Professional Documents
Culture Documents
by
Frank Kushner
Senior Consulting Engineer
Eltiott Company
Jeannette, Pennsylvania
LinMag
30
mPSI
/di v
n A
peak discharge pressure, and was found to be purely axial, similar -400
to a mild part volume surge with zero circumferential phase angle. mpai
OS 15.9844s
Amplitudes were sufficiently high and electrically noise free that
besides giving incipient surge points also provided an additional Figure 4. Time-Trace Identifying Large Pressure Change at
alarm signal for the control room. Eddy current probe and Inducer and That Upstream of Open Impeller.
accelerometer monitors were modified in turn, including band pass
filters.
REINJECTION COMPRESSOR IMPELLER TESTING
Shop verification tests also involved checking the inlet just
upstream of the inducer for instantaneous circumferential static An impeller design with extensive experience produced subsyn
pressure variations, in about 20 degree increments. Even without chronous response at a frequency varying from 70 to 80 percent of
any blade-damaging type of rotating stall being found for the rotating speed. This occurred only when approaching the surge
original and redesigned stronger impeller, the concern was that the fl ow points, similar to that described by Fulton and B lair [ 1] .
discharge volute caused a serious maldistribution of loading at the Difference evaluation did not show the cause, s o there was impetus
compliant area of blading. Data shown in Table 1 verified that up to do a special rig test. Moreover, this was the first time the manu
to and at the incipient surge point, Fourier harmonic analysis gave facturer experienced subsynchronous vibration on a unit with these
negligible excitation in the frequency area of concern, above three characteristics . In fact, there were also never any known cases of
DYNAMIC DATA ANALYSIS OF COMPRESSOR ROTATING STALL 73
psi
at rotor/bearing first critical frequency was extremely low, so that OHz AVG. 100 3"2kHz
LinMag
20
mpsi
/div
II
500 3 76. 81
Figure 9. Three-Cell Stall Time-Trace at Inlet of Reinjection A: INLET P ROBE X: 266 Hz Y: .BIU 6 mD li
B. B
Impeller; Probes 60 Degrees apart. mps i
LinMag
AT INITIATION OF STALL, WITH SOME B.P.F. BOO
PULSATIONS AT 1B4 HZ, 135 DEG. PHASE ANGLE ups i II
/div Ill II 1\
-600 Rig tests could not produce one-cell stall by changing size or
mps1 rate of discharge valve closure, nor did variations in diffuser width,
OS 31. 2195ms
eye seal, gas composition, or pressure. A last test with a vaned
Figure 10. Three-Cell Stall Time-Trace at Diffuser of Reinjection diffuser design eliminated impeller rotating stall, but gave an unac
Impeller; Probes 45 Degrees apart. ceptable flow range.
DYNAMIC DATA ANALYSIS OF COMPRESSOR ROTATING STALL 75
Incidence angle changes at the impeller inlet analytically forced angles at the design flow point. W hen off design, the vanes would
rotating stall to a much lower fl ow, below that of previous experi be at a different angle than the fl ow, but effects on efficiency were
ence with actual surge . Gas production plans for the compressor expected to be small, since there were only two vanes, 1 80 degrees
did not require rig confirmation, as critical-path delivery of new apart. Retesting verified efficiency changes at normal fl ows were
impellers was almost complete for commi ssioning. In addition, it small, but effect on rotating stall was dramatic. One-cell impeller
still could have been two-cell or three-cell stall in the rig. Surge stall at 65 to 70 percent of running speed was eliminated, along
and rotating stall occurred in the field just about simultaneously with rotor subsynchronous vibration. At low flow, stall frequency
only during control line definition with rotor vibration reduced to near 80 Hz was mild. A waterfall plot is shown in Figure 1 3 , where
acceptable levels, precluding bearing or other damage. two cells were verified by phase angle measurement, giving a prop
agating speed near 30 percent of rotor speed.
