You are on page 1of 12

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/275239842

Correlation Analysis of Multiple Sensors for Industrial Gas Turbine Compressor


Blade Health Monitoring

Article  in  Journal of Engineering for Gas Turbines and Power · November 2015


DOI: 10.1115/1.4030350

CITATIONS READS

3 287

6 authors, including:

Christopher Edison Hill Tim Lieuwen


Agilis Measurement Systems, Inc. United States Georgia Institute of Technology
1 PUBLICATION   3 CITATIONS    309 PUBLICATIONS   5,775 CITATIONS   

SEE PROFILE SEE PROFILE

Christopher Perullo
Georgia Institute of Technology
37 PUBLICATIONS   139 CITATIONS   

SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Combustion Dynamics Monitoring View project

Stability Analysis of Flames View project

All content following this page was uploaded by Tim Lieuwen on 16 June 2015.

The user has requested enhancement of the downloaded file.


Proceedings of ASME Turbo Expo 2014: Turbine Technical Conference and Exposition
GT2014
June 16 – 20, 2014, Düsseldorf, Germany

GT2014-27040

CORRELATION ANALYSIS OF MULTIPLE SENSORS FOR INDUSTRIAL GAS


TURBINE COMPRESSOR BLADE HEALTH MONITORING

Brian Kestner* Chris Hill


Georgia Institute of Technology Agilis Measurement Systems, Inc
Atlanta, GA Palm Beach Gardens, FL, USA

Leonard Angello Josh Barron Tim Lieuwen*


Electric Power Research Southern Company Georgia Institute of Technology
Institute Birmingham, AL, USA Atlanta, GA
Palo Alto, CA, USA

ABSTRACT
This paper summarizes an analysis of data obtained from NOMENCLATURE
an instrumented compressor of an operational, heavy duty BPF – Blade Passing Frequency
industrial gas turbine; the goal of the aforementioned analysis is EO – Engine Order
to understand some of the fundamental drivers which may lead F.R. – Forced Response
to compressor blade vibration. Methodologies are needed to 1.) HO – Higher Order
understand the fundamental drivers of compressor blade FFT – Fast Fourier Transform
vibration, 2.) quantify the severity of “events” which accelerate ND – Nodal Diameter
the likelihood of failure and reduce the remaining life of the NSV – Non-synchronous vibration
blade, and 3.) proactively detect when these issues are p-p – Peak to Peak
occurring so that the operator can take corrective action. The
motivation for this analysis lies in understanding the INTRODUCTION
correlations between different sensors which may be used to This paper reports the results of measurements from an
measure the fundamental drivers and blade vibrations. operational gas turbine. Methodologies are needed to 1.)
In this study, a variety of dynamic data was acquired from understand the fundamental drivers of compressor blade
an operating engine, including acoustic pressure, bearing vibration, 2.) quantify the severity of “events” which accelerate
vibration, tip timing, and traditional gas path measurements. the likelihood of failure and reduce the remaining life of the
The acoustic pressure sensors were installed on the first four blade, and 3.) proactively detect when these issues are
compressor stages, while the tip timing was installed on the occurring so that the operator can take corrective action. The
first stage only. These data show the presence of rotating stall motivation for this analysis lies in understanding the
instabilities in the front stages of the compressor, occurring correlations between different sensors which may be used to
during every startup and shutdown, and manifesting itself as measure the fundamental drivers and blade vibrations.
increased amplitude oscillations in the dynamic pressure This study aims to begin to address these needs by bridging
measurements which are manifested in blade and bearing the gap between the current lack of knowledge in this area, and
vibrations. The data that lead to these observations were a formal methodology. Before developing a formal
acquired during several startup and shutdown events, and methodology, it is necessary to understand the engine in its
clearly show that the amplitude of these instabilities and the normal operating state by analyzing the data from the available
rpm at which they occur can vary substantially.

