Professional Documents
Culture Documents
vela XL
Service and Maintenance Manual
StarMedTec GmbH reserves all rights for technical changes or Improvements to this document
®
and/or the vela XL without given prior notice.
®
This manual is part of Instruction for use. Don’t start working if the IFU vela XL is not
available. Please contact StarMedTec GmbH for a free copy of the IFU.
1 Introduction ............................................................................................................7
1.1 Working with the service and maintenance manual ..............................................................7
1.2 Structure of this manual............................................................................................................8
1.3 General safety information......................................................................................................10
2 Electrical safety ....................................................................................................11
3 Laser safety ..........................................................................................................12
3.1 Eye protection ..........................................................................................................................12
3.2 Marking the laser area .............................................................................................................13
4 Working safety and warnings..............................................................................14
5 Restrictions for technical support ......................................................................15
6 Technical Inspections ..........................................................................................16
6.1 Visual Inspection of Device and Accessories.......................................................................17
6.2 Inspection of System Functioning .........................................................................................17
6.3 Inspection of Laser Parameters .............................................................................................17
6.4 Inspection of Bleeder Resistance ..........................................................................................18
6.5 Inspection of Earth - and Enclosure Leakage Current.........................................................18
7 Technical specifications ......................................................................................19
7.1 Device specifications...............................................................................................................19
7.2 Fiber parameters ......................................................................................................................20
7.3 Electromagnetic compatibility of power cable and foot switch cable................................20
7.4 Product labels ..........................................................................................................................21
8 Technical requirements and start-up..................................................................24
8.1 Technical and spatial requirements.......................................................................................24
8.2 Start-up installation and precautions ....................................................................................25
9 Mechanics and components overview ...............................................................27
9.1 Housing .....................................................................................................................................27
9.1.1 Components overview ........................................................................................................27
9.2 Opening of the housing...........................................................................................................28
9.3 Front cover with display..........................................................................................................32
9.3.1 Components overview ........................................................................................................32
9.4 Rack...........................................................................................................................................33
9.4.1 Components overview ........................................................................................................33
10 Block diagram ...................................................................................................34
11 Electronic components, overview and exchange ..........................................35
11.1 Toroidal core isolation transformer .......................................................................................35
11.2 Transformer trip relay..............................................................................................................36
11.3 24V Power supply ....................................................................................................................37
11.4 Loudspeaker .............................................................................................................................38
11.5 Fan optic bench........................................................................................................................39
11.6 Main board ................................................................................................................................40
11.7 Module adapter.........................................................................................................................41
11.8 Display, Slave board, and Touch panel .................................................................................42
11.8.1 Display board components ............................................................................................44
11.9 Footswitch ................................................................................................................................46
12 Laser Module.....................................................................................................47
12.1 Installation ................................................................................................................................47
13 Optics.................................................................................................................57
13.1 Optical Components ................................................................................................................58
Service and Maintenance Manual Draft MOAP.doc page 4 of 101
13.1.1 Components Overview ...................................................................................................58
13.2 The incoupling block ...............................................................................................................59
13.3 Adjustment of the z-distance (Figure 10) ..............................................................................64
13.4 The Beam Combining Block ...................................................................................................65
13.5 The Pilot Laser Module............................................................................................................66
13.6 The Power Shutter ...................................................................................................................67
13.7 The Power Measuring Unit ......................................................................................................70
13.8 Mounting the collimator ..........................................................................................................71
13.9 Near field adjustment of pilot beam and working beam ......................................................74
13.10 Far field adjustment of pilot beam and working beam ....................................................76
13.11 Mounting and adjustment of the incoupling.....................................................................77
®
Energy measurement of vela XL........................................................................................................79
14 Cooling Unit ......................................................................................................84
15 Packing for Transport.......................................................................................92
16 Main Menu .........................................................................................................95
16.1 Configuration menu (set-up)...................................................................................................96
16.2 Submenus .................................................................................................................................97
16.2.1 Call-up device information .............................................................................................97
16.2.2 Open error history ...........................................................................................................97
16.2.3 Access to service level (only possible for authorised service personnel) ...............98
17 Troubleshooting and Maintenance..................................................................98
17.1 Rectifying faults .......................................................................................................................98
17.2 Error messages ........................................................................................................................99
17.3 Technical customer service ..................................................................................................101
All rights reserved. Copies or photocopies of the service manual or IFU or parts thereof or conversion into
electronic media or machine-readable formats are not permitted without the prior written authorization of
StarMedTec GmbH.
