Professional Documents
Culture Documents
MST-V 6-6-X
Technical Manual
These instructions apply to:
Material 107-7309
Copyright 2017
Belimed AG
Grienbachstrasse 11
6300 Zug, Switzerland
+41 41 449 78 88
+41 41 449 78 89
info@belimed.com | www.belimed.com
Steam sterilizer MST-V 6-6-X | © Belimed
Content
1 Introduction 8
1.1 Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.1.1 Product and type designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.1.2 Type plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.1.2.1 Type plate of sterilizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.1.2.2 Pressure equipment type plate of sterilizer. . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Change and copyright . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.5 Contact person . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.6 Symbols and notices used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 Safety 12
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.1.1 Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.3 Personnel requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 General safety instructions for maintenance . . . . . . . . . . . . . . . . . . . . . . . 12
2.5 Summary of the safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.6 Signs on the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 Description 17
3.1 Overview of the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2 Operating elements and display elements . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.1 Location of main switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.2 Location of utility shut-off valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.3 Pressure indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.4 Control panel on loading side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.2.4.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.2.4.2 Status bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.2.4.3 Communication bar. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.2.4.4 Activity bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.2.5 Display when a program is running . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.2.6 Display of the current readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.2.7 Operating panel on unloading side (2-door) . . . . . . . . . . . . . . . . . . . . . . . 28
3.2.7.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.2.7.2 Status bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.2.7.3 Activity bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.2.8 Process status indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.3 Automation unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.3.1 Batch cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.3.2 Transport cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4 Utilities 34
4.1 Utility tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.1.1 Model 6-6-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.1.2 Model 6-6-9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.1.3 Model 6-6-12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.1.4 Model 6-6-18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.2 Quality of utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.1 Cold water quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.2 Compressed air quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.3 Feed water quality for steam generator (option) . . . . . . . . . . . . . . . . . . . . . 40
4.2.4 Steam quality, external steam supply (option) . . . . . . . . . . . . . . . . . . . . . . 40
4.2.5 Cooling water quality in systems with cooling circuit
(option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5 Procedure description 42
5.1 Predefined programs and parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.2 Changing the program parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3 Pressure-time diagrams of the standard programs . . . . . . . . . . . . . . . . . . . . 42
5.3.1 Legend for the procedure phases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3.2 Abbreviations for procedure designations. . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3.3 134 °C Program for instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.3.4 134°C Programs for textiles and utensils . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.3.5 121 °C Rubber program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.3.6 134 °C warm up program with leak test . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.3.7 134 °C Bowie-Dick Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.3.8 Leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3.9 134 °C Program for heavy instruments . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.3.10 125°C porous items program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.3.11 121°C liquids program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.3.12 134° C Warm up program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
5.3.13 134 °C Bowie Dick function test with Integrated Bowie-Dick
Test (option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7 Options 62
7.1 Automation unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.1.1 Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.1.2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.2 Emergency cooling with cold water . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8 Errors 66
8.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.2 Other errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
9 Maintenance 82
9.1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
9.2 Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
9.2.1 Defining maintenance intervals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
9.3 Daily maintenance and tests (prior to start of
sterilization). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
9.4 Weekly maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
9.5 Periodic check of the pressure vessel . . . . . . . . . . . . . . . . . . . . . . . . . . 84
9.6 Maintenance plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
10 Servicing instructions 87
10.1 Maintenance work on the sterilizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.1 Checking the door drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.2 Checking the piping system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.3 Check filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.4 Checking the function of the ventilators . . . . . . . . . . . . . . . . . . . . . . . . . 87
10.1.5 Replacing the door seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
10.1.5.1 Replacing the door seal on the loading side . . . . . . . . . . . . . . . . . . . . . . . 88
10.1.5.2 Replacing the door seal on the unloading side. . . . . . . . . . . . . . . . . . . . . . 89
10.1.6 Cleaning the components in the piping. . . . . . . . . . . . . . . . . . . . . . . . . . 91
10.1.7 Replacing the ventilation filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
10.1.8 Checking the function of the valves . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
10.1.9 Checking the function of the vacuum pump . . . . . . . . . . . . . . . . . . . . . . . 92
10.1.10 Checking the function of the manometers . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1.11 Checking the function of the throttle non-return valves . . . . . . . . . . . . . . . . . 93
10.1.12 Testing the steam quality. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1.12.1 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1.12.2 Test of the non-condensable gases . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.1.13 Replacing capsule element of condensate drains . . . . . . . . . . . . . . . . . . . . 96
10.1.14 Switch cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
10.1.15 Exchanging PLC buffer battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
10.1.16 Integrated Bowie-Dick function test (option) . . . . . . . . . . . . . . . . . . . . . . 100
10.1.17 Maintenance of test device for integrated Bowie-Dick test
(option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
10.1.18 Scanner (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
10.1.19 Operate test program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
10.2 Checking the safety devices of the sterilizer . . . . . . . . . . . . . . . . . . . . . . 101
10.2.1 Checking the Emergency stop button . . . . . . . . . . . . . . . . . . . . . . . . . 101
Appendix
Appendix 1: P&I diagram 6-6-6/9/12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Appendix 2: P&I diagram, 6-6-6/9/12, with cooling water switch-over
(option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Appendix 3: P&I diagram, 6-6-18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Appendix 4: P&I diagram, 6-6-18, with cooling water switch-over
(option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Appendix 5: P&I diagram with external steam switch-over (option). . . . . . . . . . . . . . . . . . . . . . . . 131
Appendix 6: Spare parts drawing - sterilizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Appendix 7: Spare parts drawing - transport cart, 6-0-6/9/12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Appendix 8: Spare parts drawing - batch cart, 6-0-6/9/12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Appendix 9: Spare parts drawing for Integrated Bowie-Dick Test
(option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
Appendix 10: Spare parts drawing for upgrade kit TS 09 (option) . . . . . . . . . . . . . . . . . . . . . . . . . 136
Appendix 11: Spare parts drawing for extendable shelf (option) . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Appendix 12: Spare parts drawing for foot switch (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Appendix 13: Spare parts drawing for scanner (option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Appendix 14: Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
1 Introduction
1.1 Product identification
Belimed AG
Grienbachstrasse 11
6300 Zug, Switzerland
YYYY-MM-DD
1.4 Warranty
The product must only be installed and commissioned by personnel employed by Belimed AG or per-
sonnel who have been trained by Belimed AG. Belimed AG accepts no responsibility for work per-
formed by personnel neither directly employed by Belimed AG, nor trained by Belimed AG.