* Primary employer – Work performed as consultant to EPRI 1 Copyright © 2014 by ASME


sensors, thus understanding what each sensor’s data provided Rotating stall occurs when the flow around a blade, which
us, and how they affected and interacted with one another. can be viewed as an airfoil, separates and the resultant flow
through the annulus is restricted. Rotating stall can lead to
COMPRESSOR VIBRATION AND ROTATING STALL increased vibrational stresses and may produce resonance
Before developing techniques to detect compressor conditions [3]. The rotating stall typically occurs when the
problems, it is important to understand how and when blade compressor or compressor stage is operating high on its
vibration may occur. Four primary drivers for blade vibration operating line. Large stall zones, or cells, may cover multiple
are rotating stall, flutter, and synchronous vibration (or forced blade passages. These cells propagate in the direction of the
response). Depending on sources in literature, the term rotor, typically at 30 to 70% of the rotation frequency. When
nonsynchronous (NSV) may be a catch all for any type of rotating stall is present, it may show up as one or more stall
nonsynchronous vibration or it may refer to a specific type of cells. In addition, the stall cell may or may not occur across the
tip clearance driven flow separation known as rotating span of the blade. Typically only a single full span cell appears
instability[1]. Both flutter and rotating stall driven vibration are [4]. A full-span stall can extend axially through the whole
considered nonsynchronous. These events are not a function of length of a compressor. Part span stall cell only occurs near the
engine speed and appear as horizontal lines on a Campbell tip of a blade. There may be multiple cells present
diagram. When the frequency of certain process is an integer simultaneously. Unlike full span stall cells which extend axially
multiple of the engine’s RPM, these integers are known as through the compressor, part span stall cell may only exist
engine order (EO) and the resultant vibration is known as locally in a single stage. During part span rotating stall, the
synchronous. These engine orders appear as diagonal lines on number of stall cells may change. Full span typically results in
plots of frequency vs. engine speed, known as “Campbell a more noticeable overall compressor performance change.
diagrams”. The compressor blades and disk assembly have their Additionally when part span stall occurs, if the compressor is
own natural frequencies (or modes) which can intersect with further throttled, a full span stall may occur.
these excitation frequencies and lead to blade vibration. As mentioned previously multiple rotating cells may
Figure 1 shows a hypothetical Campbell diagram and what appear with each cell rotating at about 50% of the rotating
the different types of vibrations look like. In this diagram, the frequency. During a rotating stall event, the number of cells
first mode of vibration is just above 100 Hz. The second mode may vary. Figure 2 shows a Fourier transform of hot wire
of vibration is at about 300 Hz and the third mode of vibration measurements of a single and 2 cell rotating stall cell on a
occurs around 400 Hz. Both flutter and NSV do not occur at an single stage axial flow research compressor. In this figure,
engine order integer multiple, synchronous resonance occurs depending on when the snapshot of data was taken, there
when a natural frequency crosses an excitation occurring at an appear either one or two stall cells. The single stall cell appears
integer multiple of the engine speed. In this figure, synchronous as a single peak at a frequency of about 50% that of the rotating
vibrations occur around 4500 rpm at the mode 2 frequency speed. The two cell stall appears as a peak closer to the rotating
which would correspond to 4 engine orders (EO). It may be frequency. When the whole data set is aggregated, both peaks
referred to as a mode 2 4 EO synchronous vibration. On the appear. So depending on the length of sampling of the data and
other hand, flutter , rotating stall, and NSV will vibrate at a the number of stall cells present (which may range from 1 to
given frequency over a range of rpm where the flutter is a mode about 5), rotating stall may appear as discrete peaks ranging
2 vibration and the NSV is a mode 3 vibration. from 30% to 200% of the rotating frequency.

