Technical
Publications
TANZO Touch
Maintenance Manual
SMM-B-02
Technical Document
All rights reserved© 2015 By Woson Medical Instrument Co.,Ltd.
Sterilizer Maintenance Manual-TANZO Touch series
Catalogue
Part I. Basic Introduction........................................................................................................... 1
1. Electric Diagram......................................................................................................................1
2. Wiring...................................................................................................................................... 2
3. Pipeline Diagram.....................................................................................................................3
4. Temperature and Pressure Table..............................................................................................4
5. Attentions while Adjusting...................................................................................................... 4
6. Normal Cycling Status............................................................................................................ 4
7. Programs and Working Condition of Sterilizer.......................................................................5
Part II. Error Codes and Solutions............................................................................................. 7
Error Codes List.......................................................................................................................... 8
1. E01: Heat up Slow...................................................................................................................9
2. E02: Over Pressure................................................................................................................13
3. E03: Door Loose in Cycle.....................................................................................................13
4. E04: Low Pressure................................................................................................................ 14
5. E05: Release Slow.................................................................................................................15
6. E06: Vacuum Slow................................................................................................................ 16
7. E7: Atmospheric Pressure Low.............................................................................................16
8. E08: Micro-switch Broken before Close the Door............................................................... 17
9. E09: Micro-switch Broken after Close the Door.................................................................. 18
10. E10: Preheat Speed Slow.................................................................................................... 18
11. E11: Over Temperature........................................................................................................19
12. E12: Temperature Low........................................................................................................20
13. E13: Auto-filling Abnormal................................................................................................ 20
14. E14: Inner Sensors not Match............................................................................................. 21
15. E17: Water Quality Control Fail......................................................................................... 23
16. E23: Pressure not Match Temperature................................................................................24
17. E24: No Sufficient Water (not open)...................................................................................25
18. E25: Air Filter Dirty............................................................................................................ 25
19. E27: Leakage.......................................................................................................................27
20. E31: Temp. Sensor 1 Abnormal.......................................................................................... 27
21. E32: Temp. Sensor 2 Abnormal.......................................................................................... 28
22. E33:Temp. Sensor 3 Abnormal........................................................................................... 29
23. E34: Temp. Sensor 4 Abnormal.......................................................................................... 30
24. E51: Temp. Sensor 1 Output High...................................................................................... 30
25. E52: Temp. Sensor 2 Output High...................................................................................... 31
26. E53: Temp. Sensor 3 Output High...................................................................................... 31
Sterilizer Maintenance Manual-TANZO Touch series
1. Electric Diagram
回气阀
进水阀
二位三通阀
变压器AC12V输出
- +5V
蓝蓝 黑 黑 黄 白 黄 棕 棕 黄 黄 蓝 蓝 打印机
白
白 黑
微动开关传感器 风
白 黑 扇
红
红
废水箱传感器 黄
侧液位低水位传感器 蓝 黄
动加水泵
侧液位高水位传感器 蓝 黄
绿
加水泵
黄 蓝蓝 绿
白
真空泵
白
蒸汽发生器
加热棒
LCD接口
键盘
锅壁加热圈
打印机
号线 (黑)
号线 (红)
(黑)
(红)
SMM-B-02| Page 1 of 51
Sterilizer Maintenance Manual-TANZO Touch series
2. Wiring
Wiring Classification
a) The high voltage area (AC 220V): water pump wire, vacuum pump wire, power switch
wire, fuse wire, steam generator wire, heating ring wire and elementary transformer wire
(red).
DO NOT touch the high voltage area without any secure
Attention
instruments when powered on.
SMM-B-02| Page 2 of 51
Sterilizer Maintenance Manual-TANZO Touch series
3. Pipeline Diagram
AF RV
P1
EV4 EV1
CC
TS1
SM VP
EV2
EV3
WP2
R3
R2
R1
SMM-B-02| Page 3 of 51
Sterilizer Maintenance Manual-TANZO Touch series
The temperature of saturated steam (pure steam) is of certain degree under different pressure,
here is the comparison table between temperature (inside chamber) and pressure.
Pressure(Bar) Temperature(℃)
0.0 100
0.2 105
0.7 115
1.1 121
1.7 130
2.1 134
Looking into the relationship between temperature and pressure when adjusting or working
normally, you will find under normal circumstances, the lower the pressure, the larger the
deviation between temperature and pressure. For instance, when pressure is 0.0bar, the
temperature is about 97℃. But when pressure rises to 2.1bar, the temperature will be 134℃.