No repairs or modifications in or on the product must be carried out by non-authorized personnel. Un-
authorized repairs and modifications exclude any liability of the manufacturer for resulting damages
to the operator and the product. Furthermore, modifications that affect operation may violate local
laws and jeopardize insurance coverage in the event of malfunction.
Repairs and modifications must only be carried out by personnel either directly employed by
Belimed AG or trained by Belimed AG. Improperly executed repairs and modifications can entail con-
siderable danger to both patient and operator; moreover, such actions will void the warranty claim.
Furthermore, no warranty service is provided for defects caused by non-compliance with these in-
structions, improper use or normal wear and tear of the product.
Defective components must only be replaced by original spare parts from Belimed. Only if these are
used is it ensured that they meet the safety requirements. If spare parts are used that neither originate
from Belimed AG, nor have not been approved by Belimed AG, the warranty is rendered void.
DANGER
WARNING
CAUTION
NOTICE
TIP
Tips for use and useful information on the best possible utilization.
• Indicates a list.
a) Indicates a subsection of a list.
b)
2 Safety
2.1 Introduction
With this sterilizer, you have acquired a product that is state-of-the-art in terms of safety.
Nevertheless, hazards may arise from the product if it is operated incorrectly or if it is not used as
intended.
2.1.1 Conformity
By means of the EC Declaration of Conformity and the CE mark, Belimed AG affirms that this product
complies with the basic health and safety requirements.
With this sterilizer, you have acquired a product that is state-of-the-art in terms of safety.
Nevertheless, hazards may arise from the product if it is operated incorrectly or if it is not used as
intended.
• The door seal must not be manipulated when the chamber is in pressurized condition.
• If errors or damages occur, the supervisor must be informed. Errors must be resolved as de-
scribed in the error message list.
• Panel components may only be removed while the system is in switched-off state.
• Before opening screw connections, all containers and pipelines must be depressurized.
• Before switching on, check whether all panel components and protective devices are fitted in
place.
• Portable and mobile high frequency communication equipment, such as voice radio equipment or
cell phones, must not be used at distances closer than 1 m from the product. Impairment of func-
tion could otherwise result.
• The use of accessories not intended for the product is not recommended, as this can lead to an
increase of electromagnetic emissions or to a reduced immunity to electromagnetic interference.
• When working on electronic components, make sure that no electrostatic discharge is caused
when coming into contact with the components (wear ESD-compliant protective equipment and
use ESD-compliant tools).
• Only use original spare parts from Belimed AG. If other spare parts are used, the promised perfor-
mance cannot be ensured and the manufacturer's warranty becomes void.
• Always mop up and dry wet floors immediately.
DANGER
Switch the building power supply to a deenergized state and safeguard it against being switched
on again.
DANGER
In case of uncertainty with regard to correct implementation, contact the Belimed service.
WARNING
Injury hazard
If the door drive starts running during maintenance work, hands or arms can get stuck between the
chain and the cog wheel.
Before maintenance work, turn off the main switch and secure it with a padlock.
Carry the key of the padlock with you during maintenance work.
WARNING
Scalding hazard
The sterilizer, and particularly the steam lines, are under high pressure. During work on the sterilizer,
all pipes must be depressurized.
Check the manometers to verify that the entire piping, the chamber and the chamber jacket are
depressurized.
WARNING
Injury hazard
In the event of incorrect manipulation, the pressure vessel may, if it comes to the worst, burst.
Have safety valves checked only by personnel who have been trained and authorized by Belimed.
CAUTION
Injury hazard
If the slipping clutch is disengaged, the chamber door cannot slide downwards.
When working below the chamber door, block the door with the built-in door locking mechanism.
CAUTION
CAUTION
Risk of non-sterility
According to EN 285, the Bowie-Dick Test must be performed every day prior to the first product ster-
ilization cycle.
Start Bowie-Dick Test every day prior to the first product sterilization cycle.
CAUTION
Risk of damage
Use of incorrect components may lead to damage when using the thermoblock.
CAUTION
Burn hazard
The chamber must have cooled down. Do not perform any work on the hot chamber.
Dangerous voltage!
Caution! Hot!
3 Description
3.1 Overview of the product
The diagrams below show the main elements and operating elements on the loading side and on the
unloading side. The cycle documentation printer is located on the loading or unloading side, depend-
ing on the machine version.
2
3
4
1 Control panel
2 Cycle documentation printer (optionally on unloading side)
3 Chamber door
4 Emergency stop button
3
4
Fig. 3-2 Main components and operating elements on the unloading side
1 Control panel
2 Cycle documentation printer (optionally on unloading side)
3 Chamber door
4 Emergency stop button
1 Main switch
2 Switch cabinet
3 Cycle documentation printer
4 Chamber door
5 Servicing door
6 Door motor
1 Chamber door
2 Servicing door
3 Door motor
1 Main switch
1
2
In sterilizers with external steam supply, the steam shut-off valve 2 is installed on the building side,
upstream of the utility connection. Therefore, the location can vary. The utility connection is located
above the chamber behind the servicing door on the loading side.
The cold water shut-off valve is located upstream of the utility connection. The utility connection 3 is
located behind the servicing door on the loading side below the chamber, or, in the case of utility con-
nection at top 1 (option), above the chamber.
TIP
• Electronic pressure transducers that display the pressure in digital form on the operating panel.
• Mechanical pressure indicators in the servicing space.
The pressure indicators in the servicing space show the following values during operation, depending
on the program phase:
Compressed air +5 to +7 +5 to +7
Table 3-1 Typical readings of the pressure indicators in the servicing space
6
5
4
3.2.4.1 Overview
1
2
3
5
6
7
8
1 Status bar
2 button
3 Display with operating panel
4 Selection buttons
5 Activity bar
6 Confirmation button ‹OK›
7 On/Off button (standby)
8 Activity buttons
Date, time, chamber temperature, chamber pressure as well as certain settings and events are indi-
cated in the status bar 1.
Doors closed
The communication bar 3 comprises a sign and a 2-line text display. The sign indicates whether the
display is in ‹home mode› (basic display) or in ‹information mode›.
