. readable and facing outward.
• Sample was moved to a different • Ensure the sample tube is in the correct
rotor position. rotor position.
• Redo the entire test request procedure
for the sample.
C, c Control Limit Violation • A control measurement is out of • Check control parameters with data on
range. the test package insert.
• Incorrect control parameters. • Refresh controls.
• Poor control quality. • Refresh reagent.
• Poor reagent quality.
D Reagent Absorbance Deviation • The reagent absorbance Select Dilute to rerun the test using a diluted
Error / Substrate Depletion Error deviation is out of range. sample.
• Reaction was too rapid to be
measured (very high
concentration of analyte in
sample).
E Result Near Cut-off The test result is within the cut-off None
deviation range (only applies to
qualitative tests).
G General Hardware Error This error is often triggered in • Check the Error History screen for
combination with other errors. previously unresolved errors.
• Reset the system.
• If the problem persists, contact the
distributor or system support
department of the manufacturer.
k, K Results Exceeds Calibration The test result was below the lowest None
Limits calibration standard (k) or above the
highest calibration standard (K).
L Cuvette Blank Error / Lamp Error • Stained cuvette. • Perform a cuvette blank; exclude
• Poor lamp quality. cuvettes if required.
• Adjust or replace the lamp.
2014/07 8-5
Troubleshooting Viva-ProE ™ System Operator’s Guide
m, M Low / High Absorbance Limit The test result was below the lower None
Violation absorbance limit (m) or above the
upper absorbance limit (M).
o, O Cuvette Blank Error / Over Range • Stained cuvette. • Perform a cuvette blank; exclude
Reference Detector / Over Range • Poor lamp quality. cuvettes if required.
• Electronic circuit board failure. • Adjust or replace the lamp.
• Contact supplier’s service department.
R- Insufficient Reagent • Reagent bottle is empty or • Fill or replace the reagent bottle.
missing. • If bubbles have formed in the reagent,
• The liquid detection system is carefully remove bubbles using a
malfunctioning. disposable pipette.
• Ensure the correct reagent rotor is
installed.
R+ Reagent Not Taken • The fill level of the reagent is too • Remove some reagent; ensure the fill
high. level is below the neck of the bottle.
• Bubbles have formed on the • If bubbles have formed in the reagent,
reagent. carefully remove bubbles using a
disposable pipette.
T Cuvette Temperature Error • The ambient temperatures are • Ensure the ambient temperature is
out of range. within the specified working range.
• The cuvette rotor cover is • Ensure the cuvette rotor is properly
installed incorrectly. installed.
• The cuvette temperature control • Contact customer service.
system is malfunctioning.
u, U Cuvette Blank Error / Under • Stained cuvette. • Perform a cuvette blank; exclude
Range Reference Detector / • Poor lamp quality. cuvettes if required.
Under Range • Electronic circuit board failure. • Adjust or replace the lamp.
• Contact supplier’s service department.
W, w Westgard Rules Limit Violation None • Compare all quality control results for
that test.
• Prepare fresh controls.
• Rerun calibrations.
• Prepare fresh reagent.
y, Y Reference Panic Limit Error The result is below the lower panic None
limit (y) or above the upper panic limit
(Y).
8-6 2014/07
Viva-ProE ™ System Operator’s Guide Troubleshooting
Z Duplicate Difference Limit Error The difference between the results or None
repeated tests are above the limit.
# Insufficient Sample • The sample tube is empty or • Ensure the sample tube is present
missing. and has sufficient sample volume.
• The liquid detection system is • Ensure the sample has not clotted
malfunctioning. and remove any bubbles that are
present.
• Ensure no air bubbles are blocking
the sample aspiration.
!c Analysis Failure Concentration could not be calculated • Follow the !i resolution process.
due to one of the following: • Re-measure the sample to
confirm.
• Test was interrupted (combined • Re-measure the failed
with !i). photometric test.
• Interpolation on the low end of the
calibration curve was not possible.
• Measurement result of a
photometric test used in a
calculated test was not available or
did not meet the condition of the
calculated test.
!i Test Interrupted By System • The reagent probe could not be • Ensure acid solution (HCI) is
cleaned. loaded on the rotor and check the
• A hardware error has occurred. fill level.
• Check the error code and resolve
the problem.
!u Test Interrupted By User An emergency stop performed. • Resolve the reason for the
emergency stop.
• Reset the system.
2014/07 8-7
Troubleshooting Viva-ProE ™ System Operator’s Guide
Hardware Errors
Table 8-3: Hardware Error Codes, Code Description, Cause, and Resolution
E02 System Emergency Halt The arm touched during Reset the system.
measurement.
E05 No Clean Cuvette The system was reset but the cuvettes Perform the following procedure:
were not emptied.
1. Remove the Cuvette Rotor.
2. Clean and Empty the Cuvettes.
3. Reset the system.
E07 System Reset Incomplete System errors preventing the system Open the Events screen and resolve any
from resetting. system errors.