Figure 1. Hypothetical Campbell Diagram Showing Sources


of Vibration [2] Figure 2. FFT of Single and Two Cell Rotating Stall [5]

2 Copyright © 2014 by ASME


INSTRUMENTATION APPROACH long term health monitoring and offer sufficient spatial
High frequency pressure measurements have been used resolution as well as long life in industrial turbines.
over the last 20 years to enable researchers to develop methods Bearing vibration measurements have been used on
for detecting and preventing rotating stall on compressor. These industrial gas turbines for years to aid in overall gas turbine
pressure sensors measure the sound emitted by the compressor. health monitoring. These sensors have been proven to detect
Rotating machinery generates sound through several issues with the bearings as well as identify mistuned rotors.
mechanisms. Pressure measurements may be used to detect Recent literature indicates that vibration measurements may be
compressor stall and its precursors but the literature has not useful for detection of precursors to compressor blade failures
conclusively demonstrated that pressure sensors can be used to in gas turbines [13]. It is not clear in the literature as to whether
deduce the presence of blade vibration or cracks in the presence the bearing vibration sensors can be used to detect actual events
of all the other potential noise sources. Given that compressor that cause blade vibration or only large blade vibration
stall may drive blade vibration, the pressure measurements immediately prior to a catastrophic failure. For example, as
which measure the stall may be used in conjunction with other will be shown in this paper, field test results show that bearing
sensors which measure blade vibration. vibrations pick up the largest amplitude rotating stall frequency
Over the last 30 years, tip timing and tip clearance (where the pressure amplitude is highest), but do not pick up
measurement systems have been actively researched [6][7][8]. the higher harmonic where blade vibrations are maximized.
These sensors offer potential for rotating blade vibration and Other sensors ranging from more traditional case
crack detection as well as potential for active clearance control accelerometers or acoustic emission14 have been identified.
both in the compressor and turbine sections. In regards to blade However due to budget and installation constraints only the
health monitoring, it has been proven that tip timing sensors pressure, tip timing, and bearing vibration sensors were used.
can be used to capture the vibration of the blades for different Traditional gas path sensors were installed on the unit however
modes of interest as well as monitor the actual displacement the sampling rate of the sensors prevented them from being
over time. The fundamental measurement made by these used for high frequency analysis. Some of these sensors such as
systems is arrival time. Very precise arrival times are measured compressor inlet temperature were used to identify variations
for each blade passing event on each sensor. From this data a from start to start.
one per revolution sensor, the time of a revolution is
established. Knowing the circumference, all the blade positions INSTRUMENTATION SETUP
are calculated relative to the revolution reference point. From To identify potential correlation between these sensors, an
the measured blade positions, zero values are established and industrial gas turbine unit was instrumented with dynamic
the deflection is calculated relative to each blades measured pressure sensors and tip timing sensors. Bearing vibrations
zero or equilibrium position. Because the blades are only were already installed on the unit as part of its standard
measured once per revolution on each sensor, the vibrations are instrumentation package. Multiple pressure sensor
aliased. Just as the movie camera aliases the rotation of a manufacturers were also installed to compare and contrast the
wagon wheel, it may appear to be turning backwards when the pressure signals.
wagon is moving forward. Numerous analysis methodologies In addition to understanding the correlation between
have been developed to handle the aliasing that results from the sensors, another motivation driving the instrumentation setup
synchronous, under sampled, data capture rate. For NSV FFT was to compare and contrast the capabilities of different sensor
based analysis tools are useful, but multi-sensor analysis or manufacturers, while also understanding how sensitive stage
analysis over a speed change is necessary to establish the true location is to the sensor’s ability to perform diagnostics.
frequency of the vibrations. Unfortunately, most resonant Pressure sensors were installed on the first 4 stages of the
responses are synchronous with rotation because the driver is compressor. Table 1 shows the locations of the pressure probes.
fixed to the case (vanes & struts). The synchronous vibrations PCB probes were placed at the first three stages. Kulite probes
coupled with a synchronous sample rate further complicate the were installed at stage 1 and 3 with Davidson probes installed
aliasing problem. Combining data from multiple sensors at stage 2. This experimental setup has the advantage of having
located around the circumference affectively increases the both PCB and Kulite sensors located at stage 1 where the tip
sample rate. The vibrations can still be aliased, however, by timing sensors are located. This will enable a direct comparison
using multiple unequally spaced sensors one can solve for the of these sensors to any trends noted in the tip timing probes.
correct frequency, amplitude, and phase of the vibrations. One disadvantage is that no common pressure sensors are
Numerous types of tip timing sensors have been developed and located at the same stage to provide circumferential comparison
tested through the years ranging from an eddy current, optical, of the pressure measurements. Figure 3 shows the PCB sensor
and microwave type[9][10][11][12]. Optical sensors were used installation. The analysis in this paper uses the stage 1 PCB
for this testing due to the constraints of the existing case holes. measurement versus the tip timing and bearing vibration. It
Optical sensors offer the highest spatial resolution but are prone must be noted, however, that similar conclusions were drawn
to signal loss due to the accumulation of contaminants. Non- using the other sensors but detailed comparisons of these lie
optical sensors such as eddy current probes are preferred for beyond the scope of this paper.