Small deviation (±1℃) is allowed under low-pressure.
a. Each technician who tests or inspects the machine should know the working principle
very well;
b. All technicians should know exactly the positions of high and low voltage parts to avoid
electroshock;
c. Make sure the door is closed completely before running cycles;
d. When alarm appears, please record the error code, pressure and temperature before
turning off the machine;
SMM-B-02| Page 4 of 51
Sterilizer Maintenance Manual-TANZO Touch series
The engineer must know clearly the working process and status display of each cycle.
1) Sterilization Process:
(bar)
2.0
1.8
Pressure
Start End
0
Time
-0.8
-1.0
SMM-B-02| Page 5 of 51
Sterilizer Maintenance Manual-TANZO Touch series
No. Description 3 4
1 Sterilization procedure
2 Sterilization temperature 2 5
3 Sterilization time
4 Vacuum number 6 7
5 Drying time
8
6 Sterilization curve
7 Sterilization time and status 9
10
8 The total time of the program run
9 Sterilization real-time temperature
11
10 Sterilization real-time pressure 17
11 Tank logo 19 18
12 Waste water tank logo
20
13 Water quality Monitoring logo
14 Door status logo
15 Printer logo
16 USB logo
17 Standby time holding logo
18 No. sterilization times
12 13 14 15 16
19 Sterilization that day the real time
20 Operating instructions
SMM-B-02| Page 6 of 51
Sterilizer Maintenance Manual-TANZO Touch series
SMM-B-02| Page 7 of 51
Sterilizer Maintenance Manual-TANZO Touch series
SMM-B-02| Page 8 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The pressure cannot reach the set value after 30mins of heating and pressure program.
Problem Reason:
1. Water vapor appears from the cover fitted or a sound of “hissing” during heating and
pressure program, it means door gasket leaked: because too much steam during cycles,
result in leakage due to the door gasket is ejected, out of sharp, weakened sealing effect;
2. Weaken sealing effect of the safety valve bolt due to shaking or other possible reasons;
3. Weaken heating effect of the heating ring due to the port connected the heating ring and
PCB is loose, short circuit or broken;
4. Vacuum pump fails to work (no sound) during vacuum condition, result in pressure
cannot match with temperature when heating stage, and then enter adjustment condition
more than 10mins: 1. The wire of vacuum pump is loose, or broken; 2. The capacity of
vacuum pump is broken; 3. The vacuum valve(black line) is blocked or fails to open;
5. The water supply valve is blocked or broken, fails to supply water for steam generator;
6. The water pump is broken, fails to supply enough water for steam generator;
7. The air return valve is broken or leaked result in pipeline leakage during pressure stage;
8. The normal-open valve is leaked or broken result in pipeline leakage during pressure
stage;
9. The vacuum valve is leaked or broken result in pipeline leakage during pressure stage;
10. The temperature of steam generator less than 160degree due to the heating stick of steam
generator is broken, result in the water pump fails to supply water.
11. The steam generator is blocked or weakened heating effect, and fails to supply enough
steam for chamber, due to using non-designated water or non-scheduled cleaning
program.
Solution:
1. Check if the door gasket is normal, or change it according to the steps as below.
Door Gasket
SMM-B-02| Page 9 of 51
Sterilizer Maintenance Manual-TANZO Touch series
1. Hold verge of the gasket by one hand 2. Pull out the whole seal slowly once took
softly, and insert the screwdriver into the one part of the gasket out. Then check and
gap between gasket and door, take out the clean the gasket and the groove of the
seal slowly. gasket, replace the gasket if any damage.
2. Check if the relief valve is leaked during pressure stage. If yes, tighten the switch on the
valve, or change the valve.
3. Forced quit the cycle if the sterilizer fails to enter vacuuming state after 15mins. Then
wear the gloves to touch the surface of inside chamber. If no heat or less heat, check if the
lines which connect heating ring and PCB are normal. Or change the heating ring.
SMM-B-02| Page 10 of 51
Sterilizer Maintenance Manual-TANZO Touch series
4. 1) Check if the connecting lines from vacuum pump are connected well. Or replace the
capacitor, even the vacuum pump.
