You are on page 1of 37

OrthospecTM

SERVICE MANUAL 

Revision A01
Orthospec Service Manual OR‐3‐X405 
 
 
 
 
 
 
 
 

This manual is the property of MEDISPEC LTD. and may not be transferred or reproduced in 
any form without the written permission of the company.   This manual is produced under 
the copyright laws of the USA. 
 
The  treatment  procedures  and  clinical  practices  specified  in  this  manual  should  be 
considered  as  recommendations  only  and  must  be  determined  and  implemented  by  the 
physicians supervising the treatment. 
 
 
This manual may be amended by the company without prior notice at any time. 
 

Warning:  Product  liability  claims,  warranties  as  well  as  guarantees  made  by 
Medispec  with  respect  to  the  product  are  voided,  if  it  is  not  used,  serviced  or 
maintained in accordance with the instructions in this manual.  

 
  Yigal Hahamov
I am approving this document
Nov 2 2015 9:34 AM Page 2 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

Table of Contents
1 INTRODUCTION……………………………………………………………………………………………………………………….5
2 UNPACKING……………………………………………………………………………………………………………………….. .... 5
3 PARTS LIST.........………………………………………………………………………………………………………………………6
4 BEFORE STARTING TO SERVICE ....................................................................................................... 6
5 TOOLS & EQUIPMENT ..................................................................................................................... 6
6 INITIAL INSTALLATION..................................................................................................................... 7
7 TECHNICIAN SCREEN ....................................................................................................................... 7
8 INITIAL CONFIGURATION – GENERAL TAB ...................................................................................... 8
9 COUNTERS TAB ………………………………………………………………………………………………………………………9
10 CONFIGURE TAB........................................................................................................................9
10.1 GENERAL1 TAB .......................................................................................................................... 10
10.2 WATER TAB – SERVICE DOOR AND SENSOR ...................................................................................... 11
10.3 SERVICE DOOR SENSORS SYSTEM .................................................................................................... 12
11 MCU TAB 13
12 TROUBLESHOOTING ...................................................................................................................... 14
13 SYSTEM TESTS AND REPAIR PROCEDURES ................................................................................... 18
13.1 HVPS INPUT  VOLTAGE TEST ...................................................................................................... 18
13.2 CONTROL PCB  VOLTAGE TEST ...................................................................................................... 19
13.3 THERMAL SWITCH TEST ............................................................................................................... 19
13.4 HIGH VOLTAGE CALIBRATION ........................................................................................................ 20
13.5 CALIBRATION TEST ...................................................................................................................... 21
13.6 DELIVERING SHOCKWAVES VIA TECHNICIAN SCREEN .......................................................................... 21
13.7 SPARK‐GAP RECOVERY ................................................................................................................. 22
13.8 OVERRIDE PROCEDURE ................................................................................................................ 22
14 COMPONENTS REPLACEMENT GUIDE........................................................................................... 23
14.1 OVERVIEW ................................................................................................................................ 23
14.2 SYSTEM ASSEMBLY DESCRIPTION .................................................................................................... 23
14.3 EXTERNAL FUSES REPLACEMENT .................................................................................................... 26
14.4 COVERS REMOVAL ...................................................................................................................... 27
14.5 INTERNAL FUSES REPLACEMENT .................................................................................................... 27
14.6 HV POWER SUPPLY (HVPS) REPLACEMENT .................................................................................... 28
14.7 HVG REPLACEMENT ................................................................................................................... 29
14.8 CONTROL UNIT REPLACEMENT ...................................................................................................... 30
14.9 DOOR ACTUATOR REPLACEMENT ................................................................................................... 31
14.10 WATER SENSOR REPLACEMENT ..................................................................................................... 32
14.11 VALVE ACTUATOR REPLACEMENT .................................................................................................. 33
15 PREVENTIVE AND MAINTENANCE PROCEDURES ......................................................................... 34
15.1 ROUTINE USER MAINTENANCE PROCEDURES ..................................................................................... 34
15.2 PERIODICAL SERVICE PROCEDURES .................................................................................................. 34
16 APPENDIX A – BLOCK DIAGRAM ................................................................................................... 35
17 APPENDIX B – SPARE PARTS LIST .................................................................................................. 36
18 PRODUCT CONTACT INFORMATION ............................................................................................. 37
 
