Service Manual
English
Table of contents Sirona Dental Systems GmbH
Sinius Service Manual
Table of contents
1 Structure of the document ...................................................................................... 7
1.1 Identification of the danger levels ................................................................ 7
1.2 Formats and symbols used.......................................................................... 7
1.3 Symbols ....................................................................................................... 8
1.4 Abbreviations ............................................................................................... 9
3 Prerequisites ........................................................................................................... 12
63 69 149 D3561
2 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH Table of contents
Sinius Service Manual
63 69 149 D3561
D3561.076.01.08.02 10.2020 3
Table of contents Sirona Dental Systems GmbH
Sinius Service Manual
63 69 149 D3561
4 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH Table of contents
Sinius Service Manual
63 69 149 D3561
D3561.076.01.08.02 10.2020 5
Table of contents Sirona Dental Systems GmbH
Sinius Service Manual
63 69 149 D3561
6 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 1 Structure of the document
Sinius Service Manual 1.1 Identification of the danger levels
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily
injury or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTE
A possibly harmful situation which could lead to damage of the
product or an object in its environment.
IMPORTANT
Application instructions and other important information.
63 69 149 D3561
D3561.076.01.08.02 10.2020 7
1 Structure of the document Sirona Dental Systems GmbH
1.3 Symbols Sinius Service Manual
1.3 Symbols
SINIUS dentist element
Foot control
LEDview
LEDview Plus
Patient chair
Water unit
63 69 149 D3561
8 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 1 Structure of the document
Sinius Service Manual 1.4 Abbreviations
1.4 Abbreviations
AE Dentist element
AK Connection box
ASE Assistant element
DTC Dental treatment center
BL Bootloader
BL motor Brushless motor
UI User interface
DLM Download manager
DWSS Dürr wet suction system
DUMC Dental Unit Management Console
(PC program for service support)
EBL Extended bootloader
PCB Printed circuit board/board
FS Foot control
GND Ground
HW Hardware
L Line
LED Light emitting diode
SV Solenoid valve
PC Personal computer
XRIV X-ray image viewer
S Switch
SBL Standard bootloader
SDI Sirona Dental Interface
(electrical, pneumatic, hydraulic plug connection)
SINIUS Travel track device
SINIUS CS Instrument arm device (continental style)
SINIUS TS Swivel table unit
ST Patient chair
SW Software
VDC DC voltage
WU Water unit
X Connector
63 69 149 D3561
D3561.076.01.08.02 10.2020 9
2 General safety information Sirona Dental Systems GmbH
2.1 Qualification of service personnel Sinius Service Manual
63 69 149 D3561
10 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 2 General safety information
Sinius Service Manual 2.6 EMC interference suppression
63 69 149 D3561
D3561.076.01.08.02 10.2020 11
3 Prerequisites Sirona Dental Systems GmbH
Sinius Service Manual
3 Prerequisites
Take this service manual, a service notebook, the spare parts list, and
the connection documents with you on every customer visit.
You will need the DUMC (Dental Unit Management Console) to read the
service information and update the treatment station software.
The technical documentation and product software are subject to
constant updates. Download the current versions from the dealer area
of the Sirona home page on a regular basis.
Always refer to the DUMC when performing service work.
63 69 149 D3561
12 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 4 Technical data
Sinius Service Manual 2.7 Operating the treatment center with Heliodent Plus unit model
4 Technical data
Model designation: Sinius / Sinius CS / Sinius TS
Power connection: 100 – 240 V AC ± 10%
50/60 Hz
Rated current: 2.2 A at 240 V
4.35 A at 115 V
5.0 A at 100 V
also max. 6 A for external devices
Type of ground connection: TN-C-S system or TN-S system (acc.
to IEC 60364-1)
Overvoltage category: 2 acc. to IEC 60664-1
Average power consump- 0.25 kW
tion (for dimensioning an air
conditioning system):
Power consumption in the 3 W (without internal mini PC)
Standby mode:
Main building fuse: Type B automatic circuit breaker
100 - 115 VAC: 20 A medium-blow
220 – 240 V AC: 16 A medium-blow
63 69 149 D3561
D3561.076.01.08.02 10.2020 13
4 Technical data Sirona Dental Systems GmbH
4.1 Operating, transport and storage conditions Sinius Service Manual
IMPORTANT
Minimum requirements for the PC
See document "Installation instructions and system requirements for
PC configuration," (REF 61 94 075) Sivision Digital.
63 69 149 D3561
14 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 4 Technical data
Sinius Service Manual 4.2 Weight and packaging
63 69 149 D3561
D3561.076.01.08.02 10.2020 15
5 Important information Sirona Dental Systems GmbH
5.1 Data storage locations for user-specific data Sinius Service Manual
5 Important information
IMPORTANT
Reading the IP address
Before you replace the HSA board, if possible it is a good idea to read
and note down the IP address.
63 69 149 D3561
16 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.1 Data storage locations for user-specific data
63 69 149 D3561
D3561.076.01.08.02 10.2020 17
Sirona Dental Systems GmbH Table of contents
Sinius Service Manual
63 69 149 D3561
D3561.076.01.08.02 10.2020 3
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.2 Switching the treatment center on/off
63 69 149 D3561
D3561.076.01.08.02 10.2020 19
5 Important information Sirona Dental Systems GmbH
5.2 Switching the treatment center on/off Sinius Service Manual
63 69 149 D3561
20 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.2 Switching the treatment center on/off
inexistent for
SINIUS CS
63 69 149 D3561
D3561.076.01.08.02 10.2020 21
5 Important information Sirona Dental Systems GmbH
5.2 Switching the treatment center on/off Sinius Service Manual
63 69 149 D3561
22 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.2 Switching the treatment center on/off
63 69 149 D3561
D3561.076.01.08.02 10.2020 23
5 Important information Sirona Dental Systems GmbH
5.2 Switching the treatment center on/off Sinius Service Manual
63 69 149 D3561
24 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.2 Switching the treatment center on/off
63 69 149 D3561
D3561.076.01.08.02 10.2020 25
5 Important information Sirona Dental Systems GmbH
5.3 CAN BUS Sinius Service Manual
Download phase 2
The download manager starts the download process, updates the node
software via DUMC if necessary, and launches the application.