Please note:
®
The vela XL is a surgical laser for use in many different disciplines, such as general surgery, dermatology,
endoscopy, gastroenterology, gynaecology, ENT, laparoscopy, orthopaedics, paediatrics, pneumology,
urology, thoracic surgery and dentistry.
®
The vela XL is used in combination with the approved application fibers which can be used directly or with
many different application instruments, such as appropriate endoscopes or hand pieces.
Indications are dissection, ablation and coagulation of soft and hard tissue.
Please contact StarMedTec GmbH for information on applications or on our product line.
®
The vela XL emits laser beams with a wavelength of approx. 1.9 µm.
This special wavelength is transmitted by an optical fiber (application fiber) and, together with the
application-related parameters, enables efficient treatment with minimum stress to the surrounding tissue.
Various optical fibers and applicators are available for this laser system which allows to be used for many
different applications. The accessories and the types of fiber are continuously extended and refined. If you
have any questions, contact StarMedTec GmbH.
CAUTION
®
Please refer to the Instruction of use for more detailed information “how to use the vela XL”.
StarMedTec GmbH
Kreuzstraße 22
82319 Starnberg, Germany
Hotline
Please don’t start to solve difficulties by try and error techniques. Refer to the training for an effective way to
start, localize and solve troubles or problems. As a reminder the function of all major parts is described first,
second , you are given information how to work effectively on the unit.
Always take extra care for your safety as described in the “safety chapter”. Don’t start working on the unit
with not the necessary right tools or safety equipment on the hand.
Typographical Conventions
WARNING
CAUTION
Precautions alert the reader to exercise special care necessary for the safe and effective use
of the device.
NOTE
CAUTION
Use of controls or adjustments or performance of procedures other than those specified herein
may result in hazardous radiation exposure.
WARNING
CAUTION
®
Safety first, please read the safety chapters carefully before starting any work on the vela XL.
Never start working if the adequate laser safety glasses are not on hand!
CAUTION
Only StarMedTec GmbH Service Technicians or Service Technicians who have been
®
specifically trained and authorized by StarMedTec GmbH are allowed to work on the vela XL
or carry out regular technical inspections or to do service and repair.
Servicing or any kind of manipulation of the system, by non-authorized personnel will result in
a termination of the warranty and any liability on the parts of StarMedTec GmbH.
It is essential for authorized technicians to have valid certificate after CE regulations to perform
®
any technical work on the vela XL.
NOTE
Electrical equipment or electrical system installation, modification, testing, repair, and maintenance shall be
done by QUALIFIED AND AUTHORIZED electrical personnel only.
®
Only trained persons in electrical safety precautions should perform service works on the vela XL if the
housing is removed and the unit is plugged to the electrical wire system!
Please check with local authorities for national needs, certificates and training courses.
Warning
It is essential to use an earth leakage circuit breaker /RCD (Residual Current Device), release
®
current 30mA, for all electrical works on the vela XL. It is necessary for some works on the
®
vela XL to connect the system to the electrical wire system.
- Protect everyone involved and the person carrying out the laser treatment against unintended
exposure to radiation
- Protect objects in the operation area of the laser against damage, fire, explosion, etc.
CAUTION
Everyone must wear laser protection glasses or suitable eye protection in the laser area. If this
is not adhered to, irreversible eye damage may occur! Before using laser protection glasses
make sure that they are not damaged; in other words, the glasses must not have any
mechanical damage. The type described below must be used.
®
Laser protection glasses for the vela XL must fulfil the following conditions:
Protection level
CAUTION
CAUTION
The above regulations are German regulations. If the laser device is to be operated outside
Germany, the applicable national or international regulations must be complied with.
NOTE
Mark the laser area with notes on the used wavelength to avoid using wrong safety glasses. It
is recommended to provide the correct safety glasses outside the laser safety area for visitors.
Only people who are trained in laser safety, service and maintenance handlings and who have knowledge on
®
the product, should work on the vela XL.
National requirements for the setting up and operation of medical products and medical lasers must be
complied with. The operator must clarify whether the national authorities have to be notified about the
operation of laser equipment.
®
The vela XL is an active medical product in Class IIb according to Medical Device Directive 93/42/EEC.
®
If used improperly, every laser can endanger health. The vela XL is a Class 4 laser.