E11 Waste Full Waste container full. Empty the waste container.
E12 Water Running Out Treated water container empty. Fill the treated water container.
E13 Lamp Failure Lamp is near the end of its lifetime. Replace the lamp.
E16 Concentrated Waste Full Concentrated waste container full. Empty the concentrated waste container.
E17 Insufficient Treated Water Treated water container empty. Fill the treated water container.
8-8 2014/07
Viva-ProE ™ System Operator’s Guide Troubleshooting
Procedures
Unblocking the Probe
Figure 8-1: Unblocking the Probe
1 Inner Tube
3 Probe
CAUTION: Handle with care. The Probe Head and Probe are both fragile and
vital components of the system.
Note: Only use the piercing rod this is included in the analyzer
accessories kit to unblock the probe. Any other will cause
irreparable damage to the probe. Clean and store the
piercing rod after use.
1. Remove the tubing by pulling the Outer and Inner Tube [1] off the Probe
Head.
Note: Use care when pulling off the tubes. If the Inner Tube slides
too far into the Outer Tube, it may be difficult to properly
reconnect the tubing.
2. Carefully insert the Piercing Rod into the Inner Probe Tube [2]. Push
down the Piercing Rod until it comes out of the lower tip of the Probe [3].
3. Move the Piercing Rod up and down several times.
4. Using a lint free cloth remove all dirt from the lower tip of the Probe.
5. Remove the Piercing Rod.
6. Reattach the tubing by pushing the Inner and Outer Tubes over the Probe
Tip on the Sample and or Reagent Head. Ensure both the Inner and Outer
tubing slightly touch the Silicon Sealing Ring.
Note: Do not push the tubing into the Sample or Reagent Head
cover.
2014/07 8-9
Troubleshooting Viva-ProE ™ System Operator’s Guide
7. Select the Reset icon to reset the system and return to a stand-by
state.
Note: Reset can take 1 to 2 minutes to complete.
8 - 10 2014/07
Viva-ProE ™ System Operator’s Guide Data Management
2014/07 9-1
Data Management Viva-ProE ™ System Operator’s Guide
Item Description
9-2 2014/07
Viva-ProE ™ System Operator’s Guide Data Management
2014/07 9-3
Data Management Viva-ProE ™ System Operator’s Guide
6. In the Archive record older than text box, enter the number of days
after which the records should be archived.
7. In the Archive frequency text box, enter the number of days for the
frequency of the automatic archive.
9-4 2014/07
Viva-ProE ™ System Operator’s Guide Data Management
Exporting Data
Note: Only users with Administrator access can perform these
procedures.
Importing Data
Note: Only users with Administrator access can perform these
procedures.
2014/07 9-5
Data Management Viva-ProE ™ System Operator’s Guide
9-6 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
2014/07 10 - 1
System Configuration Viva-ProE ™ System Operator’s Guide
Configuration Screen
Figure 10-1: Configuration Screen
1 Configure
2 System
3 Tests
10 - 2 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Dependencies on Parameters
Parameters cannot be changed when items that depend on these parameters
are in use. For example the Edit button is disabled for tests when their
reagents are loaded on the reagent rotor. The parameters can be viewed but
not changed until all their reagents are unloaded.
Prior to making any changes to the configuration of tests or instrument
configuration items, it is recommended to check the following:
• All pending test requests were completed.
• All patient samples were unloaded.
• All issues with controls, calibrators, and blanks were resolved.
• All reagents were unloaded.
Note: Instead of unloading all reagents to unlock configuration
options, the operator can temporarily switch to an empty
reagent rotor (e.g. the Maintenance rotor) to unlock
configuration items.
2014/07 10 - 3
System Configuration Viva-ProE ™ System Operator’s Guide
Button Description
10 - 4 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Test Code A code used to represent the test in the system software.
Note: Codes starting with the digit 9 are used for user-definable
channels and cannot be edited.
Reagent Onboard The time period during when reagent can be used after it has
Stability been placed on the rotor.
Note: The On-board Stability is usually defined in the reagent
package insert.
Reagent Low Vol The percentage at which a reagent low level warning is
Warning displayed.
Service Test The test that, when selected, is not included in communications
with the LIS (the test cannot be requested from and no results
passed to the LIS when requested manually).
Note: Used only for service or maintenance of the system.
Suppress Test Results The test results that, when selected, are excluded from reports
and are not sent to the LIS.
Sample Types A list of available sample types that can be used for testing.
Note: The Disabled option is a user defined option that is
configured in the Instrument Configuration screen.
2014/07 10 - 5
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Replicates The number of times the test will be performed for each sample.
The results of these tests are averaged.
10 - 6 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Delay The delay between pipetting the sample (or the last reagent in
case of multi-reagent tests) and the first measurement.