2) Check if the core of 3-way valve (Black line)is blocked, and then clean it. Or replace
the valve.
3-way Valve
5. Abnormal sound of “Quack, Quack”, due to the water pump works without water during
heating and pressure stage, clean the water supply valve or replace it.
SMM-B-02| Page 11 of 51
Sterilizer Maintenance Manual-TANZO Touch series
7. Check and clean the core of air return valve. Or replace the valve.
8. Check and clean the core of normal-open valve. Or replace the valve.
9. Check if the core of 3-way valve (Black line) is blocked, and then clean it. Or replace the
valve.
10. Use multimeter to check if the heating stick of steam generator is normal. Or replace it.
SMM-B-02| Page 12 of 51
Sterilizer Maintenance Manual-TANZO Touch series
11. Water leaked from the bottom of sterilizer during heating stage, the silicone tube between
water pump and steam generator is broken due to steam generator blocked, replace the
steam generator.
Problem Description:
The pressure in the chamber is larger than the set value during heating or sterilization stage.
Problem Reason:
1. The temperature sensor(T2) covered;
2. Inconsistent voltage;
3. Fail to release due to the normal-open valve blocked;
4. The pressure sensor is broken.
Solution:
1. Check if T2 is covered by wrap;
2. Check if the local voltage is normal;
3. Open the normal-open valve, check and clean it. Or replace the valve;
4. Replace the PCB.
Problem Description:
Ensure the door and door handle are closed completely before cycling.
Problem Reasons:
1. Not close the door completely due to not tighten the door handle or large pressure in the
chamber;
2. Forced open the door during cycle;
3. The micro switch is loosed;
4. The connecting line which connects micro switch and PCB is loosed.
SMM-B-02| Page 13 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution:
1. Ensure the door and door handle are closed completely before cycling;
Door Handle
2. Do not open the door during cycle;
3. Check if the micro switch is loosed. Or tighten it;
Micro Switch
4. Check if the connecting line (White line) of micro switch is loosed;
Problem Description:
The pressure in the chamber is less than the set value during heating or sterilization stage.
Problem Reasons:
1. Too many articles to disinfect;
2. Inconsistent voltage, less than 190V;
3. Fail to release due to the normal-open valve blocked;
4. The pressure sensor is broken.
SMM-B-02| Page 14 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution:
1. Reduce the articles;
2. Check if the local voltage is normal;
3. Open the normal-open valve, check and clean it according to E02. Or replace the valve.
Problem Description:
Not reach 20Kpa in 20mins of vacuum stage (30mins for first vacuum stage), and then end all
cycles, not reach 20Kpa in 25mins.
Problem Reasons:
1. The chamber filter is blocked;
2. The air release valve is blocked, or broken;
3. The connecting ports which connect air release valve and PCB is loosed;
4. No electrical reaction on the connecting ports;
5. 3-1 of 3-way valve is blocked;
6. Condenser is blocked.
Solution:
1. Clean or replace the chamber filer;
Condenser
SMM-B-02| Page 15 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
Not reach -0.4bar in 5mins of vacuum stage, not reach -0.6bar in leakage test.
Problem Reasons:
1. The vacuum pump is broken;
2. Continual steam due to too many wet articles;
3. The normal-open valve blocked;
4. 2-1 of 3-way valve is blocked.
Solution:
1. Replace the vacuum pump;
2. Turn off and cool down the sterilizer, reduce the articles;
3. Check and clean the core of normal-open valve. Or replace the valve;
4. Check if the core of 3-way valve (Black line) is blocked, and then clean it. Or replace the
valve.
Solution:
1) Enter the adjustment interface: press “MENU” to enter “SYSTEM SET”, and then
choose “User Message”;
SMM-B-02| Page 16 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution: enter the setting interface, input “1111” to enter initial adjustment, choose “manual”
and save, and then restart the sterilizer.
SMM-B-02| Page 17 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Reason:
1. The micro-swift broken;
2. The motor, pressure switch or over current protector is broken.
Solution: enter the setting interface, input “1111” to enter initial adjustment, choose “manual”
and save, and then restart the sterilizer.
Problem Description: the temperature of chamber is less than 100 degree and the
temperature of steam generator is less than 110 degree after 20mins of preheat program.