Do not direct the laser beam onto reflective surfaces (metal surfaces, mirrors, etc.).
Remember that reflective surfaces or instruments can divert the laser beam in an uncontrolled manner.
Particular care must be taken with glass surfaces, but highly polished metal surfaces a few metres away can,
under certain circumstances, also cause dangerous scattered laser radiation.
During laser emission the prescribed protective glasses must be worn in the laser area.
If you are using optical equipment for observation or adjustment, you may have to use suitable protective
filters with the correct safety level.
During operation the laser area must be isolated and marked according to Safety Standard EN/IEC 60825.
When handling all accessories, disposable items or materials that were in contact with the patient, it is
important to take protective measures to prevent the transfer of germs. See Sections in the IFU chapter
Fehler! Verweisquelle konnte nicht gefunden werden. “Fehler! Verweisquelle konnte nicht gefunden
werden.” (device) and Fehler! Verweisquelle konnte nicht gefunden werden. "cleaning” (accessories).
For the treatment use only undamaged application fibers that have been sterilised according to the
instructions. Observe the specified cleaning and sterilisation cycles. Do not use damaged or dirty application
fibers. The maximum utalisation time for the application fibers for each sterilisation cycle is limited to 150
minutes.
If the laser is used in environments with enriched oxygen or with highly flammable gases or materials, there
is a risk of fire or explosion. Take the appropriate protective measures.
It is recommended that the laser is not used in environments with enriched oxygen.
®
Use only the "single use" or "reusable" application fibers approved by StarMedTec GmbH with the vela XL.
In some cases, the application fibers are subject to country-specific restrictions and, because of this, are not
approved or available everywhere.
®
To prevent damage to health, move the vela XL with at least two people; do not lift the device, not even
with the front handle.
®
Before transporting the vela XL dismantle all accessories (application fibers, power cable, foot switch).
The device should be transported with as little vibration as possible.
The life of the medical product is limited to 10 years. After this time, the adequacy of the technology must be
re-evaluated by StarMedTec GmbH.
®
For further information about use of the vela XL device, please refer to the IFU.
Note
®
The vela XL laser device has been tested according to EN / IEC 60601-1-2 (EMV). For more
information, see IFU Section 13.3 “Electromagnetic compatibility”.
CAUTION
The deionized water must be replaced at least once a year. The exchange of these parts is
part of the safety check up.
The laser parameters have to be checked. Please refer to the service maintenance protocol for the
values of the vela® XL.
Please use different fiber sizes as marked in the maintenance protocol.
NOTE
The use of different fiber sizes and check on the appropriate power levels give you hints for
proper alignment on incoupling components! Please fill in measured levels and keep the copy
for later reference.
Protective Conductor
Minimum output 5W
NOTE
®
Only approved fibers for the vela XL can be used with the device. Don’t try to install any other
fiber as this can damage the incoupling lens.
Minimal fiber deflection should not be longer than absolute necessary to avoid damage on the
coating and burn offs. Storage of fibers must be not smaller than a radius of 160mm.
NOTE
Electrical medical equipment is subject to special precautionary measures regarding EMC and
must be installed and operated according to these guidelines.
The following cables must be used to comply with interference emission and interference immunity.
Cable Length
Power cable 16 A 5m
Foot switch with cable 3m
Cable overview
CAUTION
Use of different cables and accessories may cause higher interference emission or reduced
interference immunity. The use of longer cables is not approved by StarMedTec GmbH.
If cables must be exchanged, please use same type, length and size.
NOTE
The laser warning sign on the back side of the laser device provides information about the laser class,
wavelength, rating, pulse duration and frequency of the laser beam so that appropriate safety
measures can be taken.
50 mm
ACCORDING TO EN 60825-1:2007
LASER
LASER CLASS 4 Po ≤ 150 W
λ = 1,9 µm
PILOT
LASER CLASS 2 Po < 1 mW
λ = 0,5 µm
52 mm
108
Warning and information signs on the laser device
Information sign on the back side of the laser device: foot switch & door interlock
The information sign marks the connections for the foot switch and the door interlock.
Type label
The electrical installation of the room where the vela® XL is operated must fulfil national requirements
and specifications. The sockets must be installed with fixed wiring.
The fusing and the supply line for the vela® XL must be separate. Fusing with a circuit breaker Type
C16 A or higher.