Screen in Screen in
‹home mode› ‹information mode›
Numbers New
Relocating programs
Cycle information
Program intervention
OK
3.2.7.1 Overview
1
2
3
In the status bar, certain settings and events are shown by means of signs and data:
Doors locked
Doors open
Doors closed
Light-emitting diode
Continuous green light: program active
‹Program progress›
Light-emitting diode
Continuous green light: doors locked
‹Door interlock›
The activity bar displays the functions of the activity button. What functions are available depends on
the respective program phase.
Blowing out the door seal Adjust brightness and background illumi-
nation
* = first bar
3
Fig. 3-12 Transport cart
4 Utilities
4.1 Utility tables
Utility
DL Compressed air ISO 8573-1:2010 [5:6:4] 5–7 ¼" 5 Nm3/h 0.3 Nm3
ZL Air for ventilating aggregate space via aeration channel on building side or from room
ALP Vacuum pump exhaust air freely discharged into the servicing space
Floor drain
GU 2"
Temperature resistance 100 °C
AD1 Steam release safety valve chamber ≤2.7 1" nipple 240 kg/h
AD2 Steam release safety valve jacket ≤2.7 1" nipple 240 kg/h
Utility
DL Compressed air ISO 8573-1:2010 [5:6:4] 5–7 ¼" 5 Nm3/h 0.3 Nm3
ZL Air for ventilating aggregate space via aeration channel on building side or from room
ALP Vacuum pump exhaust air freely discharged into the servicing space
AD1 Steam release safety valve chamber ≤2.7 1" nipple 240 kg/h
AD2 Steam release safety valve jacket ≤2.7 1" nipple 240 kg/h
Utility
Standard supply lines
DL Compressed air ISO 8573-1:2010 [5:6:4] 5–7 ¼" 5 Nm3/h 0.3 Nm3
ZL Air for ventilating aggregate space via aeration channel on building side or from room
ALP Vacuum pump exhaust air freely discharged into the servicing space
AD1 Steam release safety valve chamber ≤2.7 1" nipple 240 kg/h
AD2 Steam release safety valve jacket ≤2.7 1" nipple 240 kg/h
Utility
ZL Air for ventilating aggregate space via aeration channel on building side or from room
ALP Vacuum pump exhaust air freely discharged into the servicing space
AD1 Steam release safety valve chamber ≤2.7 1" nipple 240 kg/h
AD2 Steam release safety valve jacket ≤2.7 1" nipple 240 kg/h
Recommended
Maximum value
value
pH 6.5–8.0 6.0–8.5
Table 4-6 Contamination in the feed water for a dedicated steam generator
Substance/property Condensate
Steam properties
Required value
as defined in EN 285
Overheating 25 K
Recommended Maximum
Value Value
Table 4-9 Requirements with regard to the cooling water quality for cooling water switch-over (option)
5 Procedure description
5.1 Predefined programs and parameters
The sterilizer is equipped ex-factory with the procedures described in Chapter 5.3 «Pressure-time di-
agrams of the standard programs».
A broken line means that the corresponding procedure stage shown is shorter than in reality.
Pre-treatment
Post-treatment
P [bar]
0 t
-1
1 2 3
0 t
-1
1 2 3
Fig. 5-2 Pressure-time diagram for Instruments, 134 °C, with air detector (option)
TIP
Heavy instruments
If the drying time of this program is not sufficient for certain heavy instruments, then the ‹Heavy in-
struments, 134 °C› program can be used.
P [bar]
0 t
-1
1 2 3
Fig. 5-3 Pressure-time diagram for Textiles, 134 °C and Special, 134 °C
P [bar]
0 t
-1
1 2 3
Fig. 5-4 Pressure-time diagram for Textiles, 134 °C, and Special, 134 °C, with air detector (option)
P [bar]
0 t
-1
1 2 3
0 t
-1
1 2 3
Fig. 5-6 Pressure-time diagram for Rubber 121 °C with air detector (option)
P [bar]
0 t
-1
1 2 3 4
Table 5-8 Characteristics for Bowie-Dick Test and Integrated Bowie-Dick Test (option)
P [bar]
0 t
-1
1 2 3
Fig. 5-8 Pressure-time diagram for Bowie-Dick Test and Integrated Bowie-Dick Test (option)
CAUTION
Risk of non-sterility
According to EN 285, the Bowie-Dick Test must be performed every day prior to the first product ster-
ilization cycle.
Start Bowie-Dick Test every day prior to the first product sterilization cycle.
P [bar]
0 t
-1
4
NOTICE
Risk of non-sterility
The leak test cannot be completed successfully if the chamber is moist.
P [bar]
0 t
-1
1 2 3
0 t
-1
1 2 3
Fig. 5-11 Pressure-time diagram for Heavy Instruments with air detector (option)
P [bar]
0 t
-1
1 2 3
0 t
-1
1 2 3
Fig. 5-13 Pressure-time diagram for Porous items 125 °C with air detector (option)
P [bar]
2
0 t
-1 1 2 3
Fig. 5-14 Pressure-time diagram for Liquids 121°C
P [bar]
0 t
-1
1 2 3
5.3.13 134 °C Bowie Dick function test with Integrated Bowie-Dick Test (option)
Bowie Dick func- 134°C 3.5 min 5 min FRVV-VMT Empty chamber
tion test
P [bar]
0 t
-1
1 2 3
NOTICE
Risk of non-sterility
Poor steam quality can result in non-sterile goods.
Run the Bowie Dick function test once per month with heated chamber.
The message ‹Bowie-Dick Test failed› must appear on the cycle print-out.
TIP
User level
If a user is already logged on, only users with the same level or lower levels than the logged-on user
are shown.
User
User name
User password
0
User level
0
User active
Off
OK
Select the ‹User name›, then ‹User password›, then ‹User level› field via the selection buttons and
fill out each field in turn.
Set the ‹User active› field to ‹On›.
Confirm with ‹OK› button.
TIP
Deleting a user
Select the ‹User› and ‹Password› fields and delete using the button.
Set the ‹User active› field to ‹Off›.
Confirm with ‹OK› button.
Changing a password
Select the ‹User password› field.
Enter new password.
Confirm with ‹OK› button.
6.2 Settings
In addition to the settings described in the User Manual, the ‹Administrator› user can also set the fol-
lowing settings.