Problem Reason:
1. The heating ring of the chamber no heat or less heat due to the connecting ports is loosed,
short circuit or broken;
2. The heating stick of steam generator is broken, result in heat is not enough.
SMM-B-02| Page 18 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution:
1. Forced quit the cycle after shows E10, and then wear the gloves to touch the surface of
inside chamber. If no heat or less heat, check if the lines which connect heating ring and
PCB are normal. Or, change the heating ring;
2. Use multimeter to check if the heating stick of steam generator is normal. Or replace it.
Heating Stick
Problem Description:
The temperature is more 2 degree than the theoretical value, or more than the max set value
during heating or sterilization process.
Problem Reason:
1. Inconsistent voltage, more than 220V;
2. Too many articles to disinfect;
3. The temperature sensor in the chamber (T2) is broken.
Solution:
1. Check if the local voltage is normal;
2. Reduce the articles;
3. Use multimeter to check if the sensor (T2) is normal according to the PT1000, if
abnormal, change the sensor.
SMM-B-02| Page 19 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The temperature is less 2 degree than the theoretical value, or more than the min set value
during heating or sterilization process.
Problem Reason:
1. Inconsistent voltage;
2. Too many articles to disinfect;
3. The temperature sensor in the chamber (T2) is broken.
Solution:
1. Check if the local voltage is normal;
2. Reduce the articles;
3. Use multimeter to check if the sensor (T2) is normal according to the PT1000, if
abnormal, change the sensor.
Problem Description:
1. The low water level sensor does not float after auto-adding water more than 3mins;
2. The high water level sensor does not float after auto-adding water more than 4.5mins.
Problem Reason:
1. The low water level sensor broken;
2. The high water level sensor broken;
3. The ports of water level sensor on PCB is loosed;
4. The low water level sensor cannot float due to external force;
5. Less water.
Solution:
1. Check if the water level sensor is normal. It is normal to hear a sound of “dip, dip” when
toggle the high water level sensor. Or change it;
SMM-B-02| Page 20 of 51
Sterilizer Maintenance Manual-TANZO Touch series
2. Check if the ports of the water level sensor on PCB is connected well;
3. Check if external force, such as the high water level sensor may interfere the floated low
water level sensor;
4. Check if the water is enough.
Problem Description: it only fits for these machines which are with two temp. sensors.
When enter the phase of constant temp. & pressure, the value difference between T1(real
value) and T2(test value) is more than 2℃.
Problem Reason:
1. Initial setup does not adjust well;
2. The inner sensor is broken;
3. Not saturated steam in chamber, with obvious temperature difference.
SMM-B-02| Page 21 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution:
1. Initial Fine Adjustment
1) Enter the adjustment interface: press “MENU” to enter “SYSTEM SET”, and then
choose “User Message”;
2) Then input the password “1111” to “factory set”. Check the “Inn temp.” item to
adjust the inner temperature you need, then store it;
SMM-B-02| Page 22 of 51
Sterilizer Maintenance Manual-TANZO Touch series
3) Return to the standby interface to check if the problem has been solved.
2. Use multimeter to check if every sensor is normal. Or replace the sensor.
3. Obvious temperature difference in chamber, it is wrong with the PCB program, replace it.
2) Then input the password “1111” to “factory set”. Check the “Water Check” item to
adjust the value you need, then store it;
SMM-B-02| Page 23 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Reason:
1. Too much steam during cycles, result in the door gasket leakage due to the door gasket is
ejected, out of sharp, weakened sealing effect;
2. The water pump is broken, fails to supply enough water for steam generator;
3. The steam generator is blocked or weakened heating effect, fail to supply enough steam
for the chamber, result in the pressure and temperature on the display panel do not rise or
rise unobviously, due to using non-designated water or non-scheduled cleaning program;
Constantly add water to steam generator;
4. The steam generator protector is broken or adjusted improperly, result in low temperature
of the steam generator;
5. Less heat from the heating ring, results in less steam in the chamber and water pump do
not work;
6. The solenoid valve or pipeline is blocked;
7. The air release valve cannot close after releasing.
Solution:
1. Reset or replace the door gasket according to E01;
2. Check if the water pump is broken or shortage. Or replace it;
3. Check and clean the steam generator. Or replace it;
4. Check if the adjusted lever of the protector is loosed. Tight the screw if loosed, or replace
the protector;
SMM-B-02| Page 24 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The chamber pressure cannot back to normal after finishing sterilization.