 
Page 3 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

 
List of Figures 
 
 
Figure 2‐1: Unpacking the Orthospec............................................................................... 5
Figure 7‐1: Technician Screen ..........................................................................................7
Figure 8‐1: General Tab ...................................................................................................8
Figure 9‐1 Devices counters TAB ......................................................................................9
Figure 10‐1: Configure tab ...............................................................................................9
Figure 10‐2: General1 Tab for  Advanced configuration .................................................... 10
Figure 10‐3: Water Tab.................................................................................................. 11
Figure 10‐4: Limit Switches ............................................................................................ 12
Figure 11‐1: MCU TAB ................................................................................................... 13
Figure 13‐1 HVPS INPUT  Voltage Test ............................................................................ 18
Figure 13‐2: Control PCB (Control Unit)........................................................................... 19
Figure 13‐3: thermal switch element .............................................................................. 20
Figure 13‐4: MCU/KV TAB .............................................................................................. 20
Figure 13‐5:  Spark‐gap recovery .................................................................................... 22
Figure 13‐6: Override in the MCU screen ........................................................................ 22
Figure 14‐1: Internal View – Right Side ........................................................................... 23
Figure 14‐2: Internal View ‐ Left Side .............................................................................. 24
Figure 14‐3: Internal View – Back Side ............................................................................ 25
Figure 14‐4: Water System ............................................................................................ 26
Figure 14‐5: Fuse Replacement ...................................................................................... 26
Figure 14‐6: Orthospec covers disassembly ..................................................................... 27
Figure 14‐ 7 Internal Fuse replacement ........................................................................... 27
Figure 14‐8 HVPS replacement ....................................................................................... 28
Figure 14‐9 dismantling the HV power supply ................................................................. 28
Figure 14‐10: HVG Replacement .................................................................................... 29
Figure 14‐11: Control Unit  Replacement ........................................................................ 30
Figure 14‐12 Door Actuator replacement ........................................................................ 31
Figure 14‐13 Water Sensor replacement ......................................................................... 32
Figure 14‐14 water sensor controller on the Main terminal .............................................. 32
Figure 14‐15  Valve Actuator Replacement...................................................................... 33
   

 
 
Page 4 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

1 INTRODUCTION 
The Omnispec is designed to deliver extracorporeal, focused, low‐intensity shockwave 
therapy.  
 
The  device  should  be  used  only  by  qualified  and  trained  personnel  under 
supervision of the physician trained and/or experienced in the use of this device 
  as outlined in an appropriate training program. 
 
This device should be serviced by authorized personnel. Do not attempt to repair 
or service this device yourself. 
 
   
 Do not put the device in a way that will make it difficult to disconnect it from 
the Mains electric outlet.  
 To avoid risk of electric shock, this equipment must only be connected to 
 
supply mains with protective earth.
 

Keep this manual close at hand for reference  at all times 
 
2 UNPACKING 
The OrthospecTM system is supplied in a wooden box (Figure 2‐1). 
To unpack follow the instructions below:  
1. Release the screws holding the Black Cover Lockers.  
2. Remove the Black Cover Lockers with a Flat Screwdriver as shown in Figure 2‐1. 
3. Dismantle the Front and Upper Covers. Take out the small packages.  
4. Remove all the Covers. 
5. Carefully Remove the OrthospecTM from the plate and put it on the floor.   
Note: It is recommended to keep the box for future shipment. 
 

 
Figure 2‐1: Unpacking the Orthospec 

 
 
Page 5 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

3 PARTS LIST 
No.  Item Description  Qty  P/N 
1  OrthospecTM OR‐3   1  OR‐3‐A999 
2  2‐212 O‐Ring  1  OR‐2‐E003 
2  Drain Pipe  1  SW‐4‐A052 
3  5 amp Fuses  ( for 230V Inlet voltage)  1  TT‐2‐M052 
4  10 amp Fuses  ( for 115V Inlet voltage)  2  TT‐2‐M153 
5  Plastic Funnel  1  UF‐1‐E077 
6  Multi‐use Electrode  On demand  SE‐5‐A999 
7  Tips for Multi‐use Electrode  On demand  SE‐5‐C007 
 

4 BEFORE STARTING TO SERVICE 
It is important to be familiar with the device before starting to service. Be sure that you are 
familiar with the device: 
 Structure including wiring diagram 
 Components, connectors 
 Accessories, consumables 
 Operation procedure and modes  
 Graphic User Interface including all screens and windows 
 Required user maintenance actions 
 Error messages 
 

Any hardware modification shall always be done in accordance with the 
wiring diagram relevant to the model in use.  
 
 
For detailed and complete device description refer to the Orthospec Operating Manual:  
OR‐3‐X400.  
Also, a block diagram describing the system is available in the appendix of this manual. 
 
Dangerous high voltage! Before opening the device for any purpose 
it must be turned off and disconnected from the mains power for at 
  least 30 minutes. 
 

5 TOOLS & EQUIPMENT 
The following tools and equipment are needed:  
1) Allen keys of the following sizes: 2.5mm, 3/16”, 5/32”, 3/8” 
2) Philips Screwdriver (medium size) 
3) Flat Screwdriver (medium and large size) 
4) Multitester (DVM – digital multimeter) 
5) Laptop + USB extension cable 
6) Scissors 
   

 
 
Page 6 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

6 INITIAL INSTALLATION  
1) Make sure the swivel casters and brakes function properly. 
2) Make sure that the Supporting Arm with its three knobs  function properly.   
3) Make sure Main Switch is turned off, than Connect the AC power cord to the 
Orthospec and plug its other end to the mains outlet.   
4) Turn the Orthospec ON. 
5) Touch the Open Door button on the main screen and wait till the Maintenance Door 
is fully opened.  
6) Follow the “Initial preparations” described in Chapter 3 of the operating manual 
(electrode calibration/installation and reservoir filling).. 
 