SETUP key
ü All instruments are in place.
1. Hold down the “Setup“ fixed key for at least 2 seconds.
63 69 149 D3561
26 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.3 CAN BUS
63 69 149 D3561
D3561.076.01.08.02 10.2020 27
5 Important information Sirona Dental Systems GmbH
5.3 CAN BUS Sinius Service Manual
Resistance = 60 ohms
Cable run OK
Measurement of CANH → CANL
Resistance = 0 ohms
Cable run OK
Measurement of CANL source → CANL target
63 69 149 D3561
28 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.4 S-COM
NSK NAA
CAN + CAN -
5 3 X3 inexistent for 5 3
X31
SINIUS CS
GN WHGN
future connector
GN WHGN GN WHGN BU RD
X5 6 3 X5 5 3 X8 4 3
X6
4
X28
A2 3
4 mm Isolation
CAN + CAN -
A3
5.4 S-COM
On the S-COM, the PC software functions are performed for the user
interface and the user interface software is downloaded via FTP.
If the S-COM connection has been established, the S-COM LED lights
up on the HSA and HAP1 / HAP2 boards. If this LED is flashing, the
connection is being established or a malfunction has occurred. The line
connection is, however, present. If the LED goes out, no data
connection is present. In this case, it must be assumed that there is a
disruption in the line.
Unlike the CAN BUS, there is no defined terminating resistor with this
bus.
63 69 149 D3561
D3561.076.01.08.02 10.2020 29
5 Important information Sirona Dental Systems GmbH
5.4 S-COM Sinius Service Manual
NAJm HAP
S-COM + X1 SDI X22 X1 X1
B4 BN 9 9 BN 9 A5 RD A5
HSA RJ45
WHBN WHBN BU
A4 7 7 7 A4 A4
S-COM -
63 69 149 D3561
30 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.5 SDI connector
19
20
18
16
17
15
14
12
13
6 11
10
1
5
20
18
11 14
15
19
12
13
17
+
10
9
IN
LIFST
T
6
17
16
15
4
5
2
3
1
14
16
13
20
22
12
11
18
6
10
23
5
1
21
19
1 GND 9 S-COM+ 17 NC
2 NC 10 Unit ON/OFF 18 GND PK
3 CAN- 11 NC 19 NC
4 Foot_Ctrl 12 36VDC EK 20 36VDC PK
5 CAN+ 13 NC
6 Chair stop 14 GND EK
7 S-COM- 15 NC
8 NC 16 NC
63 69 149 D3561
D3561.076.01.08.02 10.2020 31
5 Important information Sirona Dental Systems GmbH
5.6 Patient chair Sinius Service Manual
63 69 149 D3561
32 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.6 Patient chair
B B
A
A
IMPORTANT
It is prohibited for any obstacles to be located in the movement range
of the chair during the reference travel.
2.
The reference travel will start only if no safety switch has been
> 2s triggered. This also requires the cuspidor bowl to be retracted.
3. To do this, press the Reference travel key in the service area of the
user interface on the dentist element. >2s
Ä The orange rectangle around the Reference travel key will flash
3. while the chair performs the movements.
IMPORTANT
Do not actuate the chair movement buttons!
Pressing any chair movement button will stop the reference travel,
e.g. the chair movement buttons on the user interface of the dentist
element, the assistant element, the 4-way foot switch, the toggle
switch or the chair buttons on the control panel.
NOTE
Not maneuverable with external voltage
The motor then receives voltage and the position encoder does not.
The position stored in the encoder then no longer agrees with the
actual motor position. This cannot even be corrected via a reference
run with the Sinius (sliding track device).
63 69 149 D3561
D3561.076.01.08.02 10.2020 33
5 Important information Sirona Dental Systems GmbH
5.7 Safety switches Sinius Service Manual
63 69 149 D3561
34 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.7 Safety switches
SS_HE SS_HE
X36
L400A X28 X15 X35
X22 X32
S14 S14a X5
X25
L418 X27
X26
S13 SS_MB X34
X29
NWEm X9
X18
X10
X8
X14
X20
F803 X12
L443 X19
X13
X3 X7
L442 X4
X1
X2
L443 L402c L441 X11
L444
S6 SS_FA
L402c
S5a SS_HV
X13
X32
SS_RL S7 X28
S5 SS_HV X31
SS_HR S3 X30
X21
HSA X4
X6
X3 X1
S3a L400A X2
SS_HR X23 X5 X19
X16 X7 X9 X26
X15
X20 X14 X10
L400A
63 69 149 D3561
D3561.076.01.08.02 10.2020 35
5 Important information Sirona Dental Systems GmbH
5.8 Water heater in dentist element and water unit Sinius Service Manual
I
No. of the safety switch on the EasyTouch user interface (fault display)
II
Identification of the safety switch according to the circuit diagram
CAUTION
The safety switch may still be hot if it was only recently triggered.
63 69 149 D3561
36 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.9 Acoustic feedback from the treatment center
63 69 149 D3561
D3561.076.01.08.02 10.2020 37
5 Important information Sirona Dental Systems GmbH
5.9 Acoustic feedback from the treatment center Sinius Service Manual
63 69 149 D3561
38 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 5 Important information
Sinius Service Manual 5.10 Overview of signal paths for camera control
Human inter-
SiroCam
face device
control
/ R232-driver
KL 1
photo sensor
control
video
KL 6
magnet
KL 5
photo sensor
Sivision
Connect
Si Video / Ethernet CAN NAJm
Sidexis NSA
TCP/IP if S1.2 = OFF
Siucom Plus
BOF
SiroCam control NSB
PC dialog Footswitch
The presence of the SiroCam AF/AF+ camera is sent to the NAJm from the
USB Video class driver via Sivision Connect.