2014/07 10 - 7
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Incubation Time The time between pipetting the sample (or the last reagent in
case of multi-reagent tests) and the measurement.
10 - 8 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Point 1, Point 2 The two required measurements that must be selected, the
points are measured from the moment the sample is pipetted.
2014/07 10 - 9
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Absorbance Limit When the result is flagged, the measured absorbance has fallen
outside this range.
Reagent Abs Limit When the result is flagged, the reagent measurement has fallen
outside this range. This measurement is performed shortly
before the sample is added.
Slope Blank If selected, the Delay for the slope blank must also be set (Kinetic
Tests or Two Point Tests using reagent R1 and R3).
Alinearity Limit Any result that exceeds the alinearity limit will be flagged
(Kinetic Tests).
10 - 10 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Absorbance Limit When the result is flagged, the measured absorbance has fallen
outside this range.
Reagent Abs Limit When the result is flagged, the reagent measurement has fallen
outside this range. This measurement is performed shortly
before the sample is added.
Reagent Abs Deviation Used to determine if the result is compromised due to depletion
or substrate in the reagent (Kinetic or Two Point Tests with one
reagent).
Substrate Depletion The factor used to determine if the result is compromised due to
substrate depletion.
Note: Valid range is -3.0 to 3.0 Abs. Displayed for Kinetic and
Two Point Tests with more than one reagent where the first
reagent volume is 220 µL or more.
2014/07 10 - 11
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Result Type The result type must be selected from either Quantitative or
Qualitative; depending on the selected type, additional
parameters are displayed.
• Qualitative Mode: Select either Increasing or Decreasing.
If increasing tests is selected, a test result above the cut-off
absorbance will be marked as positive. If decreasing tests is
selected, a test result below the cut-off absorbance will be
marked as positive.
• Qualitative Uncertainty Deviation: Sets the range in
which the test result is flagged to be viewed by the
operator. This value is used both below and above the
cut-off absorbance.
Note: View only, cannot be edited.
Primary Units The primary units in which test results are expressed.
Note: If None is selected, the absorbance is reported and the
secondary units are disabled.
Note: View only, cannot be edited.
Unit Conversion Factor A The factor used when converting values from primary to
secondary units. The primary units are converted to their base
unit, then the secondary value is calculated as A (primary base)
plus B. If both factors are set to 0, conversion to secondary units
occurs.
Note: View only, cannot be edited.
Unit Conversion Factor B The factor used when converting values from primary to
secondary units. When a test is configured in a user-definable
channel, the primary units equal the secondary units. For tests in
a predefined channel, a primary unit is predefined. To report test
results in another unit, a secondary unit can be defined.
Note: View only, cannot be edited.
10 - 12 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Item Description
Reference Ranges The reference ranges that can be enabled and set separately for
each available sample type.
Note: View only, cannot be edited.
Available Sample Types A list of available sample types. Select the double arrow to open
or close the sample type setting section.
Available sample types are Serum, Urine, Whole Blood, CSF, or
User Defined.
Note: Only the sample types enabled on the Tests-General
section are displayed.
Item Description
Concentration Limits The concentration limits that can be enabled and set separately
for each available sample type.
Note: View only, cannot be edited.
Available Sample Types A list of available sample types. Select the double arrow to open
or close the sample type setting section.
Note: Available sample types are Serum, Urine, Whole Blood,
CSF or User Defined.
Note: Only the sample types enabled on the Tests-General
section are displayed.
2014/07 10 - 13
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Calibration Type The calibration type that must be selected from either Calibrator
or Factor; depending on the selected type, additional parameters
are displayed.
Note: If the Calibration Type is set to Calibrator, tests must be
calibrated prior to running patient tests.
Correlation The Factor and Offset fields that are used to match results of
multiple systems in the same laboratory to each other.
10 - 14 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Algorithm The list that displays the available calibration methods for the
number of calibration points selected.
Note: Depending on the algorithm selected, additional
parameters may appear or disappear.
Extended Absorbance The values that determine how far the area of the measured
Limits points can be extrapolated below the lowest and above the
highest measured calibration point.
Extended Concentration The values that determine how far the area of the measured
Limits concentrations can be extrapolated below the lowest and above
the highest measured calibration concentration.
Algorithm Inaccuracy The algorithm that determines how much the modified cubic
spline is allowed to deviate from the perfect fitting curve through
all calibration points.
Item Description
2014/07 10 - 15
System Configuration Viva-ProE ™ System Operator’s Guide
Viewing Panels
Note: Only users with Administrator access can view or delete
Panels settings.
Viewing Calibrators
Note: Only users with Administrator access can view or delete
Calibrators settings.
Item Description
10 - 16 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Name The name of the control (enter for new controls, edit for existing
controls).
Creating a Control
Note: Only users with Administrator access or above can perform
these procedures.