7 TECHNICIAN SCREEN  

 
Figure 7‐1: Technician Screen 
 
The technician screen is used for settings and for system testing. It contains 4 tabs located at 
its top part: 
1. MCU tab 
2. Counters tab 
3. General tab 
4. Configure tab 
Each tab allows specific settings and tests as described in the following paragraphs. 
        To enter the Technician Screen, press on the Technician screen tab and type 15426.  
 
 
 
 
 
 
 

 
 
Page 7 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

8 INITIAL CONFIGURATION – GENERAL TAB  
For changing the default configuration use the "General" tab of the technician screen.  
After completing the system configuration it is recommended to go to the Main screen and 
verify it is working according to the selected configuration.   
 
MCU Error Log 

General 

Countdown  

Operate 

Error Log 

 
Figure 8‐1: General Tab  
 
The general tab contains 5 sections: 
 
Section  Content    Explanation  
General  Screen alignment button  Screen alignment 
Total pulse count of the device   
Countdown  Up  To set a count up mode 
Down  To set a countdown mode 
Default [number]  To  set  the  default  countdown  counter 
value. Become valid after system restart.  
Operate  Disabled   Not applicable in this product. 
Error Log  Debug  Defines the error log display. 
Info 
Warning 
Error 
MCU Error  Log On/Log Off Turn On/Off the log display 
log  Clear   clear the log display 
Save  Save the log to a file. 
 
   

 
 
Page 8 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

9 COUNTERS TAB  
This  tab  provides  information  regarding  the  value  of  the  Filter,  Plug  and  Membrane 
Counters.  
 

Do not change the counters preset values unless officially instructed by 
Medispec's Service Center 
 
 

 
Figure 9‐1 Devices counters TAB 
 
This tab contains 4 Counters Value: 
Section  Content  Explanation  
Filter  100000  The Filter must be cleaned periodically  
Plug  3800  The  Electrode  must  be  replaced  after  every 
section 
Membrane  50000  The Membrane must be replaced periodically 
Membrane Month  3  after 50000 pulses or every 3 month. 

10 CONFIGURE TAB 
This tab is password protected and is intended for technician that was specifically authorized 
by Medispec.  
When selecting this tab, two other tabs will appear:  
1) General1 
2) Water 

 
Figure 10‐1: Configure tab  
 
 
Page 9 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

10.1 General1 Tab 
This tab enables advanced system configuration. It allows: 
1) Setting the door actuators  timing 
a. Default values:  Start Open Door=7000 / Open Door=13000 
2) Setting the water drain  timing 
a. Default value: 20000 
3) Disable or enable the device height piston and shockwaves frequency during 
treatment.  
4) Setting the type of the High Voltage Power Supply. 
5) Setting the serial number of the device. 
6) Reset database after fatal error. 
7) Determining the mode of maintenance expiry warning. 
a. Checked – the warning will appear on every STOP and on system startup 
b. Unchecked ‐ the warning will appear only on system startup 
8) Determination of the date format. 
 

 
Figure 10‐2: General1 Tab for  Advanced configuration 
 

Date correction may only be done by Medispec's Personal. 
 
 

 
 
Page 10 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

10.2 Water Tab – Service door and sensor 
 

 
Figure 10‐3: Water Tab 
 
This tab allows manual control of the door Open/Close and membrane Inflate/Deflate.  
It also contains 5 indicators reflecting the limit switches: 
 
Indicator  Element  Explanation 
Plug  S2 ‐Limit Switch  Electrode presence indication (green‐yes, red‐no) 
Door sensed  S3‐ (Reed Relay)  Indication whether Maintenance Door is closed  
closed  (green – closed, red – opened) 
Spring Tension  S1‐ Limit Switch   Indication whether Maintenance Door is properly 
Senses  closed  
(green‐yes, red‐no) 
Interlock  S2 or S3  System is ready, Electrode is in and Maintenance 
Door is  
Closed (green‐yes, red‐no) 
Water in System  OR‐3‐A080  Water in reflector indication (green‐yes, red‐no)   
Water sensor electrodes  When the reflector is full with water the high 
SW‐6‐A840  voltage is enabled  
Water sensor Controller
Note: All five indicators shall be green to indicate normal status of the system 
 
 
 
 
 
 
 
 
 
 

 
 
Page 11 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

10.3 Service door sensors system 
 
When one of the indicators is red, check the relevant sensor as described in the tables 
bellow.  
 
 
#4
#1

#3

#5
#2
S2  S1 S3 

 
Figure 10‐4: Limit Switches  
 
 
 
#  Element  Description  Part Number 
S2  Plug Detector ‐ Limit Switch  OR‐3‐M120 
S3  Door sensed closed ‐Reed Relay  OR‐3‐M150 
S1  Spring Tension Senses‐ Limit Switch  OR‐3‐M120 
1.   Spring Tension  SW‐2‐B018 
2.   Spring Tension locker   
3.   Reed relay magnet  OR‐3‐M151 
4.   Limit Switch bracket *  OR‐3‐B124 
5.   Reed Relay bracket*  OR‐3‐B034 
 
 

* While replacing the element do not remove the element bracket. 
Otherwise, you will have to re‐calibrate the system.  
 