Camera in instrument holder 5 (KL 5) Camera in instrument holder 6 (KL 6)
In instrument holder 5, removal of the If the camera is located in external in-
camera is detected via a light barrier if strument holder 6 with integrated
the NAJmswitch S1.2 (camera) is set magnet (NAJm S1.2 = ON), the cam-
to OFF. era sends the removal signal to the
NAJm via USB >Siucom > Ethernet >
CAN.
● The removal signal for the camera is transferred to Sivision Connect on the
external PC via CAN > Ethernet.
The Sivision Connect sends the camera ON signal to the camera via USB.
The user interface switches over to the PC dialog box when all other
instruments have been set down.
● Sidexis/SiVideo represents the live image of the camera in video overlay,
even when a different instrument is active.
● If the camera is the active instrument, the camera functions are controlled
via the connected foot switch.
63 69 149 D3561
D3561.076.01.08.02 10.2020 39
5 Important information Sirona Dental Systems GmbH
5.11 Test of the main cut-off valve (MVAB test) Sinius Service Manual
63 69 149 D3561
40 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.1 Locations of modules and boards
LV
AE CS
ASE
FU FS WE AE ST RÖBI
Component Board/module
ST Patient chair HSA Connection box
NSB Wireless foot control base station
NSK Headrest
NSU USB connection to patient chair
FS Foot control
FU Wireless foot control SFB Wireless foot control
AE Dentist element CC Dentist element instrument deposit detection ID (5 x)
NAA ApexLocator
NAH Apex mucosal electrode
NAC/HAC / NAC BL motor control
NAJm Dentist element control module
HAP2 (SINIUS / Dentist element baseboard user interface
SINIUS TS)
HAP1 (SINIUS CS)
HPE (SINIUS CS) Support arm brake sensor
HCC (SINIUS CS) Instrument deposit detection
NAU Dentist element USB connection
XRIV X-ray image viewer NAR X-ray viewer board
63 69 149 D3561
D3561.076.01.08.02 10.2020 41
6 Overview of modules and boards Sirona Dental Systems GmbH
6.1 Locations of modules and boards Sinius Service Manual
Component Board/module
ASE Assistant element NAS Sub-board on NHE
NHEm Assistant element controller without Sprayvit
NHE Assistant element controller with Sprayvit
HTTL Assistant element user interface - keys on the left
HTTR Assistant element user interface - keys on the right
HHK Instrument deposit detection
NOP 5V power supply
WU Water unit NWEm Water unit control module
NWM Automatic tumbler filling
LV LEDview NLV Examination light control
LV S LEDview S Examination light control
PLUS
LV+ LEDview XLV Examination light control
63 69 149 D3561
42 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
63 69 149 D3561
D3561.076.01.08.02 10.2020 43
6 Overview of modules and boards Sirona Dental Systems GmbH
6.2 Patient chair Sinius Service Manual
63 69 149 D3561
44 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
63 69 149 D3561
D3561.076.01.08.02 10.2020 45
6 Overview of modules and boards Sirona Dental Systems GmbH
6.2 Patient chair Sinius Service Manual
63 69 149 D3561
46 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
NOTE
Swinging away the cuspidor bowl
If safety switch S12 is active (SS code/flashes 2x), the chair can only
move upward when the cuspidor bowl is swung away.
Motor locations
M2
M1
63 69 149 D3561
D3561.076.01.08.02 10.2020 47
6 Overview of modules and boards Sirona Dental Systems GmbH
6.2 Patient chair Sinius Service Manual
V50
V41 2
0 V30
V30 7
1
V40
5
V52 V30
3 3
V30
6
NSK
2500 2
V5V5 60623
VV
2500 2
V6V6 40
V
4
X
0115 9 1
V2V2 20212 X
VV
63 69 149 D3561
48 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
63 69 149 D3561
D3561.076.01.08.02 10.2020 49
6 Overview of modules and boards Sirona Dental Systems GmbH
6.2 Patient chair Sinius Service Manual
63 69 149 D3561
50 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
● CRC error
– V405 flash code 5x
– Replace the NSB
63 69 149 D3561
D3561.076.01.08.02 10.2020 51
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
NAJm
V
V 21
V 2 2
V 3 01
J1 V6 30403
J1 1 00
J1 2
3
X2
V
V 2
V 2 16
V 3 11
S1 V3 30 05
00 2
63 69 149 D3561
52 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
63 69 149 D3561
D3561.076.01.08.02 10.2020 53
Table of contents Sirona Dental Systems GmbH
Sinius Service Manual
63 69 149 D3561
6 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
X15
X3
MP_LIGHT-
MP_LIGHT+
X6
X14
X13
X12
X5
X1
X4
X8
X7
+5V_Iso
HAC
36
MP
H3 H1
MP MP
3V3 Ana
36V
12V
3V3
5V
Boot
Run
Error
Reset
Debug
KL
FS
Components on the board
LEDs
3 V3 (green) Operating voltage
3 V3 analog (green) Operating voltage
5 V DC (green) Operating voltage
12 V DC (green) Operating voltage
36 V DC (green) Operating voltage
5 V iso (green) Operating voltage 5 V from NAJm
Reset (red) Lights up shortly after switching on.
Error (red) ● 2 x No connection to HSA
● 3 x Self-test failed.
● 4 x Defective FRAM data.
● 5 x Critical error - reboot required.
● 6 x Error in motor behavior, motor does not start.
● 7 x Error in hardware foot control signal.
● 8 x Error in instrument detection.
● 9 x Inconsistent parameters received.
● 11 x Internal communication problem.
● 12 x Programming errors.
Run (yellow) Flashes at 1 Hz in standby
Flashes at 10 Hz when an instrument is active
Test 1 (yellow)
Test 3 (yellow)
63 69 149 D3561
D3561.076.01.08.02 10.2020 55
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
63 69 149 D3561
56 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
04
V236
V205 NAC
V212
V201
V302
V303
V312
V311
V310
V304
V303
V2
01
V9
X8
X6
X1 5
X
2
X1 3
X1 4
X1 5
X1
X3
Components on the board
LEDs
V203 3.3 V (green) Operating voltage
V204 3.3 V analog (green) Operating voltage
V205 5 V DC (green) Operating voltage
V212 12 V DC (green) Operating voltage
V236 36 V DC (green) Operating voltage
V310 Reset (red) Lights up shortly after switching on.