NOTICE
Risk of non-sterility
Do not set the time on the clock during an ongoing program.
TIP
Clock settings
Switching to Daylight Saving Time or Standard Time has to be carried out manually. If the sterilizer is
connected to the ICS 8535 documentation system (option), then the time on the sterilizer clock is syn-
chronized automatically every hour by the PC of the ICS 8535 system.
Adjust the time on the clock when the time switches to Daylight Saving Time or Standard Time.
Procedure
Check in the program list, which program numbers are not yet occupied. One of these unoccupied
program numbers must be indicated in step 5.
Select menu item ‹Copy programs› in the ‹Customize programs› menu.
Select menu item ‹Copy from program number›.
Select the program to be copied using the and buttons and confirm with ‹OK› button.
– The name of the selected program is displayed in the ‹Program name› field below.
Select menu item ‹Copy to program number›.
Select the program number using the and buttons and confirm with ‹OK› button.
– If a program number that is already occupied has been selected, the name of the program is
indicated in the menu item ‹Program name›.
If an already occupied program number has been selected, select another program number,
as the program otherwise cannot be copied.
NOTICE
Risk of mix-up
The name of a new program must be unambiguous, and it must be such that it cannot be confused
with that of the original program.
Name programs unambiguously and in such a way that they cannot be confused with other pro-
grams.
Procedure
Check in the program list, from which and to which program number the program is to be relocat-
ed. These program numbers must be indicated in step 4.
Select menu item ‹Relocate programs› in the ‹Customize programs› menu.
Select menu item ‹Relocate from position no.›.
Select the program to be relocated using the and buttons and confirm with ‹OK› button.
– The name of the selected program is displayed in the ‹Program name› field below.
Select menu item ‹Relocate to position no.›.
Select the position number using the and buttons and confirm with ‹OK› button. All po-
sitions between 1 and 20 can be selected.
– The name of the program occupying this position appears in the ‹Program name› field below.
Select button.
– The program is relocated to the specified position.
– The programs located below or above are shifted by one position respectively.
Procedure
Check in the program list, which program number is to be deleted. This program number must be
indicated in step 3.
Select menu item ‹Delete programs› in the ‹Customize programs› menu.
Procedure
Select menu item ‹Adjust parameters› in the ‹Customize programs› menu.
Select menu item ‹Program number›.
Select the program to be modified using the and buttons and confirm with ‹OK› button.
– The name of the selected program is displayed in the ‹Program name› field below.
7 Options
7.1 Automation unit
The automation unit (option) is able to automatically load or unload the sterilizer. The changeover to
automatic mode is described in the User Manual of the automation unit. The automation unit cannot
be operated manually.
7.1.1 Initialization
When the automation unit is switched to automatic mode, the slider moves into basic position. The
basic position has been defined by Belimed service in the course of commissioning and can only be
altered by Belimed service.
7.1.2 Overview
2
3
4
5
6
7
2
3
1 Sensor B72: Detects batch cart in the operating space of the sliding door of
the sterilizer
2 Sensor B74: Detects batch cart located before the sliding door
3 Sensor B71: Detects batch cart at locking pin
4 Slider safety bar: prevents jamming
5 Sensor B73: Detects docking of transport cart or queue line
NOTICE
Only switch on the emergency cooling function if the building cooling circuit is not available.
Deactivate the emergency cooling function as soon as the cooling circuit is available again.
TIP
If the sterilizer is switched off, the emergency cooling function is deactivated at the next start.
8 Errors
An internal diagnosing program continuously monitors all the important parameters and triggers an
error message as soon as a deviation is detected. The sign appears on the screen. Depending
on the type of hazard, the current program responds as follows:
Remedy
1 Emergency stop The Emergency stop button on the sterilizer has been 2
activated.
2 Vacuum pump The motor protection switch of the vacuum pump was 2
motor protection triggered due to overloading.
Remedy
Remedy
Remedy
Remedy
20 Sensor P3 faulty The jacket pressure measuring circuit P3 has been inter- 3
rupted.
Remedy
Remedy
Remedy
32 Jacket pressure Failure to reach the target jacket pressure within 60 min. 3
not reached
Check that there is sufficient steam pressure in the
supply line.
Check whether the steam generator is currently car-
rying out a self-test.
Check that the jacket steam valve V622 opens.
Check that the pilot valve V622.1 opens.
Check that the output of the steam generator is ade-
quate.
Remedy
Remedy
38 LS door - end Door on loading side failed to reach the end position 4
position not within 20 sec.
reached
Remove objects near the door opening.
Check and clean door seals and seal frames, see
Chapter 10.2.3 «Checking and cleaning the door
seals and the seal frames».
Check screws around the door and retighten them.
Replace door seal if torn.
Check the monitoring of the door sealing pressure,
see Chapter 10.2.9 «Checking the door sealing pres-
sure surveillance».
Check whether the friction clutch is loose, retighten if
necessary.
Check the closure force, see Chapter 10.2.4 «Check-
ing the closure force of the doors».
Adjust the friction clutch, see Chapter 10.2.5 «Adjust-
ing the slipping clutch».
39 US door - end Door on unloading side failed to reach the end position 4
position not within 20 sec.
reached
Remove objects near the door opening.
Check and clean door seals and seal frames, see
Chapter 10.2.3 «Checking and cleaning the door
seals and the seal frames».
Check screws around the door and retighten them.
Replace door seal if torn.
Check the monitoring of the door sealing pressure,
see Chapter 10.2.9 «Checking the door sealing pres-
sure surveillance».
Check whether the friction clutch is loose, retighten if
necessary.
Check the closure force, see Chapter 10.2.4 «Check-
ing the closure force of the doors».
Adjust the friction clutch, see Chapter 10.2.5 «Adjust-
ing the slipping clutch».
Remedy
42 LS door not The ‹Door closed› end switch on the loading side is con- 1
closed firmed. Due to excess pressure in the chamber or heat-
ing, the end switch may have shifted.
43 US door not The ‹Door closed› end switch on the unloading side is 1
closed confirmed. Due to excess pressure in the chamber or
heating, the end switch may have shifted.
49 Door not locked The door seal has not been properly pressed on. 1
51–57 – This error does not exist for MST-V with internal steam –
generator.