MID-FLOW COVERED IMPELLER RIG TEST
NITROGEN - 7800 RPM
Following design, a rig confirmation test was done on an STALL INITIATION AT PREDOMINANT 80 Hz
impeller with a change in blade profiles to obtain efficiency
improvements. Again dynamic pressure probes were used in both
the inlet and diffuser to assist in surge/stall definition. The same rig
was utilized as for the reinjection impeller test described above, so
that small nonuniform flow from stationary components should
give similar effects. At tip speeds equivalent to operation with air
below 1 000 ft/sec, a one-cell rotating stall was found prior to surge.
This stall was confirmed by CFD analysis to be due to the impeller.
Near 1 000 ft/sec equivalent speed, the same stall propagating at 65
percent of rotor speed was quickly followed by a diffuser stall,
traveling near three percent speed, but with two cells. Rotor
vibration shown in bottom trace of Figure 12 only increased for the
one-cell stall (at 50 Hz), but not for the two-cell stall (at 4.5 Hz) .
Away from stall, pulsation at once per revolution was predominant,
giving an indication of one-cell tendency. This is in sharp contrast
to the previous rig test for the reinjection compressor which did not
�50�H�z----���
-
619 . 995
mSec
have one-cell stall. As rotor speed increased, only a strong two-cell
diffuser stall without rotor vibration was found. The frequency
increased as flow was reduced, contrary to that for impeller stall. Figure I3. Initiation of Stall for Mid-Flow Impeller with Stabilizer
Vanes.
RS34A - 4!580 RPM
POINT 32
MIXED FLOW IMPELLER - UEQ - 900 FT/SEC MIXED FLOW IMPELLER - UEQ � 900 FT/SEC
ONE CELL ROTATING STALL @ 77% DF ROT. SPEED ONE CELL ROTATING STALL @ 214 HZ
LinMag \nl
,Nf'l II\ I
Real
100 200
v' 1
mpsi mps l -� .J>f" ' � ill
/div /di v "\ /1,, '�Ill 1/\\. "" 1.!' VI(
0 1
psi psi
5Hz AVG. 5 405Hz Os 15.6097ms
A: DI FF. �:===�=-X Y: 7 9
4 _: J-�-·--_-_-· 1-::
-·-,-·.J:g
- -- - ._ t-----3- 9l1_
y--f---t- ---t
- j- . __
mo/-s·i--J
mpsi" --
�---t---4--- ��-t----+---1----t-- +---�---l --
IV'\.
v
-
f--· r--
pattern. In addition, the once per revolution pressure pulsation was
still dominant;. implying that the impeller effects were controlling. 0 -:::---
psil.,-- -
Rotor vibration was also greatly reduced, comparing Figure 2 1 to OHz AVG: 23 B O O Hz
Figure 1 4 . When stall frequency occmred, it was predominantly on
only part of the 10 to 1 1 Hz, low frequency surge pulse cycle as Figure 18. Normal Spectra at Diffuser and Inlet.
DYNAMIC DATA ANALYSIS OF COMPRESSOR ROTATING STALL 77
MIXED FLOW IMPELLER : ONE-CELL STALL MIXED FLOW IMPELLER DURING SURGE
INITIATION @ 2.79 SECS.; DECAY @ 3.15 SECS. TWO SPLITTER VANES - UEG • B O O FT/SEC
Real
LinMag
-50 \) 30
mmilpp
3.245 /div
Sec
0 b&_
milpp ""- ,..,._.. .......
5Hz AVG. 110 805Hz
B: INLET 8 X: LSi Hz Y· 9 .d7i" mn ,;
400
mps 1 Hz
�
0
2.75 ps 1
SecL ------------------------------ � 5Hz AVG. 110 805Hz
--- �� - � � 7�
1 24 . 8 7 9�m s
05
AXIAL COMPRESSOR SHOP TEST, SURGE AND STALL AXIAL COMPRESSOR FREQUENCY SPECTRUM !! SURGE
WITH ORIGINAL OPEN AO�USTABLE VANE SETTING ROTATING STALL �UST BEFORE FLOW REVERSAL
Real
5
psi ·vv
1'\
/div .,....