63 69 149 D3561
D3561.076.01.08.02 10.2020 57
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
63 69 149 D3561
58 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
IMPORTANT
No BL-ISO-S motor can be operated with the NAC board!
63 69 149 D3561
D3561.076.01.08.02 10.2020 59
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
X6 3
X6 NAA 3V
NAA MP3V3 -5V
P P
M M -5V
MP
X7
X7 304 +5
V
D
X 04 MP+5VP GND
X3 1V
8
MP M GN_EK
MP1V8 P
M +5V_EK EK
P V V_
M +5 +5 _EK
MP+5V
MP
V
36V
36 X31
1
NAH
X3
V S -
36VP FSCAN- AN+
MP36MP
C +
MFPCAPNAN
M MMP MP C
P
M
X66
LEDs
63 69 149 D3561
60 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
Connector
Connector
63 69 149 D3561
D3561.076.01.08.02 10.2020 61
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
V
+5 V
, 3
+3
36
V T HAP1
SE T
RE IGH
V K L
+5 AC
B
7
X1
X1
X4
X2
R X 10
RO UN D 5
ER R AR X1
O
YB X1 X1
KE 6
0 Y
X1 KE D
Y E
ANESS 1 X3
PR X1 X3
M X1
W
_P 4
BL D X2
/S R
D4 NO ANM
LEND_2/CCO
A ED1/S
D 3/N LED
E L
L
HAP1 board
63 69 149 D3561
62 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 4 Technical data
Sinius Service Manual 4.2 Weight and packaging
63 69 149 D3561
D3561.076.01.08.02 10.2020 15
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
e
bu
g od
De sm
Te
MP2 M P1
JU JU
3
MP
JU
HAP2
HAP2
X1
4
X1
5
0 K
X1 +5 R E
V ERUNYBO
3 BL _RE RO AR
X1 _
SD PW
G R D
X1 M AN
NA PRY K
CAND ES EY
SCN _N SE
X3 X1 BA OM OR D
36 7 CK
V LIG
HT
RE
S +
X4 X2
X1 ET +3,3V
0 5V
X1
1
63 69 149 D3561
64 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
IMPORTANT
In the event of a fault during the warranty period, please replace the
complete user interface.
Exception: During the warranty period the customer damaged the
membrane keypad and wants it replaced. This is not a warranty case.
You can replace just the front plate here.
63 69 149 D3561
D3561.076.01.08.02 10.2020 65
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
63 69 149 D3561
66 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
CC
1
1 2
V 1 2
X 2
V V 1 3
3
V V 1 4
4
V V 1 5
5
V V 1
V
IMPORTANT
The function of the light barriers can be checked via the USER LED
on the NAJ.
63 69 149 D3561
D3561.076.01.08.02 10.2020 67
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
HCC
2 1 1
1 V 1
X 2 2
V V 1 3 3
V V 1 4 4
1 4
V V 1 2
X 5 5
V X V 1 3
1
V X
63 69 149 D3561
68 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.3 Dentist element
HPE
IMPORTANT
When instruments have been deposited, the function of the two
sensors can be tested via the USER LED of the NAJ. The USER LED
lights up when the sensors are approached.
63 69 149 D3561
D3561.076.01.08.02 10.2020 69
6 Overview of modules and boards Sirona Dental Systems GmbH
6.4 Water unit Sinius Service Manual
5V
X1
4
X2
US 0
OTA X1
N V1
P 03 2
_S F8
SS 6V
3 X1
3 DS
3 0
X3
X7
S3
X1 X4
X2
X1
1
63 69 149 D3561
70 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.4 Water unit
63 69 149 D3561
D3561.076.01.08.02 10.2020 71
6 Overview of modules and boards Sirona Dental Systems GmbH
6.4 Water unit Sinius Service Manual
63 69 149 D3561
72 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.4 Water unit
63 69 149 D3561
D3561.076.01.08.02 10.2020 73
6 Overview of modules and boards Sirona Dental Systems GmbH
6.4 Water unit Sinius Service Manual
63 69 149 D3561
74 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.4 Water unit
CN
1
NWEm
CN
2
LE
D1
CN
3
LE
CN D2 CN
4 5
CN
7 LE
D3
CN
8
NWM
NWM
CN
6
63 69 149 D3561
D3561.076.01.08.02 10.2020 75
6 Overview of modules and boards Sirona Dental Systems GmbH
6.5 Assistant element Sinius Service Manual
63 69 149 D3561
76 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.5 Assistant element
63 69 149 D3561
D3561.076.01.08.02 10.2020 77
6 Overview of modules and boards Sirona Dental Systems GmbH
6.5 Assistant element Sinius Service Manual
63 69 149 D3561
78 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.5 Assistant element
63 69 149 D3561
D3561.076.01.08.02 10.2020 79
6 Overview of modules and boards Sirona Dental Systems GmbH
6.5 Assistant element Sinius Service Manual
HTTL
63 69 149 D3561
80 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.5 Assistant element
HTTR
Connector
X2 > L438 > X12 NHE
63 69 149 D3561
D3561.076.01.08.02 10.2020 81
6 Overview of modules and boards Sirona Dental Systems GmbH
6.5 Assistant element Sinius Service Manual
NOP
X2
X4
X3
Components on the board
LEDs
V100 33 V DC (green) Operating voltage
V101 5 V DC (green) Operating voltage
Connector
X2 > L238 > mini LED
X3 > L436 > X3 NHE
X4 > L238 > mini LED
63 69 149 D3561
82 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.5 Assistant element
1 2 3
V V V 4 5
V V 6 7 8
V V V
Components on the board
Sensors
V1/2 Sensor for instrument holder 1
V3/4 Sensor for instrument holder 2
V5/6 Sensor for instrument holder 3
V7/8 Sensor for instrument holder 4
Connector
> ASE_Claw_Sensor_mold_L435 cable > X11 NHE
63 69 149 D3561
D3561.076.01.08.02 10.2020 83
6 Overview of modules and boards Sirona Dental Systems GmbH
6.6 LEDview Sinius Service Manual
6.6 LEDview
6.6.1 NLV board - LEDview control
63 69 149 D3561
84 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.6 LEDview
HLV
1.