58 Chamber zero The zero pressure switch has still not switched at 4
pressure surveil- 1300 mbar.
lance
Check the zero pressure switch (B335), replace if
necessary.
Remedy
59 Time base devia- Clock and PLC timer deviate from one another > 1 s/ 3
tion min.
Remedy
64 Batch control Temperature deviation too large: RIF 1–3 too large or 3
failed (T3 – T1) > 0.5°C.
Check PLC.
Check printer interface.
Check wiring between PLC and printer interface.
Remedy
78 No connection The PLC is not receiving any status queries from the 4
with ICS 8535 ICS 8535 documentation system.
system
Check computer to which the ICS 8535 is connected.
Check database.
Check data logger.
Check connection to ICS 8535.
Check switch.
Check Ethernet connecting cables.
Check IP address.
For errors without an error message, see Chapter 8.2 «Other errors».
The loading process is not started. • Connection between automation Check connectors and cables.
unit (option) and sterilizer has
Restart sterilizer.
been interrupted.
Check sensors.
Check slider according to installation
instructions.
Check output cards of the control sys-
tem on the sterilizer, and exchange if
necessary.
The unloading process is not started. • Connection between automation Check connectors and cables.
unit (option) and sterilizer has
Restart sterilizer.
been interrupted.
Test program using the web server.
Check sensors.
Check slider according to installation
instructions.
Check output cards of the control sys-
tem on the sterilizer, and exchange if
necessary.
Error during or at the end of the load- • Various causes. Check connectors and cables.
ing or unloading process.
Restart sterilizer.
Test program using the web server.
Check sensors.
Check slider according to installation
instructions.
Check output cards of the control sys-
tem on the sterilizer, and exchange if
necessary.
Scanner does not work. • Cable connection has been inter- Eliminate cause.
rupted.
Clean scanner.
• Connector has come loose. Check connectors and cables.
• Scanner is dirty. Check cabling in the switch cabinet.
If necessary, replace scanner and
cable.
For errors with an error message, see Chapter 8.1 «Error messages».
9 Maintenance
9.1 Safety instructions
The in-house technician who performs maintenance must be trained by Belimed AG and must be au-
thorized to perform maintenance on the sterilizer. Maintenance work which is intended to be per-
formed by a service technician may only be performed by in-house technicians if they have been
appropriately trained by Belimed AG.
NOTICE
Risk of damage
Unsuitable spare parts can cause damage to the product.
Only use original spare parts from Belimed AG as specified in the spare parts catalog.
WARNING
Scalding hazard
The sterilizer, and particularly the steam lines, are under high pressure. During work on the sterilizer,
all pipes must be depressurized.
Check the manometers to verify that the entire piping, the chamber and the chamber jacket are
depressurized.
CAUTION
WARNING
Injury hazard
If the door drive starts running during maintenance work, hands or arms can get stuck between the
chain and the cog wheel.
Before maintenance work, turn off the main switch and secure it with a padlock.
Carry the key of the padlock with you during maintenance work.
The following preparatory steps are mandatory before starting any maintenance work:
Press button.
Press the ‹Machine information› button.
Browse using the or button until the maintenance levels are displayed.
– It is indicated on the screen, which maintenance level is active.
After completion of maintenance, the Belimed service technician resets the message via the web ap-
plication.
Maintenance Display on the operating panel If the hours of operation are not
level reached, maintenance must be per-
formed at the latest after
NOTICE
DANGER
In case of uncertainty with regard to correct implementation, contact the Belimed service.
Maintenance level 1
Maintenance level 2
Maintenance level 3
see Chap-
Maintenance item to be performed by ter
Maintenance level 1
Maintenance level 2
Maintenance level 3
see Chap-
Maintenance item to be performed by ter
10 Servicing instructions
10.1 Maintenance work on the sterilizer
CAUTION
Burn hazard
The chamber must have cooled down. Do not perform any work on the hot chamber.
NOTICE
Risk of damage
Tools and similar aids can damage the door seal.
TIP
Wear gloves
Door seal grease is difficult to wash off.
Wear gloves.
Grease the new door seal completely with door seal grease (material no. 350 07118).
2
1 3 3 1
3 3
2 2
3 3
1 3 3 1
2
Fig. 10-1 View of door seal
Reinstall door seal.
Press the door seal 1 into the groove at all four corners.
Press the door seal 2 into the groove at the center of each of the four sides.
Press the door seal 3 into the groove at the center of each section.
Press the remainder of the door seal into the groove.
When the seal is fully seated in the groove, press the ‹Retract door seal› button.
– The vacuum pump starts up and generates a vacuum in the seal frame. The door seal is re-
tracted into the seal frame.
If the door seal is not entirely retracted, press it into the groove at the relevant place by hand and
once more activate the ‹Retract door seal› button.
TIP
After installation of the door seal, clean the surfaces of the door seal and of the seal frame.
CAUTION
Burn hazard
The chamber must have cooled down. Do not perform any work on the hot chamber.
NOTICE
Risk of damage
Tools and similar aids can damage the door seal.
TIP
Wear gloves
Door seal grease is difficult to wash off.
Wear gloves.
Grease the new door seal completely with door seal grease.
Reinstall door seal.
Press the door seal 1 into the groove at all four corners.
Press the door seal 2 into the groove at the center of each of the four sides.
Press the door seal 3 into the groove at the center of each section.
Press the remainder of the door seal into the groove.
2
1 3 3 1
3 3
2 2
3 3
1 3 3 1
2
Fig. 10-2 View of door seal
When the door seal is fully seated in the groove, press the button.
– The vacuum pump starts up and generates a vacuum in the seal frame. The door seal is re-
tracted into the seal frame.
If the door seal is not entirely retracted, press it into the groove at the relevant place by hand and
once more activate the button.
TIP
After installation of the door seal, clean the surfaces of the door seal and of the seal frame.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
NOTICE
Contamination hazard
After being unpacked, the filter is exposed to microbiological contamination.
Do not remove the filter from the packaging until shortly before installation.
Do not touch the connection thread when unpacking and installing the filter.
The ventilation filter must be replaced annually. It is located above the sterilizing chamber on the load-
ing side.
TIP
Check the time indicated on the print-out or in the documentation system (option).
R621 2 closure
R623 2 opening
R624 3 opening
If the specified time is exceeded, readjust the throttle non-return valve concerned.