3 Copyright © 2014 by ASME


Table 1. Pressure Sensor Location
Stage Unit 1
1 PCB Kulite
2 PCB Davidson
3 PCB Kulite
4 PCB

Figure 4: Tip Timing Sensor Installation

BLADE VIBRATION RESULTS


The tip timing and pressure sensors were installed during
an outage in November 2012. Since becoming operational, nine
shutdown/startup cycles were observed in the following three
months. Dynamic blade deflections caused by synchronous and
non-synchronous drivers during the startups and shutdowns
were analyzed using the Agilis Arrival Time Analysis Software
(AATAS). There were eight synchronous resonant responses
Figure 3. PCB Sensor Installation on Stage 1-4 associated with six blade vibratory modes along with a mode 1
NSV. Figure 5 shows a typical startup Campbell plot showing
Tip timing sensors were installed circumferentially around the nine dynamic deflection responses.
the 1st stage of the gas turbine using existing holes in the unit.
Figure 4 shows a picture of the tip timing sensor installation on
the top of the gas turbine casing. In all, 8 sensors were used
spaced throughout the top half of the case.

Figure 5. Dynamic Deflections Measured During Shutdown #1

4 Copyright © 2014 by ASME


Table 2 summarizes the dynamic deflections. The mode 1 Table 3: Mode 1 NSV Maximum Deflection for Startups
NSV is the largest deflection, and the mode 2 6E is the largest and Shutdowns
synchronous response. The largest amplitudes of the Mode 2 6E Transient Event Max Deflection, mils p-p
occurred during off peak operation. There are three higher order Shutdown #1 350
vibration modes detected that were not correlated with a Startup #1 930
specific resonance mode and are denoted with “HO” in Table 2. Shutdown #2 490
The following sections will discuss the mode 1 vibration in Statup #2 500
more detail. Shutdown #3 403
Startup #3 850
Table 2. Dynamic Deflections Measured During Transients
Shutdown #4 266
Approximate
Startup #4 678
Deflection
Resonance Engine Magnitude, Shutdown #5 355
Mode Order mils p-p Startup #5 484
1 NSV 500-1000 Shutdown #6 382
1 3 150 Startup #6 470
2 7 85 Shutdown #7 444
2 6 30-300 Startup #7 509
3 13 20 Shutdown #8 371
3 44 25 Startup #8 859
HO 44 40 Shutdown #9 404
HO 44 25 Startup #9 495
HO 44 30 The peak vibrations occur when low frequency peaks are
There is a large mode 1 NSV seen during all startups and seen in both the pressure and bearing vibration data. Figure 6
shutdowns. Table 3 shows the peak deflections recording shows normalized low frequency Fourier transforms of stage 1
during all shutdown and startup events. During startups the high frequency pressure measurement and the compressor X
peak deflections most often appeared at around 500 mils for six bearing measurements. The pressure data shows a low peak at
of the nine startups. However, for three of the starts the peak around 0.1 and harmonic peaks up to typically the 8th harmonic.
deflection occurred at a larger value with two starts having peak This multiple low frequency peaks is indicative of rotating stall
deflections around 900 mils. The average deflection for the 9 where the harmonic peaks may indicate multiple stall cells
starts analyzed was 641 mils with a standard deviation of 190 similar to the stall in Figure 2. The bearing vibration shows one
mils. On average, shutdowns had a much lower peak magnitude peak around 0.1 and occasionally sees a higher order peak
as compared with startups with an average value of 385 mils. around 0.5 or 0.6. Looking at these figures, there appears
The standard variation during shutdowns was 62 mils. One significant correlation between all three sensors. The following
reason that the peak magnitude is lower on shutdowns is the gas section looks into multiple starts in more detail to understand
turbine is decelerating faster on shutdowns during the rotating where there is correlation.
stall than during startups and the blade spends less time at
resonant conditions. The magnitudes of the blade deflections
seen are not indicative of an impending failure [15].