2014/07 10 - 17
System Configuration Viva-ProE ™ System Operator’s Guide
10 - 18 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Deleting a Control
Note: Only users with Administrator access or above can perform
these procedures.
Note: Verify that all tests associated with the control have been
removed before performing this procedure.
Item Description
Test The name of the test for which the calibrator level is shown.
Auto-Accept Range The limits for the calibration level that a calibration must fall
within to be automatically accepted.
Note: These values are only displayed when the check box next
to Auto Acceptance Limits is selected.
2014/07 10 - 19
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Note: The common function buttons Edit Lot, New Lot, Cancel,
and Accept are available on this screen.
10 - 20 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Test The name of the test for which the control targets are shown.
Separation The minimum difference between the cut-off value and the
control result.
Low, High When the control is flagged, the control measure has fallen
outside these limits.
Note: These parameter are available only for the Basic QC Rules.
Note: The common function buttons Edit Lot, New Lot, Cancel,
and Accept are available on this screen.
2014/07 10 - 21
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Test The test for which the probe incompatibilities or links are
defined.
10 - 22 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
2014/07 10 - 23
System Configuration Viva-ProE ™ System Operator’s Guide
10 - 24 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Test The test for which the test incompatibilities or links are defined.
2014/07 10 - 25
System Configuration Viva-ProE ™ System Operator’s Guide
10 - 26 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Test, Calculated Tests, The category for which the order of buttons is defined.
Panels
Calculated Tests
Panels
2014/07 10 - 27
System Configuration Viva-ProE ™ System Operator’s Guide
Button
Instrument
Flag Customization
Users
Button
System Parameters
Daily Operations
Result Reporting
Sample Types
Categories
Service Configuration
Reminders
Configuring Language
Select Language to choose the local interface language.
Note: Only users with Administrator access or above can perform
these procedures.
10 - 28 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Pause/Complete Audio File The sound that plays when processing is Paused or Completed.
Event Audio File The sound that plays when an event has occurred.
Alert Audio File The sound that plays when an Alert appears on the screen.
2014/07 10 - 29
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Automatic Login The selected user will be automatically logged in when the
system starts.
Inactivity Time Out The current user is automatically logged out after a defined
period of inactivity.
10 - 30 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Pre Batch Clean When enabled, the additional tests with acid solution that are
scheduled before a batch of tests.
Note: This field is shown but cannot be edited.
Item Description
Start of Day Time Un-checks the tasks on the Start of Day screen.
End of Day Time Un-checks the tasks on the End of Day screen.
2014/07 10 - 31
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Automatic Cuvette Blank Performs an automatic cuvette blank at the start of day time
when selected.
Note: If the last cuvette blank is older than 28 hours, a new cuvette
blank is required before any tests can be processed.
Note: Cuvette Blank will not perform automatically if the Treated
water container is empty or the Waste container is full.
10 - 32 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Enable Daily Automatic Performs an automatic daily backup of the database if selected.
Backup
Item Description
Print Enabled Enables print when Yes is selected and disables print when No is
selected.
Patient Report Template Enables printing of test reports for individual patient samples.
Choose between the Internal (detailed report) or External
patient reports template.
Summary Report Template Enables printing of summary test result reports. The report
includes all loaded samples for finished tests and validated test
results. Choose between the Extensive or Detailed report.
2014/07 10 - 33
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
Export Patient Names Select the checkbox to Include or deselect to Exclude patient
names from the exported data.
10 - 34 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Collated Results • Yes - Send test results for a sample when all tests are
finished and no flags remain.
• No - Send individual test results for a sample as soon as they
become available.
Extended Results Absorbance values for kinetic measurement points that can be
selected to include in the results communicated to the LIS host.
Port Number The port number on the LIS host computer over which the LIS
(TCP/IP option) communication is handled.
2014/07 10 - 35
System Configuration Viva-ProE ™ System Operator’s Guide
Item Description
10 - 36 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
2014/07 10 - 37
System Configuration Viva-ProE ™ System Operator’s Guide
10 - 38 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Item Description
Flag Lists the flags that may appear for test results.
Print on Report If selected, enables printing the flag on result reports when the
(if LIS disabled) LIS is disabled.
Print on Report If selected, enables printing the flag on result reports when the
(if LIS enabled) LIS is enabled.
Automatic Re-Run If selected, enables an automatic rerun of the test to occur when
the result is flagged.
Custom Evaluation Option Selects one of the automatic evaluation options from the list.
Note: When the Supervisor option is selected, a supervisor must
evaluate the flagged result to complete the test before the
sample can be unloaded.
2014/07 10 - 39
System Configuration Viva-ProE ™ System Operator’s Guide
Configuring Users
Note: Only users with Administrator access or above can perform
these procedures.
Item Description
Username The name to be used in the login dialog box. This name is also
registered for actions recorded in the events log.
Role The roles that determine access privileges for screens and
functions.