 
 
  Switches Positions    
Function  S1  S2  S3 
The plug is not installed  Released  Pressed  ON 
The Plug is installed  Released  Released  ON 
The door is Interrupted  Pressed  Released  OFF 
The door is opened  Pressed  Released  OFF 

 
 
Page 12 of 37 

 
 
Orthospec Service Manual OR‐3‐X405 

11 MCU TAB  
The  MCU  tab  contains  system  status  indicators  and  buttons  to  analyze  and  control  the 
microcontroller and the high voltage module of the Orthospec.  
 

 
Figure 11‐1: MCU TAB 
 

Green designates enabled/OK status 
Black designates disabled 
Red designates an error 
Note: the normal status is when all the indicators are green. 
 

   

Tech pulses = number of shockwaves performed in the technician mode 
since the Orthospec was turned ON. 
HV monitor = displays the high voltage output  
 
   
HV enable  Activates the high voltage 
Slow mode  Start to deliver shockwaves in a frequency of 30 / minute(slow) 
Single trigger  Start to deliver a single shockwave 
Trigger enable  Start delivering shockwaves (normal) 
Spark gap  Starts a "warm up" process for the spark gap (shall be done after extended 
storage ~ 3 months). Two warm up modes are available: Full and Short. 
KV  to determine the high voltage level 
Override  to bypass an error 
SBC Fail  to simulate a fault in the Watch Dog or the OMAP 
Show log  to show the system activity log 
                              
When the system activity log was opened, the following buttons appear: 
Log On  to start activity logging 
Clear  clear activity log display 
Save  save the activity log display to a file 
 
 
Page 13 of 37 

 
 
12 TROUBLESHOOTING  
Error description  Refer to section  Action required 
#4  ‐ HV Power supply fail   13.1  HVPS INPUT  Voltage Test 
#11  ‐ Calibration failure  12.4,12.5  High Voltage Calibration + Calibration Test 
#301,#101  ‐  MCU  14.4  Restart the system and if the problem persists replace the Control 
communication failure  Unit  
#301,#105 ‐ FPGA  14.4  Check in the information screen if the FPGA version appear and if 
communication failure  not, restart the system. if the problem persists replace the 
Control Unit 
 

#301,#105,#3  ‐  Terminal  13.2  Control PCB  Voltage Test 


Voltage failure  13.4,13.5  If the problem persists replace the Control Unit and perform high 
voltage calibration. 
#301,#6 ‐ Thermal failure   13.3  Thermal switch test 
Shockwaves  are not produced and the counter is  13.6  Go the MCU tab and check all the indicator leds. If a specific led is 
stuck  red act according to the instructions relevant to this led.  
Operate button is green during Idle mode  13.8  Follow override procedure.  
System cannot be turned ON  13.2   Verify that that the system is properly plugged in the mains 
13.4, 13.5  voltage. 
   Check the fuses and replace if necessary.  
   If the fuses are ok check LED1 and LED2 (low voltage power 
  supply indicators) on the Control Unit.  
 If both LEDs are ON, replace the Control Unit.  
 If one of the LEDS is OFF check the transformer connections 
and the MF1 connector on the Control Unit.  
 If the connection is ok verify 15vac between MF1‐1 to MF1‐2. 
 If there is no voltage, replace the transformer. 

 
 
Page 14 of 37 

 
Orthospec Service Manual OR‐3‐X405 

Error description  Refer to section  Action required 


Shockwaves are not produced although the  14.6  Follow  the  actions  in  the  list  below.  If  the  problem  persists 
system seems to work properly.   13.1, 13.2  continue to the next action in the list:  
13.4, 13.5   Replace spark plug. 
   Error! Not a valid result for table.  
 Check the H.V connection between the HAVES and the 
HVG.(Alden connector) 
 High voltage test and calibration test. 
 Check continuity of the cable connecting the control Unit and 
the HVG. 
 Replace the HVG. 
The system resetting during the shock wave  13.4,13.5  Follow  the  actions  in  the  list  below.  If  the  problem  persists 
continue to the next action in the list:  
 Clean the Electrode mounting area  
 Check that all the ground connections. are intact (see 
appendix).  
 If the problem persists replace the Control Unit 
10.2   Check 20Vac between TF1‐11 to TF1‐12, Check 24Vdc on the 
13.2  Value  (6P connector) 
between CON6‐1 to CON6‐2 
 
The membrane Inflate button pressed but the   Check 20Vac between TF1‐11 to TF1‐12 during Inflating ‐, if 
membrane doesn't Inflating at all.   24Vdc persists between CON6‐1 to CON6‐3 replace the 
  valve. 