Figure 6: Pressure and Vibration Fourier Magnitude Versus


RPM for Startup #2

5 Copyright © 2014 by ASME


SENSOR COMPARISON
Based on equation 1, for a -7ND resonant vibration, the
The pressure measurements capture the driver in the non-dimensional driver frequency is expected to be found in the
stationary reference frame and the tip timing measurements stationary reference plane around 0.8 to 0.875, a -6ND driver
capture the vibration in the rotating reference frame. To capture should appear around 0.69 to 0.75, and a -4 ND driver should
correlation between the sensors an equation needs to be defined appear around 0.45 to 0.5. Figure 8 shows Fourier analysis of
which relates the two measurements. Equation 1 relates the the pressure sensors at the four points of interest. For point A,
two measurements where the driver measured by the pressure there are only low frequency peaks of significant magnitude up
measurements in the stationary frame is equal to the sum of tip to around 0.4. There does not appear to be any peak in the 0.8-
timing measured vibration plus the nodal diameter multiplied 0.875 range. As the rpm is increased to point B, a couple
by the rotational speed [7]. different observations can be noted. First a large peak around
0.115 and 0.5 appears. Secondly a small peak around 0.85

  
   . .∗  (1) appears. Using equation 1, the frequencies of the observed
peaks at 0.5 and 0.85 in the pressure data match the observed
Three startups are analyzed to understand and confirm any frequency of the blade vibration. The 0.115 peak is the first
correlation between the tip timing, pressure, and bearing harmonic of the rotating stall and it is not seen in the blade
vibration. Based on the tip timing results, multiple points of vibration data. The large 0.5 peak is the -4 ND driver for the off
interest will be highlighted for further comparison. These points resonant blade vibration. The 0.85 peak is the driver for the
range from low vibration, smaller intermediate peaks, and both resonant vibration. The fact that the magnitude of the driver is
maximum blade vibrations at different nodal diameters. small indicates that not much energy from the stall is needed to
Figure 7 shows the tip timing Campbell plot focusing on excite the resonant mode. As the unit moves to point C which
the mode 1 NSV for startup #1 which had the largest maximum has the largest blade vibration, the 0.5 and 0.85 peaks remain
blade deflection of all the starts analyzed. There are four points while the 0.115 becomes smaller in magnitude. For large -7 ND
highlighted. Point A is a low deflection point. Point B is an resonance to occur at both point B and C that would imply the
intermediate deflection peak. Point C is maximum blade driver frequency has shifted to a slightly higher frequency.
deflection at -7ND, and point D is maximum blade vibration at Detailed inspection of the pressure driver does show a slight
-6ND. This start has a non-resonant vibration at -4 ND as well increase in frequency over the range of rpm. As the rpm
that occurs during points B and C. increases to point D, the -6ND blade vibration should be driven
by a 0.7 driver. This is noted in the point D pressure plot where
there is a peak at 0.7.