Problem Reason:
1. Older machines: the air return valve or the air filter is blocked;
2. New machines: 1) Initial setup does not reach ZERO state in standby condition; 2)
User’s altitude differs from factory setting.
Solution:
1. Check and clean the air filter;
Air Filter
2. Check and clean the core of the air return valve. Or replace the valve;
3. Setting standby pressure to ZERO state:
SMM-B-02| Page 25 of 51
Sterilizer Maintenance Manual-TANZO Touch series
4) Keep the door open, and then turn on the power, check the pressure;
Note If the pressure is Negative, then go to SYSTEM SET and adjust it to 0KPA.
5) Enter the adjustment interface: press “MENU” to enter “SYSTEM SET”, and then
choose “User Message”;
6) Then input the password “1111” to “factory set”. Check the “pressure” item, 10
times than the shown figure (e.g.: set it +10KPA if shows -1.0KPA), then store it;
7) Return to the standby interface to check if the pressure value is 0KPA or positive
SMM-B-02| Page 26 of 51
Sterilizer Maintenance Manual-TANZO Touch series
value.
Problem Description:
Leakage is more than 0.13bar during vacuum test.
Problem Reason:
1. The vacuum pump is broken;
2. Too many articles;
3. The normal-open valve is blocked;
4. 2-1 of 3-way valve is blocked;
5. The pipelines or connecting ports are leaked.
Solution:
1. Check and replace the vacuum pump;
2. Reduce the articles;
3. Check and clean the core of the normal-open valve. Or replace the valve;
4. Check and clean the core of the 3-way valve. Or replace the valve;
5. Check the pipelines and connecting ports.
Problem Description:
The inner temperature sensor (T2) cannot sense or bad sense the temperature.
Problem Reason:
1. Unstable voltage;
2. The connecting port of the sensor on the PCB is loosed;
3. The sensor is short circuit, or broken.
Solution:
1. Use multimeter to check if the local voltage is normal. Restart the machine if stable
voltage; change the power supply or connect voltage regulator if unstable voltage;
2. Check if the port of the sensor on the PCB is connected well;
SMM-B-02| Page 27 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The outer temperature sensor (T3) cannot sense or bad sense the temperature.
Problem Reason:
1. Unstable voltage;
2. The connecting port of the sensor on the PCB is loosed;
3. The sensor is short circuit, or broken.
Solution:
1. Use multimeter to check if the local voltage is normal. Restart the machine if stable
voltage; change the power supply or connect voltage regulator if unstable voltage;
2. Check if the port of the sensor on the chamber heating ring is connected well;
SMM-B-02| Page 28 of 51
Sterilizer Maintenance Manual-TANZO Touch series
3. Use multimeter to check if the sensor is normal. If the value is abnormal, replace the
sensor.
Problem Description:
The steam generator temperature sensor (T1) cannot sense or bad sense the temperature.
Steam Generator
Problem Reason:
1. Unstable voltage;
2. The connecting port of the sensor on the steam generator is loosed;
3. The sensor is short circuit, or broken.
Solution:
1. Use multimeter to check if the local voltage is normal. Restart the machine if stable
voltage; change the power supply or connect voltage regulator if unstable voltage;
2. Check if the port of the sensor on the steam generator chamber is connected well;
SMM-B-02| Page 29 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The inner test temperature sensor (T4) cannot sense or bad sense the temperature.
Problem Reason:
1. Unstable voltage;
2. Too many articles;
3. The connecting port of the sensor on the PCB is loosed;
4. The sensor is short circuit, or broken.
Solution:
1. Use multimeter to check if the local voltage is normal. Restart the machine if stable
voltage; change the power supply or connect voltage regulator if unstable voltage;
2. Reduce the article;
3. Check if the port of the sensor on the PCB is connected well;
4. Use multimeter to check if the sensor is normal. If the value is abnormal, replace the
sensor.
Problem Description:
The temperature in the chamber is higher than 140degree.
Problem Reason:
1. Inconsistent voltage;
2. Too many articles to disinfect;
3. The inner temperature sensor is broken;
4. The PCB is broken.
SMM-B-02| Page 30 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Solution:
1. Use multimeter to check if the local voltage is normal;
2. Reduce the articles;
3. Check the sensor according to E31;
4. Replace the PCB.
Problem Description:
The temperature of the chamber is higher than 140degree.