 
 
Page 15 of 37 

 
Orthospec Service Manual OR‐3‐X405 

Error description  Refer to section  Action required 


13.2   The OR3 reservoir is  2.5 liters and it should be filled with 
10.2  saline solution. 
 
   Check 20Vac between S21 to S22 on the water sensor 
The membrane Inflate button pressed but the  controller on TB1, Check The Fuse TB1‐3. 
button doesn't become dark.   Entered to the technician screen ‐    Water Tab under 
The Shockwaves can't be produced.  Configure tab, inflate the membrane when it's full the 
  indicator led shod be green. If not tray to recalibrate the 
sensor controller  SW‐6‐A840   
 Replace and recalibrate the unit. 
 Replace the sensor electrodes OR‐3‐A080 
10.2   Check 230V on the pump 3P connector, during the deflate 
13.2  action, if 230V persists and the pump not working replace 
the pump. 
 
The membrane deflate button pressed ,but the   Check 230V between TB1‐4 and TB1‐2, replaced the Fuse F4 
membrane doesn't deflate   if necessity. 
 Check the relay K3 on TB1 . During the deflate action 24V 
shod be measured on the relay coil, Check the wiring, 
Replace the relay. 
 If the Relay ‐ control voltage not persists replace the Control 
Unit. 
 
 
 

 
 
Page 16 of 37 

 
Orthospec Service Manual OR‐3‐X405 

Error description  Refer to section  Action required 


The  service  door  starts  closing  and  immediately  10.2   The electrode is not inserted all its way into the reflector. 
open again.    There is no adaptation between the HVG and the service door 
against the electrode. 
 There is a mechanical interference that prevents the service 
door closing. 
 The switches are not calibrated.   
 Check especially the S3 switch by using the technician screen, 
 replaced elements if needed  
The system cannot recognize that the electrode is  10.2   Check the S2 switch by using the technician screen, replaced 
in the reflector. elements if needed.
The  service  door  cannot  close  at  all  (not  13.2   Check the DC voltage on the actuator connector during door 
responding). 10.2  operation from the technician screen. If there is ~24Vdc 
change the actuator. 
 Check the AC voltage between MF6 –MF7 on the control unit, 
if there is ~20Vac change the control unit.  If there is 20Vac 
check the main transformer TF1 
 If you can operate the door actuator just from the technician 
screen, check the switch S1 replace if needed and calibrate 
the door system. 
The device Height Actuator is not responding. 13.2   Check the DC voltage on the actuator connector during height 
operation. If there is ~24Vdc change the actuator. 
 Check the AC voltage between MF6 –MF7 on the control unit, 
if there is ~20Vac change the control unit.  If there is 20Vac 
check the main transformer TF1 
Shockwaves  are  not  stable  (double  or  missed  13.7   Replace spark plug (Electrode) 
shockwaves)   Perform spark‐plug recovery (13.7) 
   Replace  the HVG / Call Medispec  Service 
 

 
 
Page 17 of 37 

 
13 SYSTEM TESTS AND REPAIR PROCEDURES 
13.1  HVPS INPUT  Voltage Test 
To analyze the source of HVPS supply failure, perform the following voltage measurements: 
1) Power supply main plug (Figure 13‐1a) 
2) "Control": between SV9‐4 to GND (see note below and Figure 13‐1b ) or TP3 on the 
Control PCB (Figure 13‐2). 
3) "Enable": between SV9‐7 to GND (see note below and Figure 13‐1b) or TP18 on the 
Control PCB (Figure 13‐2). 
4) Control unit Main Power input: between MF1‐4 to MF1‐8 (Figure 13‐1c) 
5) 230V Secondary winding 
6) 230V Primary winding 
Note: To measure the "control" and "enable" you will first have to press: "HV Enable 
button" in the MCU tab   "KV" button  type 14000  "Enter". 
After measuring, look at the table below and locate the line reflecting your measurement 
results to find the source of failure.  
 

Perform the voltage measurements in this order

1  2  3  4 5 6  
Power  "Control"   "Enable"  Control  230V Secondary  230V Primary  Source of failure 
supply  unit  winding  winding 
Main Plug  input 
=230V  ~6V ~5V  H.V. power supply
≠230V      =230V Control Unit 
≠230V      ≠230V =230V Wiring problem 
≠230V      ≠230V ≠230V =230V transformer 
≠230V      ≠230V ≠230V ≠230V Instable mains 
power or faulty 
power entry module 
Note: empty cell = no need to measure 

(a) Power supply main plug  (b) SV9 connector  (c) MF1 connector 


 
Figure 13‐1 HVPS INPUT  Voltage Test 
 
   

 
 
Page 18 of 37 

 
Orthospec Service Manual OR‐3‐X405 

13.2 Control PCB  Voltage Test 
To analyze the source of Terminal voltage problem, follow the measurements according to 
the table below. If the problem was not resolved replace the Control Unit and perform high 
voltage calibration.  

 
 
Figure 13‐2: Control PCB (Control Unit) 
 
Control unit  Transformer Measurement  Source of failure/Action required 
Measurement   on the  Secondary winding  
TP12≠12V or TP13≠5V   The 15V winding =15V  Replace  the control PCB 
TP12=12V and TP13=5V   The 15V winding ≠15V  Check the  mains power: 
 Inlet voltage, fuses (F1&F2)  , Main 
transformer, Power entry module and wiring . 