Figure 7. Startup #1 Mode 1 NSV With Points of Comparison

6 Copyright © 2014 by ASME


Figure 9. Startup #1 Compressor X Bearing Vibration Low
Figure 8. Startup #1 PCB1 Low Frequency FFT at Tip Frequency FFT at Tip Timing Points of Interest
Timing Points of Interest
The same three plots for startup #2 are shown in Figure 10
A similar comparison can be performed with the bearing through Figure 12. This start was one day later than the
vibration measurements. Figure 9 shows Fourier analysis for previous start and had peak blade deflections of about 500 mil
the compressor X bearing vibration measurement. The large p-p as compared to around 900 mils p-p of the previous start.
peak at 0.2 is the first engine order and a large magnitude peak The ambient temperature of the two starts was similar. The
is expected for this measurement. At point A there is a medium most obvious difference between the two starts was startup #1
sized peak at 0.115. At point B, the 0.115 peak becomes larger. was after a 10 day outage and startup #2 was approximately 6
At points C and D, this peak becomes smaller. There appears to hours after shutting down from full speed. This may be a
be correlation of bearing vibration with the presence of a 0.115 contributor to the differences in the two starts, but cannot be
peak in the pressure data which may excite the bearing first conclusively confirmed until more cold starts are analyzed.
critical mode. There does not appear to be a strong correlation
with blade and bearing vibration. This will become more
obvious when looking at the other startups.

Figure 10. Startup #2 Mode 1 NSV with Points of Comparison

7 Copyright © 2014 by ASME


The most obvious trend in startup #2 is the gap in blade The blade vibrations show similar trends to the previous
vibrations centered on point C. As discussed previously, the start. The 0.115 peak is present in points A, B, and D and is
gap occurs because vibrations only occur when the resonant absent in point C when the 0.115 peak disappears in the
driver matches blade resonance nodal diameter. For the pressure analysis. When looking at the Fourier plots of the
previous start, the vibrations occurred across a range of rpm pressure and vibration data plotting against rpm in Figure 6,
because it appeared the driver frequency was moving slightly there is correlation between the presence of the 0.115 in the
(from approximately 0.115 to 0.1175) as the unit accelerated. pressure data and the 0.115 peak in the vibration data. There
Point A is a point early in the -7ND mode 1 NSV which has a also appears to be some correlation between the 0.6 peak in the
relatively small amount of blade deflection. Point B is the point pressure and bearing vibration data.
which has the largest -7 ND blade vibration. Point C is a point
in between the -6ND and -7ND resonant blade vibration. And
point D is a -6ND blade vibration point.
Looking at points A and B in Figure 11 relative to the
others, there appears to be more noise in the relevant 0.8-0.9
range. This range is where a driver that causes resonance would
appear. For point A, there is not a clear peak in this region;
however, there may be a peak around 0.8 which may drive a
small amount of resonance. There is a pronounced peak at
around 0.81 at point B which causes the magnitude of the
resonant vibration to peak. At point C, the only noticeable peak
is the 0.6 peak which according to equation 1 drives the
approximate off-resonant 0.5 blade vibration. As the unit
accelerates to point D, the -6ND driver at 0.7 drives the mode 1
resonant blade vibrations.

Figure 12. Startup #2 Compressor X Bearing Vibration Low


Frequency FFT at Tip Timing Points of Interest

Figure 11. Startup #2 PCB1 Low Frequency FFT at Tip


Timing Points of Interest

8 Copyright © 2014 by ASME


Figure 13. Startup #4 Mode 1 NSV With Points of Comparison

The last start analyzed for detailed comparison of the three


sensors is startup #4 since startup #3 did not have pressure or
vibration data available. This startup had peak blade deflections
around 678 mils p-p. Figure 14 shows four points of interest
from the mode 1 NSV. Point A is a low magnitude -7ND
vibration. Point B is a higher magnitude peak. Point C is at the
largest vibrations. Point D is at smaller magnitude -7ND
vibrations. These plots will show how the pressure peaks
change during the -7ND resonant vibration.
Points A and B in Figure 14 indicate a fair amount of noise
in the 0.8 to 0.9 area of interest. Both plots may indicate a very
small peak at around 0.8. At point C where blade vibrations are
largest, a peak at 0.82 is more pronounced. At point D, a peak
around 0.8 is very pronounced, however according to equation
1, the driver frequency should be approximately 0.835 to drive
resonance. Further detailed analysis of more starts is needed to
see if this discrepancy occurs more frequently. The 0.475 peak
in points B, C, and D are the drivers for the off resonance peak
in the blade vibrations.
Figure 14. Startup #4 PCB1 Low Frequency FFT at Tip
Timing Points of Interest