Place the vela® XL only on a suitable nonslip base that can bear the weight of the device (see Section
13.1 “Device specifications”).
It is recommended that the vela® XL is used only in rooms with an ambient temperature between 18°C
and 30°C and relative humidity ≤ 75 %, non-condensing. The maximum permissible ambient
temperature of 30°C must not be exceeded.
To ensure adequate cooling, the inflow and outflow of air at the sides must not be blocked. To ensure
this, if possible, install the vela® XL freestanding. Please maintain free space of at least 40 cm beside
the laser in order to ensure proper air flow.
Ensure that the treatment room is ventilated adequately. Because the laser emits heat, ventilation of at
least 200 m³/h is recommended.
Service and Maintenance Manual Draft MOAP.doc page 24 of 101
CAUTION
Condensation on the laser device during operation may damage the device. It is important that
you allow the cool device to acclimatise at the prescribed ambient conditions in the treatment
room for several hours before use.
CAUTION
If the laser device is used improperly, it can cause irreversible damage. That is why the laser
device may be operated only by appropriately trained and experienced staff. Make sure that
stuff gets a technical advice “how to use the vela® XL”
Before using the laser device, user must ensure that the device is functionally safe and in proper
condition.
ATTENTION
Insert the plug into a fixed socket with an earth wire contact (exclusive and with at least 16 A
fuse). Do not use a multiple socket.
Install the required application fiber in the laser system. Observe the associated instructions for use
and the sterilisation conditions. See Section appropriate in the IFU and the instructions for the use of
the application fiber.
For the treatment use only undamaged application fibers that have been sterilised according to
the instructions. Observe the specified cleaning and sterilisation cycles. Do not use damaged
or dirty application fibers.
Never use any service fibers for the treatment of patient under any circumstances!
NOTE
Be careful with placement of power cable, foot switch, door interlock and application fibers to
avoid risk of damage or accidents in walkthroughs or risks in any other way.
9.1 Housing
9.1.1 Components overview
K
C D
A
M
P
N
Q
B
O
R
S
I
T
ATTENTION
Never lift machine at handle bar! The front cover can break of the heavy weight!
Use handle bar for pulling only, not for lifting!
ATTENTION
Always shut down the unit with the main switch (backside) and disconnect the power cord
before opening the housing.
To do any maintenance or repair on the system, it is required to open the housing of the unit. Take
care of the following steps to disassemble the single parts of the housing.
ATTENTION
It is important to open the housing in the direction explained in this chapter, otherwise the
display can be damaged or fall off from the unit.
Start by removing the four screws (A and B) on the backside of the unit.
(A hex key 4mm, B hex key 3mm)
B
F
C
A
NOTE
Remove display completely if you work alone on the unit. Sometime some force is necessary
to take off the side panels, use both hands to lift up! Don’t hang up the display on cables as
this will damage pins and wires! See display chapter for further information how the display is
wired up!
F G
ATTENTION
Disconnect the ground cables before removing the panels. Make sure to connect each
grounding cable back to place while reassembling the unit!
A B E
F G
C
C
D
D
F
I
Display disconnection:
- Remove the plastic connector (C) and (D) from the key switch and emergency button
A
A
B
C
C
A
A G
- Disconnect transformer wires (black wires connected to 2 and 3, Signal wires S1 and S2, output
wires red connected to post terminal) from transformer trip relay and binding post
- Remove the screws (A) of the mount plate. Like in the picture below and on the opposite side (at
all 6 screws)
4
B B
- Unscrew the transformer from the mounting plate (screw B). and take out transformer
- Put in the new transformer and do the steps in the opposite sequence to reassemble it
The transformer trip relay is used to kill high current peaks, which are produced by the toroidal core
isolation transformer.
For an exchange it is necessary to follow the next steps:
- Remove all wires from the relay, 2 black wires going to the transformer on post point 2 and 3, the
brown wire to point 1 and the blue wire on 4. The smaller cables are connected to S1 and S2
- Remove all clamp screws (A, B)
- On the new relay, adjust potentiometer (C) to “R” at “9 o’clock”
- Clamp the new relay with both of the screws
- Connect all wires like on the original relay
A
C
C
The 24V Power supply is used to supply components like the heat exchanger, the module adapter, the
mainboard and all other 24V appliances.
In case of a malfunction, exchange the 24V power supply like shown below
- Remove the air chute. The air chute is fixed with 5 screws (A,B,C,D,E)
- Disconnect all wires from the 24V power supply
ATTENTION
Make sure to check for markers or to mark all wires before anything is disconnected.