Adding a User
Note: Only users with Administrator access or above can perform
these procedures.
10 - 40 2014/07
Viva-ProE ™ System Operator’s Guide System Configuration
Deactivating a User
Note: Only users with Administrator access or above can perform
these procedures.
2014/07 10 - 41
System Configuration Viva-ProE ™ System Operator’s Guide
10 - 42 2014/07
Viva-ProE ™ System Operator’s Guide Safety
Appendix A: Safety
2014/07 11 - 1
Safety Viva-ProE ™ System Operator’s Guide
General Safety
Instrument operators must be proficient in operation and maintenance
procedures. To ensure safety, follow these basic precautions:
• Observe all warnings and cautions in the guide.
• Remove safety guards only if specifically instructed in the procedures.
Replace all guards after completing the procedures.
• For more information on the Emergency Power Shutdown procedure, see
“Emergency Shutdown” on page 15 of Chapter 2: System Operation.
• Stow cables and tubing properly to eliminate tripping hazards.
• Use only specified cleaners and follow specified cleaning procedures.
Incorrect test results or damage to system components could result if these
procedures are not followed.
11 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Safety
This summary is based on the guidelines developed by the Centers for Disease
Control, the Clinical and Laboratory Standards Institute Document M29-A3,
Protection of Laboratory Workers from Occupationally Acquired Infections, and
the Occupational Safety and Health Administration’s Bloodborne Pathogens
Standard. 1-3
2014/07 11 - 3
Safety Viva-ProE ™ System Operator’s Guide
• Do not use the biohazard sink for personal cleaning such as rinsing coffee
cups or washing hands.
To prevent needlestick injures, needles should not be recapped, purposely
bent, cut, broken, removed from disposable syringes, or otherwise
manipulated by hand.
References
1. Centers for Disease Control. 1988. Update: Universal precautions for
prevention of transmission of human immunodeficiency virus, hepatitis B
virus and other bloodborne pathogens in healthcare settings. MMWR,
37:377-382, 387, 388.
2. Clinical and Laboratory Standards Institute (formerly NCCLS). Protection of
Laboratory Workers from Occupationally Acquired Infections; Approved
Guidelines - Third Edition. CLSI Document M29-A3. [ISBN 1-56238-567-4].
Clinical and Laboratory Standards Institute, 940 West Valley Road, Suite
1400, Wayne, Pennsylvania 19087-1898 USA 2005.
3. Federal Occupational Safety and Health Administration. Bloodborne
Pathogens Standard. 29 CFR 1819. 1030.
11 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Safety
Mechanical Hazards
WARNING: Hot Surface - Do not touch the lamp when it is switch on.
If the lamp must be replaced, wait until it has cooled off
before replacing.
Biological Hazards
Chemical Hazards
2014/07 11 - 5
Safety Viva-ProE ™ System Operator’s Guide
11 - 6 2014/07
Viva-ProE ™ System Operator’s Guide Warranty and Support
2014/07 12 - 1
Warranty and Support Viva-ProE ™ System Operator’s Guide
Warranty Period
The limited warranty period generally commences upon installation of the
original instrument at the customer’s location and extends for a period of one
year thereafter, unless otherwise specifically agreed to by and between
Siemens (or its authorized distributors) and customer in a writing signed by
duly authorized representatives of both parties (sales representatives are
generally not authorized representatives of Siemens for these purposes).
Service Calls
The customer may obtain service for instruments during normal business
hours by contacting the nearest Siemens technical support provider or
authorized distributor.
Warranty or service calls generally include onsite repair or exchange of
instruments or components, travel to the location of the instrument, and
onsite labor during normal business hours. A warranty or service call is
initiated by the customer by following the instructions on how to obtain
service for the customer’s instrument. The service call is considered complete
when any defects in material or workmanship have been corrected by repair or
replacement and the instrument conforms to the applicable specifications.
When service is complete, the customer receives a copy of the documentation
detailing all work performed by the Siemens representative or authorized
distributor.
12 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Warranty and Support
2014/07 12 - 3
Warranty and Support Viva-ProE ™ System Operator’s Guide
12 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Supplies
Appendix C: Supplies
2014/07 13 - 1
Supplies Viva-ProE ™ System Operator’s Guide
Cleaning Solutions
All cleaning solutions are European Commission (EC) marked according to
directive 98/79/EC. They are for professional use only and intended for in vitro
diagnostic devices.
Note: If the solution is cloudy, discard according to local regulations
regarding waste disposal.
System Solution
• Reference (Part Number) - 10445247
• Distributed - Siemens
• Intended Use - Washing cuvettes and probes; also used for the pipetting
system (wetting effect for system fluidics)
• Preparation - Dilute with distilled water at 1/400 (12.5 mL system solution
in 5 L of distilled water)
• Composition - Aqueous solution with a detergent containing sodium azide
(<0.1%)
• Precautions - None
13 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Supplies
2014/07 13 - 3
Supplies Viva-ProE ™ System Operator’s Guide
13 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Specifications
Appendix D: Specifications
2014/07 14 - 1
Specifications Viva-ProE ™ System Operator’s Guide
System Specifications
This section summarizes the system design specifications.