When the 0.115 peaks are present in points A and B of the


pressure data, there are 0.115 peaks in the bearing vibration
data shown in Figure 15. There are no 0.115 peaks for points C
and D when the 0.115 peak is small or non-existent. At the
maximum blade vibration point C, the 0.115 peak is not present
which would indicate that there is no correlation between blade
vibration and bearing vibrations.

9 Copyright © 2014 by ASME


highlighted in this figure. The first is the -7 ND vibration region
and the second is a -6 ND vibration region. If the appropriate
driver harmonic appears in either of these regions, resonant will
occur. However, if that harmonic is not present at that rpm,
there will be no resonance.

Figure 16. Mode 1 NSV Regions of Interest


Figure 15. Startup #4 Compressor X Bearing Vibrations
SUMMARY AND CONCLUSIONS
Low Frequency FFT at Tip Timing Points of Interest
This study examines and quantifies the potential benefits in
various compressor measurement technologies. A variety of
SENSOR CORRELATION dynamic data was acquired from an operating engine, including
During all the analysis, correlation between the sensors acoustic pressure, bearing vibration, tip timing, and traditional
was noted. This correlation primarily appeared during the gas path measurements. The acoustic pressure was installed on
startup and shutdown sequences when the rotating stall driver the first four compressor stages, while the tip timing was
appeared. The correlation was strongest between the tip timing installed on the first stage only. These data show the presence
and pressure measurements. Correlation can also be seen of rotating stall instabilities in the front stages of the
between the pressure measurements and bearing vibration compressor, occurring during every startup and shutdown, and
measurements. Higher harmonics of the rotating stall appear manifesting itself as large amplitude oscillations in the dynamic
because of the stall’s complex nature. For the test points pressure measurements which drive deflections in the blades.
analyzed, typically up to the 8th harmonic was noted. For The blade vibration is primarily driven by higher order
resonance to appear the frequency of the driver must be in harmonics of the rotating stall. The vibration of the blades is the
phase with the vibration of the bladed disk assembly. Tip timing largest when the higher order pressure oscillations match the
measurements were able to measure the nodal diameter of both nodal diameter at the resonant frequency. The pressure
the driver and blade vibration using all the blade arrival oscillations are also manifested in bearing vibrations with the
measurements. When the harmonic of the driver matches the bearing vibrations primarily driven by the large first harmonic
nodal diameter of the bladed disk resonant frequency, large of the rotating stall. However the bearing vibration peaks do not
displacement vibration can be observed. correlate with blade vibration. The data that lead to these
Based on the pressure data, the main driver frequency observations were acquired during several startup and
varies from approximately 0.115 to 0.125 with up to the 8th shutdown events, and clearly show that the amplitude of these
harmonic which is approximately 1. According to the tip timing instabilities and the rpm at which they occur can vary
data, the first mode appeared at about 0.65. Knowing the rpm substantially.
region where the rotating stall occur, mode 1 resonant vibration Future work is focused on collecting data with the existing
regions of interest can be defined as shown in Figure 16. In this sensors to further develop understanding of the operational
figure, the diagonal blue line represents the frequency which drivers which lead to variation in blade vibration during each
would have to excite a -6ND resonance at 0.65. The red line transient. Currently nine sets of startup and shutdowns have
represents similar for a -7 ND resonance. There are six been analyzed with the ambient temperature ranging from 50 F
horizontal lines on the figure. The top three represent the 6th to 80 F. Of those starts only one is a cold start. It has been
harmonic of a 0.115, 0.12, and 0.125 driver, respectively. The observed that the rotating stall is a function of corrected speed
same holds true for the bottom three of the 7th harmonic of the and thus the rpm which it occurs is a function of ambient
driver. When the 6th harmonic of the driver intersects the -6 ND temperature. Collecting more starts during the fall/winter
vibration line, which would indicate that there is a risk of season will help refine any observed correlations. In addition
resonant vibration at this rpm. Two regions of interest are other future work may focus on instrumenting other similar