"\. ....... LinMag
200 B.P.F,
-25 mpsi
psi /div
Os 1.99Bs
B: ROW 1!! on:r.o"'
SfAU.
0
PSl l,
"""'\
Real
2.5 J.
psi .h
/div ...,. A 'JV'Y"'.
v� io\1\ AA LilA A. vvv· ·v
-25 0 \ ./ h.. Jl
psi psi r-
OS 499. 5117ms OHZ AVG: 111 3.2kHZ
Figure 23. Time-Trace at First Row During Surge and Recovery. Figure 25 . Spectra Showing One-Cell Pulsation at Middle Row.
....
-15 0
psi psi
Os 7.9922s OHZ AVG. 10 3.2kHZ
Figure 24. Time-Traces at First and Last Rows with Multiple Figure 26. Normal Spectra at Front and Rear Rows.
Surges.
AXIAL COMPRESSOR FREQUENCY SPECTRUM
WITH ACOUSTIC PULSATION FROM DISCHARG E PIPE
trace of rotating stall could be found as shown in Figure 25 for
open vane settings. Even though this was one fell, it was not detri
A: ROW 1 x· 376 kH Y· 766 )2 DS
mental, partly due to low sensitivity of the rotor/bearing system. 2 ..
Harmonics of stall frequency were also negligible in the range of psi
frequencies that could excite blade natural frequencies. Away from LinMag
surge points, pulsations were only at normal blade pass frequen 200
mpsi
cies, Figure 26, and stall point settings gave optimum operational /div
control line data. R
0 II
An added value of dynamic pressure probes was that commis psi
sioning also showed an acoustic resonance. It was sensed as well OHz AVG. 10 3.2kHz
B: ROW 14 X: 432 Hz Y· 256 5 3 mo ;i
with high levels at the discharge pipe and was audible, so that there 3
ps i
was concern that it could be propagating upstream at a level that
could excite the blades. Data verified frequency changed with LinMag B.P.F
speed of sound. Levels at the last rotating row plotted in Figure 27 300
mps i
showed that it was a standing acoustic wave at 432 Hz and could /div
not cause blade excitation for the calculated number of waves.
0
Amplitudes were much lower for the other probe in the last row, ps i
OHz AVG: 10 3.2kHz
showing that a standing acoustic wave is the case , rather than some
odd rotating stall. Figure 27. Spectra with 432 Hz Noise at Last Row.
DISCUSSION OF SEVERITY AND STALL MffiGATION
Relative to blade loading affects on fatigue, pulsation measure calculations can be generated to set limits on nonsynchronous rotor
ments can give values to be used for forced response analysis. vibration; as it is generally a forced, stable mechanism. For a given
Rotor vibration examples could also be used to determine whether element, severity should be a function of flow raised to the third
DYNAMIC DATA ANALYSIS OF COMPRESSOR ROTATING STALL 79
power; but for the same equivalent tip speed, would increase f8 first harmonic of stall frequency measured at a stationary
directly with power. For example, a reduced pressure performance point, Hz
test at correct equivalent tip speed showing a minor amplitude of fr first harmonic of stall frequency relative to each rotating
0.3 mil) p-p vibration could be as high as 1 .2 mil)p- for four times blade, Hz
t
the flow at full field conditions, and still be accepta le if flow is to 'i>s measured phase angle between stall cells , degrees
the left of the control line. Also, to be factored in would be the cl>m installed angle between stationary probes, degrees
frequency experienced relative to the rotor/bearing system natural a tolerance on installed angle between probes, degrees
frequency. If test conditions give a frequency much different than � tolerance on phase angle measurement, degrees
proj ected field conditions, dynamic stiffness_ would also be used to
adjust calculations. APPENDIX
A pattern from centrifugal impeller testing emerged in that one Aspects of Rotating Stall and Its Measurement
cell rotating stall correlated with indication that one per rev
Compressor surge is the well known system instability succinct
pulsation was higher than other harmonics.
ly described by Den Hartog [ 1 0] , and later by Japikse [ 1 1].