Problem Reason:
1. Inconsistent voltage;
2. Too many articles to disinfect;
3. The outer temperature sensor is broken;
4. The PCB is broken.
Solution:
1. Use multimeter to check if the local voltage is normal;
2. Reduce the articles;
3. Check the sensor according to E32;
4. Check if the chamber protector is normal;
Chamber Protector
5. Replace the PCB.
Problem Description:
The temperature of steam generator is higher than 200degree.
Problem Reason:
SMM-B-02| Page 31 of 51
Sterilizer Maintenance Manual-TANZO Touch series
1. Inconsistent voltage;
2. Too many articles to disinfect;
3. The steam generator sensor is broken;
4. The steam generator protector is broken;
5. The PCB is broken.
Solution:
1. Use multimeter to check if the local voltage is normal;
2. Reduce the articles;
3. Check the sensor according to E23;
4. Check the adjusted lever of the steam generator protector is loosed. Tight the screw if
loosed, or replace the protector;
5. Replace the PCB.
Problem Description:
The temperature in the chamber is higher than 145degree.
Problem Reason:
1. The inner temperature sensor is broken;
2. The PCB is broken.
Solution:
1. Check the sensor according to E31;
2. Replace the PCB.
Pressure Sensor
SMM-B-02| Page 32 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description: the temp. of T2 and T4 (inner sensor) are not reach 134℃.
Problem Reason:
1. The sensor is broken;
2. The PCB is broken.
Solution:
1. Check the sensor according to E31;
2. Replace the PCB.
When abnormal exit, press button for 3secs to force quit, the interface shows as below,
Attention Do not open the door if the pressure in the chamber is not “0 BAR”.
The interface shows as below after the power comes back from sudden power cut, when the
SMM-B-02| Page 33 of 51
Sterilizer Maintenance Manual-TANZO Touch series
pressure in the chamber falls to 0bar, then opens the door and restarts the machine.
Attention Do not open the door if the pressure in the chamber is not “0 BAR”.
SMM-B-02| Page 34 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The screen cannot display when turn on the power switch (the light of the buttons is
illuminated).
Problem Reasons:
1. The fuse is burned;
2. The connecting line between LCD screen and PCB is loosed;
3. LCD screen PCB is broken;
4. The power supplier is unstable;
5. Transformer is burned.
Solution:
1. Check if the power switch is light after turn on the power. And then check if the fuse left
the power switch is burned, or replace the fuse;
2. Check if the connecting line is fixed well;
26P Line
3. A sound of “dip, dip” when open/close the door, and then no display on LCD screen,
replace the LCD screen PCB;
4. Messy code or Discontinuous flicker, it means power supply is unstable, to change the
power supply or connect voltage regulator if unstable voltage;
5. Power switch lighting, power indicator on the main PCB extinguished, means no power
input, to replace the transformer.
SMM-B-02| Page 35 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
Press START button, the program cannot be started.
Problem Reasons:
1. LCD displays: Used water tank is full;
3. The illuminated means the door do not closed completely; or the micro switch is
broken;
SMM-B-02| Page 36 of 51
Sterilizer Maintenance Manual-TANZO Touch series
4. Connection problem: the buttons are out of order without any sound;
5. The PCB is broken.
Solution:
1. Drain water from used water tank;
Note: Drained water temperature should be lower than 70C. If higher than that, please
check if Fan is in good condition.
2. Fill distilled water automatically or manually;
Automatically: Plug the water pipe with adding connection, and then keep long press
the button;
Manually: add the water till hear a sound of “di, di”, and then press the button to
enter MENU.
Note: If water short alarms during sterilization cycle, don’t’ worry, water inside is
enough to complete full cycle. But fill sufficient water before next cycle.
3. Check if door handle is locked properly. And check the door micro-switch by pressing
down. Check if the micro-switch is normal if door locking icon doesn’t stop flashing, or
replace it;
4. Check if the connecting line between the screen and main PCB is fixed well;
5. Replace Master PCB.
SMM-B-02| Page 37 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Reasons:
1. Negative pressure in the chamber due to air cannot return completely after sterilization;
2. If leave the door closed after sterilization, it is a sealed space in the chamber. When the
hot chamber air colds down, it creates negative pressure, result in the door locked.
