Professional Documents
Culture Documents
PG-MEA1060-INV
Instructions for Changing the Plexiglass Part of the Perfusion Ground Plate
of the MEA1060-INV Amplifier
O-Ring-PGP PGP-O-Ring
Disconnect the eight screws fixing the plexiglass with an Allen key 2.0. Disconnect the six screws
fixing the steel perfusion IN and perfusion OUT. Detach the broken plexiglass and replace the
new plexiglass. Press the steel tubes carefully into the provided slot. Tighten the screws very
softly first and screw them not too hard.
Scope of delivery:
1 Plexiglass Part PG-MEA1060-INV
14 Screws (M3x5)
Multi Channel Systems Fon +49-7121-9 09 25- 0 © 2012 Multi Channel Systems MCS GmbH
MCS GmbH Fax +49-7121-9 09 25-11
Aspenhaustrasse 21
72770 Reutlingen info@multichannelsystems.com Product information is subject to change
Germany www.multichannelsystems.com without notice.