1 Distance required to access and remove the rear cover 20 cm (7.9 inches)
14 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Specifications
Environmental
Table 14-2: System Environmental Specifications
Electrical Requirements
Table 14-3: System Electrical Requirements
Line Frequency 50 or 60 Hz
Approvals
Table 14-4: Approvals
Approvals Description
CE European Conformity
CB Cartes Bancaires
UL Underwriters Laboratory
ICE 61326-2-6 The system complies with the emission and immunity
requirements described in the IEC 61326-2-6.
CISPR 11 Class A This equipment has been designed and tested to CISPR
11 Class A. In a domestic environment, it may cause
radio interference in this instance, it may be necessary
to take measures to mitigate the interference.
Note: The approvals listed here refer only to the instrument and
operator console, not to additional devices. For the approvals
for these devices, see the corresponding manuals.
2014/07 14 - 3
Specifications Viva-ProE ™ System Operator’s Guide
Technical Specifications
Table 14-5: Product Specification
Syringe 100 µL
Note: All positions can be used for patient samples (normal, ASAP,
and STAT), calibrators, Blank, and Controls. Normally,
controls and calibrators are placed on the inner ring and
patient samples are loaded on the two outer rings (readable
by the internal barcode scanner).
14 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Specifications
Syringe 1000 µL
Wavelengths 340, 415, 505, 546, 570, 600, 660, and 700 nm
Wavelength ± 2nm
Uncertainty
Spectral Half-width 10 ± 2 nm
Value
Temperature 37 ° C ± 0.2 °C
2014/07 14 - 5
Specifications Viva-ProE ™ System Operator’s Guide
User Interface
Table 14-9: Computer System (minimal requirements)
Technology CCD
Barcodes Code 39, Code 93, Code 128, Codabar, and Code 2
of 5 family
Technology CCD
14 - 6 2014/07
Viva-ProE ™ System Operator’s Guide Specifications
Disclaimer
The manufacturer assumes no responsibility for erroneous test results caused
by reagents kits and or test parameters that are not explicitly provided or
recommended by the manufacturer.
2014/07 14 - 7
Specifications Viva-ProE ™ System Operator’s Guide
Predefined Systems
The Viva-ProE ™ System is supplied as a predefined system (designed for
Siemens reagents). In predefined systems, the test parameters are
pre-programmed and 10 open channels can be fully configured.
Note: Tests in predefined systems cannot be edited; however, they
can be deleted.
14 - 8 2014/07
Viva-ProE ™ System Operator’s Guide Symbols
Appendix E: Symbols
2014/07 15 - 1
Symbols Viva-ProE ™ System Operator’s Guide
System Symbols
This section describes the symbols that can appear on the interior or exterior of
the system, or on the system packaging. The symbols on the system identify
the location of certain components and, where necessary, display warnings for
proper operation. The symbols on the system packaging also provide other
important information.
Table 15-1: System Symbols
Symbol Description
15 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Symbols
Symbol Description
2014/07 15 - 3
Symbols Viva-ProE ™ System Operator’s Guide
15 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Theory
Appendix F: Theory
2014/07 16 - 1
Theory Viva-ProE ™ System Operator’s Guide
Theory
This chapter contains the theoretical foundations for the measurements
performed by the instrument. It includes formulas for all measurement types,
as well as calibration algorithms and explanations of some advanced integrity
checking features.
16 - 2 2014/07
Viva-ProE ™ System Operator’s Guide Theory
Absorbance Measurements
Measuring Absorbance
Parameters are measured as spectral absorbance relative to a zero-adjustment
on water. The measured absorbance is related to the concentration following
Beers-Lambert law:
Item Description
Abs Absorbance
c Concentration
T Transmission
2014/07 16 - 3
Theory Viva-ProE ™ System Operator’s Guide
Measurement Types
When samples and reagents are mixed, a reaction takes place. In general, this
reaction follows a profile as shown in the diagram below. In the first section of
the curve, there is a linear relationship between time and change in
concentration (absorbance). In the second linear relationship between time
and change in concentration (absorbance). In the second section, the reaction
slows down until it reaches a stable endpoint, when all of the sample or
reagent is consumed.
Figure 16-2: Absorbance Versus Time Graph
The time for the reaction to reach a stable endpoint varies greatly. As the time
frame available for measurements on the system is limited, three available
types of measurements are the following:
Table 16-1: Measurement Types
Type Description
Kinetic Tests The time frame for measurement on the system corresponds to section 1 of
the curve. This measurement type assumes that the change in absorbance is
linear during the entire measurement period.