A B
H I
D
K L
F G
A warning tone comes up while Laser emission is on. You will also get a single tone for each touch on
the display by an external display speaker (not exchangeable).
To exchange the main speaker follow these steps:
- Disconnect the wires (A,B) from the speaker, remember plus and minus connector
- Remove the four screws (C,D,E,F), and take out the speaker
- Build in new speaker like the original was mounted in reverse steps
C D
A
E F
ATTENTION
It is not possible to do any repair work on the main bord pcb in field service. Please exchange
part if necessary and return the defective part to StarMedTec.
J3 Fiber identification
J17 Footswitch
„J2 Service Interface 2“
J18 Doorinterlock
J19 Speaker
J1 Slaveboard
J20 Audio in
E F
J2 Signals to Mainboard J2
J3 Signals to Mainboard J10
J9 24V Slave
J5 Engaging circuit
J1 24V
J6 Fan
NOTE
Make sure to locate each cable set point before anything is disconnected! Some plugs can be
connected wrongly or ribbon cable twisted over to the opposite side. Damages may occur to
display by wrong installation. Take care while removing the small plastic clips, connection
points with damaged clips will not work!
The display CPU module does fit only in the shown configuration, check before assembling!
1 2
3 I 4
C D
5 6
8
7
NOTE
In the case the display or parts of the display unit must be exchanged, please note serial no. of
boards after assembling! To make sure the lifetime file is up to date; all serial no. of
exchanged parts should be sent to StarMedTec.
To locate errors on the display, there are 3 LED for voltage status
- D20 bright if 12 V are available on the board
- D21 bright if 5 V are available on the board
- D22 bright if 3,3 V are available on the board
Display CPU module not assembled or wrong assembled, D21 bright, D20 and D22 off. Connection
point X 17 assembled wrongly, D 21 and D22 bright, D20 off.
D20 X18
12V
D21
X3 5V
X7
Mini USB
programming
interface
A A
A A
NOTE
The laser module unit was designed to work within the specific lifetime hours with no service at
®
all. Please return the vela XL to StarMedTec for exchange or rebuilt of the laser module. The
following steps are explained just in case of circumstances making necessary that the unit
must be installed on the spot!
12.1 Installation
- You must divide the laser rack in 2 pieces before the laser module can be installed. Take care
with the upper piece to avoid damages to sensors and cables
Back side
Store on i.e. a
chair or box beside
®
- Mainboard of the vela XL (C)
WARNING
The weight of laser source box is approx. 50kg. Do not lift the box as a single person. Use
lifting belts turned around the box, lift with at least 2 persons! Never use the water in-/outtakes
or electrical connectors to lift the unit!
WARNING
Place optical fiber safe and secure to avoid kinks, fall down or other damages. Any damage
results in return to StarMedTec for inspection and repair.
WARNING
Do not place the laser module too far to the rear as the unit can’t be moved in the tray.
Too much weight on the back can force the system to get instable or turn over.
- back side after the module (E) was placed with correct front end
- Check if foam bumper is still installed to avoid swinging noise during operations. Replace if
necessary. The foam bumper is glued to the tray (D), (G) shows the cooling unit
- Connect upper rack with frame by using stainless steel screws (J)
CAUTION
Put all screws in place; tighten all screws and nuts to avoid movement of the laser module. Do
not overtighten screws or nuts. The rack/frame system is the major component for stabilization
of the complete laser, take special care with any work performed.
Remove the protection cover by loosening three socket head cap screws.
Pilot Laser
Module Optical Tray
Fig. 2 Optical bench - components overview
Collimator Mount
Loosen the
socket head cap
screws (4x)
After disassembling the RFID- fiber identification and the angle plate (not shown here) dismount the
incoupling block by loosening the four socket head cap screws (3mm key).
SMA- Adapter
Auxiliary Tool
If the tool does not slide in easily, manipulate the micrometer screws in order to centre the carrier
slide.
Grasp the auxiliary tool and loosen the locking ring. Screw the brass lens mount out of the carrier
slide. Then you are able to disassemble the lens by loosening the inner locking ring.
The new lens has to be cleaned and the components are reassembled again.
Coupling Block
Beam Combining
Fig. 12 BC-block
The coupling block is fixed with three socket head cap screws from the downside of the optical tray.