13.3  Thermal Switch Test 
When the thermal safety led in the MCU tab is red, a thermal problem exists. The basic 
assumption is that the room temperature complies with the environmental specifications.  
To analyze the source of problem follow the measurements according to the table below: 
 
Continuity test  between  HVG / ambient   Source of failure/Action required 
MF2‐5 to MF2‐8  Thermal switches test 

ok ( closed circuit)    Replace the Control Unit 
Open circuit   Close   Check the thermal switches connections/wiring  

Open circuit   Open (HVG or ambient  Replace the thermal switch element and recheck 


or both)  continuity as explained above.   
Note: empty cell = no need to measure/check 

 
 
Page 19 of 37 

 
Orthospec Service Manual OR‐3‐X405 

 
Figure 13‐3: thermal switch element 
 
13.4 High Voltage Calibration 

 
Figure 13‐4: MCU/KV TAB   
When the high voltage test fails it is required to perform high voltage Calibration as follows:  
1) Press the HV Enable button. 
2) Press on the KV button . 
3) Type 12000 in the high voltage input field. 
4) Press Enter. 
5) The system status icon will change to "X" for few seconds and then will turn to "ok".  
6) Perform "Calibration Test" (section 13.5). 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
Page 20 of 37 

 
Orthospec Service Manual OR‐3‐X405 

13.5 Calibration Test  
1) Verify that all system led indicators are green. 
2) Press the HV Enable button. 
3) Press on the KV button. 
4) Type a value between 14000 and 20000. 
5) Press Enter. 
6)  Check the HV monitor – the value should be equal to the value you typed ±1% 
a. If ≤ 1% ‐ the test completed successfully. 
b. If > 1%  ‐ perform High Voltage Calibration (section 13.4) 
7) Repeat the test for 3 more values (you can use the ±1000 / 250 buttons). 
 
Example:  
 
KV Value  Minimum   Maximum 
14000  13860  14140 
16000  15840  16160 
18000  18180  17820 
20000  20200  19800 
 

13.6 Delivering Shockwaves via Technician Screen 
1) Verify that all system led indicators are green.  
2) Press the HV Enable button. 
3) Press the KV button. 
4) Type a value between 14000 and 20000. 
5) Press Enter. 
6) To start shockwaves in the standard frequency (120 ppm) press on the "Trigger 
Enable" button. 
7) To deliver a single shockwave press on "Single Trigger" button. 
8) To start shockwaves in the slow mode (30 ppm) press on the "Slow Mode" and then 
"Trigger Enable" button. 
9) Repeat this for different values of High Voltage in the range of 14000 to 20000  

 
 
 
 
 
 
 
 
 
 
 

 
 
Page 21 of 37 

 
Orthospec Service Manual OR‐3‐X405 

13.7 Spark‐gap recovery  
 

 
Figure 13‐5:  Spark‐gap recovery 
 
1) Press the Spark‐gap  button. 
2) Check the box near  "Don't  verify triggering" . 
3) On the right bottom side press on "Start Full Recovery" button.  

The KV will be automatically increased while delivering shockwaves in between.  

This procedure may take approximately two hours.   

13.8 Override Procedure 

 
Figure 13‐6: Override in the MCU screen 
 
1) Click on the override button. 
2) Click on the interlock button. 
3) Go back to main screen and initiate a shock wave. 
4) If the problem was resolved, perform Interlock Test.  
5) If the problem persists, override the thermal safety. 
6) If the problem was resolved, perform Thermal Switch Test.  
7) If the problem persists replace the Control Unit.  
 
 
Page 22 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14 COMPONENTS REPLACEMENT GUIDE 
14.1 Overview 
The following paragraphs describe the replacement procedure of specific components of the 
Orthospec.  
All other modifications which are not described in this manual shall not be performed.  
   
All Replacement procedures should be done while the system is turned OFF and 
while  the  power  cable  is  disconnected  from  the  main  outlet.  Wait  20  minutes 
  before removing the covers. 

14.2 System assembly description 
 

Water sensors controller  

System Counter 

Main terminal TB1 

Main Transformer TF1 

Figure 14‐1: Internal View – Right Side 
 
 
Page 23 of 37 

 
Orthospec Service Manual OR‐3‐X405 

HV Connector 

Spellman HVPS  

Figure 14‐2: Internal View ‐ Left Side 

 
 

 
 
Page 24 of 37 

 
Orthospec Service Manual OR‐3‐X405 

Control Unit 

HVG 

Spellman HVPS  

Limit Switches 

Main Transformer 

Power Entry 

 
Figure 14‐3: Internal View – Back Side 
 
    
 
 
 
 
 
 
Page 25 of 37 

 
Orthospec Service Manual OR‐3‐X405 

Water Filter 

Reservoir 
Water Valve 

Water Pump 

   
(a)  (b) 
Figure 14‐4: Water System 
 
 
14.3 External Fuses Replacement 
1) Open the External Fuse Housing F1 and F2 using a flat screwdriver. 
2) Remove the faulty fuse and replace it by a new one  
3) Use the correct fuse according to the correct wiring diagram. 
 

 
Figure 14‐5: Fuse Replacement 
     
 
 
Page 26 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.4 Covers Removal 
1) Remove the Side Covers by opening two Philips screws at the bottom of each cover. 
2) Disconnect the Fast Ground Connector. 
3) Remove the Back Cover by opening Six Philips screws. 
 