Caution Do not close the door when finish the sterilization cycle.
Solution:
1. Turn on power switch, the air return valve will open, and then try to open the door;
2. Find relief valve at rear cover, open it counter-clockwise to make the air return to the
chamber.
Caution If the door cannot open frequently, clean or replace the air return valve.
Problem Description:
1. When the clean water tank is empty and after filling water, Icon sparkles;
2. When the used water tank is full and after draining water, Icon sparkles.
Problem Reason:
1. The high or low water level sensor out of order (Clean water tank);
2. Water level sensor out of order (Used water tank).
Solution:
1. Check if the water level sensor is normal. It is normal to hear a sound of “dip, dip” when
toggle the high/low water level sensor. Or change it;
SMM-B-02| Page 38 of 51
Sterilizer Maintenance Manual-TANZO Touch series
2. Make sure that all sensor cables are well connected with Master PCB;
3. Lift the float of the water level sensor in the used tank up to the top, you will see the no
WASTER WATER OVER alarm on the screen. Or replace the single water level sensor.
Problem Reason:
1. Drainage or adding connection is blocked;
2. The valve core seal of connection is broken.
Solutions:
1. Check and clean the connections. Or replace it;
SMM-B-02| Page 39 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Problem Description:
The water pump does not work after press the automatic filling button.
Problem Reason:
1. The water pump falls off from the yellow connecting line;
2. The water pump connecting line falls off from the main PCB;
3. No voltage of output terminal on the PCB;
4. The water pump is broken.
Solutions:
1. Ensure the water pump is connected well with yellow wire;
2. Check if the connecting line between the water pump and PCB is connected well;
Problem Description:
Small amount of water left inside the chamber after sterilization.
Problem Reasons:
1. The inner filter is blocked or not close to the bottom of chamber;
2. Placed improperly, such as when the front of sterilizer is lower than the rear, much water
will stay in the front.
Solutions:
1. Clean or replace the chamber filer;
SMM-B-02| Page 40 of 51
Sterilizer Maintenance Manual-TANZO Touch series
SMM-B-02| Page 41 of 51
Sterilizer Maintenance Manual-TANZO Touch series
1. Enter the adjustment interface: press “MENU” to enter “SYSTEM SET”, and then
choose “User Message”;
2. Then input the password “1111” to “factory set”. Check the “pressure” item, 10 times
than the shown figure (e.g.: set it +10KPA if shows -1.0KPA), then store it;
3. Return to the standby interface to check if the problem has been solved.
SMM-B-02| Page 42 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Sterilization Time:
1. 105C Range: 30 - 90min;
2. 121C Range: 20 - 60min;
3. 134C Range: 04 - 40min.
SMM-B-02| Page 43 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Under maintenance interface to select vacuum test, press button to start the test.
Whole test time is 20mins, in which 05mins of vacuuming; 05mins of stability; 10mins of
keeping.
SMM-B-02| Page 44 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Helix Test
1. Test Tools:
GKE-Steri-Record
and Helix pipe
2. Put the unused helix paper into the helix tube, lock it and ensure sealed well.
Insert the test paper to the test pipe head. Set back the pipe head
3. Put the helix tube into the chamber, and start the Helix Test – 134C, 3.5mins sterilization,
3times vacuum, 8mins drying. If the color changes to black, it means pass.
Before After
SMM-B-02| Page 45 of 51
Sterilizer Maintenance Manual-TANZO Touch series
B&D Test
B&D test is to show the effect of pre-vacuum and exhausting air of Class B autoclave. It is
only for test if there is cold air in the chamber during cycle, not for the sterilization results.
Put test bag in the chamber, and start the test – 134C, 3.5mins sterilization, 3times vacuum,
8mins drying. If the color changes to black, it means pass.
B&D Test Bag Test Paper between Test Bags Test Paper
Before After
SMM-B-02| Page 46 of 51
Sterilizer Maintenance Manual-TANZO Touch series
Note:
1. Advised to rinse pipes per 6 month;
2. Make sure the inner chamber filter is well mounted, and clean it after rinsing pipes.
SMM-B-02| Page 47 of 51
Sterilizer Maintenance Manual-TANZO Touch series
SMM-B-02| Page 48 of 51
Sterilizer Maintenance Manual-TANZO Touch series