X2
X3
4.
1.
X4
X1
2.
24
10Com-32
0% p V
onon
X2
4.
HLV
X3
X4
3.
2.
X1
63 69 149 D3561
D3561.076.01.08.02 10.2020 85
6 Overview of modules and boards Sirona Dental Systems GmbH
6.7 LEDview Plus Sinius Service Manual
63 69 149 D3561
86 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.9 X-ray image viewer
Battery
Due to the high level of internal resistance, some battery types do not
provide the required peak currents.
Tested positive and working with Sirona: Varta High Energy type: 4914
Wireless module:
The wireless module operates at a frequency of 2.4 GHz.
Zigbee, Bluetooth, and WLAN do not interfere with the wireless module.
GND C
L247
NAR
KL
1
8 KL
2
1 3
2 4
5 7
6 8
63 69 149 D3561
D3561.076.01.08.02 10.2020 87
6 Overview of modules and boards Sirona Dental Systems GmbH
6.1 Locations of modules and boards Sinius Service Manual
Component Board/module
ASE Assistant element NAS Sub-board on NHE
NHEm Assistant element controller without Sprayvit
NHE Assistant element controller with Sprayvit
HTTL Assistant element user interface - keys on the left
HTTR Assistant element user interface - keys on the right
HHK Instrument deposit detection
NOP 5V power supply
WU Water unit NWEm Water unit control module
NWM Automatic tumbler filling
LV LEDview NLV Examination light control
LV S LEDview S Examination light control
PLUS
LV+ LEDview XLV Examination light control
63 69 149 D3561
42 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
7 Replacing components
NOTE
Disconnecting a component from the chair
A large volume of water can escape from the connections when
disconnecting a component.
➢ Empty the rinsing tank (see Emptying the rinsing tank [→ 119])
and the mixing tank (see Mixing tank [→ 120]).
➢ Before disconnecting the dentist element, assistant element, or
water unit from the chair, switch the treatment center off at the
main switch.
NOTE
Modules that carry water and air may still be pressurized after
switching off the treatment center.
➢ The pressure of the lines can be relieved by unscrewing the water
and air filters. The filters can be accessed via the maintenance
flap of the water unit.
NOTE
Prying or twisting cover panels causes damage
Do not open the cover by prying it with or twisting the screwdriver!
This can damage the covers.
➢ Push the screwdriver only in the axial direction and raise the
cover.
63 69 149 D3561
D3561.076.01.08.02 10.2020 89
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
NOTE
Swinging away the cuspidor bowl
If safety switch S12 is active (SS code/flashes 2x), the chair can only
move upward when the cuspidor bowl is swung away.
Motor locations
M2
M1
63 69 149 D3561
D3561.076.01.08.02 10.2020 47
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.2 Patient chair
63 69 149 D3561
D3561.076.01.08.02 10.2020 49
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
A
B
A
63 69 149 D3561
92 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
C B
A
1. Press the latching nose (A) inwards and tilt the seat upholstery
upwards.
2. Press the 2 latching noses (C) on both sides of the seat upholstery.
Ä The cushion is released from the bracket.
3. Remove the seat upholstery.
63 69 149 D3561
D3561.076.01.08.02 10.2020 93
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
A B
63 69 149 D3561
94 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
Motor-driven headrest
➢ Release the latch (A) and remove the backrest cover (B).
63 69 149 D3561
D3561.076.01.08.02 10.2020 95
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
V50
V41 2
0 V30
V30 7
1
V40
5
V52 V30
3 3
V30
6
NSK
2500 2
V5V5 60623
VV
2500 2
V6V6 40
V A
4
X
0115 9 1
NSK
V2V2 20212 X
VV
IMPORTANT
When opening the cover, make sure that the cable loop runs beneath
the fixing pin (A).
NOTE
Do not damage the half-shell covers
If you pry the outer half-shell covers open with the screwdriver, they
may be damaged.
➢ Open the half-shell covers by twisting the screwdriver.
63 69 149 D3561
96 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
63 69 149 D3561
D3561.076.01.08.02 10.2020 97
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
D B
A
ü The dentist element is open.
1. Unscrew the two screws (A) on the inner cover.
2. Slacken the three screws (B) on the fixing rail (C).
3. Slide the fixing rail (C) forward to release the user interface.
4. Angle the user interface (D) slightly downward and then pull it
forward and out.
5. Unplug connector X1 from the user interface.
63 69 149 D3561
98 D3561.076.01.08.02 10.2020
6 Overview of modules and boards Sirona Dental Systems GmbH
6.3 Dentist element Sinius Service Manual
e
bu
g od
De sm
Te
MP2 M P1
JU JU
3
MP
JU
HAP2
HAP2
X1
4
X1
5
0 K
X1 +5 R E
V ERUNYBO
3 BL _RE RO AR
X1 _
SD PW
G R D
X1 M AN
NA PRY K
CAND ES EY
SCN _N SE
X3 X1 BA OM OR D
36 7 CK
V LIG
HT
RE
S +
X4 X2
X1 ET +3,3V
0 5V
X1
1
63 69 149 D3561
64 D3561.076.01.08.02 10.2020
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
63 69 149 D3561
100 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
5. Align the upper cover (B) with the front of the dentist element and
guide it into the grooves provided.