 

 
Figure 14‐6: Orthospec covers disassembly 
 
14.5 Internal Fuses Replacement 
1) Open the Internal Fuse Housing F3 and F4, and pull down the housing. 
2) Remove the faulty fuse and replace it by a new one. Pay attention to the nominal 
        Voltage of the fuse. 
 

    
 
Figure 14‐ 7 Internal Fuse replacement 
 
 
 
 

 
 
Page 27 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.6 HV Power Supply (HVPS) Replacement 
1) Remove the back and the side covers. 
2) Disconnect the HV connector (Figure 14‐8a) 
3) Discharge the HVG Capacitor by touching the chassis with the metallic tip  
    

   
(a)  (b) 
Disconnecting the HV Connector  Discharging the HVG by contact 
Figure 14‐8 HVPS replacement 
 
4) Disconnect the AC Inlet Connector, White Control Connector and the Ground Wires 
By opening the Philips screw with a middle size Screwdriver. 
5) Remove the HVPS by opening eight Philips screws with a small Philips screwdriver. 
 

 
Figure 14‐9 dismantling the HV power supply 
 
 
6) Remove the Bottom Bracket from the "old" HVPS and assemble it to the "new" HVPS 
by opening the two small Philips screws. 
7) Replace the HVPS by a new one  
8) Perform  the  following procedures: : HV calibration (see 13.4)  and  HV calibration 
test (see 13.5). 
9) After closing the covers, turn on the Orthospec and from the Main screen let it deliver 
~1000 shockwaves.  
 
 
Page 28 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.7 HVG Replacement 
1) Make sure an Electrode is installed in the device’s Electrode Housing and the 
Maintenance Door is closed than switch the device OFF and disconnect from  
The main AC Inlet.  

To avoid electrocution:  Disconnect the HV power supply connector (white 
connector) from the HVPS and ground it to the metal chassis. 
 
 
2) Remove the back and the side covers. 
3) Disassemble the HVG Bracket by opening  two 5/32” Allen screws.      
4) Disconnect the HVG Control Connector from the HVG by opening the Connector. 
Disconnect two red fasten Connectors of the Thermal switch and the Ground fasten 
connector. 
5) Pull the HVG out of the device and replace it by a new one. Insert the new HVG gently 
all the way in. Make sure the Electrode is seated inside its hole in front of the HVG.  
6) Turn the device ON.  Assemble the HVG Bracket, don’t tighten two Allen Screws 
of the Bracket. Open the Maintenance Door by clicking on the Open Door Soft‐key 
Then click on the Close Maintenance Door.  
 Make sure the door is closing properly and only then tight the HVG Bracket.  
Turn the device OFF and connect all the cables following steps 5, 4, 2. 
7) Tests needed: Spark‐gap recovery procedure, operating the device around 1000 
shockwaves from the main screen. 
 

   
Figure 14‐10: HVG Replacement 
 
 
Page 29 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.8 Control Unit Replacement 
1) Remove the back and the side covers. 
2) Open the four Butterfly Nuts below the Control Unit  Housing. 
3) Pull the Control Unit  out of the housing and swivel it to ease cables disconnection. 
Note: to avoid scratches, insert soft material between the front of the system and          
the Control Unit .   
4) Remove the Bracket Control Unit  by opening four Philips screws with a middle 
size screwdriver and pull the bracket gently out from the Control Unit . 
5) Disconnect nine White Connectors from the Control Unit . Disconnect two D‐Type 
Connectors but first open two Philips screws of each one.    
6) Disconnect two Thermocouple and USB Connection from the Control Unit . 
7) Disconnect two D‐Type Connectors by opening two Philips screws of each one. 
8) Tests needed: Defining the device serial number, operating the all device functions 
from the main screen, makes around 1000 shockwaves from the main screen.  
 

   
 

 
 
Figure 14‐11: Control Unit  Replacement 
 
 
 
 
 
 
 
 
 
 
 

 
 
Page 30 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.9 Door Actuator Replacement 
1) Remove the back and the side covers. 
2) Open the Maintenance Door by clicking on the Open Soft‐key and  
Turn the system OFF and remove the HVG. 
3) Remove the Spring Disc with a Plier and remove the Metallic Pin. 
4) Open the Actuator Housing by opening the Screw with 13mm Wrench and pull the  
Actuator out from the Actuator Housing. 
5) Disconnect the white Connector of the Actuator. 
6) Replace the Actuator by a new one and assemble it by following step 4 and 5. 
7) Turn the device ON and click the Open Maintenance Door Soft‐key. 
Assemble the Metallic Pin . 
8) Turn the device OFF and assemble the HVG back into the system.  
9) Tests needed: Calibrating the service door limit‐ switches, operating the all device 
functions from the main screen. 
     