10.1.12.1Connections
The sterilizer is equipped with 2 connections for testing the steam quality as specified in EN 285,
Section 22.
Fig. 10-4 Connections for testing the steam quality according to EN 285, Section 22
TIP
If no withdrawal device for the steam condensate is available, this can be made with the following
materials in accordance with EN 285, Section 22.4:
1
8 2
3
7
4
6
Fig. 10-5 Schematic diagram of steam condensate withdrawal
1 Valve with impact tube, 0.6 mm restrictor bore, clamp or coupler for attach-
ment
2 Polypropylene tube, 5 m long, internal diameter 6 mm
3 Burette stand
4 Container with 8 L capacity, filled with approx. 10 kg of ice (or cold water if
need be), metal part to fix the polypropylene tube in place in the container;
ground the metal part
5 2 polypropylene bottles, each with a capacity of 250 mL
6 Connection to the condensate drain
7 Connection to the sterilizer
8 Sterilizer steam supply connection
WARNING
Scalding hazard
The sterilizer, and particularly the steam lines, are under high pressure. During work on the sterilizer,
all pipes must be depressurized.
Check the manometers to verify that the entire piping, the chamber and the chamber jacket are
depressurized.
Procedure
Connect the impact tube to the connection for the steam withdrawal according to 10.1.12.1 «Con-
nections»
Clean and rinse the polypropylene tube and bottles with distilled water and then dry them.
Attach polypropylene tube to impact tube.
Place polypropylene tube with several windings in the container filled with cold water and ice and
fix it in place with the metal part.
Attach other end of polypropylene tube to the burette stand so that the condensate can flow into
the bottle.
Open steam shut-off valve.
Do not use the first 50 mL of the condensate; thereafter, fill two bottles with 250 mL of condensate
each and close them tightly. Observe the recommendation of the laboratory. The required quantity
can deviate.
Have the condensate analyzed in a suitable laboratory.
5
6
Fig. 10-6 Individual components of the condensate drain of the steam dehydration system
1 Casing lid
2 Lid gasket
3 Spring
4 Capsule element
5 Spacer plate
6 Dowel screw
7 Valve seat gasket
8 Casing
Shut off the inlet and the outlet of the condensate drain and make sure that the sterilizer is in a
depressurized state.
Unscrew the lid 1 of the condensate drain.
Remove the spring 3, the capsule element 4 and the spacer plate 5.
Remove the dowel screw 6 and the valve seat gasket 7.
Install new dowel screw 6 and valve seat gasket 7.
Insert spacer plate 5.
Replace lid gasket 2.
TIP
Grease thread
To keep threads smooth-running, the threads must be greased.
Grease threads with door seal grease (material no. 350 07118).
DANGER
Switch the building power supply to a deenergized state and safeguard it against being switched
on again.
NOTICE
Risk of damage
Electrostatic discharges can damage electronic components.
In the case of the main contactors, make sure that these were not insufficiently tightened. Should
this be the case, then loose contacts could have led to black discolorations.
Replace contactors that show signs of wear.
NOTICE
Risk of damage
Electrostatic discharges can damage electronic components.
NOTICE
Data loss
To avoid loss of data, do not switch off the PLC.
Note: If the sterilizer is switched off on the main switch because the buffer battery is discharged, the
PLC loads the data from the flash memory the next time the sterilizer is switched on. Cycle counter,
time, and date may therefore receive an outdated status. So that the cycle counter always has the
correct value, the sterilizer must be switched off at the operating panel first. This ensures that the cur-
rent data are written to the flash memory. This takes about 30 seconds.
Wiring
Check cable for damage or breakage.
Replace damaged cable.
Check that plug connectors are securely in place.
Replace damaged connectors.
Scanner
Examine scanner for scratches.
Replaced damaged scanner.
Check green illuminated button on the scanner.
Replaced damaged scanner.
Leak test
Select and start the ‹Warm up & leak test› program.
– After servicing, the pressure increase rate must not exceed 0.5 mbar/min.
The following components become de-energized when the Emergency stop button is activated:
• Activation of the door motors via contactor Q364 and relay A385-K2
• Vacuum pump
• all control valves
• solenoid valve for releasing the door seal.
Release the Emergency stop button again and acknowledge the error message on the operating
panel.
Cancel the restart blockage by activating the ‹Continue› button.
In the case of 2-door models, follow the same procedure to check the Emergency stop button on
the unloading side.
WARNING
Injury hazard
In the event of incorrect manipulation, the pressure vessel may, if it comes to the worst, burst.
Have safety valves checked only by personnel who have been trained and authorized by Belimed.
10.2.3 Checking and cleaning the door seals and the seal frames
NOTICE
Risk of damage
Tools and similar aids can damage the door seal.
3 3
2 2
3 3
1 3 3 1
2
Fig. 10-9 View of door seal
Reinstall door seal.
Press the door seal 1 into the groove at all four corners.
Press the door seal 2 into the groove at the center of each of the four sides.
Press the door seal 3 into the groove at the center of each section.
Press the remainder of the door seal into the groove.
When the seal is fully seated in the groove, press the ‹Retract door seal› button.
– The vacuum pump starts up and generates a vacuum in the seal frame. The door seal is re-
tracted into the seal frame.
If the door seal is not entirely retracted, press it into the groove at the relevant place by hand and
once more activate the ‹Retract door seal› button.
NOTICE
Risk of damage
Tools and similar aids can damage the door seal.
2
1 3 3 1
3 3
2 2
3 3
1 3 3 1
2
Fig. 10-10 View of door seal
Press the door seal 1 into the groove at all four corners.
Press the door seal 2 into the groove at the center of each of the four sides.
Press the door seal 3 into the groove at the center of each section.
Press the remainder of the door seal into the groove.
When the door seal is fully seated in the groove, press the button.
– The vacuum pump starts up and generates a vacuum in the seal frame. The door seal is re-
tracted into the seal frame.
If the door seal is not entirely retracted, press it into the groove at the relevant place by hand and
once more activate the button.
With the door open place a weight of 15 kg on the upper rim of the door and try to close the door
automatically.
– The door should not close.
With the door open place a weight of 7.5 kg on the upper rim of the door and try to close the door
automatically.
– The door should close.
If one of these conditions is not fulfilled, the slipping clutch must be readjusted.