Set Screw
If the pilot laser needs to be changed, please loosen the set screw at the side of the pilot laser mount
and retract the old one. The new one is installed vice versa.
Shutter Housing
Shutter PCB
Detector Flag
Please take care of the shutter flag being closed properly as shown in figure 16.
Deflection Optic
It is important that the deflecting optic is clean and that the correct mounting direction is maintained.
By rotating the deflection optic, the amount of laser radiation entering the measuring spheres can be
influenced. Generally it should be set to maximum.
Beam Entry of
Measuring Sphere
Power Measuring
Mount
PCB
Potentiometers for
Adjustment of
Amplification
Cable
The power measuring unit consists of a measuring sphere in which the laser radiation is equally
distributed and a power measuring PCB with potentiometers for the adjustment of the amplification on
the front side and two detectors on the back side.
If the PCB needs to be changed, please ensure the proper fitting of the Sensors within the appropriate
holes in the measuring spheres as shown in figure 20.
Mount the collimator centric in the collimator mount as shown in figure 22.
After diligently mounting the collimator, please check the centricity of the laser beam. Therefore set
the power level to the lowest possible setting and release laser radiation. Check the centricity of the
laser beam with thermo-sensitive paper at the entry and the back side of the power shutter.
Fig. 24 Centricity check of laser beam at the entry of the power shutter
The next step needs the use of the reticule auxiliary tool.
The reticule is mounted at the front end of the BC-Block. The incoupling block must be disassembled
first.
Attach a sheet of thermo sensitive paper to the plain front end of the reticule auxiliary tool.
On the thermo sensitive paper you see the pattern of the pilot beam. Adjust the position of the near
field of the pilot beam by manipulating the black base plate as shown in figure 27.
After achieving a proper adjustment of the near field setting of the pilot beam, release laser radiation
(lowest setting possible) in order to check the centricity relative to the reticule.
The position of the working beam is adjusted by manipulating the fine thread screws of the collimator
mount. After the first shot take a new sheet of thermo sensitive paper. A proper alignment is shown in
the right hand side picture above. For the following procedures please leave the reticule in place but
remove the sheet of thermo sensitive paper
Place a sheet of plain thermo sensitive paper in a distance of approx. one meter from the BC-block.
Release laser radiation in order to create a marking of the target area of the working beam. Therefore
the output power must be increased compared to the previous procedure.
Then align the pilot beam onto the mark of the working beam by manipulating the gimbal mount of the
pilot laser module as shown in figure 30.
Gimbal mount
First, attach a small piece of adhesive tape on the front thread of the SMA-adapter in order to get a
screen to observe the pattern of the pilot beam. Second, move the loose incoupling block until the
green spot appears in the center of the SMA-adapter. Do not manipulate the micrometer screws for
that purpose! Then fasten the screws.
A diligent performance of the near- and farfield settings combined with a proper pre alignment of the
incoupling block ensures the possibility to carry on the adjustment of the incoupling.
The fine tuning of the incoupling is done by the calculation method. Therefore you need a 600µm test
fiber and your power measuring equipment. Connect the test fiber to the SMA-adapter and release
laser radiation (lowest possible power setting). Adjust the incoupling as described below.
Fiber
Lens
The focus lens pictures a small spot on the proximal surface of the fiber. Find the upper edge of the
fiber by turning the micrometer screw clockwise (green) while you release laser radiation. The reading
of the power meter will drop instantly if you reach the edge. Please note the reading of the micrometer
screw. Turn the micrometer screw anticlockwise (red) until you find the lower edge of the fiber surface.
The instant drop of the power reading will indicate that. Stop emitting laser radiation and note the
reading of the micrometer screw. Now calculate the arithmetic mean of the two readings. Turn the
micrometer screw clockwise again until you reach the calculated number.
- repeat the procedure with the second micrometer screw
- change the test fiber to a 230µm one
- repeat the procedure to ensure a proper coupling efficiency with fibers of small diameter
3 – 4cm
fig.1 fig.2
Always start with the 600µm fiber. Connect the fiber to the system.
®
Choose an energy value on the vela XL Display (for example 5W), press Laser-Ready button.
When “Ready” (fig.4) appears in the display press the footswitch for at least 5sec., until the energy
on the display gets stable! Release footswitch and write down the energy which is shown in the
USBI software. (fig.5)
fig.4 fig.5
If you can’t reach the laser energy which you should reach, check if the collimator and the focus lens
are still aligned.