10 Copyright © 2014 by ASME


units to capture variance seen from unit to unit. Lastly
instrumenting units with other sensors such as case Stator Vane Cracks”, Proceedings SPIE 8690, Industrial
accelerometers or acoustic emission to understand potential and Commercial Applications of Smart Structures
correlations could be done. Technologies, 2013.
[15] Scheibel, J., et.al., “Gas Turbine Compressor
Dependability and Risk Mitigation Measures”,
REFERENCES Proceedings of ASME Turbo Expo 2013, GT2013-95143,
San Antonio, TX, 2013.
[1] Vo, H., “Role of Tip Clearance Flow in Rotating
Instabilities and Nonsynchronous Vibrations,” Journal of
Propulsion and Power, Volume 26, 2010.
[2] Kielb, R., Hall, K., Spiker, M., and Thomas J., “Non-
Synchronous Vibration of Turbomachinery Airfoils,”
AFRL-SR-AR-TR-06-0269, 2006.
[3] Meher-Homji, C., and Gabrilles, “Gas Turbine Blade
Failures – Causes, Avoidance, and Troubleshooting”,
Proceedings of the 27th Turbomachinery Symposium,
1998.
[4] Cumpsty, N., “Compressor Aerodynamics”, Longman
Scientific, and Technical, 1989.
[5] Poensgen, C. and Gallus, H., “Rotating Stall in a Single-
Stage Axial Flow Compressor”, Journal of
Turbomachinery, Volume 118, 1996.
[6] von Flowtow, A., Mercadal, M., and Tappert, P., “Health
Monitoring and Prognostics of Blades and Disks with
Blade Tip Sensors,” IEEE Aerospace Conference
Proceedings, 2000.
[7] Heath, S., “A Survey of Blade Tip-Timing Measurement
Techniques for Turbomachinery Vibration,“ Journal of
Engineering for Gas Turbines and Power, Volume 120,
1998.
[8] Bhattacharya, A., et.al, “System to Monitor Blade Health
in Axial Flow Compressors,” IEEE Conference on
Prognostics and Health Management, 2011.
[9] Cardwell, D., Chana, K., and Russhard, P., “The Use of
Eddy Current Sensors for the Measurement of Rotor
Blade Tip Timing – Sensor Development and Engine
Testing,” Proceedings of ASME Turbo Expo 2008,
GT2008-50792, Berlin, Gernany, 2008.
[10] Lawson, C., and Ivey, P., “Compressor Blade Tip Timing
Using Capacitance Tip Clearance Probes,” Proceedings of
ASME Turbo Expo 2003, GT2003-38284, Atlanta, GA,
2003.
[11] Zielinski, M., and Ziller, G., “Noncontact Vibration
Measurements on Compressor Rotor Blades,”
Measurement Science and Technology, Volume 11, 2000.
[12] Hafner, M., Holst, T., and Billington, S., “Blade Tip
Measurement Advanced Visualization using a Three
Dimensional Representation,” Proceedings of ASME
Turbo Expo 2011, GT2011-46523, Vancouver, CA, 2011.
[13] Diallo, O., et.al., “A Data Analytics Approach to Failure
Precursor Detection of Gas Turbine”, Proceedings of
ASME Turbo Expo 2011, GT2011-46019, Vancouver,
CA, 2011.
[14] Momeni, S., et. al., “Online Acoustic Emission
Monitoring of Combustion Turbines for Compressor

11 Copyright © 2014 by ASME

View publication stats

You might also like