7.1.4 Negatoscope
7.1.4.1 Removing the NAR board
63 69 149 D3561
D3561.076.01.08.02 10.2020 101
6 Overview of modules and boards Sirona Dental Systems GmbH
6.4 Water unit Sinius Service Manual
5V
X1
4
X2
US 0
OTA X1
N V1
P 03 2
_S F8
SS 6V
3 X1
3 DS
3 0
X3
X7
S3
X1 X4
X2
X1
1
63 69 149 D3561
70 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.4 Water unit
63 69 149 D3561
D3561.076.01.08.02 10.2020 73
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
CAUTION
If external suction host cleaning (CDS60) is connected, wear
protective goggles and gloves when working on the water unit!
3 4
2 5
1 6
0 bar
3 4
2 5
1 6
0 bar
B A
1. Carefully pry the side panels (A) apart and pull them away.
2. Remove the service flap (B).
3. Unscrew the two screws from each of the covers (C); then unhook
the covers and pull them away.
4. Remove the lid (D).
D
63 69 149 D3561
104 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
7. Remove the two screws from the lid (G) and lift it off.
8. Remove the cover panel (H) of the funnel.
H NOTE
Watch out for the latching nose (I).
G I
63 69 149 D3561
D3561.076.01.08.02 10.2020 105
Sirona Dental Systems GmbH 6 Overview of modules and boards
Sinius Service Manual 6.5 Assistant element
63 69 149 D3561
D3561.076.01.08.02 10.2020 79
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
63 69 149 D3561
D3561.076.01.08.02 10.2020 107
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
63 69 149 D3561
108 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
1. Press the front cover together to release the snap lock and then
remove the cover (G).
2. Unscrew the four screws (G) in the rear cover and remove the cover
(H).
X6 X5
NLV
X1
2.
63 69 149 D3561
D3561.076.01.08.02 10.2020 109
7 Replacing components Sirona Dental Systems GmbH
7.1 Removing the cover Sinius Service Manual
IMPORTANT
Data backup
➢ Perform a treatment center data backup before replacing the
lighting system (with NLV or HLV board).
➢ Check the global software version of the treatment center after
replacing the lighting system.
If necessary, update the global software version and then restore
the data of the NLV or HLV board.
You can find information on how to perform the data backup and
the software update in the "DUMC Service Manual".
C
X1
B
L225
L225
1.
A
63 69 149 D3561
110 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
6. Press the front cover together to release the snap lock and then
remove the cover (F).
7. Unscrew the four screws (G) in the rear cover and remove the cover
(H).
HLV
X1
2.
63 69 149 D3561
D3561.076.01.08.02 10.2020 111
7 Replacing components Sirona Dental Systems GmbH
7.2 Removing boards from the swivel arm of the dentist element Sinius Service Manual
B
D
1. Release the screw (A) and open the PCB holder plate.
2. Loosen the two screws (B).
3. Loosen the grounding screw (C) and remove the grounding cable of
the NaCl pump.
4. Cut open the cable tie (D) using a pair of side cutters.
5. Pull connector X24 (E) off of the NAJm board.
6. Loosen the 4 screws and remove the NaCl pump.
63 69 149 D3561
112 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.2 Removing boards from the swivel arm of the dentist element
Removing boards
NAJm
X31
L427
NAA
X31
L254
X7
X6
NAC
HAC
X7
NAH
X2
L433
1. Pull connector X31 off of the NAJm board and the NAA board (if
present).
2. Pull connectors X6 and X7 off of the NAA board (if present).
3. Remove the NAJm board.
4. Remove the NAA board (if installed).
5. Remove the support plate.
6. Remove the NAH board (if installed).
63 69 149 D3561
D3561.076.01.08.02 10.2020 113
7 Replacing components Sirona Dental Systems GmbH
7.3 Important information when replacing boards Sinius Service Manual
7. If the NAC/HAC board (A) with heat sinks is installed, loosen the 6
metal screws and remove the board.
If the NAC/HAC board (B) without heat sinks is installed, loosen the
two plastic screws and remove the board.
Following replacement
Always check the global software using the DUMC after replacing
boards. Also update the global software if necessary.
1. Back up all data.
2. Replace the boards.
3. Perform a firmware download.
63 69 149 D3561
114 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.3 Important information when replacing boards
Turn the treatment center OFF at the power switch before replacing a
board.
Example:
63 69 149 D3561
D3561.076.01.08.02 10.2020 115
7 Replacing components Sirona Dental Systems GmbH
7.3 Important information when replacing boards Sinius Service Manual
NSK
More information can be found under: Removing the cover of the
2500 2
V5V5 60623
VV
motorized headrest on the patient chair [→ 96]-> Cable routing when
2500 2
fitting cover parts.
V6V6 40
V
4
X
0115 9 1
V2V2 20212 X
VV
S1
63 69 149 D3561
116 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.3 Important information when replacing boards
GND C
L247
NAR
KL
1
8 KL
2
1 3
2 4
5 7
6 8
IMPORTANT
While replacing the NAR board, reconnect the lines from the old to
the new board directly so that none of the lines get mixed up or
incorrectly connected.
63 69 149 D3561
D3561.076.01.08.02 10.2020 117
7 Replacing components Sirona Dental Systems GmbH
7.4 Hydraulic/pneumatic modules Sinius Service Manual
NAC
HAC
IMPORTANT
If the NaCl pump is fitted, it should be removed by unscrewing its
outer screws. This makes it easier to replace the board.
63 69 149 D3561
118 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.4 Hydraulic/pneumatic modules
Clean Setup 1.
> 2s
W27
W27
5.
W40
W40
4.
2.
3.
> 2s
The rinsing tank needs to be fully emptied before being replaced.
ü The suction hoses are connected.
1. Dentist element user interface: Hold down the “Setup“ fixed key
(>2s).
Ä The setup dialogs are displayed.
2. Hold down the "Service area" fixed key (> 2s).
Ä The service area will appear.
3. Touch the arrow key.
Ä The next screen is displayed.
4. Touch the "Empty the rinsing tank" key.
Ä The rinsing tank is emptied.
5. Wait until the flushing tank is empty.
63 69 149 D3561
D3561.076.01.08.02 10.2020 119
7 Replacing components Sirona Dental Systems GmbH
7.4 Hydraulic/pneumatic modules Sinius Service Manual
Clean Setup 1.
