BA-88A Semi-Auto Chemistry Analyzer
BA-88A Semi-Auto Chemistry Analyzer
Features
Working principles
Competitors
Operation
Price
FAQ
Features
Pump
1. long life (almost no need to change);
2. extra pump tube
3. pump calibration every ~2 months
Flowcell
1. easy to be observed when results
susceptible; easy to locate the
problem
2. easy to be installed/replaced
Easy Maintenance
Software upgrade
- via USB; easy and quick
3. …
Main Contents
Features
Working principles
Competitors
Operation
Price
FAQ
Working principles
Thermal printer
Touch screen
Aspirate button
Structure layout
Main Contents
Features
Working principles
Competitors
Operation
Price
FAQ
Improvements from BA-88
BA-88A BA-88
a. Testing mode(s)
flowcell or cuvette flowcell
b. Connection ports
c. Display
h. Patient report
Features
Working principles
Competitors
Operation
Price
FAQ
BA-88A vs RT-1904c
BA-88A
BA- RT-1904c
RT-
a. Large TFT touch screen, 7.0 a. Smaller character displaying,
inch, logic software; controlled by rolling ball;
b. Bi
Bi--chromatic or reagent blank b. No bi-
bi-chromatic or reagent blank
measurement for all kinds of measurement for Kinetics and
tests; Fixed--time tests;
Fixed
c. Flexible calibration management; c. Only 3 calibration methods,
d. Storage for 3000 sample results. impossible to change calibration
parameters;
d. Storage for 600 sample results.
BA-88A vs STAT Fax 3300
BA-88A
BA- STAT Fax 3300
a. Large TFT screen, 7.0 inch; a. Flip
Flip--up unicolour LCD display;
b. Easy to operate (by touch b. Hard to operate (by membrane
screen); buttons);
c. Room temperature, 25℃ 25℃, 37℃ for testing;
c. Only 37℃
30℃
30 ℃ or 37℃
37℃ for testing; d. Bi
Bi--chromatic measurement for
d. Bi
Bi--chromatic measurement Endpoint tests;
for all kinds of tests; e. Vacuum system, unable to pipette if
e. External peristaltic pump, leaving unused for a period (have to
easy to maintain; open the machine to loosen the tube);
f. Good heat abstractor, safe f. Temperature around cuvette / lamp
for the operators. rises high during testing, the upper
board heats.
BA-88A vs CLIMA MC-15
BA-88A
BA- MC-15
MC-
a. Large TFT screen, 7.0 inch, a. Two
Two--line unicolour display;
logic software; b. Operated by membrane
b. Easy to operate (by touch buttons;
screen); c. Measuring mode: cuvettes
c. Two measuring modes: flowcell (15 positions multicuvette
or cuvettes; rack);
d. 200 tests programmable; d. 60 tests programmable;
e. Wavelength range: 340-
340-670nm; e. Wavelength range: 320-
320-
25℃, 30℃
f. Room temperature, 25℃ 30℃ 630nm;
37℃ for testing.
or 37℃ 37℃ for testing.
f. Only 37℃
Main Contents
Features
Working principles
Competitors
Operation
Price
FAQ
BA-88A operation
1. Choose test in
software
§ Features
§ Working principles
§ Improvements from BA-
88
§ Competitors
§ Operation
§ Price
§ FAQ
Main Contents
§ Features
§ Working principles
§ Improvements from BA-
88
§ Competitors
§ Operation
§ Price
§ FAQ
Q1: When I change the light lamp of BA-88A, do I need
to change the lamp only or the whole lamp module?
A6: For end-point test (e.g. ALB, CHL…), the practical throughput would
be up to ~100 test/hour, if the operator is experienced; for fixed-time or
kinetic methods, the throughput would go up to 25~30 test/hour.
A7: The default BA-88A parameters are for Mindray chemistry reagents.
If you are using Mindray reagents, you can directly use those defauls
values; if you are using reagents other than Mindray’s, you can
change/edit the parameters on the base of the default values or just
simply add new parameters with new name (for example, if you are
setting parameter for ATS reagent of Human brand, you can add a new
parameter column with name “AST_Human”). Totally 200 tests can be
programmed.
Q8: BA-88A has a declared lamp-life of 900 hours. How is it calculated
out? Is standby mode (sleep mode) counted into this 900 hours?
A8: This 900-hour doesn't include the standby mode time. In standby
mode, BA-88A’s lamp is in a low-consume status with low intensity;
this saves life of the lamp. Thus, practically, the real life time of BA-
88A’s lamp is much longer. When BA-88A doesn’t run tests for 15
mins (default 15 mins; this time is definable), the analyzer switches
into standby mode; when needed to measure tests again, it needs
only ~3 mins to wake up (stabilize the light source to working status).
Otherwise, from shut-down status, BA-88A needs ~15 mins to
stabilize the light source.
Q9: Is it true that semi-auto chemistry analyzer has low repeatability?