Message Meaning Operation/what to do

C022 Water temperature too high Decouple patient and wait for
(treatment not possible) end cooling cycle

C032 Treatment not possible Call the service

C046 Check ECG unit Switch on ECG unit, check


connection to dormier
compact sigma

C050 Treatment possible Reached end of start-up


procedure
C051 EOC timeout Signal EOC (end of charge) is
not transmitted in time, call
dormier medtech service

C085 Key combination not allowed For example the shock wave
intensity keys+and- are
pressed simultaneously

C135 Failure in collision test of the Restart machine and repeat


localizing arm collision test when the
message “L-_?” is displayed

COLL Collision with the therapy head Correct the collision


or localization arm transducer

Ent9 Degassing procedure End mode (see page 27)


accelerated

FILL Not enough water in the water Fill water reservoir as


circuit described on page 93

L-_? Check the collision protection Active collision protection on


on and localization localization arm page 41

LEnS Change EMSE Arrange the cnage EMSE


Message appears after reset of
the shockwave counter page27
when the total numbers of
shockwaves is more than
1,000,000 shock waves
-rEL With switching on machine the Ensure that shockwave release
shock wave release button is button is not pressed or
pressed exchange hand-held control if
defective