> 2s
W27
W27
W40
W40
4.
2.
3.
> 2s
The mixing tank needs to be fully emptied before being replaced.
ü The treatment center is switched on.
1. Hold down the "Setup" fixed key on the user interface of the dentist
element for at least 2 seconds.
Ä Six setup dialogs are offered for selection.
2. Touch the "Service area" key for at least 2 seconds.
Ä The service area will appear.
3. Use the arrow key to scroll to the next screen.
4. Touch the "Empty the mixing tank" key.
Ä The mixing tank is emptied through the tumbler outlet.
5. Wait until the mixing tank is empty.
6. Turn the treatment center OFF at the power switch.
63 69 149 D3561
120 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.5 Motors/headrest in the patient chair
MVKP1
MVKP2
A55
A56
A56
X14
A56
L447
A55
A56
A55
A55
L238
Components
Solenoid valves
MVKP1 Lumbar support inflation
MVKP2 Lumbar support deflation
Connector
MVKP1 > X14 > L447 > X10 HSA
MVKP2 > X14 > L447 > X10 HSA
63 69 149 D3561
D3561.076.01.08.02 10.2020 121
7 Replacing components Sirona Dental Systems GmbH
7.6 Solenoid value for Orotol Plus dosing (CDS60) Sinius Service Manual
NOTE
Prior to performing service work on the MV51.1 valve system, the
stop cock in the terminal box must be closed.
63 69 149 D3561
122 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 8 Service area
Sinius Service Manual 8.1 User interface in the service area
8 Service area
You can switch between the screens using the arrow keys.
63 69 149 D3561
D3561.076.01.08.02 10.2020 123
8 Service area Sirona Dental Systems GmbH
8.2 Information on the treatment center Sinius Service Manual
Back
63 69 149 D3561
124 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 8 Service area
Sinius Service Manual 8.2 Information on the treatment center
3. Burr drive
Solenoid valve MV21_3 NAJm X19 A3 ' B3
Intraoral camera
NAU board present.
Camera sends info to the PC via USB
PC sends info to the HSA via Siucom/Ethernet
HSA sends information to the NAJm/user interface
NaCl pump
Jumper on NAJm X24 4 -> 5
Ultrasound
Standard function
ENDO
Dongle on NAJm J11
NAC board present
Apex locator
NAA board present
63 69 149 D3561
D3561.076.01.08.02 10.2020 125
8 Service area Sirona Dental Systems GmbH
8.2 Information on the treatment center Sinius Service Manual
NHE Hardware, bootloader and software version for the NHE board
SRAYVIT Internal NHE communication with SRAYVIT circuit.
Intraoral camera
Camera sends info to the PC via USB
PC sends info to HSA via SIUCOM plus / Ethernet
HSA sends information to NAJm / user interface
63 69 149 D3561
126 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 8 Service area
Sinius Service Manual 8.2 Information on the treatment center
HSA Hardware, bootloader and software version for the NSA board
NSK Hardware and software version of the NSK board
NSB Hardware and software version of the NSB board
Motorized headrest NSK present. Query via CAN.
63 69 149 D3561
D3561.076.01.08.02 10.2020 127
8 Service area Sirona Dental Systems GmbH
8.3 Reading service codes Sinius Service Manual
63 69 149 D3561
128 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 8 Service area
Sinius Service Manual 8.4 Maintenance display
Code categories
I = Info
W = Warning
E = Error
U = User Info (the user is shown an error message)
63 69 149 D3561
D3561.076.01.08.02 10.2020 129
8 Service area Sirona Dental Systems GmbH
8.5 Service support with the PC Sinius Service Manual
63 69 149 D3561
130 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 8 Service area
Sinius Service Manual 8.6 PC connection, networking
63 69 149 D3561
D3561.076.01.08.02 10.2020 131
9 Mechanical components Sirona Dental Systems GmbH
9.1 Setting brakes on the support arm and dentist element Sinius Service Manual
9 Mechanical components
A B B
63 69 149 D3561
132 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 9 Mechanical components
Sinius Service Manual 9.1 Setting brakes on the support arm and dentist element
2. Adjust the braking torque of the rear (B) and front (C) pivot bearings
B by tightening or loosening the two opposite screws evenly.
3. Slide the cover back on.
2. Turn the dentist element so that the screw (B) becomes visible.
3. Adjust the braking torque of the dentist element by tightening or
loosening the screw (B).
63 69 149 D3561
D3561.076.01.08.02 10.2020 133
Sirona Dental Systems GmbH 7 Replacing components
Sinius Service Manual 7.1 Removing the cover
1. Press the front cover together to release the snap lock and then
remove the cover (G).
2. Unscrew the four screws (G) in the rear cover and remove the cover
(H).
X6 X5
NLV
X1
2.
63 69 149 D3561
D3561.076.01.08.02 10.2020 109
Sirona Dental Systems GmbH 9 Mechanical components
Sinius Service Manual 9.1 Setting brakes on the support arm and dentist element
IMPORTANT
This brake is not preset. Adjust it to the required brake force.
IMPORTANT
The brakes listed here are preset
➢ Only adjust these if this is required by the user.
63 69 149 D3561
D3561.076.01.08.02 10.2020 135
9 Mechanical components Sirona Dental Systems GmbH
9.1 Setting brakes on the support arm and dentist element Sinius Service Manual
Adjusting the brake between the support arm and lighting crown
1. Increasing brake force at position (A):
Using an Allen key turn the screw (H) inwards.
2. Reducing the brake force at position (A):
Using an Allen key turn the screw (H) outwards.
IMPORTANT
The pneumatic spring is set at the factory
➢ Only adjust the pneumatic spring on the support arm if the height
of the support arm shifts automatically or if greater brake force is
required.