Endpoint Tests A stable endpoint is reached within the measurement time frame.
Two Point Tests The absorbance values at two points are compared, irrespective of which part
of the curve corresponds to the measurement time frame.
16 - 4 2014/07
Viva-ProE ™ System Operator’s Guide Theory
The factor “F” is either fixed value supplied with the reagent kit or is calculated
as follows:
2014/07 16 - 5
Theory Viva-ProE ™ System Operator’s Guide
Bichromatic Endpoint
The factor calculation and the formulas for bichromatic endpoints with
reagent blank and sample blank are identical to those used in the standard
endpoint method.
16 - 6 2014/07
Viva-ProE ™ System Operator’s Guide Theory
Calibration Algorithms
Standard Cubic Spline Method
Item Description
Cm Calculated concentration
2014/07 16 - 7
Theory Viva-ProE ™ System Operator’s Guide
The difference is the constraint of the curve that passes through all data points
is lifted. This is achieved by introducing a smoothing factor that determines
the balance between the smoothness of the curve and the proximity to the
data points. In extreme situations, this method reduces to either the linear
regression line (smoothest line) or the normal cubic spline (closest to the data
points). In the current implementation, the smoothing factor is optimized
based on a configurable maximum deviation of the curve from the data points.
The modified cubic spline is a variant of the smooth cubic spline.
See Smoothing by spline functions by C.H. Reinsch, published in Numerische
Mathematik 10, 177-183, 1967.
NLLS Algorithms
4 Parameter Logit-log
Item Description
Cm Calculated concentration
5 Parameter Logit-log
Item Description
Cm Calculated concentration
16 - 8 2014/07
Viva-ProE ™ System Operator’s Guide Theory
5 Parameter Exponential
Item Description
Cm Calculated concentration
Item Description
Cm Calculated concentration
2014/07 16 - 9
Theory Viva-ProE ™ System Operator’s Guide
Westgard Rules
The Westgard rules can only be used for single controls. When the Westgard
option is selected, the high and low limits are automatically set from the
standard deviation that is entered by the operator.
The Westgard rules are violated if one or more of the following conditions
apply:
• 1 control result is more than 3 standard deviations from the target.
• The last 2 control results are more than 2 standard deviations from the
target in the same direction (+ or -).
• The last 4 control results are more than 1 standard deviation from the
target in the same direction (+ or -).
• The last 10 control results are all located either on the “+” or the “-” side to
the target.
Note: The Westgard rules are not violated in any other case.
16 - 10 2014/07
Viva-ProE ™ System Operator’s Guide Index
A
accept control measurements 4-3
accessing configuration screen 10-3
accessing configure tests screen 10-4
accessing flag customization screen 10-39
accessing system menu screens 10-28
accuracy and precision 14-7
acid solution 13-3
activate internal barcode reader 10-30
add calibrator lot 10-19
add control lot 10-21
add probe incompatibilities 10-22
add probe incompatibilities or links 10-23
add test incompatibilities 10-24
add test incompatibilities or links 10-25
add test to a control 10-17
add user 10-40
adding controls 10-16
adjusting lamp unit 7-29
administrator 1-3
approvals 14-3
archive patient results 9-3
archiving results database 7-11
as needed 7-34
audible error signals 8-3
B
backup system database 9-2
barcode symbology 2-6
C
calculated tests 10-16
calibration 5-1
cautions 11-2
check 0.1N HCl cleaning solution 7-11
check 10% cleaning solution 2-13, 7-10
check cuvette blank measurements 7-9
checking 10% cleaning solution 2-13, 7-10
checking cuvette blank measurements 2-11
checking reagent inventory 7-11
checking reagents 3-2
checking syringes 2-11, 7-9
clean system 7-21
clean touchscreen 7-34
cleaning concentrated waste container 7-15
cleaning solutions 13-2
cleaning system 7-34
cleaning the touchscreen 7-34
cleaning touchscreen 7-34
computer 1-16
concentrated waste container 1-11
configuration 10-16
configuration screen 10-2
2014/07 Index - 1
Index Viva-ProE ™ System Operator’s Guide
D
daily setup 2-7
data management 9-1
deactivate internal barcode reader 10-30
deactivate user 10-41
delete control 10-19
delete probe incompatibilities or links 10-24
delete test from a control 10-18
delete test incompatibilities or links 10-26
dimensions and weight 14-2
E
edit calculated test parameters 10-37
edit calibrations and reagent blanks 5-2
edit calibrator lot 10-20
edit category types 10-37
edit control lot 10-21, 10-22