CAUTION
Injury hazard
If the slipping clutch is disengaged, the chamber door cannot slide downwards.
When working below the chamber door, block the door with the built-in door locking mechanism.
Release door
Turn door securing device into starting position.
Close servicing door.
Switch on the main switch.
Switch on the sterilizer at the operating panel.
Open chamber door.
Once more check the closure force.
The measurement error of the measuring instrument is determined in the calibration. No technical
intervention is performed. The measurement error of the display on the measuring instrument as
compared to the true value is determined (definition of term: DIN 58950-1, version 2011-01).
The reference gauge must have a measuring accuracy of not more than ±0.1 % and a measuring
range of 0-4000 mbar.
The reference gauge must have a measuring accuracy of not more than ±0.1 % and a measuring
range of 0-4000 mbar.
WARNING
Scalding hazard
The sterilizer, and particularly the steam lines, are under high pressure. During work on the sterilizer,
all pipes must be depressurized.
Check the manometers to verify that the entire piping, the chamber and the chamber jacket are
depressurized.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
If the sensors are within the tolerance range, reinstall the sensors.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
Relieve the pressure on the pressure vessel until it has returned to ambient pressure, then recon-
nect the pressure sensors to the jacket and the steam pipe respectively.
TIP
The reference pressure can only be entered within a range of ±200 mbar. If the deviation is larger,
the pressure sensor must be replaced.
TIP
The reference pressure can only be entered within a range of ±200 mbar. If the deviation is larger,
the pressure sensor must be replaced.
The measurement error of the measuring instrument is determined in the calibration. No technical
intervention is performed. The measurement error of the display on the measuring instrument as
compared to the true value is determined (definition of term: DIN 58950-1, version 2011-01).
The reference sensor must have a measuring accuracy of ±0,1 °C.
WARNING
Scalding hazard
The sterilizer, and particularly the steam lines, are under high pressure. During work on the sterilizer,
all pipes must be depressurized.
Check the manometers to verify that the entire piping, the chamber and the chamber jacket are
depressurized.
NOTICE
Risk of malfunctions
If the temperature sensor of the Integrated Bowie-Dick Test (option) is not reinstalled in the same po-
sition, this can cause malfunctions.
After the check has been performed, the temperature sensor must be once more immersed up to
the mark 2 in the Bowie-Dick Test device.
1 O-ring seals
2 Mark
NOTICE
Risk of malfunctions
If temperature sensor T3 (B813) of the air detector (option) is not reinstalled in the same position, this
can cause malfunctions.
Before removing, apply a mark to the sensor with a pen or adhesive tape.
After performing the check, immerse the temperature sensor in the air detector (option) to exactly
the same depth again.
Wait until the sensors indicate a stable temperature for >5 minutes.
– If the sensors for T1 (B811), T2 (B812) and T3 (B813) deviate >0,2 °C from the reference sen-
sor, then the zero point must be calibrated.
– If the sensors for T4 (B815), T5 (B817) deviate >2 °C from the reference sensor, then the sen-
sor(s) concerned must be replaced. These sensors cannot be adjusted.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
If the sensors are within the tolerance range, reinstall the sensors.
NOTICE
Risk of leaks
If a seal is used more than once, the system may develop leaks.
If the sensors are within the tolerance range, reinstall the sensors.
NOTICE
Risk of malfunctions
If the temperature sensor of the Integrated Bowie-Dick Test (option) is not reinstalled in the same po-
sition, this can cause malfunctions.
After the check has been performed, the temperature sensor must be once more immersed up to
the mark 2 in the Bowie-Dick Test device.
1 O-ring seals
2 Mark
NOTICE
Risk of malfunctions
If temperature sensor T3 (B813) of the air detector (option) is not reinstalled in the same position, this
can cause malfunctions.
Before removing, apply a mark to the sensor with a pen or adhesive tape.
After performing the check, immerse the temperature sensor in the air detector (option) to exactly
the same depth again.
TIP
Reference temperature
The reference temperature can only be entered in the range of ±2 °C with respect to the ‹Uncorrected
value›. If the sensor deviates >2 °C from the reference sensor, then the sensor or the analog compo-
nent assembly must be replaced.
Select the ‹Zero point / Gain› button and change the value to ‹Zero point› using the and
buttons. Confirm with ‹OK› button.
– The value in the ‹Corrected value› field must agree with the reference value.
CAUTION
Risk of damage
Use of incorrect components may lead to damage when using the thermoblock.
Press the button, then switch to the third page of the menu using the button.
Select the ‹System configuration› button.
Select the ‹Adjustment› button.
Select the appropriate button for the respective sensor.
Wait until the sensors indicate a stable temperature for >5 minutes. The temperature currently
measured is displayed in the ‹Uncorrected value› field.
Select the ‹Reference value› field and enter the temperature indicated on the reference sensor
using the and buttons. Confirm with ‹OK› button.
Select the ‹Zero point / Gain› button and change the value to ‹Zero point› using the and
buttons. Confirm with ‹OK› button.
Check the value in the ‹Corrected value› field.
– The value in the ‹Corrected value› field must agree with the reference value.
1 Light-emitting diode
Check on I/O-PCB A385 above the PLC, whether light-emitting diode (LED) D5 1 is lit.
– The light-emitting diode must not be lit.
Disconnect the connector from pressure switch B335.
– The error message ‹Chamber zero pressure surveillance› must appear on the operating panel.
Reconnect the connector to the pressure switch.
Acknowledge the error message and terminate the program.
Start the ‹Warm up & leak test› program and wait until the chamber pressure has reached
1500 mbar absolute pressure.
Check on I/O-PCB A385 above the PLC, whether light-emitting diode D5 is lit.
– The light-emitting diode must not be lit.
Check the position of contactor Q364.
– The contactor must be disengaged (position indicator on ‹1›).
Discontinue the program and observe light-emitting diode D5.
– When light-emitting diode D5 is lit, the chamber pressure must be 100 mbar excess pressure.
In addition, contactor Q364 must be engaged (position indicator on ‹1›)
Check whether pilot valves (M334/V334) and (M623/V623.1), and thus also steam valve V623,
are closed.
– The red position indicator of the pilot valves must not project.
Acknowledge the alarm on the operating panel and restart the ‹Leak test› program.
– The program should not start, and the error message must reappear on the operating panel
after 10 seconds.