63 69 149 D3561
136 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 9 Mechanical components
Sinius Service Manual 9.2 Adjusting the spring force of the flex arm on the dentist element
It is only once the dentist element has been fully equipped and the
desired configuration has been assembled that the spring force of the
dentist element flex arm can be set. This equipment includes:
● Instruments, including hoses
● Camera (optional)
● NaCl pump (optional)
● Tray (optional)
The support arm brake must be free of pressure so that the brake does
not impede its movement.
1. Switch the treatment center off.
2. Actuate the SPRAYVIT to deaerate the treatment center.
63 69 149 D3561
D3561.076.01.08.02 10.2020 137
9 Mechanical components Sirona Dental Systems GmbH
9.2 Adjusting the spring force of the flex arm on the dentist element Sinius Service Manual
3. Open the flex arm covers (C) towards the dentist element.
Ä This makes the adjustment screw (D) accessible (hexagon
socket head screw, 4 mm).
4. Adjust the spring force by tightening or loosening the adjustment
screw (D).
The actuating forces for lowering or lifting up should be roughly the
same; in case of doubt, the lowering movement should require
greater force.
5. Remount the covers removed previously in reverse order.
63 69 149 D3561
138 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 9 Mechanical components
Sinius Service Manual 9.3 Removing the swivel tray holder from the TS dentist's controls
2. Open the flex arm covers (C) towards the dentist element.
Ä This makes the adjustment screw (D) accessible (hexagon
socket head screw, 4 mm).
3. Adjust the spring force by tightening or loosening the adjustment
screw (D).
The actuating forces for lowering or lifting up should be roughly the
same; in case of doubt, the lowering movement should require
greater force.
4. Remount the covers removed previously in reverse order.
63 69 149 D3561
D3561.076.01.08.02 10.2020 139
9 Mechanical components Sirona Dental Systems GmbH
9.3 Removing the swivel tray holder from the TS dentist's controls Sinius Service Manual
2. Align the tray arm using the 3 studs (C) and a spirit level. The tray is
then locked in place.
63 69 149 D3561
140 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 9 Mechanical components
Sinius Service Manual 9.3 Removing the swivel tray holder from the TS dentist's controls
2. Turn the adjusting screw (D) in the + direction to increase the brake
force, in the - direction to reduce the brake force.
3. Reattach the cover (C).
63 69 149 D3561
D3561.076.01.08.02 10.2020 141
9 Mechanical components Sirona Dental Systems GmbH
9.4 Setting the Bowden wire in the motorized headrest Sinius Service Manual
E F
G
C
H
X1
X4
63 69 149 D3561
142 D3561.076.01.08.02 10.2020
7 Replacing components Sirona Dental Systems GmbH
7.4 Hydraulic/pneumatic modules Sinius Service Manual
NAC
HAC
IMPORTANT
If the NaCl pump is fitted, it should be removed by unscrewing its
outer screws. This makes it easier to replace the board.
63 69 149 D3561
118 D3561.076.01.08.02 10.2020
10 Biofilm removal Sirona Dental Systems GmbH
10.1 Approved resources for biofilm removal Sinius Service Manual
10 Biofilm removal
WARNING
Cleaning solutions must be mixed in accordance with the usage
instructions from the manufacturer of the cleaning agent!
10.2 Preparation
IMPORTANT
External devices must be disconnected.
WARNING
! ü
Wear gloves when performing the following work.
63 69 149 D3561
144 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 10 Biofilm removal
Sinius Service Manual 10.2 Preparation
10. Remove the filter and the sealing washer from AF1 (A).
11. Remove the filter and the sealing washer from AF6 (B) if this is
present.
B
12. Remove the filter and the sealing washer from AF (C).
13. Remove the connection box cover (see Removing the chair covers
[→ 89]).
63 69 149 D3561
D3561.076.01.08.02 10.2020 145
10 Biofilm removal Sirona Dental Systems GmbH
10.3 Execution Sinius Service Manual
IMPORTANT
With older treatment centers there may still be cardboard gaskets on
the sanitary hose. Replace these or replace the entire hose.
15. Connect the biofilm removal device (pump) to the sanitary hose.
16. Connect a hose with valve on the external device port if there is one
present.
10.3 Execution
Preparation / cleaning solution 1
WARNING
Wear safety gloves, safety goggles and mouth and nose protection
when performing the following work.
IMPORTANT
Collect the colored cleaning liquid as applicable in a suitable
container for subsequent inspection.
63 69 149 D3561
146 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 10 Biofilm removal
Sinius Service Manual 10.4 Re-processing
IMPORTANT
The solution in the collection container must be clear following the
rinsing process. If the solution is still cloudy, fill the tanks with a new
cleaning solution 1 again and complete cleaning steps once again.
The cleaning liquid can then be removed via the cuspidor.
IMPORTANT
Collect the colored cleaning liquid as applicable in a suitable
container for subsequent inspection.
IMPORTANT
The solution in the collection container must be clear following the
rinsing process. If the solution is still cloudy, fill the tanks with a new
cleaning solution 2 again and complete cleaning steps once again.
The cleaning liquid can then be removed via the cuspidor.
10.4 Re-processing
1. Reduce the water pressure in the treatment center, by removing the
Sprayvit and operating the water button until no more water
emerges.
2. Switch the treatment center off.
3. Reduce the air pressure in the treatment center by operating the air
button.
4. Remove the biofilm removal device.
63 69 149 D3561
D3561.076.01.08.02 10.2020 147
10 Biofilm removal Sirona Dental Systems GmbH
10.4 Re-processing Sinius Service Manual
WARNING
Drops of the cleaning solution contain acids / alkalines and may affect
surrounding parts.
Carefully loosen the adapter from the sanitary hose.
63 69 149 D3561
148 D3561.076.01.08.02 10.2020
Sirona Dental Systems GmbH 10 Biofilm removal
Sinius Service Manual 10.4 Re-processing
Sanitation procedure
➢ Complete the sanitation procedure as described in the operating
instructions.
IMPORTANT
The biofilm removal is now complete.
63 69 149 D3561
D3561.076.01.08.02 10.2020 149
We reserve the right to make any alterations which may be required due to technical improvements.