Thus, a potential method is planned for testing wherein a
Sometimes a more localized phenomenon, rotating stall, can be
multiple cell structure can be forced to occur. Hopefully, this idea,
found just before or simultaneously with surge. Generally, stall is
used for patent application, will compliment the use of stabilizer
at the point where pressure rise peaks in one or more components
vanes and diffuser vanes.
and starts to give a positive slope on the system pressure vs flow
CONCLUSIONS FROM TESTING curve. A sketch to indicate cells of low pressure alternating
circurnferentially is shown in Figure 28 with those cells higher
• Measurements using dynamic pressure probes are useful in iden than the average pressure. These revolve at £08 , some fraction of
tifying the true nature of stall inception, propagation and effects on rotor speed ror, almost always in the same direction from a station
system surge definition. Other flow and pressure instrumentation ary viewpoint. Relative to each rotating blade, the cells are
should be used in conjunction with audible noise, sensing of propagating backwards at a speed of (ror -£08) . Thus sections of
vibration, speed fluctuations, and load changes at peak head, etc. higher flow in tum provide more flow into blade areas that just had
• A volute stage can be properly designed to give negligible alter experienced a low flow condition. A steady state with this pattern
nating blade loading on highly-loaded open inducer impellers, can exist prior to in-phase reversal (surge). Each blade experiences
without damaging static pressure nonuniformity away from stall. a peak pressure at a rate of n multiplied by (ror - £08); but any
one-cell condition would give a net fluctuating force on the rotor at
• Data show that centrifugal impeller stall frequency (propagating
frequency of f8•
speed) decreased as fl ow is reduced, opposite to that for diffuser
For stationary. measurements in a casing near the rotating
stall. Future tests should eventually better identify cause and effect
component (s), dynamic probes can be used to determine number
relative to sources giving improved designs and easier modifica
of cells and relative amplitudes , including harmonics for
tions for problem-solving.
waveforms, not purely sinusoidal. Usually, two probes are used
•Additional advances in CFD analytical techniques can be bench with one giving a circumferential reference for phase angle mea
marked using dynamic pressure probe data. surements. If only one probe is used, a frequency of f8 is obtained,
• Use of diffuser stabilizer vanes can eliminate concern for one but the number of cells and their propagating speed can be in
cell rotating stall caused by centrifugal impellers, and give question.
efficiency improvements for volutes. Normal diffuser vane designs A phase angle measurement value near zero would show an
could also be used to change characteristics, generally only if flow axial pulsation, which would indicate a local or full surge type flow
range limitation is not a severe detriment. reversal. Usually a phase angle of± 1 0 degrees for the second
probe can be easily detected even for low amplitudes using
• Data for centrifugal compressors also showed that predominate frequency response functions (FRFs) from two-channel fast
one per revolution pulsations may be an indicator of propensity for Fourier transform (FFf) analy ses. Typical projected results are
one-cell, in lieu of multiple-cell rotating stall. Only one-cell stall shown in Table 3 for selected separation angles of the two probes.
can give significant unbalanced loads to cause rotor vibration. A separation much less than 4 5 degrees could give questionable
• Surge tests, especially for the axial compressor showed extreme data in that pulsation frequency could really be surge, near zero
pulsation levels, once again highlighting the concern for limiting degrees. With consideration of accuracy, 45 degree probe separa
occurrences. Adjusting discharge valve with dynamic pressure tion, installed to ± 2.5 degrees, can determine one to five cells, as
probes for surge point definition can be useful in defining true shown for Case B in Table 4 using Equation ( 1 ) .
setpoints for automatic controls , minimizing number of severe
(1)
surges . Setpoint data are best quickly logged and flow increased to
remove any hysterisis.