Increase the door sealing pressure to 3 bar again on the pressure controller.
As soon as the end switch 3 switches off, stop the door and measure the distance ‹A› between
the door leaf 2 and the plastic stopper 1.
– The distance must be 1–3 mm.
1
A
Fig. 10-15 Distance ‹A› between the door leaf 2 and the plastic stopper 1
Slowly open the door again while at the same time observing the voltmeter.
As soon as the end switch 3 switches on again, stop the door and once more measure the dis-
tance ‹A› between the door leaf 2 and the plastic stopper 1.
– The distance must be 10 mm.
Fig. 10-16 Distance ‹B› between the door leaf 2 and the upper edge of the seal frame 1
1
A
Fig. 10-17 Distance ‹A› between the door leaf 2 and the plastic stopper 1
Slowly open the door again while at the same time observing the voltmeter.
As soon as the end switch 3 switches on again, stop the door and once more measure the dis-
tance ‹A› between the door leaf 2 and the plastic stopper 1.
– The distance must be 10 mm.
As soon as the end switch 3 switches off, stop the door and measure the distance ‹B› between
the door leaf 2 and the upper edge of the seal frame 1.
– The distance must be 22–26 mm.
Fig. 10-18 Distance ‹B› between the door leaf 2 and the upper edge of the seal frame 1
12 Technical data
12.1 General
Series MST
Height 660
Height 563
Weight [kg] 12 15 18
per
Maximum loading capacitya[kg] 30 45 60
support grid
Height 393
Height 29
Width 600
Each extendable shelf [mm]
Depth 600 900
Recommended
Maximum value
value
pH 6.5–8.0 6.0–8.5
Recommended Maximum
Value Value
Table 12-6 Requirements with regard to the cooling water quality for cooling water switch-over (option)
Substance/property Condensate
Steam properties
Required value
as defined in EN 285
Overheating 25 K
Frequency [Hz] 50 / 60
Pre-fuse [A] 10 16
IP protection type IP 20
Table 12-9 Electrical connection specifications for steam sterilizer (3N~200–230 V, 50/60 Hz)
Frequency [Hz] 50 / 60
Pre-fuse [A] 10
IP protection type IP 20
Table 12-10 Electrical connection specifications for steam sterilizer (3N~380–420 V, 50/60 Hz)
12.5.1 Operation
Use Indoors
12.5.2 Storage
Storage Indoors
Atmosphere Non-corrosive
Roll diameter 50 mm
B B
"NO"
C C
DAMPF / SD DAMPF ZUM STERILISATOR /
STEAM DN 20 STEAM TO STERILIZER
© Belimed 2017
YS
D Q1003 D
Appendix 5: P&I diagram with external steam switch-over (option)
Q1001
E E
Project Equipment Format Symbols standard PCE standard
---- ---- A3 ISO 10628 EN 62424
Verifier Originator Document type
PCH SWI R&I Schema / P&I-diagram
Customer,
Machine type: MST-V 09.02
Belimed AG
Zelgstrasse 8 Fremddampfumschaltung FD/ED / External steam switch-over
CH-8583 Sulgen
www.belimed.com Part number Document number
F The reproduction, distribution and utilization of this document as well
--- 1005-3801 F
as the communication of its contents to others without express Revision index Revision number Date of issue Sheet
authorization is prohibited. Offenders will be held liable for the
payment of damages. Copyright © BELIMED 01 - 14.08.2017 1/1
1 2 3 4 5 6 7 8
P&I diagram with external steam switch-over (option)
131/148
Spare parts drawing - sterilizer
Appendix 6: Spare parts drawing - sterilizer
© Belimed 2017
9 11 13
10
1 4
12
22 11 14
20 21
14
Das Urheberrecht an dieser Zeichnung und allen Beilagen, die dem Empfänger persönlich anvertraut sind, verbleibt jederzeit bei
unserer Firma. Ohne unsere schriftliche Genehmigung dürfen sie nicht kopiert oder vervielfältigt, auch niemals Dritten mitgeteilt
oder zugänglich gemacht werden.
133/148
134/148
Detail A
4
5
10
Spare parts drawing - batch cart, 6-0-6/9/12
6
Appendix 8: Spare parts drawing - batch cart, 6-0-6/9/12
1
3
11 A
© Belimed 2017
12
9
7
5
2
11
Das Urheberrecht an dieser Zeichnung und allen Beilagen, die dem Empfänger persönlich anvertraut sind, verbleibt jederzeit bei
unserer Firma. Ohne unsere schriftliche Genehmigung dürfen sie nicht kopiert oder vervielfältigt, auch niemals Dritten mitgeteilt
oder zugänglich gemacht werden.
1
1
10
© Belimed 2017
Appendix 11: Spare parts drawing for extendable shelf (option)
11 11
3
9 10
4
7 5 13 2
12 6
Detail A 8 Detail C
Detail B Detail D
Das Urheberrecht an dieser Zeichnung und allen Beilagen, die dem Empfänger persönlich anvertraut sind, verbleibt jederzeit bei
unserer Firma. Ohne unsere schriftliche Genehmigung dürfen sie nicht kopiert oder vervielfältigt, auch niemals Dritten mitgeteilt
oder zugänglich gemacht werden.
1
000-314-617
137/148
138/148
Detail A
4
Spare parts drawing for foot switch (option)
Appendix 12: Spare parts drawing for foot switch (option)
© Belimed 2017
3
2
1
Das Urheberrecht an dieser Zeichnung und allen Beilagen, die dem Empfänger persönlich anvertraut sind, verbleibt jederzeit bei
unserer Firma. Ohne unsere schriftliche Genehmigung dürfen sie nicht kopiert oder vervielfältigt, auch niemals Dritten mitgeteilt
A oder zugänglich gemacht werden.
Verfahrenstechnik / Process technology: MST-V ohne Kühlkreislauf / without cooling circuit 60_300.18233
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 1 / 7
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 2 / 7
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 3 / 7
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 4 / 7
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 6 / 7
1004-5011 V05 ETLi MST-V 09.02 DGRL 2017-10-09 SAP 1004-5011 Version 05 Seite / page: 7 / 7
Belimed AG | Grienbachstrasse 11 | 6300 Zug | Switzerland | +41 41 449 78 88 | +41 41 449 78 89 | www.belimed.com