edit flag customization 10-40
edit operator reminders 10-38
edit patient tests 4-5
edit probe incompatibilities 10-22
edit probe incompatibilities or links 10-24
edit test incompatibilities 10-24
edit test incompatibilities or links 10-25
edit test order 10-27
editing controls 10-16
editing test parameters 10-18
electrical requirements 14-3
Index - 2 2014/07
Viva-ProE ™ System Operator’s Guide Index
F
fill treated water container 2-9, 7-7
filling cooling fluid 7-37
G
general care and cleaning 7-3
general error condition icons 8-4
H
hardware errors 8-8
hardware overview 1-4
home screen 1-20
I
import data 9-5
install reagent rotor 3-7
Instrument Removal 11-4
intended use 1-2
internal barcode reader 1-14, 14-6
K
keyboard 1-17
L
lamp unit 1-12
load and process sample tubes with internal barcode reader 4-6
load and process sample tubes without internal barcode reader 4-7
load calibration screen 1-25
load calibration screen detailed view 1-27
load calibrators 5-2
load controls screen 1-29
load patients screen 1-32
load reagent 3-4
load reagent blanks 5-2
login 2-7
login superuser 1-18
logout superuser 1-18
2014/07 Index - 3
Index Viva-ProE ™ System Operator’s Guide
M
maintenance 7-1, 7-34
maintenance - as needed 7-34
maintenance - daily 7-5
maintenance - every 10,000 tests 7-32
maintenance - every 3 months 7-17
maintenance - every 6 months 7-19
maintenance - monthly 7-13
maintenance - weekly 7-12
maintenance - yearly 7-23
maintenance logs 7-39
maintenance schedule 7-4
manually accept cuvette blank results 2-12
manually accept results 5-3
manually archive patient results 9-4
manually backup system database 9-2
manually process samples 4-8
manually reject results 5-3
N
notes 11-2
O
operator 1-3
P
patients 4-5
pause sample processing 4-10
perform calibrations and reagent blanks 5-2
pipettor 1-9
precautions for operation 2-2
preparing samples 2-3
prime system 7-11
print daily report 2-13, 7-10
process STAT samples 4-9
Q
quality control 6-1
quality control screen 6-2
R
reagent info screen 1-24
reagent probe 1-5
reagent rotor 1-8
reagent rotor image size indicators 1-23
reagent rotor image status colors 1-22
reagent screen 1-22
reagents 3-1
refill 10% cleaning solution 2-13, 7-10
refilling 10% cleaning solution 2-13, 7-10
reject control measurements 4-3
reject cuvette blank results 2-12
remove reagent 3-5
Index - 4 2014/07
Viva-ProE ™ System Operator’s Guide Index
S
safety 11-1
sample barcode label properties and placement 2-5
sample management 4-1
sample probe 1-5
sample rotor 1-7
search patient results 4-11
service 12-1
service calls 12-2
sodium hydroxide 13-3
software overview 1-18
specifications 14-1
start of day 2-7
start of day screen 1-21
stopping the system 2-15
superuser login 1-18
supervisor 1-3
supplies 13-1
support 12-1, 12-2
switch reagent rotor 3-5
switch user 1-18
system access 1-18
system cleaning solution 13-2
system configuration 10-1
system operation 2-1
system shutdown 2-15
system software login 1-18
system solution 13-2
system specifications 14-2
system symbols 15-2
2014/07 Index - 5
Index Viva-ProE ™ System Operator’s Guide
T
technical specifications 14-4
test related errors 8-5
tests 10-16
theory 16-2
touchscreen 1-16
transfer backup database to external media device 9-3
troubleshooting 8-1
troubleshooting error codes 8-3
turn on system 2-7
turn on the system 2-15
U
unblocking the probe 8-9
unload all calibrators 5-4
unload all controls 4-4
unload all reagent blanks 5-4
unload all samples 4-12
unload calibrators 5-3
unload controls 4-4
unload reagent 3-5
unload reagent blanks 5-3
unload sample 4-11
unload screen 1-41
unloading reagent 3-5
unloading reagents 3-5
user accounts 1-18
user interface 14-6
user interface screens 1-20
using gel tubes 2-3
using pediatric sample cups 2-4
V
view calibrator lot 10-19
view control lot 10-21
view daily report 2-13, 7-10
view probe incompatibilities 10-22
view quality control 6-4
view test incompatibilities 10-24
view test order 10-27
viewing calibrators 10-16
viewing controls 10-16
viewing tests 10-5
viewing tests - absorbance checks 10-10
viewing tests - calibration 10-14
viewing tests - measurement endpoint tests 10-8
viewing tests - measurement kinetic tests 10-7
viewing tests - measurement two point tests 10-9
viewing tests - processing 10-6
viewing tests - result calculation 10-12
viewing tests - sample blank absorbance checks 10-11
Index - 6 2014/07
Viva-ProE ™ System Operator’s Guide Index
W
warnings 11-2
warranty 12-1, 12-2
waste system 1-10
water system 1-10
windows login 1-18
2014/07 Index - 7
Index Viva-ProE ™ System Operator’s Guide
Index - 8 2014/07