Other factors t o consider:
• A procedure such as described for rotating stall severity calcula
Measured amplitudes should also be about the same for each
tions could save the industry from performing unnecessary full
•
NOMENCLATURE (3)
rotor operating speed, Hz
stall propagating speed in direction of rotor speed, Hz Examples 1 and 5 in Table 5 show that there is some small risk
number of rotating cells that instead of one cell, there are nine cells . For more accuracy, a
80 PROCEEDINGS OF THE TWENTY-FIFTH TURBOMACHINERY SYMPOSIUM
ONE-CELL STALL Table 3. Stall Number Determination for Separation Angle of Two
Probes.
Implied Nnmber
Probe Separation Measured Phase Anlde , of Cells
(degrees) (degrees)
90 90 1,5,9,...
180 2,6,10,
0
••.
360 4,8,12,...
60 [60] [1],7,...
120 2,8,...
180 3,9, ...
240 4,10,...
300 5,11,...
360 6,12,...
45 [45] [1],9,...
90 2,10,...
135 3,11,...
180 4,12,...
225 5, 13,...
TWO-CELL STALL 270 6,14,...
315 7,15,...
360 8,16,...
0
� CaseB
350-10 350-10 0
30-60 32.5.57.5 1
70-110 75-105 2
110-160 117.5.152.5 3
150 • Z10 160.200 4
190.260 202.5-247.5 5
230.310 245.295 6
270-0 287.5 • 342.5 7
310.5 0 ' 330.30 8
350-100 12.5.77.5 9
30-150 55-125 10
Figure 28. Sketch for One and Two Rotating Stall Cells. Case A. a= 5•; � = 10 Degrees
Case J1 a= 2.5"; � = 10 Degrees
third probe spaced 60 degrees from the reference probe gives a
match if it shows near 60 degrees measurement needed for one Examples 3 and 5 in Table 5 could show concern for an axial
cell; see bracketed values in Table 3 . Nine stall cells would require compressor or open (non-covered) inducer type impeller, e.g., the
this probe to indicate 1 80 degrees. However, rotor vibration also frequency relative to blading for Example 3 is approaching three
found during stall would add to certainty that there is only one stall per revolution; the second harmonic at 300 Hz could especially be
cell, which gives unbalanced forces on the rotor as shown in Figure near a fundamental blade mode. Excitation at rotating stall could
28 at stall frequency, f8• Likewise, if only one probe is used and the add to other sources, especially critical if the blade mode is also at
frequency is the same as rotor subsynchronous vibration, one cell a harmonic of rotor speed. Again, more probes at different spacing
can also be safely implied. could be necessary for convincing analysis.
DYNAMIC DATA ANALYSIS OF COMPRESSOR ROTATING STALL 81
Table 5. Examples of Rotating Stall Analysis. 6. Lynghjem, A., Svendsen, 0., and Underbakke , H., 'The Offshore
Application of a Dual-Mode Inj ection Centrifugal Compressor
Two Probes 45 Degrees Apart and Improvements to Rotating Stall,'' ASME 96-GT-322
( 1 996).
Example No. 1 2 3 4 5 7. Ariga, 1., Masuda, S., and Ookita, A., "Inducer Stall in a
Centrifugal Compressor With Inlet Distortion," ASME 86-GT-
Rotor Operating Speed -Hz 60 60 60 60 60 1 39 ( 1 986).
8 . Van Den Braembussche, R., "Surge and Stall In Centrifugal
Measured Frequency -Hz 45 84 30 6 45
Compressors," VKI Lecture S eries 1 984-07-Flow In
Centrifugal Compressors, von Karman Institute for Fluid
Measured Phase Angle -Degs. 45 90 135 45 45 Dynamics, Rhode Saint Genese, B elgium (May 1 984).