Note:
Any measurement head has a specific tolerance of max. +/-20%. So it is possible that you will get
an energy value which may not be the value you have expected.
For example, if you choose 30W on the laser display, the measurement display will show an
external energy between 24W and 36W.
A E
A
C
B D
After filling up the tank the first time, it is needed to run the heat exchanger must run in “standalone mode”.
Therefore we need the bridge for the heat exchanger (Fig. 4). Put it into “X1” of the signal plugs.
Switch on main switch on the backside of the unit. The heat exchanger now should be “on” but not running.
Put the loosen cable from the bridge plug into “pin2” of the plug for about 2-3 seconds. The heat exchanger
should start running. Now the water level in the filler neck will sink. Fill in water to the tank till it is between
Min. and Max. level again. Let the heat exchanger run for about 10 minutes and check the water level
several times. After the last refill, please let the heat exchanger run for another 10 minutes to get sure that
there are no more air bubbles in the cooling circuit.
In order to replace the heat exchanger first we have to drain out all the water from the system. When the
water tank is empty the first time, first you have to unplug the water inlet from the laser module. Now the
water from the laser module can drain out to the water tank. Also drain out the rest of the water from the
water tank. Now you can also unplug the water outlet (be careful, there could be still some water coming
out).
A C
Remove holding pin (E) for side panel and remove screws (F), (G), (H), (I) and (K).
F G
H I K
After replacing the heat exchanger it is necessary to configure the temperature Limit comparators (LC)
of the heat exchanger.
Service and Maintenance Manual Draft MOAP.doc page 87 of 101
INFORMATION
The particle filter is installed in the water outlet hose of the chiller. The filter can be replaced as follows:
ATTENTION
ATTENTION
®
Never lift on the vela XL handle bar or metal cavity!
Place footswitch
here
- Check if all sides are protected with foam bumpers against movement or bounce
NOTE
Don’t pack any loose items in the box! During transport loose items may damage painting,
wheels or display. Use adequate cardboard for any equipment sent or packed with the laser.
When the device is switched on, the StarMedTec GmbH logo is displayed during the system test.
Note
The reduction of the heat input into the tissue is generated by reduction of the maximum
power permitted and a special modulation of the laser.
Touch the configuration icon in the main menu to open the configuration menu (set-up).
There, you can open the following basic information and settings:
On this page you can call-up data and information: Serial number, software version, hardware
version, total output of the laser source, remaining life until the laser source has to be replaced
On this page you can read error codes and error-relevant data.
Faults are either displayed on the device display or the display turns blank when the device switches
off automatically.
If no message is displayed, the difficulty could be due to one of the causes listed below:
Error # Description
1001 Life time laser-source almost reached
1002 Water temperature high warning
1003 Device temperature too high
1006 Information about the applicator’s remaining reusability
1007 Information about last usage of applicator
1008 Water Level Warning
1010 Fiber not connected.
1101 Information: Applicator can’t be used anymore.
1103 Fiber identification failed
1104 Release foot switch
1105 Check door interlock
1106 Data of fiber identification
1107 Login fiber identification
1110 Time limit for fiber usage
1203 Device temperature too low
1204 Water temperature high error
1205 Water temperature low warning
1206 Water temperature low warning after error
1301 Laser power too low
1302 Laser power too high
1303 Mean laser power too low
1304 Mean laser power too high
1305 Shutter out of order
1306 Shutter open
1307 Shutter closed
1308 Power measurement out of order, channel deviation
1309 Power measurement out of order, offset
1310 Water temperature sensor out of order
1311 Water temperature low error
1313 Life time of laser-source reached
1314 Timeout fiber identification
1315 Touch panel out of order
1316 Device over temperature
1317 Air temperature sensor out of order
1318 Flow out of valid Range
1319 Cooling defect
1404 “Main power supply on” signal defect
1405 “Emission on” signal defect
1501 MIPS Dataformat
StarMedTec GmbH offers an extensive warranty and support program to ensure that your laser device
continues performing reliable for many years. This program includes our service hotline. Highly qualified staff
in the service team will quickly help you resolving any difficulties you may encounter. If necessary, you can
obtain information how to use our repair service.
Our foreign customers should first contact the service hotline of their local dealer.
StarMedTec GmbH
Kreuzstraße 22
82319 Starnberg, Germany
Hotline