 
Figure 14‐12 Door Actuator replacement 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
Page 31 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.10 Water Sensor Replacement 
 
1) Remove the back and the side covers. 
2) Remove the Left Water System Cover by opening three Philips screws. 
3) Remove the two Locking Rings by opening the rings with a wrench. 
4) Remove the Electrical White Connector than disconnect the Sensor Pipes 
by pulling the Flexible Pipes out from the Blue Pipes.  
Notice: It is easier to insert or to pull‐out the Flexible Pipes after heating it is by a            
              Blower.   
5) Disconnect the white Connector than disconnect the Sensor Pipes. 
6) Replace the Water Sensor.  
7) Tests needed: calibrating the water sensor controller if needed, operating the all 
device functions from the main screen especially the water system. 
 
 

       
Figure 14‐13 Water Sensor replacement 
 
 

Figure 14‐14 water sensor controller on the Main terminal 
 
 
 
 
 

 
 
Page 32 of 37 

 
Orthospec Service Manual OR‐3‐X405 

14.11 Valve Actuator Replacement 
 
1) Remove the back and the side covers. 
2) Open the Maintenance Door by clicking on the Open Door Soft‐key. 
3) Make sure the Membrane is empty and remove the Electrode from its 
housing. 
4) Disconnect the Valve actuator Connector. 
5) Remove the Valve actuator by opening two Allen screws with 9/64 Allen Key.   
6) Replaced the Valve actuator . 
7)  Tests needed: operating the all device functions from the main screen especially the 
water system. 
 
 
   

 
Figure 14‐15  Valve Actuator Replacement 
   

 
 
Page 33 of 37 

 
Orthospec Service Manual OR‐3‐X405 

15 PREVENTIVE AND MAINTENANCE PROCEDURES 
15.1 Routine user maintenance procedures 
 
The Orthospec user is required to perform several maintenance procedures that are detailed 
and explained in the operating manual.  
During service it is recommended to review these procedures with the user and verify they 
are being done properly.  

The safety of this equipment depends on ongoing maintenance and regular safety 
inspections.  
Electrical leakage current and ground impedance of this equipment needs to be 
 
checked by a Medispec authorized technician every 6 months.  
 

15.2 Periodical service procedures 
The following procedures shall be carried out once in a six months by Medispec qualified 
service technician.  
1) Use Alcohol to clean the electrode contacts of the HVG while the Service door is 
opened. (**) 
2) Check and replace if necessary the O‐ring (NR‐2‐E040) located in the reflector. 
Lubricate it with coupling solution. (*) 
3) Clean the Reflector using alcohol. (**) 
4) Check the water system. 
5) Electrical safety test: 
a. Check the Grounding 
b. Check the Leakage current 
6) General inspection of the installation /Cables 
 
(*) Orthospec Power shall be OFF. 
(**) Orthospec Power OFF and power cord disconnected. 
   

 
 
Page 34 of 37 

 
Orthospec Service Manual OR‐3‐X405 

16 APPENDIX A – BLOCK DIAGRAM 
 

 
 
Page 35 of 37 

 
Orthospec Service Manual OR‐3‐X405 

17 APPENDIX B – SPARE PARTS LIST 
 
#  Part Description  Part Number 
1. Orthospec System Computerizes  OR‐3‐A999b 
2. HVG for Orthospec Digital Model  VG‐6‐A999c 
3. HVPS/Switching HV Power Supply (Spellman)  SW‐5‐M175 
4. Control Unit   OR‐3‐A430 
5. Valve Operator / Water Control Ball Valve CD  OR‐3‐A855 
6. Ball Valve 3/4"  SW‐4‐E500 
7. Water Sensor  OR‐3‐A080 
8. Door Actuator   OR‐3‐L525 
9. Height Actuator  OR‐3‐A402 
10. Multi‐Use Electrode   SE‐5‐A999 
11. Tip  SE‐5‐C007 
12. Electrode O‐Ring  NR‐2‐E040 
13. Wheel Colson 4" (X4)  MW‐4‐E026 
14. Door Actuator  OR‐3‐L525 
15. Water Pump  SW‐6‐A850 
16. Electrode O‐Ring  NR‐2‐E040 
17. Water Filter Assy.  SW‐4‐E045 
18. Membrane Fastening O‐Ring  OR‐3‐E002 
19. Lower Apparatus System  OR‐3‐A041 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
Page 36 of 37 

 
Orthospec Service Manual OR‐3‐X405 

18 PRODUCT CONTACT INFORMATION 
If  an  adverse  event  occurs,  such  as  a  malfunctioning  of  the  device,  a  mistake  in  using  the 
device,  or  an  injury  relating  to  the  use  of  the  device,  report  the  occurrence  immediately.  
Alert the physician of any patient health issues that occur while using the OR‐5 Omnispec.  
For  troubleshooting  assistance,  complaints  or  additional  questions  regarding  the  OR‐5 
Omnispec device, contact the Service Department: 
 
Medispec Ltd. 
203 Perry Parkway, Suite 6 
Gaithersburg, MD 20877 
Tel: 301‐944‐1575 
Toll Free:  1‐888‐663‐3477 
Fax: 301‐972‐6098 
USA: servicedept@medispec.com 
Rest of world:  service‐int@medispec.com 
Site: www.medispec.com 

 
 

 
 
Page 37 of 37 

You might also like